K Number
K182690
Date Cleared
2018-12-21

(85 days)

Product Code
Regulation Number
866.3365
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GenMark ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on GenMark's ePlex Instrument for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The ePlex BCID-FP Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organism.

The following fungal organisms are identified using the ePlex BCID-FP Panel: Candida albicans, Candida auris, Candida dubliniensis, Candida famata, Candida glabrata, Candida guilliermondii, Candida kefyr, Candida krusei, Candida lusitaniae, Candida parapsilosis, Candida tropicalis, Cryptococcus gattii, Cryptococcus neoformans, Fusarium and Rhodotorula.

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from the ePlex BCID-FP Panel are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a suspected bloodstream infection may be due to infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by the ePlex BCID-FP Panel may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by ePlex BCID-FP Panel, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

Device Description

The ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel is based on the principles of competitive nucleic acid hybridization using a sandwich assay format, wherein a single-stranded target binds concurrently to a sequence-specific solution-phase signal probe and a solid-phase electrode-bound capture probe. The test employs nucleic acid extraction, target amplification via polymerase chain reaction (PCR) or reverse transcription PCR (RT-PCR) and hybridization of target DNA. In the process, the double-stranded PCR amplicons are digested with exonuclease to generate single-stranded DNA suitable for hybridization.

Nucleic acid extraction from biological samples occurs within the cartridge via cell lysis, nucleic acid capture onto magnetic beads, and release for amplification. The nucleic acid extraction is processed through microfluidic liquid handling. Once the nucleic acid targets are captured and inhibitors are washed away, the magnetic particles are delivered to the electrowetting environment on the printed circuit board (PCB) and the targets are eluted from the particles and amplified.

During hybridization, the single-stranded target DNA binds to a complementary, single-stranded capture probe immobilized on the working gold electrode surface. Single-stranded signal probes (labeled with electrochemically active ferrocenes) bind to specific target sequence / region adjacent to the capture probe. Simultaneous hybridization of target to signal probes and capture probe is detected by alternating current voltammetry (ACV). Each working electrode on the array contains specific capture probes, and sequential analysis of each electrode allows detection of multiple analyte targets.

AI/ML Overview

This document describes the analytical and clinical performance of the GenMark ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel, an in vitro diagnostic test. The information provided is sufficient to extract the requested details about acceptance criteria and study proving the device meets them.

1. Table of acceptance criteria and reported device performance:

The document implicitly defines acceptance criteria through the reported performance characteristics. While no explicit "acceptance criteria" table is provided, the clinical performance (Sensitivity/PPA and Specificity/NPA) tables against a comparator method serve as the primary evidence of meeting performance expectations. Analytical performance characteristics also define a form of acceptance criteria (e.g., LOD, exclusivity).

Here’s a table summarizing the reported device performance for each target organism, which represents the device meeting its performance objectives. The device demonstrates high sensitivity (positive percent agreement) and specificity (negative percent agreement) across various sample types (prospective, retrospective, and contrived).

TargetSample TypeSensitivity/PPA % (95% CI)Specificity/NPA % (95% CI)
Candida albicansOverall97.1 (89.9-99.2)99.9 (99.3-100)
Candida aurisOverall100 (92.7-100)100 (99.5-100)
Candida dubliniensisOverall100 (93.1-100)100 (99.5-100)
Candida famataOverall100 (93.0-100)100 (99.5-100)
Candida glabrataOverall98.3 (91.1-99.7)99.8 (99.1-99.9)
Candida guilliermondiiOverall98.0 (89.5-99.6)100 (99.5-100)
Candida kefyrOverall100 (93.0-100)99.8 (99.1-99.9)
Candida kruseiOverall100 (92.9-100)100 (99.5-100)
Candida lusitaniaeOverall98.0 (89.3-99.6)99.9 (99.3-100)
Candida parapsilosisOverall98.3 (91.1-99.7)99.9 (99.3-100)
Candida tropicalisOverall100 (92.9-100)99.9 (99.3-100)
Cryptococcus gattiiOverall100 (92.9-100)100 (99.5-100)
Cryptococcus neoformansOverall100 (93.7-100)100 (99.5-100)
FusariumOverall98.6 (92.3-99.7)100 (99.5-100)
RhodotorulaOverall96.2 (87.0-98.9)99.9 (99.3-100)

2. Sample size used for the test set and the data provenance:

The test set for evaluating clinical performance consisted of:

  • Prospective Samples: 21 evaluable samples (11 fresh, 10 frozen) collected at 6 clinical sites. These samples are from patients of all ages and genders. Collection dates are specified from May 2015 through July 2016 (frozen) and July through August 2018 (fresh). The country of origin is not explicitly stated but implied to be the US given the FDA submission. This data is prospective.
  • Retrospective Samples: 120 samples collected from 9 sites. This data is retrospective.
  • Contrived Samples: 725 evaluable samples prepared by spiking isolates into blood culture bottles. These are laboratory-prepared samples.

The total number of samples evaluated for clinical performance was 866 (11 fresh prospective + 10 frozen prospective + 120 retrospective + 725 contrived).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not specify the number of experts or their qualifications (e.g., radiologists with 10 years of experience) used to establish the ground truth. Instead, it relies on standard laboratory procedures and analytically validated PCR assays followed by bi-directional sequencing as the comparator methods (ground truth). This implies that the ground truth was established through a combination of traditional microbiological methods and molecular techniques, not through expert consensus reading of images or other subjective assessments.

4. Adjudication method for the test set:

The document describes the "comparator method" as the gold standard. For specific targets like Candida auris, Fusarium, Rhodotorula, and to confirm Candida parapsilosis, PCR/sequencing was used to determine the presence or absence of the organism, effectively acting as an adjudication step for these cases. For other organisms, standard laboratory procedures (culture, MALDI-TOF IVD, etc.) defined the ground truth. There is no mention of a traditional reader adjudication process (e.g., 2+1 or 3+1) as would be common in image-based AI studies, as this is a molecular diagnostic test. For discrepant results (e.g., section "Co-detections in Clinical Samples"), PCR/sequencing was used to investigate.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-assisted diagnostic tools that involve human interpretation of images. The ePlex BCID-FP Panel is an in vitro diagnostic (IVD) test that directly detects and identifies genetic material, so human readers are not involved in its direct interpretation in the way they would be in an AI imaging study. Therefore, there is no effect size of how much human readers improve with AI vs without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, the primary clinical performance evaluation is a standalone performance of the algorithm (the ePlex BCID-FP Panel) against a defined ground truth (comparator methods). The reported sensitivity/PPA and specificity/NPA values are purely the device's performance.

7. The type of ground truth used:

The ground truth for the clinical performance evaluation was established using:

  • Standard laboratory procedures: This includes traditional and automated culture, MALDI-TOF IVD (Matrix-Assisted Laser Desorption/Ionization Time-Of-Flight Mass Spectrometry), and other microbiological and biochemical techniques.
  • Analytically validated PCR assays followed by bi-directional sequencing: This advanced molecular method was used for specific targets (Candida auris, Fusarium, Rhodotorula) and to confirm certain identifications (Candida parapsilosis).

This represents a combination of expert consensus (through standard lab practices) and molecular outcomes data.

8. The sample size for the training set:

The document does not explicitly state the sample size for a "training set" in the context of machine learning. As this is a molecular diagnostic assay using nucleic acid hybridization and PCR, not a machine learning algorithm that learns from data in the same way, the concept of a distinct 'training set' for the device's core functionality specification might not apply directly in the conventional AI sense. The development of such assays involves analytical validation using numerous strains and concentrations (analytical reactivity, LOD, exclusivity), which implicitly serve as a form of "training" or optimization data during product development, but this is distinct from machine learning model training. The provided data focuses on the performance evaluation (test set) for regulatory approval.

9. How the ground truth for the training set was established:

Given that there isn't a "training set" in the typical machine learning sense, the way "ground truth" would be established for the development of such a molecular assay would involve:

  • Known Reference Strains: Use of well-characterized microbial strains (e.g., ATCC, CBS, CDC strains) with confirmed identities. These are used in analytical studies like Limit of Detection (LOD) and Analytical Reactivity (Inclusivity), as well as Competitive Inhibition studies. Table 21 ("Contrived Sample Summary") and Table 27 ("Analytical Reactivity (Inclusivity) Results") list numerous specific strains and their origins (e.g., ATCC, CBS, NCPF, CDC#) used in testing.
  • Sequencing and Phenotypic Characterization: During the assay's development, target sequences would be determined through genome sequencing, and phenotypic characteristics would be confirmed through established microbiological methods.

Therefore, the "ground truth" during device development (analogous to training/development data in AI) relies on well-characterized laboratory standards and molecular gold standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2018

GenMark Diagnostics, Incorporated Beth Stofka Sr. Regulatory Affairs Specialist 5964 La Place Court Carlsbad, California 92008

Re: K182690

Trade/Device Name: ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel Regulation Number: 21 CFR 866.3365 Regulation Name: Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures Regulatory Class: Class II Product Code: PEO Dated: September 26, 2018 Received: September 27, 2018

Dear Beth Stofka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Uwe Scherf -S

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182690

Device Name

ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel

Indications for Use (Describe)

The GenMark ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on GenMark's ePlex Instrument for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The ePlex BCID-FP Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organism.

The following fungal organisms are identified using the ePlex BCID-FP Panel: Candida albicans, Candida auris, Candida dubliniensis, Candida famata, Candida glabrata, Candida guilliermondii, Candida kefyr, Candida krusei, Candida lusitaniae, Candida parapsilosis, Candida tropicalis, Cryptococcus gattii, Cryptococcus neoformans, Fusarium and Rhodotorula.

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from the ePlex BCID-FP Panel are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a suspected bloodstream infection may be due to infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by the ePlex BCID-FP Panel may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by ePlex BCID-FP Panel, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Summary of Safety and Effectiveness

Submitter Information

Submitter:GenMark Diagnostics, Incorporated5964 La Place CourtCarlsbad, CA 92008
Manufacturer:GenMark Diagnostics, Incorporated5964 La Place CourtCarlsbad, CA 92008
Establishment Registration Number:3008632402
Contact:Beth StofkaSr. Regulatory Affairs Specialist
Phone:760-579-4778
Fax:760-683-6961
E-mail:Beth.Stofka@genmarkdx.com
Alternate Contact:Alan Maderazo, Ph.D., RACVice President, Quality, Regulatory and Clinical Affairs
Phone:760-448-4308
Fax:760-683-6961
E-mail:Al.Maderazo@genmarkdx.com
Date Prepared:September 27, 2018

Name of Device and Classification

ePlex® Blood Culture Identification Fungal Pathogen (BCID-FP) Panel Product Name: Device Classification: 866.3980, Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures, Class II Product Code(s): PEO

{5}------------------------------------------------

Predicate Device

Predicate:

FilmArray Blood Culture Identification Panel; BioFire Diagnostics; K130914

Device Description

The ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel is based on the principles of competitive nucleic acid hybridization using a sandwich assay format, wherein a single-stranded target binds concurrently to a sequence-specific solution-phase signal probe and a solid-phase electrode-bound capture probe. The test employs nucleic acid extraction, target amplification via polymerase chain reaction (PCR) or reverse transcription PCR (RT-PCR) and hybridization of target DNA. In the process, the double-stranded PCR amplicons are digested with exonuclease to generate single-stranded DNA suitable for hybridization.

Nucleic acid extraction from biological samples occurs within the cartridge via cell lysis, nucleic acid capture onto magnetic beads, and release for amplification. The nucleic acid extraction is processed through microfluidic liquid handling. Once the nucleic acid targets are captured and inhibitors are washed away, the magnetic particles are delivered to the electrowetting environment on the printed circuit board (PCB) and the targets are eluted from the particles and amplified.

During hybridization, the single-stranded target DNA binds to a complementary, single-stranded capture probe immobilized on the working gold electrode surface. Single-stranded signal probes (labeled with electrochemically active ferrocenes) bind to specific target sequence / region adjacent to the capture probe. Simultaneous hybridization of target to signal probes and capture probe is detected by alternating current voltammetry (ACV). Each working electrode on the array contains specific capture probes, and sequential analysis of each electrode allows detection of multiple analyte targets.

{6}------------------------------------------------

Intended Use/Indications for Use

The GenMark ePlex® Blood Culture Identification Fungal Pathogen (BCID-FP) Panel is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on GenMark's ePlex Instrument for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The ePlex BCID-FP Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organism.

The following fungal organisms are identified using the ePlex BCID-FP Panel: Candida albicans, Candida auris, Candida dubliniensis, Candida famata, Candida glabrata, Candida guilliermondii, Candida kefyr, Candida krusei, Candida lusitaniae, Candida parapsilosis, Candida tropicalis, Cryptococcus gattii, Cryptococcus neoformans, Fusarium and Rhodotorula.

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from the ePlex BCID-FP Panel are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a suspected bloodstream infection may be due to infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by the ePlex BCID-FP Panel may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by ePlex BCID-FP Panel, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

{7}------------------------------------------------

Summary of Technological Characteristics of the Device Compared to the Predicate Device

The GenMark ePlex Blood Culture Identification Fungal Pathogen (BCID-FP) Panel ("Subject Device") and the legally marketed device, FilmArray Blood Culture Identification Panel, K130914, ("Predicate Device") are described below:

CharacteristicSubject DevicePredicate Device
Product NameePlex BCID-FP PanelFilmArray BCID Panel
ManufacturerGenMark Diagnostics, Inc.BioFire Diagnostics, Inc.
OrganismsDetected• Candida albicans• Candida auris• Candida dubliniensis,• Candida famata• Candida glabrata• Candida guilliermondii• Candida kefyr• Candida krusei• Candida lusitaniae• Candida parapsilosis• Candida tropicalis• Cryptococcus gattii• Cryptococcus neoformans• Fusarium• RhodotorulaEnterococci, Listeria monocytogenes,commonly encountered Staphylococci(including specific differentiation ofStaphylococcus aureus), commonlyencountered Streptococci (withspecific differentiation ofStreptococcus agalactiae,Streptococcus pneumoniae, andStreptococcus pyogenes),Acinetobacter baumannii, commonlyencountered Enterobacteriaceae(including specific differentiation ofthe Enterobacter cloacae complex,Escherichia coli, Klebsiella oxytoca,Klebsiella pneumoniae, Proteus, andSerratia marcescens), Haemophilusinfluenzae, Neisseria meningitidis(encapsulated), Pseudomonasaeruginosa, Candida albicans,Candida glabrata, Candida krusei,Candida parapsilosis, and Candidatropicalis.
Indication forUseThe ePlex BCID-FP Panel is indicatedas an aid in the diagnosis ofbloodstream infections. The use ofadditional laboratory testing (e.g. sub-culturing of positive blood cultures foridentification of organisms not detectedby the ePlex BCID-FP Panel and forsusceptibility testing, differentiation ofmixed growth) and clinical presentationmust be taken into consideration in thefinal diagnosis of blood streaminfection.FilmArray BCID is indicated as an aidin the diagnosis of specific agents ofbacteremia and fungemia and resultsshould be used in conjunction withother clinical and laboratory findings.
CharacteristicSubject DevicePredicate Device
Specimen TypeBlood culture samples identified aspositive by a continuous monitoringblood culture system and whichcontain fungal organism.Blood culture samples identified aspositive by a continuous monitoringblood culture system that demonstratesthe presence of organisms asconfirmed by Gram stain.
ChemistryReagents on cartridge include: samplelysis and nucleic acid extraction, PCRamplification and hybridization-basedelectrochemical detection reagents.The FilmArray BCID pouch containsfreeze-dried reagents to performnucleic acid purification and nested,multiplex PCR with DNA meltanalysis.
HardwareGenMark ePlex Instrument & SingleUse CartridgeFilmArray Instrument and assay pouch
SoftwareInterfaceResult ReportingGenMark ePlex System Software GenMark ePlex BCID-FP PanelSoftwareThe FilmArray Software automaticallyinterprets the results of each DNAmelt curve analysis and combines thedata with the results of the internalpouch controls to provide a test resultfor each organism and antimicrobialresistance gene on the panel.

{8}------------------------------------------------

{9}------------------------------------------------

Summary of Performance Data

EXPECTED VALUES

A prospective, multicenter clinical study was conducted to evaluate the clinical performance of the ePlex BCID-FP Panel in positive blood culture samples. A total of 447 positive blood culture samples were collected at 6 clinical sites in 2 phases from patients of all ages and genders. Samples were collected and frozen for future testing from May 2015 through July 2016. Samples were collected from July through August 2018 and tested fresh (never frozen). Of these 447 samples, 21 had a Gram stain result indicating fungal organism. The expected values of individual analytes based on the ePlex BCID-FP Panel results in the 21 prospective samples are summarized by age group and by site in Tables 1 and 2 below.

{10}------------------------------------------------

TargetAll Ages(N=21)n (%)Age <1(N=1)n (%)Age 1-17(N=2)n (%)Age 18-44(N=4)n (%)Age 45-64(N=11)n (%)Age 65-84(N=2)n (%)Age 85+(N=1)n (%)
Candida albicans4 (19.0)1 (100)0 (0.0)0 (0.0)2 (18.2)1 (50.0)0 (0.0)
Candida auris0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida dubliniensis1 (4.8)0 (0.0)0 (0.0)0 (0.0)1 (9.1)0 (0.0)0 (0.0)
Candida famata0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida glabrata6 (28.6)0 (0.0)1 (50.0)1 (25.0)3 (27.3)1 (50.0)0 (0.0)
Candida guilliermondii0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida kefyr0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida krusei2 (9.5)0 (0.0)0 (0.0)0 (0.0)2 (18.2)0 (0.0)0 (0.0)
Candida lusitaniae0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida parapsilosis2 (9.5)0 (0.0)1 (50.0)0 (0.0)1 (9.1)0 (0.0)0 (0.0)
Candida tropicalis2 (9.5)0 (0.0)0 (0.0)1 (25.0)1 (9.1)0 (0.0)0 (0.0)
Cryptococcus gattii0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Cryptococcus neoformans0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Fusarium0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Rhodotorula1 (4.8)0 (0.0)0 (0.0)1 (25.0)0 (0.0)0 (0.0)0 (0.0)

Table 1: Expected Value by Age Group (Prospective Samples)

Table 2: Expected Value by Collection Site (Prospective Samples)

All Sites(N=21)Site 1(N=1)Site 2(N=8)Site 3(N=2)Site 4(N=4)Site 5(N=4)Site 6(N=2)
Targetn (%)n (%)n (%)n (%)n (%)n (%)n (%)
Candida albicans4 (19.0)1 (100)0 (0.0)2 (100)1 (25.0)0 (0.0)0 (0.0)
Candida auris0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida dubliniensis1 (4.8)0 (0.0)0 (0.0)0 (0.0)1 (25.0)0 (0.0)0 (0.0)
Candida famata0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida glabrata6 (28.6)0 (0.0)2 (25.0)0 (0.0)1 (25.0)2 (50.0)1 (50.0)
Candida guilliermondii0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida kefyr0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida krusei2 (9.5)0 (0.0)2 (25.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida lusitaniae0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Candida parapsilosis2 (9.5)0 (0.0)0 (0.0)0 (0.0)0 (0.0)1 (25.0)1 (50.0)
Candida tropicalis2 (9.5)0 (0.0)2 (25.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Cryptococcus gattii0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Cryptococcus neoformans0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Fusarium0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)0 (0.0)
Rhodotorula1 (4.8)0 (0.0)0 (0.0)0 (0.0)0 (0.0)1 (25.0)0 (0.0)

{11}------------------------------------------------

PERFORMANCE CHARACTERISTICS

Clinical Performance

Samples with a Gram stain result indicating fungal organism, a final, valid investigational test result, and a valid comparator result were evaluable and included in summaries and analyses of demographics, expected values (positivity rate), and performance characteristics. Evaluable samples included 11 prospective fresh and 10 prospective frozen samples as well as 120 retrospective samples and 725 contrived samples.

Comparator Method

The performance of the ePlex BCID-FP Panel was compared to standard laboratory procedures, including traditional and automated culture, MALDI-TOF IVD, and microbiological and biochemical techniques. In addition, all prospective samples were tested with analytically validated PCR assays followed by bi-directional sequencing to determine the presence or absence of Candida auris, Fusarium, and Rhodotorula. Identification for samples with Candida parapsilosis identified by standard laboratory procedures was confirmed using analytically validated PCR assays followed by bi-directional sequencing.

The comparator method(s) results were used to determine the Detected / Not Detected status for each target organism on the ePlex BCID-FP Panel. The comparator methods for each target are summarized in Table 3.

{12}------------------------------------------------

TargetComparator Method
Candida albicans
Candida dubliniensis
Candida famata
Candida glabrata
Candida guilliermondii
Candida kefyrStandard laboratory procedures for organism identification.
Candida krusei
Candida lusitaniae
Candida tropicalis
Cryptococcus gattii
Cryptococcus neoformans
Candida parapsilosisStandard laboratory procedures for organism ID. PCR/sequencing toconfirm C. parapsilosis or identify C. metapsilosis, C. orthopsilosis.
Candida auris, Fusarium, andRhodotorulaStandard laboratory procedures for organism identification.PCR/sequencing in prospective samples.

Table 3: Comparator Method(s) by ePlex BCID-FP Panel Target

Demographics of Clinical Samples

Clinical performance was evaluated in samples prospectively and retrospectively collected. Prospective samples were collected at 6 clinical sites, with 21 evaluable samples. Sample with final, valid, ePlex BCID-FP Panel results and valid comparator results were considered evaluable. Demographic information for prospectively-collected samples is described in Table 4. Subjects enrolled in this study were from a diverse demographic distribution and represent the intended patient population.

To supplement the results of the prospective collection, 120 samples were collected retrospectively from a total of 9 sites, and 725 evaluable samples were contrived for organisms with low prevalence. Demographic information for retrospectively-collected samples is described in Table 5.

{13}------------------------------------------------

All Sites(N=21)Site 1(N=1)Site 2(N=8)Site 3(N=2)Site 4(N=4Site 5(N=4)Site 6(N=2)
n (%)n (%)n (%)n (%)n (%))n (%)n (%)
Sex, n (%)
Male14(66.7)1(100)7(87.5)1(50.0)3(75.0)1(25.0)1(50.0)
Female7(33.3)0(0.0)1(12.5)1(50.0)1(25.0)3(75.0)1(50.0)
Age (years)
<1 yr1(4.8)0(0.0)0(0.0)0(0.0)1(25.0)0(0.0)0(0.0)
1-17 yrs2(9.5)0(0.0)0(0.0)0(0.0)0(0.0)2(50.0)0(0.0)
18-44 yrs4(19.0)0(0.0)2(25.0)0(0.0)1(25.0)1(25.0)0(0.0)
45-64 yrs11(52.4)1(100)4(50.0)1(50.0)2(50.0)1(25.0)2(100)
65-84 yrs2(9.5)0(0.0)1(12.5)1(50.0)0(0.0)0(0.0)0(0.0)
85+ yrs1(4.8)0(0.0)1(12.5)0(0.0)0(0.0)0(0.0)0(0.0)

Table 4: Demographic Data for Clinical Samples by Collection Site (Prospective Collection)

{14}------------------------------------------------

All Sites(N=120)Site 1(N=13)Site 2(N=14)Site 3(N=17)Site 4(N=4)Site 5(N=3)Site 6(N=13)Site 7(N=16)Site 8(N=5)Site 9(N=35)
n (%)n (%)n (%)n (%)n (%)n (%)n (%)n (%)n (%)n (%)
Sex, n (%)
Male68(56.7)10(76.9)8(57.1)8(47.1)1(25.0)2(66.7)8(61.5)9(56.3)3(60.0)19(54.3)
Female52(43.3)3(23.1)6(42.9)9(52.9)3(75.0)1(33.3)5(38.5)7(43.8)2(40.0)16(45.7)
Age (years)
<1 yr2(1.7)0(0.0)0(0.0)0(0.0)0(0.0)0(0.0)0(0.0)0(0.0)0(0.0)2(5.7)
1-17 yrs8(6.7)1(7.7)0(0.0)0(0.0)0(0.0)0(0.0)5(38.5)0(0.0)0(0.0)2(5.7)
18-44 yrs27(22.5)4(30.8)2(14.3)2(11.8)1(25.0)0(0.0)3(23.1)3(18.8)1(20.0)11(31.4)
45-64 yrs39(32.5)2(15.4)6(42.9)6(35.3)1(25.0)2(66.7)2(15.4)7(43.8)1(20.0)12(34.3)
65-84 yrs39(32.5)6(46.2)6(42.9)8(47.1)2(50.0)0(0.0)2(15.4)5(31.3)2(40.0)8(22.9)
85+ yrs5(4.2)0(0.0)0(0.0)1(5.9)0(0.0)1(33.3)1(7.7)1(6.3)1(20.0)0(0.0)

Table 5: Demographic Data for Clinical Samples by Collection Site (Retrospective Collection)

{15}------------------------------------------------

Clinical Performance

Sensitivity or positive percent agreement (PPA) was calculated by dividing the number of true positive (TP) results by the sum of TP and false negative (FN) results, while specificity or negative percent agreement (NPA) was calculated by dividing the number of true negative (TN) results by the sum of TN and false positive (FP) result being defined as a sample where the detected ePlex BCID-FP Panel result matched the detected comparator method result, while a TN result was one where a negative ePlex BCID-FP Panel result matched a negative comparator method result. The two-sided 95% confidence interval was also calculated.

A total of 866 positive blood culture samples with a Gram stain result indicating fungal organism consisting of 11 fresh prospective, 10 frozen prospective, 120 retrospective, and 725 contrived samples were evaluated for the ePlex BCID-FP Panel targets. Contrived samples were prepared by spiking an isolate into a blood culture bottle and growing until flagged positive by a continuously monitoring blood culture system. Samples were removed from the system within 8 hours of positivity and stored frozen until the time of testing. PPA and NPA results are summarized by target in Tables 6-20 below and the strains used to contrive samples are summarized in Table 21.

{16}------------------------------------------------

TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)TN/TN+FPSpecificity/NPA% (95% CI)
Candida albicansProspective (Fresh)2/2100 (34.2-100)9/9100 (70.1-100)
Prospective (Frozen)2/2100 (34.2-100)8/8100 (67.6-100)
Prospective (All)4/4100 (51.0-100)17/17100 (81.6-100)
Retrospective49/5098.0 (89.5-99.6)70/70100 (94.8-100)
Prospective/Retrospective53/5498.1 (90.2-99.7)87/87100 (95.8-100)
Contrived13/1492.9 (68.5-98.7)710/71199.9 (99.2-100)
Overall66/6897.1 (89.9-99.2)797/79899.9 (99.3-100)

Table 6: Clinical Performance for Candida albicans

CI= Confidence Interval

Table 7: Clinical Performance for Candida auris
-------------------------------------------------------------
TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)TN/TN+FPSpecificity/NPA% (95% CI)
Candida aurisProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective0/0---120/120100 (96.9-100)
Prospective/Retrospective0/0---141/141100 (97.3-100)
Contrived49/49100 (92.7-100)676/676100 (99.4-100)
Overall49/49100 (92.7-100)817/817100 (99.5-100)

{17}------------------------------------------------

TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)TN/TN+FPSpecificity/NPA% (95% CI)
CandidadubliniensisProspective (Fresh)1/1100 (20.7-100)10/10100 (72.2-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)1/1100 (20.7-100)20/20100 (83.9-100)
Retrospective3/3100 (43.9-100)117/117100 (96.8-100)
Prospective/Retrospective4/4100 (51.0-100)137/137100 (97.3-100)
Contrived48/48100 (92.6-100)677/677100 (99.4-100)
Overall52/52100 (93.1-100)814/814100 (99.5-100)

Table 8: Clinical Performance for Candida dubliniensis

Table 9: Clinical Performance for Candida famata

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
Candida famataProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective0/0---120/120100 (96.9-100)
Prospective/Retrospective0/0---141/141100 (97.3-100)
Contrived51/51100 (93.0-100)674/674100 (99.4-100)
Overall51/51100 (93.0-100)815/815100 (99.5-100)

{18}------------------------------------------------

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
Candida glabrataProspective (Fresh)4/4100 (51.0-100)7/7100 (64.6-100)
Prospective (Frozen)2/2100 (34.2-100)8/8100 (67.6-100)
Prospective (All)6/6100 (61.0-100)15/15100 (79.6-100)
Retrospective37/3897.4 (86.5-99.5)80/8297.6 (91.5-99.3)
Prospective/Retrospective43/4497.7 (88.2-99.6)95/97A97.9 (92.8-99.4)
Contrived16/16100 (80.6-100)709/709100 (99.5-100)
Overall59/6098.3 (91.1-99.7)804/80699.8 (99.1-99.9)

Table 10: Clinical Performance for Candida glabrata

A. C. glabrata was detected in 2/2 false positive clinical samples using PCR/sequencing.

Table 11: Clinical Performance for Candida guilliermondii
-----------------------------------------------------------------------
TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)TN/TN+FPSpecificity/NPA% (95% CI)
CandidaguilliermondiiProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
CandidaguilliermondiiRetrospective0/0---120/120100 (96.9-100)
CandidaguilliermondiiProspective/Retrospective0/0---141/141100 (97.3-100)
CandidaguilliermondiiContrived49/5098.0 (89.5-99.6)675/675100 (99.4-100)
CandidaguilliermondiiOverall49/5098.0 (89.5-99.6)816/816100 (99.5-100)

{19}------------------------------------------------

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+F P% (95% CI)
Candida kefyrProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective0/0---120/120100 (96.9-100)
Prospective/Retrospective0/0---141/141100 (97.3-100)
Contrived51/51100 (93.0-100)672/67499.7 (98.9-99.9)
Overall51/51100 (93.0-100)813/81599.8 (99.1-99.9)

Table 12: Clinical Performance for Candida kefyr

Table 13: Clinical Performance for Candida krusei

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
Candida kruseiProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)2/2100 (34.2-100)8/8100 (67.6-100)
Prospective (All)2/2100 (34.2-100)19/19100 (83.2-100)
Retrospective2/2100 (34.2-100)118/118100 (96.8-100)
Prospective/Retrospective4/4100 (51.0-100)137/137100 (97.3-100)
Contrived46/46100 (92.3-100)679/679100 (99.4-100)
Overall50/50100 (92.9-100)816/816100 (99.5-100)

{20}------------------------------------------------

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+F P% (95% CI)
Candida lusitaniaeProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective3/475.0 (30.1-95.4)116/116100 (96.8-100)
Prospective/Retrospective3/475.0 (30.1-95.4)137/137100 (97.3-100)
Contrived45/45100 (92.1-100)679/68099.9 (99.2-100)
Overall48/4998.0 (89.3-99.6)816/81799.9 (99.3-100)

Table 14: Clinical Performance for Candida lusitaniae

Table 15: Clinical Performance for Candida parapsilosis

TargetSample TypeTP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
CandidaparapsilosisProspective (Fresh)2/2100 (34.2-100)9/9100 (70.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)2/2100 (34.2-100)19/19100 (83.2-100)
Retrospective16/1794.1 (73.0-99.0)102/10399.0 (94.7-99.8)
Prospective/Retrospective18/1994.7 (75.4-99.1)121/122A99.2 (95.5-99.9)
Contrived41/41100 (91.4-100)684/684100 (99.4-100)
Overall59/6098.3 (91.1-99.7)805/80699.9 (99.3-100)

C. parapsilosis was detected in 1/1 false positive clinical sample using PCR/sequencing. A.

{21}------------------------------------------------

TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)Specificity/NPATN/TN+FP% (95% CI)
Candida tropicalisProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)2/2100 (34.2-100)8/8100 (67.6-100)
Prospective (All)2/2100 (34.2-100)19/19100 (83.2-100)
Retrospective3/3100 (43.9-100)116/11799.1 (95.3-99.8)
Prospective/Retrospective5/5100 (56.6-100)135/136A99.3 (96.0-99.9)
Contrived45/45100 (92.1-100)680/680100 (99.4-100)
Overall50/50100 (92.9-100)815/81699.9 (99.3-100)

Table 16: Clinical Performance for Candida tropicalis

A. C. tropicalis was detected in 1/1 false positive clinical sample using PCR/sequencing.

Table 17: Clinical Performance for Cryptococcus gattii
------------------------------------------------------------------
TargetSample TypeTP/TP+FN% (95% CI)TN/TN+F P% (95% CI)
Cryptococcus gattiiProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective0/0---120/120100 (96.9-100)
Prospective/Retrospective0/0---141/141100 (97.3-100)
Contrived50/50100 (92.9-100)675/675100 (99.4-100)
Overall50/50100 (92.9-100)816/816100 (99.5-100)

{22}------------------------------------------------

TargetSample TypeTP/TP+FNSensitivity/PPA% (95% CI)TN/TN+FPSpecificity/NPA% (95% CI)
CryptococcusneoformansProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective5/5100 (56.6-100)115/115100 (96.8-100)
Prospective/Retrospective5/5100 (56.6-100)136/136100 (97.3-100)
Contrived52/52100 (93.1-100)673/673100 (99.4-100)
Overall57/57100 (93.7-100)809/809100 (99.5-100)

Table 18: Clinical Performance for Cryptococcus neoformans

Table 19: Clinical Performance for Fusarium

Sensitivity/PPASpecificity/NPA
TargetSample TypeTP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
FusariumProspective (Fresh)0/0---11/11100 (74.1-100)
Prospective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)0/0---21/21100 (84.5-100)
Retrospective0/0---120/120100 (96.9-100)
Prospective/Retrospective0/0---141/141100 (97.3-100)
Contrived69/7098.6 (92.3-99.7)655/655100 (99.4-100)
Overall69/7098.6 (92.3-99.7)796/796100 (99.5-100)

{23}------------------------------------------------

TargetSample TypeSensitivity/PPASpecificity/NPA
TP/TP+FN% (95% CI)TN/TN+FP% (95% CI)
Prospective (Fresh)1/1100 (20.7-100)10/10100 (72.2-100)
RhodotorulaProspective (Frozen)0/0---10/10100 (72.2-100)
Prospective (All)1/1100 (20.7-100)20/20100 (83.9-100)
Retrospective1/1100 (20.7-100)119/119100 (96.9-100)
Prospective/Retrospective2/2100 (34.2-100)139/139100 (97.3-100)
Contrived48/5096.0 (86.5-98.9)674/67599.9 (99.2-100)
Overall50/5296.2 (87.0-98.9)813/81499.9 (99.3-100)

Table 20: Clinical Performance for Rhodotorula

Table 21: Contrived Sample Summary

TargetOrganismStrainIndependentContrived SamplesTested
Candida albicansCandida albicansATCC 102312
ATCC 140532
ATCC 244332
ATCC 900285
ATCC MYA-44413
Candida albicans total14
Candida aurisCandida aurisATCC 109131
ATCC 123721
ATCC 127661
CBS 109133
CBS 123723
CBS 123732
CBS 127663
CBS 127673
CBS 127682
CDC#03855
CDC#03865
CDC#03875
CDC#03885
CDC#03895
CDC#03905
Candida auris total
ATCC MYA-5776
TargetOrganismStrainIndependentContrived SamplesTested
CandidadubliniensisCandida dubliniensisATCC MYA-57812
Candida dubliniensisATCC MYA-57912
Candida dubliniensisATCC MYA-58213
Candida dubliniensisNCPF39495
Candida dubliniensis total48
Candida famataDebaryomyces fabryiCBS 78921
Debaryomyces hanseniiCBS 19613
Debaryomyces subglobosusCBS 179627
Candida famata total51
Candida glabrataCandida glabrata128-40581
128-40721
ATCC 151262
ATCC 155452
ATCC 20011
ATCC 660324
ATCC 900302
Candida glabrataATCC MYA-29503
Candida glabrata total16
CandidaguilliermondiiCandida guilliermondiiATCC 2201713
ATCC 626012
Meyerozyma guilliermondiiATCC 9019710
ATCC 901989
ATCC 901996
Candida guilliermondii total50
Candida kefyrCandida kefyrATCC 2040934
ATCC 251210
ATCC 413513
ATCC 6602812
ATCC 855312
Candida kefyr total51
Candida kruseiCandida kruseiATCC 142438
ATCC 229859
ATCC 288709
ATCC 3219610
ATCC 3413510
Candida krusei total46
TargetOrganismStrainIndependentContrived SamplesTested
CandidalusitaniaeCandida lusitaniaeATCC 262875
ATCC 3444910
ATCC 427209
ATCC 6603511
ATCC MYA-76610
Candida lusitaniae total45
CandidaparapsilosisCandida parapsilosisATCC 2201911
ATCC 284745
ATCC 2847510
ATCC 588957
ATCC 900188
Candida parapsilosis total41

{24}------------------------------------------------

{25}------------------------------------------------

Candida tropicalisCandida tropicalisATCC 13699
ATCC 1380312
ATCC 2013809
ATCC 2013817
ATCC 7508
Candida tropicalis total45
CryptococcusgattiiCryptococcus gattiiATCC 1424811
ATCC 76108
ATCC MYA-4138
ATCC MYA-4560
ATCC MYA-48779
Cryptococcus gattii total50
CryptococcusneoformansCryptococcus neoformans var. grubiiATCC 141169
ATCC208821(H99)
ATCC 90112
NCPF8195
NCPF8299
NCPF8357
Cryptococcus neoformans var. neoformans(Filobasidiella bacillispora Kwon-Chung,teleomorph (serotype D))ATCC 348733
Cryptococcus neoformans var. neoformans(serotype D)ATCC 365563
ATCC MYA-5659
Cryptococcus neoformans total52
FusariumBisifusarium dimerumCBS 1089443
CBS 110317
CBS 116520

{26}------------------------------------------------

TargetOrganismStrainIndependentContrived SamplesTested
Fusarium moniliformeATCC 3815911
Fusarium oxysporumCBS 1166111
Fusarium sacchariATCC 2437910
CBS 11982811
Fusarium solaniATCC 3603111
Fusarium verticillioidesCBS 10031211
Fusarium total70
RhodotorulaRhodotorula glutinisATCC 327653
Rhodotorula glutinisATCC 327663
Rhodotorula glutinis (Fresenius) HarrisonATCC 963654
Rhodotorula mucilaginosaATCC 6603421
ATCC 944919
Rhodotorula total50

{27}------------------------------------------------

Genus and Group Assay Species Stratification

The ePlex BCID-FP Panel reports genus level results for Fusarium and Rhodotorula targets. Sensitivity/PPA of these genus and group level targets from all clinical and contrived samples tested are summarized in Table 22 below.

Target Species DetectedAll Samples
TP/TP+FN% (95% CI)
Fusarium69/7098.6 (92.3-99.7)
Bisifusarium dimerum14/1593.3 (70.2-98.8)
Fusarium moniliforme11/11100 (74.1-100)
Fusarium oxysporum1/1100 (20.7-100)
Fusarium sacchari21/21100 (84.5-100)
Fusarium solani11/11100 (74.1-100)
Fusarium verticillioides11/11100 (74.1-100)
Rhodotorula50/5296.2 (87.0-98.9)
Rhodotorula*2/2100 (34.2-100)
Rhodotorula glutinis6/6100 (61.0-100)
Rhodotorula glutinis (Fresenius) Harrison4/4100 (51.0-100)
Rhodotorula mucilaginosa38/4095.0 (83.5-98.6)

Table 22: Species Detected by Genus Assays

*Results are from 2 clinical samples (1 retrospective). All other Fusarium and Rhodotorula samples were contrived.

{28}------------------------------------------------

Co-detections in Clinical Samples

The ePlex BCID-FP Panel did not identify any fungal co-detections in prospective samples, and 6 fungal co-detections were identified in the retrospective samples. Of the 120 retrospective samples, 114 (95.0%) had single detections and 6 (5.0%) had double detections. In the 6 codetections, 4 included 1 organism that comparator method(s) did not detect. See Table 27 below for a summary of co-detections in retrospective samples.

Distinct Organism Combinations Detected bythe ePlex Panel in Retrospective SamplesNumber of Samples(Number Discrepant)DiscrepantOrganism(s)A,B
Target 1Target 2
Candida albicansCandida dubliniensis1 (0)
Candida albicansCandida parapsilosis1 (0)
Candida dubliniensisCandida parapsilosis1 (1)C. parapsilosis (1)
Candida dubliniensisCandida tropicalis1 (1)C. tropicalis (1)
Candida glabrataCandida lusitaniae2 (2)C. glabrata (2)

Table 23 Co-Detections Identified by the ePlex BCID-FP Panel (Retrospective Samples)

A discrepant organism is defined as one that was detected by the BCID-FP Panel but not by the comparator method(s). A. B.

4/4 organisms were investigated using PCR/sequencing and the discrepant organism was detected in 4/4 cases.

i. In 1/1 sample, C. parapsilosis was detected.

ii. In 1/1 sample, C. tropicalis was detected.

iii. In 2/2 samples, C. glabrata was detected.

Summaries of additional distinct fungal co-detections detected by the comparator method in prospective and retrospective samples are provided in Table 29. These tables include additional distinct fungal co-detections not included in the co-detections identified by the BCID-FP Panel; they have 1 or more organism not detected by the BCID-FP Panel or an offpanel fungal organism.

Table 24: Co-Detections Identified by the Comparator Method (Prospective Samples
Distinct Organism Combinations Detected by
Distinct Organism Combinations Detected bythe Comparator Method in ProspectiveSamplesNumber of Samples(Number Discrepant)DiscrepantOrganism(s)A
Target 1Target 2
Candida metapsilosis*Trichosporon asahii*1 (0)

A. A discrepant organism is defined as one that was detected by the comparator method(s) but not by the BCID-FP Panel (excludes organisms not targeted by the BCID-FP Panel).

  • Off-panel organism not targeted by the BCID-FP Panel.

{29}------------------------------------------------

Distinct Organism Combinations Detected by theComparator Method in Retrospective SamplesNumber of Samples(NumberDiscrepant)DiscrepantOrganism(s)A
Target 1Target 2Target 3
CandidaalbicansCandidadubliniensisCandida glabrata1 (1)C. glabrata (1)
CandidaalbicansCandidaparapsilosis1 (1)C. parapsilosis (1)

Table 25: Co-Detections Identified by the Comparator Method (Retrospective Samples)

A discrepant organism is defined as one that was detected by the comparator method(s) but not by the BCID-FP Panel A. (excludes organisms not targeted by the BCID-FP Panel).

{30}------------------------------------------------

Clinical Study ePlex Instrument Performance

A total of 867 samples (including prospective, retrospective, and contrived samples) were initially tested in the clinical evaluations. Of these, 7/867 (0.8%) did not complete the run and the sample was retested. After repeat testing, all 867 sampleted testing and 839/867 (96.8%, 95% CI: 95.4%-97.8%) generated valid results and 28/867 (3.2%, 95% CI: 2.2%-4.6%) generated invalid results on the first completed attempt.

Upon repeat testing of the 28 samples with initially invalid results, all completed the run and 27/28 (96.4%) generated valid results. Overall, after final testing, 1/867 (0.1%, 95% CI: 0.0%-0.7%) had final, invalid results, resulting in a final validity rate of 866/867 (99.9%, 95% CI: 99.3%-100%).

Analytical Performance Characteristics

Limit of Detection (Analytical Sensitivity)

The Limit of Detection (LOD), or analytical sensitivity, was identified and verified for each assay on the BCID-FP Panel using at least two quantified reference strains in simulated blood culture sample matrix, which is defined as whole blood with EDTA added to a blood culture bottle in the same ratio as the manufacturer recommends and incubated for 8 hours. At least 20 replicates per target were tested for each condition. The limit of detection was defined as the lowest concentration of each target that is detected in ≥95% of tested replicates. The confirmed LOD for each ePlex BCID-FP Panel organism is shown in Table 26.

{31}------------------------------------------------

Table 26: LOD Results Summary

TargetOrganismStrainLODConcentration(CFU/mL)
Candida albicansCandida albicansATCC 14053$1 x 10^6$
Candida albicansATCC 24433$1 x 10^5$
Candida aurisCandida aurisCBS 10913$1 x 10^5$
Candida aurisCBS 12766$1 x 10^5$
Candida dubliniensisCandida dubliniensisATCC MYA-577$1 x 10^4$
Candida dubliniensisNCPF 3949$1 x 10^5$
Candida famataCandida famataCBS 767$1 x 10^3$
Candida famataCBS 766$1 x 10^4$
Candida glabrataCandida glabrataATCC 2001$1 x 10^6$
Candida glabrataATCC 15545$1 x 10^6$
Candida guilliermondiiCandida guilliermondiiATCC 22017$1 x 10^5$
Candida guilliermondiiATCC 6260$1 x 10^5$
Candida kefyrCandida kefyrATCC 4135$1 x 10^3$
Candida kefyrATCC 8553$1 x 10^4$
Candida kruseiCandida kruseiATCC 22985$1 x 10^5$
Candida kruseiATCC 28870$1 x 10^6$
Candida lusitaniaeCandida lusitaniaeATCC 34449$1 x 10^6$
Candida lusitaniaeATCC 66035$1 x 10^5$
Candida parapsilosisCandida parapsilosisATCC 28474$1 x 10^4$
Candida parapsilosisATCC 28475$1 x 10^5$
Candida tropicalisCandida tropicalisATCC 13803$1 x 10^5$
Candida tropicalisATCC 1369$1 x 10^6$
Cryptococcus gattiiCryptococcus gattiiATCC MYA-4877$1 x 10^3$
TargetOrganismStrainLODConcentration(CFU/mL)
Cryptococcus gattiiATCC MYA-41381 x 103
Cryptococcus neoformansCryptococcus neoformansATCC 2088211 x 105
Cryptococcus neoformansATCC MYA-5651 x 105
FusariumFusarium oxysporumCBS 1166111 x 106spores/mL
Fusarium solaniATCC 363011 x 106spores/mL
RhodotorulaRhodotorula mucilaginosaATCC 40581 x 105
Rhodotorula glutinisATCC 327651 x 105
TargetOrganismStrain
Candida albicansCandida albicansATCC 10231
Candida albicansATCC 90028
Candida albicansATCC MYA-4441
Candida aurisCandida aurisCDC#385
Candida aurisCDC#386
Candida aurisCDC#387
Candida aurisCDC#388
Candida aurisCDC#389
Candida aurisCDC#390
Candida aurisCBS 12766
Candida dubliniensisCandida dubliniensisATCC MYA-578
Candida dubliniensisATCC MYA-579
Candida dubliniensisATCC MYA-582
Candida famataCandida famataATCC 20850
Candida famataCBA 1961
Candida famataCBS 789
Candida glabrataCandida glabrataATCC 15126
Candida glabrataATCC 66032
Candida glabrataATCC MYA-2950
Candida guilliermondiiCandida guilliermondiiATCC 90197
Candida guilliermondiiATCC 90198
Candida guilliermondiiATCC 90199
Candida kefyrCandida kefyrATCC 204093
Candida kefyrATCC 2512
Candida kefyrATCC 66028
Candida kruseiCandida kruseiATCC 14243
Candida kruseiATCC 32196
Candida kruseiATCC 34135
Candida lusitaniaeCandida lusitaniaeATCC 42720
Candida lusitaniaeATCC MYA-766
Candida lusitaniaeZ010
Candida parapsilosisCandida parapsilosisATCC 22019
Candida parapsilosisATCC 58895
Candida parapsilosisATCC 90018
Candida tropicalisCandida tropicalisATCC 201380
Candida tropicalisATCC 201381
TargetOrganismStrain
Candida tropicalisATCC 750
Cryptococcus gattiiCryptococcus gattiiATCC 14248
Cryptococcus gattiiCryptococcus gattiiATCC 4560
Cryptococcus gattiiCryptococcus gattiiATCC 76108
Cryptococcus neoformansCryptococcus neoformansATCC 14116
Cryptococcus neoformansCryptococcus neoformansATCC 90112
Cryptococcus neoformansCryptococcus neoformansNCPF 8299
FusariumBisifusarium dimerumCBS 110317
FusariumFusarium moniliformeATCC 38159
FusariumFusarium proliferatumCBS 131570
FusariumFusarium sacchariCBS 119828
FusariumFusarium verticillioidesCBS 100312
RhodotorulaRhodotorula glutinisATCC 96365
RhodotorulaRhodotorula mucilaginosaATCC 66034
RhodotorulaRhodotorula mucilaginosaATCC 9449

{32}------------------------------------------------

Analytical Reactivity (Inclusivity)

A panel of 51 strains/isolates, representing the genetic, temporal and geographic diversity of each target on the ePlex BCID-FP Panel was evaluated to demonstrate analytical reactivity. Each strain was tested in triplicate at concentrations approximating bottle positivity (1 x 10° CFU/mL for Candida and Rhodotorula, 1 x 107 CFU/mL for Cryptococcus, and 1 x 108 spores/mL for Fusarium). All organisms tested were detected at bottle positive concentrations. Results of the analytical reactivity are summarized in Table 27. An additional 29 unique strains were detected as a part of the Limit of Detection (Analytical Sensitivity) study and are summarized in Table 26.

{33}------------------------------------------------

Table 27: Analytical Reactivity (Inclusivity) Results

{34}------------------------------------------------

{35}------------------------------------------------

Predicted (in silico) Reactivity for Genus and Group Assays

Note: the performance of the ePlex BCID-FP Panel has not been established for all of the organisms listed in the tables below. See the Analytical Reactivity (Inclusivity) and Limit of Detection (Analytical Sensitivity) sections of the package insert for data on organisms for which performance characteristics have been established (indicated with an asterisk in Tables 32 and 33). Some species were not assessed in silico due to lack of sequence data, though they may appear in the analytical sensitivity or specificity studies.

In addition to species-specific assays, the ePlex BCID-FP Panel contains two broader genuslevel assays: Fusarium and Rhodotorula. Table 28 and Table 29 highlight the predicted (in silico) reactivity (inclusivity) for these assay targets.

Detection Predicted for ≥95% of target sequences
Fusarium acaciae-mearnsiiFusarium cortaderiaeFusarium napiforme
Fusarium acuminatumFusarium culmorumFusarium nisikadoi
Fusarium acutatumFusarium denticulatumFusarium nygamai
Fusarium aethiopicumBisifusarium dimerum*Fusarium oxysporum*
Fusarium ananatumFusarium dlaminiiFusarium palustre
Fusarium andiyaziFusarium equisetiFusarium phyllophilum
Fusarium anthophilumFusarium falciformeFusarium poae
Fusarium armeniacumFusarium foetensFusarium polyphialidicum
Fusarium asiaticumFusarium fujikuroiFusarium proliferatum*
Fusarium austroamericanumFusarium gaditjirriFusarium pseudoanthophilum
Fusarium avenaceumFusarium globosumFusarium pseudocircinatum
Fusarium aywerteFusarium guttiformeFusarium pseudograminearum
Fusarium bactridioidesFusarium hostaeFusarium pseudonygamai
Fusarium begoniaeFusarium incarnatumFusarium ramigenum
Fusarium beomiformeFusarium inflexumFusarium sacchari*
Fusarium boothiiFusarium konzumFusarium secorum
Fusarium brachygibbosumFusarium lacertarumFusarium sinensis
Fusarium brasilicumFusarium lactisFusarium solani*
Fusarium brevicatenulatumFusarium langsethiaeFusarium sporotrichioides
Fusarium bulbicolaFusarium lichenicolaFusarium sterilihyphosum

Table 28: Predicted (in silico) Reactivity (Inclusivity) Results for Fusarium

{36}------------------------------------------------

(Cylindrocarpon lichenicola)
Fusarium bullatumFusarium louisianenseFusarium subglutinans
Fusarium camptocerasFusarium lunulosporumFusarium temperatum
Fusarium cerealisFusarium mangiferaeFusarium thapsinum
Fusarium circinatumFusarium meridionaleFusarium udum
Fusarium communeFusarium mesoamericanumFusarium verticillioides*
Fusarium concentricumFusarium mexicanum
Fusarium concolorFusarium musae
Detection Predicted for 85%-94% of target sequences
Fusarium torulosumFusarium xylarioides
Detection Predicted for <85% of target sequences
Fusarium chlamydosporum (66.7%)Fusarium graminearum (59.4%)Fusarium longipes (25.0%)
Fusarium coeruleum (50.0%)Fusarium lateritium (50.0%)Fusarium nelsonii (16.7%)
Detection Not Predicted
Fusarium kyushuenseFusarium sambucinumFusarium venenatum
Fusarium miscanthiFusarium stilboides
Fusarium redolensFusarium succisae

{37}------------------------------------------------

Detection Predicted for ≥95% of target sequences
Rhodotorula araucariaeRhodotorula graminisRhodotorula taiwanensis
Rhodotorula glutinis*Rhodotorula mucilaginosa*
Detection Predicted for 85%-94% of target sequences
None Identified
Detection Predicted for <85% of target sequences
None Identified
Detection Not Predicted
Rhodotorula acheniorumRhodotorula fragariaeRhodotorula marina
Rhodotorula acutaRhodotorula fujisanensisRhodotorula minuta
Rhodotorula armeniacaRhodotorula hinnuleaRhodotorula muscorum
Rhodotorula aurantiacaRhodotorula hordeaRhodotorula nothofagi
Rhodotorula auriculariaeRhodotorula hylophilaRhodotorula philyla
Rhodotorula bacarumRhodotorula ingeniosaRhodotorula phylloplana
Rhodotorula bogoriensisRhodotorula javanicaRhodotorula pilati
Rhodotorula buffoniiRhodotorula lactosaRhodotorula pustula
Rhodotorula ferulicaRhodotorula lignophilaRhodotorula sonckii

Table 29: Predicted (in silico) Reactivity (Inclusivity) Results for Rhodotorula

{38}------------------------------------------------

Analytical Specificity (Cross-Reactivity and Exclusivity)

Cross-reactivity of on-panel and off-panel analytes was evaluated with the BCID-FP Panel. Onpanel organisms were tested in triplicate at concentrations approximating bottle positivity (see Analytical Reactivity (Inclusivity) section of the package insert for additional details). Offpanel organisms were tested at concentrations of ≥1x10° CFU/mL for bacteria and ≥1x107 CFU/mL or spores/mL for fungi unless otherwise noted in Table 30. If the target concentration could not be reached, the organism was diluted 2-fold from stock for use.

No cross reactivity was observed for any of the organisms tested. See Table 30 for a summary of the on-panel strains tested and Table 31 for a summary of off-panel strains tested.

{39}------------------------------------------------

On-Panel Exclusivity

OrganismStrainOrganism
Candida albicansATCC 10231Candida krusei
Candida albicansATCC 90028Candida krusei
Candida albicansATCC MYA-4441Candida lusitaniae
Candida aurisCBS 12766Candida lusitaniae
Candida aurisCDC#385Candida lusitaniae
Candida aurisCDC#386Candida parapsilosis
Candida aurisCDC#387Candida parapsilosis
Candida aurisCDC#388Candida parapsilosis
Candida aurisCDC#389Candida tropicalis
Candida aurisCDC#390Candida tropicalis
Candida dubliniensisATCC MYA-578Candida tropicalis
Candida dubliniensisATCC MYA-579Cryptococcus gattii
Candida dubliniensisATCC MYA-582Cryptococcus gattii
Candida famataATCC 20850Cryptococcus gattii
Candida famataCBA 1961Cryptococcus neoformans
Candida famataCBS 789Cryptococcus neoformans
Candida glabrataATCC 15126Cryptococcus neoformans
Candida glabrataATCC 66032Bisifusarium dimerum
Candida glabrataATCC MYA-2950Fusarium lichenicola(Cylindrocarpon lichenicola)
Candida guilliermondiiATCC 90197Fusarium moniliforme
Candida guilliermondiiATCC 90198Fusarium proliferatum
Candida guilliermondiiATCC 90199Fusarium sacchari
Candida kefyrATCC 204093Fusarium verticillioides
Candida kefyrATCC 2512Rhodotorula glutinis
Candida kefyrATCC 66028Rhodotorula mucilaginosa
Candida kruseiATCC 14243Rhodotorula mucilaginosa
OrganismStrain
Acinetobacter IwoffiiATCC 15309
Acremonium kilienseATCC 4301
Aspergillus fumigatusATCC 204305A
Bacteroides fragilisATCC 25285
Bordetella pertussisATCC 9340
Candida bracarensisCBS 10154
Candida carpophilaCBS 5256
Candida duobushaemuloniiCDC#394
Candida haemuloniiCDC#393
Candida inconspicuaATCC 16783
Candida lambicaATCC 24750
Candida lipolyticaATCC 20177
Candida metapsilosisATCC 96144
Candida nivariensisCBS 9984
Candida norvegensisATCC 22977
Candida orthopsilosisATCC 96139
Candida pelliculosaATCC 10262
Candida rugosaCBS 96275
Candida sakeATCC 22021
Candida utilisATCC 9256
Citrobacter freundiiATCC 6879
Clostridium perfringensATCC 13124
Corynebacterium striatumATCC 7094
Enterobacter aerogenesATCC 29751
Enterobacter cloacaeATCC 23373
Enterococcus faeciumATCC 31282
Exophiala jeanselmeiATCC 12734
Filobasidium elegansCBS 7637
Filobasidium globisporumCBS 7642
Klebsiella oxytocaATCC 43165
Kluyveromyces lactisATCC 10689
OrganismStrain
Kodamaea ohmeriCDC#0396
Lactobacillus rhamnosusATCC 53103
Malassezia furfurATCC 12078
Malassezia furfurATCC 14521
Malassezia furfurCBS 7710
Malassezia globosaATCC MYA-4612
Malassezia restrictaATCC MYA-4611
Malassezia sympodialisATCC 44031
Meyerozyma caribbica(Candida fermentati)ATCC 20296
Micrococcus luteusATCC 19212
Morganella morganiiATCC 25830
Mucor velutinosusATCC MYA-4766
Penicillium marneffeiATCC 200050
Proteus mirabilisATCC 35659
Rhodotorula minutaATCC 36236
Saccharomyces cerevisiaeATCC 18824
Salmonella enterica (Typhi)ATCC 19430
Scedosporium prolificansATCC 200543
Schizosaccharomyces pombeLPY 02387
Serratia marcescensATCC 43861
Sporidiobolus salmonicolorATCC 24217
Sporothrix schenckiiATCC 18616
Staphylococcus hominisATCC 27844
Staphylococcus intermediusATCC 29663
Staphylococcus saprophyticusATCC 15305
Streptococcus agalactiaeATCC 12401
Streptococcus anginosusATCC 9895
Streptococcus pyogenesATCC 12384
Trichosporon asahiiATCC 201110
Trichosporon asteroidesATCC 90043
Trichosporon dermatisATCC 204094

Table 30: On-Panel Organisms Assessed for Cross-reactivity with the ePlex BCID-FP Panel (Exclusivity)

{40}------------------------------------------------

Off-Panel Exclusivity

Table 31: Off-Panel Organisms Assessed for Cross-Reactivity

with the ePlex BCID-FP Panel (Exclusivity)

A. Tested at 1 x 106 spores/mL

{41}------------------------------------------------

Bottle Positivity

Several representative fungal organisms were spiked into blood culture bottles along with the manufacturer's recommended volume of human whole blood and grown to positivity in a commercially-available continuously monitoring blood culture system. Bottles were removed from the incubator within two hours of being identified as positive as well as eight hours after bottle positivity. At least two independent positive blood culture replicates were quantified for each organism on culture plates. Organisms tested and approximate bottle positivity concentrations are summarized in Table 32. Concentrations shown below represent approximate levels that may be observed in a clinical setting. All estimated bottle positivity concentrations are equivalent or greater than the established Limit of Detection (LOD) for each of the assays of the ePlex BCID-FP Panel.

OrganismStrain IDMean Bottle PositivityConcentrationMean Bottle Positivity+8 hours Concentration
Candida albicansATCC 900821.6 x 106 CFU/mL1.4 x 106 CFU/mL
Cryptococcus neoformans var. grubiiATCC 141161.3 x 107 CFU/mL6.5 x 107 CFU/mL
Fusarium solaniATCC 360319.6 x 106 spores/mL7.7 x 106 spores/mL
Rhodotorula mucilaginosaATCC 660341.6 x 106 CFU/mL4.2 x 106 CFU/mL

Table 32: Bottle Positivity Concentrations

{42}------------------------------------------------

Reproducibility

Two positive mixes including 5 on-panel organisms at two concentrations and one negative mix including an off-panel organism were tested. Concentrations in the positive mixes reflected those observed at time of bottle positivity (BP) and time of bottle positivity plus 8 hours or one log higher than that expected at bottle positivity (BP+8) and one mix containing an off-panel organism grown to bottle positivity, which is expected to yield a negative result. Bottle concentrations used in this study are summarized in Table 32. Testing occurred at three sites, with two operators testing the mixes over six days using three cartridge lots.

OrganismBottle PositivityConcentrationBottle Positivity +8Hours Concentration
Candida albicans1 x 106 CFU/mL1 x 107 CFU/mL
Candida kefyr1 x 106 CFU/mL1 x 107 CFU/mL
Cryptococcus neoformans1 x 107 CFU/mL1 x 108 CFU/mL
Fusarium sacchari6.5 x 106 spores/mL6.1 x106 spores/mL
Rhodotorula mucilaginosa1 x 106 CFU/mL1 x 107 CFU/mL

Table 32: Bottle Positivity Concentrations

The percent agreement of each target with the expected result is summarized in Tables 33-37.

{43}------------------------------------------------

Concentration ofCandida albicansSiteAgreement with Expected Results
Agreed / N%95% CI
Bottle Positive+ 8 Hours(1x107 CFU/mL)134/3597.1(85.5-99.5)
235/3697.2(85.8-99.5)
336/36100(90.4-100)
All105/10798.1(93.4-99.5)
Bottle Positive(1x106 CFU/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Negative1108/108100(96.6-100)
2108/108100(96.6-100)
3108/108100(96.6-100)
All324/324100(98.8-100)

Table 33: Percent Agreement for Candida albicans

Table 34: Percent Agreement for Candida kefyr

Concentration ofCandida kefyrSiteAgreement with Expected Results
Agreed / N%95% CI
Bottle Positive+ 8 Hours(1x107 CFU/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Bottle Positive(1x106 CFU/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Negative1107/107100(96.5-100)
2108/108100(96.6-100)
3108/108100(96.6-100)
All323/323100(98.8-100)

{44}------------------------------------------------

Concentration ofSiteAgreement with Expected Results
Cryptococcus neoformansAgreed / N%95% CI
Bottle Positive+ 8 Hours(1x108 CFU/mL)135/35100(90.1-100)
236/36100(90.4-100)
336/36100(90.4-100)
All107/107100(96.5-100)
Bottle Positive(1x107 CFU/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Negative1108/108100(96.6-100)
2108/108100(96.6-100)
3108/108100(96.6-100)
All324/324100(98.8-100)

Table 35: Percent Agreement for Cryptococcus neoformans

Table 36: Percent Agreement for Fusarium

Concentration ofFusarium sacchariSiteAgreement with Expected Results
Agreed / N%95% CI
Bottle Positive+ 8 Hours(6.1x106 spores/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Bottle Positive(6.5x106 spores/mL)136/36100(90.4-100)
236/36100(90.4-100)
336/36100(90.4-100)
All108/108100(96.6-100)
Negative1107/107100(96.5-100)
2108/108100(96.6-100)
3108/108100(96.6-100)
All323/323100(98.8-100)

{45}------------------------------------------------

Concentration ofSiteAgreement with Expected Results
Rhodotorula mucilaginosaAgreed / N%95% CI
Bottle Positive+ 8 Hours(1x107 CFU/mL)136/36100(90.4-100)
234/3694.4(81.9-98.5)
335/3697.2(85.8-99.5)
All105/10897.2(92.1-99.1)
Bottle Positive(1x106 CFU/mL)135/3697.2(85.8-99.5)
236/36100(90.4-100)
335/3697.2(85.8-99.5)
All106/10898.1(93.5-99.5)
Negative1107/107100(96.5-100)
2108/108100(96.6-100)
3108/108100(96.6-100)
All323/323100(98.8-100)

Table 37: Percent Agreement for Rhodotorula

{46}------------------------------------------------

Interfering Substances and Sample Matrix Equivalency (Bottle Evaluation)

Two organism mixes consisting of 5 on-panel organisms and negative blood matrix were used to assess potentially interfering substances and bottle types for interference. The concentration of each organism tested is summarized in Table 38.

OrganismConcentration
Candida albicans1 x 106 CFU/mL
Candida kefyr1 x 106 CFU/mL
Cryptococcus neoformans1 x 107 CFU/mL
Fusarium sacchari6.5 x 106 spores/mL
Rhodotorula mucilaginosa1 x 106 CFU/mL

Table 38: Interfering Substance and Bottle Equivalency Concentrations

Interfering Substances

Eighteen substances were used to assess the ePlex BCID-FP Panel for potential interference. The organisms in Table 38 were spiked into negative blood matrix and tested in triplicate with and without each potentially interfering substance. Negative blood matrix was tested to control for potential positive interference. Potentially interfering substances are summarized in Table 39. None of the eighteen substances commonly found in blood culture specimens or as medications commonly used to treat skin or blood infections were found to inhibit the ePlex BCID-FP Panel at the clinically relevant concentrations. The effect of interfering substances has only been evaluated for the organisms listed in Table 38. Interference due to substances other than those described in this section can lead to erroneous results.

{47}------------------------------------------------

Endogenous SubstancesTesting Concentration
Bilirubin60 µg/mL
Hemoglobin0.6 g/L
Human Genomic DNA6 x 105 copies/mL
Triglycerides1000 mg/dL
γ-globulin0.85 g/dL
Exogenous SubstancesTesting Concentration
Amoxicillin/Clavulanate3.5 µg/mL
Amphotericin B2 µg/mL
Caspofungin5 µg/mL
Ceftriaxone0.23 mg/mL
Ciprofloxacin3 mg/L
Fluconazole25 mg/L
Flucytosine90 µg/mL
Gentamicin sulfate3 µg/mL
Heparin0.9 U/mL
Imipenem83 µg/mL
Sodium Polyanethol Sulfonate(SPS)0.25% w/v
Tetracycline5 mg/L
Vancomycin30 mg/L

Table 39: Potentially Interfering Substances: Substance List

{48}------------------------------------------------

Sample Matrix Equivalency (Bottle Evaluation)

Fifteen bottle types were tested for interference with each of the organisms listed in Table 38. Five replicates of each organism were tested in each of two bottle lots. Negative blood matrix was run as a negative control. Thirteen of the bottle types tested showed no interference for any of the targets tested. One lot of the BACT/Alert® PF Plus bottle type showed lower sensitivity for Rhodotorula and one lot BACT/Alert® FA Plus bottle type showed lower sensitivity for Candida albicans. A summary of the bottle types assessed and the study outcomes is found in Table 40.

{49}------------------------------------------------

ManufacturerBottle BrandBottle TypeStudy Outcome
BDBACTEC™Plus AerobicNo interference observed
BDBACTECPlus AnaerobicNo interference observed
BDBACTECStandard AerobicNo interference observed
BDBACTECStandard AnaerobicNo interference observed
BDBACTECPeds Plus™No interference observed
BDBACTECLytic AnaerobicNo interference observed
BDBACTECMycoNo interference observed
bioMérieuxBACT/ALERT®SA Standard AerobicNo interference observed
bioMérieuxBACT/ALERTSN Standard AnaerobicNo interference observed
bioMérieuxBACT/ALERTFA PlusA false negative result wasobserved for the Candidaalbicans target in one lot.
bioMérieuxBACT/ALERTFN PlusNo interference observed
bioMérieuxBACT/ALERTPF PlusFalse negative results wereobserved for the Rhodotorulatarget in one lot.
bioMérieuxBACT/ALERTMP MycobacteriaNo interference observed
Thermo Scientific™VersaTREK™REDOX™ 1 EZ DrawAerobicNo interference observed
REDOX 2 EZ DrawAnaerobicNo interference observed

Table 40: Sample Matrix Equivalency (Bottle Evaluation) Bottle Types

{50}------------------------------------------------

Carryover and Cross-Contamination

Carryover and cross-contamination were evaluated for the ePlex BCID-FP Panel within and between runs by alternating high positive and negative samples across multiple runs over 5 rounds of testing. Fusarium sacchari was grown to bottle positivity +8 hours and spiked with 1 x 107 CFU/mL Candida albicans to simulate clinically relevant high positive samples for positive testing. Negative blood culture matrix was used to represent negative samples. No false positives were detected, indicating that no carryover or cross-contamination was observed between bays or within bays with the ePlex BCID-FP Panel when testing consecutively or in adjacent bays.

{51}------------------------------------------------

Competitive Inhibition Study

Competitive inhibition was evaluated for the ePlex BCID-FP Panel by pairing twelve clinically relevant organisms (including 9 off-panel organisms) in thirteen simulated dual infection sample mixes. Candida albicans was tested at low titer (concentration expected at bottle positivity) while in the presence of other organisms at higher titer (concentrations expected at 8 hours beyond bottle positivity, or one log higher than that expected at bottle positivity). Candida glabrata and Candida parapsilosis were also tested at the concentration expected at bottle positivity in the presence of Candida albicans at a higher titer concentration. No competitive inhibition was observed in any of the sample mixes evaluated at the concentrations listed in Table 41.

On-panel OrganismsHighConcentrationLowConcentration
Candida albicans1 x 107 CFU/mL1 x 106 CFU/mL
Candida glabrata1 x 107 CFU/mL1 x 106 CFU/mL
Candida parapsilosis1 x 107 CFU/mL1 x 106 CFU/mL
Off-panel OrganismsHigh Concentration
Acinetobacter baumannii1 x 109 CFU/mL
Cutibacterium acnes1 x 109 CFU/mL
Enterococcus faecalis1 x 108 CFU/mL
Escherichia coli1 x 109 CFU/mL
Klebsiella pneumoniae1 x 109 CFU/mL
Staphylococcus aureus1 x 108 CFU/mL
Staphylococcus epidermidis1 x 108 CFU/mL
Streptococcus pneumoniae4 x 108 CFU/mL
Pseudomonas aeruginosa4 x 108 CFU/mL

Table 41: Competitive Inhibition Organisms and Concentrations Tested

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).