K Number
K183160

Validate with FDA (Live)

Date Cleared
2019-05-09

(175 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TEG® 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and TEG 6s Citrated: K, RT, FF Assay Cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots.

The Citrated: K, RT, FF Assay Cartridge contains three independent assays (CK, CRT and CFF) and the system output consists of a table of numerical values for parameters R, LY30, and MA.

The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and L Y 30 (fibrinolysis after 30 minutes of reaching maximum clot strength).

The CRT assay monitors the hemostasis process via both the intrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 65 Hemostasis System use is with adult patients (18 years and older) where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluation with the TEG 6s Hemostasis System using the Citrated: K, RT, FF Assay Cartridge is used to assess clinical conditions in a trauma setting to assess hemorrhage or thrombosis conditions.

For professional use only.

Device Description

The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitations of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a dysfunction or coagulopathy is present.

To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.

AI/ML Overview

This document describes the TEG® 6s Hemostasis System, an in vitro diagnostic device used to provide semi-quantitative indications of the hemostasis state of a venous blood sample.

Here’s an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for performance in a pass/fail format with exact numerical thresholds for all evaluated aspects. Instead, it details various performance characteristics and states that the data "supports a substantial equivalence determination." However, based on the performance testing presented, we can infer the implicit acceptance criteria by observing the reported results and the conclusions drawn.

Acceptance Criteria (Inferred from Performance Testing)Reported Device Performance
Electrical Safety & EMC Compliance: Meets relevant IEC and UL standards.The system complies with IEC 60601-2-10, IEC 60601-2-10, IEC 60601-2-101 and UL 61010-1 for safety, and IEC 60601-1-2, EN61326-1, EN61326-2-6, EN61000-3-2, EN61000-3-3 and EN55011 for EMC.
Software Verification & Validation: Adheres to FDA guidance for "moderate" level of concern software.Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance. The software was considered "moderate" level of concern.
Reference Ranges: Established according to CLSI C28-A3c.Reference ranges were established using citrated whole blood from normal donors. Non-parametric analysis determined the following ranges: - CK R (minutes): 4.6 - 9.1 (n=157) - CK LY30 (percent): 0.0 - 2.6 (n=132) - CRT MA (mm): 52 - 70 (n=152) - CFF MA (mm): 15 - 32 (n=151)
Analytical Precision: Demonstrated across different coagulation levels (hypo, normal, hyper) and sample types (natural, contrived, patient-derived), using CLSI EP5-A2 guidance with acceptable Standard Deviations (SD) and Coefficients of Variation (%CV).Three precision studies were conducted. Results (SD and %CV) were reported across reagent lots, operators, instruments, and days for various parameters and sample types (Hypo, Normal, Hyper donors; Contrived samples with Dabigatran/Cytochalasin D/ReoPro, Kaolin/RiaStap; Patient-derived samples with Dabigatran/Warfarin; and low/high/no tPA for LY30). The values are presented in Tables 5, 6, 7, and 8, indicating the device maintains precision across these conditions. (Specific numeric thresholds for "acceptable" are not given, but the presentation implies the results are within acceptable industry limits for such a device).
Interference: Identifies interfering factors and limits for CRT and CFF assays.For CK assay, no interfering factors were found among: Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution, Direct Oral Anticoagulants (FXa and direct thrombin inhibitors), Antiplatelet Drug (P2Y12 inhibitor). For CRT assay, Hemolysis and Hemodilution above 30% were found to be interfering factors. For CFF assay, Hemodilution above 40% was found to be an interfering factor.
Method Comparison (Equivalency to Predicate Device): Slopes of linear regression lines close to 1.0 (with 95% CI containing 1.0), acceptable predictive bias relative to acceptance criteria, and high Pearson linear correlation (>0.9 for identical parameters, 0.86 for CRT MA vs CK MA).The linear regression slope estimates for all between-device comparisons were close to 1.0, with their respective 95% confidence intervals all containing 1.0 (range 0.99 to 1.06). The assessment of predictive bias and its 95% confidence interval relative to the bias acceptance criteria supports equivalency according to CLSI EP09-A3. Predicted biases at AMR limits were consistent with reference range limits. Pearson linear correlation estimates were above 0.9 for all identical parameters. The correlation for CRT MA (TEG® 6s) and CK MA (TEG® 5000) was 0.86. Conclusion: The method comparison data strongly supports the correlation between TEG® 6s and the TEG® 5000.

2. Sample Size Used for the Test Set and Data Provenance

  • Reference Ranges:

    • CK R: 157 samples
    • CK LY30: 132 samples
    • CRT MA: 152 samples
    • CFF MA: 151 samples
    • Provenance: Citrated whole blood from normal donors, representative of normal population distributions (age, gender, race), with no known coagulopathies and not taking any drugs that would potentially affect patient hemostasis. (Implies prospective collection for range establishment).
  • Analytical Precision (First Study):

    • Sample Size: Blood from 4 donors (2 Hypo, 1 Normal, 1 Hyper). Run in duplicate for 5 non-consecutive days, across 3 reagent lots, by 2 operators, using 12 analyzers. Each data point in Table 5 represents n=120 replicates.
    • Provenance: Donors with natural coagulation levels (Hypo, Normal, Hyper). (Implies prospective collection from specific donor profiles).
  • Analytical Precision (Second Study - Contrived/Patient-Derived):

    • Sample Size:
      • 3 contrived samples (Hypo, Hyper types)
      • 4 patient-derived dabigatran samples
      • 2 patient-derived warfarin samples
      • Each run in duplicate for 5 non-consecutive days, by 3 operators, using up to 12 analyzers. Each data point in Tables 6 and 7 represents N=12 replicates per parameter/sample/patient ID.
    • Provenance: Contrived samples (blood from normal donors spiked with drugs/reagents). Patient-derived samples (blood from clinical patients being treated with therapeutic levels of anticoagulants). (Combination of prospective collection for spiking and retrospective/prospective from clinical patients).
  • Analytical Precision (Third Study - CK LY30):

    • Sample Size: 3 sample types (no tPA, low tPA, high tPA). 12 replicates per sample type, by 3 operators, using 3 reagent lots for 5 days. Each data point in Table 8 represents N=58 to N=60 replicates.
    • Provenance: Normal donors (for no tPA) and normal donor blood spiked to create low/high tPA samples. (Implies prospective collection for spiking).
  • Method Comparison:

    • Sample Size:
      • CFF MA: 450 samples
      • CK R: 405 samples
      • CK LY30: 86 samples
      • CRT MA vs. CK MA: 336 samples
    • Provenance: Conducted at 12 US clinical sites, enrolling adult patients (18+ years) who met full or limited trauma team criteria of the American College of Surgeons or similar institutional guidelines. (This indicates prospective, multi-center, real-world clinical data collection).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

For this device, the "ground truth" is typically established by the inherent physical properties of the blood samples (clotting characteristics) and comparison to an established predicate device (TEG® 5000), not by human expert interpretation of results in the traditional sense of image analysis or diagnostic decision-making.

  • Reference Ranges: Established using samples from "normal donors" with "no known coagulopathies." The determination of "normal" is likely based on standard medical criteria, possibly verified by physicians or lab personnel, but not explicitly stated as "expert consensus" in the document.
  • Precision Studies: Ground truth is the measured value of the parameters (R, MA, LY30) from the samples themselves, along with their known characteristics (e.g., "hypo," "hyper," spiked with certain drugs). This is measured intrinsically by the device itself, not by external human experts.
  • Method Comparison: The predicate device, TEG® 5000, serves as the "ground truth" or reference for comparison. The study assesses how closely the TEG® 6s results correlate with the TEG® 5000, rather than comparing to a diagnostic expert's opinion. Clinical experts (physicians) would ultimately interpret both TEG® 5000 and TEG® 6s results to make patient diagnoses, but they are not directly establishing the "ground truth" for the device's functional performance in this context.

Therefore, the document does not mention human experts establishing ground truth for the test sets in the way AI/ML studies often describe.

4. Adjudication Method for the Test Set

Not applicable in the conventional sense. The studies described are performance evaluations of an in vitro diagnostic device, involving quantitative measurements. There is no mention of an adjudication process (e.g., 2+1, 3+1 consensus) because the "ground truth" for the performance tests (precision, method comparison) is based on instrumental measurements and comparison to a predicate, not subjective human interpretations that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic tool that provides numerical parameters describing blood hemostasis. It is not an AI-assisted diagnostic imaging device or a tool designed to directly improve human reader performance for tasks like lesion detection in images. The output is a graphical tracing and a table of numerical values for R, LY30, and MA, which clinicians then interpret.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The performance studies described are standalone (algorithm/device only) in nature. The TEG® 6s Hemostasis System automatically measures and outputs parameters without human-in-the-loop diagnostic assistance within the measurement process itself. The interpretation of these results for patient diagnosis is then performed by a professional. The studies evaluate the device's ability to accurately and precisely perform these measurements.

7. The Type of Ground Truth Used

The types of "ground truth" used are:

  • Defined Reference Ranges: Established from "normal donors" (i.e., empirically derived from a healthy population).
  • Known Sample Characteristics: For precision and interference studies, samples were classified as Hypo, Normal, Hyper, or specifically manipulated (spiked with drugs/reagents like Dabigatran, Warfarin, Kaolin, RiaStap, tPA) to represent known coagulation states.
  • Predicate Device Measurements: For method comparison, the measured values from the legally marketed Thrombelastograph® Coagulation Analyzer (TEG®) – 5000 served as the reference for determining substantial equivalence.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" in the context of machine learning. The TEG® 6s Hemostasis System is described as a measurement device based on physical principles (resonant frequency, modulus of elasticity) rather than a machine learning algorithm that is "trained" on data. Therefore, the concept of a separate training set, as typically defined in AI/ML, is not applicable or discussed here. The reference ranges and performance characteristics are established via analytical studies on various sample types.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, there is no explicit "training set" in the AI/ML sense. The device's operational characteristics and parameter calculations are based on established biophysical principles of coagulation, not on machine learning from a labeled training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 9, 2019

Haemonetics Corporation Mark Anzalone Manager, Regulatory Affairs 400 Wood Road Braintree, Massachusetts 02184

Re: K183160

Trade/Device Name: TEG 6s Hemostasis System Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: JPA Dated: November 14, 2018 Received: November 15, 2018

Dear Mark Anzalone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, vou may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183160

Device Name TEG® 6s Hemostasis System

Indications for Use (Describe)

The TEG® 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and TEG 6s Citrated: K, RT, FF Assay Cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots.

The Citrated: K, RT, FF Assay Cartridge contains three independent assays (CK, CRT and CFF) and the system output consists of a table of numerical values for parameters R, LY30, and MA.

The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and L Y 30 (fibrinolysis after 30 minutes of reaching maximum clot strength).

The CRT assay monitors the hemostasis process via both the intrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 65 Hemostasis System use is with adult patients (18 years and older) where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluation with the TEG 6s Hemostasis System using the Citrated: K, RT, FF Assay Cartridge is used to assess clinical conditions in a trauma setting to assess hemorrhage or thrombosis conditions.

For professional use only.
Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored blocks in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol to the right of the word. The blocks are light blue, olive green, gray, and red, respectively.

510(k) Summary

Date: April 8, 2019

SUBMITTER

Haemonetics Corporation 400 Wood Road Braintree, MA 02184

CONTACT

Mark Anzalone Manager, Regulatory Affairs Phone: 781-356-9912 Fax: 781-356-3558 Email: mark.anzalone@haemonetics.com

DEVICE INFORMATION

Trade Name:TEG® 6s Hemostasis System / TEG® 6s Citrated: K, RT, FF Assay Cartridge
Common Name:Whole Blood Hemostasis System
Classification Name:Multipurpose System for in vitro Coagulation Studies
Regulation Number:21 CFR 864.5425
Product Code:JPA, GGN
Device Class:2

PREDICATE DEVICE

Thrombelastograph® Coagulation Analyzer (TEG®) – 5000, K993678/K002177, Product Code JPA (System, Multipurpose, for In Vitro Coagulation Studies), Haemoscope Corporation

DESCRIPTION OF THE DEVICE

System Description

The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitations of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a dysfunction or coagulopathy is present.

{5}------------------------------------------------

HAEMONET

To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

The TEG® 6s Measurement Technique

The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.

Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG® 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.

In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change.

{6}------------------------------------------------

HAEMONET

Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the test tube (ring) and the resonant frequency will rise with increasing firmness of the clot. The TEG® 6s Hemostasis Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot. Results are presented in a format similar to the TEG® 5000.

Both the TEG® 5000 and TEG® 6s Hemostasis System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. Like the TEG® 5000, the TEG® 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity. Method Comparison testing has been performed, yielding data from12 clinical sites. These data include the applicable parameters for the tests in the TEG® 6s Citrated: K, RT, FF Assay Cartridge. The TEG® 5000 System measures the changes of the linkage forces of the clot between the cup and pin over time, while the TEG® 6s Hemostasis System measures the changes of the resonant frequency of the clot over time. Despite the difference in the way the 2 analyzers measure the changes in the modulus of elasticity of the clot over time, they measure the same physical phenomenon and produce the same result when converted from their specific units of measurement (forces for the TEG® 5000 System and frequency for the TEG® 6s Hemostasis System) to millimeters. The fact that these results are similar across all reagents used in both technologies is demonstration that the 2 technologies measure the same phenomenon, the changes in elastic modulus. Table 1 provides the following definitions that apply to calculated parameters in the TEG® 6s Hemostasis System.

TEG® 6sParameterDefinitionParameter Relation toHemostasis
RR is the time from the start of the test untilinitial fibrin formation. This represents theenzymatic portion of coagulation.Normal / reduced / increasedspeed of coagulationinitiation
MAMA, or Maximum Amplitude, represents themaximum firmness of the clot during the test.Normal / reduced / increasedclot elasticity/strength
LY30LY30 is a measurement of the rate of fibrinolysis30 minutes after MA is reached. The LY30measurement is based on the reduction of thetracing area that occurs between the time thatMA is measured until 30 minutes after the MA isdefined.Normal / reduced clotstability; clot dissolution

Table 1. TEG® 6s parameter definitions

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "HAEMONETICS" in red, with a registered trademark symbol next to it. Above the word are four colored squares in a row. The colors of the squares are light blue, olive green, gray, and red.

Citrated Assays

CK assay

The CK assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via the intrinsic pathway in citrated whole blood specimens on the TEG® 6s Hemostasis System. The CK assay consists of Kaolin is used in the assay for activation of coagulation. It is combined with calcium chloride to neutralize the sodium citrate used to anticoagulate the blood sample. The CK hemostasis profile resulting from kaolin activation provides a measure of the time it takes for the first measurable clot to be formed, the kinetics of clot formation, the strength of the clot and the breakdown of the clot, or fibrinolysis.

The clotting characteristics of the CK generated hemostasis profile are described by the functional parameters Clotting Time (R) and Clot Lysis measured thirty minutes after MA (LY30). Since it may take an hour or more for a non-activated whole blood sample to reach maximum amplitude MA, Kaolin is essential to reduce run time and variability associated with running non-activated whole blood samples.

CRT assay

The CRT assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process via both the intrinsic and extrinsic pathway in citrated whole blood specimens on the TEG® 6s Hemostasis System. The CRT is an accelerated assay consisting of Kaolin and Tissue Factor. As described in the CK assay, Kaolin is used for activation of coagulation and is combined with Calcium Chloride to neutralize sodium citrate in the blood sample. The addition of Tissue Factor is used for coagulation activation that would be classically described as extrinsic. The CRT hemostasis profile resulting from Kaolin and Tissue Factor activation provides a measure of the strength of the clot and the breakdown of the clot, or fibrinolysis.

The clotting characteristics of the CRT generated hemostasis profile are described by the functional parameter Maximum Clot Strength (MA). The CRT assay produces an accelerated clotting time which allows for an earlier MA result compared to the CK assay. Therefore, in the TEG® Hemostasis System, the CRT assay is simultaneously run along with the CK assay to provide a fast way to reach a stable value for MA (CRT) while still measuring the timedependent parameters (CK).

CFF assay

The CFF assay is a semi-quantitative in vitro diagnostic assay for monitoring the hemostasis process after blocking platelet contributions to clot strength in citrated whole blood specimens on the TEG® 6s Hemostasis System. The CFF assay consists of Tissue Factor and abciximab (ReoPro®). It is combined with Calcium Chloride to neutralize sodium citrate in the blood sample. Tissue Factor is used for coagulation that would be classically described as extrinsic,

{8}------------------------------------------------

HAEMONE

with platelet aggregation inhibited by abciximab (a GPIIb/IIIa inhibitor), excluding its contribution to clot strength, and thereby measuring fibrinogen contribution to clot strength.

The clotting characteristics of the CFF generated hemostasis profile are described by the functional parameter Maximum Clot Strength (MA) and measures the part of clot strength that is contributed by fibrinogen in the blood sample.

INTENDED USE/INDICATIONS FOR USE:

TEG® 6s Hemostasis System for use with the TEG® 6s Citrated: K, RT, FF Assay Cartridge Intended Use/Indications for Use

The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and TEG 6s Citrated: K, RT, FF Assay Cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots.

The Citrated: K, RT, FF Assay Cartridge contains three independent assays (CK, CRT and CFF) and the system output consists of a table of numerical values for parameters R, LY30, and MA.

The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and LY30 (fibrinolysis after 30 minutes of reaching maximum clot strength).

The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s Hemostasis System use is with adult patients (18 years and older) where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluation with the TEG 6s Hemostasis System using the Citrated: K, RT, FF Assay Cartridge is used to assess clinical conditions in a trauma setting to assess hemorrhage or thrombosis conditions.

For professional use only.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored squares in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol next to it. The squares are light blue, light green, gray, and red, respectively. The text is in a bold, sans-serif font.

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARING THE TEG® 6s HEMOSTATSIS SYSTEM TO THE TEG® 5000 PREDICATE DEVICE

Tables 2 and 3 provide a summary of the similarities and differences of the TEG® 6s Hemostasis system and the TEG® 5000 predicate device, respectively.

ItemTEG® 5000 System PredicateAnalyzerTEG® 6s Hemostasis System
Technological PurposeMonitoring the response of aclot to low levels of appliedstrain (resonance frequency)Monitoring the response of aclot to low levels of appliedstrain (resonance frequency)
MeasurementChanges in physical clotelasticity over timeChanges in physical clotelasticity over time
Initial Warm Up Time5 min5 min
Time to Complete a TestVaries with assayVaries with assay
Measurement OutputGraphical tracings ofresonant frequency perreagent type; table ofparametersGraphical tracings ofresonant frequency perreagent type; table ofparameters
Assay and Reagents
Citrated Kaolin (CK)Kaolin and CaCl2Kaolin and CaCl2, samematerials as TEG® 5000
Citrated RapidTEG™ (CRT)Tissue Factor (TF), Kaolin andCaCl2Tissue Factor (TF), Kaolin andCaCl2 , same materials asTEG® 5000
Citrated FunctionalFibrinogen(CFF)Abciximab, Tissue Factor andCaCl2Abciximab, Tissue Factor andCaCl2, same materials asTEG® 5000

Table 3. Table of Differences

ItemTEG® 5000 System PredicateTEG® 6s Hemostasis System
AnalyzerThrombelastographyanalyzer, separate computerand softwareFully integratedThrombelastographyanalyzer
Measuring TechniqueDirect-contact measurementof shear elasticity of acoagulating sampleNon-contact measurement ofshear elasticity of acoagulating sample
ItemTEG® 5000 System PredicateTEG® 6s Hemostasis System
Measuring ChannelsTwo, each independent and interchangeable (can be used with any approved reagent)Four, each independent and interchangeable (can be used with any approved reagent)
Signal TransducerElectromechanical detection (rotary variable inductive transformer) of rotary motion of a pin suspended in the sampleOptical detection (silicon photodiode) of the motion of a free surface of the sample
Temperature Control20° to 40°C20° to 50°C
Sample Volume (per channel)360-380 μl20 μl
Total Reaction Volume (single channel)360-380 μl20 μl
Mains Supply Voltage120V, 60Hz and 220V, 50Hz model available100-240V, 50-60Hz (international power supply)
Analyzer Input Voltage24 volts AC, 30 watts max12 volts DC, 60 watts max
EnvironmentLevel and vibration free position, no solar radiation Operating temperature: 10° to 35 °C Storage Temperature: -30° to +50 °C (analyzer only) Relative humidity 20-80% (non-condensing)Stable and level surface Operating Temperature 10° to 32°C Storage Temperature: -20° to 50°C (analyzer only) Relative humidity 20 to 80% (non-condensing)
Sample PreparationPerformed by the operator using pipettes to reconstitute reagents and mix reagents with the samplePerformed under analyzer control within the disposable cartridge
PipettingManual accurate pipettes (10, 20, 50, 100, 340, 360, 500, 1000μl)Unmetered transfer pipette or syringe; blood sample is added until it fills to a level above the line marked on the blood intake well of the cartridge
Analyzer SoftwareThrombelastography analyzer, separate computer and softwareFully integrated Thrombelastography analyzer
ItemTEG® 5000 System PredicateTEG® 6s Hemostasis System
ConsumablesCups & Pins (acrylic plastic)Carrier (acrylic plastic) withmicrofluidics laminate andtest rings (acrylic plastic)

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image displays the logo for Haemonetics. The logo consists of four colored squares in the top left corner, with the colors being light blue, light green, gray, and red. Below the squares is the company name, "HAEMONETICS®", in a red serif font. The "®" symbol is located to the right of the company name.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored blocks in a row, followed by the word "HAEMONETICS" in red. The blocks are light blue, olive green, gray, and red. The word "HAEMONETICS" is in a bold, sans-serif font, and the registered trademark symbol is next to the word.

SUMMARY OF PERFORMANCE DATA

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the TEG® 6s Hemostasis analyzer. The system complies with the IEC 60601-2-10, IEC 60601-2-10, IEC 60601-2-101 and UL 61010-1 standards for safety and the IEC 60601-1-2, EN61326-1, EN61326-2-6, EN61000-3-2, EN61000-3-3 and EN55011 standards for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since there is no direct patient contact, any possible injury to a patient is indirect.

Performance Testing

Reference Ranges

Reference Ranges were established according to CLSI C28-A3c. Citrated whole blood from normal donors (representative of normal population distributions -age, gender, race) with no known coagulopathies and not taking any drugs that would potentially affect patient hemostasis was used. Non-parametric method for analysis was used to determine the reference range for each assay parameter. Table 4 contains the reference range data for each assay parameter.

Reagent/ParameterMinMaxSample size
CK R (minutes)4.69.1157
CK LY30 (percent)0.02.6132
CRT MA (mm)5270152
CFF MA (mm)1532151

Analytical Precision

Testing was performed for precision, using CLSI EP5-A2 as guidance. Three types of donor citrated whole blood were used in this precision testing:

  • . Hypo: donors with natural coagulation levels of R parameter near the upper limit of the reference range and MA parameter near the lower limit of the reference range;

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored squares in the top left corner, with the colors being light blue, light green, gray, and red. The word "HAEMONETICS" is written in red, with the registered trademark symbol next to it.

  • Normal: donors with natural coagulation levels of R and MA parameters near the center . of the reference ranges;
  • . Hyper: donors with natural coagulation levels of R parameter near the lower limit of the reference range and MA parameter near the upper limit of the reference range

Testing was performed with blood from 4 donors (2 Hypo, 1 Normal and 1 Hyper) run in duplicate, 2 operators using 3 reagent lots and 12 analyzers for 5 non-consecutive days

Precision test estimates by test, parameter and donor sample test level are shown in Table 5.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for HAEMONETICS. The logo consists of four colored squares in a row, followed by the company name in red. The squares are light blue, olive green, gray, and red, respectively.

TestParameterLevelnMeanReagent LotOperatorInstrument*Day**RepeatabilityTotal^
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
CKRHypol1208.900.00.293.300.00.647.10.859.50.910.0
CKRHypo21206.600.000.000.00.7612.00.7712.00.7712.0
CKRNorm1206.500.000.000.00.284.30.497.60.497.6
CKRHyper1205.20.132.500.000.00.315.80.6613.00.6713.0
CRTMAHypol12063.800.00.090.10.090.10.280.40.330.50.350.6
CRTMAHypo212052.40.390.70.220.400.00.921.80.591.10.741.4
CRTMANorm12062.700.00.150.20.140.20.370.60.310.50.370.6
CRTMAHyper12069.400.000.000.00.911.30.20.30.20.3
CFFMAHypol12021.000.00.150.70.080.40.331.60.311.50.361.7
CFFMAHypo212014.80.110.800.000.00.292.00.332.20.352.3
CFFMANorm12020.600.00.130.600.00.31.50.221.10.251.2
CFFMAHyper12032.700.000.000.02.718.30.531.60.531.6

Table 5. Precision Summary

  • within Operator, Reagent Lot; ** within instrument, operator, reagent lot, ^ within day

{14}------------------------------------------------

HAEMONET

A second precision study was performed to supplement the above precision data. Three types of citrated whole blood samples were used in this precision testing:

  • Hypo (contrived): blood from normal donors with natural coagulation levels of R and ● MA parameters near the center of the reference ranges, spiked with high concentrations of Dabigatran to increase the R parameter and Cytochalasin D and ReoPro® to reduce the MA parameter.
  • . Hyper (contrived): blood from normal donors with natural coagulation levels of R and MA parameters near the center of the reference ranges, spiked with Kaolin solution to reduce the R parameter and RiaStap® (fibrinogen) to increase the MA parameter.
  • . Hypo (patient-derived): blood from clinical patients being treated with therapeutic levels of anticoagulants (Dabigatran or Warfarin (Coumadin))

Testing was performed with 3 contrived samples for the Hypo (contrived) and Hyper (contrived) types, 4 patient-derived dabigatran samples and 2 patient-derived warfarin samples run in duplicate, 3 operators using 3 reagent lots and up to 12 analyzers.

Precision test estimates by test, parameter and donor sample test level are shown in Tables 6 and 7.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored rectangles in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol to the right of the word. The rectangles are light blue, green, gray, and red.

TestParameterContrivedSampletypePatientIDNMeanLotOperatorInstrumentRepetitionTotal
SDCVSDCVSDCVSDCVSD
CKRhyperD101122.10.00.00.00.00.14.50.13.10.15.5
CKRhyperD108122.70.00.00.01.00.13.50.12.40.14.3
CKRhyperD169122.20.12.80.00.00.13.00.12.50.14.6
CKRhypoD1411216.10.00.01.16.91.38.00.74.31.710.8
CKRhypoD1691215.30.85.30.00.01.610.21.17.12.013.2
CKRhypoD1081212.10.11.00.10.90.97.20.75.61.19.3
CRTMAhyperD1081270.80.20.30.00.00.10.20.10.20.20.3
CRTMAhyperD1211272.50.20.30.00.00.40.50.40.50.50.8
CRTMAhyperD1451273.00.20.20.00.00.20.30.20.20.30.4
CRTMAhypoD1411241.30.00.00.00.00.71.80.51.20.92.2
CRTMAhypoD1691244.40.00.00.30.70.61.30.40.90.81.7
CRTMAhypoD1081245.50.00.00.40.91.02.10.51.11.12.5
CFFMAhyperD1011249.90.10.10.00.00.50.90.30.50.51.1
CFFMAhyperD1411244.30.00.00.00.00.51.20.30.60.61.3
CFFMAhyperD1451247.00.00.00.00.00.71.40.51.00.81.7
CFFMAhypoD1011210.20.21.50.00.00.33.00.21.70.43.6
CFFMAhypoD125129.50.00.00.21.80.44.00.32.70.55.0
CFFMAhypoD169126.50.00.00.710.10.34.90.46.00.811.6

Table 6. Precision summary for contrived samples

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains the logo for Haemonetics. The logo consists of four colored rectangles in a row, followed by the word "HAEMONETICS" in red font. The colors of the rectangles are light blue, light green, gray, and red. The registered trademark symbol is located to the right of the word "HAEMONETICS".

TestParameterPatient IDNMeanLotOperatorInstrumentRepetitionTotal
SDCVSDCVSDCVSDCVSDCV
CKRMH-DB00112130.00.00.00.00.43.21.28.91.29.4
CKRMH-DB0021221.30.00.00.00.01.04.72.110.12.311.0
CKRMH-DB0031214.40.00.00.00.00.00.02.416.92.416.9
CKRMH-DB0041211.80.54.00.54.30.00.01.613.91.714.8
CKRMH-WR001129.90.77.00.32.70.32.60.54.90.88.5
CKRMH-WR0021210.11.010.20.00.00.33.21.010.01.413.6
CRTMAMH-DB0011264.90.10.10.00.00.10.10.10.20.10.2
CRTMAMH-DB0021267.70.10.10.00.00.00.00.20.20.20.2
CRTMAMH-DB0031266.00.00.00.00.00.00.00.10.20.10.2
CRTMAMH-DB0041273.80.10.10.00.00.10.10.20.20.20.3
CRTMAMH-WR0011264.00.10.10.00.00.00.00.20.30.20.3
CRTMAMH-WR0021264.70.00.00.00.00.00.00.30.40.30.4
CFFMAMH-DB0011220.70.10.40.00.00.31.30.10.60.31.4
CFFMAMH-DB0021223.70.31.30.31.40.00.00.20.70.41.6
CFFMAMH-DB0031223.40.21.00.00.00.00.00.31.50.41.7
CFFMAMH-DB0041251.91.12.20.00.00.00.00.40.81.12.0
CFFMAMH-WR0011220.70.00.00.00.00.00.00.41.70.41.7
CFFMAMH-WR0021224.70.10.40.31.20.00.00.31.30.41.7

Table 7. Precision summary for patient-derived samples

DB – Dabigatran; WR – Warfarin

{17}------------------------------------------------

HAEMONETIC

A third precision study was performed to support the precision of the CK LY30 parameter. Testing was performed for precision, using CLSI EP5-A2 as guidance. Three types of citrated whole blood samples were used in this precision testing:

  • no tPA: blood from normal donors ●
  • . low tPA (contrived): blood from a normal donor spiked to create a low tPA blood sample
  • high tPA (contrived): blood from a normal donor spiked to create a high tPA blood sample

Testing was performed with the 3 sample types (no tPA, low tPA, high tPA) 12 replicates/sample type using 3 operators using 3 reagent lots for 5 days. Precision test estimates by test, parameter and sample test level are shown in Table 8.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored blocks in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol next to it. The blocks are light blue, light green, gray, and red.

DonorSampleNMeanWithin DayBetween OperatorBetween CartridgeBetween DayTotalTotal (Descriptive)
SDCV%SDCV%SDCV%SDCV%SDCV%SDCV%
146High Lysis6026.231.2250.0000.7431.3551.9671.867
146Low Lysis607.441.08140.0000.2840.84111.40191.2917
146No Lysis601.010.30300.0980.0000.22210.38380.3535
147High Lysis6025.151.1950.0001.1453.21133.61143.2213
147Low Lysis605.090.0000.2350.0000.91181.27251.0821
147No Lysis600.460.23490.12260.0000.15330.30650.2863
100287High Lysis5917.821.1060.5930.4931.1261.6891.689
100287Low Lysis585.030.63130.0000.54110.96191.27251.1423
100287No Lysis581.20.28230.0760.0000.17140.32270.3227

Table 8. CK LY30 Precision Summary

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of a series of colored squares in the upper portion of the image. The squares are light blue, green, gray, and red. Below the squares is the company name, "HAEMONETICS," in red font.

Interference

Testing was performed for interference following CLSI EP7-A2 as guidance.

For the CK assay, potential interfering factors tested were Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution, Direct Oral Anticoagulants (FXa and direct thrombin inhibitors), and Antiplatelet Drug (P2Y12 inhibitor). None were found to be interfering factors.

For the CRT assay, potential interfering factors tested were Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution, Direct Oral Anticoagulants (FXa and direct thrombin inhibitors), and Antiplatelet Drug (P2Y12 inhibitor). Only Hemolysis and Hemodilution above 30% were found to be interfering factors.

For the CFF assay, potential interfering factors tested were Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution, Direct Oral Anticoagulants (FXa and direct thrombin inhibitors), and Antiplatelet Drug (P2Y12 inhibitor). Only Hemodilution above 40% was found to be an interfering factor.

Measurement Interpretation Guidance

The measurement interpretation table below is based on in vitro studies that examined individual values of assays and parameters with respect to their reference ranges. Only one or a few variables influencing TEG results were systematically varied while other variables were kept constant.

The measurement interpretation guidance table is not intended to be comprehensive of all variables that could influence test results, but addresses key variables based on literature review and clinical experience. As with any hemostasis test, TEG 6s test results should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. Refer to Table 9.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored rectangles in a row, with the colors being light blue, light green, gray, and red. Below the rectangles is the word "HAEMONETICS" in red, with a registered trademark symbol next to it.

Parameter(Units)RefRange(RR)ParameterReadoutHemostaticSignificance ofIndividualParameterInterpretation ofParameter Readout forConsiderationReferences
Assay
CKR (min)4.6-9.1CK R > RRHypocoagulable↓ Coagulation factoractivity and/or presenceof heparin at sufficientlyhigh concentrations+1-4, 17, 26
CK R < RRHypercoagulable24, 28
LY30 (%)0-2.6CK LY30 > RRHypocoagulableHyperfibrinolysis5-11, 16, 19-20,23
CRTMA (mm)52-70CRT MA < RRHypocoagulable↓ Fibrinogen or ↓platelet contribution ++8, 10, 12, 13, 18,21, 25, 27
CRT MA > RRHypercoagulable↑ Plateletcontribution++
CFFMA (mm)15-32CFF MA < RRHypocoagulable↓ Fibrinogen++7, 10, 12, 14-15,22, 25, 27
CFF MA > RRHypercoagulable↑ Fibrinogen++

Table 9. TEG® 6s Measurement Interpretation Guidance

Note: These findings were supported by in vitro studies.

  • Detection thresholds were 0.1 IU/mL for unfractionated heparin and 0.3 IU/mL for low molecular weight heparin, respectively in in vitro experiments. Actual threshold values in patients may differ. Heparinization can be exogenous or endogenous 26

++ Other factors, such as factors of the coagulation cascade and FXIII, may contribute to clot stiffness.

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the logo for HAEMONETICS. The logo consists of four colored squares in the top row, with the colors being light blue, light green, gray, and red. The text "HAEMONETICS" is in red and is located below the squares. A registered trademark symbol is located to the right of the text.

Citrated Kaolin (CK)

The standard Citrated Kaolin TEG® assay uses kaolin for activation of coagulation. Kaolin activation has traditionally been described as intrinsic pathway activation. The hemostasis profile resulting from kaolin activation provides a measure of the time it takes for the first measurable clot to be formed, the kinetics of clot formation, the strength of the clot and the breakdown of the clot, or fibrinolysis.

CK R

Kaolin R is the time in minutes elapsing between sample activation and the point in time where clotting provides enough resistance to produce a 2 mm amplitude reading on the TEG® analyzer tracing. The CK R parameter represents the initiation phase of coagulation triggered by enzymatic clotting factors and culminating with the initial fibrin formation. A prolonged R value is indicative of slow clot formation, and a shortened R value indicative of fast clot formation.

Clinical Value. A prolonged R value is indicative of slow clot formation, due to coagulation factor deficiencies or heparin.1-4,1726 A shortened CK R time has been observed in patients post traumatic injury.28

CK LY30

Kaolin LY30 is the percent lysis based on the reduction of the tracing area that occurs between the time maximal amplitude (MA) is measured until 30 minutes after the MA is defined. After the clot has formed, it is degraded by fibrinolytic factors within the blood and consequently the amplitude decreases over time. By measuring the extent of amplitude reduction over time, clot lysis can be assessed.

Clinical Value. LY30 provides information about patient fibrinolysis and potential pathologies. 5-11-6,9,20,23

Citrated RapidTEG™ (CRT)

The RapidTEG assay incorporates both tissue factor and kaolin, which simultaneously activates the intrinsic and extrinsic coagulation pathways. The assay accelerates coagulation compared to the conventional Kaolin test.

CRT MA

RapidTEG™ MA is the point of maximal amplitude of the TEG® tracing, measured in mm, and reflects the maximum clot strength. The strength of the clot is primarily a result of platelet-fibrin interactions via the GPIIb/IIIa receptors.15

Clinical Value. The MA provides information, in combination with CFF MA, of the contribution of platelets to the overall strength of the clot. A decreased MA is indicative of low clot strength, which could be due to decreased platelet contribution or decreased fibrinogen, whereas an increased MA is indicative of high clot strength, which could be due to increased platelet contribution. 8,10,12,13,27

Citrated Functional Fibrinogen (CFF)

The Citrated Functional Fibrinogen assay activates the extrinsic pathway using tissue factor and inhibits platelet aggregation using a platelet inhibitor that binds to GPIIb/Illa receptors.

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored squares in the top left corner, with the colors being light blue, green, gray, and red. Below the squares is the company name "HAEMONETICS" in red, with a registered trademark symbol next to the "S".

CFF MA

The Functional Fibrinogen reagent inhibits platelet aggregation via the GPIIb/IIIa receptor, excluding its contribution to clot strength (MA) and thereby primarily measures the fibrinogen contribution to clot strength.

Clinical Value: CFF MA provides the overall contribution of fibrinogen and/or platelet contribution to clot strength. In conjunction with CRT, this assay enables the contributions of fibrin and platelets to clot strength to be determined. 7,10,12,14,15,22,25,27

Method Comparison

A method comparison study was conducted at 12 US clinical sites collecting patient samples following CLSI EP09-A3 Guidelines. Enrolled were adult patients (male or females 18 years of age and older) who met the full or limited trauma team criteria of the American College of Surgeons or similar criteria established per institutional guidelines.

The assessment of equivalency between the two devices was primarily based on the assessment of predicted bias at the reference range limits relative to the predefined acceptable limits of the bias. Further assessment of equivalency was based on evaluating the estimate of the slope of the linear regression line. An additional evaluation was based on the assessment of the predicted bias at the limits of the analytical measurement range (AMR) and the estimate of the Pearson linear correlation.

The linear regression slope estimates for all between device comparisons were close to 1.0 with their respective 95% confidence intervals all containing 1.0. The slope estimates for all parameters ranged from 0.99 to 1.06.

For all parameters, the assessment of predictive bias and its 95% confidence interval relative to the bias acceptance criteria supports equivalency according to the CLSI EP09-A3. Predicted biases at the AMR limits were consistent with bias predictions at the reference range limits.

Pearson linear correlation estimates were above 0.9 for all identical parameters. In addition, the between device correlation of CRT MA (TEG® 6s) and CK MA (TEG® 5000) was 0.86.

In summary, the method comparison data strongly supports the correlation between TEG® 6s and the TEG® 5000 in patients with known or suspected traumatic injury. Refer to Table 10.

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored squares in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol next to it. The squares are light blue, olive green, gray, and red.

Assay ParameterNSpikeSlopeCorrelationAMR Point 1AMR Bias 1AMR Point 2AMR Bias 2Ref Point 1Ref Bias 1Ref Point 2Ref Bias 2
CFF MA450300.99[0.94, 1.03]0.95[0.94, 0.96]4-2.01[-2.77, -1.23]52-2.6[-4.29, -1.08]15-2.14[-2.44, -1.87]32-2.35[-3.04, -1.75]
CK R405361.05[1, 1.1]0.9[0.88, 0.91]0.40.55[0.31, 0.79]171.33[0.7, 1.95]4.60.75[0.64, 0.85]9.10.96[0.72, 1.2]
CK LY3086171.01[0.91, 1.1]0.91[0.87, 0.94]00.48[-0.12, 1.08]220.6[-1.22, 2.41]00.48[-0.12, 1.08]2.60.49[0, 0.98]
CRT MAvs.CK MA*33601.06[0.99, 1.12]0.86[0.83, 0.89]40-5.31[-6.95, -3.66]75-3.31[-4.12, -2.51]52-4.62[-5.51, -3.73]70-3.6[-4.13, -3.06]

Table 10. Method Comparison Summary Table for TEG® 6s and TEG® 5000

AMR - Analytical Measurement Range; Ref - Reference Range; Point 1/Bias 1 - Lower Limit; Point 2/Bias 2 - Upper Limit

  • The TEC® 6S CRT MA parameter was compared to demosstate that the CRT MA parmeer is equivalent to the CK MA Parameter but to final MA value is reached more quickly using the CRT assay.

{24}------------------------------------------------

HAEMONETIC

CONCLUSIONS DRAWN FROM PERFORMANCE TESTING

The performance data and information provided in this submission support a substantial equivalence determination for the TEG® 6s Hemostasis System and the TEG® 5000 predicate device.

REFERENCES

    1. Da Luz LT, Nascimento B, Shankarakutty AK, Rizoli S, Adhikari NK. Effect of thromboelastography (TEG®) and rotational thromboelastometry (ROTEM®) on diagnosis of coagulopathy, transfusion guidance and mortality in trauma: descriptive systematic review. Crit Care. 2014;18(5):518.
    1. Kang Y. Thromboelastography in liver transplantation. Semin Thromb Hemost. 1995;21(S04): 034-044.
    1. Nielsen VG, Cohen BM, Cohen E. Effects of coagulation factor deficiency on plasma coagulation kinetics determined via thrombelastography : critical roles of fibrinogen and factors II, VII, X and XII. Acta Anaesthesiologica Scandinavica. 2005;49:222-231.
    1. Zmuda K, Neofotistos D, Ts'ao C. Effects of unfractionated heparin, low-molecular-weight heparin, and heparinoid on thromboelastographic assay of blood coagulation. American Journal of Clinical Pathology. 2000;113(5):725-731.
    1. ACS TQIP massive transfusion in trauma guidelines. https://www.facs.org/~/media/files/quality programs/trauma/tqip/massive transfusion in trauma guildelines.ashx. Accessed February 6, 2019.
    1. Cotton BA, Minei KM, et al. Admission rapid thrombelastography predicts development of pulmonary embolism in trauma patients. Journal of Trauma and Acute Care Surgery. 2012;72(6):1470—1477.
    1. Gonzalez E, Moore EE, Moore HB, et al. Trauma-induced coagulopathy: an institution's 35 year perspective on practice and research. Scand J Surg. 2014;103(2):89-103.
    1. Gonzalez E, Moore EE, Moore HB. Management of trauma-induced coagulopathy with thrombelastography. Crit Care Clin. 2017;33(1):119-134.
    1. Gonzalez E, Moore EE, Moore HB, et al. Goal-directed hemostatic resuscitation of traumainduced coagulopathy: a pragmatic randomized clinical trial comparing a viscoelastic assav to conventional coagulation assays. Ann Surg. 2016;263(6):1051-9.
    1. Johansson Pl, Sørensen AM, Larsen CF, et al. Low hemorrhage-related mortality in trauma patients in a Level i trauma center employing transfusion packages and early thromboelastography-directed hemostatic resuscitation with plasma and platelets. Transfusion. 2013;53(12):3088-3099.
    1. Chapman MP, Moore EE, Moore HB, et al. Overwhelming tPA release, not PAI-1 degradation, is responsible for hyperfibrinolysis in severely injured trauma patients. J Trauma Acute Care Surg. 2016;80(1):16-23.

{25}------------------------------------------------

HAEMONETIC

    1. Kornblith LZL, Kutcher MEM, Redick BJ, Calfee CS, Vilardi RF, Cohen MJ. Fibrinogen and platelet contributions to clot formation: Implications for trauma resuscitation and thromboprophylaxis. J Trauma Acute Care Surg. 2014;76(2):255-6-3.
    1. Oshita K, Az-Ma T, Osawa Y, Yuge O. Quantitative measurement of thromboelastography as a function of platelet count. 1999.
    1. Harr JN, Moore EE, Ghasabyan A, et al. Functional fibrinogen assay indicates that fibrinogen is critical in correcting abnormal clot strength following trauma. Shock. 2013;39(1):45-49.
    1. Kawasaki J, Katori N, Kodaka M, Miyao H. Electron microscopic evaluations of clot morphology during thrombelastography. Anesth Analg. 2004;99(5):1440-4.
    1. Chapman MP, Moore EE, Ramos CR, et al. Fibrinolysis greater than 3% is the critical value for initiation of antifibrinolytic therapy. J Trauma Acute Care Surg. 2013;75(6):961-7.
    1. Nascimento B, Al Mahoos M, Callum J, et al. Vitamin K-dependent coagulation factor deficiency in trauma: a comparative analysis between international normalized ratio and thromboelastography. Transfusion. 2012;52(1):7-13.
    1. Bowbrick VA, Mikhailidis DP, Stansby G. Influence of platelet count and activity on thromboelastography parameters. Platelets. 2003;14(4):219-224.
    1. Cotton B a, Harvin J a, Kostousouv V, et al. Hyperfibrinolysis at admission is an uncommon but highly lethal event associated with shock and prehospital fluid administration. J Trauma Acute Care Surg. 2012;73(2):365-70.
    1. Harvin JA, Peirce CA, Mims MM, et al. The impact of tranexamic acid on mortality in injured patients with hyperfibrinolysis. J Trauma Acute Care Surg. 2015;78(5):905-9.
    1. Liu C, Guan Z, Xu Q, et al. Relation of thromboelastography parameters to conventional coagulation tests used to evaluate the hypercoagulable state of aged fracture patients. Medicine (Baltimore). 2016;95(24):e3934.
    1. Meyer M a S, Ostrowski SR, Sørensen AM, et al. Fibrinogen in trauma, an evaluation of thrombelastography and rotational thromboelastometry fibrinogen assays. J Surg Res. 2014:1-10.
    1. Khan M, Jehan F, Bulger EM, et al. Severely injured trauma patients with admission hyperfibrinolysis: Is there a role of tranexamic acid? Findings from the PROPPR trial. J Trauma Acute Care Surg. 2018;85(5):851-857.
    1. Toukh M, Siemens DR, Black A, et al. Thromboelastography identifies hypercoagulability and predicts thromboembolic complications in patients with prostate cancer. Thromb Res. 2014;133(1):88-95.
    1. Erdoes G, Schloer H, Eberle B, et al. Next generation viscoelasticity assays in cardiothoracic surgery: Feasibility of the TEG6s system. PLoS One. 2018;13(12).
    1. Ostrowski SR, Johansson PI. Endothelial glycocalyx degradation induces endogenous heparinization in patients with severe injury and early traumatic coagulopathy. J Trauma Acute Care Surg. 2012;73(1):60-66.

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for Haemonetics. The logo consists of four colored blocks in a row, followed by the word "HAEMONETICS" in red, with a registered trademark symbol to the right of the word. The blocks are light blue, green, gray, and red, respectively.

    1. Noorman F, Hess JR. The contribution of the individual blood elements to the variability of thromboelastographic measures. Transfusion. 2018;58(10):2430-2436.
    1. Schreiber MA, Differding J, Thorborg P, et al. Hypercoagulability is most prevalent early after injury and in female patients. J Trauma. 2005;58(3):475-80.

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.