K Number
K070960

Validate with FDA (Live)

Date Cleared
2007-07-11

(97 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nova StatStrip Xpress Glucose Hospital Meter System is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, arterial, and neonate whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control.

Nova StatStrip Glucose Test Strips are intended for use only with Nova StatStrip Glucose Hospital Meters for quantitative tests. The glucose meter is intended to quantitatively measure glucose (sugar) in whole blood. The Glucose Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatStrip Glucose Test Strips are for testing outside the body (in vitro diagnostic use only).

Nova StatStrip Control Solutions is intended for use with Nova StatStrip Glucose Hospital Meters and Nova StatStrip Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Level 2, and Level 3). These solutions will be offered for sale separately from the meter.

Nova StatStrip Glucose Linearity Kit solutions are used to check the linearity of Nova StatStrip Glucose Hospital Meter Systems. There are five levels of linearity solutions, (Level 2, Level 3, Level 4, and Level 5).

Device Description

Not Found

AI/ML Overview

I am sorry, but the provided text does not contain information about acceptance criteria, device performance, or any detailed study results. The document is a 510(k) premarket notification letter from the FDA, along with an "Indications for Use" statement for the Nova StatStrip Xpress Glucose Hospital Meter.

It states that the FDA has reviewed the notification and determined the device is substantially equivalent to legally marketed predicate devices. It grants permission to market the device subject to general controls provisions of the Act.

Therefore, I cannot extract the specific information requested in your prompt regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 1 1 2007

Nova Biomedical Corp. c/o Dr. Paul W. MacDonald Chief Quality Assurance and Regulatory Affairs Officer 200 Prospect Street Waltham, MA 02454-9141

Re: K070960

Trade/Device Name: Nova StatStrip Xpress Glucose Hospital Meter Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: Class II Product Code: CGA Dated: June 19, 2007 Received: June 20, 2007

Dear Dr. MacDonald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{1}------------------------------------------------

Page 2 -

This letter will allow you to hegin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Coopes, M.S., D.V.M.

Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): K070960

Device Name: Nova StatStrip Xpress Glucose Hospital Meter

Indications For Use:

The Nova StatStrip Xpress Glucose Hospital Meter System is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of glucose in capillary, venous, arterial, and neonate whole blood. It is indicated for use in a clinical setting by healthcare professionals as an aid to monitor the effectiveness of diabetes control.

Nova StatStrip Glucose Test Strips are intended for use only with Nova StatStrip Glucose Hospital Meters for quantitative tests. The glucose meter is intended to quantitatively measure glucose (sugar) in whole blood. The Glucose Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatStrip Glucose Test Strips are for testing outside the body (in vitro diagnostic use only).

Nova StatStrip Control Solutions is intended for use with Nova StatStrip Glucose Hospital Meters and Nova StatStrip Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Level 2, and Level 3). These solutions will be offered for sale separately from the meter.

Nova StatStrip Glucose Linearity Kit solutions are used to check the linearity of Nova StatStrip Glucose Hospital Meter Systems. There are five levels of linearity solutions, (Level 2, Level 3, Level 4, and Level 5).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Page 1 of 1

Office of In Vitro Diagnostic Device

510(k) K070960

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.