K Number
K172644
Date Cleared
2017-11-03

(63 days)

Product Code
Regulation Number
882.5880
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the stimulator to the fascia or interspinous/supra-spinous ligament.

Device Description

The Stimwave Technologies Incorporated (Stimwave) SandShark Injectable Anchor (SIA) System is used to fixate the Stimwave Freedom Stimulator to surrounding tissue. The System is comprised of a carbothane anchor (SandShark Anchor) that is transferred onto the deployment handle (SandShark Injectroducer) with the Loading Rod and Loading Base. The SIA System is provided sterile. The SandShark Injectroducer is used to deploy the SandShark Anchor onto the Stimulator.

AI/ML Overview

The provided text describes the Stimwave SandShark Injectable Anchor (SIA) System and its substantial equivalence to predicate devices, but it does not contain information about specific acceptance criteria or a study proving that the device meets those criteria with numerical performance values relevant to AI/ML devices.

The document is a 510(k) summary for a medical device (an anchor system for spinal cord stimulators), not an AI/ML device. Therefore, the questions related to AI/ML performance metrics, sample sizes for AI/ML models, expert ground truth adjudication, MRMC studies, and standalone AI performance are not applicable to the content of this document.

The document focuses on demonstrating substantial equivalence to predicate devices through:

  • Intended Use and Technological Characteristics Comparison: Showing that the SIA System has the same intended use and similar technological characteristics as existing legally marketed devices.
  • Biocompatibility Data: Confirming that the materials meet biological safety and compatibility requirements (ISO 10993-1:2009).
  • Non-Clinical Performance Data: Verifying that the device meets system design requirements and applicable voluntary standards for physical performance (e.g., protection from temperature change, atmospheric pressure change as per AAMI ANSI ISO 14708-3:2008).

Since the device is not an AI/ML system, the requested information in the prompt, which is tailored for AI/ML device evaluations (e.g., sensitivity, specificity, expert ground truth, training/test sets, etc.), cannot be extracted from this document.

Therefore, I am unable to provide a table of acceptance criteria and reported device performance, or details about studies proving device performance with respect to AI/ML metrics, as this information is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 3, 2017

Stimwave Technologies Incorporated c/o Elizabeth Greene Vice President of Quality Assurance and Regulatory Affairs 1310 Park Central Boulevard South Pompano Beach, Florida 33064

Re: K172644

Trade/Device Name: Sandshark Injectable Anchor System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZB Dated: August 30, 2017 Received: September 1, 2017

Dear Elizabeth Greene:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

William J. Heetderks -S 2017.11.03 08:43:03 -04'00'

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Device Name: SandShark Injectable Anchor (SIA) System

Indications For Use:

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the stimulator to the fascia or interspinous/supra-spinous ligament.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Stimwave Technologies. The word "stim" is in dark blue, "wave" is in green, and "TECHNOLOGIES" is in dark blue with a blue line above it. To the right of the word "wave" are several blue concentric arcs.

510(k) Summary

for

SandShark Injectable Anchor (SIA) System

1. Submission Sponsor

Stimwave Technologies Incorporated 1310 Park Central Boulevard South Pompano Beach FL, 33064 USA Phone: 800.965.5134 Fax: 800.965.5134 Contact: Elizabeth Greene, Vice President of Quality Assurance and Regulatory Affairs

2. Date Prepared

August 30, 2017

3. Device Identification

Trade/Proprietary Name:SandShark Injectable Anchor System
Classification Name:Stimulator, Spinal-Cord, Implanted (Pain Relief)
Classification Regulation:882.5880
Product Code:GZB
Device Class:Class II
Classification Panel:Neurology

4. Legally Marketed Predicate Device(s)

Primary: Boston Scientific (Anulex Technologies Inc.) Fixate Suturing Device (K113805)

Anchor Innovation Medical (A.I.M.) Lead Loop Suturing Device (K150924)

5. Device Description

The Stimwave Technologies Incorporated (Stimwave) SandShark Injectable Anchor (SIA) System is used to fixate the Stimwave Freedom Stimulator to surrounding tissue. The System is comprised of a carbothane anchor (SandShark Anchor) that is transferred onto the deployment handle (SandShark Injectroducer) with the Loading Rod and Loading Base. The SIA System is provided sterile. The SandShark Injectroducer is used to deploy the SandShark Anchor onto the Stimulator.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the two words is the word "TECHNOLOGIES" in blue, inside of a blue box with rounded corners. To the right of the words is a blue graphic that looks like a wave.

SandShark Injectable Anchor (SIA) System

InjectroducerAn acrylonitrile butadiene styrene (ABS) handle and cannula(stainless steel 304V, KR01 Phillips K-Resin) anchor deploymentdevice that is used to secure the SandShark Anchor onto thestimulator to the fascia or interspinous/supra-spinous ligament.
SandSharkAnchorA carbothane (80A) anchor that is deployed by the Injectroducer ontothe stimulator securing the device to the fascia or interspinous/supra-spinous ligament. Four (4) SandShark Anchors are provided in theSIA System, pre-loaded onto the Loading Rod.
Loading RodAn assembly that is used with the Loading Base to transfer theSandShark Anchor onto the cannula of the Injectroducer. The handleis constructed of ABS and the rod is stainless steel 304V.
Loading BaseAn acrylonitrile butadiene styrene (ABS) base that holds the LoadingRod in place while transferring the SandShark Anchor onto thecannula of the Injectroducer.

6. Indication for Use Statement

The SandShark Injectable Anchor (SIA) System is intended to be an accessory to the stimulator component of the Stimwave Freedom Spinal Cord Stimulator (SCS) System to secure the stimulator to the fascia or interspinous/supra-spinous ligament.

7. Substantial Equivalence Discussion

The following table compares the Stimwave SIA System to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

Table SA. Comparison of Characteristics
ComparatorStimwave SandSharkInjectable Anchor (SIA)SystemBoston Scientific FixateSuturing Device(K113805)A.I.M. Lead LoopSuturing Device(K150924)
Product CodeGZBGZBGZB
Regulation No.882.5880882.5880882.5880
Regulation NameStimulator, Spinal-Cord,Implanted (Pain Relief)Stimulator, Spinal-Cord,Implanted (Pain Relief)Stimulator, Spinal-Cord,Implanted (Pain Relief)
Intended UseAccessory for securing aspinal cord stimulator tosurrounding tissueSame as SandSharkSame as SandShark
Implant SiteFascia or inter-spinous/supra-spinous ligamentSame as SandSharkSame as SandShark
Environmental UseHospital or AmbulatorySurgical Center OnlySame as SandSharkSame as SandShark
Intended ClinicianOrthopedic, Neurosurgeon,AnesthesiologistSame as SandSharkSame as SandShark
Anchor MaterialCarbothane 80APEEK or PETPEEK with Polyethyleneor Polyester braid
Cannula MaterialStainless Steel 304VStainless SteelStainless Steel
Table 5A. Comparison of Characteristics
---------------------------------------------

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is underlined. To the right of the words is a blue circular design that looks like radio waves.

ComparatorStimwave SandSharkInjectable Anchor (SIA)SystemBoston Scientific FixateSuturing Device(K113805)A.I.M. Lead LoopSuturing Device(K150924)
Anchor Length1.5 inchNANA
Anchor Outer Diameter0.087 inchN/AN/A
Method of IntroductionPercutaneous and AnchorIncisionPercutaneous and AnchorIncisionPercutaneous and AnchorIncision
Tissue ContactYesSame as SandSharkSame as SandShark
SterilizationEthylene Oxide (EO)Same as SandSharkSame as SandShark
LabelingLabeled as Sterile, SingleUse, Prescription DeviceSame as SandSharkSame as SandShark
SterileYes - ethylene oxideSame as FreedomSame as Freedom
Single-UseYesYesYes
Shelf Life1 yearUnknownUnknown
Complies with ISO 10993-1YesYesYes
Safety Testing PassedYesYesYes

(*) asterisk denotes that formulas were used for the calculations.

8. Biocompatibility Data

The materials, construction and intended use of the SIA System is comparable to the predicate device, and have a long history of safety with respect to biocompatibility. The biological safety of the SandShark Anchor was evaluated in accordance to ISO 10993-1:2009, guidance document Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing for cytotoxicity. sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, implantation (4, 8, and 13 weeks), subacute and subchronic toxicity, chronic toxicity, carcinogenicity, extractables and leechables demonstrated no negative impacts from the materials that are used in the SIA System. The SIA System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.

9. Non-Clinical Performance Data

The SIA System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The SIA System complies with all design requirements and applicable voluntary standards.

AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the SIA System was functional, receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the SIA System was functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a dark blue font, and the word "wave" is in a green font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller, dark blue font. To the right of the words is a graphic of concentric circles in a light blue color.

Stimwave Technologies Incorporated Traditional 510(k) Premarket Submission SandShark Injectable Anchor (SIA) System

Stimwave completed a number of tests for the SIA System that demonstrates substantial equivalence to the legally marketed predicate devices. The SIA System meets all the requirements for overall design, sterilization, and biocompatibility confirms that the output meets the design inputs and specifications. The SIA System passed all testing stated above as shown by the acceptable results obtained.

10. Clinical Performance Data

There was no clinical testing required to support the medical device, as the indications for use are equivalent to the legally marketed predicate devices. These types of devices, including the legally marketed predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to legally marketed predicate devices when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. The SIA System has the same intended use as the legally marketed predicates devices. Performance tested verified that the SIA System complies with all applicable voluntary standards such as AAMI ANSI ISO 14708-3. The SIA System also meets the design requirements where no applicable standard could be used. This included anchor durability testing, as well as biocompatibility and sterilization validation of the SIA System. There were no recognized performance standards for this device. There was no clinical testing performed on this device since performance testing demonstrated similar performance as the legally marketed predicate devices, and materials for the SIA System are the same as the legally marketed predicate devices.

It has been shown in this 510(k) submission that the difference between the SIA System and the legally marketed predicate devices do not raise any questions regarding its safety and effectiveness as compared to legally marketed predicate devices. The SIA System, as designed and manufactured, is determined to be substantially equivalent to the referenced legally marketed predicate devices.

§ 882.5880 Implanted spinal cord stimulator for pain relief.

(a)
Identification. An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.(b)
Classification. Class II (performance standards).