AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is an ambidextrous patient examination glove that is non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs listed below

Minimum Breakthrough detection time (minutes)
Carmustine (BCNU) (3.3 mg/ml)15.9
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240
Cytarabine (100mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Paclitaxel (Taxol) (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)11.1
Vincristine Sulfate (1.0 mg/ml)>240

Do not use with Carmustine and Thiotepa.

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 15.9 minutes and Thiotepa:11.1 minutes.

Device Description

The Powder-free Nitrile Patient Examination Glove, Green Colored, Non-sterile, Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs meets all the requirements of ASTM standard D6319-10. It is an ambidextrous medical glove, which is a disposable device, provided in x-small, medium, large and xlarge sizes. In addition, these gloves were tested for use with chemotherapy drugs per ASTM D6978-05 "Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drug," and Modified-Draize-95 Test, per FDA guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products," 1999.

AI/ML Overview

The provided text describes the acceptance criteria and performance of a medical device, specifically a "Powder-free Nitrile Patient Examination Glove." This document is a 510(k) summary submitted to the FDA, demonstrating substantial equivalence to a predicate device.

It's important to note that this document does not describe an AI/ML medical device. The "device" in question is a physical glove, and the "tests" are for its physical properties and chemical resistance, not for an algorithm's performance on a dataset. Therefore, many of the requested fields related to AI/ML device testing (e.g., sample size for training set, number of experts for ground truth, MRMC study, standalone algorithm performance) are not applicable to this document.

However, I can extract the relevant information regarding the acceptance criteria and how the glove's performance was proven based on the provided text.


Acceptance Criteria and Device Performance (for a physical medical glove)

The document primarily focuses on demonstrating that the subject device (the green nitrile glove) is substantially equivalent to a predicate device (blue or white nitrile gloves) by meeting established standards for patient examination gloves and resistance to chemotherapy drugs.

1. Table of Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance Criteria (Standard Reference and/or General Expectation)Reported Device PerformanceComments on Meeting Criteria
DimensionsASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application) - max allowable defective samples met"The number of defective samples was 0"Meets
Tensile StrengthASTM D6319-10 - max allowable defective samples met"The number of defective samples was 0"Meets
Residual PowderASTM D6319-10 - max allowable defective samples met"The number of defective samples were 0"Meets
Water Leak TestASTM D6319-10 - max allowable defective samples met (2 per lot)"In three of the five lots, the number of defective samples was 1; less or equal than the maximum allowable. In two of the three lots, the number of defective samples was zero."Meets
Low Dermatitis Potential (Modified Draize Test)No clinical evidence of residual chemical additives inducing Type IV allergy in un-sensitized general user populations"no clinical evidence of the presence of residual chemical additives that may induce Type IV allergy..."Meets
Chemotherapy Drug Permeation (ASTM D6978-05)Specific breakthrough times for various chemotherapy drugs, ideally >240 minutes for safe use.See detailed table below.Meets for most drugs, with specific warnings for Carmustine and Thiotepa due to low breakthrough times.
Biocompatibility (Cytotoxicity)ANSI/AAMI/ISO 10993-5 compliant"Cytotoxicity compliant as per ANSI/AAMI/ISO 10993-5:2009/(R)2014"Meets
Biocompatibility (Irritation/Sensitization)ISO 10993-10 compliant"irritation/sensitization...successfully tested in accordance with ISO 10993-10 Third Edition 2010-08-01"Meets

Chemotherapy Drug Permeation Specifics (from "Indications for Use" and "Technological Characteristics" tables):

Chemotherapy DrugAcceptance Criteria (Predicate Device) - Minimum Breakthrough detection time (minutes)Reported Device Performance (Subject Device) - Minimum Breakthrough detection time (minutes)
Carmustine (BCNU) (3.3 mg/ml)10.415.9
Cisplatin (1.0 mg/ml)>240>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240>240
Cytarabine (100mg/ml)>240>240
Dacarbazine (DTIC) (10.0 mg/ml)>240>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240>240
Etoposide (20.0 mg/ml)>240>240
Fluorouracil (50.0 mg/ml)>240>240
Ifosfamide (50.0 mg/ml)>240>240
Methotrexate (25.0 mg/ml)>240>240
Mitomycin C (0.5 mg/ml)>240>240
Mitoxantrone (2.0 mg/ml)>240>240
Paclitaxel (Taxol) (6.0 mg/ml)>240>240
Thiotepa (10.0 mg/ml)90.511.1
Vincristine Sulfate (1.0 mg/ml)>240>240

Important Note: The document explicitly states "Do not use with Carmustine and Thiotepa" due to their extremely low permeation times, even though the subject device still improved upon Carmustine's breakthrough time compared to the predicate.


Regarding the AI/ML-specific questions (which are not applicable to this physical glove device):

  1. Sample sizes used for the test set and the data provenance: Not applicable. Testing was on physical glove samples from various production lots, not on a dataset in the AI/ML context. Data provenance is implied to be from the manufacturer's testing of their product.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties (e.g., hole in glove, chemical permeation) is established through standardized laboratory tests, not human expert consensus. "Experts" would be laboratory technicians or chemists.
  3. Adjudication method for the test set: Not applicable. Laboratory tests follow predefined protocols; no human adjudication is mentioned or implied for test results.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical device, not an AI/ML algorithm.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical device.
  6. The type of ground truth used: For physical and chemical tests, the ground truth is objective measurement against established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards. For biocompatibility, it's the biological response in standardized in-vitro or in-vivo (human repeat insult patch testing) models.
  7. The sample size for the training set: Not applicable. This is not an AI/ML device.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

October 4, 2019

Semperit Investments Asia PTE LTD. % Jay Mansour Principal Mansour Consulting LLC 845 Aronson Lake Court Roswell, Georgia 30075

Re: K183441

Trade/Device Name: Powder-free Nitrile Patient Examination Glove, Green Colored, Non-sterile, Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: August 29, 2019 Received: September 3, 2019

Dear Jay Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183441

Device Name

Powder-free Nitrile Patient Examination Glove, Green Colored, Non-sterile, Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

This device is an ambidextrous patient examination glove that is non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs listed below

Minimum Breakthrough detection time (minutes)
Carmustine (BCNU) (3.3 mg/ml)15.9
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240
Cytarabine (100mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Paclitaxel (Taxol) (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)11.1
Vincristine Sulfate (1.0 mg/ml)>240

Do not use with Carmustine and Thiotepa.

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 15.9 minutes and Thiotepa:11.1 minutes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) # K183441

510k Summary

As required by 21 CFR 807.92

1- Date summary prepared: September 26, 2019

Owner/Submitter/Sponsor/Applicant information: 2-

SEMPERIT INVESTMENTS ASIA PTE. LTD. 8 Jurong Town Hall Road, #29-03 To 06 The JTC Summit Singapore 609434

Contact person Name: Jay Mansour, MSQA, BE, RAC Title: regulatory consultant- Mansour Consulting LLC – Principal Phone: (678) 908-8180 Email address: jay@mansourconsulting.com

  • 3- Device information:
    Common/usual/classification name:

Powder-Free Nitrile Patient Examination Glove

Device name:

Powder-free Nitrile Patient Examination Glove, Green colored, Non-sterile, Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs

FDA 3 letter codeLZA, LZC
FDA regulation number: 21 CFR880.6250
Regulation medical specialtyGeneral Hospital
Review panelGeneral Hospital
ClassI

4- Predicate device:

510k numberTrade or Proprietary or Model NameManufacturer
K151824--Powder Free Nitrile Patient Examination Glove, BlueColored, Non-Sterile, Low Dermatitis Potential, andTested for use with Chemotherapy Drugs--Powder Free Nitrile Patient Examination Glove, WhiteColored, Non-Sterile, Low Dermatitis Potential, andTested for use with Chemotherapy DrugsKOSSAN INTERNATIONALSDN. BHD.

5- Description of the device:

{4}------------------------------------------------

The Powder-free Nitrile Patient Examination Glove, Green Colored, Non-sterile, Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs meets all the requirements of ASTM standard D6319-10. It is an ambidextrous medical glove, which is a disposable device, provided in x-small, medium, large and xlarge sizes. In addition, these gloves were tested for use with chemotherapy drugs per ASTM D6978-05 "Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drug," and Modified-Draize-95 Test, per FDA guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products," 1999.

  • 6- Indications for use
    This device is an ambidextrous patient examination glove that is non-sterile, single use, disposable device intended for medical purposes, worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs listed below
Carmustine (BCNU) (3.3 mg/ml)15.9
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (Cytoxan) (20.0 mg/ml)>240
Cytarabine (100mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Paclitaxel (Taxol) (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)11.1
Vincristine Sulfate (1.0 mg/ml)>240

Minimum Breakthrough detection time (minutes)

{5}------------------------------------------------

Do not use with Carmustine and Thiotepa.

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 15.9 minutes and Thiotepa:11.1 minutes.

7- Technological Characteristics:

Shown below is the technological characteristics comparison of the subject device and the predicate device.

Subject devicePredicate DeviceComments
ManufacturerSEMPERIT INVESTMENTS ASIA PTE. LTD.KOSSAN INTERNATIONAL SDN. BHD.N/A
510(k) numberK183441K151824N/A
Device trade orproprietary namePowder-free Nitrile Patient Examination Glove,Green Colored, Non-sterile, Low DermatitisPotential, and Tested for Use withChemotherapy DrugsPowder Free Nitrile Patient Examination Glove,Blue Colored, Non-Sterile, Low Dermatitis Potential,and Tested for use with Chemotherapy DrugsPowder Free Nitrile Patient Examination Glove,White Colored, Non-Sterile, Low DermatitisPotential, and Tested for use with ChemotherapyDrugsIdentical, except thatsubject device is ingreen, whilepredicate device is inblue or white.
Device ClassificationName/ Regulation numberPatient Examination Glove21 CFR Part 880.6250Patient Examination Glove21 CFR Part 880.6250Identical
Product CodeLZA, LZCLZA, LZCIdentical
Intended UseThis device is an ambidextrous patientexamination glove that is a non-sterile, singleuse, disposable device intended for medicalpurposes, worn on the examiner's hand toprevent contamination between patient andexaminer. The tested chemotherapy drugs are:A patient examination glove is a disposable deviceintended for medical purposes that is worn on theexaminer's hand to prevent contamination betweenpatient and examiner. The tested chemotherapydrugs are:Identical, except forlower breakthroughtime for Thiotepa.(Both devices areambidextrous)
CARMUSTINE (BCNU)3.3 MG/MLCISPLATIN1.0 MG/MLCYCLOPHOSPHAMIDE(CYTOXAN)20.0 MG/MLCYTARABINE100 MG/MLDACARBAZINE (DTIC)10.0 MG/MLDOXORUBICINHYDROCHLORIDE2.0 MG/MLextremely lowpermeation timesof 15.9 minutes.no breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesCARMUSTINE (BCNU)3.3 MG/MLCISPLATIN1.0 MG/MLCYCLOPHOSPHAMIDE(CYTOXAN)20.0 MG/MLCYTARABINE100 MG/MLDACARBAZINE (DTIC)10.0 MG/MLDOXORUBICINHYDROCHLORIDE2.0 MG/MLextremely lowpermeation timesof 10.4 minutes.no breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutes

{6}------------------------------------------------

ETOPOSIDE20.0 MG/MLFLUOROURACIL50.0 MG/MLIFOSFAMIDE50.0 MG/MLMETHOTREXATE25.0 MG/MLMITOMYCIN C0.5 MG/MLMITOXANTRONE2.0 MG/MLPACLITAXEL(TAXOL)6.0 MG/MLTHIOTEPA10.0 MG/MLVINCRISTINESULFATE1.0 MG/MLno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesno breakthroughup to 240 minutesextremely lowpermeation timesof 11.1 minutesno breakthroughup to 240 minutesETOPOSIDE20.0 MG/MLFLUOROURACIL50.0 MG/MLIFOSFAMIDE50.0 MG/MLMETHOTREXATE25.0 MG/MLMITOMYCIN C0.5 MG/MLMITOXANTRONE2.0 MG/MLPACLITAXEL(TAXOL)6.0 MG/MLTHIOTEPA10.0 MG/MLVINCRISTINESULFATE1.0 MG/MLno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesno breakthrough upto 240 minutesbreakthrough after90.5 minutesno breakthrough upto 240 minutes
MaterialsNitrileNitrileIdentical
ColorGreenBlue or WhiteSimilar (greenversus blue)
Standards usedASTM D6319-10ASTM D5151-06ASTM D6124-06ASTM D6978-05(2013)ASTM D412-16ASTM D3767-03ANSI/AAMI/ISO 10993-5ISO 10993-10ISO 2859-1ASTM D6319-10ASTM D5151-11ASTM D6124-11ASTM D6978-05 (Reapproved 2013)ISO 10993-10IdenticalNote: ASTM D412-16,D3767-03 and ISO2859-1 aresupporting standards
Non-Sterile, Single Use,disposableYesYesIdentical
PackagingPacked in Dispenser BoxesPacked in Dispenser BoxesIdentical
Labeling FeaturesNon-SterilePowder FreeExamination GlovesAmbidextrous, by SizeSingle use onlyDevice ColorManufactured by:Lot Number:Quantity by WeightMade in MalaysiaNon-SterilePowder FreeExamination GlovesAmbidextrous, by SizeSingle use onlyDevice ColorManufactured by:Lot Number:Quantity by WeightMade in MalaysiaIdentical
SizesX-Small, Small, Medium, Large, X-LargeX-Small, Small, Medium, Large, X-LargeIdentical
Labeling Claim1. Low Dermatitis Potential2. Tested for Use with Chemotherapy Drugs1. Low Dermatitis Potential2. Tested for Use with Chemotherapy DrugsIdentical
BiocompatibilityCytotoxicity compliant as per ANSI/AAMI/ISO10993-5Sensitization and irritation compliant as per ISO10993-10Sensitization and irritation compliant as per ISO10993-10Identical forsensitization andirritation. Our subjectdevice also tested forCytotoxicity
Modified Draize TestUnder the conditions of the study, there was noclinical evidence of the presence of residualUnder the conditions of the study, there was noclinical evidence of the presence of residualIdentical

{7}------------------------------------------------

chemical additives that may induce Type IV allergy in the un-sensitized general user populations in the tested articleschemical additives that may induce Type IV allergy in the un-sensitized general user populations in the tested articles
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    1. Summary of Non-Clinical Performance Test
      The subject device was tested successfully against:
  • a) Measurement of Dimensions according to Standard Practice for Rubber—Measurement of Dimensions ASTM D3767-03 (2014):
    Testing was conducted on lot numbers of each size. All acceptance criteria according Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6319-10 (2015) were met. The number of defective samples was 0; less or equal than the maximum allowable.

  • b) Measurement of Tensile test according to Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension ASTM D412-16: Testing was conducted on lot numbers of each size. All acceptance criteria according Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6319-10 (2015) were met. The number of defective samples was 0; less or equal than the maximum allowable.

  • Measurement of Residual powder according to Standard Test Method for Residual Powder on Medical c) Gloves ASTM D6124-06 (2017): Testing was conducted on lot numbers of each size. All acceptance criteria according Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6319-10 (2015) were met. The number of defective samples were 0; less or equal than the maximum allowable.

d) Measurement of Water leak test according to Standard Test Method for Detection of Holes in Medical Gloves ASTM D5151-06 (2015): Testing was conducted on lot numbers of each size. All acceptance criteria according Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6319-10 (2015) were met. In three of the five lots, the number of defective samples was 1; less or equal than the maximum

allowable (which is 2 per lot). In two of the three lots, the number of defective samples was zero; less or equal than the maximum allowable.

  • e) Low dermatitis potential according to Standard Test Method For Human Repeat Insult Patch Testing Of Medical Gloves ASTM D6355-07: Under the conditions of the study, there was no clinical evidence of residual chemical additives that may induce Type IV allergy in the un-sensitized general user populations in the tested articles.
  • f) Permeation by Chemotherapy Drugs according to Standard Practice For Assessment Of Resistance Of Medical Gloves To Permeation By Chemotherapy Drugs ASTM D6978-05 (Reapproved 2013) Under the conditions of the study, 15 chemotherapy drugs passed the test, while Carmustine (BCNU) (3.3 mg/ml) failed at 15.9 breakthrough detection time (in minutes), and Thiotepa (10.0 mg/ml) failed at 11.1.
  • Biocompatibility testing was successfully tested in accordance with ANSI AAMI ISO 10993-5:2009/(R)2014 g) (Cytotoxicity) and ISO 10993-10 Third Edition 2010-08-01 (irritation/sensitization)

{8}------------------------------------------------

9. Summary of Clinical Test

Not applicable.

10. Conclusion

The conclusions drawn from the nonclinical (discussed above) demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.