(88 days)
Not Found
No
The description focuses on traditional signal processing and spectroscopy techniques, with no mention of AI or ML algorithms for data analysis or interpretation.
No.
The device is a monitor used to measure regional hemoglobin oxygen saturation; it does not treat or directly alleviate a medical condition.
Yes
The device is intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood, which involves collecting data to assess a physiological state, fitting the definition of a diagnostic device.
No
The device description explicitly states that the device consists of hardware components including a multi-channel touch screen display, preamplifier, cables, and a single use sensor, in addition to utilizing near infrared diffuse reflectance spectroscopy technology.
Based on the provided information, the INVOS™ Patient Monitor, model PM7100, is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use describes the device as a "noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor." This involves measuring physiological parameters directly from the patient's body.
- Device Description: The device description details a system that uses near infrared light applied to the skin to measure oxygen saturation in tissue. This is a non-invasive method applied to the living body.
- Lack of In Vitro Activity: An IVD device is specifically designed to examine specimens derived from the human body (like blood, urine, tissue samples) outside of the body to provide information about a physiological state, disease, or condition. The INVOS™ system does not process or analyze samples in vitro.
The INVOS™ system is a non-invasive physiological monitoring device that measures parameters directly from the patient's body.
N/A
Intended Use / Indications for Use
The INVOS™ Patient Monitor, model PM7100, is a noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use on individuals greater than 40 kg at risk for reduced-flow or no-flow ischemic states.
The INVOS™ Adult rSO2 Sensor is indicated for single patient use when cerebral/somatic monitoring of site-specific regional oxygen saturation (rSO2) is required in patients weighing >40 kilograms. This sensor is only intended to be used with INVOS™ Near Infrared Spectroscopy (NIRS) technology including monitoring systems and devices integrated with INVOS™ NIRS technology. For additional information regarding setup and use of the INVOS™ System including indications for use, contraindications, warnings and cautions, consult the Monitoring System Operator's Manual.
Product codes (comma separated list FDA assigned to the subject device)
MUD, QEM
Device Description
The INVOS™ PM7100 Patient Monitor is a cerebral/somatic tissue oximeter that utilizes a near infrared diffuse reflectance spectroscopy system employing near infrared light at four wavelengths. Two wavelengths are used to estimate the percentage of hemoglobin saturated with oxygen in tissue underneath the sensor. Near infrared light at two additional wavelengths are used for the sensor on/off detection algorithm. The subject device is non-sterile and consists of a multi-channel touch screen display, preamplifier, cables, and a single use sensor. The device utilizes up to four detachable sensors to collect signals. Up to two preamplifiers receive signals from the sensors, digitize the signals, process the data and then estimate the current rSO2 at each sensor site. The preamplifiers then transmit the measured parameter data to the monitor where the information is displayed.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Near infrared diffuse reflectance spectroscopy / Near Infrared Spectroscopy (NIRS)
Anatomical Site
Brain or in other tissue beneath the sensor.
Indicated Patient Age Range
Individuals greater than 40 kg / patients weighing > 40 kilograms.
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
NONCLINICAL TESTING:
The INVOS™ PM7100 Patient Monitor was designed and bench tested against the performance categories noted below. When available, the applicable standards and guidance documents were used for design and testing in support of substantial equivalence.
Bench Test Categories: Essential Performance, Electrical Safety, Electromagnetic Compatibility, Usability, Alarm, Optical Safety, Software, Biocompatibility, Mechanical and Environmental, Cybersecurity.
CLINICAL TESTING:
Three clinical studies were conducted and completed: rSO2 trend equivalence, sensor backward compatibility and sensor on/off.
- An rSO2 trend equivalence clinical study was performed to demonstrate the rSO2 trend precision between the INVOS™ PM7100 and the predicate device. A secondary analysis of the data for the system comparison of accuracy and precision to support device equivalency was conducted and demonstrated that rSO2 values of subject and predicate are comparable.
- A sensor backwards compatibility clinical test was performed to demonstrate that the new PMSENS71-A sensor trend accuracy is equivalent to the trend accuracy of the SAFB-SM sensor when used with the INVOS™ 5100C. A secondary analysis of the system comparison of accuracy and precision to support PMSENS71-A sensor equivalency was conducted and demonstrated that rSO2 values of both sensors are comparable.
- A sensor on/off detection clinical study was conducted to demonstrate that the new sensor and the new sensor on/off algorithm are capable of detecting sensor off conditions on clinical study subjects.
Key results:
The clinical studies paired with a bench study using simulated patient data demonstrate that the INVOS™ PM7100 Patient Monitor meets the same trend performance as the predicate device, the new sensor on/off feature functions as intended, and the new sensor is backwards compatible with the predicate device and is therefore deemed substantially equivalent to the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Covidien llc Samir Ibrahim Regulatory Affairs Manager 6135 Gunbarrel Avenue Boulder, Colorado 80301
January 8, 2019
Re: K182868
Trade/Device Name: INVOS PM7100 Patient Monitor, INVOS Adult rSO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD. OEM Dated: October 11, 2018 Received: October 12, 2018
Dear Samir Ibrahim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Neil R Neil R Ogden -S Ogden -S 10:02:10 -05'00'
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K182868
Device Name INVOS™ PM7100 Patient Monitor INVOSTM Adult rSO2 Sensor
Indications for Use (Describe)
The INVOS™ Patient Monitor, model PM7100, is a noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use on individuals greater than 40 kg at risk for reduced-flow or no-flow ischemic states.
The INVOS™ Adult rSO2 Sensor is indicated for single patient use when cerebral/somatic monitoring of site-specific regional oxygen saturation (rSO2) is required in patients weighing >40 kilograms. This sensor is only intended to be used with INVOS™ Near Infrared Spectroscopy (NIRS) technology including monitoring systems and devices integrated with INVOS™ NIRS technology. For additional information regarding setup and use of the INVOS™ System including indications for use, contraindications, warnings and cautions, consult the Monitoring System Operator's Manual.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801.109) |
| X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, provided is the 510(k) summary for INVOS™ PM7100 Patient Monitor.
SUBMITTER INFORMATION:
| Submitted By: | Covidien IIc
6135 Gunbarrel Avenue
Boulder, CO 80301 |
|----------------------|------------------------------------------------------------|
| Contact: | Samir Ibrahim
Manager Regulatory Affairs |
| Phone: | (720) 253-2752 |
| Fax: | (303) 305-2212 |
| Date of Preparation: | October 11, 2018 |
Device Name:
Trade Name(s): | INVOS TM PM7100 Patient Monitor |
---|---|
Common/Usual Name: | Cerebral Somatic Tissue Oximeter |
Classification: | Class II |
Classification Name: | Oximeter |
CFR Reference: | 21 CFR 870.2700 |
Product Code: | MUD and QEM |
PREDICATE DEVICE: | |
Manufacturer: | Covidien llc |
Manufacturer: | Covidien llc |
---|---|
Device Name: | INVOS TM Cerebral/Somatic Oximeter System, Model 5100C |
510(k) Number: | K082327 |
Clearance Date: | April 3, 2009 |
Device Description:
The INVOS™ PM7100 Patient Monitor is a cerebral/somatic tissue oximeter that utilizes a near infrared diffuse reflectance spectroscopy system employing near infrared light at four wavelengths. Two wavelengths are used to estimate the percentage of hemoglobin saturated with oxygen in tissue underneath the sensor. Near infrared light at two additional wavelengths are used for the sensor on/off detection algorithm.
The subject device is non-sterile and consists of a multi-channel touch screen display, preamplifier, cables, and a single use sensor.
4
The device utilizes up to four detachable sensors to collect signals. Up to two preamplifiers receive signals from the sensors, digitize the signals, process the data and then estimate the current rSO2 at each sensor site. The preamplifiers then transmit the measured parameter data to the monitor where the information is displayed.
Accessories:
INVOSTM Adult rSO2 Sensor, PM7100 and 5100C | PMSENS71-A |
---|---|
INVOSTM Reusable Sensor Cable for PM7100 | PMAC71RSC |
INVOSTM Docking Station, PM7100 | PMAC71DOC |
INVOSTM Patient Monitor Stand, PM7100 | PMAC71STAND |
INVOSTM Reusable Sensor Cable for 5100C, Channel 1 | PMAC71RSC-L-CH1 |
INVOSTM Reusable Sensor Cable for 5100C, Channel 2 | PMAC71RSC-L-CH2 |
INVOSTM Reusable Sensor Cable for 5100C, Channel 3 | PMAC71RSC-L-CH3 |
INVOSTM Reusable Sensor Cable for 5100C, Channel 4 | PMAC71RSC-L-CH4 |
INDICATIONS FOR USE:
The INVOS™ Patient Monitor, model PM7100, is a noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use on individuals greater than 40 kg at risk for reduced-flow or no-flow ischemic states.
The INVOS™ Adult rSO2 Sensor is indicated for single patient use when cerebral/somatic monitoring of site-specific regional oxygen saturation (rSO2) is required in patients weighing > 40 kilograms. This sensor is only intended to be used with INVOS™ Near Infrared Spectroscopy (NIRS) technology including monitoring systems and devices integrated with INVOS™ NIRS technology. For additional information regarding setup and use of the INVOS™ System including indications for use, contraindications, warnings and cautions, consult the Monitoring System Operator's Manual.
CONTRAINDICATIONS
The INVOS™ Adult rSO2 sensor is contraindicated for use on patients who exhibit allergic reactions to the adhesive tape.
TECHNOLOGICAL CHARACTERISTICS SUMMARY:
The INVOS™ PM7100 Patient Monitor is a four wavelength, diffuse reflectance spectroscopy system employing near infrared light. Two wavelengths of near infrared light are used to estimate the percentage of hemoglobin saturated with oxygen in tissue underneath the sensor and two additional wavelengths are used for sensor on/off detection. An adhesive sensor containing four LED light sources and two photodiodes is applied to the skin over cerebral and somatic tissue and the returning light from two LEDs is analyzed for oxyhemoglobin and deoxyhemoglobin light absorption signals from the photodiode closer to the light source are subtracted from the farther photodiode where the returning photons penetrate more deeply in the tissue. This removes the impact of absorption events originating in the outer layers of tissue that are common to both photodiodes, including the effects of skin pigmentation and subcutaneous tissues.
SUBSTANTIAL EQUIVALENCE STATEMENT:
5
The INVOS™ PM7100 Patient Monitor has the same intended use and indications for the adult population > 40 kg, principles of operation and technological characteristics as the predicate device. The changes in the display, preamplifier and senor do not raise any new questions of safety or effectiveness. Performance data demonstrates that the INVOS™ PM7100 Patient Monitor and the INVOS™ Adult rSO₂ Sensor are equivalent to the predicate.
Performance Data Summary:
NONCLINICAL TESTING:
The INVOS™ PM7100 Patient Monitor was designed and bench tested against the performance categories noted below. When available, the applicable standards and guidance documents were used for design and testing in support of substantial equivalence.
| Bench Test
Category | Guidance/Standard |
---|---|
Essential | |
Performance | EN 60601-1 |
Electrical Safety | IEC 60601-1 |
Electromagnetic | |
Compatibility | IEC 60601-1-2 |
FDA Guidance Electromagnetic Compatibility of Electrically-Powered Medical Devices | |
Usability | IEC 60601-1-6 |
IEC 62366 | |
FDA Guidance Applying Human Factors and Usability Engineering to Medical Devices | |
Draft Guidance | |
Alarm | IEC 60601-1-8 |
Optical Safety | IEC 62471 |
Software | IEC 62304 |
FDA Guidance for the Content of Premarket Submissions for Software Contained in | |
Medical Devices | |
FDA Guidance General Principles of Software Validation | |
FDA Guidance Off-The-Shelf Software Use in Medical Devices | |
Biocompatibility | ISO 10993-1 |
Mechanical and | |
Environmental | IEC 60601-1 |
Cybersecurity | FDA Guidance Content of Premarket Submissions for Management of Cybersecurity in |
Medical Devices |
CLINICAL TESTING:
Three clinical studies were conducted and completed, rSO2 trend equivalence, sensor backward compatibility and sensor on/off in support the INVOS™ PM7100 Patient Monitor.
An rSOz trend equivalence clinical study was performed to demonstrate the rSO2 trend precision between the INVOS™ PM7100 and the predicate device. A secondary analysis of the data for the system comparison of accuracy and precision to support device equivalency was conducted and demonstrated that rSO2 values of subject and predicate are comparable. The INVOS™ PM7100 maintains the same fundamental rSO2 technology as its predicate i.e., the same rSO2 algorithm and the same optical and
6
electrical signal chain.
A sensor backwards compatibility clinical test was performed to demonstrate that the new PMSENS71-A sensor trend accuracy is equivalent to the trend accuracy of the SAFB-SM sensor when used with the INVOS™ 5100C. A secondary analysis of the system comparison of accuracy and precision to support PMSENS71-A sensor equivalency was conducted and demonstrated that rSO2 values of both sensors are comparable. The test supports the use of the PMSENS71-A sensor with the INVOS™ 5100C.
A sensor on/off detection clinical study was conducted to demonstrate that the new sensor and the new sensor on/off algorithm are capable of detecting sensor off conditions on clinical study subjects.
The clinical studies paired with a bench study using simulated patient data demonstrate that the INVOS™ PM7100 Patient Monitor meets the same trend performance as the predicate device, the new sensor on/off feature functions as intended, and the new sensor is backwards compatible with the predicate device and is therefore deemed substantially equivalent to the predicate device.
CONCLUSION
Results from comprehensive verification and validation testing of a system design analysis, a bench-top tissue phantom test, trend performance validation clinical study, and a backward compatibility validation clinical study demonstrated that the INVOS™ PM7100 Patient Monitor is substantially equivalent to the predicate device with respect to clinical use case, characteristics, and performance. This equivalence allows for the subject device to support the same claims (adult only) and indications for use as the predicate device.
The result from the sensor on/off clinical study proves that the INVOS™ PM7100 Patient Monitor can effectively detect when the sensor is on and when the sensor is off the patient.
The results from the human factors validation have shown that the updated monitor/user interface with a touch screen display improves user-friendliness of the system but doesn't introduce unforeseen risk to the subject device.
The information provided in this 510(k) demonstrates the INVOS™ PM7100 Patient Monitor is substantially equivalent to the predicate device with respect to performance.