K Number
K190499
Date Cleared
2019-09-06

(189 days)

Product Code
Regulation Number
878.4300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VESSEL EVERTER System is indicated for use with the Microvascular Anastomotic COUPLER and FLOW COUPLER Device in the anastomosis of only arteries normally encountered in microsurgical procedures only in the peripheral vascular system. The VESSEL EVERTER System is indicated for use with COUPLER and FLOW COUPLER System sizes from 2.0 to 4.0 mm.

Device Description

Two accessory devices comprise the Vessel Everter System: Vessel Everter and Sizing Guide. The Vessel Everter system is an accessory device to the GEM FLOWCOUPLER System, a pair of implantable rings that are used for end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures.

The Vessel Everter System are two non-implantable, sterile, single use hand-held devices that facilitate sizing and use of the GEM FLOWCOUPLER System. The Vessel Everter is comprised of a handle with two soft silicone end effectors (tips), one on each end. The instrument handle features two separate silicone end effectors, each of a different size, intended to be used at the surgeon's discretion. The Vessel Everter end effector is designed with a narrowing tip to allow for engagement with multiple sizes of vessels, corresponding with a coupler size of 2.0-4.0 mm rings. The Vessel Everter is used to press vessel tissue onto the GEM FLOWCOUPLER rings. To accomplish this, the surgeon presses the Vessel Everter into the tissue, flaring the tissue onto and over the Coupler locking pins.

The Sizing Guide is a hard, plastic handle featuring holes on either end through which the vessel's outer diameter may be measured. The holes' diameters are intended to match the inner diameters of the various GEM FLOWCOUPLER rings. The holes correlate to vessel diameters of 2.0 mm to 4.0 mm in 0.5 mm increments. The surgeon may use the sizing guide to guide the selection of which coupler they will use for each particular anastomosis. The Sizing Guide is used by placing the vessel on the sizing guide and comparing the vessel diameter to the holes in the Sizing Guide.

The Vessel Everter and Sizing Guide are sterilized by ethylene oxide and are provided sterile for-single-patient use. The Vessel Everter and Sizing Guide are not implantable and are disposed of after single use.

The parent device, GEM FLOWCOUPLER System, remains unchanged and is not addressed in this submission.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the GEM FlowCoupler System accessories (Vessel Everter and Sizing Guide). It does not contain information about a study proving the device meets acceptance criteria in the context of an AI/ML device. Instead, it details performance testing for a medical device's physical and biological properties.

Therefore, many of the requested fields cannot be filled as they pertain to AI/ML device validation studies. However, I can extract the available information regarding the device's performance testing for the physical accessories.

Here's a breakdown based on the provided text, acknowledging the limitations for AI/ML specific information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Test)Reported Device Performance (Result)
Silicone tip retention - tensile forceSubject devices met acceptance criteria
Retention Barb - Cantilever LoadingSubject devices met acceptance criteria
Coupler Engagement and Disengagement ForceSubject devices met acceptance criteria
Intimal Integrity (porcine vessels)Subject devices met acceptance criteria
Biocompatibility: Cytotoxicity (L929 MEM Elution)Not cytotoxic per ISO10993-5 (PASS)
Biocompatibility: Sensitization (Kligman Maximization test)Non-sensitizer per ISO10993-10 (PASS)
Biocompatibility: Irritation (intracutaneous injection in rabbit)Negative per ISO10993-10 (PASS)
Biocompatibility: Toxicity (systemic injection in mouse)Negative for toxicity per ISO10993-11 (PASS)
Biocompatibility: Rabbit blood hemolysis testNon-hemolytic per ASTM F756 (PASS)

2. Sample size used for the test set and the data provenance

The text mentions "porcine vessels" for the Intimal Integrity Verification Testing but does not specify the sample size. It does not mention any human data or specific provenance for these tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The tests performed are physical and biological evaluations of the device's materials and mechanical properties, not an AI/ML diagnosis where expert ground truth is established.

4. Adjudication method for the test set

Not applicable. This concept is relevant for AI/ML performance evaluation against a human-adjudicated ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the performance tests outlined typically refers to established standards and measurement techniques for mechanical properties (e.g., force measurements), material characteristics (e.g., cytotoxicity assays), and biological responses (e.g., hemolysis tests). For the intimal integrity test, the "ground truth" would be the observed integrity of the porcine vessel intima after device use.

8. The sample size for the training set

Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

September 6, 2019

Baxter/ Synovis Micro Companies Alliance Inc. Julie Carlston Sr. Dept. Specialist/ R & D 2875 University Ave. West St. Paul, Minnesota 55114

Re: K190499

Trade/Device Name: GEM FlowCoupler System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: MVR, DPW Dated: August 1, 2019 Received: August 9, 2019

Dear Julie Carlston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cindy Chowdhury, Ph.D., M.B.A Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190499

Device Name GEM FlowCoupler System

Indications for Use (Describe)

The VESSEL EVERTER System is indicated for use with the Microvascular Anastomotic COUPLER and FLOW COUPLER Device in the anastomosis of only arteries normally encountered in microsurgical procedures only in the peripheral vascular system. The VESSEL EVERTER System is indicated for use with COUPLER and FLOW COUPLER System sizes from 2.0 to 4.0 mm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. Name of Submitter, Contact Person, and Date Prepared:

Name:Baxter/ Synovis Micro Companies Alliance Inc.
Address:2875 University Ave. WestSt. Paul, MN 55114
Telephone:651-796-7511
Contact PersonJulie Carlston
Telephone:571-299-9806 Mobile
Email:julie_carlston@baxter.com
Date Prepared:August 1, 2019

II. DEVICE

  • a. 510(k) Number: K190499
  • Trade Name: GEM FlowCoupler System b.
  • Common Name: Implantable clip and accessories C.
  • d. Classification Name: Device, Anastomotic, Microvascular
  • Classification Regulation: 21 CFR 878.4300 e.
  • Device Classification: Class II f.
  • Classification Product Code: MVR and DPW g.
  • h. Panel: General and Plastic Surgery

III. PREDICATE DEVICE:

K143589 GEM FlowCoupler System This predicate has not been subject to a design-related recall. PERFORMANCE STANDARDS

None required under Section 514 of the Act

IV. DEVICE DESCRIPTION

Two accessory devices comprise the Vessel Everter System: Vessel Everter and Sizing Guide. The Vessel Everter system is an accessory device to the GEM FLOWCOUPLER System, a pair of implantable rings that are used for end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures.

The Vessel Everter System are two non-implantable, sterile, single use hand-held devices that facilitate sizing and use of the GEM FLOWCOUPLER System. The Vessel Everter is comprised of a handle with two soft silicone end effectors (tips), one on each end. The instrument handle features two separate silicone end effectors, each of a different size, intended to be used at the surgeon's discretion. The Vessel Everter end effector is designed with a narrowing tip to allow for engagement with multiple sizes of vessels, corresponding with a coupler size of 2.0-4.0 mm rings. The Vessel Everter is used to press vessel tissue onto the GEM FLOWCOUPLER rings. To accomplish this, the surgeon presses the Vessel Everter into the tissue, flaring the tissue onto and over the Coupler locking pins.

The Sizing Guide is a hard, plastic handle featuring holes on either end through which the vessel's outer diameter may be measured. The holes' diameters are intended to match the

{4}------------------------------------------------

inner diameters of the various GEM FLOWCOUPLER rings. The holes correlate to vessel diameters of 2.0 mm to 4.0 mm in 0.5 mm increments. The surgeon may use the sizing guide to guide the selection of which coupler they will use for each particular anastomosis. The Sizing Guide is used by placing the vessel on the sizing guide and comparing the vessel diameter to the holes in the Sizing Guide.

The Vessel Everter and Sizing Guide are sterilized by ethylene oxide and are provided sterile for-single-patient use. The Vessel Everter and Sizing Guide are not implantable and are disposed of after single use.

The parent device, GEM FLOWCOUPLER System, remains unchanged and is not addressed in this submission.

V. INDICATIONS FOR USE

The accessory device Indications for Use are as follows:

The VESSEL EVERTER System is indicated for use with the Microvascular Anastomotic COUPLER and FLOW COUPLER Device in the anastomosis of only arteries normally encountered in microsurgical procedures only in the peripheral vascular system. The VESSEL EVERTER System is indicated for use with COUPLER and FLOW COUPLER System sizes from 2.0 to 4.0 mm.

The indications for use for the parent device (GEM FLOWCOUPLER System) remain unchanged:

The FlowCOUPLER Device is a single use. implantable device that is intendedto be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-f it onto one of the rings. When the FlowCOUPLER Device is used in conjunction with the FlowCOUPLER Monitor, the FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomnotic site. Postoperatively, blood flow can be detected on an as needed basis for up to 7 days. The FlowCOUPLER Doppler probe is not intended to be a permanent implant and should be removed 3 to 14 days post-operatively.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Comparison to Predicate Device: This premarket notification describes addition of two accessories for single-patient-use: a disposable single use Vessel Everter and Sizing Guide. The system currently includes reusable Vessel Everter and Sizing Guide accessory devices (the predicate devices) which the disposable accessories will replace.

The subject Vessel Everter device has the same intended use as the predicate, reusable Vessel Everter accessory; both assist the surgeon in placing the vessel over the Coupler.

{5}------------------------------------------------

The subject single-patient-use Sizing Guide has the same intended use as the predicate reusable Sizing Guide to assist the surgeon in determining appropriate coupler size. The subject and predicate devices differ in materials (plastic vs. metal) and the subject device is provided sterile unlike the predicate which requires the user to sterilize prior to use. The subject and predicate accessories are equivalent with regard to intended use, target population, anatomical site, use setting, performance, compatibility with the environment and other devices, and biocompatibility. The subject device accessories differ from the predicate device accessories primarily with respect to sterilization method) and materials.

TestTest Method SummaryResults
Silicone tip retentiontensile forceverification testingMeasurement of tension forcerequired to completelydissociate the silicon tip fromthe barbed handle.Subject devices metacceptance criteria
Retention BarbCantilever LoadingVerification TestingMeasurement of forcerequired to dissociate thesilicon tip from the barbedhandle tested by applying acompressive loadperpendicular (90°) to the axisof the retention barbSubject devices metacceptance criteria
Coupler Engagementand DisengagementForce VerificationMeasurement of forcerequired to disengage evertervessel silicone tip fromcoupler pins, force to piercethe coupler pins, and numberof penetrations onto thecoupler barbsSubject devices metacceptance criteria
Intimal IntegrityVerification TestingUse of the device in porcinevessels. Evaluation of forcethat anastomosis canwithstand without a full-thickness dissection of theintima in porcine vesselsSubject devices metacceptance criteria

PERFORMANCE TESTING VII.

VIII. BIOCOMPATIBILITY TESTING

Test NameResultDisposition
Cytotoxicity - L929 MEM ElutionNot cytotoxic per ISO10993-5PASS
Sensitization – KligmanMaximization testNon-sensitizer per ISO10993-10PASS

{6}------------------------------------------------

Test NameResultDisposition
Irritation - intracutaneousinjection in rabbitNegative per ISO10993-10PASS
Toxicity - systemic injection inmouseNegative for toxicity per ISO10993-11PASS
Rabbit blood hemolysis testNon-hemolytic per ASTM F756PASS

IX. CLINICAL STUDIES

Clinical studies were not required to support substantial equivalence between the reusable accessories and the single-patient-use accessories.

X. SUBSTANTIAL EQUIVALENCE

The subject device accessories have the same intended use as the predicate accessories and the same technological characteristics. Testing in porcine vessels demonstrates that the device may be used per its intended use. There are no new types of questions of safety and effectiveness and the Vessel Everter and Sizing Guide are substantially equivalent to the predicate accessories.

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.