K Number
K183602
Manufacturer
Date Cleared
2019-06-27

(183 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL). The test is intended for use as an aid in the determination of a man's fertility status. For in vitro, over-the-counter use.

The number of progressively motile sperm cells is only one factor that contributes to a male's fertility status. The SwimCount™ Sperm Quality Test does not provide a complete evaluation of a male's fertility status. For a comprehensive assessment of male fertility status then the patient should consult a physician.

Device Description

The SwimCount™ Sperm Quality Test is intended for in vitro, over-the-counter, use and is a prescreening sperm quality test that measures the number of Progressive Motile Sperm Cells per mL (PMSCs/mL) .

The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL).

The SwimCount™ Sperm Quality Test cut-off value is derived from the WHO laboratory manual for the Examination and processing of human semen, 5th edition guidelines 2010.

The SwimCount™ Sperm Quality Test Kit includes the following components:

  • SwimCount™ Sperm Quality Test Box ●
  • SwimCount™ Sperm Quality Test Device ●
  • Semen Collection Cup (used to collect the sperm sample is to remain in the sample cup for 30 minutes)
  • . Semen Transfer Pipette (syringe) (used to stir the semen sample 10 times before adding the sample to the Device and further used to collect 0.5 mL semen sample from the sperm sample placed in the cup)
  • SwimCount™ Sperm Quality Test Instructions for Use .
AI/ML Overview

The acceptance criteria and study proving the device meets those criteria are detailed below, primarily extracted from the "Summary - SwimCount™ Sperm Quality Test" section of the provided document.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a separate section with specific numerical targets. However, the performance outcomes of key studies (Multi-Center Clinical Study and Lay-User Usability/Interpretation Studies) serve as the implicitly accepted performance metrics for substantial equivalence. The primary clinical performance metrics are Sensitivity, Specificity, and Accuracy against the manual sperm count reference. For usability, the emphasis is on the percentage of correct interpretations by lay-users.

Here's a table summarizing the reported device performance, effectively serving as the demonstrated acceptance criteria through the provided study results:

Performance Metric (Implicit Acceptance Criteria)Reported Device PerformanceStudy Context
Sensitivity (Clinical)95.83% (95% CI: 88.30%-99.13%)Multi-Center Clinical Study
Specificity (Clinical)90.68% (95% CI: 86.23%-94.07%)Multi-Center Clinical Study
Accuracy (Clinical)91.88% (95% CI: 89.90%-95.81%)Multi-Center Clinical Study
Lay-User Correct Calls (Usability - Performance)97.53%Lay-User Usability Study (performance of test completion)
Lay-User Correct Calls (Usability - Interpretation)97.65% (95% CI: 96.70%-98.81%)Consumer Interpretation Study (picture interpretation)
Repeatability (% correct)98.8% Overall (75-100% per concentration)Repeatability Study
Reproducibility (% correct)100% at 30 min & 60 min after collection at all tested concentrationsReproducibility Study

2. Sample Size Used for the Test Set and Data Provenance

  • Multi-Center Clinical Study (Test Set):

    • Sample Size: 308 net semen samples across four centers. A total of 323 samples were initially collected, but 15 were excluded for analysis due to various reasons (e.g., <2 days abstinence, globospermia, hematospermia, leucospermia, samples <0.5 mL, too viscous).
    • Data Provenance: Prospective collection. The study was conducted at 4 different sites:
      • 2 sites in Denmark (Fertility Clinic Ciconia, Nordic Cryo Bank/European Sperm Bank)
      • 1 site in Spain (Fertility Clinic IVI)
      • 1 site in Turkey (Acibadem Fulya Hospital)
    • Patient Category: Males attending fertility clinics and donors from a sperm bank.
  • Lay-User Usability Study (Test Set):

    • Sample Size: 81 Lay-Users. Each Lay-User performed the test with their own semen sample.
    • Data Provenance: Implied prospective, as lay-users performed the test and provided feedback. Country of origin not explicitly stated, but given the company's location (Denmark) and the clinical study sites, it's likely European or a mix.
  • Consumer Interpretation Study (Test Set):

    • Sample Size: 81 participants (Lay-Users), who evaluated 10 SwimCount™ Sperm Quality Test result pictures each, resulting in a total of 810 picture interpretations.
    • Data Provenance: Implied prospective, involves users interpreting pre-existing images. Country of origin not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Multi-Center Clinical Study:

    • Experts: The document states "compared the test results obtained by standard microscopy" and "Manual Sperm Count Reference." While the exact number of experts (e.g., lab technicians, embryologists, andrologists) involved in performing the "standard microscopy" is not specified, it implies that trained laboratory professionals at each of the four clinical sites established the ground truth.
    • Qualifications: Not explicitly stated beyond "standard microscopy" implying trained laboratory personnel at fertility clinics and sperm banks. The reference to "Manual Sperm Count Reference" used in the comparison implies established laboratory protocols and trained personnel.
  • Lay-User Usability Studies (Performance & Interpretation):

    • Experts: For the "Lay-User Usability Study," it states that "Laboratory Professionals" judged the 81 Lay-User results from uploaded pictures, with 79 out of 81 agreeing (97.53%). For the "Consumer Interpretation Study," it also implies laboratory professionals or a defined standard for the "Manual Sperm Count Reference" to which the interpretations of the 810 pictures were compared to establish ground truth for correct vs. incorrect calls.
    • Qualifications: "Laboratory Professionals" are mentioned, suggesting individuals with relevant training in laboratory procedures and result interpretation, likely associated with sperm analysis. No specific years of experience or board certifications are provided.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for resolving discrepancies in the ground truth establishment for the primary clinical test set. The ground truth for the clinical study was established by "standard microscopy." It is common for such laboratory procedures to have internal quality control and potentially double-checking mechanisms, but these are not detailed.

For the Lay-User studies, ground truth for their interpretations was established by "Laboratory Professionals" analyzing the pictures. The adjudication in this context seems to be the comparison of the Lay-User's interpretation against the expert-determined ground truth for each picture/test.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not explicitly mentioned or performed. The studies focused on the device's performance (sensitivity, specificity, accuracy) against a "manual sperm count reference" (ground truth) and lay-user's ability to use and interpret the device. There is no information provided about human readers improving with AI vs. without AI assistance, as this is a qualitative over-the-counter test and not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The SwimCount™ Sperm Quality Test is a qualitative, over-the-counter, in vitro diagnostic device that relies on a user's visual interpretation of a color change. It is not an algorithm-driven device or an automated system that provides a numerical output for human interpretation. Therefore, a "standalone algorithm-only" performance study as typically understood for AI/ML devices is not applicable here. The device's performance is intrinsically linked to the user's ability to follow instructions and interpret the color change, which is why the "Lay-User Usability Study" is crucial.

7. The Type of Ground Truth Used

  • Multi-Center Clinical Study: The ground truth was established by "standard microscopy" for Progressive Motile Sperm Cells per mL (PMSCs/mL). This method is considered a laboratory reference standard in semen analysis.
  • Repeatability/Reproducibility Studies: The ground truth for PMSC concentration levels was determined by "conventional microscopy."
  • Lay-User Usability & Consumer Interpretation Studies: The ground truth for the test results (Low vs. Normal Sperm Quality) was established by "Laboratory Professionals" or based on the "Manual Sperm Count Reference."

8. The Sample Size for the Training Set

The document does not detail a separate "training set" as it would for a machine learning or AI model development. This device appears to be a chemical/biological assay with a visual readout, not a system that requires a distinct data-driven training phase in the context of AI/ML. The development and optimization of the assay itself would have involved laboratory work and testing, but not in the sense of an "AI training set."

9. How the Ground Truth for the Training Set Was Established

As noted above, a "training set" in the context of data-driven models is not described for this device. The accuracy and performance of the device's mechanism (staining mitochondria, color change correlation) would have been established through traditional laboratory R&D, likely using internal samples with ground truth determined by standard microscopy. The document focuses on the validation studies (clinical, usability, analytical) that demonstrate the device's performance against predefined standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 27, 2019

MotilityCount ApS Jacob Mollenbach Co-Founder G1. Koge Landevej 55 Valby, DK-2500 Dk

Re: K183602

Trade/Device Name: SwimCount Sperm Quality Test Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: POV Dated: January 2, 2019 Received: January 4, 2019

Dear Jacob Mollenbach:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Takeesha Taylor-Bell Branch Chief Hematology Devices Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183602

Device Name SwimCountTM Sperm Quality Test

Indications for Use (Describe)

The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL). The test is intended for use as an aid in the determination of a man's fertility status. For in vitro, over-the-counter use.

The number of progressively motile sperm cells is only one factor that contributes to a male's fertility status. The SwimCount™ Sperm Quality Test does not provide a complete evaluation of a male's fertility status. For a comprehensive assessment of male fertility status then the patient should consult a physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Summary - SwimCount™ Sperm Quality Test – Summary

Version: 8

Date: 30th April 2019


510(k) Summary

Submitter:MotilityCount ApS
Address:Gl. Køge Landevej 55
DK-2500 Valby,
Denmark
Contact:Jacob Møllenbach
Co-Founder
Telephone Number:45-21445919
Email:jm@motilitycount.com
Summary Preparation Date:30th April 2019
Device Description:The SwimCount™ Sperm Quality Test is intended for home useand is a pre-screening sperm quality test that measures thenumber of Progressive Motile Sperm Cells per mL (PMSCs/mL)
Trade Name:SwimCount™ Sperm Quality Test
Common Name:Semen Analysis Device
Classification Name:Automated Differential Cell Counter
Regulatory Class:Class II
Product Code:POV
C.F.R. Section:21 CFR 864.5220

{4}------------------------------------------------

Predicate Device

ManufacturerBrand Name510(k)Number
PrimaryGenosis LtdFertell Male Fertility TestK041039

Device Description

The SwimCount™ Sperm Quality Test is intended for in vitro, over-the-counter, use and is a prescreening sperm quality test that measures the number of Progressive Motile Sperm Cells per mL (PMSCs/mL) .

The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL).

The SwimCount™ Sperm Quality Test cut-off value is derived from the WHO laboratory manual for the Examination and processing of human semen, 5th edition guidelines 2010.

The SwimCount™ Sperm Quality Test Kit includes the following components:

  • SwimCount™ Sperm Quality Test Box ●
  • SwimCount™ Sperm Quality Test Device ●
  • Semen Collection Cup (used to collect the sperm sample is to remain in the sample cup for 30 minutes)
  • . Semen Transfer Pipette (syringe) (used to stir the semen sample 10 times before adding the sample to the Device and further used to collect 0.5 mL semen sample from the sperm sample placed in the cup)
  • SwimCount™ Sperm Quality Test Instructions for Use .

Indications for use

The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL). The test is intended for use as an aid in the determination of a man's fertility status. For in vitro, over-the-counter use.

The number of progressively motile sperm cells is only one factor that contributes to a male's fertility status. The SwimCount™ Sperm Quality Test does not provide a complete evaluation of a male's fertility status. For a comprehensive assessment of male fertility status the patient should consult a physician.

{5}------------------------------------------------

Technological Similarities and Differences to the Predicate

Similarities
ItemDevice -SwimCount™ SpermQuality TestPredicate -Fertell MaleFertility Test
Indications for UseThe SwimCount™Sperm Quality Test is aqualitative test thatdetects spermconcentration above orbelow 5,000,000Progressive MotileSperm Cells per mL(PMSCs/mL).The Fertell MaleFertility Testintended use is tomeasure motilesperm in semen asan adjunctive screenof male fertility.
Intended UseThe SwimCount™Sperm Quality Test isintended for use as anaid in the determinationof a man's fertilitystatus. For in vitro,over-the -counter useSame
Test TypeQualitativeSame
Sample TypeHuman SemenSame
Test SettingFor in vitro, over-the-counter useSame
TechnologyThe PMSCs areseparated from the restof the semen sample,stained and captured ona detection filter. Themore PMSCs the darkerthe test resultSperm migration

{6}------------------------------------------------

Differences
ItemDevice –SwimCount™ SpermQuality TestPredicate –Fertell MaleFertility Test
Assay principleThe test result is readas a color intensity andcompared to tworeference colors. Thedarker the color thehigher the number ofPMSCs/mL.Visual line; coloredlabel bound to sperm.
Primary cut-off Level5 M PMSCs/mL(derived from WHO5th edition guidelines2010)10 M/mL motilesperm cells (lowerreference limit, WHO4th edition guidelines1999)
Test TemperatureAmbient98.6° F ~ 37° Celsius
Result ReadingThe test result is readas a color intensity andcompared to tworeference colors. Thedarker the color thehigher the number ofPMSCs/mL.Visual line; coloredlabel bound to sperm
Primary cut-off Level5 M PMSCs/mL(derived from WHO5th edition guidelines2010)10 M/mL motilesperm cells (lowerreference limit, WHO4th edition guidelines1999)

1. Discussion of Technological Similarities between the SwimCount™ Sperm Quality Test and the Primary Predicate Device: Fertell Male Fertility Test

The SwimCount™ Sperm Quality Test and the Primary Predicate Device have the same indication for use, intended use, sample type and test setting.

2. Discussion of Technological Differences between SwimCount™ Sperm Quality Test and the Primary Predicate Device: Fertell Male Fertility Test

Indications for Use - The SwimCount™ Sperm Quality Test is a qualitative test that detects sperm concentration above or below 5,000,000 Progressive Motile Sperm Cells per mL (PMSCs/mL). The test is intended for use as an aid in the determination of a man's fertility status. For in vitro, overthe-counter use.

The number of progressively motile sperm cells is only one factor that contributes to a male's fertility status. The SwimCount™ Sperm Quality Test does not provide a complete evaluation of a

{7}------------------------------------------------

male's fertility status. Sperm concentration and semen volume are only two factors that could impact a man's fertility status and time to pregnancy. For a comprehensive assessment of male fertility status, please consult a physician.

The concentration of progressively motile sperm cells (PMSCs) has consistently been shown to be the most predictive factor with regards to outcome. Around 64% of studies suggest that a reasonable chance of success with artificial insemination requires at least 5x106 motile sperm and this is supported by the WHO's revised reference range for natural conception. (Tomlinson et al. Human Fertility, 2013; 16(3): 175-193).

The SwimCount™ Sperm Quality Test and the Primary Predicate Device both measure the concentration of progressively motile sperms in a fresh semen sample. The difference is that the SwimCount™ Sperm Quality Test has a cut-off level of 5,000,000 PMSCs/mL and the Primary Predicate Device has a cut-off value of 10,000,000 PMSCs/mL. The Primary Predicate Device is derived from the WHO 4th edition guidelines 1999 lower reference level and also mentioned as the reference in their 510(k) decision summary.

The 10,000,000 PMSCs/mL calculation is derived from WHO 4th edition guidelines 1999 as follows:

Total sperm concentration per mL of 20 M x 50% (progressive motile sperm cells) = 10 M PMSCs/mL.

For the SwimCount™ Sperm Quality Test we have used the same method of calculation, however, based on and derived from the WHO 5th edition guidelines 2010 as follows:

The SwimCount™ Sperm Quality Test has a primary cut-off Level of 5 M PMSCs/mL which is derived from WHO's 5th edition guidelines 2010, where in section 2.5.4. it is stated that "The lower reference limit for total motility (PR + NP) is 40% (5th centile, 95% CI 38-42).

"The lower reference limit for progressive motility (PR) is 32% (5th centile, 95% CI 31-34).

The primary cut-off value is calculated using the WHO 5th edition section 2.8.6.: "The lower reference limit for sperm concentration spermatozoa is 15 x 106 spermatozoa per mL (5th centile, 95% CI 12-16 x 106).

The above WHO 5th edition clinical method of calculating the lower reference cut-off value for Progressive Motile Sperm Cells (PMSCs)/mL is then calculated as follows:

Total sperm concentration per mL of 15 M x 32% = 4.8 M PMSCs/mL which is rounded to 5 M PMSCs/mL.

As mentioned, the difference in the %-level to be used is due to SwimCount™ Sperm Quality Test using the latest WHO guidelines and whereas the Primary Predicate Device uses an older WHO version (4th edition guidelines 1999).

{8}------------------------------------------------

Please see below a comparison table of the Lower Reference Levels when calculations are based on the WHO 5th edition guidelines 2010 and the WHO 4th edition guidelines 1999:

ParameterLower ReferenceLimits (WHO 2010)Reference Value(WHO 1999)
Semen Volume1.5 ml≥ 2.0 ml
Sperm Concentration15 million sperm/ ml≥ 20 million sperm/ ml
Progressive Motility32%≥ 50%*
Total Motility40%^n/a
Vitality58% live≥ 75% live
Morphology (Strict criteria)4% normal form(≥ 15% normal form)*

Reference: https://specialty.mims.com/topic/semen-analysis---what-a-clinician-should-knowmy?topic-grouper=cme

The difference in indications for use does not raise any new issues of safety and effectiveness.

Assay Principle - The dye used in the SwimCount™ Sperm Quality Test stains the mitochondria of the spermatozoa and works by entering the mitochondria where the dye is cleaved to a colored product (only in living cells) by an enzyme. In practicality the color changes from yellow to purple/blue and the darker the purple/blue color equals a higher sperm detection level. The Primary Predicate device's assay test unit is positioned onto the liquefied sample and the pressing of a button releases a solution of sodium hyaluronate buffer solution over the semen sample and starts heating the fluid to 37℃. Motile sperm swim-up through the sodium hyaluronate for 30 minutes before a valve is opened, by turning a knob, allowing the buffer solution, and motile sperm present, to flow along a capillary channel. Anti-CD59 monoclonal antibody that is conjugated with colloidal gold is released from an absorbent pad in the channel and reacts with the sperm forming an immunocomplex of gold-labeled sperm. The fluid containing this complex, flows onto a nitrocellulose strip where the gold-labeled sperm are trapped, forming a red line. Unreacted conjugate is washed from the strip by the flow of excess buffer. The appearance of a clear red line (test result) indicates motile sperm in the semen sample at a concentration of ≥ 10 M/mL. The differences in assays raise no new issues of safety and effectiveness.

{9}------------------------------------------------

Test Temperature - The SwimCount™ Sperm Quality Test's is performed at ambient temperature whereas the Primary Predicate Device is performed at 37° Celsius.

Reading Results - The SwimCount™ Sperm Quality Test provides a color chart with 2 possible outcomes; 1) Light Purple/Blue = LOW Sperm Quality = below 5,000,000 PMSCs/mL and 2) Purple/Blue = NORMAL Sperm Quality = above 5.000.000 PMSCs/mL. The Primary Predicative device provides a clear red line (test result) which indicates motile sperm in the semen sample at a concentration of > 10 M/mL. This level is indicative of normally expected motile sperm concentrations.

The differences in reading results raise no new issues of safety and effectiveness.

Non-Clinical Performance Testing

A series of studies were performed that evaluated the traditional laboratory performance characteristics; a summary of each study follows:

Repeatability/Reproducibility

Our precision study (repeatability/reproducibility) was carried out by three operators using three SwimCount™ Sperm Quality Test batches (Operator 1 using SwimCount™ Sperm Quality Test batch 1; Operator 2 using SwimCount™ Sperm Quality Test Batch 2 and Operator 3 using SwimCount™ Sperm Quality Test Batch 3), two replicates per run and two runs per timepoint i.e. 3 x 2 x 2 = 12 samples per time adding up to a total of 36 replicates per sperm concentration (3 time periods). The three Operators were placed at three different locations.

Repeatability

The repeatability (same PMSCs/mL and device lot/operator/location) is shown in the table below (all data at T=30, i.e. 30 minutes after collection of the semen samples):

0 M PMSCs/mLCorrectNumber of tests% correct
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%
1.9 MPMSCs/mLCorrectNumber of tests% correct
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%

{10}------------------------------------------------

4.1 MPMSCs/mLCorrectNumber of tests% correct
Operator 13475%
Operator 244100%
Operator 344100%
SUM111291,7%
6.2 MCorrectNumber of tests% correct
PMSCs/mL
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%
8.4 MPMSCs/mLCorrectNumber of tests% correct
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%
10.1 MCorrectNumber of tests% correct
PMSCs/mL
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%
97.2 MCorrectNumber of tests% correct
PMSCs/mL
Operator 144100%
Operator 244100%
Operator 344100%
SUM1212100%
OverallCorrectNumber of tests% correct
Operator 1272896.4%
Operator 22828100%
Operator 32828100%
SUM838498.8%

The repeatability was found to vary between 75% to 100%. Operator 1 misjudged one sample at 4.1 M PMSCs/mL and as this sample is very close to the cut-off value of 5.0 M PMSCs/mL it is concluded, that the repeatability of the SwimCount™ Sperm Quality Test is highly satisfactory.

{11}------------------------------------------------

Reproducibility

The reproducibility of the three operators/batches and locations was very good, but the raw data indicates, that there was a trend towards reduced reproducibility over time.

A new study was performed. This time the actual progressive motile sperm concentration was determined by conventional microscopy at each time points 30 and 60 minutes were investigated.

Semen Sample PMSCConcentration Levels30 min. after collection60 min. after collection
0 M PMSCs/mLActual PMSC concentration (counted in a Microscope)00
Reproducibility of SwimCount™ Sperm Quality Test100%100%
1.7 M PMSCs/mLActual PMSC concentration (counted in a Microscope)1.7 M1.7 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%
3.9 M PMSCs/mLActual PMSC concentration (counted in a Microscope)3.9 M3.9 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%
6.2 M PMSCs/mLActual PMSC concentration (counted in a Microscope)6.2 M6.2 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%
9.1 M PMSCs/mLActual PMSC concentration (counted in a Microscope)9.1 M9.1 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%
11.4 M PMSCs/mLActual PMSC concentration (counted in a Microscope)11.4 M11.4 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%
116.5 M PMSCs/mLActual PMSC concentration (counted in a Microscope)116.5 M116.5 M
Reproducibility of SwimCount™ Sperm Quality Test100%100%

Conclusion

It is concluded, that the Reproducibility of the SwimCount™ Sperm Quality Test is very good when analyzing the semen samples within 60 minutes after collection.

{12}------------------------------------------------

Semen Sample Stability Study

Semen Sample PMSCConcentration Levels30 min. aftercollection60 min. aftercollection
0 M PMSCs/mLActual PMSC concentration (countedin a Microscope)00
1.7 M PMSCs/mLActual PMSC concentration (countedin a Microscope)1.7 M1.7 M
3.9 M PMSCs/mLActual PMSC concentration (countedin a Microscope)3.9 M3.9 M
6.2 M PMSCs/mLActual PMSC concentration (countedin a Microscope)6.2 M6.2 M
9.1 M PMSCs/mLActual PMSC concentration (countedin a Microscope)9.1 M9.1 M
11.4 M PMSCs/mLActual PMSC concentration (countedin a Microscope)11.4 M11.4 M
116.5 M PMSCs/mLActual PMSC concentration (countedin a Microscope)116.5 M116.5 M

As can be seen from the above table, using the produced semen sample within 60 minutes of producing the sample the PMSC concentrations are stable.

Because of the obtained results it is emphasized in the SwimCount™ Sperm Quality Test Instructions for Use that the sperm sample should be analyzed 30-60 minutes after semen collection.

Evaluation of Test Performance. Cut-off level.

A pooled semen sample from sperm donors was diluted with semen plasma to progressive motile sperm concentrations of 3.9 M, 4.0 M, 5.0 M, 6.0 M, and 6.2 M PMSCs/mL. Each concentration was tested in 40 replicates. The 4.0 M PMSCs/mL, 5.0 M PMSCs/mL and 6.0 M PMSC/mL concentrations were tested on two SwimCount™ Sperm Quality Test lots (i.e. 80 replicates per sample). The results support the 5.0 M PMSCs/mL cut-off level.

{13}------------------------------------------------

Concentration(PMSCs/mL)Percent positive(Positive/Total)Percent negative(Negative/Total)
3.9 M0 (0/40)100 (40/40)
4.0 M7.5 (6/80)92.5 (74/80)
5.0 M55.0 (44/80)45.0 (36/80)
6.0 M90.0 (72/80)10.0 (8/80)
6.2 M100 (40/40)0 (0/40)

Analytical Specificity (interference study)

The purpose of this study is to identify potential cross-reactivity and interference substances and to investigate if any of these interfere with the test result.

Substances in this context are defined as substances contaminating the sample leading to faulty results.

Such substances include:

    1. Bacteria present on skin surfaces and in the sample cup. As model bacteria a gram-positive and a gram-negative bacteria strain are included and in addition microorganisms responsible for sexually transmitted diseases (Neisseria gonorrhoeae and Chlamydia trachomatis) was tested. The concentration tested is set to 104 bacteria per mL, as bacteriospermia is defined as > 103 bacteria per mL by WHO.
  • Saliva and urine which potentially may contaminate the sperm sample. The tested 2. concentration is set to 10%, which is far above the theoretical contamination concentration.
    1. Blood contamination (hematospermia) is very seldom found in semen samples (4) but was investigated at a spike concentration of 1% which corresponds to 107 Red Blood cells (RBC)/ml. The concentration of RBC in blood is approximately 108 RBC/mL.
    1. Possible interfering substances from infections in glands and ureter was investigated using 5 x 106 White Blood Cells /mL. The concentration of WBC in blood is approximately 100 WBC/mL.
    1. Hormones were tested in 10 x the upper reference concentrations found in male serum.

{14}------------------------------------------------

Overview of cross-reactivity and interfering testing of the SwimCount™ Sperm Quality Test

SubstanceSource and catalog no.Test concentration/mL
Gram positive bacteria:Staphylococcus epidermisATCC 14990$10^4$
Gram negative bacteria:Escherichia coliATCC 35218$10^4$
Neisseria gonorrhoeaeATCC 19424$10^4$
Chlamydia trachomatisATCC VR-885$10^4$
Urine-10 %
Saliva-10 %
Red Blood Cells (RBC)Fractionated from donorblood at OdenseUniversity Hospital,Denmark$10^7$
White Blood Cells (WBC)Fractionated from donorblood at OdenseUniversity Hospital,Denmark$5 x 10^6$
D-NorgestrelCayman Chemical, Catno. 1000631710 ng/mL
β-EstradiolSigma E8875600 pg/mL
TestosteroneAcros Organics164410050120 ng/mL

The test was done in triplicate. Each potential interfering substance was spiked in semen samples from vasectomized men (containing semen plasma, only) to investigate any additive effect of interfering substance and in semen samples containing approximately 20 M progressive spermatozoa/mL to investigate any negative effect.

The following substances were identified as substances that may interfere on the SwimCount™ Sperm Quality Test result:

  • Urine ●
  • Red Blood Cells (RBC)
  • . White Blood Cells (WBC)

Urine, Red Blood Cells (RBC) and White Blood Cells (WBC) were included as possible interfering substances in the SwimCount™ Sperm Quality Test Instructions for Use.

Substances which have been identified NOT to interfere with the test:

  • Hormones (D-Norgestrel, B-Estradiol and Testosterone)
  • Microorganism (Staphylococcus epidermis, Escherichia coli, Neisseria gonorrhoeae, ●

{15}------------------------------------------------

Chlamydia trachomatis)

  • Saliva

Summary of results of the Multi-Center Clinical Study of the SwimCount™ Sperm Quality Test

MotilityCount ApS conducted a Multi-Center Clinical Study at 4 different sites in Denmark (2), Spain (1) and Turkey (1) of the SwimCount™ Sperm Ouality Test.

The subjects participating in this study were males attending fertility clinics and donors from a sperm bank with the following Age Brackets:

Multi-Center Clinical Study of the SwimCount™ Sperm Quality Test - No. of Patients per Age Bracket and Category
Number of SamplesNumber of Samples excludedTotal No. of Net Samples
AgeBracketGenderLOW SpermQuality ~ <5 MPMSCs/mLNORMAL SpermQuality ~ ≥5 MPMSCs/mLTotal No.ofSamples< 2 daysAbstinencePeriodGlobospermiaHaematospermiaLeucospermiaSamples<0.5 mLSamplestooviscousTotalCorrected
0-20Male31720020
21-30Male238410711114103
31-40Male27115142351110132
41-58Male1044541153
Total6326032341152215308

The objective of the Multi-Center Clinical Study was to investigate the performance of the SwimCount™ Sperm Quality Test and compare the test results obtained by standard microscopy.

Results - The SwimCount™ Sperm Quality Test appeared as a useful tool for male fertility assessment. The results obtained were compatible with those obtained with conventional sperm analysis. A Sensitivity result of 95.83% and a Specificity result of 90.68% and an Accuracy of 91.88% was obtained based on 308 semen samples.

SwimCount™ Sperm Quality Test - Sensitivity and Specificity for each individual site:

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (< 5 M PMSCs/mL)NORMAL (≥ 5 M PMSCs/mL)
LOW (< 5 M PMSCs/mL)74
NORMAL (≥ 5 M PMSCs/mL)140
95% Conf. interval
Sensitivity87.50%47.35%-99.68%
Specificity90.91%78.33%-97.47%

Center 1 (N=52) - Fertility Clinic Ciconia - Denmark

{16}------------------------------------------------

Center 2 (N=135) - Fertility Clinic IVI - Spain

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (< 5 M PMSCs/mL)NORMAL (≥ 5 M PMSCs/mL)
LOW (< 5 M PMSCs/mL)3814
NORMAL (≥ 5 M PMSCs/mL)083
95% Conf. interval
Sensitivity100.00%90.75%-100.00%
Specificity85.57%76.97%-91.88%

Center 3 (N=91) - Nordic Cryo Bank/European Sperm Bank - Denmark

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (< 5 M PMSCs/mL)NORMAL (≥ 5 M PMSCs/mL)
LOW (< 5 M PMSCs/mL)163
NORMAL (≥ 5 M PMSCs/mL)171
95% Conf. interval
Sensitivity94.12%71.31%-99.85%
Specificity95.95%88.61%-99.16%

Center 4 (N=30) - Acibadem Fulya Hospital - Turkey

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (< 5 M PMSCs/mL)NORMAL (≥ 5 M PMSCs/mL)
LOW (< 5 M PMSCs/mL)81
NORMAL (≥ 5 M PMSCs/mL)120
95% Conf. interval
Sensitivity88.89%51.75%-99.72%
Specificity95.24%76.18%-99.88%

Multi-Center Clinical Test - Calculation based on Line Data

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (<5 M PMSCs/mL)NORMAL (≥5 M PMSCs/mL)
LOW (<5 M PMSCs/mL)6922
NORMAL (≥5 M PMSCs/mL)3214
(4.17% false negative)(9.32% false positive)
95% Conf. Interval
Sensitivity95.83%88.30%-99.13%
Specificity90.68%86.23%-94.07%
Accuracy*91.88%89.90%-95.81%
Total No. of Semen Samples323
Otherwise excluded for analysis15
Included Net No. of Semen Samples308

{17}------------------------------------------------

Summary of results of the SwimCount™ Sperm Quality Test from the Lay-User Usability Study

81 Lay-Users participated in the SwimCount™ Sperm Quality Test of the Lay-User Usability Study and were asked to provide their opinion of the test, labelling and interpreting the result of the test. This was accomplished via a Questionnaire including 18 Questions.

The aim of the Lay-User Usability study was to demonstrate and clarify that the SwimCount™ Sperm Quality Test is simple, accurate and poses no significant risk of erroneous results.

The Lay-User Usability Study was divided into two main parts:

  • A. Lay-User's ability to perform the test as intended and described in the SwimCount™ Sperm Quality Test Instructions for Use
    and

  • B. That Lay-Users can interpret the SwimCount™ Sperm Quality Test result correctly

Consumer Use Study

Lay-Users who participated in the Lay-User Usability Study (Consumer Use Study) provided their opinion on a scale ranging from 'Very Easy'=5 to 'Very Difficult'=1 of the test, labelling and results.

This was accomplished via a Questionnaire and the 6 most relevant out of 18 questions posed to the 81 Lay-User participants are provided in the tables below along with the results.

Question # 1
FOLLOWING AND UNDERSTANDING the SwimCount™ Sperm Quality Test Instructions for Use
#QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
5Very Easy4353.09
43543.21
333.70
200.00
1Very Difficult00.00
TOTAL81100%
Question # 2
APPLYING THE SEMEN SAMPLE correctly to the device according to the SwimCount™ Sperm Quality Test Instructions for Use
#QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
5Very Easy5264.20
4Image: Up Down Arrow2429.63
356.17
200.00
1Very Difficult00.00
TOTAL81100%

{18}------------------------------------------------

Question # 3
FINDING THE RESULT WINDOW according to the SwimCount™ Sperm Quality Test Instructions for Use
#QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
5Very Easy5972.84
41417.28
367.41
211.23
1Very Difficult11.23
TOTAL81100.0%
Question # 4
WHAT DOES A LIGHT BLUE COLOR RESULT mean according to the SwimCount™ Sperm Quality Test Instructions for Use
QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
LOW Sperm Quality (<5 M PMSCs/mL)81100.00 (correct answer)
NORMAL Sperm Quality (≥5 M PMSCs/mL)00.00
TOTAL81100%
Question # 5
WHAT DOES A DARK BLUE COLOR RESULT mean according to the SwimCount™ Sperm Quality Test Instructions for Use
QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
LOW Sperm Quality (<5 M PMSCs/mL)00.00
NORMAL Sperm Quality (≥5 M PMSCs/mL)81100.00 (correct answer)
TOTAL81100%
Question # 6
DETERMINE the EASE-OF-USE WHEN READING YOUR SwimCount™ Sperm Quality Test Result
#QUESTIONNUMBER OF PARTICIPANTS RESPONDINGRESPONSE %
5Very Easy4353.09
4Image: Up Down Arrow2834.57
367.41
222.47
1Very Difficult22.47
TOTAL81100%

The overall summary of results of the SwimCount™ Sperm Quality Test Lay-User Usability Study was as follows:

  • A very good and easy-to-use device (some even called it Great, Super easy and Simple) ● with a score of 4.33 out of 5 to read their own SwimCount™ Sperm Quality Test result
  • 79 out of the 81 participants read their own result correctly ●
  • Out of the 81 read results 15 results were LOW (<5 M PMSCs/mL) and 66 were NORMAL (>5 M PMSCs/mL) as judged by Laboratory Professionals
  • The Lay-Users uploaded a picture of their own SwimCount™ Sperm Quality Test result in . the survey
  • . Thus, in the Lay-User study carried out at home with the Lay-User's own semen sample, there was a 97.53% agreement (% of Correct Calls) between the SwimCount™ Sperm Quality Test result read by the Lay-User at home and those obtained by Laboratory Professionals reading and analyzing the same uploaded picture of the SwimCount™ Sperm Quality Test result

{19}------------------------------------------------

SwimCount™ Sperm Quality Test - % of Correct Calls - Consumer Use Study
StudyNo.ofLay-UsersTotalNo. ofSamplesTotal No.ofResponsesper Lay-User/IDCorrectedfor errorsTotal No.of NetResponsesfrom Lay-Users/IDCorrectAnswersNotCorrectAnswers% ofCorrectCallsby Lay-Users
ConsumerUse Study- Lay UserStudy(Year2018)8181108179297.53%
  • Nothing was confusing as the Pre-Performing phase of the SwimCount™ Sperm Quality . Test had 100% of the Participants reading and fully understanding the SwimCount™ Sperm Quality Test Instructions for Use
  • Regarding the Post Reading of the SwimCount™ Sperm Quality Test Instructions for Use ● 100% of the Participants could perform the SwimCount™ Sperm Quality Test in a satisfactory way

Consumer Interpretation Study

In addition to the Consumer Use Study a Consumer Interpretation Study was carried out.

81 Participants were asked to provide their opinion on 10 SwimCount™ Sperm Quality Test result pictures (N=810). This was accomplished via a Questionnaire including 10 Pictures of a SwimCount™ Sperm Quality Test result each including both LOW and NORMAL Sperm Quality Test Results.

791 out of the 810 pictures were read correctly and 19 pictures were not read correctly by the participants equal to 97.65% of correct calls by lay-users.

{20}------------------------------------------------

Consumer Interpretation Study - Lay User Study

Manual Sperm Count Reference
SwimCount™ Sperm Quality TestLOW (<5 M PMSCs/mL)NORMAL (≥5 M PMSCs/mL)
LOW (<5 M PMSCs/mL)32217
NORMAL (≥5 M PMSCs/mL)2469
(0.62% false negative)(3.50% false positive)
95% Conf. Interval
Correct Calls by Lay-Users97.65%96.70%-98.81%
Total Number of Participants81
Pictures Read Per Participant10
Total Number of Pictures Read810

Conclusions from Non-clinical and Clinical testing

The data from the Non-clinical testing indicates that the SwimCount™ Sperm Quality Test is safe and effective for its intended use.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”