K Number
K182667
Date Cleared
2019-06-18

(266 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Med-link Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitors of Philips Model Intelli Vue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL.

These devices are used by qualified medical professional only.

Device Description

The proposed devices are used for patient temperature measurement for multi-patient use. The temperature probes are reusable and consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement.

AI/ML Overview

The Med-link Temperature Probes are intended for monitoring temperature for multi-patient use. They are reusable and designed for use with specific monitors (Philips Model IntelliVue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL).

Here's an analysis of the acceptance criteria and supporting study information:

1. Acceptance Criteria and Reported Device Performance

The core performance acceptance criterion for the Med-link Temperature Probes appears to be accuracy.

Acceptance CriteriaReported Device Performance
Accuracy: ±0.1 °CAccuracy: ±0.1 °C
Measurement Range: 25-45°CMeasurement Range: 25-45°C

Other relevant criteria and performance:

  • Type of Use: Prescription Use
  • Operating Principle: Resistance of thermistor based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement.
  • Measurement Site: Skin, Esophageal, and Rectal
  • Component: Reusable
  • Thermistor Resistance: 2.25K@25°C
  • Sterilization: Non-sterile
  • Electrical Safety: Comply with IEC 60601-1
  • Performance Standard Compliance: ISO 80601-2-56

2. Sample Size and Data Provenance for Test Set

The document does not explicitly state the specific sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for any clinical studies.

The document mentions "performance bench testing in accordance with ISO 80601-2-56 for all models of the subject device with compatible monitors was conducted and the results met the requirements." This indicates that the testing was likely laboratory-based bench testing rather than clinical trials with human subjects.

3. Number of Experts and their Qualifications for Ground Truth

The document does not indicate the involvement of experts in establishing ground truth for a test set, as the testing described is primarily bench testing against recognized standards rather than clinical validation requiring expert interpretation or diagnosis.

4. Adjudication Method for the Test Set

Not applicable, as the document describes bench testing against technical standards, not a study requiring adjudication of expert interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission focuses on device equivalence based on technical standards and performance specifications, not on human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study in the form of bench testing was conducted. The document states: "performance bench testing in accordance with ISO 80601-2-56 for all models of the subject device with compatible monitors was conducted and the results met the requirements." This testing evaluates the device's technical performance against a recognized standard without human interpretation in the measurement process.

7. Type of Ground Truth Used

The ground truth used was established by recognized international technical standards and specifications. Specifically, the device's performance was compared against the requirements of:

  • ISO 80601-2-56: Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
  • The stated accuracy of ±0.1°C and measurement range of 25-45°C are the "ground truth" against which the device's measured performance was assessed during bench testing.

8. Sample Size for the Training Set

The document does not describe any machine learning or AI components that would require a "training set." Therefore, this question is not applicable. The device is a traditional electronic thermometer.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a training set for machine learning or AI.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

June 18, 2019

Shenzhen Med-link Electronics Tech Co., Ltd. Baihan Feng Regulatory Affairs Specialist 4th Floor, Building A, Yingtailong Industrial Park Dalang South Road, Longhua District Shenzhen, Guangdong 518109 China

Re: K182667

Trade/Device Name: Med-link Temperature Probes Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 14, 2019 Received: May 16, 2019

Dear Baihan Feng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Tina Kiang, Ph.D Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182667

Device Name Med-link Temperature Probes

Indications for Use (Describe)

Med-link Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitors of Philips Model Intelli Vue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL.

These devices are used by qualified medical professional only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MED-LINKET. The logo has the word "MED" in red, with a blue swoosh underneath. There is a red circle in the middle of the swoosh. To the right of the circle is the word "LINKET" in blue, with a small "TM" symbol in the upper right corner.

510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. Type of submission: Traditional The assigned 510(k) number is: K182667

1. Submitter information

Manufacturer Name: Shenzhen Med-link Electronics Tech Co., Ltd. Address: 4th Fl, Bldg A, Yingtailong Industrial Park, Dalang South Road, Longhua District, Shenzhen, Guangdong, CHINA, 518109 Tel: 0086-755-61568825 Fax: 0086-755-61120055 Establishment Registration Number: 3006636961

2. Correspondent

Baihan Feng (Regulatory Affairs Specialist, Primary Contact) E-mail: MDL001 @medlinket.com

Liu Fei E-mail: USER22@med-linket.com

    1. Data of Preparation
      14th, May 2019

4. Identification of the Device

Trade Name: Med-link Temperature Probes

Common Name: Temperature Probe

Classification Regulation: 21 CFR 880.2910

Classification Name: Clinical Electronic Thermometer

Product Code: FLL

Class: II

Review Panel: General Hospital

Identification of the Predicate Device 5.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MED Linket. The logo has the letters "MED" in red, with a blue swoosh underneath. To the right of the letters is a red circle, and the word "LINKET" in blue. There is a small "TM" symbol in the upper right corner of the word "LINKET".

No.Device NameCommonNameManufactureClassification andCodeClassificationregulation510(k)number
1Unimed TemperatureProbe (Unimed SkinTemperature Probe,Unimed GeneralPurpose TemperatureProbe)TemperatureProbeUnimedMedicalsupplies Inc.Class II,FLL21 CFR880.2910K121427

Table 1 Predicate Device Information

Indications for Use of the Subject device 6.

Med-link Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitors of Philips Model IntelliVue MP50, Mindray Model PM-9000, Datex-Ohmeda Model Cardiocap II, Drager Model Infinity Gamma XL.

These devices are used by qualified medical professional only.

Device Description 7.

The proposed devices are used for patient temperature measurement for multi-patient use. The temperature probes are reusable and consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement.

NoModelDescriptionMeasurementrangeAccuracyCompatiblemonitors
1W0003AAdult Reusable Skin-surface Probe25-45°C±0.1 °CPhilips
2W0003BAdult Reusable Esophageal/Rectal Probe25-45°C±0.1 °CIntelli VueMP50(K040304)
3W0001AAdult Reusable Skin-surface Probe25-45°C±0.1°CMindray

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MedLinket. The logo features the letters "MED" in red, with a blue orbit-like design surrounding the letters. To the right of the letters is a red circle, followed by the word "LINKET" in blue. The letters "TM" are in superscript to the right of the word "LINKET".

4W0001BAdult Reusable Esophageal/Rectal Probe25-45°C±0.1 °CPM-9000(K070791)
5W0013AAdult Reusable Skin-surface Probe25-45°C±0.1 °CMindray
6W0013BAdult Reusable Esophageal/Rectal Probe25-45°C±0.1 °CBeneVisionN12(K182075)
7W0007AAdult Reusable Skin-surface Probe25-45°C±0.1°CDatex-Ohmeda
8W0007BAdult Reusable Esophageal/Rectal Probe25-45°C±0.1 °CCardiocap II(K912530)
9W0008AAdult Reusable Skin-surface Probe25-45°C±0.1 °CDrager Infinity
10W0008BAdult Reusable Esophageal/Rectal Probe25-45°C±0.1 °CGamma XL(K053484)

8. Non-clinical Test

A series of safety, essential performance and biocompatibility tests were performed to assess the safety and effectiveness of Med-link Temperature Probes. The tests listed below were conducted in accordance with

IEC 60601-1 Medical electrical equipment-Part 1:General requirements for basic safety, and essential performance.

ISO 80601-2-56 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

ISO 10993-5 Biological Evaluation of medical devices-Part 5: Test for in vitro cytotoxicity

ISO 10993-10 Biological evaluation of medical devices- Part 10: Tests For Irritation And Skin Sensitization

  • Cytotoxicity Test
  • Irritation Test .
  • . Sensitization Test

FDA Guidance, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labelling document issued on: March 17, 2015.

9. Comparison to the Predicate Device

ItemProposed DevicePredicate DeviceVerdict
Trade nameMed-link Temperature ProbesUnimed Temperature Probe(Unimed Skin Temperature Probe, Unimed General Purpose Temperature Probe)/
ItemProposed DevicePredicate DeviceVerdict
510(K) SubmitterShenzhen Med-link ElectronicsTech Co., Ltd.Unimed Medical suppliesInc./
510(K) NumberK182667K121427/
ClassificationRegulation21CRF 880.291021CRF 880.2910SE
Classification andCodeClass II,FLLClass II,FLLSE
Common nameTemperature ProbeTemperature ProbeSE
Type of UsePrescriptionPrescriptionSE
Intended useMed-link Temperature Probes areintended to be used formonitoring temperature formulti-patient use. The temperatureprobes are reusable and designedfor use with monitors of PhilipsModel IntelliVue MP50, MindrayModel PM-9000, Datex-OhmedaModel Cardiocap II, DragerModel Infinity Gamma XL.These devices are used byqualified medical professionalonly.Unimed Temperature Probesare intended to be used formonitoring temperature. Thetemperature probes arereusable and designed for usewith monitors of Philips,Marquette, Mindray,Spacelabs, Siemens,Artema/S&W and othermonitors compatible withYSI 400 series temperatureprobes.These devices are indicatedfor used by qualified medicalpersonnel only.SENote 1
OperatingPrincipleResistance of thermistor based onthe metal conductor increaseswith temperature decrease, andthe linear changes to thecharacteristics of the temperaturemeasurement.Resistance of thermistorbased on the metal conductorincreases with temperaturedecrease, and the linearchanges to the characteristicsof the temperaturemeasurement.SE
Measurement SiteSkin, Esophageal and RectalSkin, Esophageal and RectalSE
Measurement25-45°C25-45°CSE
ItemProposed DevicePredicate DeviceVerdict
Range
Accuracy±0.1°C±0.1°CSE
ComponentReusableReusableSE
Thermistorresistance2.25K@25°C2.252K@25°CSE
MaterialCable: PVCProbe end: Epoxy, S304 StainlessSteelNot-providedSENote 2
CompatibleMonitorsPhilips Model IntelliVue MP50,Mindray Model PM-9000,Datex-Ohmeda Model CardiocapII, Drager Model Infinity GammaXL.Philips,Marquette, Mindray,Spacelabs, Siemens,Artema/S&W and othermonitors compatible withYSI 400 series temperatureprobes.SENote 1
SterilizationNon-sterileNon-sterileSE
OperationEnvironmentTemperature: +5°C~+40°C;Atmospheric Pressure: 86 kPa to106 kPaRelative humidity range:0 % to80 %, non-condensing (% RH)Not-providedSENote 3
StorageEnvironmentTemperature:-10°C to +40°CNot-providedSENote 3
Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1SE
PerformanceISO 80601-2-56EN 12470-4SENote 4
BiocompatibilityEvaluationAll the patient contractingmaterials are evaluated by thebiocompatibility standard ISO10993-5, ISO 10993-10.All the patient contractingmaterials are evaluated by thebiocompatibility standardISO 10993-5, ISO 10993-10.SE

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MED LINKET. The word "MED" is in red, and the word "LINKET" is in blue. There is a red circle between the two words. The letters are bold and easy to read.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MEDLINKET. The logo features the letters "MED" in red, with a blue swoosh underneath. To the right of the letters is a red circle, and the word "LINKET" in blue. The letters "TM" are in the upper right corner of the word "LINKET".

Note 1

The compatible monitors of proposed devices are different from the predicate device. The core component of temperature probe in the subject device is Negative Temperature Coefficient (NTC)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for MEDLINKET. The logo features the word "MED" in red, with a blue swoosh underneath. To the right of "MED" is a red circle, followed by the word "LINKET" in blue. The letters "TM" are in the upper right corner of the word "LINKET".

which is identical to the NTC used in predicate device. The NTC determines the accuracy and range of temperature measurement. The performance bench testing in accordance with ISO 80601-2-56 for all models of the subject device with compatible monitors was conducted and the results met the requirements. Therefore, the intended use of proposed devices is the same as predicate device. This difference does not raise different questions of safety or effectiveness for the subject devices.

Note 2

Although patient contact material are different for proposed device and predicate device, they are both complied with ISO 10993-5 and ISO 10993-10. The differences do not affect the safety and effectiveness.

Note 3

Although some specifications of operating & storage conditions are different for proposed device and predicate device, they are both complied with IEC 60601-1 and ISO 80601-2-56. The differences do not affect the safety and effectiveness.

Note 4

EN 12470-4 as old performance standard was replaced by new performance standard ISO 80601-2-56. Therefore, both of them met the applicable performance requirements. The differences do not affect the safety and effectiveness.

10. Conclusion

Based on the comparison and analysis in this submission, it can be concluded that: Med-link Temperature Probes are substantially equivalent to the predicate devices.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.