K Number
K182892
Date Cleared
2019-11-26

(407 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including:

A. Implant placement , including

    1. Preparation of the osteotomy site
    1. Bone contouring , osteoplasty
  • B. Periodontal surgeries
    1. Bone contouring & alveoplasty around living teeth
    1. Removal of exostosis
  • C. Bone grafting
    1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
    1. Harvesting autogen living bone
    1. Sinus elevation & grafting of alveolar sockets

D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions

Device Description

The TRAUS Dental Handpieces; TRAUS CRB46LN and TRAUS CRB46NN are gear driven hand-held dental handpieces with Gear Ratio of 20:1. They can be driven by torque adjustable electrical motors for surgery treatment. They are attached to drive via ISO 3964 coupling. The head clamp accepts instrument complying with ISO 1797-1. They have contra-angle attachment intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

AI/ML Overview

The provided text is an FDA 510(k) Premarket Notification for a dental handpiece. It details the device's indications for use, technological characteristics, and comparison to a predicate device to establish substantial equivalence.

However, the document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/human-in-the-loop performance study.

The "studies" mentioned are:

  • Biocompatibility testing (following ISO standards) to ensure material safety.
  • Bench testing (to ISO 14457) to ensure mechanical performance and compliance.
  • Sterilization validation (following FDA guidance, ASTM, and ISO standards) to confirm effective reprocessing.

These are standard engineering and safety tests for a medical device like a dental handpiece, not performance studies assessing diagnostic accuracy or human performance with or without AI assistance. The device in question is a physical dental instrument, not a software algorithm for diagnostic image analysis.

Therefore, I cannot fulfill the request to describe acceptance criteria and a study proving the device meets them in the context of AI/human-in-the-loop performance, as the provided text does not contain this information for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 26, 2019

Saeshin Precision Co., Ltd. Choi Jong Woo Quality Manager 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun Daegu, 42921 Republic of Korea

Re: K182892

Trade/Device Name: TRAUS Dental Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EGS Dated: August 21, 2019 Received: September 3, 2019

Dear Choi Jong Woo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182892

Device Name TRAUS Dental Handpieces

Indications for Use (Describe)

The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including:

A. Implant placement , including

    1. Preparation of the osteotomy site
    1. Bone contouring , osteoplasty
  • B. Periodontal surgeries
    1. Bone contouring & alveoplasty around living teeth
    1. Removal of exostosis
  • C. Bone grafting
    1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
    1. Harvesting autogen living bone
    1. Sinus elevation & grafting of alveolar sockets

D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for SAESHIN. The logo features a blue geometric shape above the company name, which is also in blue. Below the logo is the company's address, which includes "echeon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu".

52, Secheon-ro 1-gil, Dasa-eup, Dalseong-Tel: +82 53 587 2345 / Fax: +82 53 580 0939

510(k) Summary

[As required by 21 CFR 807.92]

1. 510(k) Number

K182892

2. Date Prepared [21 CFR 807.92(a)(1)]

November 20, 2019

3. Submitter's Information [21 CFR807.92(a)(1)]

  • . Name of Applicant: Saeshin Precision Co., Ltd.
    • 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Address: -Daegu, 42921, Republic of Korea

  • Contact Name: Jong Woo, Choi (Mr.) / Quality Manager .
    • Telephone No. : +82 53 587 2345 -
    • -Fax No. : +82 53 580 0939
    • Email Address : ksqc@saeshin.com -
  • . Registration Number: 3007958831

4. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:TRAUS Dental Handpieces
Classification Name:Dental Handpieces and Accessories
Common Name:Handpiece, Contra- And Right-Angle Attachment,Dental
Classification Panel:Dental
Classification Regulation:21 CFR 872.4200
Product Code:EGS
Device Class:I

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345 and the fax number is +82 53 580 0939.

5. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follows:

  • 510(k) Number: K171436
  • Applicant: Saeshin Precision Co., Ltd. .
  • Common Name: . Dental Handpieces and Accessories
  • Device Name: STRONG Dental Handpieces

There are no significant differences between the Model TRAUS Dental Handpieces and the predicate device. It is substantially equivalent to these devices in device design, composition of materials and technical specifications.

6. Description of the Device [21 CFR 807.92(a)(4)]

The TRAUS Dental Handpieces; TRAUS CRB46LN and TRAUS CRB46NN are gear driven hand-held dental handpieces with Gear Ratio of 20:1. They can be driven by torque adjustable electrical motors for surgery treatment. They are attached to drive via ISO 3964 coupling. The head clamp accepts instrument complying with ISO 1797-1. They have contra-angle attachment intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

7. Indications for Use [21 CFR 807.92(a)(5)]

The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including:

  • A. Implant placement, including
      1. Preparation of the osteotomy site
      1. Bone contouring, osteoplasty
  • B. Periodontal surgeries
      1. Bone contouring & alveoplasty around living teeth
      1. Removal of exostosis
  • C. Bone grafting
      1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
      1. Harvesting autogen living bone
      1. Sinus elevation & grafting of alveolar sockets
  • D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a blue geometric shape above the word "SAESHIN" in a stylized blue font. The geometric shape is made up of several smaller shapes, creating a three-dimensional effect. The overall design is modern and corporate.

8. Technological Characteristics [21 CFR 807.92(a)(6)]

TRAUS CRB46LNTRAUS CRB46NN
Revolu-tionMax Speedin Rpm40,000 rpm40,000 rpm
After the gearspeed reduction2,000 rpm2,000 rpm
Gear Ratio20:120:1
Weightapprox.87gapprox.88g
SizeØ20 x 94.8mmØ20 x 94.8mm
ArticlesLow speed angleLow speed angle
Standard CouplingISO 3964ISO 3964

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345 and the fax number is +82 53 580 0939.

Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] 9.

When compared to the predicate device (K171436), the STRONG Dental Handpieces presented in this submission has the same:

  • Indications for Use .
  • Device Design .
  • Composition of materials .
  • · Technical Specifications
ParameterProposedTRAUS Dental Handpieces (Model Nos.:TRAUS CRB46LN, TRAUS CRB46NN)PredicateSTRONG Dental Handpieces (Model Nos.:ACL(B)-611, ACL(B)-621, ACL(B)-631,ACL(B)-651, ACL(B)-04C)
510(k) NumberUnknownK171436
ManufacturerSaeshin Precision Co., LtdSaeshin Precision Co., Ltd.
Indications for UseThe TRAUS Dental Handpieces are indicated for wide range of dental procedures.A. Implant placement, including1. Preparation of the osteotomy site2. Bone contouring, osteoplastyB. Periodontal surgeries1. Bone contouring & alveoplasty around living teeth2. Removal of exostosisC. Bone grafting1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)2. Harvesting autogen living bone3. Sinus elevation & grafting of alveolar socketsThe Strong Dental Handpieces are indicated for wide range of dental procedures.A. Implant placement, including1. Preparation of the osteotomy site2. Bone contouring, osteoplastyB. Periodontal surgeries1. Bone contouring & alveoplasty around living teeth2. Removal of exostosisC. Bone grafting1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)2. Harvesting autogen living bone3. Sinus elevation & grafting of alveolar socketsD. Removal and sectioning of teeth and teeth
ParameterProposedTRAUS Dental Handpieces (Model Nos.:TRAUS CRB46LN, TRAUS CRB46NN)PredicateSTRONG Dental Handpieces (Model Nos.:ACL(B)-61I, ACL(B)-62I, ACL(B)-631,ACL(B)-651, ACL(B)-04C)
D. Removal and sectioning of teeth and teethbone for e.g. impacted third molars andcomplicated extractionsbone for e.g. impacted third molars andcomplicated extractions
Device Design
Operational ModeGearGear
Gear Ratio20:120:1, 16:1, 64:1, 32:1, 1:1
Length94.8 mm96.9, 96.9, 99.5, 96.9, 83.3 mm
DiameterØ 20Ø 19.6
Dia. of tool2.35 mm2.35 mm
Max. Overall Length ofRotary Instrument30 mm30 mm
Head Height14 mm14.45,14.2 mm
Head DiameterØ 9.6 mmØ 8, Ø 7.7 mm
ShankBy ISO 1797-1By ISO 1797-1
Length of Shank9 – 12 mm9 – 12 mm
Type of ChuckPush-Button LockingPush-Button Locking
Coupling DimensionBy ISO 3964By ISO 3964
Type of ConnectorNot AppliedNot Applied
AccessoriesSpray NozzleSpanner
Composition of Materials/Surface treatment
GearSUS420F/Vacuum Heat TreatmentSUS420F/Vacuum Heat Treatment
ShankN/ASUS304
Head and NozzleSUS303F/Hard Chrome coatedSUS303F/Hard Chrome coated
ChuckSUS420F/Vacuum Heat TreatmentSUS420F/Vacuum Heat Treatment
HandleSUS303F/Hard Chrome coatedAL6061/Anodizing
PipeSUS304/Hard Chrome coatedSUS303F/Hard Chrome coated
N/A (ACL(B)-04C)
Patient-ContactingChuck, HeadChuck, Head
Operator-ContactingHandle, HeadHandle, Head
Technical Specification
Chuck DesignType1 Push-Button Locking by ISO 1797-1Type1 Push-Button Locking by ISO 1797-1
ParameterProposedTRAUS Dental Handpieces (Model Nos.:TRAUS CRB46LN, TRAUS CRB46NN)PredicateSTRONG Dental Handpieces (Model Nos.:ACL(B)-61I, ACL(B)-62I, ACL(B)-631,ACL(B)-65I, ACL(B)-04C)
Bur Extraction Force(N)Max. TorquesMax. Water PressureMax. Speed in rpmShank ConformanceCoupling DimensionLubricantChemical CompositionBiocompatibleDelivery system45 N80 NcmN/A40,000 rpmBy ISO 1797-1By ISO 3964N/A510k#45 N50 NcmN/A50,000 rpm, 35,000 rpmBy ISO 1797-1By ISO 3964N/A
SterilizationSteam Heat 132 °C / 4minutesSteam Heat 132 °C / 4minutes
Operating principleThe TRAUS Dental Handpieces are gear drivenhand-held dental handpieces with gear ratio of20:1. It can be driven by torque adjustableelectrical motors for surgery treatment. It isattached to drive via ISO 3964 coupling. Thehead clamp accepts instrument complying withISO 1797-1. They have contra angle attachmentfor difficult to reach areas intended to preparedental cavities for restorations, such as fillings,and for cleaning teeth.The STRONG Dental Handpieces are geardriven hand-held dental handpieces with gearratio of 20:1, 16:1, 64:1, 32:1 and 1:1. It can bedriven by torque adjustable electrical motorsfor surgery treatment. It is attached to drive viaISO 3964 coupling. The head clamp acceptsinstrument complying with ISO 1797-1. Theyhave contra angle attachment for difficult toreach areas intended to prepare dental cavitiesfor restorations, such as fillings, and forcleaning teeth.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a geometric shape made up of blue triangles above the company name, which is also in blue. Below the company name is the text "Dasa-eup, Dalseong-gun, Daegu".

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for SAESHIN, which is written in a stylized blue font. Above the company name is a geometric blue logo. Below the logo is the address: # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu.

23921, Korea
Tel: +82 53 587 2345 / Fax: +82 53 580 0939

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the contact information for a company called SAESHIN. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345, and the fax number is +82 53 580 0939.

10. Summary of Non-Clinical Performance Data

Biocompatibility

Patient contact materials of the TRAUS Dental Handpieces are as follows:

1) TRAUS CRB46LN, TRAUS CRB46NN

PartMaterialSurfacetreatmentPatientcontactDuration of Contactby ISO 10993-1Bio-compatibility
Head andnozzleSUS303FHard ChromeCoatedYesLimited (<24 hours)Yes
ChuckSUS420FVacuum heattreatmentYesLimited (<24 hours)Yes
  • Cytotoxicity per ISO 10993-5
  • . Sensitization per ISO 10993-10
  • . Irritation per ISO 10993-10

Bench Testing

The bench tests were performed in order to ensure the performance of the TRAUS Dental Handbieces: TRAUS CRB46LN and TRAUS CRB46NN verify conformity to ISO 14457 and demonstrate substantial equivalence to the predicates.

TRAUS CRB46LN and TRAUS CRB46NN samples were compliant with ISO 14457: 2017 Dentistry - Handpieces and Motors and demonstrated substantial equivalence to the predicates.

I Sterilization

The TRAUS Dental handpieces, TRAUS CRB46LN and TRAUS CRB46NN are provided nonsterile and labeled for sterilization by the end user, the Instructions for use include the following parameters:

Sterilization TypeTemperatureTimeLoadCharacteristicsDry Time
Steam Sterilization(Pre-vacuum Type)132 °C4 min.Sterilization Bag30 min.

The sterilization method was validated per :

  • . Reprocessing validation per the FDA Guidance Document entitled, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" dated March 17, 2015
  • Cleaning Process Validation per ASTM F2314 and ISO 15883-5
  • Sterilization Validation per ISO 17665-1 and ISO 17665-2

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The contact information includes the address: #52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea, as well as the telephone number: +82 53 587 2345 and fax number: +82 53 580 0939.

11.Conclusion [21 CFR 807.92(b)(3)]

In accordance with the Federal Food, Drug and Cosmetic, 21 CFR Part 807, and based on the information provided in this premarket notification Saeshin Precision Co., Ltd. Concludes that the TRAUS Dental Handpieces are substantially equivalent to predicate devices as described herein.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.