(187 days)
No
The device description and performance studies focus on the physical properties and surgical function of a staple line reinforcement material, with no mention of AI or ML capabilities.
Yes
The device is described as an "absorbable staple line reinforcement material" that provides "support to soft tissue during the healing process" and is used as "a buttress to provide support during the healing process," indicating a therapeutic function in aiding tissue repair and healing.
No
Explanation: The device is a surgical buttress for reinforcing staple lines during soft tissue transection or resection procedures, providing support during the healing process. Its function is interventional/therapeutic, not diagnostic.
No
The device description clearly states that the device is a physical, absorbable staple line reinforcement material composed of fibers and films, intended for implantation during surgery. It is a hardware device, not software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for surgical procedures involving soft tissue transection or resection with staple line reinforcement. This is a surgical device used during a procedure on the patient's body.
- Device Description: The description clearly states it's a "staple line reinforcement, also known as a buttress, for use in the surgical environment for the purpose of reinforcing a staple line." It's an implantable material that provides support during the healing process.
- Lack of IVD Characteristics: An IVD is a medical device used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. The provided text does not mention any testing of samples or diagnostic purposes.
Therefore, the ECHELON ENDOPATH Staple Line Reinforcement is a surgical implant/device, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
Product codes
OXC
Device Description
Echelon Endopath Staple Line Reinforcement is a staple line reinforcement, also known as a buttress, for use in the surgical environment for the purpose of reinforcing a staple line.
The Subject Device is to be used Echelon Endopath GST 60 mm reloads in conjunction with the Echelon family of 60 mm endoscopic linear cutters. The Echelon 60 mm Stapler in conjunction with the Echelon Endopath GST60 mm reload places staggered rows of titanium staples with a reinforcement material, and simultaneously divide the tissue and the reinforcement material between the stapled rows. The Subject Device is an absorbable staple line reinforcement material which is secured to the stapler anvil and reload with a synthetic attachment material. The Subject Device contains an applicator and the implantable device which consists of 3 materials: the Vicryl material, the Polydioxanone film and the attachment material. Echelon Endopath Staple Line Reinforcement is an implanted material which works as an adjunct to surgical staples after transection, to provide support to soft tissue during the healing process.
The Echelon Endopath Staple Line Reinforcement Subject Device and Predicate Device are both synthetic, absorbable staple line reinforcement materials to be applied to staplers to provide support to the staple line during the tissue healing process. The Subject Device is to be applied to the Ethicon Echelon Flex Powered Plus Stapler. Both materials degrade via hydrolysis and have been evaluated for biocompatibility in alignment with ISO 10993-1 and are deemed biocompatible.
Echelon Endopath Staple Line Reinforcement is composed of undyed multifilament fibers identical to those used in Vicryl surgical sutures. The polyglactin 910 implant is knitted by a process which interlocks each fiber juncture to prevent unraveling. The implanted material consists of a Vicryl layer laminated with a PDO film on each side, and a Buttress Attachment Material which is a synthetic material applied to the top outer surface of each implant. The staple line reinforcement is used as a buttress to provide support during the healing process. Both Subject and Predicate materials are hydrolyzed by body fluids and absorbed after serving the intended function. The subject Device absorption is essentially complete at 120 days.
The primary difference between the Subject and Predicate Devices is the feature for loading and delivery of the device to the implantation site. A thin layer of bioabsorbable adhesive is coated onto one surface of the implantable device and is composed of a synthetic polymer/poloxamer blend to attach the Subject Device to the stapler. Whereas the Predicate Device utilized a sleeve construction and suture pullcord to secure the reinforcement material on the stapler.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
lung, bariatric, gastric, small bowel and colorectal procedures
Indicated Patient Age Range
Not Found
Intended User / Care Setting
surgical environment
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance tests were performed ex-vivo and in-vivo, as appropriate for the required endpoint. to demonstrate the ECHELON ENDOPATH Staple Line Reinforcement is substantially equivalent to the Predicate Device.
Ex-vivo evaluations include design verification testing of the ECHELON ENDOPATH Staple Line Reinforcement, including:
- Device Implant Characterization
- Thickness
- Porosity
- Density
- Tensile
- Stiffness
- Tear Resistance
- Burst strength
- Staple pull out strength
- Device Compatibly with Staplers and Reloads
- Force to Close
- Staple Height
- Staple Form Quality
- Staple Line Integrity
- Release of stapled reinforcement material from Endocutter
- Manipulation of device on tissue, buttress security on surgical stapler.
Preclinical in-vivo evaluations include studies with animal models which support the intended use of the ECHELON ENDOPATH Staple Line Reinforcement, demonstrating performance in the indicated gastric, thoracic, small bowel, and colon/colorectal procedures. The surgical functionality of the Subject Device, such as device attachment to the surgical stapler and usability, was demonstrated in animal studies. In-vivo testing evaluations include:
- Hemostasis, bleeding at the staple line
- Tissue healing response, survival studies: Gastric, thoracic tissue absorption, Perpendicular firings, L-shaped firings, Sequential Firings
- Usability of the Subject Device
- Abrasion evaluation on tissue, survival study
Animal tests were performed to verify that the Subject Device meets the definition of substantial equivalence to the Predicate Device.
Stability studies were conducted, and the shelf life of the device was demonstrated. Biocompatibility evaluation was performed and confirmed that the proposed device is biocompatible for the intended patient contact profile. Subject Device is a medical device implant for use on tissue with long-term (>30 days) patient exposure. The patient-contacting materials used in the Subject Device are categorized as Long-term patient contacting materials.
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
Key results: The conclusions of the testing criteria demonstrate that ECHELON ENDOPATH Staple Line Reinforcement is as safe and effective and performs as well as the legally marketed Predicate Device, Gore Seamguard K032865 in terms of intended use, design, materials, biocompatibility, sterility, and performance.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.3300 Surgical mesh.
(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 14, 2019
Ethicon Endo-Surgery LLC Emily Nesbitt Associate Director, Regulatory Affairs 4545 Creek Rd Cincinnati, Ohio 45242
Re: K190937
Trade/Device Name: Echelon Endopath Staple Line Reinforcement Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXC Dated: September 10, 2019 Received: September 13, 2019
Dear Emily Nesbitt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K190937
Device Name ECHELON ENDOPATH Staple Line Reinforcement
Indications for Use (Describe)
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
Type of Use (Select one or both, as applicable) |
---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) |
☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary
Company
Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
Contact
Emily Nesbitt (née Kruetzkamp), Associate Director, Regulatory Affairs Ethicon Endo-Surgery, Inc. Telephone: 513.337.1546 Email: enesbitt@its.jnj.com
Date Prepared: | April 9, 2019 |
---|---|
Trade Name: | Echelon Endopath Staple Line Reinforcement |
Common Name: | Staple Line Reinforcement Material |
Classification Name: | Surgical mesh |
Classification: | 21CFR 878.3300 |
Product Code: | OXC |
Device Class: | Class II |
Panel: | General and Plastic Surgery |
Predicate Device | Gore Seamguard Bioabsorbable Staple Line Reinforcement cleared under K032865 |
Reference Device | Vicryl (Polyglactin 910) Mesh, K810428 |
Device Description
Echelon Endopath Staple Line Reinforcement is a staple line reinforcement, also known as a buttress, for use in the surgical environment for the purpose of reinforcing a staple line.
The Subject Device is to be used Echelon Endopath GST 60 mm reloads in conjunction with the Echelon family of 60 mm endoscopic linear cutters. The Echelon 60 mm Stapler in conjunction with the Echelon Endopath GST60 mm reload places staggered rows of titanium staples with a reinforcement material, and simultaneously divide the tissue and the reinforcement material between the stapled rows. The Subject Device is an absorbable staple line reinforcement material which is secured to the stapler anvil and reload with a synthetic attachment material. The Subject Device contains an applicator and the implantable device which consists of 3 materials: the Vicryl material, the Polydioxanone film and the attachment material. Echelon Endopath Staple Line Reinforcement is an implanted material which works as an adjunct to surgical staples after transection, to provide support to soft tissue during the healing process.
The Echelon Endopath Staple Line Reinforcement Subject Device and Predicate Device have identical intended use: to support staple line reinforcent in soft tissue. The Subject Device has a
4
smaller list of anatomical sites than the Predicate Device, and the Subject Device is a subset of the Predicate Indications.
The Echelon Endopath Staple Line Reinforcement Subject Device and Predicate Device are both synthetic, absorbable staple line reinforcement materials to be applied to staplers to provide support to the staple line during the tissue healing process. The Subject Device is to be applied to the Ethicon Echelon Flex Powered Plus Stapler. Both materials degrade via hydrolysis and have been evaluated for biocompatibility in alignment with ISO 10993-1 and are deemed biocompatible.
Echelon Endopath Staple Line Reinforcement is composed of undyed multifilament fibers identical to those used in Vicryl surgical sutures. The polyglactin 910 implant is knitted by a process which interlocks each fiber juncture to prevent unraveling. The implanted material consists of a Vicryl layer laminated with a PDO film on each side, and a Buttress Attachment Material which is a synthetic material applied to the top outer surface of each implant. The staple line reinforcement is used as a buttress to provide support during the healing process. Both Subject and Predicate materials are hydrolyzed by body fluids and absorbed after serving the intended function. The subject Device absorption is essentially complete at 120 days.
The primary difference between the Subject and Predicate Devices is the feature for loading and delivery of the device to the implantation site. A thin layer of bioabsorbable adhesive is coated onto one surface of the implantable device and is composed of a synthetic polymer/poloxamer blend to attach the Subject Device to the stapler. Whereas the Predicate Device utilized a sleeve construction and suture pullcord to secure the reinforcement material on the stapler.
Indications for Use
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
Summary of Similarities and Differences in Technological Characteristics
Staple line reinforcement is the technological principle for both the Subject Endopath Echelon Staple Line Reinforcement and the Predicate Gore® Seamguard® Reinforcement device. It is based on the use of applying an absorbable material, also known as a buttress, as a reinforcement to the transected surgical staple line in soft tissue. At a high level, the Subject and Predicate Devices are based on the following same technological elements:
- Staple line reinforcement device is applied to a surgical stapler
- Staple line reinforcement and stapler are positioned on the desired anatomical area to be transected
- Stapler is fired, knife proceeds to transect tissue, and staples with buttress are applied to ● the tissue
- Absorbable implant reinforces tissue and helps to secure the staple line ●
5
The Endopath Echelon Staple Line Reinforcement differs from Predicate Gore® Seamguard® Reinforcement device in the Steps for Use to load the buttress device onto the surgical stapler. The Subject Device uses an attachment material to permit temporary attachment to the surgical stapler.
Both devices are sterile, single use, absorbable implants that absorb over time. The Subject Device absorbs over approximately 120 days, the Predicate Device absorbs over approximately 180 days.
Performance Data
Performance tests were performed ex-vivo and in-vivo, as appropriate for the required endpoint. to demonstrate the ECHELON ENDOPATH Staple Line Reinforcement is substantially equivalent to the Predicate Device.
Ex-vivo evaluations include design verification testing of the ECHELON ENDOPATH Staple Line Reinforcement, including:
- Device Implant Characterization ●
- Thickness O
- O Porosity
- Density O
- Tensile O
- Stiffness O
- Tear Resistance O
- Burst strength o
- Staple pull out strength O
- Device Compatibly with Staplers and Reloads ●
- Force to Close o
- Staple Height O
- Staple Form Quality O
- Staple Line Integrity O
- Release of stapled reinforcement material from Endocutter O
- Manipulation of device on tissue, buttress security on surgical stapler .
Preclinical in-vivo evaluations include studies with animal models which support the intended use of the ECHELON ENDOPATH Staple Line Reinforcement, demonstrating performance in the indicated gastric, thoracic, small bowel, and colon/colorectal procedures. The surgical functionality of the Subject Device, such as device attachment to the surgical stapler and usability, was demonstrated in animal studies. In-vivo testing evaluations include:
- Hemostasis, bleeding at the staple line
- . Tissue healing response, survival studies: Gastric, thoracic tissue absorption, Perpendicular firings, L-shaped firings, Sequential Firings
6
- Usability of the Subject Device
- Abrasion evaluation on tissue, survival study ●
Animal tests were performed to verify that the Subject Device meets the definition of substantial equivalence to the Predicate Device.
Stability studies were conducted, and the shelf life of the device was demonstrated. Biocompatibility evaluation was performed and confirmed that the proposed device is biocompatible for the intended patient contact profile. Subject Device is a medical device implant for use on tissue with long-term (>30 days) patient exposure. The patient-contacting materials used in the Subject Device are categorized as Long-term patient contacting materials.
Consensus Standards
All components of the Echelon Endopath Staple Line Reinforcement are comprised of materials which are in accordance with ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
Clinical
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
Conclusion
The conclusions of the testing criteria demonstrate that ECHELON ENDOPATH Staple Line Reinforcement is as safe and effective and performs as well as the legally marketed Predicate Device, Gore Seamguard K032865 in terms of intended use, design, materials, biocompatibility, sterility, and performance.