(90 days)
No
The device description and performance studies focus on the mechanical and hydraulic properties of the aspiration system components, with no mention of AI or ML.
Yes
The device is indicated for revascularization in patients with acute ischemic stroke, which is a medical treatment aimed at restoring blood flow and thus falls under the definition of a therapeutic device.
No
The device is a treatment device, specifically designed for revascularization in patients with acute ischemic stroke by restoring blood flow. It facilitates clot removal through aspiration and does not provide diagnostic information.
No
The device description clearly outlines multiple hardware components including a catheter, tubing, pump, and liner set. The device is a physical system used for aspiration, not a software-only solution.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to revascularize patients with acute ischemic stroke by removing intracranial large vessel occlusions. This is a therapeutic intervention performed in vivo (within the living body).
- Device Description: The device components are designed for mechanical aspiration of blood clots from blood vessels. This is a physical process, not a diagnostic test performed on a sample in vitro (outside the living body).
- Lack of Diagnostic Function: There is no mention of the device being used to analyze a sample (like blood, tissue, etc.) to provide diagnostic information about a patient's condition.
IVD devices are used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device's function is to physically remove an obstruction from a blood vessel, which is a treatment, not a diagnostic test.
N/A
Intended Use / Indications for Use
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Product codes (comma separated list FDA assigned to the subject device)
NRY
Device Description
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The AXS Universal Aspiration System is designed for use within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries. The AXS Universal Aspiration System is composed of the following components:
- AXS Catalyst™ Distal Access Catheter
- AXS Universal Aspiration Tubing
- Medela Dominant Flex Pump
- AXS Universal Liner Set
The AXS Catalyst™ Distal Access Catheter is a sterile, single lumen, variable stiffness catheter. The catheter shaft has a hydrophilic coating to reduce friction during use, includes a radiopaque marker on the distal end for angiographic visualization, and includes a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostastic Valve (RHV), Tuohy Borst Valve with Sideport, and Peel Away Introducer. The Rotating Hemostastic Valve and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst. The AXS Catalyst Distal Access Catheter is the only component of the AXS Universal Aspiration System that is used intravascularly.
The AXS Universal Aspiration Tubing serves as a conduit to supply vacuum from the Medela Dominant Flex Pump to the distal tip of the AXS Catalyst Distal Access Catheter. The AXS Universal Aspiration Tubing provides a connection between the sterile and nonsterile environments. The proximal end of the AXS Universal Aspiration Tubing is connected to the AXS Universal Liner Set (outside of the sterile environment) while the distal end of the AXS Universal Aspiration Tubing is connected to the AXS Catalyst Distal Access Catheter (inside the sterile environment). The AXS Universal Liner Set is connected to the Medela Dominant Flex Pump (also outside of the sterile environment).
The Medela Dominant Flex Pump is designed to generate vacuum for the AXS Universal Aspiration System. When used as part of the AXS Universal Aspiration System, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from the Medela Dominant Flex Pump. The Medela Dominant Flex Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The AXS Universal Liner Set is provided non-sterile and consists of an individually packaged canister liner and a ClotFinder specimen cup. The AXS Universal Liner Set is offered with and without a desiccant. The AXS Universal Liner Set is single-use and the repository for aspirated material.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Physician trained in interventional endovascular procedures.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance Data - Bench Testing:
- In-vitro Simulated Use (Direct Aspiration): Evaluated the performance of the Subject Device to Primary Predicate when aspirating clot in tortuous anatomical model. All test samples met acceptance criteria.
- In-vitro Usability Study: Multiple User evaluation of the clot retrieval, durability and kink resistance of the Subject and Primary Predicate Devices in a tortuous anatomical model. All test samples met acceptance criteria.
- Aspiration Flow Rate: Determined the aspiration flow rate and lumen integrity of the system when no occlusion is present. All test samples met acceptance criteria.
- Tip Buckling: Determined the maximum force a catheter tip could withstand before buckling. All test samples met acceptance criteria.
- Track Force: Determined the maximum force a test catheter exerts on the tortuous vessel model as it tracks to the M2 for both advancing and retracting. All test samples met acceptance criteria.
- Catheter Tip and Lumen Patency (Direct Aspiration): Tested resistance to tip and lumen collapse during direct aspiration and test tip integrity to tears and missing material. All test samples met acceptance criteria.
Performance Data - Bench Testing Previously Cleared (from K151667):
- Dimensional Verification: Met acceptance criteria.
- Tip Configuration: Met acceptance criteria.
- Surface Integrity: Met acceptance criteria.
- Tip Buckling: Met acceptance criteria.
- Catheter lubricity and durability: Met acceptance criteria.
- Trackability: Met acceptance criteria.
- Tensile Strength: Met acceptance criteria.
- Liquid Leak Resistance: Of catheter met acceptance criteria.
- Air Leak Resistance: Of catheter met acceptance criteria.
- Catheter Torsional Bond Strength: Met acceptance criteria.
- Flexural Fatigue: Met acceptance criteria.
- Catheter Kink Radius: Met acceptance criteria.
- Chemical Compatibility: Met acceptance criteria.
- Hub Gauging: Met acceptance criteria.
Performance Data - Animal Study:
- In-vivo Efficacy and Safety Evaluation (Arm 1): Assessed vessel revascularization and adverse events associated with mechanical thrombectomy procedure via direct aspiration comparing Subject Device to Primary Predicate Device. Conclusion: Subject Device is equivalent to the Primary Predicate in efficacy and safety.
- In-vivo Efficacy and Safety Evaluation (Arm 2): Assessed the vascular safety profile of treatment by aspiration and navigation of the Subject Device compared to the Primary Predicate Device. Conclusion: Subject Device is equivalent to the Primary Predicate in safety.
- In-vivo Vascular Response (Direct Aspiration): Assessed the vascular response of direct aspiration through the Subject Device compared to Primary Predicate device. Conclusion: Subject Device is equivalent to the Primary Predicate in safety.
Performance Data - Clinical:
No clinical study was conducted. A review of published clinical study articles featuring the Primary Predicate Device demonstrated that equivalent devices had clinical outcomes supporting the proposed indication for use and product code.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K090752 (Penumbra Reperfusion Catheter), K160449 (Penumbra Max Pump, Penumbra Hi-Flo Aspiration Tubing, Penumbra MAX Canister)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
K151667 (AXS Catalyst Distal Access Catheter)
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2018
Stryker Neurovascular Angelica Beckmann Director, Regulatory Affairs 47900 Bayside Parkway Fremont, California 94538
Re: K173841
Trade/Device Name: AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II
Product Code: NRY Dated: December 14, 2017 Received: December 18, 2017
Dear Angelica Beckmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
1
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K173841
Device Name
AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System
Indications for Use (Describe)
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) |
---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Introduction:
According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.
Submitter Name, Address and Content:
| Submitter: | Stryker Neurovascular
47900 Bayside Parkway
Fremont, CA 94538-6515
(FDA Registration Number: 3008853977) |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Angelica Beckmann
Director, Regulatory Affairs
Phone: 510-413-2900
Fax: 510-413-2588
Email: angelica.beckmann@stryker.com |
| Date Prepared: | March 12, 2018 |
| Device Name and Classification: | |
| Trade/Proprietary Name: | AXS Catalyst ™ Distal Access Catheter as part of the AXS Universal
Aspiration System |
| Common Name: | Catheter, Thrombus Retriever |
| Classification Name: | Percutaneous Catheter, 21CFR 870.1250 – Class II |
| Product Code: | NRY |
4
| Legally
Marketed
Predicate
Device(s): | Primary Predicate | Reference Predicate | |
---|---|---|---|
Catheter | Penumbra Reperfusion | ||
Catheter | |||
(K090752) | AXS Catalyst Distal Access | ||
Catheter | |||
(K151667) | |||
Aspiration | |||
System | |||
Components | Penumbra Max Pump | ||
Penumbra Hi-Flo Aspiration | |||
Tubing | |||
Penumbra MAX Canister | |||
(K160449) | N/A |
Device Description
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The AXS Universal Aspiration System is designed for use within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries. The AXS Universal Aspiration System is composed of the following components:
- AXS Catalyst™ Distal Access Catheter ●
- AXS Universal Aspiration Tubing
- . Medela Dominant Flex Pump
- AXS Universal Liner Set
The AXS Catalyst™ Distal Access Catheter is a sterile, single lumen, variable stiffness catheter. The catheter shaft has a hydrophilic coating to reduce friction during use, includes a radiopaque marker on the distal end for angiographic visualization, and includes a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostastic Valve (RHV), Tuohy Borst Valve with Sideport, and Peel Away Introducer. The Rotating Hemostastic Valve and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst. The AXS Catalyst Distal Access Catheter is the only component of the AXS Universal Aspiration System that is used intravascularly.
The AXS Universal Aspiration Tubing serves as a conduit to supply vacuum from the Medela Dominant Flex Pump to the distal tip of the AXS Catalyst Distal Access Catheter. The AXS Universal Aspiration Tubing provides a connection between the sterile and nonsterile environments. The proximal end of the AXS Universal Aspiration Tubing is connected to the AXS Universal Liner Set (outside of the sterile environment) while the distal end of the AXS Universal Aspiration Tubing is connected to the AXS Catalyst Distal Access Catheter (inside the sterile environment). The AXS Universal Liner Set is connected to the Medela Dominant Flex Pump (also outside of the sterile environment).
5
The Medela Dominant Flex Pump is designed to generate vacuum for the AXS Universal Aspiration System. When used as part of the AXS Universal Aspiration System, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from the Medela Dominant Flex Pump. The Medela Dominant Flex Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The AXS Universal Liner Set is provided non-sterile and consists of an individually packaged canister liner and a ClotFinder specimen cup. The AXS Universal Liner Set is offered with and without a desiccant. The AXS Universal Liner Set is single-use and the repository for aspirated material.
Indications for Use
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Technological Characteristics and Product Feature Comparison
Stryker Neurovascular has demonstrated the AXS Catalyst Distal Access Catheters when used as part of the AXS Universal Aspiration System are substantially equivalent to the primary predicate device (K090752 and K160449) and reference predicate device (K151667) based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the subject device with the primary predicate and reference predicate device is summarized in Table 1 below.
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
---|---|---|
Detail | Submission | |
Subject Device | ||
AXS Catalyst Distal Access | ||
Catheter | Primary | |
Predicate Device | ||
Penumbra Reperfusion | ||
Catheter | ||
Manufacturer | Stryker Neurovascular | Penumbra, Inc. |
510(k) Number | K173841 | K090752 |
Device Trade Name | AXS Catalyst TM Distal Access | |
Catheter | Penumbra Reperfusion | |
Catheter (aka 5MAX ACE or | ||
5MAX or ACE60) | ||
Regulation Number | 21 CFR 870.1250 | Same as Subject Device |
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
Detail | Submission | |
Subject Device | ||
AXS Catalyst Distal Access | ||
Catheter | Primary | |
Predicate Device | ||
Penumbra Reperfusion | ||
Catheter | ||
Regulation Name | Percutaneous Catheter | Same as Subject Device |
Regulatory Class | II | Same as Subject Device |
Product Code | NRY | NRY |
Intended Use/ | ||
Indications for Use | The AXS Catalyst Distal Access | |
Catheter as part of the AXS | ||
Universal Aspiration System is | ||
indicated for use in the | ||
revascularization of patients | ||
with acute ischemic stroke | ||
secondary to intracranial large | ||
vessel occlusive disease (in the | ||
internal carotid, middle | ||
cerebral - M1 and M2 | ||
segments, basilar, and | ||
vertebral arteries) within 8 | ||
hours of symptom onset. | ||
Patients who are ineligible for | ||
intravenous tissue plasminogen | ||
activator (IV t-PA) or who failed | ||
IV t-PA are candidates for | ||
treatment. | The Penumbra System is | |
intended for use in the | ||
revascularization of patients | ||
with acute ischemic stroke | ||
secondary to intracranial large | ||
vessel occlusive disease (in the | ||
internal carotid, middle | ||
cerebral - M1 and M2 | ||
segments, basilar, and | ||
vertebral arteries) within 8 | ||
hours of symptom onset. | ||
Device Description | The AXS Catalyst Distal Access | |
Catheter when used as part of | ||
the AXS Universal Aspiration | ||
System is designed to restore | ||
blood flow in patients with | ||
acute ischemic stroke | ||
secondary to intracranial large | ||
vessel occlusive disease. The | ||
AXS Universal Aspiration | ||
System is designed for use | ||
within the internal carotid, | ||
middle cerebral - M1 and M2 | ||
segments, basilar, and | ||
vertebral arteries. The AXS | ||
Universal Aspiration System is | ||
composed of the following | ||
components: | Per the associated | |
DFU: The Penumbra System | ||
consists of three devices that | ||
work as a system to remove | ||
thrombus including the | ||
Penumbra Reperfusion | ||
Catheter, Penumbra Separator | ||
and Aspiration Tubing. The | ||
Penumbra System is used with | ||
the Penumbra Aspiration | ||
Pump. | ||
(Note: The use of the | ||
separator may not be needed | ||
for catheters of inner diameter | ||
of .054" or greater). | ||
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
Detail | Submission | |
Subject Device | Primary | |
Predicate Device | ||
AXS Catalyst Distal Access | ||
Catheter | Penumbra Reperfusion | |
Catheter | ||
AXS Catalyst™ Distal Access Catheter AXS Universal Aspiration Tubing AXS Universal Liner Set Medela Dominant Flex Pump | ||
Accessory Devices | ||
Provided (not in | ||
direct contact with | ||
patient) | Rotating Hemostatic Valve, | |
Tuohy Borst Valve with | ||
sideport (2) Peel Away Sheaths | Peelable Sheath, Rotating | |
Hemostasis Valve, Shaping | ||
Mandrel | ||
Outer Jacket | Pebax with Nylon | Not available |
Reinforcement | Stainless Steel with Nitinol wire | |
and polymer fiber | Stainless Steel Flat wire | |
(0.002x0.007") | ||
Strain Relief | Thermoplastic rubber | Stainless Steel, 304 |
Inner Layer | PTFE | Not available |
Catheter Hub | Nylon | Proximal Hub Grilamid (TR55- |
LX) | ||
Marker Band | Platinum/Iridium | Same as Subject Device |
Adhesive | Cyanoacrylate | Not available |
Outer Jacket | ||
Coating | Hydrophilic Coating | SRDX Harmony (proprietary) |
coating | ||
Labeled Shaft Outer | ||
Diameter | Distal OD: | |
5.3F, 5.4F | Distal OD: | |
5MAX 5F | ||
5MAX ACE 5.4F | ||
Proximal OD: | ||
5.6F, 6.0 F | Proximal OD: | |
5MAX 6F | ||
5MAX ACE 6F | ||
Effective Lengths | 115 cm, 132 cm | 125 cm, 127 cm, 132 cm |
Distal ID (") | 0.058, 0.060 | 0.054, 0.060 |
Proximal ID (") | 0.058, 0.060 | 0.064, 0.068 |
Packaging Materials | ||
and Configuration | Polyethylene Tube and HDPE | |
Packaging Card | Polyester / Polyethylene / | |
Tyvek and Polyethylene | ||
Packaging Card | ||
Sterilization | EO Sterilization | Same as Subject Device |
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
Detail | Submission | |
Subject Device | ||
AXS Catalyst Distal Access | ||
Catheter | Primary | |
Predicate Device | ||
Penumbra Reperfusion | ||
Catheter | ||
Method | ||
Non-Reusable | ||
Device | Single Use device | Same as Subject Device |
Principles of | ||
Operation | The AXS Catalyst Distal Access | |
Catheter is advanced into the | ||
neuro vasculature by a | ||
physician trained in | ||
interventional endovascular | ||
procedures using a compatible | ||
sheath or guide catheter, and | ||
over an appropriately sized | ||
guide wire. A peel away sheath | ||
is provided in the package to | ||
provide support and facilitate | ||
the introduction of the AXS | ||
Catalyst Catheter tip into the | ||
sheath/guide catheter valve. | ||
Once the catheter is inserted, | ||
the peel away sheath can be | ||
removed. Under fluoroscopic | ||
guidance, the catheter can be | ||
advanced through the | ||
vasculature to the desired | ||
location. The catheter is | ||
designed to remove thrombus | ||
from the vasculature using | ||
aspiration provided by an | ||
external source device, the | ||
Medela Dominant Flex Pump. | ||
The aspiration system is | ||
attached to the catheter via the | ||
RHV (or Tuohy Borst), and the | ||
vacuum is turned on. The | ||
aspiration tubing clamp is | ||
opened to apply aspiration. | ||
Upon completion of the | ||
procedure, the AXS Catalyst | ||
Distal Access Catheter, the AXS | Confirm vessel diameter and | |
select an appropriate size | ||
Penumbra Reperfusion | ||
Catheter. The catheter tip may | ||
be shaped using the steam | ||
shaping mandrel provided. | ||
Attach the rotating hemostatis | ||
valve provided to the catheter. | ||
Insert the catheter into the | ||
rotating hemostasis valve | ||
connected to the proximal hub | ||
of a guide catheter. If a guide | ||
catheter is not used, insert the | ||
catheter through the valve of | ||
the long femoral sheath using | ||
the peelable sheath. After | ||
inserting the catheter, remove | ||
the peelable sheath from the | ||
vascular sheath, and peel from | ||
the catheter shaft. Using | ||
conventional catheterization | ||
techniques under fluoroscopic | ||
guidance, advance the catheter | ||
into the target vessel over an | ||
appropriate neurovascular | ||
guidewire. Position the | ||
catheter proximal to the | ||
thrombus. Remove the | ||
guidewire from the catheter. | ||
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
Detail | Submission | |
Subject Device | ||
AXS Catalyst Distal Access | ||
Catheter | Primary | |
Predicate Device | ||
Penumbra Reperfusion | ||
Catheter | ||
Universal Aspiration Tubing, | ||
and the AXS Universal Liner Set | ||
are disposed of per the | ||
institution's procedures. The | ||
Medela Dominant Flex Pump is | ||
cleaned per the instructions | ||
provided with the pump. | ||
How Supplied | Sterile/Single Use | Same as Subject Device |
Aspiration Pump | Medela AG Dominant Flex | |
Pump | ||
K170329 | ||
• Adjustable vacuum pressure | ||
dial | ||
• Three selectable air flow | ||
rates: | ||
40,50,60 L/min | ||
• Non-Sterile | ||
• Reusable | ||
• Cleanable | ||
• Minimum Vacuum Pressure | ||
for aspiration: | ||
-68 kPa [-20.08 inHg] | Penumbra MAX Pump | |
K160449 | ||
• Adjustable vacuum pressure | ||
dial | ||
• Non-adjustable air flow rate | ||
• Non-Sterile | ||
• Reusable | ||
• Cleanable | ||
• Minimum Vacuum Pressure | ||
for aspiration: | ||
-68 kPa [-20.08 inHg] | ||
Aspiration Tubing | Medela AG AXS Universal | |
Aspiration Tubing | ||
• Class II Exempt | ||
• EO Sterilized | ||
• Non-reusable | ||
• ID: 0.218" | ||
• Length: 300cm | ||
• Flow clamp | Penumbra Hi-Flo Aspiration | |
Tubing | ||
• K160449 | ||
• EO Sterilized | ||
• Non-reusable | ||
• ID: 0.110" | ||
• Length: 284.5cm | ||
• Flow switch | ||
Aspiration Canister | Medela AG Dominant Flex | |
Pump utilizes a reusable | ||
polycarbonate suction jar and a | ||
non-reusable liner. The AXS | ||
Universal Liner Set consists of 3 | ||
individually packaged poly-bags | ||
containing one canister liner. | Penumbra MAX Canister is a | |
non-reusable polycarbonate | ||
jar. | ||
• Class I Exempt | ||
• Non-Sterile | ||
Table 1: Product Feature Comparison of Subject Device to Predicate Device | ||
Detail | Submission | Primary |
Subject Device | Predicate Device | |
AXS Catalyst Distal Access | Penumbra Reperfusion | |
Catheter | Catheter | |
and one ClotFinder specimen | ||
cup. The AXS Universal Liner | ||
Set if offered with and without | ||
a desiccant. | ||
• Class I Exempt | ||
• Non-Sterile |
6
7
8
9
10
The differences between the devices are not critical as demonstrated above and through the testing referenced below.
Risk Assessment
Risk assessment of the AXS Catalyst Distal Access Catheter when used as part of the AXS Universal Aspiration System has been conducted in accordance with EN ISO 14971. A warning has been added to the instructions for use to instruct users as to appropriate use of aspiration. Results of testing are appropriate for determining that the AXS Catalyst Distal Access Catheter when used as part of the AXS Universal Aspiration System is substantially equivalent to the legally marketed predicate devices.
Testing Summary
Performance Data - Bench Testing
The results of design verification and design validation testing conducted on the AXS Catalyst Distal Access Catheter when used as part of the AXS Universal Aspiration System demonstrate that it performs as designed, is suitable for the indication for use, and is substantially equivalent to the legally marketed Primary Predicate device. The following design validation tests outlined below in Table 2 were performed on the subject device in support of the indication for use:
11
Table 2: Overview of Design Validation Testing Results – Clot Aspiration | ||
---|---|---|
Test | Test Method Summary | Conclusions |
In-vitro Simulated Use | ||
(Direct Aspiration) | Purpose: To evaluate the performance | |
of the Subject Device to Primary | ||
Predicate when aspirating clot in | ||
tortuous anatomical model. |
Method: Simulated use testing uses a
physiological neurovascular model
where clot is aspirated following the
written protocol. | All test samples met
acceptance criteria. |
| In-vitro Usability Study | Purpose: Multiple User evaluation of
the clot retrieval, durability and kink
resistance of the Subject and Primary
Predicate Devices in a tortuous
anatomical model.
Method: Users performed a direct
aspiration of a clot procedure, tracked
the devices to the site of the occlusion
using a neurovascular model that
replicated the tortuosity, diameter and
location of the arteries in the
neurovasculature. | All test samples met
acceptance criteria. |
Design Verification testing was performed comparing the flow rates of the subject aspiration system to the primary predicate aspiration system. Tip buckling and track forces (insertion and retraction) for the subject device catheters were also compared to the predicate device catheters. Design Verification testing was also performed to assess catheter tip and lumen patency under direct and adjunctive aspiration using the Medela Dominant Flex Pump.
The design verification bench testing is summarized in Table 3 below.
12
Table 3: Performance Data - Bench Testing | ||
---|---|---|
Test | Test Method Summary | Conclusions |
Aspiration Flow Rate | Purpose: To determine the aspiration | |
flow rate and lumen integrity of the | ||
system when no occlusion is present. |
Method: The volume of water an
aspiration system (catheter + associated
pump and accessories) could aspirate in
20s was measured. This volume in mL
was divided by 20 to give the flow rate in
mL/s. | All test samples met
acceptance criteria. |
| Tip Buckling | Purpose: To determine the maximum
force a catheter tip could withstand
before buckling.
Method: The test catheter was soaked in
37°C water prior to testing. A mandrel
was then inserted into the distal end of
the catheter and the catheter with
mandrel was placed into the test fixture
on a tensile test machine. The test
catheter was compressed against a load
cell until the distal tip buckled. The
compression force was recorded. | All test samples met
acceptance criteria. |
| Track Force | Purpose: To determine the maximum
force a test catheter exerts on the
tortuous vessel model as it tracks to the
M2 for both advancing and retracting.
Method: The test catheter was soaked in
37°C water prior to testing. The catheter
was tracked to the M2 through the
smallest ID compatible guide in the
neurovascular challenge path model and
the force was recorded. The catheter
was then retracted through the same
model path and the force was recorded. | All test samples met
acceptance criteria. |
| Test | Test Method Summary | Conclusions |
| Catheter Tip and Lumen
Patency
(Direct Aspiration) | Purpose: To test resistance to tip and
lumen collapse during direct aspiration
and test tip integrity to tears and missing
material.
Method: Prepare test sample and
simulated use model. Insert plug in
catheter tip. Place test sample in the
model to a specified location following
procedural instructions outlined in the
Instructions for Use. Aspirate test
sample using 60cc syringe and aspiration
pump. Visually inspect test sample to
verify indication of no tip or lumen
collapse. | All test samples met
acceptance criteria. |
13
The following bench tests were submitted and cleared as part of K151667 for the AXS Catalyst Distal Access Catheter and remain in effect. The results are provided below in Table 4 for reference purposes.
Table 4: Performance Data - Bench Testing Previously Cleared | |||
---|---|---|---|
Test | Test Method Summary | Conclusions | |
Dimensional Verification | Purpose: To describe the procedure and | ||
technique of making dimensional | |||
measurements using various | |||
measurement equipment. | |||
Method: Verify dimensions using | |||
specified measurement tool. Record | |||
measurements. | Dimensional verification | ||
met acceptance criteria | |||
and was cleared as part | |||
of K151667. | |||
Tip Configuration | Purpose: To verify that the catheter tip is | ||
smooth, rounded, tapered or similarly | |||
finished in order to minimize trauma to | |||
vessels during use per EN ISO 10555-1. | |||
Method: Visually inspect distal tip at 10X | |||
magnification to verify distal tip end is | |||
smooth, rounded, tapered or similarly | |||
finished. Record results. | Tip configuration met | ||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Table 4: Performance Data - Bench Testing Previously Cleared | |||
Test | Test Method Summary | Conclusions | |
Surface Integrity | Purpose: To determine if external | ||
surface of the catheter is free from | |||
extraneous matter, process and surface | |||
defects, and does not have drops of | |||
lubricant fluids. | |||
Method: Visually inspect external | |||
surface of catheter for extraneous | |||
matter, process and surface defects, and | |||
drops of lubricant fluids. Record results. | Surface integrity met | ||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Tip Buckling | Purpose: To measure the maximum | ||
force required to cause a test sample to | |||
buckle. | |||
Method: Prepare sample for test. Use | |||
buckling tester to measure the maximum | |||
force required to cause a test sample to | |||
buckle. Record results. | Tip buckling met | ||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Catheter lubricity and | |||
durability | Purpose: To determine the lubricity and | ||
durability of the coating on the catheter | |||
outer shaft. | |||
Method: Prepare sample for test. Use | |||
friction tester to measure the frictional | |||
force of the device sample when pulled | |||
between two clamped pads. Record the | |||
peak frictional force after 5 cycles. | Coating lubricity and | ||
durability met | |||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Trackability | Purpose: To measure track advance | ||
force of catheter over microcatheter. | |||
Method: A neurovascular model is | |||
placed in a re-circulating water bath at | |||
37°C to simulate human arterial | |||
circulation. The sample is inserted | |||
through model over a microcatheter and | |||
attached to a tensile tester. Advance | |||
catheter through model and determine | |||
peak tracking force. Record results. | Track advance force met | ||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Table 4: Performance Data - Bench Testing Previously Cleared | |||
Test | Test Method Summary | Conclusions | |
Tensile Strength | Purpose: To determine tensile force | ||
tensile force required to induce failure of | |||
fused joints, shaft junctions, and marker | |||
band for non-hydratable catheters based | |||
on EN ISO 10555-1. | Tensile strength met | ||
acceptance criteria and | |||
was cleared as part of | |||
K151667. | |||
Method: Identify joint and prepare | |||
sample for test. Use tensile tester to | |||
determine applied peak tensile force. | |||
Record results. | |||
Liquid Leak Resistance | Purpose: |
- To determine whether catheter
meets the freedom from leakage-liquid
leak requirement 4.7.1 of EN ISO 10555-
- To determine if catheter hub meets
the liquid leakage requirement 4.2.1 of
EN 1707. | Liquid leak resistance of
catheter met
acceptance criteria and
was cleared as part of
K151667. | |
| | Method: Connect test hub sample to
fixture and flush with water to expel air.
Occlude distal tip. Apply pressure of
300kPa minimum and maintain pressure
for 30s. Visually inspect catheter/hub
joint and catheter shaft for leaks.
Record results. | | |
14
15
16
Table 4: Performance Data - Bench Testing Previously Cleared | ||
---|---|---|
Test | Test Method Summary | Conclusions |
Air Leak Resistance | Purpose: |
- To determine whether catheter meets
the freedom from leakage-air aspiration
requirement of 4.7.2 of EN ISO 10555-1. - To determine if catheter hub meets
the air leakage requirement 4.2.2 of EN
Method: Connect test hub sample to a
partially filled syringe. With the nozzle of
the syringe pointing down towards the
ground, withdraw the plunger to the
10cc mark. Hold for 15 seconds and
examine the water in the syringe for the
formation of air bubbles. Record results. | Air leak resistance of
catheter met
acceptance criteria and
was cleared as part of
K151667. |
| Catheter Torsional Bond
Strength | Purpose: To measure the strength of a
catheter shaft when torque is applied.
Torque strength is defined as number of
rotations before failure occurs.
Method: Prepare test sample and insert
into torsional bond strength test fixture
with tortuous path model. Apply torque
to catheter shaft and observe number of
360-degree rotations before failure
occurs. Record results. | Catheter torsional bond
strength met
acceptance criteria and
was cleared as part of
K151667. |
17
Table 4: Performance Data - Bench Testing Previously Cleared | ||
---|---|---|
Test | Test Method Summary | Conclusions |
Flexural Fatigue | Purpose: To determine the flexural | |
fatigue on the catheter shaft. | ||
Method: Prepare test sample. Advance | ||
entire assembly of guide wire, | ||
microcatheter, and test sample into test | ||
model and track it through test model. | ||
While holding the guide wire, | ||
microcatheter, and test sample, pull the | ||
whole assembly pack proximally until it | ||
exits the models. Repeat for nine more | ||
runs. After run number ten, remove | ||
guide wire and microcatheter out of test | ||
sample and inspect for kink or damage. | ||
Record results. | Flexural fatigue met | |
acceptance criteria and | ||
was cleared as part of | ||
K151667. | ||
Catheter Kink Radius | Purpose: To measure the kink radius of a | |
catheter at its distal and specific mid- | ||
shaft joint section. | ||
Method: Prepare test sample. Thread | ||
test sample through fixture loop and lock | ||
down test sample. Pull both ends of test | ||
sample until test sample kinks. Calculate | ||
kink radius using measurement of 2nd to | ||
final loop OD and sample OD. Record | ||
results. | Catheter kink radius met | |
acceptance criteria and | ||
was cleared as part of | ||
K151667. |
18
Table 4: Performance Data - Bench Testing Previously Cleared | ||
---|---|---|
Test | Test Method Summary | Conclusions |
Chemical Compatibility | Purpose: To determine visual and | |
dimensional integrity of catheter | ||
following exposure to saline, non-ionic | ||
and ionic contrast liquids. |
Method: Prepare sample for test. Flush
sample with appropriate chemical.
Measure ID and OD using RAM optical
measurement system. Insert mandrel
through sample to verify inner lumen
integrity. Repeat with second mandrel
and record results. Visually inspect distal
end of sample for any chemical effects
on the shaft, inner lumen and cross-
sectional areas. Record results. | Chemical compatibility
met acceptance criteria
and was cleared as part
of K151667. |
| Hub Gauging | Purpose: To determine if catheter hub
meets gauging requirement 4.1 of EN
1707.
Method: Using the appropriate gauge,
the gauge was applied to the conical
fitting with a total axial force of 5N
without the use of torque. The axial load
was then removed and the sample
inspected. | Hub gauging met
acceptance criteria and
was cleared as part of
K151667. |
Performance Data – Animal Study
Animal studies were conducted in compliance with applicable requirements in the GLP regulation (21 CFR Part 58) to evaluate performance design attributes of the AXS Catalyst™ Distal Access Catheter when used as part of the AXS Universal Aspiration System through user evaluation in an animal model as outlined below in Table 5.
19
Table 5: Performance Data – Animal Study | ||
---|---|---|
Test | Test Method Summary | Conclusions |
In-vivo Efficacy and | ||
Safety Evaluation | ||
(Arm 1) | Purpose: To assess vessel revascularization | |
and adverse events, if any, associated with a | ||
mechanical thrombectomy procedure | ||
performed via direct aspiration using the | ||
Subject Device compared to the Primary | ||
Predicate Device. | Subject Device is | |
equivalent to the | ||
Primary Predicate in | ||
efficacy and safety. | ||
Method: Porcine test subjects were exposed | ||
to aspiration treatment using the AXS | ||
Universal Aspiration System and the | ||
predicate Penumbra System after a vascular | ||
occlusion was artificially induced. Vascular | ||
response was assessed by contrast | ||
angiography and histopathology. | ||
In-vivo Efficacy and | ||
Safety Evaluation | ||
(Arm 2) | Purpose: To assess the vascular safety | |
profile of treatment by aspiration and | ||
navigation of the Subject Device compared | ||
to the Primary Predicate Device. | Subject Device is | |
equivalent to the | ||
Primary Predicate in | ||
safety. | ||
Method: Porcine test subjects were exposed | ||
to aspiration treatment using the AXS | ||
Universal Aspiration System and the | ||
predicate Penumbra System under worst- | ||
case aspiration force and treatment | ||
duration conditions. Vascular response was | ||
assessed by contrast angiography and | ||
histopathology. | ||
In-vivo Vascular | ||
Response | ||
(Direct Aspiration) | Purpose: To assess the vascular response of | |
direct aspiration through the Subject Device | ||
compared to Primary Predicate device. | Subject Device is | |
equivalent to the | ||
Primary Predicate in | ||
safety. | ||
Method: A porcine model was used to | ||
evaluate acute and chronic vessel damage | ||
after direct aspiration in vessels sized | ||
appropriately to simulate the human M2. | ||
Vascular response was assessed by contrast | ||
angiography and histopathology. |
20
Performance Data - Clinical
No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the Primary Predicate Device. The review demonstrated that equivalent devices published clinical outcomes that supported the proposed indication for use, and the corresponding product code. Given the equivalency of the Subject Device and Primary Predicate Device, the Subject Device is therefore suitable for the proposed indication for use and the associated NRY Product Code.
Shelf Life Testing
The labeled shelf life for the AXS Catalyst Distal Access Catheter is two years. Shelf life testing (product and packaging) and Distribution Shipping Challenge Conditioning and testing were performed on the subject device and the results met established criteria. The Shelf Life testing protocol was cleared as part of the reference predicate under K151667 and is ongoing.
The shelf life for the AXS Universal Aspiration Tubing, AXS Universal Liner Set, and Medela Dominant Flex Pump are established by Medela.
Sterilization
The AXS Catalyst Distal Access Catheter and all system components (including the Rotating Hemostatic Valve, the Tuohy Borst Valve and the Peel Away Introducer Sheath) are sterilized with 100% Ethylene Oxide. The AXS Catalyst Distal Access Catheter and the accessories packaged with the Catheter (RHV, TBV, and the Peel Away Introducer Sheath) are provided sterile to a sterility assurance level (SAL) of 10th, and are for single use only.
Ethylene oxide (EO) residuals on a sample representative of the AXS Catalyst Distal Access Catheter are less than the maximum allowed for EO residuals per EN ISO 10993-7 for a limited contact delivery system -externally communicating.
As reported in the reference predicate clearance as part of K151667, results are:
- . Ethylene Oxide Results: 2.29 mg/device extracted residuals (EN ISO 10993-7 requirement is average daily dose