(329 days)
Yes
The device description explicitly mentions an "image recognition algorithm running on the Backend" which is a common application of AI/ML in medical imaging analysis.
No
The device is described as an in-vitro diagnostic system intended for analysis of urine to aid in diagnosis, not for treatment or therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states, "The ACR | LAB Urine Analysis Test System is intended for in-vitro diagnostic use by a healthcare professional in a point of care setting. These results may be used in conjunction with clinical evaluation as an aid in the diagnosis for kidney function."
No
The device description explicitly states it is comprised of a smartphone application, a proprietary Color-Board, and ACR Reagent Strips, indicating it includes hardware components (Color-Board and Reagent Strips) in addition to the software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The "Intended Use / Indications for Use" section explicitly states: "The ACR | LAB Urine Analysis Test System is intended for in-vitro diagnostic use by a healthcare professional in a point of care setting."
- Function: The device is designed to perform a test on a biological sample (urine) outside of the body ("in-vitro") to provide information for diagnosis ("aid in the diagnosis for kidney function").
- Components: The system includes reagent strips, which are a common component of IVD tests.
- Setting and User: It is intended for use by a healthcare professional in a point-of-care setting, which is a typical environment for IVD testing.
The description clearly aligns with the definition of an In Vitro Diagnostic device.
No
The letter does not state that the FDA has reviewed and approved or cleared a Predetermined Change Control Plan (PCCP) for this specific device.
Intended Use / Indications for Use
The ACR | LAB Urine Analysis Test System is comprised of a smartphone application, a proprietary Color-Board, and ACR Reagent Strips. It is intended for the semi-quantitative detection of albumin and creatinine in urine, as well as the presentation of their ratio. The ACR | LAB Urine Analysis Test System is intended for in-vitro diagnostic use by a healthcare professional in a point of care setting. These results may be used in conjunction with clinical evaluation as an aid in the diagnosis for kidney function.
Product codes (comma separated list FDA assigned to the subject device)
JFY, JIR, KQO
Device Description
The ACR | LAB Urine Analysis Test System is comprised of a smartphone application, a proprietary Color-Board and ACR Reagent Strips. It is intended for the semi-quantitative detection of albumin and creatinine in urine, as well as the presentation of their ratio.
The device is provided as a kit that is comprised of a canister of 100 FDA-cleared urine test strips (ACON Laboratories Inc. Mission Urinalysis Reagent Strips (Microalbumin/Creatinine) K150330), 10 Color-Boards, and a User Manual. The ACR | LAB Urine Analysis Test System also consists of a smartphone application for use on iPhone 7 device (iOS 12), and an image recognition algorithm running on the Backend*.
The software component of the ACR | LAB consists of both an application (App) and a Backend server (Backend). The App instructs the professional user how to accurately perform the test. The App conducts a series of boundary condition analyses, and if the scan is approved, sends the information to the Backend for complete analysis and results classification. Once analyzed, the results are securely transmitted to a patient Electronic Medical Record for review by a healthcare professional. The patients do not have access to the results at any point during the testing process.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Reflectance Photometry
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
healthcare professional in a point of care setting.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Analytical Performance Testing:
- Precision: Comprised of two sub-studies:
- Repeatability:
- Study Type: Repeatability
- Sample Size: 189 total tests (3 sites X 3 solutions X 3 lots X 7 strips X 1 phones)
- Key Results: Recorded an exact match of 100% for each analyte.
- Reproducibility:
- Study Type: Reproducibility
- Sample Size: 1,080 total tests (3 sites x 3 solutions X 3 lots x 20 days x 1 device)
- Key Results: Recorded an exact match of 99.8% for each analyte.
- Repeatability:
- Interference:
- Study Type: Interference Testing
- Key Results: Interfering substances and their lowest concentrations causing interference were identified and listed in the accompanying User Manual.
- Limit of Detection:
- Study Type: Limit of Detection
- Key Results: The cut-off for each block is defined as the lowest concentration at which >55% of the results were positive.
- Linearity:
- Study Type: Linearity
- Key Results: Both analytes demonstrated a linear relationship between variables, with no biases disrupting the overall pattern of linearity.
- Timing Flex Study:
- Study Type: Timing Flex (carry-over, dipping, wetting (blotting), assay-time)
- Key Results:
- Carry-Over: None of the carry-over time durations impacted the ACR | LAB's ability to measure accurate results.
- Wetting (Blotting) Time: For creatinine, no impact on accuracy. For albumin, waiting 60-seconds to blot resulted in False Positive results.
- Dipping Time: For creatinine, no impact on accuracy. For albumin, at 5- and 10-seconds there was an increase in False Positive results.
- Assay Time: None of the assay time durations had any impact on the ACR | LAB's ability to accurately measure albumin or creatinine.
- Stability:
- Study Type: Stability (real time and accelerated aging, transportation conditions)
- Key Results: The ACR | LAB device was not affected by any of these parameters and its stability was therefore confirmed.
Clinical Performance Testing:
- Method Comparison:
- Study Type: Method Comparison to predicate device
- Sample Size: 375 native urine samples and 60 contrived samples
- Key Results: Demonstrated high levels of accuracy and substantial equivalence with the predicate device.
Software Validation:
- Study Type: Software Validation
- Key Results: Software meets the design requirements and all potential risks have been mitigated. Cyber security analysis report and validation test report summarize assessment process for cyber security and demonstrates the software meets the required level of cyber security protection.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
- Method Comparison - Overall agreement (% exact match):
- Albumin: 89%
- Creatinine: 84%
- Albumin-creatinine ratio: 93%
- Method Comparison - ± 1 color block % agreement:
- Albumin: 100%
- Creatinine: 100%
- Albumin-creatinine ratio: 100%
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1225 Creatinine test system.
(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
July 26, 2019
Healthy.io Ltd Ron Zohar Chief Product Officer 2 Ibn Gabirol Street Tel Aviv, 6407702 Israel
Re: K182384
Trade/Device Name: ACR | LAB Urine Analysis Test System Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: JFY, JIR, KQO Dated: June 25, 2019 Received: June 25, 2019
Dear Ron Zohar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K182384
Device Name ACR | LAB Urine Analysis Test System
Indications for Use (Describe)
The ACR | LAB Urine Analysis Test System is comprised of a smartphone application, a proprietary Color-Board, and ACR Reagent Strips. It is intended for the semi-quantitative detection of albumin and creatinine in urine, as well as the presentation of their ratio. The ACR | LAB Urine Analysis Test System is intended for in-vitro diagnostic use by a healthcare professional in a point of care setting. These results may be used in conjunction with clinical evaluation as an aid in the diagnosis for kidney function.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image contains the logo for Healthy.io. The logo consists of a colorful square made up of a red rectangle on the left, a white rectangle on the top right, and a yellow rectangle on the bottom right. To the right of the square is the text "Healthy.io" in a simple, sans-serif font.
Attachment 3
510(K) SUMMARY
Healthy.io's ACR | LAB Urine Analysis Test System K182384
Submitter
Healthy.io Ltd. 2 Ibn Gabirol St. Tel Aviv, 6407702 Israel Phone: +972-54-445-4514 Fax: +972-77-470-4808 Contact Person: Ron Zohar
Date Prepared: July 2019
Name of Device: ACR | LAB Urine Analysis Test System
Common or Usual Name: Smartphone enabled albumin-creatinine ratio analyzer
Regulation Section and Classification Name:
Class I: Albumin Class II: Creatinine 21 CFR § 862.1645 Urinary protein or albumin (non-quantitative) test system 21 CFR § 862.1225 Creatinine test system 21 CFR § 862.2900 Automated urinalysis system
Product Code:
JIR JFY KQO
Classification Panel: Clinical Chemistry
Predicate Device:
The ACR | LAB is substantially equivalent to the following predicate device:
Manufacturer | Device | 510(k) Number |
---|---|---|
ACON Laboratories Inc. | Mission U120 Ultra Urine Analyzer | |
and Mission Urinalysis Reagent Strips | ||
(Microalbumin/Creatinine) | K142391 |
4
Image /page/4/Picture/0 description: The image shows the logo for Healthy.io. The logo consists of a square divided into four smaller squares on the left side, with the colors red and yellow. To the right of the square is the text "Healthy.io" in a sans-serif font. The text is black and the background is white.
Device Description:
The ACR | LAB Urine Analysis Test System is comprised of a smartphone application, a proprietary Color-Board and ACR Reagent Strips. It is intended for the semi-quantitative detection of albumin and creatinine in urine, as well as the presentation of their ratio.
The device is provided as a kit that is comprised of a canister of 100 FDA-cleared urine test strips (ACON Laboratories Inc. Mission Urinalysis Reagent Strips (Microalbumin/Creatinine) K150330), 10 Color-Boards, and a User Manual. The ACR | LAB Urine Analysis Test System also consists of a smartphone application for use on iPhone 7 device (iOS 12), and an image recognition algorithm running on the Backend*.
The software component of the ACR | LAB consists of both an application (App) and a Backend server (Backend). The App instructs the professional user how to accurately perform the test. The App conducts a series of boundary condition analyses, and if the scan is approved, sends the information to the Backend for complete analysis and results classification. Once analyzed, the results are securely transmitted to a patient Electronic Medical Record for review by a healthcare professional. The patients do not have access to the results at any point during the testing process.
Device Components:
The ACR | LAB Urine Analysis Test System is comprised of the following components:
-
- A canister of 100 Mission Urinalysis Reagent Strips U031-021 (ACON Laboratories Inc., K150330)
-
- Ten individually wrapped Color-Boards
-
- A smartphone App
-
- A backend, cloud-based server
-
- A user manual
-
- iPhone 7 with iOS 12 (not provided)
Intended Use/Indication for Use:
The ACR | LAB Urine Analysis Test System is comprised of a smartphone application, a proprietary Color-Board, and ACR Reagent Strips. It is intended for the semi-quantitative detection of albumin and creatinine in urine, as well as the presentation of their ratio. The ACR | LAB Urine Analysis Test System is intended for in-vitro diagnostic use by a healthcare professional in a point of care setting. These results may be used in conjunction with clinical evaluation as an aid in the diagnosis for kidney function.
Comparison of ACR | LAB and ACON Laboratories' Mission U120 Ultra Urine Analyzer:
Table 1 below summarizes similarities and variabilities between the ACR | LAB and its predicate device, ACON Laboratories' Mission U120 Ultra Urine Analyzer and Mission Urinalysis Reagent Strips (Microalbumin/Creatinine) (K142391).
1
Backend will henceforth refer to the system's cloud-based servers, which run the server-side algorithms, and turn the algorithms' analysis into clinical results. These are then securely transmitted to a patient Electronic Medical Record. The ACR | LAB Backend is HIPAA compliant and uses advanced security protocols for data handling and transmission.
5
Image /page/5/Picture/0 description: The image shows the logo for Healthy.io. The logo consists of a square divided into four sections on the left side. The top left section is red, the top right section is white, and the bottom section is yellow. To the right of the square is the text "Healthy.io" in a simple, sans-serif font.
Table 1: Comparison between ACR LAB Urine Analysis Test System and the ACON Laboratories' Mission |
---|
U120 Ultra Urine Analyzer (K142391) |
Feature | Healthy.io's ACR LAB Urine Analysis Test System (K182384) | ACON Laboratories' Mission U120 Ultra Urine Analyzer and strips (K142391) |
---|---|---|
Intended Use | The ACR LAB Urine Analysis Test | |
System is comprised of a | ||
smartphone application, a | ||
proprietary Color-Board, and ACR | ||
Reagent Strips. It is intended for the | ||
semi-quantitative detection of | ||
albumin and creatinine in urine, as | ||
well as the presentation of their | ||
ratio. The ACR LAB Urine Analysis | ||
Test System is intended for in-vitro | ||
diagnostic use by a healthcare | ||
professional in a point of care | ||
setting. These results may be used | ||
in conjunction with clinical | ||
evaluation as an aid in the diagnosis | ||
for kidney function. | The Mission® U120 Ultra Urine | |
Analyzer is a uranalysis instrument | ||
intended for in vitro diagnostic use. It | ||
is intended for professional use only at | ||
point-of-care locations. The Mission | ||
U120 Ultra Urine Analyzer is intended | ||
to read Mission® Urinalysis Reagent | ||
Strips (Microalbumin/Creatinine) for | ||
the semi-quantitative measurement | ||
of albumin and creatinine. These | ||
measurements are used to assist | ||
diagnosis for kidney function. | ||
Test Specimen | Urine | Urine |
Detection | ||
Methodology | Reflectance Photometry | Reflectance Photometry |
Detection device | Photosensitive Diode | Photosensitive Diode |
Operating Conditions | 0-35°C (32-95°F); ≤93% Humidity | |
(non-condensing) | 0-40°C (32-104ºF); ≤85% Humidity | |
(non-condensing) | ||
Strips to be used | Mission Urinalysis Reagent Strips | |
U031-021 (K150330) | Mission Urinalysis Reagent Strips | |
U031-021 | ||
Strip Operating | ||
Conditions | 15-30°C [59°-86°F] | |
20%-80% Relative humidity (non- | ||
condensing) | 15-30°C [59°-86°F] | |
20%-80% Relative humidity (non- | ||
condensing) | ||
Strip Incubation Time | 1 minute | 1 minute |
Calibration | Automatic | Automatic |
Power Source | Not Applicable | 6 AA batteries 100- 240 VAC |
(adapter), (46-65 Hz± 1HZ) | ||
Line Leakage Current | Not Applicable | 55% of the results were positive. |
Linearity. The linearity study of the ACR | LAB Urine Analysis Test System was designed in accordance with guidance provided by the Clinical and Laboratory Standards Institute (CLSI) document EP6-A – Evaluation of Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. |
- The Linearity study tested the ability of the ACR | LAB Urine Analysis Test System to provide results that are directly proportional to the concentrations of analytes within each test sample. The study tested the entire range of values that the ACR | LAB Urine Analysis Test System is designed to measure. These expected values were presented against actual device results. Both of the analytes demonstrated a linear relationship between variables, with no biases disrupting the overall pattern of linearity.
8
Image /page/8/Picture/0 description: The image shows the logo for Healthy.io. The logo consists of a square divided into four sections on the left, with a red vertical rectangle on the left, a white square in the upper right, and an orange rectangle in the bottom right. To the right of the square is the text "Healthy.io" in a simple, sans-serif font.
Timing Flex Study. This experiment essentially consists of four different studies: 1. carry-over; 2. dipping; 3. wetting (blotting); and 4. assay-time; and was designed to evaluate how the timing of dipping, wetting, and blotting, as well as carry-over of the Strips impacts the ability of the ACR | LAB Urine Analysis Test System to accurately measure the levels of albumin and creatinine in a urine sample. The framework of this study is based primarily on similar studies of devices with similar indications for use.
- . Carry-Over. In this study, strips were held in both the "up" (with the creatinine on top) and "down" (with the albumin on top) positions for different time intervals after being dipped in spiked solutions. None of the carry-over time durations, in both "up" and "down" configurations had any impact on the ACR | LAB's ability to measure accurate results. The study used the most extreme cases, testing for high-positive "run-off" into low-positive patches.
- Wetting (Blotting) Time: This experiment tested six different "blotting" times, i.e. the tester dipped the Strip and then waited a specific amount of time (0-, 5-, 10-, 20-, 40- and 60-seconds) before blotting the Strip and scanning it with the ACR | LAB device. For creatinine, none of the various dipping times impacted the accuracy of the device. For albumin, waiting 60-seconds to blot the strip resulted in False Positive results. The User Manual and in-app instructions direct the professional user to dip, blot, and place the strip on the Color-Board in quick succession.
- Dipping Time. Strips were dipped in spiked solutions for four different time durations (1-, 2-, 5-, and 10-seconds) to test the impact of various dipping times on the ACR | LAB's ability to perform accurate measurements. For creatinine, none of the various dipping times impacted the accuracy of the device. For albumin, at times 5- and 10-seconds there was an increase in False Positive results. The User Manual and in-app instructions direct the professional user to dip the ACR Reagent Strip for 1-second.
- Assay Time: This experiment tested how different assay times impacted the ACR | LAB's ability to measure accurate results. Strips were dipped and then immediately blotted, per the device's instructions. The user then waited a predetermined amount of time (60-, 80-, 100-, or 120- seconds) before scanning the strip. None of the assay time durations had any impact on the ACR | LAB's ability to accurately measure albumin or creatinine. The User Manual and the in-app instructions direct the professional user to scan the Strip and Color-Board once the in-app timer has indicated that 60-seconds have elapsed.
Stability. This stability experiment was designed to test the ACR | LAB ability to provide accurate measurements after subjecting the device kit to both real time and accelerated aging timeframes and conditions that could arise during transportation (e.g. temperature, humidity and vibration). In addition, an open vial experiment was also conducted. After exposure to these transportation conditions and timeframes, the performance of the kits was evaluated through a number of functional tests. The ACR | LAB device was not affected by any of these parameters and its stability was therefore confirmed.
Clinical Performance Testing:
Method Comparison. The objective of the Method Comparison study was to test the ACR | LAB's degree of agreement with its predicate device, when operated by a professional user at a point-of-care location
The primary acceptance criteria for the study were the percent of exact match and ±1 color block match between the ACR | LAB Urine Analysis Test System and the predicate. Throughout the testing, the ACR |
9
LAB Urine Analysis Test System demonstrated high levels of accuracy as demonstrated by its percent of agreement with the predicate device.
Study Design
The study evaluated native urine samples from 375 subjects as well as 60 contrived samples at three U.S. clinical sites. The ACR | LAB device was provided to the study coordinators. All subjects were asked to provide a urine sample. Two separate lab technicians were responsible for measuring each urine sample. The first lab technician measured the urine sample using the iPhone 7 device; the second technician measured the urine sample using the predicate device, the ACON Laboratories' Mission U120 Ultra Urine Analyzer (U120 Ultra). Each urine sample was tested twice in total; once with the iPhone 7 device and then subsequently with the predicate device (U120 Ultra). The results of the ACR | LAB device and results measured by the predicate (U120 Ultra) were compared.
Study Results – Accuracy
The overall agreement (% exact match) between the ACR | LAB Urine Analysis Test System and the predicate device was 89% for albumin and 84% for creatinine. The overall agreement between the ACR | LAB Urine Analysis Test System and the predicate device when measuring the albumin-creatinine ratio was 93%. The ± 1 color block % agreement between the ACR | LAB Urine Analysis Test System and the predicate device was 100% for albumin, 100% for creatinine, and 100% for the albumin-creatinine ratio.
The high rates with which the ACR | LAB measured the same results as the predicate ACON Laboratories' Mission U120 Ultra Urine Analyzer demonstrate substantial equivalence between the two devices.
Software Validation:
Validation of the ACR | LAB device software was performed according to clause 14 of IEC 60601-1 (third edition) Software requirements, IEC 62304:2004 to IEC 62304 - 2006/AC 2008, Medical device, and Software life-cycle processes standards. The software related documents were composed according to the specific IEEE standards and the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The Software Validation documents summarize the validation assessment and demonstrate that the software meets the design requirements and all potential risks have been mitigated.
In accordance with the FDA's guidance for management of cyber-security in medical devices, the OCR's HIPAA regulation and ISO 27001:2013, Healthy.io conducted a comprehensive hazard analysis. The cyber security analysis report, together with the cyber security validation test report summarizes the assessment process for cyber security and demonstrates that the software meets the required level of cyber security protection.
Conclusions:
The clinical and analytical performance study results demonstrate that the ACR | LAB Urine Analysis Test System is substantially equivalent to the predicate device Mission® U120 Ultra Urine Analyzer and Mission Urinalysis Reagent Strips (Microalbumin/Creatinine) (FDA 510(k): K142391).