AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Single and Multi-Drug Rapid Test Cup With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

The Single and Multi-Drug Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

The Single and Multi-Drug Rapid Test Panel With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

The Single and Multi-Drug Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

The Single Drug Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

The Single and Multi-Drug Home Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

The Single and Multi-Drug Home Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, 50ng/mL, Methadone 300ng/mL, Methamphetamine Mariiuana 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

The Single Drug Home Drug Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, 50ng/mL. Methadone Marijuana 500ng/mL. Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1.5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

Device Description

The Candidate Drug Screen Tests are rapid lateral flow immunoassays in which drugprotein conjugates in the test device compete with drugs or drug metabolites that may be present in urine.

On each test strip, a drug-protein conjugate is added to the test band of the membrane known as the test region (T), and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. Anti-drug antibodies derived from sheep and/or mice are used on the candidate tests. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody- gold conjugates then complexes with the drug-protein conjugates to form visible lines.

Therefore, the formation of the visible precipitant in the test band indicates a negative result. If the target drug level exceeds its cut-off concentration, the drug/metabolite antigen competes with drug protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result. A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

AI/ML Overview

The provided text describes several rapid chromatographic immunoassay devices for the qualitative detection of drugs and drug metabolites in urine. The 510(k) summary (K182738) states that the devices are substantially equivalent to a predicate device (K173303) from Guangzhou Wondfo Biotech Co., Ltd.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" or provide a table of performance data for the submitted device (Hangzhou AllTest Biotech Co., Ltd). Instead, it states that "The results of all verification and validation activities demonstrate that the candidate device is substantially equivalent to the cleared predicate devices." This implies that the device's performance met the standards required for substantial equivalence, which would align with the predicate device's established performance. The specific performance metrics (e.g., sensitivity, specificity, accuracy) are not detailed.

However, the "Cutoff Levels" section under "Comparison to Predicate Devices" acts as a form of performance characteristic.

AnalyteCutoff Level (Candidate Device) (ng/mL)Cutoff Level (Predicate Device) (ng/mL)
Amphetamine500500
Secobarbital (Barbiturates for Predicate)300300
Benzodiazepines300300
Buprenorphine1010
Cocaine150150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300300
Methylenedioxymethamphetamine (Ecstasy/MDMA)500500
Marijuana5050
Methadone300300
Methamphetamine500500
Morphine300 and 2000300 and 2000
Phencyclidine2525
Nortriptyline (Tricyclic Antidepressants)1,0001,000
Oxycodone100100

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample size used for the test set, the country of origin of the data, or whether the study was retrospective or prospective. It summarily states "The results of all verification and validation activities demonstrate that the candidate device is substantially equivalent to the cleared predicate devices."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not specify the number or qualifications of experts used to establish the ground truth. It mentions that Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method for obtaining a confirmed analytical result, implying that GC/MS provides the ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any adjudication method.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is an immunoassay device, not an AI-powered diagnostic imaging device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This refers to the performance of the immunoassay device itself, which operates as a standalone test. The document states it is a "rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine." While the specific performance metrics are not given, the whole 510(k) submission relates to the standalone performance of this device. The phrase "This assay provides only a preliminary analytical test result" indicates its standalone (screening) utility.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The preferred confirmatory method mentioned is Gas Chromatography/Mass Spectrometry (GC/MS). This indicates that GC/MS results serve as the ground truth for confirming the presence and concentration of drugs and drug metabolites in urine samples.

8. The sample size for the training set

The document does not explicitly mention a training set or its sample size. Immunoassays typically undergo analytical validation rather than machine learning training.

9. How the ground truth for the training set was established

As there is no mention of a training set in the context of machine learning, this question is not applicable. For analytical validation of immunoassays, ground truth (if a "training set" were to be conceptualized as samples used for assay development/optimization) would be established by highly accurate methods like GC/MS.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name in blue text, with the word "FDA" in a blue square.

March 27, 2019

Hangzhou AllTest Biotech Co., Ltd % Feng-Yu Lee, Principal Regulatory Consultant IVDD Regulatory Consultant 29222 Rancho Viejo Rd., Suite 218 San Juan Capistrano, CA 92675

Trade/Device Name:Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)Single and Multi-Drug Rapid Test Panel (Urine)Single and Multi-Drug Rapid Test Cup With Adulteration (Urine)Single and Multi-Drug Rapid Test Cup (Urine)Single Drug Rapid Test Dipstick (Urine)Single and Multi-Drug Home Rapid Test Panel (Urine)Single and Multi-Drug Home Rapid Test Cup (Urine)Single Drug Home Rapid Test Dipstick (Urine)
Regulation Number:21 CFR 862.3100
Regulation Name:Amphetamine test system
Regulatory Class:Class II
Product Code:DKZ, NFT, DIS, PTH, JXM, NFV, DJG, NGL, DIO, NFY, DJR, PTG, DJC, NGG, LDJ,NFW, LAF, DNK, NGI, LCM, NGM, LFG, QAW
Dated:February 18, 2019
Received:February 19, 2019

Dear Feng-Yu Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K182738

Device Name

Single and Multi-Drug Rapid Test Cup With Adulteration (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Rapid Test Cup With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

Type of Use ( Select one or both, as applicable )
----------------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K182738

Device Name Single and Multi-Drug Rapid Test Cup (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K182738

Device Name

Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Rapid Test Panel With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

Type of Use ( Select one or both, as applicable )
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K182738

Device Name

Single and Multi-Drug Rapid Test Panel (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known) K182738

Device Name Single Drug Rapid Test Dipstick (Urine)

Indications for Use (Describe)

The Single Drug Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine

150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known) K182738

Device Name

Single and Multi-Drug Home Rapid Test Cup (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Home Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter Use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known) K182738

Device Name

Single and Multi-Drug Home Rapid Test Panel (Urine)

Indications for Use (Describe)

The Single and Multi-Drug Home Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter Use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{9}------------------------------------------------

510(k) Number (if known) K182738

Device Name Single Drug Home Rapid Test Dipstick (Urine)

Indications for Use (Describe)

The Single Drug Home Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/ mL, and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter Use.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{10}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K182738

1. Submitter's Identification:

Hangzhou AllTest Biotech Co., Ltd #550, Yinhai Street Hangzhou Economic & Technological Development Area, Hangzhou 310018, P.R. China

c/o IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-949-218-0928 Date Summary Prepared: March 27, 2019

Name of the Device: 2.

For Prescription Use

Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)

Single and Multi-Drug Rapid Test Panel (Urine)

Single and Multi-Drug Rapid Test Cup With Adulteration (Urine)

Single and Multi-Drug Rapid Test Cup (Urine)

Single Drug Rapid Test Dipstick (Urine)

{11}------------------------------------------------

For Over-the-Counter

Single and Multi-Drug Home Rapid Test Panel (Urine)

Single and Multi-Drug Home Rapid Test Cup (Urine)

Single Drug Home Rapid Test Dipstick (Urine)

3. Common or Usual Name: Drugs of Abuse Test

Panel: Toxicology

AnalyteProductCodeProfessional /OTCRegulation Section
AmphetamineDKZ / NFT21CFR 862.3100,Amphetamine Test System
SecobarbitalDIS / PTH21 CFR 862.3150,Barbiturates Test System
BenzodiazepinesJXM / NFV21 CFR 862.3170,Benzodiazepines Test System
BuprenorphineDJG / NGL21CFR 862.3650, OpiateTest System
CocaineDIO / NFY21 CFR 862.3250, Cocaineand metabolites Test System
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidineDJR / PTG21 CFR 862.3620,Methadone Test System
MethylenedioxymethamphetamineDJC / NGG21 CFR 862.3610,Methamphetamine Test
MarijuanaLDJ / NFW21 CFR 862.3870,Cannabinoids Test System
MethadoneDJR / PTG21 CFR 862.3620,Methadone Test System
MethamphetamineLAF / NGG21 CFR 862.3610,Methamphetamine Test
Morphine (300 and 2000)DNK / NGI21 CFR 862.3640,Morphine Test System

{12}------------------------------------------------

OxycodoneDJG / NGL21 CFR 862.3650, OpiateTest System
PhencyclidineLCM / NGMUnclassified, EnzymeImmunoassay Phencyclidine
NortriptylineLFG / QAW21 CFR 862.3910,Tricyclic AntidepressantDrugs Test System

4. Device Description:

The Candidate Drug Screen Tests are rapid lateral flow immunoassays in which drugprotein conjugates in the test device compete with drugs or drug metabolites that may be present in urine.

On each test strip, a drug-protein conjugate is added to the test band of the membrane known as the test region (T), and the anti-drug antibody-colloidal gold conjugate pads are placed at the forward end of the membrane. Anti-drug antibodies derived from sheep and/or mice are used on the candidate tests. If target drugs are present in the urine specimen below its cut-off concentration, the solution of the colored antibody-colloidal gold conjugates moves along with the sample solution by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody- gold conjugates then complexes with the drug-protein conjugates to form visible lines.

Therefore, the formation of the visible precipitant in the test band indicates a negative result. If the target drug level exceeds its cut-off concentration, the drug/metabolite antigen competes with drug protein conjugates on the test band region for the limited antibody on the colored drug antibody-colloidal gold conjugate pad. The drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result. A band should form in the control region (C) of the devices regardless of the presence of drug in the sample to indicate that the test has been performed properly.

ട്. Intended Use:

(I) For Prescription Use

- Single and Multi-Drug Rapid Test Cup With Adulteration (Urine)

The Single and Multi-Drug Rapid Test Cup With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL.

{13}------------------------------------------------

Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL. Nortriptyline 1.000ng/mL. Oxycodone 100ng/mL. and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

- Single and Multi-Drug Rapid Test Cup (Urine)

The Single and Multi-Drug Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

- Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)

The Single and Multi-Drug Rapid Test Panel With Adulteration (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

{14}------------------------------------------------

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

- Single and Multi-Drug Rapid Test Panel (Urine)

The Single and Multi-Drug Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Prescription Use.

- Single Drug Rapid Test Dipstick (Urine)

The Single Drug Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following concentrations: cut-off 300ng/mL. Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana Methadone 50ng/mL, Methamphetamine 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

{15}------------------------------------------------

For Prescription Use.

(II) For Over-the-Counter

- Single and Multi-Drug Home Rapid Test Cup (Urine)

The Single and Multi-Drug Home Rapid Test Cup (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, Marijuana 50ng/mL, Methadone 300ng/mL, Methamphetamine 500ng/mL. Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

- Single and Multi-Drug Home Rapid Test Panel (Urine)

The Single and Multi-Drug Home Rapid Test Panel (Urine) are rapid chromatographic immunoassay for the qualitative detection of single or multiple drugs and drug metabolites in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, 50ng/mL, Methadone 300ng/mL, Methamphetamine Mariiuana 500ng/mL, Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

- Single Drug Home Rapid Test Dipstick (Urine)

{16}------------------------------------------------

The Single Drug Home Drug Rapid Test Dipstick (Urine) are rapid chromatographic immunoassay for the qualitative detection of individual drug and drug metabolite(s) in urine at the following cut-off concentrations: Amphetamine 500ng/mL, Secobarbital 300ng/mL, Benzodiazepines 300ng/mL, Buprenorphine 10ng/mL, Cocaine 150ng/mL, 50ng/mL. Methadone Marijuana 500ng/mL. Methylenedioxymethamphetamine 500ng/mL, Morphine 300ng/mL and 2,000ng/mL, Phencyclidine 25ng/mL, Nortriptyline 1,000ng/mL, Oxycodone 100ng/mL, and 2ethylidene-1.5-dimethyl-3,3-diphenylpyrrolidine 300ng/mL.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

For Over-The-Counter use.

6. Predicate Device Information:

The subject devices are substantially equivalent to the following 510(k) cleared device noted below.

Device Company:
510(K) Number:GUANGZHOU WONDFO BIOTECH CO., LTD.; andK173303

Device Name:

INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)

INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use)

    1. Comparison to Predicate Devices (Substantial Equivalence Comparison):
Similarities and Differences
ItemCandidate Device(s)Predicate Device(s)
Intended UseSameFor the qualitative detection ofdrugs of abuse in human
ResultsSameQualitative

{17}------------------------------------------------

MethodologySameLateral flow, competitive bindingimmunoassay based on the principleof antigen and antibody
IntendedUsersSamePrescription and over the counterusers
CutoffLevels(ng/mL)Amphetamine – 500Secobarbital - 300Benzodiazepines – 300Buprenorphine – 10Cocaine - 150Marijuana - 50Methadone - 300Methamphetamine – 500Methylenedioxymethamphetamine– 500Morphine - 300 and 2000Phencyclidine – 25Nortriptyline - 1,000Oxycodone - 1002-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine – 300Amphetamine – 500Barbiturates – 300Benzodiazepines – 300Buprenorphine – 10Cocaine - 150EDDP - 300Ecstasy (MDMA) – 500Marijuana - 50Methadone – 300Methamphetamine – 500Morphine - 300Opiates - 2000Oxycodone - 100Phencyclidine – 25Tricyclic Antidepressants – 1000
FormatCup, Panel / DipstickCup, Panel / Dip Card, Cassette,and Dipstick

Conclusions from Verification and Validation Activities: 8.

The results of all verification and validation activities demonstrate that the candidate device is substantially equivalent to the cleared predicate devices.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).