(277 days)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs is a patient examination glove made from nitrile compound. non-sterile (as per 21 CFR 880.6250. Class I). The patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
This document is a 510(k) Premarket Notification from the FDA, evaluating a medical device for substantial equivalence to a legally marketed predicate device. The information provided heavily focuses on bench testing and comparison to standards, rather than clinical studies with human participants or AI-driven performance metrics typically found in software as a medical device (SaMD) clearances.
Therefore, many of the requested details, such as the use of a test set with ground truth established by experts, MRMC studies, or training set details for an AI model, are not applicable to this document. The "device" in this context is a physical product (nitrile examination gloves) and not an AI/ML software.
The acceptance criteria and performance are related to the physical and chemical properties of the gloves, not AI performance.
Here's an attempt to answer the questions based only on the provided text, noting where information is not applicable (N/A) or not present.
Acceptance Criteria and Device Performance for Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are established by adherence to various ASTM standards and 21 CFR regulations for physical and chemical properties, and specifically ASTM D6978-05 for chemotherapy drug permeation. The reported performance is the device's measured characteristics against these standards.
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance | Comparison |
|---|---|---|---|
| Dimension | ASTM D6319-10 (Reapproved 2015) | ||
| Length | ≥230mm | Complies with ASTM D6319-10 (230mm min.) | Met |
| Width (X Small) | 70 ± 10 mm | Complies with ASTM D6319-10 (70±10 mm) | Met |
| Width (Small) | 80 ± 10 mm | Complies with ASTM D6319-10 (80±10 mm) | Met |
| Width (Medium) | 95 ± 10 mm | Complies with ASTM D6319-10 (95±10 mm) | Met |
| Width (Large) | 110 ± 10 mm | Complies with ASTM D6319-10 (110±10 mm) | Met |
| Width (X Large) | 120 ± 10 mm | Complies with ASTM D6319-10 (120±10 mm) | Met |
| Thickness - Finger tip | ≥0.05mm | Complies with ASTM D6319-10 (0.05mm min.) | Met |
| Thickness - Palm | ≥0.05mm | Complies with ASTM D6319-10 (0.05mm min.) | Met |
| Physical Properties | ASTM D6319-10 (Reapproved 2015) | ||
| Tensile strength (Before aging) | ≥14MPa | ≥14MPa | Met |
| Tensile strength (After aging) | ≥14MPa | ≥14MPa | Met |
| Elongated rate (Before aging) | ≥500% | ≥500% | Met |
| Elongated rate (After aging) | ≥400% | ≥400% | Met |
| Freedom from pinholes | 21 CFR 800.20, ASTM D6319-10 (Reapproved 2015), Test method: ASTM D5151-06 (Reapproved 2015) | Passed Standard Acceptance Criteria (AQL 2.5) | Met |
| Powder Residual | ASTM D6319-10 (Reapproved 2015), Test method: D6124-06 (Reaffirmation 2011) | < 2 mg/glove | Met |
| Biocompatibility | ISO 10993-10: Third Edition 2010-08-01, ISO 10993-5: Third Edition 2009-06 | ||
| Primary Skin Irritation | Passes | "Not a primary skin irritant" | Met |
| Dermal Sensitization | Passes | "Not a primary skin sensitizer" | Met |
| In vitro cytotoxicity | Passes | "Not cytotoxic" | Met |
| Chemotherapy Permeation | ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs (Max testing time 240 minutes) | Listed below for selected drugs, with others being > 240 minutes. | Met (except for stated exceptions) |
| Carmustine (BCNU) (3.3 mg/ml) | N/A (Comparative, less than 240 mins expected) | 59.4 minutes | N/A |
| Thiotepa (10.0 mg/ml) | N/A (Comparative, less than 240 mins expected) | 118.5 minutes | N/A |
| Most other 34 chemotherapy drugs | > 240 minutes | > 240 minutes | Met |
Note: For Carmustine and Thiotepa, the "acceptance criteria" is not explicitly defined as a threshold, but rather it is noted that these drugs have "extremely low permeation time," and the observed times are reported. The comparison is against the predicate device's performance for these specific drugs, which the proposed device improved upon (59.4 vs 12.4 for Carmustine, 118.5 vs 24.4 for Thiotepa).
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document does not specify the exact sample sizes for each specific test (e.g., how many gloves were tested for pinholes, or how many samples for each chemotherapy drug). It references adherence to ASTM standards, which would dictate appropriate sample sizes for these types of bench tests.
- Data Provenance: The tests were conducted to evaluate the qualities of the "Disposable Powder Free Nitrile Examination Glove, Blue Color" manufactured by Ever Growth (Vietnam) Co. Ltd. (Vietnam). These are bench tests on manufactured products, not data from human subjects or collected retrospectively/prospectively in a clinical setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- N/A. The "ground truth" for the performance of these gloves is established by objective, standardized bench testing methods (e.g., ASTM standards for material properties, chemical permeation tests). There is no "expert consensus" or human adjudication involved in determining these physical and chemical properties.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- N/A. No adjudication method is applicable as these are objective bench test measurements, not subjective human interpretations that require reconciliation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- N/A. This device is a physical medical glove, not an AI-powered diagnostic or assistive technology. No MRMC study was conducted or is relevant.
6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done
- N/A. This device is a physical medical glove. "Standalone performance" in the context of an algorithm is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the device's performance is determined by objective measurements obtained through standardized bench testing methods (e.g., ASTM International standards for physical properties, chemical permeation analysis). For example, a "pinhole" is objectively detected using a water leak test, and "tensile strength" is measured mechanically.
8. The sample size for the training set
- N/A. This is a physical device, not an AI/ML model for which a "training set" would be used.
9. How the ground truth for the training set was established
- N/A. Not applicable, as there is no training set for a physical product.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 23, 2019
Ever Growth (Vietnam) Co. Ltd. % Elizabeth Deng U.S. Representative Elizabeth Deng 5748 Eaglewood Place Ranch Cucamonga, California 91730
Re: K190736
Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: April 2, 2019 Received: September 30, 2019
Dear Elizabeth Deng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190736
Device Name
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs
Indications for Use (Describe)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.
| Test Chemotherapy Drug | Concentration(mg/ml) | Minimum Breakthrough Detection Time ( Min ) |
|---|---|---|
| 01 Bleomycin Sulfate | 15.0 | > 240 |
| 02 Busulfan | 6.0 | > 240 |
| 03 Carboplatin | 10.0 | > 240 |
| 04 Carmustine (BCNU) | 3.3 | 59.4 |
| 05 Chloroquine | 50.0 | > 240 |
| 06 Cisplatin | 1.0 | > 240 |
| 07 Cyclophosphamide | 20.0 | > 240 |
| 08 Cyclosporin A | 100.0 | > 240 |
| 09 Cytarabine | 100.0 | > 240 |
| 10 Dacarbazine (DTIC) | 10.0 | > 240 |
| 11 Daunorubicin | 5.0 | > 240 |
| 12 Docetaxel | 10.0 | > 240 |
| 13 Doxorubicin Hydrochloride | 2.0 | > 240 |
| 14 Epirubicin | 2.0 | > 240 |
| 15 Etoposide (Toposar) | 20.0 | > 240 |
| 16 Fludarabine | 25.0 | > 240 |
| 17 Fluorouracil | 50.0 | > 240 |
| 18 Gemcitabine | 38.0 | > 240 |
| 19 Idaribicin | 1.0 | > 240 |
| 20 Ifosfamide | 50.0 | > 240 |
| 21 Irinotecan | 20.0 | > 240 |
| 22 Mechlorethamine HCl | 1.0 | > 240 |
| 23 Melphalan | 5.0 | > 240 |
| 24 Methotrexate | 25.0 | > 240 |
| 25 Mitomycin C | 0.5 | > 240 |
| 26 Mitoxantrone | 2.0 | > 240 |
| 27 Oxaliplatin | 2.0 | > 240 |
| 28 Paclitaxel (Taxol) | 6.0 | > 240 |
| 29 Paraplatin | 10.0 | > 240 |
| 30 Retrovir | 10.0 | > 240 |
| 31 Rituximab | 10.0 | > 240 |
| 32 Thiotepa | 10.0 | 118.5 |
| 33 Topotecan HCl | 1.0 | > 240 |
| 34 Trisonex | 1.0 | > 240 |
| 35 Velcade (Bortezomib) | 1.0 | > 240 |
| 36 Vincristine | 1.0 | > 240 |
{3}------------------------------------------------
The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
| Carmustine (BCNU) | 3.3 mg/ml | 59.4 minutes |
|---|---|---|
| Thiotepa | 10.0 mg/ml | 118.5 minutes |
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
区 Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
K190736
1.0 Submitter:
| Submitter's name : | Ever Growth (Vietnam) Co., Ltd. |
|---|---|
| Submitter's address : | Long Khanh Industrial Zone, Binh Loc Ward, Long Khanh Township, Dong Nai, Vietnam |
| Phone number : | 84-61-3514025 |
| Fax number : | 84-61-3514023 |
| Name of contact person: | Ming Lee |
| Summary Preparation Date: | Nov. 26th, 2019 |
2.0 US Agent:
| US representative name | Elizabeth Deng |
|---|---|
| Company address | 5748 Eaglewood Place, Rancho Cuamonga, California, Rancho Cucamonga, CA 91739 |
| Telephone number | 909 4659188 |
| Contact email | Baxianunited48@Yahoo.Com |
3.0 Name of the Device
| Proprietary/Trade name: | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs |
|---|---|
| Common Name: | Nitrile Examination Gloves |
| 510(k) Number | K190736 |
| Classification Name: | Patient Examination Glove |
| Device Classification: | Class I |
| Regulation Number: | 21 CFR 880.6250 |
| Product Code: | LZA, LZC |
4.0 Predicate device
| Device Name: | Central Medicare Blue Non Sterile Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs |
|---|---|
| Company name: | Central Medicare Sdn. Bhd |
| 510(K) Number: | K173942 |
5.0 Device Description:
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs is a patient examination glove made from nitrile compound. non-sterile (as per 21 CFR 880.6250. Class I). The
patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
6.0 Indications for Use:
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs:
| Test Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time (Min.) | |
|---|---|---|
| 1 | Bleomycin Sulfate 15.0 mg/ml | > 240 |
| 2 | Busulfan 6.0 mg/ml | > 240 |
| 3 | Carboplatin 10.0 mg/ml | > 240 |
| Table 5.1 Nitrile Blue color test for 36 chemotherapy drugs | |||
|---|---|---|---|
| ------------------------------------------------------------- | -- | -- | -- |
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs
{5}------------------------------------------------
| 4 | Carmustine (BCNU), 3.3 mg/ml | 59.4 |
|---|---|---|
| 5 | Chloroquine 50.0 mg/ml | > 240 |
| 6 | Cisplatin 1 mg/ml | > 240 |
| 7 | Cyclophosphamide 20 mg/ml | > 240 |
| 8 | Cyclosporin A 100.0 mg/ml | > 240 |
| 9 | Cytarabine 100.0 mg/ml | > 240 |
| 10 | Dacarbazine (DTIC), 10.0 mg/ml | > 240 |
| 11 | Daunorubicin 5.0 mg/ml | > 240 |
| 12 | Docetaxel 10.0 mg/ml | > 240 |
| 13 | Doxorubicin Hydrochloride, 2.0 mg/ml | > 240 |
| 14 | Epirubicin 2.0 mg/ml | > 240 |
| 15 | Etoposide (Toposar), 20.0 mg/ml | > 240 |
| 1 | Fludarabine 25.0 mg/ml | > 240 |
| 17 | Fluorouracil, 50.0 mg/ml | > 240 |
| 18 | Gemcitabine 38.0 mg/ml | > 240 |
| 19 | Idaribicin 1.0 mg/ml | > 240 |
| 20 | Ifosfamide 50.0 mg/ml | > 240 |
| 21 | Irinotecan 20.0 mg/ml | > 240 |
| 22 | Mechlorethamine HCl 1.0 mg/ml | > 240 |
| 23 | Melphalan 5 mg/ml | > 240 |
| 24 | Methotrexate 25 mg/ml | > 240 |
| 25 | Mitomycin C 0.5 mg/ml | > 240 |
| 26 | Mitoxantrone 2.0 mg/ml | > 240 |
| 27 | Oxaliplatin 2.0 mg/ml | > 240 |
| 28 | Paclitaxel (Taxol), 6.0 mg/ml | > 240 |
| 29 | Paraplatin 10 mg/ml | > 240 |
| 30 | Retrovir 10 mg/ml | > 240 |
| 31 | Rituximab 10 mg/ml | > 240 |
| 32 | Thiotepa, 10.0 mg/ml | 118.5 |
| 33 | Topotecan HCl 1 mg/ml | > 240 |
| 34 | Trisonex 1 mg/ml | > 240 |
| 35 | Velcade (Bortezomib) 1 mg/ml | > 240 |
| 36 | Vincristine 1.0 mg/ml | > 240 |
The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
| Carmustine (BCNU), 3.3 mg/ml | 59.4 minutes |
|---|---|
| Thiotepa, 10.0 mg/ml | 118.5 minutes |
7.0 Summary of the Technological Characteristics of the Device:
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs are
{6}------------------------------------------------
summarized with the following technological characteristics compared to ASTM or equivalent standard. Table 5.3 Summary of the Technological Characteristics
| Characteristics | Standard |
|---|---|
| Dimension | ASTM standard D 6319-10(Reapproved 2015) |
| Length ≥230mm | |
| Width X Small 70 ± 10 mm | |
| Small 80 ± 10 mm | |
| Medium 95 ± 10 mm | |
| Large 110 ± 10 mm | |
| X large 120 ± 10 mm | |
| Thickness | |
| Finger tip ≥0.05mmPalm ≥0.05mm | |
| Physical Properties | ASTM standard D 6319-10 (Reapproved 2015) |
| Tensile strength (Before aging) ≥14MPa | |
| Tensile strength (After aging) ≥14MPa | |
| Elongated rate (Before aging) ≥500% | |
| Elongated rate (After aging) ≥400% | |
| Freedom frompinholes | 21 CFR 800.20ASTM standard D 6319-10 (Reapproved 2015)Test method in accordance with ASTMD5151-06(Reapproved 2015)Passed Standard Acceptance Criteria |
| Powder Residual | ASTM standard D6319-10(Reapproved 2015)Test method in accordance withD6124-06(Reaffirmation 2011)< 2 mg/glove |
| Biocompatibility | Primary Skin Irritation in rabbitsISO 10993-10: Third Edition2010-08-01PassesUnder the conditions of the study, thesubject device is not a primary skin irritant. |
| Dermal sensitization in the guinea pigISO 10993-10: Third Edition2010-08-01PassesUnder the conditions of the study, thesubject device is not a primary skinsensitizer. | |
| In vitro cytotoxicity accordance withISO 10993-5: Third Edition 2009-06PassesUnder the conditions of the study, thesubject device is not cytotoxic. |
8.0 Based on Assessment of Non-Clinical Performance Data:
The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. Central Medicare Blue Non Sterile Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs made by Central Medicare Sdn. Bhd. met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing:
- A Dimension per ASTM D6319-10 (Reapproved 2015)
-
Tensile strength (Before aging/After aging) and Elongation (Before aging/After aging) per ASTM D6319-10(Reapproved 2015)
-
Water leak test on pinhole per ASTM D6319-10(Reapproved 2015) and per 21 CFR 800.20.
- A Powder Residual tests per ASTM D6319-10(Reapproved 2015)
- A Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01.
-
Assessment Of Resistance To Permeation By Chemotherapy Drugs per ASTM D6978-05(R 2013)
9.0 Based on Assessment of Clinical Performance Data:
Clinical data was not needed to demonstrate that the subject glove is substantially equivalent to the predicate glove. So determination of substantial equivalence is not based on an assessment of clinical performance data.
{7}------------------------------------------------
10.0 Substantial Equivalence Comparison:
Table 5.4 SE compare
| DeviceCharacteristic | Predicate Device | Proposed Device | Comparison | |
|---|---|---|---|---|
| Product name | Central Medicare Blue Non SterilePowder Free Nitrile Examination GlovesTested For Use With ChemotherapyDrugs | Disposable Powder Free NitrileExamination Glove, Blue Color, Tested ForUse With Chemotherapy Drugs | N/A | |
| 510(K) No. | K173942 | K190736 | N/A | |
| ProductOwner | Central Medicare Sdn. Bhd. | Ever Growth Enterprise Corporation | Different | |
| Product Code | LZA, LZC | LZA, LZC | same | |
| Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | same | |
| Class | I | I | same | |
| Intended Use | Blue Non Sterile Powder-Free NitrileExamination glove Tested For Use WithChemotherapy Drugs is a disposabledevice intended for medical purposesthat is worn on the examiner's hand orfinger to prevent contaminationbetween patient and examiner. Gloveshave been tested for use withchemotherapy drugs using ASTMD6978-05 and will be labeled with astatement of compliance and a summaryof the testing result. | A patient examination gloves is adisposable device intended for medicalpurpose that is worn on the examiner'shand or fingers to prevent contaminationbetween patient and examiner. Inaddition, these gloves were tested for usewith chemotherapy drugs in accordancewith ASTM D6978-05 Standard Practicefor Assessment of Medical gloves toPermeation by Chemotherapy Drugs. | similar | |
| Power free | Yes | Yes | same | |
| Size | X Small/Small/ Medium/Large/X Large | X Small/ Small/ Medium/Large/X Large | similar | |
| Single Use | YES | YES | same | |
| Non-Sterile | YES | YES | same | |
| Dimensions-Length | Complies with ASTM D6319-10230 mm min. | Complies with ASTM D6319-10230 mm min. | same | |
| Dimensions-Palm Width | Complies with ASTM D6319-10X Small 70±10Small 80 ±10Medium 95±10Large 110 ±10X large 120 ±10 | Complies with ASTM D6319-10X Small 70±10Small 80 ±10Medium 95±10Large 110 ±10X large 120 ±10 | same | |
| Dimensions-Thickness | Complies with ASTM D6319-10Palm - 0.08 mm min.Finger - 0.10 mm min. | Complies with ASTM D6319-10Palm - 0.05mm min.Finger - 0.05 mm min. | similar | |
| PhysicalProperties | Tensile Strength:Before Aging 15 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min. | Tensile Strength:Before Aging 14 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min. | same | |
| Residualpowder | Complies with ASTM D6319-10 | Complies with ASTM D6319-10 | same | |
| Freedomfrom Holes | In accordance with ASTM D6319-10 andASTM D5151-06(reapproved 2011), G-1,AQL 2.5 | In accordance with ASTM D6319-10 andASTM D5151-06(reapproved 2011), G-1,AQL 2.5 | same | |
| Bio-compatibility | AAMI/ANSI/ISO 10993-10PassesNot a skin irritant & Not a skin sensitizer | AAMI/ANSI/ISO 10993-10Skin sensitization test & Skin Irritationtest:PassesAAMI/ ANSI/ ISO 10993-5In vitro cytotoxicity test:Passes | Similar | |
| Chemotherapy drugs tested | Breakthrough Detection Time in Minutes | comparison | ||
| Device Characteristic | Predicate device | Proposed Device | ||
| Blue | Blue | |||
| Bleomycin Sulfate 15.0 mg/ml | > 240 | > 240 | same | |
| Busulfan 6.0 mg/ml | > 240 | > 240 | same | |
| Carboplatin 10.0 mg/ml | > 240 | > 240 | same | |
| Carmustine (BCNU), 3.3 mg/ml | 12.4 | 59.4 | similar | |
| Chloroquine 50.0 mg/ml | > 240 | > 240 | same | |
| Cisplatin 1 mg/ml | > 240 | > 240 | same | |
| Cyclophosphamide 20 mg/ml | > 240 | > 240 | same | |
| Cyclosporin A 100.0 mg/ml | > 240 | > 240 | same | |
| Cytarabine 100.0 mg/ml | > 240 | > 240 | same | |
| Dacarbazine (DTIC), 10.0 mg/ml | > 240 | > 240 | same | |
| Daunorubicin 5.0 mg/ml | > 240 | > 240 | same | |
| Docetaxel 10.0 mg/ml | > 240 | > 240 | same | |
| Doxorubicin Hydrochloride, 2.0 mg/ml | > 240 | > 240 | same | |
| Epirubicin 2.0 mg/ml | > 240 | > 240 | same | |
| Etoposide (Toposar), 20.0 mg/ml | > 240 | > 240 | same | |
| Fludarabine 25.0 mg/ml | > 240 | > 240 | same | |
| Fluorouracil, 50.0 mg/ml | > 240 | > 240 | same | |
| Gemcitabine 38.0 mg/ml | > 240 | > 240 | same | |
| Idaribicin 1.0 mg/ml | > 240 | > 240 | same | |
| Ifosfamide 50.0 mg/ml | > 240 | > 240 | same | |
| Irinotecan 20.0 mg/ml | > 240 | > 240 | same | |
| Mechlorethamine HCl 1.0 mg/ml | > 240 | > 240 | same | |
| Melphalan 5 mg/ml | > 240 | > 240 | same | |
| Methotrexate 25 mg/ml | > 240 | > 240 | same | |
| Mitomycin C 0.5 mg/ml | > 240 | > 240 | same | |
| Mitoxantrone 2.0 mg/ml | > 240 | > 240 | same | |
| Oxaliplatin 2.0 mg/ml | > 240 | > 240 | same | |
| Paclitaxel (Taxol), 6.0 mg/ml | > 240 | > 240 | same | |
| Paraplatin 10 mg/ml | > 240 | > 240 | same | |
| Retrovir 10 mg/ml | > 240 | > 240 | same | |
| Rituximab 10 mg/ml | > 240 | > 240 | same | |
| Thiotepa, 10.0 mg/ml | 24.4 | 118.5 | similar | |
| Topotecan HCl 1 mg/ml | > 240 | > 240 | same | |
| Trisonex 1 mg/ml | > 240 | > 240 | same | |
| Velcade (Bortezomib) 1 mg/ml | > 240 | > 240 | same | |
| Vincristine 1.0 mg/ml | > 240 | > 240 | same |
{8}------------------------------------------------
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs
રાજ
{9}------------------------------------------------
11.0 Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.