AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites: Amphetamines, Barbiturates, Buprenorphine, Benzodiazepines, Cocaine, Methamphetamine, Methadone, Phencyclidine, Tricyclic Antidepressants, Cannabinoids, MDMA, Morphine, Oxycodone. These assays may yield positive results when barbiturates, benzodiazepines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs. This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

The INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites: Amphetamines, Barbiturates, Buprenorphine, Benzodiazepines, Cocaine, Methamphetamine, Methadone, Phencyclidine, Tricyclic Antidepressants, Cannabinoids, MDMA, Morphine, Oxycodone. These assays may yield positive results when barbiturates, benzodiazenines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs. This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Device Description

These devices are one-step lateral flow chromatographic immunoassays consisting of any combination of one (1) to thirteen (13) individual test strip(s). Each test strip in the device consists of 1) a conjugate pad containing colloidal gold coupled with the anti-drug antibodies and 2) nitrocellulose membrane containing a test line (T line) coated with the conjugated drug antigen and a control line (C line). The C line serves as an internal quality control of the system and appears as a burgundy-colored band during the test regardless of the presence of the drug.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes two types of performance studies: a "precision" study (analytical performance) and a "lay user study" (user performance for OTC use). The acceptance criteria for the precision study appear to be implicit in the results, suggesting that all negative samples below a certain threshold should be negative, and all positive samples above a certain threshold should be positive. For the lay user study, the acceptance criteria are for the interpretation of results by the lay users, implicitly aiming for high agreement with the expected result.

Analytical Performance (Precision Study - Example for Amphetamines)

AnalyteConcentration (ng/mL)% of CutoffExpected ResultReported Results (Neg/Pos) Lot 1Reported Results (Neg/Pos) Lot 2Reported Results (Neg/Pos) Lot 3Acceptance Criteria (Implicit)
Amphetamines00%Negative50/050/050/0100% negative
250-75%Negative50/050/050/0100% negative
500-50%Negative50/050/050/0100% negative
750-25%Negative50/050/050/0100% negative
1000CutoffMixed/Transition2/481/491/49Expected to show a transition from negative to positive results, with some variability around the cutoff.
1250+25%Positive0/500/500/50100% positive
1500+50%Positive0/500/500/50100% positive
1750+75%Positive0/500/500/50100% positive
2000+100%Positive0/500/500/50100% positive
Summary of performance from table for analytes (e.g. Amphetamines): For concentrations at or below -25% of cutoff, all results were negative (150/0 across 3 lots). For concentrations at or above +25% of cutoff, all results were positive (150/0 across 3 lots). At the cutoff point (1000 ng/mL), a mix of negative and positive results were observed across the three lots (e.g., Lot 1: 2 negative, 48 positive), indicating sensitivity around the cutoff as expected for a qualitative assay.

Lay User Study Performance (Example for Amphetamines)

DrugConcentration (ng/ml)Expected ResultNumber of samplesReported NegativeReported PositiveAcceptance Criteria (Implicit)
AMP (1000)0Negative3503500High accuracy (e.g., >95%) in interpretation for clearly negative samples.
750 (75%)Negative1091High accuracy in interpretation for samples below cutoff. A small number of false positives might be acceptable, but ideally none.
1250 (125%)Positive1019High accuracy in interpretation for samples above cutoff. A small number of false negatives might be acceptable, but ideally none.
1500 (150%)Positive30030High accuracy in interpretation for clearly positive samples.
Summary of performance from table for AMP: For 0% concentration, 100% negative results (350/350). For 75% concentration, 90% negative (9/10). For 125% concentration, 90% positive (9/10). For 150% concentration, 100% positive (30/30). This generally indicates accurate interpretation by lay users.
Surveys and labeling assessments: For all categories (intended use, directions for test, performing test, interpreting results), a very high percentage of participants rated the ease of understanding as "very easy to understand" or "easy to understand" (96.5% to 100%). This suggests the labeling and instructions were clear for lay users.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Analytical/Precision Study Test Set:

    • For each of the 13 analytes, there were 9 different concentration levels tested.
    • For each concentration level, 50 samples were tested per Lot (Neg/Pos).
    • There were 3 Lots, for a total of 150 tests per concentration level per analyte.
    • Total tests per analyte: 9 concentrations * 50 samples/conc/lot * 3 lots = 1350 tests.
    • Total tests across all 13 analytes: 1350 tests/analyte * 13 analytes = 17,550 tests.
    • Data Provenance: Not explicitly stated, but typically these types of precision studies are conducted in a controlled laboratory setting (e.g., by the manufacturer). It is a retrospective analysis of device performance under controlled conditions.
  • Lay User Study Test Set:

    • For each analyte, various concentrations at 0%, 75%, 125%, and 150% of the cutoff were tested.
    • The number of samples varied per concentration level per analyte. For example, for Amphetamines, there were 350 samples at 0%, 10 at 75%, 10 at 125%, and 30 at 150%. This totals 400 samples for Amphetamines.
    • Since there were 13 analytes, this would suggest a very large number of total tests if each participant tested all analytes. However, the text states "each participant was provided one (1) package insert, one (1) blind labeled test solution, and one (1) test device. Test solutions were randomly picked for participants, one for each." This implies each participant (total 400) tested one specific drug concentration. Therefore, the overall sample size for the lay user study is 400 participants/tests.
    • Data Provenance: The study was conducted with 400 human participants with "diverse educational backgrounds and ranged in age from 18 to >60." This is prospective data collection with human subjects. The country of origin is not explicitly stated, but given it's an FDA submission, it's typically in the US or in a country following similar regulatory standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Analytical/Precision Study: The ground truth for the test concentrations (e.g., 0 ng/mL, 250 ng/mL, cutoff, etc.) is established by laboratory standards and precise dilution methods, not by human experts. The concentrations were precisely prepared.
  • Lay User Study: Similar to the analytical study, the ground truth for these samples was based on the known, prepared concentrations of the drug within the test solutions. No human experts were used to establish the ground truth; it was based on the controlled spiking of urine samples.

4. Adjudication Method for the Test Set

  • Analytical/Precision Study: No adjudication method involving multiple readers is mentioned. The results were recorded as Neg/Pos based on the device's output and presumably read by the operators performing the tests.
  • Lay User Study: No formal adjudication method is mentioned. Each participant performed one test and reported their result. The study assessed how well the individual lay users interpreted the result, rather than having multiple readers interpret the same result for adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. This device is a rapid, qualitative immunoassay for drug detection, and the studies performed focus on its analytical performance and lay user interpretability, not on reader improvement with AI assistance. The device itself is not an AI-assisted diagnostic tool for human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • This device is not an AI algorithm. It is a chemical immunoassay device. The results are visually read by a human. Therefore, a "standalone" algorithmic performance study without human-in-the-loop is not applicable in the context of AI. However, the "precision study" can be considered a standalone performance assessment of the device's chemical reactions, irrespective of user interpretation (though a human still reads the lines).

7. Type of Ground Truth Used

  • Analytical/Precision Study: The ground truth was based on known, precisely prepared drug concentrations in the urine matrix. This is a form of "spiked" samples, where the true concentration is known by the experimenters.
  • Lay User Study: The ground truth was also based on known, precisely prepared drug concentrations in the test solutions given to participants.

8. Sample Size for the Training Set

  • This device is not an AI/ML algorithm that requires a training set. It is a chemical immunoassay. Therefore, the concept of a "training set" is not applicable.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

February 23, 2018

Alfa Scientific Designs, Inc Naishu Wang CTO 13200 Gregg Street Poway, CA 92064

Re: K173303

Trade/Device Name: INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ, DIS, DJG, JXM, DIO, DJC, DJR, LCM, LFG, LDJ, DNK, NFT, PTH, NFV, NFY, NGG, PTG, NGM, NFW, NGI, NGL, QAW Dated: January 16, 2018 Received: January 22, 2018

Dear Naishu Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173303

Device Name

INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)

Indications for Use (Describe)

The INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites:

AnalyteCalibratorCutoff (ng/mL)
Amphetaminesd-Amphetamine1000
BarbituratesSecobarbital200
BuprenorphineBuprenorphine10
BenzodiazepinesOxazepam300
CocaineBenzoylecgonine300
Methamphetamined-Methamphetamine1000
MethadoneMethadone300
PhencyclidinePhencyclidine25
TricyclicAntidepressantsNortriptyline1000
Cannabinoids11-nor-Δ9-THC-9-COOH50
MDMAMethylenedioxymethamphetamine500
MorphineMorphine2000
OxycodoneOxycodone300

These assays may yield positive results when barbiturates, benzodiazepines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs.

This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K173303

Device Name

INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use)

Indications for Use (Describe)

The INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites:

AnalyteCalibratorCutoff (ng/mL
Amphetaminesd-Amphetamine1000
BarbituratesSecobarbital200
BuprenorphineBuprenorphine10
BenzodiazepinesOxazepam300
CocaineBenzoylecgonine300
Methamphetamined-Methamphetamine1000
MethadoneMethadone300
PhencyclidinePhencyclidine25
Tricyclic
AntidepressantsNortriptyline1000
Cannabinoids11-nor-A9-THC-9-COOH50
MDMAMethylenedioxymethamphetamine500
MorphineMorphine2000
OxycodoneOxycodone300

These assays may yield positive results when barbiturates, benzodiazenines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs.

This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

510(k) Summary

Safety and effectiveness as required by 21 CFR 807.92

Manufacturerand SubmitterName:Alfa Scientific Designs, Inc.
Address:13200 Gregg StreetPoway, CA 92064Telephone: 858-513-3888Fax: 858-513-8388
Contact Person:Naishu Wang, MD, PhDEmail: wnss@alfascientific.com
Device NameTrade Name:INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use)
Common Name:Immunoassay, Drug of Abuse Screening Urine Test
Classification Name:Amphetamine Test System, Barbiturate Test System,Benzodiazepine Test System, Cocaine and Cocaine MetaboliteTest System, Methamphetamine Test System, Morphine TestSystem, Cannabinoid Test System, Methadone Test System,Tricyclic Antidepressant Drugs Test System, Phencyclidine TestSystem, Opiate Test System
AnalyteRegulationNumberProduct code(OTC use)Product code(prescription use)
Amphetamines862.3100NFTDKZ
Barbiturates862.3150PTHDIS
Buprenorphine862.3650NGLDJG
Benzodiazepines862.3170NFVJXM
Cocaine862.3250NFYDIO
Methamphetamine862.3610NGGDJC
Methadone862.3620PTGDJR
PhencyclidineunclassifiedNGMLCM
TricyclicAntidepressants862.3910QAWLFG
Cannabinoids862.3870NFWLDJ
MDMA862.3610NGGDJC
Morphine862.3640NGIDNK
Oxycodone862.3650NGIDJG

{7}------------------------------------------------

Date ofPreparation02/22/2018
PredicateDevicesK152122Instant-View Multi-Drug Urine Test Cup(Home Use), Instant-View Multi-DrugUrine Test Panel (Home Use)
DeviceDescriptionThese devices are one-step lateral flow chromatographic immunoassaysconsisting of any combination of one (1) to thirteen (13) individual teststrip(s). Each test strip in the device consists of 1) a conjugate padcontaining colloidal gold coupled with the anti-drug antibodies and 2)nitrocellulose membrane containing a test line (T line) coated with theconjugated drug antigen and a control line (C line). The C line serves as aninternal quality control of the system and appears as a burgundy-coloredband during the test regardless of the presence of the drug.

Intended Use

The INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites:

AnalyteCalibratorCutoff (ng/mL)
Amphetaminesd-Amphetamine1000
BarbituratesSecobarbital200
BuprenorphineBuprenorphine10
BenzodiazepinesOxazepam300
CocaineBenzoylecgonine300
Methamphetamined-Methamphetamine1000
MethadoneMethadone300
PhencyclidinePhencyclidine25
Tricyclic AntidepressantsNortriptyline1000
Cannabinoids11-nor-Δ9-THC-9-COOH50
MDMAMethylenedioxy-methamphetamine500
MorphineMorphine2000
OxycodoneOxycodone300

These assays may yield positive results when barbiturates, benzodiazepines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs.

This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical

{8}------------------------------------------------

consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

The INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) device is a rapid, qualitative immunoassay device for the detection of one or more drugs or metabolites at the designated cutoff concentrations in human urine. The device can detect up to 13 drugs or their metabolites:

AnalyteCalibratorCutoff (ng/mL)
Amphetaminesd-Amphetamine1000
BarbituratesSecobarbital200
BuprenorphineBuprenorphine10
BenzodiazepinesOxazepam300
CocaineBenzoylecgonine300
Methamphetamined-Methamphetamine1000
MethadoneMethadone300
PhencyclidinePhencyclidine25
Tricyclic AntidepressantsNortriptyline1000
Cannabinoids11-nor-Δ9-THC-9-COOH50
MDMAMethylenedioxy-methamphetamine500
MorphineMorphine2000
OxycodoneOxycodone300

These assays may yield positive results when barbiturates, benzodiazepines, or tricyclic antidepressants are ingested at or above therapeutic doses. There are no uniformly recognized cutoff levels for barbiturates, benzodiazepines, or tricyclic antidepressants in urine. The assays are not intended to distinguish between prescription use or abuse of these drugs.

This device provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

{9}------------------------------------------------

Comparison to Predicate Devices

Similarities
ItemDevicePredicate
Intended UseSameQualitative detection ofdrugs of abuse in urine
Test PrincipleSameLateral flowimmunochromatographic
MatrixSameUrine
Number of strips per deviceSame1 – 13 depending uponconfiguration
Analyte cutoffs (ng/mL)SameAMP - 1000BAR – 200BUP – 10BZD – 300COC - 300MDMA - 500MET – 1000MTD – 300MOR – 2000OXY - 300PCP – 25TCA - 1000THC – 50
Shelf lifeSame24 months
Differences
ItemDevicePredicate
Sample application procedureUser urinates into cupand urine contacts thetest strips immediately.User urinates into cup, butthe urine sample does notcontact test strips until aknob is pushed, allowingthe sample to flow to thebottom of the cup.
Storage conditionsSame15° – 30° C

{10}------------------------------------------------

Performance Characteristics

To assess precision, each analyte was tested at the following concentrations: Negative, -75%, -50%, -25%, cutoff, +25%, +50%, +75%, and +100% of the cutoff for each analyte. The panels were blinded and randomized prior to testing. Testing was performed using three lots of test strips and was performed by 10 operators over ten non- consecutive days, and the results of this testing are summarized as follows for each analyte.

Amphetamines

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
250-75%50/050/050/0
500-50%50/050/050/0
750-25%50/050/050/0
1000cutoff2/481/491/49
1250+25%0/500/500/50
1500+50%0/500/500/50
1750+75%0/500/500/50
2000+100%0/500/500/50

Barbiturates

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
50-75%50/050/050/0
100-50%50/050/050/0
150-25%50/050/050/0
200cutoff1/493/472/48
250+25%0/500/500/50
300+50%0/500/500/50
350+75%0/500/500/50
400+100%0/500/500/50

{11}------------------------------------------------

Buprenorphine

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
2.5-75%50/050/050/0
5-50%50/050/050/0
7.5-25%50/050/050/0
10cutoff1/492/482/48
12.5+25%0/500/500/50
15+50%0/500/500/50
17.5+75%0/500/500/50
20+100%0/500/500/50

Benzodiazepines

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
75-75%50/050/050/0
150-50%50/050/050/0
225-25%50/050/050/0
300cutoff2/482/482/48
375+25%0/500/500/50
450+50%0/500/500/50
525+75%0/500/500/50
600+100%0/500/500/50

Cocaine

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
75-75%50/050/050/0
150-50%50/050/050/0
225-25%50/050/050/0
300cutoff1/492/481/49
375+25%0/500/500/50
525+75%0/500/500/50
600+100%0/500/500/50

{12}------------------------------------------------

Methamphetamine

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
250-75%50/050/050/0
500-50%50/050/050/0
750-25%50/050/050/0
1000cutoff1/492/481/49
1250+25%0/500/500/50
1500+50%0/500/500/50
1750+75%0/500/500/50
2000+100%0/500/500/50

Methadone

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
75-75%50/050/050/0
150-50%50/050/050/0
225-25%50/050/050/0
300cutoff2/482/482/48
375+25%0/500/500/50
450+50%0/500/500/50
525+75%0/500/500/50
600+100%0/500/500/50

Phencyclidine

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
6.25-75%50/050/050/0
12.5-50%50/050/050/0
18.75-25%50/050/050/0
25cutoff1/492/482/48
31.25+25%0/500/500/50
37.5+50%0/500/500/50
43.75+75%0/500/500/50
50+100%0/500/500/50

{13}------------------------------------------------

Tricyclic Antidepressants

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
250-75%50/050/050/0
500-50%50/050/050/0
750-25%50/050/050/0
1000cutoff3/472/481/49
1250+25%0/500/500/50
1500+50%0/500/500/50
1750+75%0/500/500/50
2000+100%0/500/500/50

Cannabinoids

Cannabinoids
Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
12.5-75%50/050/050/0
25-50%50/050/050/0
37.5-25%50/050/050/0
50cutoff1/491/491/49
62.5+25%0/500/500/50
75+50%0/500/500/50
87.5+75%0/500/500/50
100+100%0/500/500/50

MDMA

MDMA
Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
125-75%50/050/050/0
250-50%50/050/050/0
375-25%50/050/050/0
500cutoff3/472/482/48
625+25%0/500/500/50
750+50%0/500/500/50
875+75%0/500/500/50
1000+100%0/500/500/50

{14}------------------------------------------------

Morphine
Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
500-75%50/050/050/0
1000-50%50/050/050/0
1500-25%50/050/050/0
2000cutoff2/481/492/48
2500+25%0/500/500/50
3000+50%0/500/500/50
3500+75%0/500/500/50
4000+100%0/500/500/50

Oxycodone

Conc(ng/mL)% ofcutoffResults(Neg/Pos)Lot 1Results(Neg/Pos)Lot 2Results(Neg/Pos)Lot 3
0050/050/050/0
75-75%50/050/050/0
150-50%50/050/050/0
225-25%50/050/050/0
300cutoff2/482/482/48
375+25%0/500/500/50
450+50%0/500/500/50
રુડર+75%0/500/500/50
600+100%0/500/500/50

A lay user study was also performed. A total of four-hundred (400) participants were recruited, and each participant was provided one (1) package insert, one (1) blind labeled test solution, and one (1) test device. Test solutions were randomly picked for participants, one for each. Every participant was requested to read the provided materials, perform the testing, and then filled out the forms. Participants recruited were with diverse educational backgrounds and ranged in age from 18 to >60. Results were as follows:

{15}------------------------------------------------

Drug(cutoffng/ml)CutoffConcentration%(ng/ml)NumberofsamplesNegativePositive
AMP(1000)0% (0)3503500
75% (750)1091
125% (1250)1019
150% (1500)30030
BAR(200)0% (0)3503500
75% (150)1091
125% (250)1019
150% (300)30030
BUP(10)0% (0)20200
50% (5)60600
75% (7.5)60573
125% (12.5)1209111
150% (15)1400140
BZD(300)0% (0)3503500
75% (225)1091
125% (375)1019
150% (450)30030
COC(300)0% (0)3503500
75% (225)1091
125% (375)10010
150% (450)30030
MET(1000)0% (0)3503500
75% (750)1091
125% (1250)1019
150% (1500)30030
MTD(300)0% (0)3503500
75% (225)1082
125% (375)10010
150% (450)30030
PCP(25)0% (0)3503500
75% (18.75)10100
125% (31.25)1019
150% (37.5)30030
TCA(1000)0% (0)3503500
75% (750)1091
125% (1250)1019
150% (1500)30030
THC(50)0% (0)3503500
75% (37.5)10100
125% (62.5)1028
150% (75)30030
XTC(MDMA)(500)0% (0)3503500
75% (37.5)10100
125% (625)1019
150% (750)30030
MOR(2000)0% (0)3503500
75% (1500)10100
125% (2500)1028
150% (3000)30030
OXY(300)0% (0)3503500
75% (225)10100
125% (375)1028
150% (450)30030

{16}------------------------------------------------

Surveys and labeling assessments

Very easy tounderstandEasy tounderstandUnderstandw/ somedifficultyDifficult orimpossible tounderstand
Explanationof intendeduse of the test171(42.75%)215(53.75%)14(3.5%)0
Directions todo the test226(59%)163(40.75%)11(2.75%)0
Performingthe test186(46.5%)202(50.5%)12(3%)0
Direction tointerpret theresults246(61.5%)151(37.75%)13(3.25%)0
Actualinterpretationof the testresults182(45.5%)212(53%)16(4%)0

Conclusion The INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use) and INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) are substantially equivalent to the predicate device.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).