K Number
K190275
Device Name
GenePOC Carba
Manufacturer
Date Cleared
2019-05-10

(91 days)

Product Code
Regulation Number
866.1640
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GenePOC™ Carba assay, performed on the revogene™ instrument, is a qualitative in vitro diagnostic test designed for the detection and differentiation of the blakec. blaNDM, blackA-48-like, and blamp gene sequences associated with carbapenem-nonsusceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar or MacConkey agar. The test utilizes automated real-time Polymerase Chain Reaction (PCR).

The GenePOC™ Carba assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing. A negative GenePOC™ Carba assay result does not preclude the presence of other resistance mechanisms.

The GenePOC™ Carba assay is intended as an aid for infection control in the detection of carbapenem-non-susceptible bacteria that colonize patients in healthcare settings. The identification of a bland, blaym or blaviM metallo-B-lactamase gene (i.e., the genes that encode the IMP, NDM and VIM metallo-ß-lactamases, respectively) may be used as an aid to clinicians in determining appropriate therapeutic strategies for patients with known or suspected carbapenem non-susceptible infections.

Device Description

The GenePOC™ Carba assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection and differentiation of the blaype, blayDM, blavin, blaoxA-48-ike, and blamp gene sequences in carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa grown on blood agar or MacConkey agar. The GenePOC™ Carba assay is designed to be performed on the revogene™ instrument. The revogene™ is an instrument that automates sample homogenization, sample dilution, cells lysis, DNA amplification, and detection of the amplified PCR products.

The GenePOC™ Carba assay kit is comprised of single-use, disposable microfluidic cartridges (PIEs), Sample Buffer Tubes (SBT), and Disposable Transfer Tools (DTT). These components are used to dilute the sample, extract, amplify, and simultaneously detect blayDM, blavin, blaoxA-48-like, and blamp DNA. A Process Control (PrC) is also incorporated into each PIE to verify sample processing and amplification steps. The PrC allows for the verification of potential inhibitor substances as well as microfluidic, instrument or reagent failure.

Each GenePOC™ Carba assay kit contains 24 individual pouches. Each pouch has components for one (1) test including one (1) Carba PIE, one (1) SBT, and one (1) DTT. The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of the blaker, blands, blaoxa-48-like, and blanyp gene sequences from isolates of pure cultures of carbapenem-non-susceptible gram-negative bacteria in approximately 70 minutes. User intervention is only required for preparing the standardized 0.5 McFarland bacterial suspension from characterized carbapenem-non-susceptible isolated colonies, inoculating the bacterial suspension into the Sample Buffer Tube (SBT), transferring the sample into the single-use disposable PIE, and loading/unloading the PIEs into the revogene™ carousel. Each PIE is a completely integrated closed device in which a sample is dispensed and processed through different microfluidic chambers and channels that allow for the sample processing and subsequent real-time PCR steps.

Upon completion of a run, the results are interpolated by the revogene™ from measured fluorescent signals and embedded calculation algorithms. The output results include positive, negative, indeterminate, and unresolved. Upon completion of a run, the user removes the used cartridges and disposes of them in normal biological waste. The results are automatically generated at the end of the process in a report that can be viewed and printed.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the GenePOC™ Carba assay, based on the provided FDA 510(k) summary:

Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the clinical study's objective: to establish the performance characteristics (sensitivity and specificity) of the GenePOC™ Carba assay compared to a Reference Method for detecting carbapenemase genes. No explicit numerical acceptance criteria (e.g., "Sensitivity >= 95%") are formally stated in the provided text as pass/fail thresholds for clearance, but the reported performance represents what was achieved and found acceptable for substantial equivalence.

Table of Acceptance Criteria (Implied) and Reported Device Performance

The summary presents performance broken down by each target gene. The "Reference Method" in the clinical study refers to an FDA-cleared Nucleic Acid Amplification Test (NAAT) whose performance was established against PCR/bidirectional sequencing.

Performance on Isolated Colonies Grown on Blood Agar Relative to Reference Method (Overall by Target)

TargetImplied Acceptance Criterion (High Sensitivity & Specificity)Reported Sensitivity (95% CI)Reported Specificity (95% CI)
NDMHigh Sensitivity & Specificity98.9% [96.2 - 99.7%]99.4% [97.8 - 99.8%]
KPCHigh Sensitivity & Specificity99.1% [95.2 - 99.8%]99.2% [97.8 - 99.7%]
OXA-48-likeHigh Sensitivity & Specificity100.0% [94.4 - 100.0%]99.3% [98.0 - 99.8%]
IMPHigh Sensitivity & Specificity100.0% [87.5 - 100.0%]96.1% [94.0 - 97.5%]
VIMHigh Sensitivity & Specificity100.0% [93.1 - 100.0%]99.8% [98.8 - 100.0%]

Performance on Isolated Colonies Grown on MacConkey Agar Relative to Reference Method (Overall by Target)

TargetImplied Acceptance Criterion (High Sensitivity & Specificity)Reported Sensitivity (95% CI)Reported Specificity (95% CI)
NDMHigh Sensitivity & Specificity98.9% [96.2 - 99.7%]99.4% [97.8 - 99.8%]
KPCHigh Sensitivity & Specificity100.0% [96.7 - 100.0%]99.2% [97.8 - 99.7%]
OXA-48-likeHigh Sensitivity & Specificity100.0% [94.4 - 100.0%]99.6% [98.4 - 99.9%]
IMPHigh Sensitivity & Specificity100.0% [87.5 - 100.0%]95.7% [93.5 - 97.2%]
VIMHigh Sensitivity & Specificity100.0% [93.1 - 100.0%]99.8% [98.8 - 100.0%]

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Size: 512 fully compliant bacterial isolates.
      • 475 clinical stock isolates.
      • 57 prospectively collected fresh isolates.
    • Data Provenance: Three clinical centers (one in Canada and two in the United States of America).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth was established by a "Reference Method" (an FDA-cleared Nucleic Acid Amplification Test) whose performance was previously established against PCR/bidirectional sequencing. The document does not specify the number or qualifications of experts involved in running or interpreting this reference method or the PCR/bidirectional sequencing, nor in the initial characterization of the clinical isolates.
  3. Adjudication method for the test set:

    • Discordant analysis: For samples with discrepant results between the GenePOC™ Carba assay and the Reference Method, further testing was performed using alternative PCR for each of the five assay analytes, followed by bi-directional Sanger sequencing. The results of this discrepant testing were then used to reconcile the initial findings (effectively re-evaluating the "true" status of the sample).
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving human readers with/without AI assistance was not performed. This device is an automated in vitro diagnostic test, not an AI-assisted diagnostic imaging device targeting human interpretation.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the clinical study assessed the "standalone" performance of the GenePOC™ Carba assay. The device is an automated real-time PCR platform that extracts, amplifies, and detects target DNA, and "the results are interpolated by the revogene™ from measured fluorescent signals and embedded calculation algorithms." User interaction is primarily for sample preparation and loading/unloading. The reported sensitivity and specificity values represent this standalone algorithmic performance against the defined Reference Method.
  6. The type of ground truth used:

    • The ground truth for the clinical study was established by a Reference Method, which was an FDA-cleared Nucleic Acid Amplification Test (NAAT). This NAAT itself had its performance previously established in comparison to PCR/bidirectional sequencing. For discordant results, alternative PCR followed by bi-directional Sanger sequencing was used for final adjudication, indicating that molecular sequencing data served as the ultimate ground truth for discrepant cases.
  7. The sample size for the training set:

    • The document describes analytical and clinical performance studies, but it does not specify a separate "training set" for the algorithm's development in the context of machine learning. For an in vitro diagnostic like this, the "training" involves assay design, optimization, and establishment of thresholds. The document states, "The assay cut-offs were determined by testing a total of 599 samples" (Section N.1.g), which could be considered part of the internal development/training phase for establishing the algorithm's decision boundaries.
  8. How the ground truth for the training set was established:

    • As noted above, for the "training" (assay cut-off determination), ground truth would have been established through well-characterized samples with known presence or absence of the target genes. The analytical reactivity (inclusivity) study also contributes to this, using 58 carbapenem-non-susceptible isolates with known resistance genes and variants, determined by their collection number, geographical/temporal origin, and reported resistance gene/variant. For the 599 samples used to determine assay cut-offs, the specific method for establishing their ground truth is not detailed, but it is implied to be based on the known genetic makeup of the strains, likely through sequencing or other established molecular methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 10, 2019

GenePOC Inc. Dany Leblanc VP QA/RA 360 rue Franquet Quebec, G1P 4N3 Ca

Re: K190275

Trade/Device Name: GenePOC Carba Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: PMY, OOI Dated: February 6, 2019 Received: February 8, 2019

Dear Dany Leblanc:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(K) SUMMARY 5.0

{3}------------------------------------------------

510K SUMMARY

510(k) Number: K190275

A. GENERAL INFORMATION

Submission Date: April 29, 2019

Submitter Information:

Submitted By:GenePOC Inc.360 rue FranquetQuébec (Québec) G1P 4N3
Contact Person:Dany LeblancVP Quality Assurance and Regulatory AffairsGenePOC Inc.Telephone: +1 418 650-3535Email: dany.leblanc@genepoc.ca

B. PURPOSE FOR SUBMISSION

To obtain a substantial equivalence determination for the GenePOC™ Carba assay on the revogene™ instrument for the qualitative detection and differentiation of the blaxer, blayDM, blayM, blaoxA-48-ike, and blamp gene sequences from carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa.

C. MEASURAND

Target DNA sequence of the following genes: blakec, blandM, blaving, blaoxA-48like, and blanp

D. TYPE OF TEST

Qualitative real-time Polymerase Chain Reaction (PCR) assay

E. APPLICANT

GenePOC Inc.

F. PROPRIETARY AND ESTABLISHED NAMES

Proprietary Name: GenePOC™ Carba Common Name : GenePOC™ Carba assay

{4}------------------------------------------------

G. REGULATORY INFORMATION

Trade Name:GenePOC™ Carba
Classification:Class II
Regulation:21 CFR 866.1640
Regulation Name:Antimicrobial susceptibility test powder
Product Code:PMY, OOI
Panel:83 - Microbiology

H. INTENDED USE

    1. Intended Use and Indications for Use:
      The GenePOC™ Carba assay, performed on the revogene™ instrument, is a qualitative in vitro diagnostic test designed for the detection and differentiation of the blakec. blaNDM, blackA-48-like, and blamp gene sequences associated with carbapenem-nonsusceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar or MacConkey agar. The test utilizes automated real-time Polymerase Chain Reaction (PCR).

The GenePOC™ Carba assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing. A negative GenePOC™ Carba assay result does not preclude the presence of other resistance mechanisms.

The GenePOC™ Carba assay is intended as an aid for infection control in the detection of carbapenem-non-susceptible bacteria that colonize patients in healthcare settings. The identification of a bland, blaym or blaviM metallo-B-lactamase gene (i.e., the genes that encode the IMP, NDM and VIM metallo-ß-lactamases, respectively) may be used as an aid to clinicians in determining appropriate therapeutic strategies for patients with known or suspected carbapenem non-susceptible infections.

    1. Special conditions for use statement(s):
      Prescription Use Only

Organisms should be identified and carbapenem non-susceptibility status should be determined prior to testing with the GenePOCTM Carba assay.

The GenePOC™ Carba assay detects blakec, blandM, blavisM, blaoxA-48-like, and blamp resistance genes from pure colonies and is not for bacterial identification.

The performance of the GenePOC™ Carba assay with bacteria other than Enterobacteriaceae, Acinetobacter baumannii or Pseudomonas aeruginosa, has not been evaluated.

{5}------------------------------------------------

The GenePOC™ Carba assay is not a sub-typing tool and does not report variants of the blakpc, blandm, blaym, blaoxa-48-like, and blandp

    1. Special instrument requirements:
      The GenePOC™ Carba assay uses PCR technology on the revogene™ instrument platform, which extracts, amplifies, and detects the target DNA.

I. INDICATIONS FOR USE

Same as Intended Use.

J. DEVICE DESCRIPTION

The GenePOC™ Carba assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection and differentiation of the blaype, blayDM, blavin, blaoxA-48-ike, and blamp gene sequences in carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa grown on blood agar or MacConkey agar. The GenePOC™ Carba assay is designed to be performed on the revogene™ instrument. The revogene™ is an instrument that automates sample homogenization, sample dilution, cells lysis, DNA amplification, and detection of the amplified PCR products.

The GenePOC™ Carba assay kit is comprised of single-use, disposable microfluidic cartridges (PIEs), Sample Buffer Tubes (SBT), and Disposable Transfer Tools (DTT). These components are used to dilute the sample, extract, amplify, and simultaneously detect blayDM, blavin, blaoxA-48-like, and blamp DNA. A Process Control (PrC) is also incorporated into each PIE to verify sample processing and amplification steps. The PrC allows for the verification of potential inhibitor substances as well as microfluidic, instrument or reagent failure.

Each GenePOC™ Carba assay kit contains 24 individual pouches. Each pouch has components for one (1) test including one (1) Carba PIE, one (1) SBT, and one (1) DTT. The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of the blaker, blands, blaoxa-48-like, and blanyp gene sequences from isolates of pure cultures of carbapenem-non-susceptible gram-negative bacteria in approximately 70 minutes. User intervention is only required for preparing the standardized 0.5 McFarland bacterial suspension from characterized carbapenem-non-susceptible isolated colonies, inoculating the bacterial suspension into the Sample Buffer Tube (SBT), transferring the sample into the single-use disposable PIE, and loading/unloading the PIEs into the revogene™ carousel. Each PIE is a completely integrated closed device in which a sample is dispensed and processed through different microfluidic chambers and channels that allow for the sample processing and subsequent real-time PCR steps.

Upon completion of a run, the results are interpolated by the revogene™ from measured fluorescent signals and embedded calculation algorithms. The output results include positive, negative, indeterminate, and unresolved. Upon completion of a run, the user removes the used

{6}------------------------------------------------

cartridges and disposes of them in normal biological waste. The results are automatically generated at the end of the process in a report that can be viewed and printed.

K. SUBSTANTIAL EQUIVALENCE INFORMATION

    1. Predicate Device Name: Xpert® Carba-R Assay
    1. Predicate 510(k) Number: K152614
    1. Comparison with Predicate:

The GenePOC™ Carba assay is substantially equivalent to the Xpert® Carba-R Assay (K152614). The GenePOC™ Carba and the Xpert Carba-R assays both detect target gene sequences from antibiotic-non-susceptible bacteria and use real-time PCR amplification and fluorogenic target-specific hybridization detection. The performance of the GenePOC™ Carba assay was determined in a multi-site clinical study. The table below shows the similarities and differences between the GenePOC™ Carba assay and the predicate device (Xpert® Carba-R Assay).

ItemGenePOCT™ Carba assayXpert® Carba-R Assay(Predicate Device)
K NumberK190275K152614
SIMILARITIES
ClassificationClass IIClass II
Intended UseThe GenePOC™ Carba assay, performed onthe revogene™ instrument, is a qualitativein vitro diagnostic test designed for thedetection and differentiation of the blaKPC,blaNDM, blaVIM, blaOXA-48-like, and blaIMPgene sequences associated withcarbapenem-non-susceptible pure coloniesof Enterobacteriaceae , Acinetobacterbaumannii , or Pseudomonas aeruginosa ,when grown on blood agar or MacConkeyagar. The test utilizes automated real-timePolymerase Chain Reaction (PCR).The GenePOC™ Carba assay should beused in conjunction with other laboratorytests including phenotypic antimicrobialsusceptibility testing. A negativeGenePOC™ Carba assay result does notpreclude the presence of other resistancemechanisms.The GenePOC™ Carba assay is intended asan aid for infection control in the detectionof carbapenem-non-susceptible bacteria thatcolonize patients in healthcare settings. Theidentification of a blaKPC, blaNDM or blaVIMThe Xpert® Carba-R Assay,performed on the GeneXpert®Instrument Systems, is a qualitativein vitro diagnostic test for thedetection and differentiation of theblaKPC, blaNDM, blaVIM, blaOXA-48,and blaIMP gene sequencesassociated with carbapenem-non-susceptible pure colonies ofEnterobacteriaceae , Acinetobacterbaumannii , or Pseudomonasaeruginosa grown on blood agar orMacConkey agar. The test utilizesautomated real-time polymerasechain reaction (PCR).A negative Xpert Carba-R Assayresult does not preclude thepresence of other resistancemechanisms. The Xpert Carba-RAssay should be used in conjunctionwith other laboratory tests includingphenotypic antimicrobialsusceptibility testing. The XpertCarba-R Assay is intended as an aidfor infection control in detecting
ItemGenePOCT™ Carba assayXpert® Carba-R Assay(Predicate Device)
metallo-β-lactamase gene (i.e., the genesthat encode the IMP, NDM and VIMmetallo-β-lactamases, respectively) may beused as an aid to clinicians in determiningappropriate therapeutic strategies forpatients with known or suspectedcarbapenem non-susceptible infections.and differentiating genetic markersof resistance to monitor the spreadof carbapenem-non-susceptibleorganisms in healthcare settings.The Xpert Carba-R Assay is notintended to guide or monitortreatment for carbapenem-non-susceptible bacterial infections.
TechnologicalPrinciplesAutomated nucleic acid amplification; real-time PCRSame
Test CartridgeDisposable single-use, multi-chamberedfluidic cartridgeSame
Detection ProbesTaqMan® ProbesSame
Assay TargetsDetects blaKPC, blaNDM, blaVIM, blaOXA-48-like,and blaIMP genes.Same
Sample PreparationAutomatedSame
Interpretation of TestResultsAutomatedSame
Sample TypeBacterial isolates from cultureSame
OrganismsEnterobacteriaceae, Pseudomonasaeruginosa, Acinetobacter baumanniiSame
DIFFERENCES
Instrument Systemrevogene™GeneXpert Instrument System(includes GeneXpert Dx, Infinity-48, Infinity-48s, and Infinity-80)
Time to Obtain TestResultsApproximately 70 minutesApproximately 50 minutes
KPC-2, 3, 4KPC-2, 3, 4
Variant Type(s)Detected (based onanalytical studies) 1NDM-1, 4 to 7NDM-1, 2, 4, 5
IMP-1, 4, 8, 9, 11IMP-1, 2, 4, 6, 10, 11
OXA-48, 181, 204, 232OXA-48, 181
VIM-1, 2, 10, 19VIM-1, 2, 4, 10, 19
Variant Types(s) NotDetected (based onanalytical studies) 1NoneIMP-7, 13, 14
KPC-5 to 38KPC-5 to 16
NDM-2, 3, 8 to 24NDM-3, 6 to 9
Additional DetectableVariant Types(s)Predicted from insilico analysis1IMP-2, 5, 6, 10, 13 to 20, 23 to 26, 28 to 30,32, 33, 37, 38, 40, 42, 45, 47 to 49, 53 to 56,59, 60, 62, 66, 69 to 72, 74 to 79IMP-3, 8, 9, 13, 19 to 22, 24, 25, 27,28, 30, 31, 33, 37, 40, 42
OXA-162, 199, 244, 245, 252, 370, 484,505, 514, 515, 519, 546, 547, 566OXA-162, 163, 204, 232, 244, 245,247
VIM-3 to 6, 8, 9, 11, 12, 14 to 18, 20, 23 to46, 48 to 50, 52 to 55, 57, 59, 60VIM 5 to 9, 11 to 18, 20, 23 to 38

{7}------------------------------------------------

1 Note: The variants listed reflect the respective labeling of each device and the analytical studies and in silico analyses conducted at the time of 510(k) clearance.

{8}------------------------------------------------

The GenePOC™ Carba assay has the same general intended use as the predicate device and has the same technological characteristics as the predicate device. The differences between the GenePOC™ Carba assay and the predicate device do not raise new questions of safety and effectiveness.

L. STANDARDS/GUIDANCE DOCUMENTS REFERENCED

  • CLSI Guideline EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; . Approved Guideline, 2009.
  • . CLSI EP28-A3c, Guideline Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, 2008.
  • CLSI M02, Performance Standards for Antimicrobial Disk Susceptibility Tests; Approved . Guideline, 13th edition, 2018.
  • CLSI M100, Performance Standards for Antimicrobial Susceptibility Testing; Approved . Guideline, 28th edition, 2018.

M. TEST PRINICIPLE

The GenePOC™ Carba assay is a single-use test for the qualitative detection and differentiation of the blake, blaNDM, blavin, blaoxA-48-ike, and blaIMP gene sequences in carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa grown on blood agar or MacConkey agar that utilizes an automated sample preparation and real-time polymerase chain reaction (PCR) technology with fluorogenic detection of the amplified DNA. The GenePOC™ Carba assay utilizes the GenePOC™ Carba microfluidic cartridge (PIE) for the simultaneous detection of the blaxer, bland, blaym, blaoxa-48-like, and blamp DNA and the internal process control (PrC) DNA (to verify sample processing, amplification, and the absence of reaction inhibitors).

Well-characterized Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, carbapenem-non-susceptible strain is streaked for isolation on a blood agar plate (BAP) or a MacConkey agar plate with a 10-ug meropenem disc on the first quadrant. The plate is then incubated for 18-24 hours at 35°C ± 2°C under aerobic conditions. A standardized 0.5 McFarland (McF) suspension is prepared in saline solution by collecting isolated colonies selected from the incubated agar plate with a sterile swab. Fifteen (15) uL of the 0.5 McF suspension are transferred into the Sample Buffer Tube (SBT) containing 1 mL of Sample Buffer (SB). After a 15-second vortex step at maximal speed, a volume of SB lying between the two marks of the DTT is sampled from the SBT and transferred into a GenePOC™ Carba PIE to perform the assay. The loaded GenePOC™ Carba PIE is placed into the revogene™ for further sample processing. No operator intervention is necessary once the sample is loaded onto the revogene™.

Each Carba PIE is a completely integrated and self-contained device. Each sample is sequentially transferred by centrifugation from one microfluidic chamber to the next and all reagents specific for the PCR reaction are incorporated and dried within the PCR wells. The stepwise process includes sample homogenization, lysis of cells, and sample dilution followed by the subsequent real-time PCR steps. An internal PrC is contained in the homogenization chamber and is therefore present in every test to verify critical steps of the analytical process

{9}------------------------------------------------

(including sample lysis, dilution and nucleic acid amplification and detection) for the presence of potential inhibitory substances as well as system or reagent failures. The amplified products are detected in real time using target-specific TagMan® chemistry-based probes. The results are interpolated by the system from measured fluorescent signals and embedded calculation algorithms. Results may be viewed, printed, transferred, and/or stored by the user.

N. PERFORMANCE CHARACTERISTICS

1. Analytical Performance

  • Reproducibility/Precision a.
    The Reproducibility and Precision study was performed by testing a total of ten (10) positive bacterial carbapenem-non-susceptible strains, each one harboring resistance gene targeted by the GenePOC™ Carba assay (n=2 strains per resistance gene, i.e. blakec, blaNDM, blaOXA-48-like, and blamp) and one (1) negative bacterial carbapenem-non-susceptible strain that does not carry any of the resistance genes targeted by the assay from standardized 0.5 McF suspensions.

The Between-Laboratory Reproducibility study was performed by two (2) operators at three (3) sites with one (1) GenePOC™ Carba assay kit lot over a total of five (5) days (consecutive or not) per panel member tested. Two (2) runs were performed by each operator on each day. On each day of testing, four (4) panel members of each positive strain were tested on each site and by each operator for a total of 120 replicates per strain (5 days of testing x 4 panel members per strain x 3 sites x 2 operators = 120 replicates per strain). Eight (8) negative panel members were also tested on each site and by each operator, on each respective day of testing for a total of 240 replicates (5 days of testing x 8 panel members x 3 sites x 2 operators = 240 replicates).

To assess the Between-Lot Reproducibility, testing was pursued at one (1) site with two (2) additional kit lots for a total of 15 days of testing (five (5) days per kit lot, consecutive or not) per panel member tested for each kit lot. Two (2) runs per kit lot were performed by each operator on each day. On each day of testing, four (4) panel members of each positive strain were tested by each operator for a total of 120 replicates per strain (5 days of testing x 4 panel members per strain x 3 kit lots x 2 operators = 120 replicates per strain). Eight (8) negative panel members were also tested with each kit lot and by each operator, on each respective day of testing for a total of 240 replicates (5 days of testing x 8 panel members x 3 kit lots x 2 operators = 240 replicates).

The Within-Laboratory Precision for each panel member was calculated based on one (1) GenePOC™ Carba assay kit lot (kit lot 1) that was tested by two (2) operators at site 1 over a total of five (5) days (consecutive or not). A total of 40 replicates for each of the ten (10) positive bacterial carbapenem-non-susceptible strains, each one harboring one (1) resistance gene targeted by the GenePOC Carba assay (n=2 strains per resistance gene), were considered for the analysis (5 days of testing x 4 panel members per strain x 2 operators = 40 replicates). A total of 80 replicates of the

{10}------------------------------------------------

negative bacterial carbapenem-non-susceptible strain that does not carry any of the resistance genes targeted by the assay were also considered (5 days of testing x 8 panel members x 2 operators = 80 replicates).

For the negative samples tested, there was agreement of 98.8% (237/240; 95% CI: [96.4-99.6%], 99.6% (239/240; 95% CI: [97.7-99.9%]) and 100.0% (80/80; 95% CI: [95.4-100.0%], respectively for the Between-Laboratory Reproducibility study, the Between-Lot Reproducibility study and for the Within-Laboratory Precision study. For the positive samples, agreement of 100.0% (120/120; 95% CI: [96.9-100.0%]), 100.0% (120/120; 95% CI: [96.9-100.0%]) and 100.0% (40/40; 95% CI: [91.2-100.0%]), was achieved for each panel member in the Between-Laboratory Reproducibility study, the Between-Lot Reproducibility study and for the Within-Laboratory Precision study, respectively.

{11}------------------------------------------------

Results/TotalOverall %Results/TotalOverall %Agreement
ResistanceSite 1Site 2Site 3AgreementResistanceKit Lot 1Kit Lot 2Kit Lot 3
Gene andVariantOperator AOperator BSiteOperator AOperator BSiteOperator AOperator BSite[95% CI]Gene andVariantOperator AOperator BSiteOperator AOperator BSiteOperator AOperator BSite[95% CI]
IMP-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]IMP-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]
IMP-420/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]IMP-420/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%[96.9-100.0]
KPC-420/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]KPC-420/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%[96.9-100.0]
KPC-220/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]KPC-220/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]
NDM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]NDM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%[96.9-100.0]
NDM-520/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]NDM-520/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%(96.9-100.0]
OXA-18120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]OXA-18120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]
OXA-4820/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]OXA-4820/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]
VIM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]VIM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]
VIM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%)[96.9-100.0]VIM-120/2020/2040/4020/2020/2040/4020/2020/2040/40120/120 (100.0%[96.9-100.0]
Negative40/4040/4080/8039/4040/4079/8039/4039/4078/80237/240 (98.8%)[96.4-99.6]Negative40/4040/4080/8040/4039/4079/8040/4040/4080/80239/240 (99.6%)[97.7-99.9]

Tested Target Results and Overall Agreement for the Between-Laboratory Reproducibility Qualitative Analysis

{12}------------------------------------------------

Tested Target Results and Overall Agreement for the Between-Lot Reproducibility Qualitative Analysis

{13}------------------------------------------------

Tested Target Results and Overall Agreement for the Within-Laboratory Precision Qualitative Analysis

Resistance Geneand VariantResults/TotalSite 1 and Kit Lot 1Overall % Agreement[95% CI]
Operator AOperator B
IMP-120/2020/2040/40 (100.0%) [91.2-100.0]
IMP-420/2020/2040/40 (100.0%) [91.2-100.0]
KPC-420/2020/2040/40 (100.0%) [91.2-100.0]
KPC-220/2020/2040/40 (100.0%) [91.2-100.0]
NDM-120/2020/2040/40 (100.0%) [91.2-100.0]
NDM-520/2020/2040/40 (100.0%) [91.2-100.0]
OXA-18120/2020/2040/40 (100.0%) [91.2-100.0]
OXA-4820/2020/2040/40 (100.0%) [91.2-100.0]
VIM-120/2020/2040/40 (100.0%) [91.2-100.0]
VIM-120/2020/2040/40 (100.0%) [91.2-100.0]
Negative40/4040/4080/80 (100.0%) [95.4-100.0]

{14}------------------------------------------------

ResistanceGene andVariantMean Ct¹NBetween-LaboratoryBetween-OperatorBetween-DayRepeatabilityOverall
SDCV%SDCV%SDCV%SDCV%SDCV%
IMP-129.71201.24.00.62.20.00.00.93.11.65.5
IMP-429.61200.72.20.41.40.20.71.03.31.34.3
KPC-430.51201.34.10.72.30.20.81.65.12.17.0
KPC-230.41201.65.20.41.40.10.21.55.02.27.3
NDM-130.31200.30.90.10.20.00.00.93.01.03.1
NDM-530.21200.62.10.51.60.20.51.24.01.54.8
OXA-18128.51200.93.20.30.90.20.81.03.41.44.8
OXA-4830.51201.44.50.51.50.30.91.13.71.96.1
VIM-127.61200.31.10.41.30.00.00.93.11.03.5
VIM-128.51200.31.00.51.70.00.01.13.81.24.3
Negative34.92371.13.10.72.00.41.21.43.91.95.5

Ct Values for the Between-Laboratory Reproducibility Quantitative Analysis

N: Number; SD: Standard Deviation; CV: Coefficient of Variation

1 For resistance genes and variants, Ct values reported are for negative samples, Ct values reported are for the PC.

{15}------------------------------------------------

ResistanceGene andVariantMean Ct¹NBetween-LotBetween-OperatorBetween-DayRepeatabilityOverall
IMP-129.01200.82.60.31.20.10.40.72.51.13.8
IMP-429.21200.82.80.30.90.10.41.13.61.44.7
KPC-429.61201.65.40.31.10.00.01.55.02.27.5
KPC-230.81201.23.90.31.10.31.01.96.12.37.4
NDM-129.51201.03.20.10.30.10.41.13.61.44.8
NDM-528.91201.13.70.31.20.00.01.03.41.55.1
OXA-18128.21201.24.10.30.90.20.80.93.01.55.2
OXA-4828.61201.34.60.20.70.00.01.03.31.65.7
VIM-127.51200.41.50.41.40.10.50.82.91.03.6
VIM-128.11200.93.00.31.00.00.01.03.61.44.8
Negative33.02392.57.50.72.10.41.31.33.92.98.8

Ct Values for the Between-Lot Reproducibility Quantitative Analysis

N: Number; SD: Standard Deviation; CV: Coefficient of Variation

1 For resistance genes and variants, Ct values reported are for negative samples, Ct values reported are for the PC.

{16}------------------------------------------------

ResistanceBetween-DayBetween-RunRepeatabilityOverall
Gene andVariantMean Ct1NSD%CVSD%CVSD%CVSD%CV
IMP-130.0400.00.00.00.00.82.50.82.5
IMP-430.2400.51.50.00.01.03.41.13.7
KPC-431.5400.00.00.00.01.65.01.65.0
KPC-232.2400.10.40.00.01.75.11.75.1
NDM-130.6400.30.90.10.30.93.11.03.2
NDM-530.2400.41.40.00.00.92.81.03.2
OXA-18129.5400.41.20.00.00.93.21.03.4
OXA-4830.1400.00.00.51.61.03.51.13.8
VIM-128.1400.10.30.00.00.93.30.93.3
VIM-129.1400.00.00.20.71.03.41.03.5
Negative35.9800.61.70.41.11.85.12.05.5

Ct Values for the Within-Laboratory Precision Quantitative Analysis

N: Number; SD: Standard Deviation; CV: Coefficient of Variation

1 For resistance genes and variants, Ct values reported are for negative samples, Ct values reported are for the PC.

b. Linearity/Assay Reportable Range

Not Applicable.

  • C. Traceability, Stability, Expected Values (controls, calibrators, or methods)

Internal Process Control (PrC)

Each PIE contains a PrC that controls for amplification inhibition, assay reagents, and sample processing effectiveness.

External Controls

External Controls should be prepared by the end user. These are recommended but not required to perform the GenePOC™ Carba assay. The user should select the most appropriate controls for their laboratory quality control program to comply with applicable regulations and the requirements of accrediting agencies.

{17}------------------------------------------------

As a Positive External Control, GenePOC recommends using a standardized 0.5 McF suspension of Klebsiella pneumoniae or Escherichia coli prepared from a commercially available strain that harbors one of the targeted carbapenemase genes (NCTC 13476 for blamp.1, CCUG 59348 for blakpc.2, ATCC® BAA-2146™ for blaNDM-1, ATCC® BAA-2523TM for blaoxA-48 and NCTC 13440 for blavIM-1).

As a Negative External Control, GenePOC recommends using a standardized 0.5 McF suspension of isolated colonies from a representative carbapenem-non-susceptible bacterial strain of Enterobacteriaceae, A. baumannii or P. aeruginosa that does not carry any of the resistance genes targeted by the GenePOC™ Carba assay.

Sample Stability in the PIE

The GenePOC™ Carba assay PIE (microfluidic cartridge) stability study was verified by testing five (5) positive carbapenem-non-susceptible bacterial strains, each carrying a different carbapenemase gene (blakpc, blayDM, blayM, blaoxA-48-like, and blamp) targeted by the GenePOC™ Carba assay. In addition, a carbapenem-non-susceptible Enterobacter cloacae strain that does not carry any of the carbapenemase resistance genes targeted by the assay was also tested.

Both positive and negative samples were tested with the GenePOC™ Carba assay from standardized 0.5 and 4 McF suspensions, respectively. The results of the study support the recommended maximum interval of one (1) hour at 25 ± 2℃ from opening of the PIE pouch and sample addition into the GenePOC™ Carba assay PIE to processing on the revogene™ instrument.

Sample Stability

Sample stability was evaluated with Sample Buffer Tube (SBT) inoculated with a standardized 0.5 McF suspension of a bacterial strain carrying one of the resistance genes targeted by the GenePOC™ Carba assay (blakec, blaNDM, blaoxA-48-like, and blaMP) and with a standardized 0.5 McF suspension of a carbapenem-nonsusceptible bacterial strain that does not carry any of the resistance genes targeted by the GenePOC™ Carba assay. The product labeling indicates that SBT inoculated with a standardized 0.5 McF suspension prepared from carbapenem non-susceptible bacterial isolates can be stored at 2-8°C for up to seven (7) days or at 25 ± 2°C for up to four (4) days prior to testing with the GenePOC™ Carba assay.

d. Detection Limit

Not applicable.

e. Analytical Reactivity

The analytical reactivity (inclusivity) of the GenePOC™ Carba assay was evaluated with 58 carbapenem-non-susceptible isolates of Enterobacteriaceae. Pseudomonas aeruginosa and Acinetobacter baumannii from various geographic origins representing

{18}------------------------------------------------

temporal diversity and harboring variants of the blakec, blandM, blaoxA-48-like, and blamp carbapenemase genes:

  • Two (2) strains with two (2) resistance genes; -
  • 11 strains (including five [5] variants) with the blamp resistance gene; -
  • 11 strains (including at least three [3] variants) with the blagge resistance gene; -
  • 14 strains (including five [5] variants) with the blaNDM resistance gene; -
  • -11 strains (including three [3] variants) with the blaoxA-48-like resistance gene;
  • 9 strains (including four [4] variants) with the blavM resistance gene. -

Each strain was tested from a standardized 0.5 McF bacterial suspension. Three (3) replicates per strain were tested using three (3) different GenePOC™ Carba kit lots (one (1) replicate per kit lot). The ability of the GenePOC™ Carba assay to detect multiple variants of the blaxec, blaNDM, blaoxA-48-like, and blaiMP carbapenemase genes in the 58 Enterobacteriaceae. Pseudomonas aeruginosa and Acinetobacter baumannii strains carrying these genes was demonstrated.

Carbapenem-Non-susceptible Bacterial Isolates Tested for Analytical Reactivity (Inclusivity) with the GenePOC™ Carba Assay

SpeciesCollection NumberGeographical and Temporal OriginReported Resistance Gene and Variant (When Available)
Isolates Harboring Multiple Genes
PseudomonasaeruginosaATCC® BAA-2793TMChile2014 KPC-2, VIM-2
Klebsiella pneumoniaeCCRI-23061Switzerland22015 OXA-232, NDM-1
KPC Isolates
Klebsiella pneumoniaeNCTC 13438N/A1N/A1 KPC-3
Klebsiella pneumoniaeATCC® BAA-1705TMUSA2007 KPC-2
PseudomonasaeruginosaCCRI-21587Canada22011 KPC-2
Enterobacter cloacaeCCRI-21578Canada2011 KPC-4
Klebsiella oxytocaCCRI-21581Canada2011 KPC-3
Klebsiella pneumoniaeCCRI-19587Canada22009 KPC-3
Klebsiella pneumoniaeCCRI-19570USA2003 KPC-2
Klebsiella pneumoniaeCCUG 59413Sweden2010 KPC-3
Klebsiella pneumoniaeCCUG 59348Norway2010 KPC-2
Klebsiella pneumoniaeCCUG 56233Sweden2088 KPC-2
Escherichia coliATCC® BAA-2340TMUSA (State Public Healthlaboratories)N/A1 KPC
NDM Isolates
Klebsiella pneumoniaeNCTC 13443N/A1N/A1 NDM-1
Klebsiella pneumoniaeATCC® BAA-2146TMUSA (State Public Healthlaboratories)N/A1 NDM-1
Escherichia coliCCRI-22255France22013 NDM-1
Klebsiella pneumoniaeCCRI-21711Canada2011 NDM-1
Klebsiella pneumoniaeCCRI-22199Canada2012 NDM-1
Providencia rettgeriCCRI-22257France22013 NDM-1
Providencia stuartiiCCRI-22256France22013 NDM-1
Klebsiella pneumoniaeCCRI-22254France22013 NDM-4
Escherichia coliCCRI-23064Switzerland22015 NDM-5
SpeciesCollection NumberGeographical and Temporal OriginReported Resistance Gene and Variant (When Available)
Escherichia coliCCRI-23464Canada2016NDM-5
Escherichia coliCCRI-23065Switzerland22015NDM-6
Escherichia coliCCRI-23066Switzerland22015NDM-7
Enterobacter cloacaeATCC® BAA-2468TMUSA (State Public Health laboratories)N/A1NDM-1
Klebsiella pneumoniaeCCUG 60138Sweden2010NDM-1
VIM Isolates
Pseudomonas aeruginosaNCTC 13437N/A1N/A1VIM-10
Klebsiella pneumoniaeNCTC 13439N/A1N/A1VIM-1
Klebsiella pneumoniaeNCTC 13440N/A1N/A1VIM-1
Klebsiella pneumoniaeCCRI-19585France2009VIM-1
Klebsiella pneumoniaeCCRI-22258France22013VIM-1
Pseudomonas aeruginosaCCRI-21588Canada22011VIM-2
Serratia marcescensCCRI-22261France22013VIM-2
Pseudomonas aeruginosaCCRI-22720Argentina2014VIM-2
Klebsiella pneumoniaeCCRI-22259France22013VIM-19
OXA-48-like Isolates
Klebsiella pneumoniaeNCTC 13442N/A1N/A1OXA-48
Klebsiella pneumoniaeCCRI-22263France22013OXA-48
Escherichia coliCCRI-22265France22013OXA-48
Enterobacter cloacaeCCRI-22266France22013OXA-48
Klebsiella pneumoniaeCCRI-22264France22013OXA-181
Providencia rettgeriCCRI-22267France22013OXA-181
Klebsiella pneumoniaeCCRI-23060Switzerland22015OXA-204
Citrobacter freundiiCCRI-23374Canada2016OXA-204
Escherichia coliATCC® BAA-2523TMhuman clinical isolatesN/A1OXA-48
Klebsiella pneumoniaeATCC® BAA-2524TMhuman clinical isolatesN/A1OXA-48
Klebsiella pneumoniaeCCUG 64452Sweden2013OXA-48
IMP Isolates
Escherichia coliNCTC 13476N/A1N/A1IMP-1
Acinetobacter baumanniiCCRI-19488Canada22003IMP-1
Klebsiella pneumoniaeCCRI-19569Japan2003IMP-1
Klebsiella pneumoniaeCCRI-19582Turkey2009IMP-1
Pseudomonas aeruginosaCCRI-21589Canada22011IMP-1
Klebsiella pneumoniaeCCRI-19583Taiwan2009IMP-4
Klebsiella pneumoniaeCCRI-19588Taiwan2009IMP-4
Citrobacter youngaeCCRI-21591Canada22011IMP-4
Klebsiella pneumoniaeCCRI-19584Taiwan2009IMP-8
Pseudomonas aeruginosaCCRI-21590China2000IMP-9
Serratia marcescensCCRI-22262France22013IMP-11

{19}------------------------------------------------

¹ N/A = Not Available
² Isolation country

^2 Isolation country

{20}------------------------------------------------

An in silico analysis was performed to assess inclusivity of the primers and probes of the GenePOC™ Carba assay. For each target, all variants documented in the National Center for Biotechnology Information (NCBI) nucleotide database were analyzed on November 7th, 2018. One representative sequence of each known variant was aligned. Variants for each target, predicted to be detected, are summarized in the table below.

TargetedResistanceGenes# ofSamplesVariantsDetectedVariantsNotDetectedDetectable1PotentiallyDetectable2NotDetectable
blaKPC122, 3, 4None2 to 38NoneN/A
blaNDM151, 4, 5, 6, 7None1 to 24NoneN/A
blaIMP111, 4, 8, 9, 11None1, 2, 4 to 6, 8 to 10,13 to 20, 23 to 26, 28to 30, 32, 33, 37, 38,40, 42, 45, 47 to 49,53 to 56, 59, 60, 62,66, 69 to 72, 74 to 793, 73, 114, 21, 22,27, 34, 41, 43,44, 51, 52, 58,61, 64, 67, 68,7312, 31, 35, 63
blaOXA-48-like1248, 181,204, 232None48, 162, 181, 199,204, 232, 244, 245,2525, 370, 484, 505,5145, 5155, 519, 5465,5475, 566None545, 1636,2476, 4056,4165, 436,4386, 4396,517, 5355,5385, 567
blaVIM101, 2, 10, 19None1 to 6, 8 to 12, 14 to20, 23 to 46, 48 to 50,52 to 55, 57, 59, 6051, 56, 587, 13, 47

GenePOC™ Carba Assay Target Variants Coverage

1 Based on alignments with identity and query cover ≥ 95% and E-values < 0.01.

2 Based on alignments presenting not more than two (2) nucleotides mismatches.

3 A recombinant strain carrying a blance ; gene was tested with GenePOC™ Carba, in complement to the in silico study, and its detection was confirmed.

4 A clinical strain carrying a blank-11 gene was tested with GenePOC™ Carba in the analytical inclusivity study and its detection was confirmed.

5 Variants only identified in the rare opportunistic human pathogens of the Shewanella genus.

6 Variants presenting a deletion in their sequence resulting in no carbapenamase activity.

{21}------------------------------------------------

f. Analytical Specificity

The analytical specificity study was evaluated by testing 50 bacterial strains of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa exhibiting non-susceptibility or susceptibility to the four (4) carbapenems (ertapenem, meropenem, imipenem, doripenem), but not carrying any of the resistance genes targeted by the GenePOC™ Carba assay (blakec, blaNDM, blavm, blaoxA-48-ike, and blamp). Among the 50 strains which carried blactix-M, blasmoc, blasme, blatEM and/or blaspm B-lactamase resistance genes that are different from those targeted by the GenePOCTM Carba assay, only one (1) was observed to cross-react with the GenePOCTM Carba assay. The Klebsiella pneumoniae CCUG 59359 blaTEM-52 strain yielded one positive result when tested at 1.13x107 CFU/mL of SB. However, no crossreactivity was observed for a 10-fold dilution of the same sample (1.13x106 CFU/mL of SB). All other strains analyzed yielded negative results.

Bacterial strains were characterized for their ß-lactamase gene content by whole genome sequencing, except for the Pseudomonas aeruginosa C72 strain which was characterized using an SPM gene-specific PCR method.

Species(Species identified followingsequencing, if different)CollectionNumberβ-Lactamase Gene(s) Identified(100% Coverage)Resistance Phenotype(S/I/R)1ETP1MEM1IMP1DOR1
Enterobacter cloacaeCCRI-3852ACT-7ISSS
Enterobacter cloacaeCCRI-3854ACT-42ISIS
Proteus mirabilisCCRI-21789No β-Lactamase Gene IdentifiedSSIS
Serratia marcescensCCRI-21537SRT-1SSIS
Acinetobacter baumanniiCCRI-1016TEM-90, Mbl, BlaA2, OXA-653,Zn-dependent hydrolaseN/A4SSS
Acinetobacter baumanniiCCRI-1017TEM-206, SCO-1, Mbl, BlaA2,OXA-673, Zn-dependent hydrolaseN/A4SSS
Enterobacter aerogenes(Klebsiella aerogenes)CCRI-19495SHV-5, AmpCRRRR
Enterobacter amnigenus(Enterobacter cloacae)CCRI-22353ACT-15RRRR
Enterobacter cloacaeCCRI-23473No β-Lactamase Gene IdentifiedRRRR
Enterobacter cloacaeCCRI-21536ACT-5RRRR
Enterobacter cloacaeCCRI-21540ACT-7RSIS
Enterobacter cloacaeCCRI-21603ACT-7RRRR
Enterobacter cloacaeCCRI-21692ACT-14RSRI
Species(Species identified followingsequencing, if different)CollectionNumberβ-Lactamase Gene(s) Identified(100% Coverage)Resistance Phenotype(S/I/R)1
ETP1MEM1IMP1DOR1
Enterobacter cloacaeCCRI-22075ACT-7RRRR
Enterobacter cloacaeCCRI-22097ACT-16RRRR
Enterobacter cloacaeCCRI-23318CTX-M-15, TEM-206, CMH-1RRRR
Escherichia coliCCRI-21970AmpC1, AmpC2, MrdA, AmpH,CMY-44RIRI
Escherichia coliCCUG58541CTX-M-14, TEM-104, MrdA,AmpC2, AmpHRSSS
Klebsiella pneumoniaeCCUG58546SHV-44, AmpHRSSS
Pseudomonas aeruginosaCCRI-873OXA-503N/A4SSS
Pseudomonas aeruginosaCCRI-1228OXA-503N/A4SSS
Pseudomonas aeruginosaC72SPMN/A4RRR
Serratia marcescensCCRI-23334SME-4, SRT-1RRRR
Enterobacter aerogenesCCRI-3853AmpCSSSS
Enterobacter aerogenesCCRI-3879AmpCSSSS
Escherichia coliNCTC13441CTX-M-15, TEM-198, MrdA, OXA-13, AmpC2SSSS
Escherichia coliCCRI-21710AmpC2, MrdA, CTX-M-15, AmpH,OXA-13SSSS
Escherichia coliCCRI-778AmpC2, MrdASSSS
Escherichia coliCCRI-779TEM-206, AmpC1, MrdA, AmpC2,AmpHSSSS
Escherichia coliCCRI-785TEM-206, AmpC1, MrdA, AmpC2,AmpHSSSS
Escherichia coliCCRI-878AmpC2, MrdA, AmpHSSSS
Escherichia coliCCUG55970CTX-M-9, AmpC2, TEM-206,MrdA, AmpC1, AmpHSSSS
Escherichia coliCCUG55971CTX-M-15, TEM-143, AmpC2SSSS
Escherichia coliCCUG55972CTX-M-2, AmpC1, AmpC2, AmpHSSSS
Escherichia coliCCUG58540CTX-M-15, AmpC2, TEM-206,MrdA, AMPH, OXA-13SSSS
Escherichia coliCCUG58542CTX-M-15, MrdA, OXA-13,AmpC2SSSS
Klebsiella pneumoniaeNCTC13465SHV-85, TEM-206, AmpHSSSS
Klebsiella pneumoniaeCCUG54718CTX-M-15, TEM-33, OXA-13,AmpHSSSS
Klebsiella pneumoniaeCCUG59358SHV-14, OXA-13, LAP-2SSSS
Species(Species identified followingsequencing, if different)CollectionNumberβ-Lactamase Gene(s) Identified(100% Coverage)Resistance Phenotype(S/I/R)1
ETP1MEM1IMP1DOR1
Klebsiella pneumoniaeCCUG59349CTX-M-15, AmpH, OXA-13, TEM-105, SHV-11SSSS
Klebsiella pneumoniaeCCUG59359TEM-15, SHV-70, AmpHSSSS
Klebsiella pneumoniaeCCUG59360SHV-12, TEM-168, AmpH, OXA-93SSSS
Klebsiella pneumoniaeCCRI-784SHV-27, AmpHSSSS
Klebsiella pneumoniae(Klebsiellaquasipneumoniae)CCRI-806OKP-B-11SSSS
Klebsiella pneumoniaeCCRI-1015TEM-171, SCO-1, PER-2, OXA-93,SHV-39, AmpHSSSS
Proteus mirabilisCCRI-825TEM-33, CTX-M-2, OXA-23SSSS
Proteus mirabilisCCRI-826TEM-215SSSS
Proteus mirabilisCCRI-831TEM-206, CTX-M-2, OXA-23SSSS
Salmonella sp.CCRI-8892CTX-M-5, TEM-166, OXA-13SSSS
Salmonella sp.CCRI-8893CTX-M-5, TEM-95, OXA-13SSSS

Bacterial Strains Tested for Analytical Specificity with the GenePOC™ Carba Assay

{22}------------------------------------------------

{23}------------------------------------------------

1 S/I/R: Sensitive/Intermediate/Resistant, ETP: Ertapenem, IMP: Imipenem, DOR: Doripenem

2 Resistance gene identified with 99.9% homology.

3 These blaox variants are not part of the blaoxA-48-like family, but are part of the Ambler class D.

4 N/A: Not Applicable, A. baumannii and P. aeruginosa possess intrinsic resistance to ertapenem.

The cross-reactivity with primers and probes of the GenePOC™ Carba assay was evaluated by an in silico analysis performed on sequences contained in the National Center for Biotechnology Information (NCBI) database on September 6th, 2018. No other resistance gene than the assay target genes (i.e. blaxpc, blaNDM, blaym, blaoxA-48like, and blamp) was found to have significant level of homology with the primers and probes of the GenePOC™ Carba assay.

  • g. Assay Cut-off
    Thresholds and cut-offs for the GenePOC™ Carba assay are embedded within the Assay Definition File that also encodes the instrument settings required to perform the test. The assay cut-offs were determined by testing a total of 599 samples.

  • h. Interfering Substances
    The interfering substances study for the GenePOC™ Carba assay evaluated the potential interference effects of substances from culture media and/or saline (colony diluent) on GenePOC™ Carba assay performance. A total of 12 potentially interfering

{24}------------------------------------------------

combinations of solid media plates and sterile saline solutions including two (2) types of agar plates (i.e. Blood Agar Plates [BAP] and MacConkey Agar Plates [MAP]) each from three (3) manufacturers and sterile saline solutions from two (2) manufacturers were used during this study.

Five (5) positive bacterial strains harboring each of the resistance genes detected by the GenePOCTM Carba assay and one (1) negative bacterial carbapenem-non-susceptible strain that does not carry any of the resistance genes targeted by the assay were tested from standardized 0.5 McF suspensions.

List of Combinations of Agar Plates and Saline Solutions Tested with the GenePOC™ Carba

Combinations of Agar Plates and Saline Solutions (Manufacturer)
Colombia Blood Agar 5% (Hardy Diagnostics) / BBL™ Prepared Saline Solution (BD)
Colombia Blood Agar 5% (Hardy Diagnostics) / Saline 0.85% (Thermo Scientific™Oxoid™)
Prepared Media BD BBL™ Columbia Agar with 5% Sheep Blood (BD) / BBL™ PreparedSaline Solution (BD)
Prepared Media BD BBL™ Columbia Agar with 5% Sheep Blood (BD) / Saline 0.85%(Thermo Scientific™ Oxoid™)
Columbia Agar with 5% Sheep Blood (bioMérieux) / BBL™ Prepared Saline Solution(BD)
Columbia Agar with 5% Sheep Blood (bioMérieux) / Saline 0.85% (Thermo Scientific™Oxoid™)
MacConkey Agar (Hardy Diagnostics) / BBL™ Prepared Saline Solution (BD)
MacConkey Agar (Hardy Diagnostics) / Saline 0.85% (Thermo Scientific™ Oxoid™)
MacConkey Agar Medium (Thermo Scientific™ Remel™) / BBL™ Prepared SalineSolution (BD)
MacConkey Agar Medium (Thermo Scientific™ Remel™) / Saline 0.85% (ThermoScientific™ Oxoid™)
MacConkey Agar (Thermo Scientific™ Oxoid™) / BBL™ Prepared Saline Solution (BD)
MacConkey Agar (Thermo Scientific™ Oxoid™) / Saline 0.85% (Thermo Scientific™Oxoid™)

The interfering substances study demonstrated the compatibility of the GenePOCTM Carba assay with culture media and colony diluent from different manufacturers.

  • i. Carry-over and Cross-Contamination Studies
    The purpose of the carry-over and cross-contamination study was to determine the carry-over rate of contamination in negative samples due to the nucleic acid amplification of high positive samples. Klebsiella pneumoniae CCUG 59348, a carbapenem-non-susceptible strain, which harbors the blakec gene, was used to prepare positive samples. Enterobacter cloacae CCRI-22760, a representative carbapenemnon-susceptible strain that does not carry any of the carbapenemase resistance genes targeted by the assay, was used to prepare negative samples. The concentration of bacterial suspensions was standardized to 4 McF (≥1.14x10' CFU/mL of SB), higher than the nominal concentration of 0.5 McFarland that is specified for use in the assay.

{25}------------------------------------------------

For the carry-over study, a run of eight (8) replicates of high positive samples followed by a run of eight (8) replicates of negative samples was performed by two (2) operators with the GenePOC™ Carba assay, for a total of ten (10) runs on one (1) revogene.

For the cross-contamination study, a total of ten (10) runs were performed by two (2) operators on one (1) revogene. Testing was performed with alternating high positive and negative samples.

No unexpected positive results for the blaxpc target were detected in negative samples tested in the carry-over study (n=40) nor in the cross-contamination study (n=40).

2. Comparison Studies

  • a. Method Comparison with predicate device
    Refer to Section 3 Clinical Studies

  • b. Matrix Comparison
    Not Applicable.

    1. Clinical Studies
      The clinical study was designed to assess the clinical performance of the GenePOC™ Carba assay for its use in the detection and differentiation of the blaxpc, blaNDM, blagxa-48.Jike, and blanyp gene sequences associated with carbapenem-non-susceptibility on isolated colonies. Three (3) clinical centers (one (1) in Canada and two (2) in the United States of America) participated in this clinical study as testing sites. The Reference Method for the study consisted of a high-performing FDA-cleared Nucleic Acid Amplification Test (NAAT) for the detection of the targeted carbapenemase genes from isolated colonies of known carbapenem non-susceptible organisms, the performance of which was established in comparison to PCR/bidirectional sequencing and which was used according to the manufacturer's instructions. All isolates in the study were tested with the Reference Method and with the GenePOCTM Carba assay. Organism susceptibility status (susceptible, intermediate, or resistant) to meropenem, ertapenem, and/or imipenem was determined using CLSI standard test methods (M02, 13th Edition) and the interpretive criteria found in the FDA drug labels and CLSI M100 28th Edition.

For Reference Method testing, well-isolated colonies grown on blood agar plates were diluted to a standardized 0.5 McFarland (McF) suspension before testing. For GenePOC™ Carba assay testing, well-isolated colonies grown on each of the agar types were suspended to a standardized 0.5 McF suspension. If discordant results between the GenePOC™ Carba assay and Reference Method were observed, discrepant testing was performed using alternative PCR for each of the five (5) assay analytes followed by bi-directional sequencing on isolates from blood agar plates.

{26}------------------------------------------------

A total of 532 bacterial isolates (475 clinical stock isolates and 57 prospectively collected fresh isolates) were enrolled in the clinical study. Sixteen (16) isolates were excluded from the analysis of performance for the following reasons: eleven (11) did not meet the inclusion criteria for species identification (either they were not among the targeted organisms or organism identification was inconclusive), three (3) were unavailable for analysis due to laboratory error and two (2) were found to be susceptible to all four (4) carbapenems tested. Four (4) additional isolates were also excluded because they were associated with a Negative External Control failure on initial testing and produced Indeterminate results upon repeat. Thus, 512 compliant isolates remained in the final analysis.

A total of 512 fully compliant isolates were tested with both the Reference Method and the GenePOC™ Carba assay for the performance estimation study. The primary objective of the trial was to establish the performance characteristics of the GenePOC™ Carba assay for its use in determining the presence of blaype, blaym, blaoxA-48-like, and blann gene DNA sequences in carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar and MacConkey agar, and to establish the sensitivity and the specificity in comparison to the Reference Method.

External controls for the GenePOC™ Carba assay consisted of one (1) negative control and five (5) different positive controls each containing a single target analyte of the assay. The clinical sites ran a negative control and rotated testing one (1) of the positive controls on each day that study samples were tested (i.e. one specific positive control every five (5) days of testing). Study sample results were not valid until expected results were obtained for each control. External control data were compiled across all sites and overall QC results were acceptable.

Isolates grown on both blood agar and MacConkey agar types were evaluated with the GenePOC™ Carba assay for a total of 165 runs. Performance of the GenePOC™ Carba assay was assessed separately for each type of agar and resistance gene target relative to Reference Method. The study showed that 98.4% (1016/1032) of isolates yielded results on the first run. The initial non-reportable rates (combining Unresolved and Indeterminate rates), at the media level, were 1.7% (9/516) from blood agar, and 1.4% (7/516) from MacConkey agar. At the target level, from blood agar, the rates were 1.7% (9/516) for blaker, blands, bland and blaoxA-48-like gene targets and 1.6% (8/516) for blaving gene target, with an overall rate of 1.7% (9/516). From MacConkey agar, the non-reportable rates were 1.4% (7/516) across all targets. Out of the 16 isolates that initially vielded nonreportable results, eight (8) resulted in a non-reportable result upon repeat testing. The final non-reportable rates were 0.8% (4/516) across all media and gene targets.

Discrepant analysis was performed for every sample that produced discordant results between the GenePOC™ Carba assay and the Reference Method. The discrepant analysis comprised five (5) alternative PCR assays, one (1) for each resistance gene targeted by the GenePOCTM Carba assay, followed by bi-directional Sanger sequencing.

{27}------------------------------------------------

Of 31 isolates grown on blood agar with discrepant results, 21 were positive by the alternative PCR/bidirectional sequencing for the same target as the GenePOC Carba assay (see table footnotes below for details on all discrepancies). Seventeen (17) out of 19 samples with False Positive results for IMP target were found to be IMP positive by alternative PCR and bi-directional sequencing, including one (1) variant of IMP-4 (from Australia [2010]), 11 IMP-13/IMP-37 variants (one (1) from Argentina [2006], one (1) from North America [2014], and nine (9) from Europe [2005-2015]), one (1) variant of IMP-27/IMP-64 (from Canada [2017]), one (1) variant of IMP-15 (from Argentina [2004]), one (1) variant of IMP-16 (from Brazil [2004]), and two (2) variants of IMP-62 (from Argentina [2006]). The results of the discrepant analysis suggest that the GenePOC™ Carba assay and Reference Method may exhibit different coverage with regard to variants of IMP. Similar results were observed with colonies grown on MacConkey agar.

A total of 24 isolates contained more than one carbapenemase gene as determined by the GenePOC™ Carba assay. The reported carbapenemase gene content of 21 of these isolates was confirmed by the Reference Method. However, the Reference Method only detected one carbapenemase gene in the other three (3) isolates.

Performance by target and performance by organism group for the GenePOCTM Carba assay with isolates grown on blood agar is shown in the two (2) following tables.

{28}------------------------------------------------

TargetNTPFPTNFNSensitivity[95% CI]Specificity[95% CI]
NDM512186213222698.9%[96.2 - 99.7%]99.4%[97.8 - 99.8%]
KPC51211332395199.1%[95.2 - 99.8%]99.2%[97.8 - 99.7%]
OXA-48-like51265334440100.0%[94.4 - 100.0%]99.3%[98.0 - 99.8%]
IMP512271944660100.0%[87.5 - 100.0%]96.1%[94.0 - 97.5%]
VIM51252154590100.0%[93.1 - 100.0%]99.8%[98.8 - 100.0%]

Performance of GenePOC™ Carba Assay on Isolated Colonies Grown on Blood Agar Relative to Reference Method

N: Number; TP: True Positive; FP: False Positive; TN: False Negative; CI: Confidence Interval

Note: Discrepant testing consisted of five (5) alternative PCR followed by bi-directional sequencing and was performed for every discrepant target result. The results from discrepant testing of 21/31 samples agreed with those of the Carba assay. The notes below summarize the discrepant testing results for each target.

1 One (1) out of two (2) was NDM-1 positive.

2 Two (2) out of three (3) were KPC-3/KPC-38 positive.

3 One (1) out of three (3) was OXA-48 positive. Investigation suggested an OXA cross-contamination at the step of sample preparation in one (1) out of three (3) isolates. Discrepant testing did not produce a sequence match with the OXA-48-like target.

  • 4 17 out of 19 were found IMP positive including one (1) variant of IMP-4 (from Australia [2010]), 11 variants IMP-13/IMP-37 (one (1) from Argentina [2006], one (1) from North America [2014], and nine (9) from Europe [2005-2015]), one (1) variant of IMP-27/IMP-64 (from Canada [2017]), one (1) variant of IMP-15 (from Argentina [2004]), one (1) variant of IMP-16 (from Brazil [2004]), and two (2) variants of IMP-62 (from Argentina [2006]). The discrepant analysis pointed out potential differences in IMP variant coverage between the GenePOC™ Carba assay and the Reference Method. Investigation suggested an IMP crosscontamination at the step of sample preparation in two (2) out of 19 isolates for which discrepant testing did not produce a sequence match with the IMP target.
    5 Investigation suggested a VIM cross-contamination at the step of sample preparation. Discrepant testing did not produce a sequence match with the VIM target but produced a sequence match for the NDM target.

  • 6 Discrepant testing did produce a sequence match with the NDM-1 target in one (1) isolates and produced a sequence match for the OXA-48-like target in one (1) out of two (2) isolates. The OXA-48- positive isolate was classified as FP.

{29}------------------------------------------------

Performance of GenePOC™ Carba Assay by Organism Category and by Target Gene, for Isolated Colonies Grown on Blood Agar Relative to Reference Method

OrganismsTargetNTPFPTNFNSensitivity[95% CI]Specificity[95% CI]
Entero-bacteriaceaeNDM306851219198.8%[93.7 - 99.8%]99.5%[97.5 - 99.9%]
KPC3061123190199.1%[95.2 - 99.8%]98.4%[95.5 - 99.5%]
OXA-48-like3066432390100.0%[94.3 - 100.0%]98.8%[96.4 - 99.6%]
IMP3061452870100.0%[78.5 - 100.0%]98.3%[96.1 - 99.3%]
VIM3061202940100.0%[75.8 - 100.0%]100.0%[98.7 - 100.0%]
PseudomonasaeruginosaNDM107261800100.0%[87.1 - 100.0%]98.8%[93.3 - 99.8%]
KPC107001070-100.0%[96.5 - 100.0%]
OXA-48-like107101060100.0%[20.7 - 100.0%]100.0%[96.5 - 100.0%]
IMP107514880100.0%[56.6 - 100.0%]86.3%[78.3 - 91.6%]
VIM107390680100.0%[91.0 - 100.0%]100.0%[94.7 - 100.0%]
AcinetobacterbaumanniiNDM9975023198.7%[92.9 - 99.8%]100.0%[85.7 - 100.0%]
KPC9910980100.0%[20.7 - 100.0%]100.0%[96.2 - 100.0%]
OXA-48-like9900990-100.0%[96.3 - 100.0%]
IMP9980910100.0%[67.6 - 100.0%]100.0%[96.0 - 100.0%]
VIM9911970100.0%[20.7 - 100.0%]99.0%[94.4 - 99.8%]
TargetNTPFPTNFNSensitivity[95% CI]Specificity[95% CI]
NDM512186213222698.9%[96.2 - 99.7%]99.4%[97.8 - 99.8%]
KPC512114323950100.0%[96.7 - 100.0%]99.2%[97.8 - 99.7%]
OXA-48-like51265234450100.0%[94.4 - 100.0%]99.6%[98.4 - 99.9%]
IMP512272144640100.0%[87.5 - 100.0%]95.7%[93.5 - 97.2%]
VIM51252154590100.0%[93.1 - 100.0%]99.8%[98.8 - 100.0%]
OrganismsTargetNTPFPTNFNSensitivity[95% CI]Specificity[95% CI]
Entero-bacteriaceaeNDM306851219198.8%[93.7 - 99.8%]99.5%[ 97.5 - 99.9%]
KPC30611331900100.0%[96.7 - 100.0%]98.4%[95.5 - 99.5%]
OXA-48-like3066422400100.0%[94.3 - 100.0%]99.2%[97.0 - 99.8%]
IMP3061452870100.0%[78.5 - 100.0%]98.3%[96.1 - 99.3%]
VIM3061212930100.0%[75.8 - 100.0%]99.7%[98.1 - 99.9%]
PseudomonasaeruginosaNDM107261800100.0%[87.1 - 100.0%]98.8%[93.3 - 99.8%]
KPC107001070-100.0%[96.5 - 100.0%]
OXA-48-like107101060100.0%[20.7 - 100.0%]100.0%[96.5 - 100.0%]
IMP107514880100.0%[56.6 - 100.0%]86.3%[78.3 - 91.6%]
VIM107390680100.0%[91.0 - 100.0%]100.0%[94.7 - 100.0%]
AcinetobacterbaumanniiNDM9975023198.7%[92.9 - 99.8%]100.0%[85.7 - 100.0%]
KPC9910980100.0%[20.7 - 100.0%]100.0%[96.2 - 100.0%]
OXA-48-like9900990-100.0%[96.3 - 100.0%]
IMP9982890100.0%[67.6 - 100.0%]97.8%[92.3 - 99.4%]
VIM9910980100.0%[20.7 - 100.0%]100.0%[96.2 - 100.0%]

N: Number; TP: True Positive; FP: False Positive; FN: False Negative; CI: Confidence Interval Multiple target results were observed for some isolates.

{30}------------------------------------------------

Performances with isolates grown on MacConkey agar are shown in the following two (2) tables.

Performance of GenePOC™ Carba Assay on Isolated Colonies Grown on MacConkey Agar Relative to Reference Method

N: Number; TP: True Positive; FP: False Positive; TN: False Negative; CI: Confidence Interval Note: Discrepant testing consisted of five alternative PCR followed by bi-directional sequencing and was performed for every discrepant target result. The results from discrepant testing of 21/31 samples agreed with those of the GenePOC™ Carba assy. The notes below summarize the discrepant testing results for each target.

1 One (1) out of two (2) was NDM-1 positive.

2 Two (2) out of three (3) were KPC-3/KPC-38 positive.

3 One (1) out of two (2) was OXA-48 positive. Investigation suggested an OXA cross-contamination at the step of isolate preparation in one (1) out of two (2) isolates. Discrepant testing did not produce a sequence match with the OXA-48-like target.

4 17 out of 21 were found IMP positive including one (1) variant of IMP-4 (from Australia [2010]), 11 variants IMP-13/IMP-37 (one (1) from Argentina [2006], one (1) from North America [2014], and nine (9) from Europe [2005-2015]), one (1) variant of IMP-27/IMP-64 (from Canada [2017]), one (1) variant of IMP-15 (from Argentina [2004]), one (1) variant of IMP-16 (from Brazil [2004]), and two (2) variants of IMP-62 (from Argentina [2006]). The discrepant analysis pointed out potential differences in IMP variant coverage between the GenePOC™ Carba assay and the Reference Method. Investigation suggested an IMP crosscontamination at the step of isolate preparation in four (4) out of 21 isolates for which discrepant testing did not produce a sequence match with the IMP target.

5 Investigation suggested a VIM cross-contamination at the step of sample preparation. Discrepant testing did not produce a sequence match with the VIM target but produced a sequence match for the KPC target.

6 Discrepant testing did produce a sequence match with the NDM-1 target in one (1) isolates and produced a sequence match for the OXA target in one (1) out of two (2) isolates. The OXA-48 positive isolate was classified as FP.

{31}------------------------------------------------

Performance of GenePOC™ Carba Assay by Organism Category and by Target Gene, on Isolated Colonies Grown on MacConkey Agar Relative to Reference Method

N: Number; TP: True Positive; FP: False Positive; TN: False Negative; CI: Confidence Interval Multiple target results were observed for some isolates.

4. Clinical Cut-off

Not Applicable.

{32}------------------------------------------------

5. Expected Values/Reference Range

A total of 512 isolates of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa covering a large geographical and temporal diversity and reported to be carbapenem-non-susceptible based on conventional phenotypic (AST) methods have been tested with GenePOC™ Carba assay. After their growth on blood agar, 449 isolates were determined to have one (1) or more of the blaKPC, blaNDM, blaOXA-48-like, and blaIMP gene targets by the GenePOC™ Carba assay. Results obtained from isolates grown on MacConkey agar plates were similar, with 449/512 carbapenem non-susceptible isolates positive for blakec, blandM, blayIM, blaoxA-48-like, or blamp gene targets.

O. INSTRUMENT NAME

revogene™M

P. SYSTEM DESCRIPTIONS

    1. Modes of Operation:
      Real-time PCR with fluorogenic detection of amplified DNA.
    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes X_ or No _________________________________________________________________________________________________________________________________________________________

    1. Specimen Identification:
      Barcodes are used to identify patient specimens. The GenePOC™ Carba assay's Sample Buffer Tube (SBT) and microfluidic cartridge (PIE) are both pre-labeled with a unique barcode to identify both the specimen and assay. The instrument has two barcode readers to identify reagents and patient specimens. It provides traceability of the sample ID to the PIE ID, SBT ID, and assay ID.

4. Specimen Sampling and Handling:

User intervention is required for preparing the standardized 0.5 McFarland (McF) bacterial suspension from characterized carbapenem-non-susceptible isolated colonies, inoculating the bacterial suspension into the Sample Buffer Tube (SBT), transferring the sample into the microfluidic cartridge (PIE), and loading/unloading the microfluidic cartridge into the revogene. All further sample handling is automated.

{33}------------------------------------------------

5. Calibration:

The system is factory calibrated by the manufacturer. The calibration is verified annually. Upon the verification, maintenance is performed if required.

6. Quality Control:

An Internal Process Control (PrC) is provided in each microfluidic cartridge (PIE) of the GenePOC™ Carba assay. The PrC is lysed, amplified, and detected along with each sample tested and verifies the efficacy of the DNA extraction and PCR amplification processes.

Commercially available strains (NCTC 13476 for blamp-1, CCUG 59348 for blaxe-2, ATCC® BAA-2146™ for blandM-1, ATCC® BAA-2523™ for bla0XA-48 and NCTC 13440 for blaym-1) can be used as a Positive External Control. A carbapenem non-susceptible strain of Enterobacteriaceae, A. baumannii or P. aeruginosa that does not carry any of the resistance genes targeted by the GenePOC™ Carba assay can be used as a Negative External Control.

Q. OTHER SUPPORTIVE INSTRUMENT PERFORMANCE CHARACTERISTICS DATA NOT COVERED IN THE "PERFORMANCE CHARACTERISTICS" SECTION

Not applicable.

R. PROPOSED LABELING

The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

S. CONCLUSION

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).