K Number
K182552
Date Cleared
2018-12-13

(87 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

Device Description

The StatStrip Xpress Glucose Hospital Meter is a hand-held, battery-powered, in vitro diagnostic laboratory instrument that works in conjunction with Nova Biomedical glucose electrochemical test strips to measure glucose in a whole blood sample, Quality Control (QC), linearity, or proficiency solutions. In addition to measuring glucose, the meter stores patient test data, and other information relating to patient sample, operator, reagents, and the meter. The meter can store up to 400 patient and/or quality control test results. The user can review all stored test results on screen. Functions and data selection are accomplished by 3 push buttons. The meter has a built-in beeper for audible alerts and prompts.

The StatStrip Glucose Hospital Meter Test Strips are intended for use with the StatStrip Xpress Glucose Hospital Meter System.

The Nova StatStrip Glucose Control Solutions are used as a quality control check to ensure that the StatStrip Xpress Glucose Hospital Meter and the Glucose Test Strips are working properly as a system.

The Nova StatStrip Glucose Linearity Solutions are used to check the linearity of the Nova StatStrip Xpress Glucose Hospital Meter.

AI/ML Overview

This is a 510(k) summary for the StatStrip Xpress Glucose Hospital Meter System. This document focuses on the claim of substantial equivalence to a predicate device and an expanded indication for use.

Here's an analysis based on your request, but it's important to note that this document is primarily a clearance letter and 510(k) summary, not a detailed study report. Therefore, much of the information regarding specific study methodologies (like sample sizes for test sets, expert details, adjudication methods, MRMC studies, or standalone performance study details) is not present in this document.

Device: StatStrip Xpress Glucose Hospital Meter System

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a table format for performance metrics. Instead, it relies on the device being "substantially equivalent" to a predicate device and having an expanded indication that was already cleared for the predicate.

The reported performance characteristics (which are the same as the predicate device) are:

ItemReported Device Performance (StatStrip Xpress Glucose Hospital Meter System)
Intended UseQuantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens in all hospital and professional healthcare settings, including patients receiving intensive medical intervention/therapy. (Note: This is the expanded indication based on the predicate device's clearance).
Operating PrincipleElectrochemical biosensor
Measuring TechnologyEnzyme, Amperometric Glucose Enzyme (Aspergillus sp., >1.0 IU)
Sample TypeCapillary whole blood (finger stick), venous whole blood, arterial whole blood, neonatal heel stick, and neonatal arterial whole blood samples
Sample Size1.2 μL
Sample ApplicationTest strip capillary draw
Measuring Range10-600 mg/dL
Hematocrit Range20-65%
Reported Outputmg/dL
Time to Result~ 6 seconds
CalibrationAutomatic, no Calibration Code
Test Strip Active ReagentGlucose Oxidase
Quality Control3 levels
Linearity5 levels
HandheldYes
Data Storage400 patient, quality control, linearity, and proficiency tests (Note: This differs from the predicate which stored 1000 patient tests, 200 QC, 4000 operators)
BarcodeNo (Note: This differs from the predicate which had a barcode feature)
Power Source3V dc Li coin cell battery (Note: This differs from the predicate which used a Rechargeable 3.7 volt Lithium battery)
Operating Temp41° to 104°F (5° to 40°C) (Note: This differs from the predicate which was 59°F - 104°F (15°C - 40°C))

Note on Acceptance Criteria: The primary "acceptance criterion" for this 510(k) submission is demonstrating substantial equivalence to the predicate device (StatStrip Glucose Hospital Meter System, K181043) and justifying that the expanded indication (for capillary whole blood specimens in all hospital and healthcare settings) was already cleared for the predicate. The document states "No changes have been made to the candidate device, the software, the test strips, controls or linearity solutions since the K161856 clearance therefore; there are no additional safety and efficacy concerns to the device." This implies that the performance data previously used to clear K161856 and K181043 are sufficient. Specific numerical acceptance criteria for accuracy (e.g., % within +/-X mg/dL) are not detailed here.

2. Sample Size for the Test Set and Data Provenance

The document does not provide details on the sample size used for a specific test set in this submission. It focuses on the fact that no changes were made to the device itself since previous clearances (K161856), and the expanded indication was already cleared for the predicate (K181043). Therefore, specific new test data for this 510(k) submission are not presented.

The data provenance is not mentioned. Typically, clinical studies for medical devices are conducted in multiple geographical locations sometimes, but this is not specified here. It's likely within the US, but not confirmed. The document does not specify if data was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth

This information is not provided in this 510(k) summary. For a blood glucose meter, the "ground truth" (reference method) is usually a laboratory-based glucose analyzer, not an expert consensus.

4. Adjudication Method for the Test Set

This information is not applicable/provided in this 510(k) summary. Adjudication methods are typically relevant for subjective assessments, such as image interpretation, not for quantitative measurements like blood glucose.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study is not applicable for a blood glucose meter, as it is a quantitative measurement device, not an interpretation device that benefits from human readers.

6. Standalone (Algorithm Only) Performance Study

This document describes a device (blood glucose meter and test strips) that performs a direct measurement. Its "standalone performance" is its accuracy and precision against a reference method. While the document implies this was evaluated for previous clearances, no specific standalone study details (e.g., accuracy against a lab reference method) are provided within this particular 510(k) summary. It refers to prior clearances (K150461 and K161856) for performance.

7. Type of Ground Truth Used

For blood glucose measurements, the ground truth is typically established by a laboratory reference method (e.g., a central laboratory analyzer using an enzymatic reference method). This is standard for glucose meters.

8. Sample Size for the Training Set

This information is not provided in this 510(k) summary. Given that the device relies on an electrochemical biosensor and enzyme-based measurement, rather than a complex algorithm developed through machine learning, the concept of a "training set" in the context of AI/ML is not directly applicable. The device's calibration and performance are established through manufacturing controls, quality control, and clinical validation studies (which typically would use patient samples, but these are not referred to as a "training set" for an algorithm in this context).

9. How the Ground Truth for the Training Set Was Established

As mentioned above, the concept of a "training set" in the AI/ML sense is not directly applicable here. The ground truth for establishing the performance of such a device would be based on comparison to a highly accurate laboratory reference method for glucose measurement, which would be part of the validation for the electrochemical biosensor system. Specific details for this device are not in this document.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 13, 2018

Nova Biomedical Corporation Khushboo Surendran 200 Prospect Street Waltham, MA 02454

Re: K182552

Trade/Device Name: StatStrip Xpress Glucose Hospital Meter System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: PZI Dated: September 14, 2018 Received: September 17, 2018

Dear Khushboo Surendran:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182552

Device Name

StatStrip Xpress Glucose Hospital Meter System

Indications for Use (Describe)

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healtheare settings, including patients receiving intensive medical intervention/therapy.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

510(k) Number: K182552

510(k) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.
Waltham, MA 02454
Fax Number:784-891-4806
Contact Person:Khushboo Surendran
Phone:781-894-0800
Date Prepared:September 14, 2018

Proprietary Name: StatStrip Xpress Glucose Hospital Meter System

Common or Usual Name: Blood Glucose Meter

Product Code: PZI

Classification Name:

Classification NameClass No.Regulation No.Class
Prescription Use Blood Glucose Meter ForNear-Patient Testing75 PZI862.1345II

Predicate Device: StatStrip Glucose Hospital Meter System, K181043.

Device Description:

The purpose of this submission is to change the labeling the indication for use statement to include the use of capillary whole blood specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy. This expanded indication was previously cleared for the predicate device, StatStrip Glucose Hospital Meter System in K181043. No changes have been made to the candidate device, the test strips, controls or linearity solutions since the clearance in K161856.

Meter:

The StatStrip Xpress Glucose Hospital Meter is a hand-held, battery-powered, in vitro diagnostic laboratory instrument that works in conjunction with Nova Biomedical glucose electrochemical test strips to measure glucose in a whole blood sample, Quality Control (QC), linearity, or proficiency solutions. In addition to measuring glucose, the meter stores patient test data, and other information relating to patient sample, operator, reagents, and the meter. The meter can store up to 400 patient and/or quality control test results. The user can review all stored test results on screen. Functions and data selection are accomplished by 3 push buttons. The meter has a built-in beeper for audible alerts and prompts.

Test Strips:

The StatStrip Glucose Hospital Meter Test Strips are intended for use with the StatStrip Xpress Glucose Hospital Meter System. No changes have been made to the form, fit, function or packaging of StatStrip Glucose Hospital Meter Test Strips since K070960 clearance and they are not a subject of this submission.

Control Solutions:

The Nova StatStrip Glucose Control Solutions are used as a quality control check to ensure that the StatStrip Xpress Glucose Hospital Meter and the Glucose Test Strips are working properly as a system.

{4}------------------------------------------------

No changes have been made to the form, fit, function or packaging of the Nova StatStrip Glucose Control Solution since K070960 clearance and they are not a subject of this submission.

Linearity Solutions:

The Nova StatStrip Glucose Linearity Solutions are used to check the linearity of the Nova StatStrip Xpress Glucose Hospital Meter. No changes have been made to the form, fit, function or packaging of the Nova StatStrip Glucose Linearity Solution since K070960 clearance and they are not a subject of this submission.

Principle of Test

The Test Strip is designed with an electrode that measures glucose (sugar) in the blood sample mixes with reagent on the test strip that produces an electric current. The amount of current that is produced depends on how much glucose is in the blood. The StatStrip Xpress Glucose Hospital Meter System with StatStrip Glucose Hospital Meter Test Strip measures the glucose level in blood. The glucose result is displayed on the screen of the meter. The test strip is calibrated against plasma.

Intended Use:

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

lt is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysqlycemia.

Summary of the Technological Characteristics:

The StatStrip Xpress Glucose Hospital Meter System uses the exact same technology, operating principle, software measurement algorithm and user workflow as the predicate device, StatStrip Glucose Hospital Meter System (K181043).

The StatStrip Xpress Glucose Hospital Meter System was recently cleared under K150461 by demonstrating substantially equivalent performance to the StatStrip Glucose Hospital Meter System.

The only proposed change is to the labeling of the device to allow the product to be used on capillary whole blood specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.

{5}------------------------------------------------

Comparison of Predicate and Candidate devices

ItemPredicate Device (K181043)StatStrip Glucose Hospital MeterSystemCandidate DeviceStatStrip Xpress Glucose Hospital MeterSystem
Intended UseThe StatStrip Glucose Hospital Meter Systemis intended for point-of-care, in vitrodiagnostic, multiple-patient use for thequantitative determination of glucose incapillary finger stick, venous whole blood,arterial whole blood, neonate arterial wholeblood, and neonate heel stick specimensthroughout all hospital and all professionalhealthcare settings, including patientsreceiving intensive medical interven-tion/therapy.The system should only be used with single-use, auto-disabling lancing devices whenperforming a capillary finger stick or neonateheel stick.It is not intended for use with neonate cordblood specimens.It is not intended for the screening ordiagnosis of diabetes mellitus but is indicatedfor use in determining dysglycemia.The StatStrip Xpress Glucose Hospital MeterSystem is intended for point-of-care, in vitrodiagnostic, multiple-patient use for thequantitative determination of glucose incapillary finger stick, venous whole blood,arterial whole blood, neonate arterial wholeblood, and neonate heel stick specimensthroughout all hospital and all professionalhealthcare settings, including patientsreceiving intensive medical interven-tion/therapy.The system should only be used with single-use, auto-disabling lancing devices whenperforming a capillary finger stick or neonateheel stick.It is not intended for use with neonate cordblood specimens.It is not intended for the screening ordiagnosis of diabetes mellitus but is indicatedfor use in determining dysglycemia.
OperatingPrincipleElectrochemical biosensorSame
MeasuringTechnologyEnzyme, Amperometric Glucose Enzyme(Aspergillus sp., >1.0 IU)Same
Sample typeCapillary whole blood (finger stick), venouswhole blood, arterial whole blood, neonatalheel stick, and neonatal arterial whole bloodsamplesSame
Sample size1.2 μLSame
SampleapplicationTest strip capillary drawSame
Measuringrange10-600 mg/dLSame
Hematocritrange20-65%Same
ItemPredicate Device (K181043)StatStrip Glucose Hospital MeterSystemCandidate DeviceStatStrip Xpress Glucose Hospital MeterSystem
Reportedoutputmg/dLSame
Time to Result~ 6 secondsSame
CalibrationAutomatic, no Calibration CodeSame
Test stripactive reagentGlucose OxidaseSame
Quality Control 3 levelsSame
Linearity5 levelsSame
HandheldYesSame
Data storage1000 patient tests, 200 QC tests, 4000Operators400 patient, quality control,linearity, and proficiency tests
BarcodeYesNo
Power sourceRechargeable 3.7 volt Lithium battery3v dc Li coin cell battery
OperatingTemp59°F - 104°F (15°C - 40°C)41° to 104°F (5° to 40°C)

{6}------------------------------------------------

Conclusion:

FDA had previously determined that the StatStrip Xpress Glucose Hospital Meter System is substantially equivalent to that of the predicate StatStrip Glucose Hospital Meter System (K070960, K152461). The expanded indication proposed in this submission was previously cleared for the predicate device (K181043). No changes have been made to the candidate device, the software, the test strips, controls or linearity solutions since the K161856 clearance therefore; there are no additional safety and efficacy concerns to the device.

The candidate device with the proposed revised labeling is substantially equivalent to the predicate StatStrip Glucose Hospital Meter System (K181043), and supports a CLIA waiver designation.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.