K Number
K183687
Date Cleared
2019-09-25

(271 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LifeWindow™ LW8 Lite is a multi-parameter physiological patient monitor indicated for continuous measurement of the following parameters:

  • · ECG waveforms from 3 or 5 lead measurements
  • · Heart Rate (from ECG)

· Non-invasive blood pressure (NIBP) automatically measures systolic and diastolic pressure and pulse rate, while calculating mean arterial pressure (MAP). The equation used to calculate MAP provides an approximate value

  • · End-tidal CO2 concentration (EtCO2) Mainstream or Sidestream
  • · Inspired CO2 concentration (insCO2) Mainstream or Sidestream
  • · Fractional Inspired Oxygen (FiO2)
  • · Acoustic Respiration Rate (RRa)
  • · Respiration Rate (from bioimpedance or EtCO2 measurement, or RRa)
  • Two Temperature channels
  • · Two Invasive Pressure channels (IP1, IP2)
  • · Pulse Rate in BPM (from Pulse CO-Oximetry and/or NIBP)

· Pulse CO-Oximetry and pulse rate - the LifeWindow LW8 Lite Pulse CO-Oximeter functions and accessories incorporating Masimo technology are indicated for the continuous nomitoring of the following parameters during both no motion and motion conditions, and for patients who are well or poorly perfused:

  • o Functional oxygen saturation of arterial hemoglobin (SpO2)
  • o Carboxyhemoglobin saturation (SpCO)
  • o Methemoglobin saturation (SpMet)
  • o Total hemoglobin concentration (SpHb)

The defined patient population is adult through neonatal patients with appropriate accessories. The monitor displays a range of physiological waveforms and indices and provides alarms.

The LifeWindow™ LW8 Lite multi-parameter patient monitor may be used in hospital-type facilities, surgery centers, and emergency facilities. For professional use only. It is not intended for helicopter transport. hospital ambulance, or home use.

Device Description

The LifeWindow™ LW8 Lite Multi-Parameter Monitor is designed for the measurement and monitoring of multiple physiological parameters simultaneously in a single patient.

LifeWindow™ LW8 Lite Multi-Parameter Monitor is constructed with the following features:

  • I Compact monitor design in a rugged aluminum chassis
  • Integration of cleared vital signs modules
  • 트 Patients from neonates through adults
  • I High Definition 8.4" TFT screen with touch screen operation
  • 트 Multi-channel vital signs trace display with variable sweep speeds
  • I Hardcopy printing via an integrated strip chart or compatible printer
  • 트 Waveforms sweep speeds from 6.25mm/s up to 50mm/s
  • . Mainstream and sidestream Capnography options
  • I Specification of cleared third-party patient-applied accessories
AI/ML Overview

This document describes the 510(k) premarket notification for the LifeWindow™ LW8 Lite multi-parameter patient monitor. The submission primarily relies on demonstrating substantial equivalence to predicate devices through non-clinical testing against recognized standards and guidance documents. No clinical studies were submitted.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are derived from recognized national and international standards and guidance documents regarding medical device safety, electromagnetic compatibility, usability, alarms, and specific physiological parameter monitoring. The reported device performance indicates compliance with these standards.

Parameter/CharacteristicAcceptance Criteria (from Standards/Guidance)Reported Device Performance (Compliance)
General Safety & Performance
Electrical SafetyIEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) or IEC 60601-1: 2012Met acceptance criteria for compliance to the standards.
Electromagnetic CompatibilityIEC 60601-1-2:2007, 3rd Edition; AIM Standard 7351731:2016, Rev. 1Met acceptance criteria for compliance to the standards.
Usability/Human FactorsIEC 60601-1-6:2010; IEC 62366:2007; Applying Human Factors and Usability Engineering To Medical Devices - GuidanceUsability evaluation conducted, device found safe and effective for intended users, uses, and environments.
Medical AlarmsIEC 60601-1-8:2006+A1:2012Met acceptance criteria for compliance to the standards.
Risk ManagementISO 14971Met acceptance criteria for residual risks.
Software LifecycleIEC 62304:2006 (First Edition) + A1:2015; IEC 62304 with software safety class C; FDA Content of Premarket Submissions for Management of CybersecuritySoftware verified to requirements and validated to meet specified intended use(s).
Specific Physiological Parameters
ECG MonitorsIEC 60601-2-27:2011, 3rd Edition; Cardiac Monitor Guidance (including Cardiotachometer and Rate Alarm)Met acceptance criteria for compliance to the standards and guidance.
NIBP (Automated)IEC 80601-2-30:2013, Edition 1.1; FDA Non-invasive Blood Pressure (NIBP) Guidance DocumentMet acceptance criteria for compliance to the standards and guidance.
Invasive Blood PressureIEC 60601-2-34:2011, 3rd EditionMet acceptance criteria for compliance to the standards.
Multifunction Patient MonitorsIEC 60601-2-49:2011Met acceptance criteria for compliance to the standards.
Respiratory Gas MonitorsISO 80601-2-55:2011, 1st EditionMet acceptance criteria for compliance to the standards.
Clinical ThermometersISO 80601-2-56:2009; Guidance on the Content of Premarket Notification [510(k)] Submissions for Clinical Electronic ThermometersMet acceptance criteria for compliance to the standards and guidance.
Pulse OximeterISO 80601-2-61:2011, 1st Edition; Pulse Oximeters Premarket Notification Submissions [510(k)s] GuidanceMet acceptance criteria for compliance to the standards and guidance.
Specific Performance Values (Comparison to Predicate)
ECG Heart Rate Measurement Range10 - 300 BPM (Adult); 10 - 350 BPM (Pedi/Neo)Same
ECG Heart Rate Resolution1 BPMSame
ECG Heart Rate Accuracy± 1% or 1 BPM, whichever is greaterSame
Resp Rate (Impedance) Range0 to 150 BPMSame
Resp Rate (Impedance) Resolution1 BPMSame
Resp Rate (Impedance) Accuracy±2 BPM or ±2%, whichever greaterSame
NIBP Systolic RangeAdult: 40 - 260; Pediatric: 40 - 230; Neonatal: 40 - 130Same
NIBP Diastolic RangeAdult: 20 – 200; Pediatric: 20 - 160; Neonatal: 20 - 100Same
NIBP MAP RangeAdult: 26 – 220; Pediatric: 26 - 183; Neonatal: 26 - 110Same
NIBP Pressure AccuracyMeets accuracy requirements of ANSI/AAMI SP10:2002(R)2008, EN1060-4:2004 and ISO 81060-2:2009 (Clinical Accuracy)Same
NIBP Pressure Transducer Acc.±3mmHg between 0 mmHg and 300 mmHgSame
NIBP PR Range30 - 220 BPMSame
NIBP Pulse Rate Accuracy2% or 3 bpm, whichever is greaterSame
SpO2 Measurement Range0 - 100%Same
SpO2 Resolution1%Same
SpO2 Accuracy (No Motion)Adults, Pediatric, Infants: 60–80 ± 3%, 70–100 ± 2%; Neonates: 70–100 ± 3%Same
SpO2 Accuracy (Motion)70–100 ± 3%Same
SpO2 Accuracy (Low Perfusion)70–100 ± 2%Same
Pulse Rate (from SpO2) Range25–240 BPMSame
Pulse Rate (from SpO2) acc. (No motion)± 3 BPMSame
Pulse Rate (from SpO2) acc. (Motion)± 5 BPMSame
Pulse Rate (from SpO2) acc. (Low Perf.)± 3 BPMSame
Pulse Rate (from SpO2) Resolution1 BPMSame
SpCO Measurement Range0–99%Same
SpCO Accuracy (No Motion)Adults, Pediatric, Infants: 1 - 40 ± 3%Same
SpMet Measurement Range0 - 99.9%Same
SpMet AccuracyAdults, Pediatric, Infants, Neonates: 1 - 15 ± 1%Same
Perfusion Index (PI) Range0.02 – 20%Same
Perfusion Index (PI) Resolution0.01 (PI < 1.00); 0.1 (PI < 10); 1 (PI > 10)Same
SpHb Measurement Range0 - 25 g/dLSame
SpHb AccuracyAdults, Pediatric: 8 - 17 ± 1 g/dLSame
SpHb Resolution0.1 g/dLSame
SpHb Alarm Range1.0 - 24.5 g/dLSame
SpOC Measurement Range0 - 35 ml/dLSame
SpOC Resolution1 ml/dLSame
PVI Measurement Range0 - 100%Same as predicate Radical-7, not implemented on primary predicate A3.
PVI Resolution1 %Same as predicate Radical-7, not implemented on primary predicate A3.
RRa (Acoustic) Range4 - 70 bpmSame as predicate Radical-7, not implemented on primary predicate A3.
RRa (Acoustic) AccuracyAdults/Pediatric: 4 - 70 ± 1 BPMSame as predicate Radical-7, not implemented on primary predicate A3.
RRa (Acoustic) Resolution1 BPMSame as predicate Radical-7, not implemented on primary predicate A3.
Temperature AccuracyRated Range: 35.0 - 42.0°C (95.0 - 107.6°F): ± 0.3 °C; Extended Range: -1.0 - 34.9°C (30.2 – 94.8°F) and 42.1 – 45.0°C: ± 0.4 °CEquivalent Function, LifeWindow complies with current industry standard.
Temperature Resolution0.1°C (0.2°F)Same
IBP Pressure Measurement Range-40 to +300 mmHgEquivalent Function with slightly different range - claims of the LW8 are within the claims of the predicate.
IBP Pressure Measurement Acc.± 4 mmHg or ±4% of the reading, whichever is greaterSame
IBP Resolution1 mmHgSame
EtCO2 (Mainstream) RangePercent: 0 – 19.7; mmHg: 0 – 150; kPa: 0 – 20Same
EtCO2 (Mainstream) Resolution1 mmHgSame
EtCO2 (Mainstream) Accuracy0 - 40 mmHg ± 2 mmHg; 41 - 70 mmHg ± 5%; 71 - 100 mmHg ± 8%; 101 - 150 mmHg ± 10%Same
awRR (Mainstream) Range0 - 150 BPMSame
awRR (Mainstream) Accuracy± 1 BPMSame
InsCO2 (Mainstream) Range3 to 50 mmHgLifeWindow specification is per cleared CO2 module.
InsCO2 (Mainstream) Resolution1 mmHgEquivalent Function not offering decimal point resolution (predicate offers 0.1 mmHg)
InsCO2 (Mainstream) Accuracy3 - 40 mmHg ± 2 mmHg; 41 - 50 mmHg ± 5%Same
EtCO2 (Sidestream) Range0 - 19.7%; 0 – 150 mmHg; 0 – 20 kPaSame
EtCO2 (Sidestream) Resolution1 mmHgSame
EtCO2 (Sidestream) Accuracy0 - 40 mmHg ± 2 mmHg; 41 - 70 mmHg ± 5%; 71 - 100 mmHg ± 8%; 101 - 150 mmHg ± 10% (when RR > 80 rpm, all range is ±12%)Same
awRR (Sidestream) Range2 - 150 BPMSame
awRR (Sidestream) Accuracy± 1 BPMSame
InsCO2 (Sidestream) Range3 to 50 mmHgSame
InsCO2 (Sidestream) Accuracy3 - 40 mmHg ± 2 mmHg; 41 - 50 mmHg ± 5%Same
InsCO2 (Sidestream) Resolution1 mmHgEquivalent Function with no decimal point resolution (predicate offers 0.1 mmHg)
FiO2 Measurement Range0-100% OxygenSame
FiO2 Display Precision1% (Resolution)Equivalent Function with no decimal point resolution (predicate offers 0.1%)
FiO2 Accuracy & Repeatability<1% vol. O2 if calibrated with 100% vol. O2Same

2. Sample Size Used for the Test Set and the Data Provenance

The submission indicates reliance on non-clinical bench testing and simulations, not clinical studies with patient data. Therefore, there is no "test set" in the context of patient data, nor is there information on data provenance (country of origin, retrospective/prospective).

  • For the Usability / Human Factors summative study, there were 15 representative professional users (USA) in a simulated critical care scenario with vital signs generation. This is a simulation study, not a clinical data study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. As described above, the studies conducted were non-clinical bench tests and a usability study. There were no "ground truth" experts in the traditional sense of medical image or diagnosis interpretation for patient data. For the usability study, "representative professional users" were involved, but their qualifications beyond being "professional users" are not further detailed.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring expert adjudication was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No MRMC comparative effectiveness study was done. The device is a multi-parameter patient monitor, not an AI-assisted diagnostic tool for interpretation of medical cases by human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a standalone monitor. The non-clinical tests assess the performance of the device's various physiological measurement algorithms against established standards and predicate device specifications. These tests essentially evaluate the "standalone" performance of the embedded algorithms for each parameter.

7. The Type of Ground Truth Used

For the non-clinical tests, the "ground truth" was established by:

  • Reference standards and test equipment: Compliance with international and national standards (e.g., IEC 60601 series, ISO 80601 series, ANSI/AAMI SP10) typically involves testing the device against calibrated reference instruments and simulated physiological signals that represent known "ground truth" values.
  • Predicate device specifications: Performance characteristics were compared directly to the specifications of the identified predicate devices, which serve as a reference for equivalence.

8. The Sample Size for the Training Set

Not applicable. This device is a physiological patient monitor and does not describe the use of machine learning algorithms that require a "training set" in the context of large datasets for model development. The software development lifecycle adheres to IEC 62304 standards for medical device software.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a training set for machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 25, 2019

Digicare Biomedical Technology, Inc. Pedro Miranda VP of Software Engineering 107 Commerce Rd. Boynton Beach, Florida 33426

Re: K183687

Trade/Device Name: LifeWindow LW8 Lite Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DRT, DSB, BZQ, DXN, DSK, FLL, DQA, DPZ, CCK, CCL Dated: August 21, 2019 Received: August 23, 2019

Dear Pedro Miranda:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for LT Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183687

Device Name

Digicare Biomedical LifeWindow™ LW8 Lite multi-parameter patient monitor and accessories

Indications for Use (Describe)

The LifeWindow™ LW8 Lite is a multi-parameter physiological patient monitor indicated for continuous measurement of the following parameters:

  • · ECG waveforms from 3 or 5 lead measurements
  • · Heart Rate (from ECG)

· Non-invasive blood pressure (NIBP) automatically measures systolic and diastolic pressure and pulse rate, while calculating mean arterial pressure (MAP). The equation used to calculate MAP provides an approximate value

  • · End-tidal CO2 concentration (EtCO2) Mainstream or Sidestream
  • · Inspired CO2 concentration (insCO2) Mainstream or Sidestream
  • · Fractional Inspired Oxygen (FiO2)
  • · Acoustic Respiration Rate (RRa)
  • · Respiration Rate (from bioimpedance or EtCO2 measurement, or RRa)
  • Two Temperature channels
  • · Two Invasive Pressure channels (IP1, IP2)
  • · Pulse Rate in BPM (from Pulse CO-Oximetry and/or NIBP)

· Pulse CO-Oximetry and pulse rate - the LifeWindow LW8 Lite Pulse CO-Oximeter functions and accessories incorporating Masimo technology are indicated for the continuous nomitoring of the following parameters during both no motion and motion conditions, and for patients who are well or poorly perfused:

  • o Functional oxygen saturation of arterial hemoglobin (SpO2)
  • o Carboxyhemoglobin saturation (SpCO)
  • o Methemoglobin saturation (SpMet)
  • o Total hemoglobin concentration (SpHb)

The defined patient population is adult through neonatal patients with appropriate accessories. The monitor displays a range of physiological waveforms and indices and provides alarms.

The LifeWindow™ LW8 Lite multi-parameter patient monitor may be used in hospital-type facilities, surgery centers, and emergency facilities. For professional use only. It is not intended for helicopter transport. hospital ambulance, or home use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary in accordance with 21 CFR 807.92

(a)(1)Submitted by:Digicare Biomedical Technology, Inc.107 Commerce Rd.Boynton BeachFlorida 33426Tel.: +1 (561) 689-0408Fax: +1 (561) 689-0021pedro(at)digicarebiomedical.com
Contact Person:Mr. Pedro Miranda
Position/Title:Vice President of Software Engineering
Date of Preparation:September 19, 2019
(2)Trade Name:LifeWindow™ LW8 Lite Multi-ParameterMonitor

Common/Classification Name: Monitor, Physiological, Patient

Product Code(s):

Regulation No.Classification NameProductCodeDeviceClass
Main Product Code
21 CFR §870.2300Monitor, Physiological, Patient (Without ArrhythmiaDetection or Alarms)MWIClass II
Subsequent Product Code(s)
21 CFR § 870.2300Cardiac monitor (including cardiotachometer andrate alarm) Cardiac monitorDRTClass II
21 CFR § 870.2770Impedance plethysmographDSBClass II
21 CFR § 868.2375Breathing frequency monitorBZQClass II
21 CFR § 870.1130Non-invasive blood pressure measurement systemDXNClass II
21 CFR § 870.1110Blood pressure computerDSKClass II
21 CFR § 880.2910Clinical Electronic ThermometerFLLClass II
21 CFR § 870.2700Pulse OximeterDQAClass II
21 CFR § 870.2710Ear OximeterDPZClass II
21 CFR § 868.1400Carbon Dioxide Gas AnalyzerCCKClass II
21 CFR § 868.1720Oxygen Gas AnalyzerCCLClass II

{4}------------------------------------------------

(3) Predicate Device(s): Substantial Equivalence to:

K NumberModelManufacturer
K131898Biolight A Series and Q Series Patient MonitorsGuangdong BiolightMeditech Co., Ltd.
K122290EnviteC MySign® O OxygenMeasuring DeviceEnviteC-Wismar GmbH
K110028Radical-7 Pulse CO-OximeterMasimo Corporation

(4) Description of Device:

The LifeWindow™ LW8 Lite Multi-Parameter Monitor is designed for the measurement and monitoring of multiple physiological parameters simultaneously in a single patient.

LifeWindow™ LW8 Lite Multi-Parameter Monitor is constructed with the following features:

  • I Compact monitor design in a rugged aluminum chassis
  • Integration of cleared vital signs modules
  • 트 Patients from neonates through adults
  • I High Definition 8.4" TFT screen with touch screen operation
  • 트 Multi-channel vital signs trace display with variable sweep speeds
  • I Hardcopy printing via an integrated strip chart or compatible printer
  • 트 Waveforms sweep speeds from 6.25mm/s up to 50mm/s
  • . Mainstream and sidestream Capnography options
  • I Specification of cleared third-party patient-applied accessories

(5) Intended use:

Multi-parameter monitoring of patient vital signs has been a standard of care for over 25 years and is established globally. Multi-parameter patient monitors provide clinicians with a range of continuous physiological measurements with alarms.

The intended use for the LifeWindow™ LW8 Lite Multi-Parameter Monitor is the same as the predicate device: measurement and monitoring of multiple physiological parameters, in an adult through neonatal patient population, with appropriate patient applied accessories. The LifeWindow™ LW8 Lite Monitor is specified for use in professional healthcare environments by clinical staff.

Indications for Use:

{5}------------------------------------------------

The LifeWindow™ LW8 Lite is a multi-parameter physiological patient monitor indicated for continuous measurement of the following parameters:

  • I ECG waveforms from 3 or 5 lead measurements
  • Heart Rate (from ECG)
  • l Non-invasive blood pressure (NIBP) automatically measures systolic and diastolic pressure and pulse rate. while calculating mean arterial pressure (MAP). The equation used to calculate MAP provides an approximate value.
  • End-tidal CO2 concentration (EtCO2) - Mainstream or Sidestream
  • 트 Inspired CO2 concentration (insCO2) - Mainstream or Sidestream
  • I Fractional Inspired Oxygen (FiO2)
  • I Acoustic Respiration Rate (RRa)
  • 트 Respiration Rate (from bioimpedance or EtCO2 measurement, or RRa)
  • Two Temperature channels
  • I Two Invasive Pressure channels (IP1, IP2)
  • 트 Pulse Rate in BPM (from Pulse CO-Oximetry and/or NIBP)
  • I Pulse CO-Oximetry and pulse rate - the LifeWindow LW8 Lite Pulse CO-Oximeter functions and accessories incorporating Masimo technology are indicated for the continuous noninvasive monitoring of the following parameters during both no motion and motion conditions, and for patients who are well or poorly perfused:
    • Functional oxygen saturation of arterial hemoglobin (SpO2) o
    • Carboxyhemoglobin saturation (SpCO) o
    • Methemoglobin saturation (SpMet) o
    • Total hemoglobin concentration (SpHb) O

The defined patient population is adult through neonatal patients with appropriate accessories. The monitor displays a range of physiological waveforms and indices and provides alarms.

The LifeWindow LW8 Lite multi-parameter patient monitor may be used in hospitals, hospital-type facilities, surgery centers, and emergency facilities. For professional use only. It is not intended for helicopter transport, hospital ambulance, or home use.

Prescription device.

(6) Technological Characteristics:

LifeWindow™ LW8 Lite Multi-Parameter Monitor utilize the same technological principles as the predicate devices to measure patient vital signs: the incorporation of physiological measurement modules, with appropriate accessories. Refer to the following comparison table:

{6}------------------------------------------------

Page 4 of 24

Comparison of Technological Features to Predicate Devices:

Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison AnalysisIndications for use:The LifeWindow™ LW8Lite is a multi-parameterphysiological patientmonitor indicated forcontinuous measurementof the followingparameters:• ECG waveforms from 3or 5 lead measurements• Heart Rate (from ECG)• Non-invasive bloodpressure (NIBP)automatically measuressystolic and diastolicpressure and pulse rate,while calculating meanarterial pressure (MAP).The equation used tocalculate MAP providesan approximate value.• End-tidal CO2concentration (EtCO2) –Mainstream orSidestream• Inspired CO2concentration (insCO2) –Mainstream orSidestream• Fractional InspiredOxygen (FiO2)• Acoustic RespirationRate (RRa)• Respiration Rate (frombioimpedance or EtCO2measurement, or RRa)• Two Temperaturechannels• Two Invasive Pressurechannels (IP1, IP2)• Pulse Rate in BPM(from Pulse CO-Oximetry and/or NIBP)Patient monitors areintended to be used formonitoring, displaying,reviewing, storing andalarming of multiplephysiological parametersincluding ECG (3-lead or5-lead or 12-leadselectable), arrhythmiadetection, ST segmentanalysis, Heart Rate(HR), Respiration Rate(RESP), Temperature(TEMP), Pulse OxygenSaturation (SpO2), PulseRate (PR), Non-invasiveBlood Pressure (NIBP),Invasive Blood Pressure(IBP), Carbon dioxide(C02), Anesthetic Gas(AG), ImpedanceCardiograph (ICG),Cerebral State Index(CSI), Bispectral Index(HIS), TotalHemoglobin(SpHb),Carboxyhemoglobin(SpCO), andMethemoglobin(SpMet).The arrhythmiadetection, ST segmentanalysis only applied to asingle adult patient.The monitors are to beused in healthcarefacilities by clinicalphysicians or appropriatemedical staff under thedirection of physician.It is not intended forhelicopter transport,This oxygen measuringdevice MySign® O isdesigned for continuousor spot monitoring ofinspired oxygenconcentrations inbreathing gas.MySign® O can beused for monitoring thebreathing gasesdispensed by thefollowing devices:Anaesthesia breathingsystemsRespiratory equipmentInfant incubatorsOxygen therapysystemsThe system is suitablefor use inside hospitalsas well as duringtransport (except byair), emergencies, andartificial respirationprovided at home.The Masimo Radical 7Pulse CO-Oximeter andaccessories areindicated for thecontinuous non-invasivemonitoring of functionaloxygen saturation ofarterial hemoglobin(SPO2), pulse rate,carboxyhemoglobinsaturation (SpCO),methemoglobinsaturation (SpMet), totalhemoglobinconcentration (SpHb),and/or respiratory rate(RRa). The MasimoRadical 7 Pulse CO-Oximeter andaccessories areindicated for use withadult, pediatric, andneonatal patients duringboth no motion andmotion conditions, andfor patients who are -well or poorly perfusedin hospitals, hospital-type facilities, mobile,and homeenvironments. Inaddition, the MasimoRadical 7 Pulse CO-Oximeter andaccessories areindicated to provide thecontinuous non-invasivemonitoring dataobtained from theMasimo Radical 7 PulseCO-Oximeter endaccessories offunctional oxygensaturation of arterialSee above- equivalentmonitoring function forspecified parametersand sites
ManufacturerDigicare BiomedicalTechnology, Inc.Guangdong BiolightMeditech Co., Ltd.EnviteC-Wismar GmbHby HoneywellMasimo Corporation
Model Number(s)LW8A3 SeriesMySign® ORadical-7
510(k) Number(pending - thissubmission)K131898K122290K110028
Application/Intended use:Monitor multiplephysiologicalparametersMonitor multiplephysiologicalparametersMonitor fractionalinspired Oxygen(FiO2)Monitor pulse co-oximetry and acousticrespiration rate (RRa)Equivalent monitoringfunction for specifiedparameters and sites

{7}------------------------------------------------

Page 5 of 24

{8}------------------------------------------------

Page 6 of 24

• Pulse CO-Oximetry andhospital ambulance, orhemoglobin (SpO2) and
pulse rate - thehome use.pulse rate to multi-
LifeWindow LW8 Liteparameter devices for
Pulse CO-Oximeterthe display of those
functions anddevices.
accessories
incorporating Masimo
technology are indicated
for the continuous
noninvasive monitoring
of the following
parameters during both
no motion and motion
conditions, and for
patients who are well or
poorly perfused:
o Functional oxygen
saturation of arterial
hemoglobin (SpO2)
o Carboxyhemoglobin
saturation (SpCO)
o Methemoglobin
saturation (SpMet)
o Total hemoglobin
concentration (SpHb)
The defined patient
population is adult
through neonatal patients
with appropriate
accessories. The monitor
displays a range of
physiological waveforms
and indices and provides
alarms.
The LifeWindow LW8 Lite
multi-parameter patient
monitor may be used in
hospitals, hospital-type
facilities, surgery centers,
and emergency facilities.
For professional use only.
lt is not intended for
helicopter transport,

{9}------------------------------------------------

Page 7 of 24

Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
hospital ambulance, orhome use.
AlarmsYESYESYESYESSame
Application Site(s)Patient sites vary byparameter measuredPatient sites vary byparameter measuredInspired airway pathPulse co-oximetrysite: digit, toe, ear,foot, forehead;acoustic respirationEquivalent monitoringfunction at specifiedsites
Patient PopulationAdult through neonatalpatientsAdult through neonatalpatients; (adults only forarrhythmia and STsegment)Adult through neonatalpatientsAdult through neonatalpatientsEquivalent monitoringfunction, LW8 doesnot claim Arrhythmiaand ST segment.
Cardiac Monitor (ECG)3/5 Lead3/5/12 Lead----Equivalent functionClaims of the LW8 arewithin the claims of thepredicate.
Detector and Alarm,Arrhythmia--Implemented/offered----Arrhythmia detectionnot claimed by subjectdevice - claims of theLW8 are within theclaims of the predicate
Monitor, ST Segment withAlarm--Implemented/offered----ST Segmentmonitoring not claimedby subject device -claims of the LW8 arewithin the claims of thepredicate
Impedance respirationImplemented with ECGImplemented w/ ECG----Equivalent functionwith same technology

{10}------------------------------------------------

of248Page
-----------------------
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Acoustic respirationMasimo (RRa)----Masimo (RRa)LW8 Acoustic respirationrate utilizes the MasimoRRa which is the sametechnology utilized inpredicate Radical-7
Invasive Blood pressureYESYES----Equivalent functionwith samemeasurement principle
Non-invasive bloodpressureOscillometricOscillometric----Equivalent functionwith same technology
Patient TemperatureYSI 400YSI 400----Equivalent functionwith samemeasurement principle
Pulse co-oximetryMasimo technologyMasimo technology--Masimo technologyEquivalent functionwith same technology
Carbon Dioxide GasAnalyzerMainstream andsidestreamMainstream andsidestream----Equivalent functionwith same technology
Inspired Oxygen (FiO2)EnviteC technologyParamagneticEnviteC technology--Same as the MySignO predicate device.Primary predicate A3offers a different typeof technology formeasuring FiO2.
Enflurane gas analyzer--Implemented/offered----LW8 does not
Halothane gas analyzer--Implemented/offered----measure nor claimanesthetic agents.

{11}------------------------------------------------

Page 9 of 24
--------------------
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Nitrous Oxide gasanalyzer--Implemented/offered----therefore there are noclaims related to thesegas measurements -claims of the LW8 arewithin the claims of thepredicate
Cerebral State Index(EEG)--Implemented/offered----LW8 does not measurenor claim cerebral stateindex (CSI) by EEGmethod, are no claimsrelated to this parametermeasurements - claimsof the LW8 are within theclaims of the predicate
Environment of UseMay be used in hospitals,hospital-type facilities,surgery centers, andemergency facilities. Forprofessional use only. Notintended for helicoptertransport, hospitalambulance, or home use.(Professional healthcareenvironments). notintended for helicoptertransport, hospitalambulance, or home useFor use inside hospitalsas well as duringtransport (except byair), emergencies, andartificial respirationprovided at homeFor use inside hospitals,hospital-type facilities,mobile, and homeenvironmentsBoth LW8 and primarypredicate A3 monitorsare specified forprofessional healthcarefacilities, and nottransport or home.MySign® O and Radical-7 use includes home andtransport environments.
Scenario of UsePatient connected monitorfor continuous monitoringof vital signs with alarmsPatient connectedmonitor for continuousmonitoring of vital signswith alarmsPortable device forcontinuous monitoringof inspired Oxygen withalarmsPortable device forcontinuous monitoringof inspired Pulse-COOximetry and BreathingFrequency (RRa) withalarmsSame context of use ascontinuous alarmingmonitor
Moisture Protection ClassIP21IPX1IP54IPX1Equivalent to PrimaryPredicate A3. The LW8provides higher IPclassification thanprimary predicate.
Classification according tothe IEC60601-1
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Electrical Protection Class -MainframeIIIIISame electricalprotection class asprimary predicate
Electrical ProtectionCF Defibrillation proofCF Defibrillation proofBFBFSame protection class forLW8 and primarypredicate A3 monitor
Mode of OperationContinuousContinuousContinuousContinuousSame
ECG Heart Rate
Measurement range10 - 300 BPM (Adult)10 - 350 BPM (Pedi/Neo)10 - 300 BPM (Adult)10 - 350 BPM(Pedi/Neo)----Same
Resolution1 BPM1 BPM----Same
Accuracy± 1% or 1 BPM,whichever is greater± 1% or 1 BPM,whichever is greater----Same
Resp Rate
Measurement Range0 to 150 BPM0 to 150 BPM----Same
Resolution1 BPM1 BPM----Same
Accuracy±2 BPM or ±2%,whichever greater±2 BPM or ±2%,whichever is the greater----Same
NIBP (Suntech)
Systolic Measurement range(mmHg)Adult: 40 - 260 Pediatric:40 - 230Neonatal: 40 - 130Adult: 40 - 260Pediatric: 40 - 230Neonatal: 40 - 130----Same
Diastolic Measurementrange (mmHg)Adult: 20 – 200Pediatric: 20 - 160Neonatal: 20 - 100Adult: 20 – 200Pediatric: 20 - 160Neonatal: 20 - 100----Same

{12}------------------------------------------------

Page 10 of 24

{13}------------------------------------------------

of241Page1
----------------------------
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
MAP Measurement range(mmHg)Adult: 26 – 220Pediatric: 26 - 183Neonatal: 26 - 110Adult: 26 – 220Pediatric: 26 - 183Neonatal: 26 - 110----Same
Pressure MeasurementAccuracyClinical Accuracy:Meets accuracyrequirements ofANSI/AAMISP10:2002(R)2008,EN1060-4:2004 and ISO81060-2:2009± 3 mmHg (± 0.4 kPa) or2 % of the reading above200 mmHg for theambient temperaturerange of 50°F to 104°F(10℃ to 40℃), thehumidity range of 15 %to 90 %----Same
Pressure transduceraccuracy±3mmHg between 0mmHg and 300 mmHg foroperating conditionsbetween 0°C and 50°C.±3mmHg between 0mmHg and 300 mmHgfor operating conditionsbetween 0°C and 50°C.----Same
PR range30 - 220 BPM30 - 220 BPM----Same
Pulse Rate Accuracy2% or 3 bpm, whicheveris greater2% or 3 bpm, whicheveris greater----Same
SpO2 (Masimo)Pulse CO-Oximetry
SpO2
Measurement range0 - 100%0 - 100%--0 – 100%Same
Resolution1%1%--1%Same
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
AccuracyNo Motion Adults,Pediatric, Infants:60–80 ± 3%70–100 ± 2%No Motion Neonates:70–100 ± 3%Motion:70–100 ± 3%Low Perfusion:70–100 ± 2%0% to 69%, unspecifiedNo Motion Adults,Pediatric, Infants:60–80 ± 3%70–100 ± 2%No Motion Neonates:70–100 ± 3%Motion:70–100 ± 3%Low Perfusion:70–100 ± 2%0% to 69%, unspecified--No Motion Adults,Pediatric, Infants:60–80 ± 3%70–100 ± 2%No Motion Neonates:70–100 ± 3%Motion:70–100 ± 3%Low Perfusion:70–100 ± 2%0% to 69%,unspecifiedSame
Measurement Range25–240 BPM25–240 BPM--25–240 BPMSame
AccuracyNo Motion: ± 3 BPMMotion: ± 5 BPMLow Perfusion: ± 3 BPMNo Motion: ± 3 BPMMotion: ± 5 BPMLow Perfusion: ± 3 BPM--No Motion: ± 3 BPMMotion: ± 5 BPMLow Perfusion: ± 3 BPMSame
Resolution1 BPM1 BPM--1 BPMSame
Measurement range0–99%0–100%--0–99%Same
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
AccuracyNo Motion:Adults, Pediatric, Infants:$1 - 40 \pm 3 %$>40%, unspecifiedNo Motion:Adults, Pediatric, Infants:$1 - 40 \pm 3 %$>40%, unspecified--No Motion:Adults, Pediatric,Infants:$1 - 40 \pm 3 %$>40%, unspecifiedSame
SpMet--
Measurement range0 - 99.9%0 – 100%--0 - 99.9%Same
AccuracyAdults, Pediatric, Infants,Neonates:$1 - 15 \pm 1 %$>15%, unspecified0% to 15%: ±1%(non-motion conditions)>15%,unspecified--Adults, Pediatric,Infants, Neonates:$1 - 15 \pm 1 %$>15%, unspecifiedSame
Perfusion Index (PI)--
Measurement Range0.02 – 20%0.05 — 20%--0.02 – 20%Same
ResolutionPI < 1.00: 0.01PI < 10: 0.1PI > 10: 1Information Not Available--PI < 1.00: 0.01PI < 10: 0.1PI > 10: 1Same
SpHb--
Measurement range0 - 25 g/dL0 - 25 g/dL--0 – 25 g/dLSame
AccuracyAdults, Pediatric:$8 - 17 \pm 1$ g/dL8 g/dl to 17 g/dl: ±1g/dl(non-motion conditions)<8 g/dl or >17 g/dl,unspecified--Adults, Pediatric:$8 - 17 \pm 1$ g/dLSame
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Resolution0.1 g/dLInformation Not Available--0.1 g/dLSame
Alarm Range1.0 - 24.5 g/dLInformation Not Available--1.0 - 24.5 g/dLSame
SpOC--
Measurement range0 - 35 ml/dL0 - 35 ml/dL--0 - 35 ml/dLSame
Resolution1 ml/dLInformation Not Available--1 ml/dLSame
Pleth Variability Index(PVI)--
Measurement Range0 - 100%----0 - 100%Same as predicateRadical-7, notimplemented on primarypredicate A3.
Resolution1 %----1 %Same as predicateRadical-7, notimplemented on primarypredicate A3.
Respiration Rate (RRa)--
Measurement Range4 - 70 bpm----4 - 69 bpmSame as predicateRadical-7, notimplemented on primarypredicate A3.
AccuracyAdults/Pediatric:4 - 70 ± 1 BPM----Adults/Pediatric:4 - 70 ± 1 BPMSame as predicateRadical-7, notimplemented on primarypredicate A3.
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Resolution1 BPM----1 BPMSame as predicateRadical-7, notimplemented on primarypredicate A3.
Temperature--
AccuracyRated Range:35.0 - 42.0°C (95.0 -107.6°F): ± 0.3 °C (±0.5°F)Extended Range:-1.0 - 34.9°C (30.2 –94.8°F) and 42.1 – 45.0°C(107.8 - 113°F): ± 0.4 °C(± 0.7°F)Accuracy ±0.1°C or ±1°F(exclusive of probe)----Equivalent Function,LifeWindow complieswith current industrystandard.
Resolution0.1°C (0.2°F)0.1°C or 1°F----Same
IBP (Invasive BloodPressure)--
Pressure measurementrange-40 to +300 mmHg-50 mmHg to +350mmHg----Equivalent Functionwith slightly differentrange - claims of theLW8 are within theclaims of the predicate
Pressure Measurementaccuracy± 4 mmHg or ±4% of thereading, whichever isgreater.±4 mmHg or ±4% of thereading, whichever is thegreater (inclusion oftransducer)----Same

{14}------------------------------------------------

Page 12 of 24

{15}------------------------------------------------

Page 13 of 24

{16}------------------------------------------------

Page 14 of 24

{17}------------------------------------------------

Resolution

1 mmHg

Page 15 of 24

--

Same

--

(Transducer included)

1 mmHg

{18}------------------------------------------------

Page16 of 24
--------------------
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Mainstream CO2 module(CAPNOSTAT5)Respironics----------
Measurement range (EtCO2)Percent (%): 0 – 19.7mmHg: 0 – 150kPa: 0 – 200% to 19.7 %0 mmHg to 150 mmHg----Same
Resolution (EtCO2)1 mmHg0.1% or 1mmHg----Same
Accuracy (EtCO2)0 - 40 mmHg ± 2 mmHg41 - 70 mmHg ± 5% ofreading71 - 100 mmHg ± 8% ofreading101 - 150 mmHg ± 10% ofreading0 - 40 mmHg ± 2 mmHg41 - 70 mmHg ± 5% ofreading71 - 100 mmHg ± 8% ofreading101 - 150 mmHg ± 10%of reading----Same
awRR measurement range(Resp Rate)0 - 150 BPM0 rpm to 150 rpm----Same
awRR measurementaccuracy (Resp Rate)± 1 BPM±1 rpm----Same
Measurement range(InsCO2)3 to 50 mmHg3 to 50 mmHg----LifeWindow specificationis per cleared CO2module.
Resolution (InsCO2)1 mmHg0.1 mmHg 0 to 50 mmHg----Equivalent Function notoffering decimal pointresolution
Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Accuracy (InsCO2)3 - 40 mmHg ± 2 mmHg41 - 50 mmHg ± 5% ofreading3 - 40 mmHg ± 2 mmHg41 - 50 mmHg ± 5% ofreading----Same
Respironics CO2 module(LoFlo) Respironics
Measurement range (EtCO2)0 - 19.7%0 – 150 mmHg0 – 20 kPa0% to 19.7 %0 mmHg to 150 mmHg----Same
Resolution (EtCO2)1 mmHg0.1% or 1 mmHg----Same
Accuracy (EtCO2)0 - 40 mmHg ± 2 mmHg41 - 70 mmHg ± 5% ofreading71 - 100 mmHg ± 8% ofreading101 - 150 mmHg ± 10% ofreading(when RR > 80 rpm, allthe range is ±12% ofreading) Gas temperatureat 25°C0 - 40 mmHg ± 2 mmHg41 - 70 mmHg ± 5% ofreading71 - 100 mmHg ± 8% ofreading101 - 150 mmHg ± 10%of reading(when RR > 80 rpm, allthe range is ±12% ofreading) Gastemperature at 25°C----Same
awRR measurementrange (Resp Rate)2 - 150 BPM2 rpm to 150 rpm----Same
awRR measurementaccuracy (Resp Rate)± 1 BPM±1 rpm----Same

{19}------------------------------------------------

Page 17 of 24

{20}------------------------------------------------

Product/FeatureLifeWindow™ LW8 LiteMulti-Parameter MonitorA3 Series Multi-Parameter MonitorMySign® O OxygenMeasuring DeviceRadical-7 Pulse CO-OximeterComparison Analysis
Measurement Range(insCO2)3 to 50 mmHg3 to 50 mmHg----Same
Accuracy (InsCO2)3 - 40 mmHg ± 2 mmHg41 - 50 mmHg ± 5% ofreading3 - 40 mmHg ± 2 mmHg41 - 50 mmHg ± 5% ofreading----Same
Resolution (InsCO2)1 mmHg0.1 mmHg 0 to 50 mmHg----Equivalent Function withno decimal pointresolution
Resp Rate Resolution1 BPM1 BPM----Same
Inspired Fraction ofOxygen (FiO2)
Measurement Range0-100% Oxygen--0-100% Oxygen--Same
Display Precision(Resolution)1%--0.1%--Equivalent Function withno decimal pointresolution
Accuracy and Repeatability(Precision)<1% vol. O2 if calibratedwith 100% vol. O2--<1% vol. O2 if calibratedwith 100% vol. O2--Same

{21}------------------------------------------------

Summary points from the device claims comparison:

  • . The intended use for the LifeWindow™ LW8 Lite and the primary predicate device Biolight A Series Patient Monitor are the same: multi-parameter patient vital signs monitoring.
  • The LifeWindow™ LW8 Lite Monitor does not claim arrhythmia detection & ● monitoring (product code DSI). ST seament monitoring (product code MLD). nor cerebral state by EEG (product code GWQ). LifeWindow™ LW8 Lite Monitor does not provide support for the measurement of anesthetic agents for halothane product code CBS), enflurane (product code CBQ), nor nitrous Oxide (CBR). The predicate Biolight A Series Patient Monitor does claim these monitoring parameters and classifications. The LifeWindow™ LW8 Lite Monitor utilizes parameter regulatory classifications that are within those utilized by the predicate Biolight A Series Patient Monitor.
  • The LifeWindow™ LW8 Lite Monitor and predicate MySign O Oxygen ● Measuring Device are both indicated for continuous or spot monitoring of inspired oxygen concentrations in breathing gas.
  • . The LifeWindow™ LW8 Lite Monitor and predicate Masimo Radical-7 Pulse CO-Oximeter are both indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SPO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa).
  • . The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the referenced predicate devices are specified for adult through neonatal patient types (based upon appropriate patient-applied accessories utilized for each parameter measurement). Application sites are equivalent and also based upon the selection of appropriate patient-applied accessories.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and primary predicate ● device Biolight A Series Patient Monitor are specified for use in professional care environments. The EnviteC MySign O Oxygen Measuring Device is specified for hospital/professional, home, and (non-air) out transport environments. The Masimo Radical-7 Pulse CO-Oximeter is specified for professional and home environments. The claims of the LifeWindow™ LW8 Lite Multi-Parameter Monitor for professional use environments are within those of the referenced predicates.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and predicate devices . are prescription devices specified for professional use. As described above, the Masimo Radical-7 Pulse CO-Oximeter is additionally specified for home use.
  • Like the predicate monitors, the LifeWindow™ LW8 Lite Multi-Parameter . Monitor is a non-sterile device. Specific third-party patient-applied accessories such as invasive blood pressure transducers may be packaged/provided sterile by their respective manufacturers.
  • Per the indications for use LifeWindow™ LW8 Lite Multi-Parameter Monitor . claims a monitoring feature set equivalent to the Biolight A Series Patient Monitor for the following parameters:
    • o ECG waveforms and heart rate from 3 or 5 lead (Biolight A Series includes 12 lead)
    • Non-invasive blood pressure (NIBP): systolic, diastolic, and mean o arterial pressure
    • End-tidal CO2 concentration (EtCO2) Mainstream or Sidestream O

{22}------------------------------------------------

  • Inspired CO2 concentration (insCO2) Mainstream or Sidestream O
  • Fractional Inspired Oxygen (FiO2) O
  • Respiration Rate (from bioimpedance or EtCO2 measurement) O
  • Two Temperature channels O
  • Two Invasive Pressure channels (IP1, IP2)

Pulse CO-Oximetry and pulse rate as previously described above с Summary points from specification comparisons:

  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary . predicate device Biolight A Series Patient Monitor measure the same vital parameters and specify similar ranges for performance characteristics.
  • Both the LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary . predicate device Biolight A Series Patient Monitor offer the cleared Biolight electrocardiogram (ECG) measurement technology, which includes respiration rate by thoracic bioimpedance. ECG measurement principles and performance levels are equivalent, with some differences in waveform display.
  • The LifeWindow™ LW8 Lite Monitor does not implement the EGC arrhythmia ● or ST seqment functionality, the predicate device Biolight A Series Patient Monitor does. The LifeWindow™ LW8 Lite Monitor offers 3 lead and 5 lead ECG, the Biolight A Series Patient Monitor offers 3 lead, 5 lead, and 12 lead monitoring.
  • . Both the LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary predicate Biolight A Series Patient Monitor offer the cleared Masimo parameter measurement technology of multi-wavelength pulse co-oximetry, and offer similar third party Masimo sensor configurations (finger, disposable, etc.). Pulse co-oximetry measurement principles and performance levels are equivalent.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate . Masimo Radical-7 Monitor both offer acoustic respiration rate measurement (RRa) using cleared Masimo technology and third party sensors, the Biolight A Series Patient Monitor does not claim this feature.
  • Both the LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary . predicate Biolight A Series Patient Monitor offer the cleared Suntech noninvasive oscillometric (NIBP) parameter measurement technology. NIBP measurement principles and performance levels are equivalent.
  • Both the LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary . predicate Biolight A Series Patient Monitor offer the cleared Respironics sidestream and mainstream end tidal CO2 (EtCO2) parameter measurement technology. EtCO2 measurement principles and performance levels are equivalent.
  • . The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary predicate Biolight A Series Patient Monitor both measure inspired Oxygen (FiO2). The measurement principles are different for the devices - the Biolight A Series Patient Monitor uses a paramagnetic Oxygen sensor and the LifeWindow™ LW8 Lite Monitor uses a cleared EnviteC electro-galvanic Oxygen cell. The measurement ranges are equivalent.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary . predicate Biolight A Series Patient Monitor both offer the measurement of invasive blood pressure (IBP) parameter using cleared third party blood pressure transducers. Pressure measurement principles and performance levels are equivalent.

{23}------------------------------------------------

  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the primary predicate Biolight A Series Patient Monitor both offer the measurement of patient temperature parameter via cleared third party thermistor sensors meeting the YSI-400 temperature specification. Temperature measurement principles and performance levels are equivalent. The LifeWindow™ LW8 Lite Monitor offers two temperature monitoring channels, the Biolight A Series Patient Monitor offers up to eight channels by adding measurement modules.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate ● Biolight A Series Patient Monitor both provide physiological alarms and have similar low and high level alarm settings and functions. Both monitors provide visual and audible alarms. Audible alarms may be suspended.
  • The environmental conditions of the LifeWindow™ LW8 Lite Multi-Parameter . Monitor and the primary predicate Biolight A Series Patient Monitors specify similar ranges. The EnviteC MySign® O Oxygen Measuring Device has been tested to meet non-air out of hospital transport requirements.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate . Biolight A Series Patient Monitor both specify third party patient-applied accessories. Third party patient-applied accessories specified for use with the LifeWindow™ LW8 Lite Monitor have been 510(k) cleared by their respective manufacturers.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate ● Biolight A Series Patient Monitor both utilize color TFT LCD displays and have similar sizes (8.4 inch, 8.0 inch, respectively).
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate Biolight A Series Patient Monitor both offer key input and touch screen input with similar menu functionality.
  • . The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate Biolight A Series Patient Monitor both offer indicator LED's for the mains and battery charge status.
  • . The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate Biolight A Series Patient Monitor both offer an integrated thermal dot array printer with thermal paper width of 50 mm.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate ● Biolight A Series Patient Monitor both may be connected to an external VGA display - the LifeWindow™ LW8 Lite Monitor specifies isolation components (i.e. isolation transformer) to be used when connecting to external components.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate ● Biolight A Series Patient Monitor both offer USB and RJ45 network connector to support external printers. For a USB or RJ45 connected printer the LifeWindow™ LW8 Lite Monitor specifies isolation components (i.e. port isolators, transformer) to be used when connecting to external components.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate . Biolight A Series Patient Monitor both have similar mains power requirements including a universal power supply input range of 100-240 VAC, 50/60Hz.
  • The LifeWindow™ LW8 Lite Multi-Parameter Monitor incorporates an internal ● rechargeable sealed lead-acid battery and the predicate Biolight A Series Patient Monitor utilizes an internal Li-Ion battery source - the functionality of the batteries is the same, operation when disconnected from mains.
  • The physical size and weight of the LifeWindow™ LW8 Lite Multi-Parameter Monitor and the predicate Biolight A Series Patient Monitors are similar - the

{24}------------------------------------------------

LifeWindow™ LW8 Lite Multi-Parameter Monitor is slightly smaller in size at 235 x 175 x 165 mm, and at 4.5kg it weighs slightly more compared to 4.2 kg for the predicate Biolight Monitor.

In conclusion, the LifeWindow™ LW8 Lite Multi-Parameter Monitor is equivalent in its intended use to the predicate devices, the Biolight A Series and Q Series Patient Monitors, EnviteC MySign® O Oxygen Measuring Device, and Masimo Radical-7 Pulse CO-Oximeter.

In summary, the LifeWindow™ LW8 Lite Multi-Parameter Monitor utilizes equivalent technology and has similar technical specifications as the predicate devices and the reference devices, the Biolight A Series and Q Series Patient Monitors, EnviteC MySign® O Oxygen Measuring Device, and Masimo Radical-7 Pulse CO-Oximeter, respectively. The claims of the subject device are within those of the predicate and reference devices.

Therefore, in consideration of the above, the technological characteristics of the devices do not raise different questions of safety and effectiveness when used as labeled and do not impact the Intended Use of the device for use as a multiparameter monitor.

Non-Clinical Tests Submitted: (b) (1)

LifeWindow™ LW8 Lite Multi-Parameter Monitor was laboratory tested to current applicable standards for medical device electrical safety and electromagnetic compatibility as well as particular standards for multi-function patient monitors. The following standards were utilized in compliance testing:

  • Electrical safety testing per IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 ● (2007) + AM1 (2012) or IEC 60601-1: 2012 (FDA recognition number 19-4)
  • . Electromagnetic compatibility testing per IEC 60601-1-2:2007, 300 Edition (FDA recognition number 19-1)
  • Usability evaluation per IEC 60601-1-6:2010 (FDA recognition number 5-89) ● and IEC 62366:2007 (FDA recognition number 5-114) for professional use
  • . Medical alarm systems testing per IEC 60601-1-8:2006+A1:2012 (FDA recognition number 5-76)
  • . Particular requirements for Electrocardiographic monitors per IEC 60601-2-27:2011, 3rd Edition (FDA recognition number 3-126)
  • . Particular requirements for automated non-invasive sphygmomanometers per IEC 80601-2-30:2013, Edition 1.1 (FDA recognition number 3-152)
  • Particular requirements for invasive blood pressure monitors per IEC 60601-2-● 34:2011, 31d Edition (FDA recognition number 3-115)
  • Particular requirements for multifunction patient monitors per IEC 60601-2-49:2011 (FDA recognition number (none))
  • Particular requirements for respiratory gas monitors per ISO 80601-2-55:2011. . 1st Edition (FDA recognition number 1-96)
  • . Particular requirements for clinical thermometers per ISO 80601-2-56:2009 (FDA recognition number 6-403)
  • Particular requirements for pulse oximeter equipment per ISO 80601-2-61:2011, 1st Edition (FDA recognition number 1-85)

{25}------------------------------------------------

  • Medical Electrical Equipment and System Electromagnetic Immunity Test for . Exposure To Radio Frequency Identification Readers, AIM Standard 7351731:2016, Rev. 1 (FDA recognition number 19-21)
    The monitor met the acceptance criteria for compliance to the standards.

The monitor was evaluated to the requirements of the following quidance documents:

  • . Cardiac Monitor Guidance (including Cardiotachometer and Rate Alarm), Issued November 5, 1998
  • FDA Non-invasive Blood Pressure (NIBP) Guidance Document, March 10, 1997
  • . Guidance on the Content of Premarket Notification [510(k)] Submissions for Clinical Electronic Thermometers, March 1993
  • Pulse Oximeters Premarket Notification Submissions [510(k)s] Guidance for . Industry and Food and Drug Administration Staff March 4, 2013
  • Applying Human Factors and Usability Engineering To Medical Devices -● Guidance For Industry And Food And Drug Administration Staff, Feb 3, 2016

The monitor met the requirements and/or recommendations of the quidance documents.

Risk management, risk and hazard analysis of the probes was performed to the following standard:

  • . Application of risk management to medical devices per ISO 14971
    The monitor met acceptance criteria for residual risks.

The LifeWindow™ LW8 Lite Multi-Parameter Monitor software was developed in accordance with FDA guidelines for MAJOR level of concern devices. The software lifecvcle process was evaluated to meet:

  • Medical device software Software life-cvcle processes per IEC 62304:2006 ● (First Edition) + A1:2015 (FDA recognition number 13-79)
  • Medical device software lifecycle process per IEC 62304 with software safety . class C (equivalent to MAJOR level of concern).
  • FDA Content of Premarket Submissions for Management of Cybersecurity in . Medical Devices, October 2, 2014.

The device software was verified to requirements and validated to meet the specified intended use(s).

A usability / Human Factors/Usability summative study was conducted on the LifeWindow™ LW8 Lite Multi-Parameter Monitor with 15 representative professional users (USA) in a simulated critical care scenario with vital signs generation on the subject software. The study followed the FDA usability quidance. The subject device has been found to be safe and effective for the intended users. uses and use environments.

In summary, the LifeWindow™ LW8 Lite Multi-Parameter Monitor met test criteria for standards conformance to the applicable standards and guidance documents, and residual risks met criteria for acceptability for the intended use.

(2) Clinical Tests Submitted:

(none)

{26}------------------------------------------------

(3) Conclusions from Tests:

As described in (b)(1) and (b)(2) above, the LifeWindow™ LW8 Lite Multi-Parameter Monitor is equivalent to the predicate devices as supported by compliance, laboratory, and biocompatibility testing.

The results of all tests demonstrate that the LifeWindow™ LW8 Lite Multi-Parameter Monitor meet specified requirements and criteria for substantial equivalence to the referenced predicate devices.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).