K Number
K150461

Validate with FDA (Live)

Date Cleared
2015-05-20

(86 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens.

The StatStrip Xpress Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal heel stick and neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

Device Description

The StatStrip Xpress Glucose Hospital Meter System is a hand-held testing device that works in conjunction with the StatStrip Glucose Test Strips. Meter operation is self-prompting using a segmented liquid crystal display (LCD) and icons. Function and data selection is accomplished using 3 push buttons. The handheld meter supports audible alerts and prompts with a built-in beeper. In addition to measuring glucose, the meter also stores up to 400 patient test records. The user can recall and review test results. A single coin battery powers the device, and is expected to perform up to 600 tests before needing to be replaced. A low battery prompt will appear when it is time to replace the battery. All test data is stored in a non-volatile form to prevent data loss.

The StatStrip Glucose Hospital Meter Test Strips are designed with an electrode that measures Glucose levels. Glucose in the blood sample mixes with reagent on the test strip that produces an electric current. The amount of current that is produced depends on how much Glucose is in the blood. The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result.

The StatStrip Control Solutions are aqueous solutions that contain no products of human origin. There are three levels of control solutions (Levels 1-3). StatStrip Glucose Control Solutions are intended to verify the accuracy of the blood Glucose test results.

The StatStrip Glucose Linearity Solutions are aqueous materials with a known concentration of glucose intended to verify performance of the Nova Biomedical Analyzers. There are five levels of linearity solutions (Levels 1-5). The Linearity Solution is intended for monitoring the performance (linearity) of the StatStrip Glucose Hospital Meter System.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the StatStrip Xpress Glucose Hospital Meter System, based on the provided text:

Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It specifically states: "The StatStrip Xpress Glucose Hospital Meter System that is the subject of this submission is identical in form, fit and function, to the device originally cleared in K070960." And also, "The StatStrip Glucose Hospital Meter Test Strips that are the subject of this submission are identical in form, fit, function and packaging, to the qlucose test strips originally cleared for use with the system in K070960."

Therefore, the performance testing described is likely a verification that the new model performs identically to the predicate, rather than a full de novo validation of a completely new device. This means the detailed acceptance criteria and standalone performance would have been established during the clearance of the predicate device (K132121) or the original device (K070960). The current document reports that the new device meets the same performance standards as that predicate.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" for the StatStrip Xpress Glucose Hospital Meter System in this 510(k) submission. Instead, it highlights that the device is "identical in form, fit and function" to its predicate and that "The results of the performance verification testing confirmed that the StatStrip Xpress Glucose Hospital Meter System is safe and effective for its intended purpose and that the StatStrip Xpress Glucose Hospital Meter System is substantially equivalent to the predicate StatStrip Glucose Hospital Meter System (K132121)."

The comparison table (Table 2-1) emphasizes the sameness between the predicate and the modified device for key technical characteristics, implying that the performance met the same standards as the predicate.

Here's an inferred table based on the common characteristics of blood glucose meters and the affirmation of substantial equivalence:

CharacteristicAcceptance Criteria (Inferred from Predicate's Performance)Reported Device Performance (StatStrip Xpress)
Measuring Range10-600 mg/dL (as per predicate)Same (10-600 mg/dL)
Hematocrit Range20-65% (as per predicate)Same (20-65%)
Sample Size1.2 µL (as per predicate)Same (1.2 µL)
Time to Result~6 seconds (as per predicate)Same (~6 seconds)
Accuracy/PrecisionMeets the clinical and analytical performance claimed for the predicate device (K132121 and K070960). While specific metrics (e.g., % within +/-15mg/dL) are not given here, this would be a core requirement."The results of the performance verification testing confirmed that the StatStrip Xpress Glucose Hospital Meter System is safe and effective for its intended purpose and that the StatStrip Xpress Glucose Hospital Meter System is substantially equivalent to the predicate..."
CalibrationAutomatic, no calibration code (as per predicate)Same (Automatic, no Calibration Code)
Quality ControlOperates with 3 levels for QC (as per predicate)Same (3 levels)
LinearityOperates with 5 levels for linearity (as per predicate)Same (5 levels)
Intended Use AdherenceMatches the predicate's intended useMatches the predicate's intended use

2. Sample Size Used for the Test Set and Data Provenance

The document states: "The results of the performance verification testing confirmed that the StatStrip Xpress Glucose Hospital Meter System is safe and effective for its intended purpose and that the StatStrip Xpress Glucose Hospital Meter System is substantially equivalent to the predicate StatStrip Glucose Hospital Meter System (K132121)."

However, the specific sample sizes used for the performance verification testing of the StatStrip Xpress Glucose Hospital Meter System are not detailed in this 510(k) summary. It refers to "performance verification testing" but does not provide details on the number of samples, patients, or the specific types of clinical studies (e.g., prospective, retrospective, country of origin).

Given the context of a 510(k) where the device is described as "identical in form, fit and function" to a cleared predicate, the new submission primarily validates that the manufacturing process or minor changes do not negatively impact performance, usually through bench testing and often a limited clinical bridging study. The extensive clinical performance data would have been submitted with the original (K070960) or predicate (K132121) devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. For blood glucose meters, the "ground truth" (or reference method) is typically established by laboratory-based glucose analyzers, which are themselves highly standardized and calibrated. Expert consensus is generally not the primary method for establishing ground truth in this context.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in imaging studies where subjective expert interpretation is required. For a quantitative device like a blood glucose meter, the ground truth is established by a reference laboratory measurement, and thus adjudication methods are not applicable nor mentioned in this context.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study is mentioned. MRMC studies are relevant for reader-dependent diagnostic tools, often in imaging. A blood glucose meter is a direct quantitative measurement device, not one that relies on human interpretation for its primary output.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

For a blood glucose meter, the device itself is the "standalone" algorithm. The performance of the meter system (meter + test strip) is what is always evaluated. Therefore, yes, a standalone performance study was implicitly done as part of the "performance verification testing" where the device's measurements are compared against a reference method. The results of this testing led to the conclusion of substantial equivalence.

7. Type of Ground Truth Used

The ground truth for blood glucose measurements is almost universally established by a reference laboratory method, typically an enzymatic reference method (e.g., hexokinase method) performed on a laboratory analyzer, which is highly accurate and precise. While not explicitly stated for this specific submission's "performance verification," this is the standard for glucose meter ground truth. The document supports this by stating the device is "intended to provide plasma equivalent results to laboratory methods."

8. Sample Size for the Training Set

The document is a 510(k) submission for a device described as "identical in form, fit and function" to a predicate. This implies that the underlying algorithms and device design were already well-established and "trained" during the development of the original (K070960) and predicate (K132121) systems.

Therefore, the concept of a separate "training set" for this specific submission is not directly applicable or discussed. Any "training" would have occurred during the development phase of the prior, substantially equivalent devices, not for this specific K150461 submission, which verified performance of an identical design.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, the notion of a 'training set' for this specific submission is not relevant as it's a re-submission of an "identical" device. For the original development of such a device, the ground truth for any data used in algorithm development (if machine learning was involved, which is unlikely for a simple electrochemical biosensor calculation) would also have been established via reference laboratory methods, as described in point 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 1, 2019

Nova Biomedical Corporation Paul MacDonald Chief Quality Assurance and Regulatory Affairs Officer 200 Prospect Street Waltham, MA 02454-9141

Re: K150461

Trade/Device Name: StatStrip Xpress Glucose Hospital Meter System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: PZI Dated: February 20, 2015 Received: February 23, 2015

Dear Paul MacDonald:

This letter corrects our substantially equivalent letter of May 20, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150461

Device Name

StatStrip Xpress Glucose Hospital Meter System

Indications for Use (Describe)

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens.

The StatStrip Xpress Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

510(K) Owner:Nova Biomedical Corporation
Registration Number:1219029
Address:200 Prospect St.Waltham, MA 02454
Phone:781-894-0800
Fax Number:784-891-4806
Contact Person:Paul W. MacDonald
Date Prepared:8 May, 2015

Proprietary Name: StatStrip Xpress Glucose Hospital Meter System

Common or Usual Name: Blood Glucose Meter

Classification Name: Multiple

Classification Names:Class No.Reg. No.Class
Glucose Oxidase75CGA862.1345II

Product Codes: CGA, Glucose Oxidase, Glucose

Predicate Device: K132121 - StatStrip Glucose Hospital Meter System

Device Description:

Intended Use:

The StatStrip Xpress Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens.

The StatStrip Xpress Glucose Hospital Meter System is also intended for use in the quantitative determination of glucose in venous whole blood, arterial whole blood, neonatal heel stick and neonatal arterial whole blood samples throughout all hospital and all professional healthcare settings.

The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.

It is not intended for use with neonate cord blood specimens.

It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.

{4}------------------------------------------------

Summary of the Technological Characteristics:

Hospital Meter

The StatStrip Xpress Glucose Hospital Meter System that is the subject of this submission is identical in form, fit and function, to the device originally cleared in K070960. The StatStrip Xpress Glucose Hospital Meter System is specifically designed to meet the bedside and point-of-care glucose testing needs in today's hospital environment. The system is intended for in vitro diagnostic use by health care professionals for both clinical and point-of-care usage for the quantitative determination of glucose in whole blood. It is intended to provide plasma equivalent results to laboratory methods. The StatStrip Xpress Glucose Hospital Meter System is intended for use in a clinical setting by healthcare professionals as an aid to monitor qlucose levels in the management of dysqlycemia.

The StatStrip Xpress Glucose Hospital Meter is a hand-held testing device that works in conjunction with the StatStrip Glucose Test Strips. Meter operation is self-prompting using a segmented liguid crystal display (LCD) and icons. Function and data selection is accomplished using 3 push buttons. The handheld meter supports audible alerts and prompts with a built-in beeper. In addition to measuring glucose, the meter also stores up to 400 patient test records. The user can recall and review test results.

A single coin battery powers the device, and is expected to perform up to 600 tests before needing to be replaced. A low battery prompt will appear when it is time to replace the battery. All test data is stored in a non-volatile form to prevent data loss.

Test Strips

The StatStrip Glucose Hospital Meter Test Strips that are the subject of this submission are identical in form, fit, function and packaging, to the qlucose test strips originally cleared for use with the system in K070960. The Test Strip is designed with an electrode that measures Glucose levels. Glucose in the blood sample mixes with reagent on the test strip that produces an electric current. The amount of current that is produced depends on how much Glucose is in the blood.

The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be sold in cartons of 100 strips (50 strips/vial).

Contents:

Each glucose test strips contain a reaction layer that contains glucose enzyme (Aspergillus sp.) >1.0 IU, mediator >20 µg, and other nonreactive substances.

Function:

StatStrip Glucose Test Strips are intended for quantitative determination of Glucose in fresh whole blood specimens. StatStrip Glucose Test Strips are for use only with the StatStrip Family of Meters.

Storage Conditions:

Store the test strips in the vial between 34-86°F (1-30°C); between 10-90% relative humidity (non-condensing). Ensure that the vial is closed between uses. Once opened the test strips in the vial may be used for 180 days or until the expiration date printed on the label, which ever comes first.

Controls and Linearity Solutions

The StatStrip Control and Linearity Solutions that are intended for use with the system are identical in formulation and packaging, to the Control and Linearity Solutions originally cleared for use with the system in K070960.

{5}------------------------------------------------

Control Solutions

The StatStrip Control Solutions are aqueous solutions that contain no products of human origin. The controls solutions are for use with all of the StatStrip Meters.

Contents

Each vial contains approximately 4mLs of a buffered aqueous control solution containing glucose, ß-ketone, preservative, viscosity-adjusting agent and other non-reactive ingredients (dye).

There are three levels of control solutions (Levels 1-3).

StatStrip Glucose Control Solutions are intended to verify the accuracy of the blood Glucose test results.

Storage Conditions

Store the control solution in the vial at room temperature. Ensure that the vial is closed between uses. Once opened the control solution may be used for up to 9 months or until the expiration date printed on the label, which ever comes first.

Linearity Solutions

The StatStrip Glucose Linearity Solutions are aqueous materials with a known concentration of glucose intended to verify performance of the Nova Biomedical Analyzers. Assay values for expected ranges are included on every bottle of linearity standards. If the results obtained are outside the expected range, the system may not be performing correctly.

Contents

Each vial contains 4mLs of a buffered glucose, ß-ketone, preservative, viscosity-adjusting agent and other non-reactive ingredients (dye). They contain no products of human origin.

There are five levels of linearity solutions (Levels 1-5).

Function

The Linearity Solution is intended for monitoring the performance (linearity) of the StatStrip Glucose Hospital Meter System.

Storage conditions:

Store the linearity solution in the vial at room temperature (below 86°F/30°C). Ensure that the vial is closed between uses. Once opened the control solution may be used for up to 3 months or until the expiration date printed on the label, whichever comes first.

Summary of Performance Testing:

The results of the performance verification testing confirmed that the StatStrip Xpress Glucose Hospital Meter System is safe and effective for its intended purpose and that the StatStrip Xpress Glucose Hospital Meter System is substantially equivalent to the predicate StatStrip Glucose Hospital Meter System (K132121).

{6}------------------------------------------------

Table 2-1: Comparison of Predicate and Proposed devices
ItemPredicate Device (K132121)Modified Device
StatStrip Glucose Hospital Meter SystemStatStrip Xpress Glucose Hospital Meter System
Intended UseThe StatStrip Glucose Hospital Meter System is intended forpoint-of-care, in vitro diagnostic, multiple-patient use for thequantitative determination of glucose in capillary finger stick,venous whole blood, arterial whole blood, neonate arterialwhole blood and neonate heel stick specimens.The StatStrip Glucose Hospital Meter System is alsointended for use in the quantitative determination of glucosein venous whole blood, arterial whole blood, neonatal heelstick and neonatal arterial samples throughout all hospitaland all professional healthcare settings.The system should only be used with single-use, auto-disabling lancing devices when performing a capillary fingerstick or neonate heel stick.It is not intended for use with neonate cord blood specimensIt is not intended for the screening or diagnosis of diabetesmellitus but is indicated for use in determining dysglycemia.The StatStrip Xpress Glucose Hospital Meter System isintended for point-of-care, in vitro diagnostic, multiple-patientuse for the quantitative determination of glucose in capillaryfinger stick, venous whole blood, arterial whole blood,neonate arterial whole blood and neonate heel stickspecimens.The StatStrip Xpress Glucose Hospital Meter System is alsointended for use in the quantitative determination of glucosein venous whole blood, arterial whole blood, neonatal heelstick and neonatal arterial samples throughout all hospitaland all professional healthcare settings.The system should only be used with single-use, auto-disabling lancing devices when performing a capillary fingerstick or neonate heel stick.It is not intended for use with neonate cord blood specimensIt is not intended for the screening or diagnosis of diabetesmellitus but is indicated for use in determining dysglycemia.
Operating PrincipleElectrochemical biosensorSame
Sample typeCapillary whole blood (finger stick), venous whole blood,arterial whole blood, neonate heel stick, and neonate arterialwhole blood specimensVenous whole blood, arterial whole blood, neonatal heelstick, and neonatal arterial whole blood samples throughoutall hospital and all professionalSame
Sample size1.2 µLSame
Sample applicationTest strip capillary drawSame
Measuring range10-600 mg/dLSame
ItemPredicate Device (K132121)Modified Device
StatStrip Glucose Hospital Meter SystemStatStrip Xpress Glucose Hospital Meter System
Hematocrit range20-65%Same
Reported outputmg/dLSame
Time to Result~ 6 secondsSame
CalibrationAutomatic, no Calibration CodeSame
Test strip activereagentGlucose OxidaseSame
Quality Control3 levelsSame
Linearity5 levelsSame
Handheld?YesSame
Data storage1000 patient test results and 200 QC test results400 test results
BarcodeYesNo
Power sourceRechargeable 3.7 volt Lithium battery3v dc Li coin cell battery

{7}------------------------------------------------

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.