K Number
K190749
Manufacturer
Date Cleared
2019-07-15

(112 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 087 Balloon Guide Catheter System is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vasculature. The balloon provides temporary vascular occlusion during such procedures.

Device Description

The 087 Balloon Guide Catheter System is a sterile, single-use intravascular catheter. The system consists of:

  • 087 Balloon Guide Catheter
  • 1cc Inflation Syringe
  • 8Fr Peel Away Introducer
  • Hub Extension
  • Three Way Stopcock

The 087 Balloon Guide Catheter is offered in three effective lengths, 90, 95, and 100 cm. The 087 Balloon Guide Catheter is an 8 French variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and dual radiopaque distal marker bands on each side of the balloon. The catheter shaft is stainless steel coil reinforced. The hub central port is positioned coaxial to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port and is used with the accessory Inflation Syringe to facilitate inflating and deflating the balloon. The 087 Neurovascular Balloon Guide Catheter uses a distal hydrophilic coating to reduce friction between the catheter shaft and the vessel wall.

AI/ML Overview

The provided document is a 510(k) Premarket Notification summary for the Q'Apel Medical LLC 087 Balloon Guide Catheter System. It describes the device, its indications for use, and the performance testing conducted to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Key Takeaway: The entire submission focuses on demonstrating the substantial equivalence of a new medical device (a catheter system) to an already legally marketed predicate device, rather than proving the efficacy of a new AI/software-based medical device. Therefore, many of the requested points related to AI/MRMC studies, expert ground truth adjudication, and training set details are not applicable to this document. The "studies" here are primarily benchtop and biocompatibility tests on a physical medical device.


Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally "Pass" indicating that "All samples met the pre-determined acceptance criteria" for each listed test. The reported performance is a statement of compliance with these criteria.

Performance Test Study NameDescriptionReference StandardAcceptance Criteria & Reported Device Performance
Conditioning, Distribution, and Shelf Life Aging VerificationTo demonstrate that the product meets the packaging strength and packaging integrity following accelerated aging to a 6-month shelf life equivalent.ASTM F88, ASTM F2096Pass: All samples met the pre-determined acceptance criteria.
Packaging Visual InspectionTo demonstrate that the product meets the packaging visual inspection requirements given.N/A (Internal protocol likely)Pass: All samples met the pre-determined acceptance criteria.
Visual Surface RequirementsTo demonstrate the product satisfies the visual surface requirements.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Dimensional InspectionTo demonstrate that the product meets the dimensional specifications.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Inflation Volume vs Balloon DiameterTo demonstrate that the product meets the inflation volume vs balloon diameter specifications.In consideration of ISO 10555-4:2013Pass: All samples met the pre-determined acceptance criteria.
Balloon Burst VolumeTo demonstrate that the Balloon is capable of withstanding an injection volume of 1.0 CC.In consideration of ISO 10555-4:2013Pass: All samples met the pre-determined acceptance criteria.
Tip Deflection, FG 00100To demonstrate that the stiffness of the distal end of the product is similar to other marketed devices.N/A (Comparison to predicate likely)Pass: All samples met the pre-determined acceptance criteria.
Torque TestingTo demonstrate that the product is capable of 720 degrees of rotation about the central lumen axis without failure.N/A (Internal protocol likely)Pass: All samples met the pre-determined acceptance criteria.
Peak TensileTo demonstrate the product satisfies the peak tensile requirements.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Flow RateTo demonstrate that the flow rate is comparable to the predicate device.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Small Bore Connector Compliance with StandardTo demonstrate that the product meets the requirements.ISO 80369-7:2016Pass: All samples met the pre-determined acceptance criteria.
Corrosion ResistanceTo demonstrate the product satisfies the corrosion resistance requirements.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
RadiopacityTo determine the radiopaque characteristics of the device.ISO 10555-1:2013, ASTM F640-12Pass: All samples met the pre-determined acceptance criteria.
Hub Extension Liquid Leakage Under PressureTo demonstrate that the product meets the liquid leakage requirements.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Hub Extension Hub Aspiration Air LeakageTo demonstrate that the product meets the liquid leakage requirements.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Particulates, Coating IntegrityThis study was conducted to determine the quantity and size of particles generated during simulated use.AAMI TIR42:10, USP <788>, FDA CTQ: Hydrophilic Coated and Hydrophobic Coated Vascular and Neurological Devices, August 2015Pass: All samples met the pre-determined acceptance criteria.
Liquid Leakage Under PressureTo demonstrate that the product meets the liquid leakage requirements given in ISO 10555-1.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Hub Aspiration Air LeakageTo demonstrate that the product meets the hub aspiration air leakage requirements given in ISO 10555-1.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Balloon Inflation FatigueTo demonstrate that there is no degradation of the Balloon after 20 inflation cycles.In consideration of ISO 10555-4:2013Pass: All samples met the pre-determined acceptance criteria.
Simulated UseEvaluation of device performance under simulated use conditions compared to the predicate device.In consideration of ISO 10555-4:2013Pass: All samples met the pre-determined acceptance criteria.
Flex FatigueTo demonstrate that the product does not lose structural integrity when used in the tortuous path model.ISO 10555-1:2013Pass: All samples met the pre-determined acceptance criteria.
Simulated Use Evaluation of the Q'Apel Balloon Guide CatheterSimulated use under in vitro conditions in a cerebral vascular model.N/A (Internal protocol likely)Pass: All samples met the pre-determined acceptance criteria.

Biocompatibility Testing Summary:

Test NameTest MethodAcceptance Criteria & Reported Device Performance
CytotoxicityISO 10993-5Pass: Noncytotoxic according to the predetermined acceptance criteria.
Intracutaneous IrritationISO 10993-10Pass: Test requirements for intracutaneous reactivity were met.
SensitizationISO 10993-10 (Kligman Maximization Test)Pass: Did not elicit a sensitization response.
Systemic ToxicityISO 10993-11Pass: Test requirements for systemic toxicity were met.
Material Mediated PyrogenicityISO 10993-11, USP 40 <151> Pyrogen TestPass: Nonpyrogenic.
HemolysisASTM F756-17, ISO 10993-4Pass: Non-hemolytic.
In Vitro HemocompatibilityISO 10993-4Pass: Not expected to result in adverse effects in vivo.
Complement ActivationISO 10993-4 (SC5b-9 Complement Activation)Pass: Does not activate the complement system.
Un-activated Partial Thromboplastin TimeISO 10994-4, ASTM F2382-04Pass: Does not have an effect on coagulation of human plasma.
ThrombogenicityISO 10994-4Pass: Demonstrates similar thromboresistance characteristics as the control device.

Study Details (Applicability to a Physical Medical Device, Not AI/Software)

  1. Sample sizes used for the test set and the data provenance:

    • The document states "All samples met the pre-determined acceptance criteria" for each test. However, the specific number of samples tested for each benchtop performance or biocompatibility test is not explicitly provided in the summary.
    • Data provenance: These are laboratory/benchtop tests and in vitro biocompatibility tests, not data derived from patients or clinical sources. Therefore, concepts like "country of origin of the data" or "retrospective/prospective" studies are not applicable. The tests are conducted in a controlled lab environment.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For a physical device like a catheter, "ground truth" is established by adherence to engineering specifications and international standards (e.g., ISO, ASTM) through objective physical testing, not by expert consensus on interpretations of data.
    • Experts, if involved, would be internal engineers, quality control personnel, or third-party testing labs specializing in medical device testing, not typically medical experts like radiologists establishing diagnostic "ground truth."
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are relevant for subjective interpretations (e.g., image reading, clinical diagnoses) where multiple experts might disagree. These are objective engineering and biological tests with defined pass/fail criteria.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This type of study is specifically for evaluating the impact of AI algorithms on human performance in diagnostic tasks (e.g., radiology). This submission is for a physical catheter, not an AI software. The document explicitly states: "No animal or clinical studies were required to demonstrate substantial equivalence."
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This relates to the performance of an AI algorithm alone. The device is a physical catheter system, not a software algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground truth for this device is based on engineering specifications, physical measurements, and adherence to established international standards (e.g., ISO 10555-1, ASTM F88, ISO 10993 series). It's objective, quantitative data from benchtop and in vitro (laboratory) tests.
  7. The sample size for the training set:

    • Not applicable. This concept pertains to machine learning models (AI). There is no "training set" for a physical medical device. The device itself is manufactured to specifications.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, there is no training set for this type of device.

In summary, the provided document details the testing and acceptance criteria for a physical medical device (a catheter system) to demonstrate its substantial equivalence to a predicate device for FDA 510(k) clearance. It does not involve AI or software, and therefore, many of the questions related to AI validation methodologies (e.g., MRMC studies, expert adjudication, training/test sets) are not relevant to this specific submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 15, 2019

Q'Apel Medical LLC % Michele Lucey President Lakeshore Medical Device Consulting LLC 128 Blye Hill Landing Newbury, New Hampshire 03255

Re: K190749

Device Name: 087 Balloon Guide Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, MJN Dated: June 12, 2019 Received: June 14, 2019

Dear Michele Lucey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K190749

Device Name 087 Balloon Guide Catheter System

Indications for Use (Describe)

The 087 Balloon Guide Catheter System is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vasculature. The balloon provides temporary vascular occlusion during such procedures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY As required by 21 CFR 807.92

Submitter Information:

Submitter's Name:Q'Apel Medical LLC
Address:330 Wilshire Boulevard
Santa Monica CA 90401
Telephone:310-395-3950
Fax:310-395-3950
Contact Person:Michele Lucey
Telephone:603-748-1374
Date Prepared:July 15, 2019
Device Trade Name:087 Balloon Guide Catheter System
Classification:Class II
Classification Name:Percutaneous Catheter
Product Code(s):DQY, MJN
Regulation Number(s):870.1250
Predicate Devices:Concentric Medical, FlowGate2™ (K153729)

Indication for Use:

The 087 Balloon Guide Catheter System is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vasculature. The balloon provides temporary vascular occlusion during such procedures.

Device Description

The 087 Balloon Guide Catheter System is a sterile, single-use intravascular catheter. The system consists of:

  • 087 Balloon Guide Catheter ●
  • 1cc Inflation Syringe ●
  • 8Fr Peel Away Introducer ●
  • Hub Extension ●
  • Three Way Stopcock ●

The 087 Balloon Guide Catheter is offered in three effective lengths, 90, 95, and 100 cm. The 087 Balloon Guide Catheter is an 8 French variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and dual radiopaque distal marker bands on each side of the balloon. The catheter shaft is stainless steel coil reinforced. The hub central port is positioned

{4}------------------------------------------------

coaxial to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port and is used with the accessory Inflation Syringe to facilitate inflating and deflating the balloon. The 087 Neurovascular Balloon Guide Catheter uses a distal hydrophilic coating to reduce friction between the catheter shaft and the vessel wall.

Performance Testing Summary:

The 087 Balloon Guide Catheter System has successfully completed the following relevant performance testing demonstrating that the device is suitable for its intended use.

Performance Test Summary
Study NameDescriptionReference StandardResults
Conditioning, Distribution, and Shelf Life Aging VerificationTo demonstrate that the product meets the packaging strength and packaging integrity following accelerated aging to a 6-month shelf life equivalentASTM F88 Seal Strength of Flexible Barrier materialsASTM F2096 Standard Test Method for Detecting Gross Leaks in Medical PackagingPassAll samples met the pre-determined acceptance criteria
Packaging Visual InspectionTo demonstrate that the product meets the packaging visual inspection requirements givenN/APassAll samples met the pre-determined acceptance criteria
Visual Surface RequirementsTo demonstrate the product satisfies the visual surface requirementsISO 10555-1:2013Intravascular catheters - Sterile and single-use catheters - Part 1:General requirementsPassAll samples met the pre-determined acceptance criteria
Dimensional InspectionTo demonstrate that the product meets the dimensional specificationsISO 10555-1:2013Intravascular catheters - Sterile and single-use catheters - Part 1:General requirementsPassAll samples met the pre-determined acceptance criteria
Inflation Volume vs Balloon DiameterTo demonstrate that the product meets the inflation volume vs balloon diameter specificationsIn consideration of ISO 10555-4:2013Intravascular Catheters - Sterile and Single-Use Catheters - Part 4:Balloon Dilatation CathetersPassAll samples met the pre-determined acceptance criteria
Balloon Burst Volume,To demonstrate that the Balloon is capable of withstanding an injection volume of 1.0 CC.In consideration of ISO 10555-4:2013Intravascular Catheters - Sterile and Single-Use Catheters - Part 4:Balloon Dilatation CathetersPassAll samples met the pre-determined acceptance criteria
Tip Deflection, FG 00100To demonstrate that the stiffness of the distal end of the product is similar to other marketed devices.N/APassAll samples met the pre-determined acceptance criteria
Performance Test Summary
Study NameDescriptionReference StandardResults
Torque TestingTo demonstrate that the product iscapable of 720 degrees of rotationabout the central lumen axiswithout failure.N/APassAll samples metthe pre-determinedacceptance criteria
Peak TensileTo demonstrate the productsatisfies the peak tensilerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Flow RateTo demonstrate that the flow rateis comparable to the predicatedevice.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Small Bore ConnectorCompliance withStandardTo demonstrate that the productmeets the requirementsISO 80369-7 2016,Small-bore connectorsfor liquids and gases inhealthcare applications- Part 7, Connectorsfor intravascular orhypodermicapplications.PassAll samples metthe pre-determinedacceptance criteria
Corrosion ResistanceTo demonstrate the productsatisfies the corrosion resistancerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
RadiopacityTo determine the radiopaquecharacteristics of the device.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsASTM F640-12Standard Test Methodsfor DeterminingRadiopacity forMedical UsePassAll samples metthe pre-determinedacceptance criteria
Hub Extension LiquidLeakage Under PressureTo demonstrate that the productmeets the liquid leakagerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Hub Extension HubAspiration Air LeakageTo demonstrate that the productmeets the liquid leakagerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Performance Test Summary
Study NameDescriptionReference StandardResults
Particulates, CoatingIntegrityThis study was conducted todetermine the quantity and size ofparticles generated duringsimulated useAAMI TIR42:10Evaluation ofparticulates associatedwith vascular medicaldevicesUSP <788> ParticulateMatter in InjectionsFDA CTQ: HydrophilicCoated andHydrophobic CoatedVascular andNeurological Devices,August 2015PassAll samples metthe pre-determinedacceptance criteria
Liquid Leakage UnderPressureTo demonstrate that the productmeets the liquid leakagerequirements given in ISO 10555-1.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Hub Aspiration AirLeakageTo demonstrate that the productmeets the hub aspiration airleakage requirements given inISO 10555-1.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Balloon Inflation FatigueTo demonstrate that there is nodegradation of the Balloon after20 inflation cycles.In consideration of ISO10555-4:2013Intravascular Catheters- Sterile and Single-UseCatheters - Part 4:Balloon DilatationCathetersPassAll samples metthe pre-determinedacceptance criteria
Simulated UseEvaluation of device performanceunder simulated use conditionscompared to the predicate deviceIn consideration of ISO10555-4:2013Intravascular Catheters- Sterile and Single-UseCatheters - Part 4:Balloon DilatationCathetersPassAll samples metthe pre-determinedacceptance criteria
Flex FatigueTo demonstrate that the productdoes not loose structural integritywhen used in the tortuous pathmodel.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1:General requirementsPassAll samples metthe pre-determinedacceptance criteria
Simulated Use Evaluationof the Q'Apel BalloonGuide CatheterSimulated use under invitroconditions in a cerebral vascularmodelN/APassAll samples metthe pre-determinedacceptance criteria

{5}------------------------------------------------

{6}------------------------------------------------

No animal or clinical studies were required to demonstrate substantial equivalence.

{7}------------------------------------------------

Biocompatibility Testing Summary

Categorized as Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours), per ISO 10993-1, the following testing was conducted:

Test NameTest MethodResults
CytotoxicityTested in accordance with ISO 10993-5, BiologicalEvaluation of Medical Devices - Part 5: Tests for in vitro toxicity, Neutral Red Uptake MethodPassNoncytotoxic according to thepredetermined acceptance criteria
IntracutaneousIrritationTested in accordance with ISO 10993-10, BiologicalEvaluation of Medical Devices - Part 10: Tests forIrritation and Skin SensitizationPassTest requirements forintracutaneous reactivity were metaccording to the predeterminedacceptance criteria
SensitizationTested in accordance with ISO 10993-10, BiologicalEvaluation of Medical Devices - Part 10 Tests forIrritation and Skin Sensitization, KligmanMaximization TestPassdid not elicit a sensitizationresponse according to thepredetermined acceptance criteria
Systemic ToxicityTested in accordance with ISO 10993-11, BiologicalEvaluation of Medical Devices - Part 11: Tests forSystemic ToxicityPassTest requirements for systemictoxicity were met according to thepredetermined acceptance criteria
Material MediatedPyrogenicityTested in accordance with ISO 10993-11, BiologicalEvaluation of Medical Devices - Part 11: Tests forSystemic Toxicity and USP 40 <151> Pyrogen TestPassNonpyrogenic, met thepredetermined acceptance criteria
HemolysisTested in accordance with ASTM F756-17,Standard Practice for Assessment of HemolyticProperties of Materials and ISO 10993-4, BiologicalEvaluation of Medical Devices - Part 4: Selection ofTests for Interactions with Blood, Tests forHemolytic Properties, Direct and Indirect MethodsPassNon-hemolytic, met thepredetermined acceptance criteria
In VitroHemocompatibilityTested in accordance with ISO 10993-4, BiologicalEvaluation of Medical Devices - Part 4: Selection ofTests for Interactions with Blood,Hemocompatibility, Direct Contact MethodPassNot expected to result in adverseeffects in vivo , met thepredetermined acceptance criteria
ComplementActivationTested in accordance with ISO 10993-4, BiologicalEvaluation of Medical Devices - Part 4: Selection ofTests for Interactions with Blood, SC5b-9Complement ActivationPassDoes not activate the complementsystem, met the predeterminedacceptance criteria
Un-activated PartialThromboplastin TimeTested in accordance with ISO 10994-4, BiologicalEvaluation of Medical Devices - Part 4: Selection ofTests for Interactions with Blood and ASTM F2382-04, Standard Test Method for Assessment ofIntravascular Medical Device Materials on PartialThromboplastin Time (UPTT)PassDoes not have an effect oncoagulation of human plasma, metthe predetermined acceptancecriteria
ThrombogenicityTested in accordance with ISO 10994-4, BiologicalEvaluation of Medical Devices - Part 4: Selection ofTests for Interactions with BloodPassDemonstrates similarthromboresistance characteristicsas the control device, met thepredetermined acceptance criteria

{8}------------------------------------------------

Predicate Device Comparison

The following table provides a comparison of the key characteristics of the 087 Balloon Guide Catheter System to the predicate device.

Comparison Chart
FeatureSubject Device087 Balloon GuideCatheter SystemPredicate devicethe 8F FlowGate2TMBalloon Guide CatheterComparison
Regulatory Clearance/Approval ReferenceK190749K153729NA
FDA ClassificationClass IIClass IISame
Product Code(s)DQYDQYSame
Regulation Number870.1250870.1250Same
Regulation NamePercutaneous CatheterPercutaneous CatheterSame
Anatomical LocationsPeripheral and neurovasculaturePeripheral and neurovasculatureSame
MaterialCommonly used medicalgrade plastics andstainless steel,Commonly usedmedical grade plasticsand stainless steelSimilar, differencedoes not raise newquestions regardingsafety and efficacy,confirmed throughbiocompatibility andperformance testing
Reinforced CatheterShaft ReinforcementStainless steel coilreinforcedStainless steel braidreinforcedSimilar, minordifference does notraise new questionsregarding safety andefficacy, bothdevices havereinforced catheterwalls
Injection PortYesYesSame
RadiopacityYes, shaft is visible dueto coil reinforcement,distal tip Pt-Ir markerbandsYes, shaft materialcontains barium sulfate,distal tip Pt-Ir markerbandSimilar, minordifference does notraise new questionsregarding safety andefficacy, bothdevices areradiopaque
Marker Band Locations0.06-in (1.5mm) from thedistal tip edge (distal ofthe balloon)0.66-in (16.8mm) fromthe distal tip edge(proximal of the balloon).08-in (2mm) from thedistal tip edgeSimilar, minordifference does notraise new questionsregarding safety andefficacy, bothdevices utilizeradiopaque markerbands
Compliant BalloonYes, low durometerurethaneYes, siliconeSimilar, materialdifferences do not
Comparison Chart
FeatureSubject Device087 Balloon GuideCatheter SystemPredicate devicethe 8F FlowGate2TMBalloon Guide CatheterComparison
raise new questionsof safety andefficacy, bothmaterials are usedfor compliantballoons forintravascular use
Labeled Shaft OuterDiameter0.110-in (max)(2.79mm)8Fr0.106 in(2.7mm)8FrSimilar, minordifferences do notraise new questionsof safety andefficacy
Labelled Shaft InnerDimension.087-in(2.21mm)6.6Fr0.084-in(2.11mm)6.4FrSimilar, minordifferences do notraise new questionsof safety andefficacy
Effective length90, 95, 100 cm(35.4, 37.4, 39.4 in)90cm, 100cm(35.4, 39.4 in)Same
Tip ShapeStraightStraightSame
Maximum BalloonVolume0.6ml0.6mlSame
Injection PortYesYesSame
RadiopaqueDistal Tip has radiopaquemarker bands, stainlesssteel reinforcement in thecatheter shaft renders theshaft visible onfluoroscopyDistal Tip has aradiopaque markerband, stainless steelreinforcement in thecatheter shaft rendersthe shaft visible onfluoroscopySame
CoatingHydrophilic Coating –Proximal to the balloon,on the distal PortionNoneDifference does notraise new questionsof safety andefficacy.Hydrophiliccoatings arecommonly used tolubricate vascularcatheters. Bothdevices do not havecoating on theballoon region
Internal ConstructionMulti Lumen catheterwallCoaxial lumenDifferences do notraise new questionsof safety andefficacy
Comparison Chart
FeatureSubject Device087 Balloon GuideCatheter SystemPredicate devicethe 8F FlowGate2TMBalloon Guide CatheterComparison
Reinforced ShaftStainless steel reinforcedshaftStainless steelreinforced shaftSame
Accessories Supplied3-Way Stopcock, PeelAway Sheath, Hubextension, 1ml SyringeDilator, RotatingHemostasis Valve,Tuohy Borst Valvewith sideport, PeelAway Sheaths, Luer-Activated ValvesSimilar, minordifferences do notraise new questionsof safety andefficacy.
How SuppliedSterile, single useSterile, single useSame
Sterilization MethodEtOEtOSame
Sterility AssuranceLevel10-610-6Same
efficacy. Bothdesigns are intendedto create a pathwayto the side arm forballoon inflation

{9}------------------------------------------------

{10}------------------------------------------------

Conclusion:

Q'Apel Medical LLC concludes through a review of the benchtop assessments, the comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility that the 087 Balloon Guide Catheter System is substantially equivalent to the predicate device, the FlowGate2™ Balloon Guide Catheter. Any differences are minor and do not raise different questions of safety and effectiveness.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).