K Number
K182367
Device Name
DIASAFEplusUS
Date Cleared
2019-01-10

(132 days)

Product Code
Regulation Number
876.5665
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate from pre-treated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g., RO system, central delivery system).

The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE® plusus.

Device Description

The DIASAFE®plusus (P/N F00007039) is a non-sterile dialysis fluid filter that produces ultrapure dialysate as defined in ANSI/AAMI/ISO 11663. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment. The DIASAFE® plusus is installed and exchanged on 2008 Series hemodialysis machines using the DIAFIX™ Lock System. The DIAFIX™ Lock System is a standard feature on 2008 Series hemodialysis machines and is installed during machine production.

Mixed dialysate is forced through an open filter port across the fibers of DIASAFE® plusus. A bypass valve at the other end is closed which forces the dialysate across the fiber membrane. Dialysate passes through the porous membrane and into the filtrate compartment where it is allowed to flow through the uncapped side ports and into the hemodialyzer. Microorganisms and molecules too large to pass through the membrane are trapped in the fibers until they are flushed out, and/or disinfected during a routine cleaning cycle

AI/ML Overview

The provided text describes the regulatory clearance (510(k)) for the DIASAFE®plusUS, a water purification system for hemodialysis. The document outlines acceptance criteria and the results of various performance tests conducted to demonstrate the device's substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

Test ConductedAcceptance CriteriaReported Device Performance
Bacteria RetentionThe filter shall retain bacteria to produce ultrapure dialysate with a microbial count <0.1 CFU/mL (ANSI/AAMI/ISO 11663)Pass, results within acceptance criteria
Endotoxin RetentionThe filter shall retain endotoxin to produce ultrapure dialysate with an endotoxin count <0.03 EU/mL (ANSI/AAMI/ISO 11663)Pass, results within acceptance criteria
Ultrafiltration RateThe aqueous ultrafiltration rate shall be >300 mL/hr mmHg (3.75 L/min bar) at 37°CPass, results within acceptance criteria
Dialysis Fluid CompositionDialysate composition shall not be adversely affected after passing through the DIASAFE®plusUSPass, results within acceptance criteria
Filter IntegrityThe filter connectors, caps, housing, and welding shall not leak at a filtration pressure <2 barPass, results within acceptance criteria
Membrane IntegrityThe membrane shall be intact at a transmembrane pressure <2 bar and at a temperature of 20 ± 5°CPass, results within acceptance criteria
Chemical ContaminantsThe DIASAFE®plusUS should not contribute unacceptable levels of chemical contaminantsPass, results within acceptance criteria for each defined element/ion

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (number of devices tested) for each performance test. It mentions "DIASAFE®plusUS filter samples" for the Ultrafiltration Rate test and generally indicates that tests were "conducted" or "evaluated."

The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, based on the context of a 510(k) submission for a medical device, these would be lab-based, prospective studies performed by the manufacturer to demonstrate product performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the type of device and testing described. The DIASAFE®plusUS is a physical filter for water purification. Its performance is measured through objective, quantifiable laboratory tests (e.g., microbial counts, endotoxin levels, flow rates) rather than human interpretation or expert consensus on medical images or patient conditions. Therefore, there's no "ground truth" established by experts in the context of medical image review.

4. Adjudication method for the test set

This information is not applicable. Since the tests are objective, quantifiable measurements of physical and chemical properties (e.g., bacteria count, endotoxin levels), there is no need for an adjudication method among human readers or experts to determine a ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This is not an AI/software-based device that would assist human readers in, for example, diagnosing medical conditions. It is a physical filtration device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical filter, not an algorithm. Its "standalone" performance is literally its ability to perform the filtration function, which is what the performance tests evaluate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the performance tests relies on quantitative laboratory measurements against established standards and specifications. For example:

  • Bacteria and Endotoxin Retention: The ground truth is the measured microbial count and endotoxin level in the filtered dialysate falling below the specified threshold (<0.1 CFU/mL and <0.03 EU/mL respectively, as per ANSI/AAMI/ISO 11663).
  • Ultrafiltration Rate: The ground truth is the measured flow rate exceeding the specified minimum (>300 mL/hr mmHg).
  • Integrity/Composition: The ground truth is the observation that the device components remain intact and the dialysate composition is not adversely affected, matching the engineering specifications.

These are not based on expert consensus, pathology, or outcomes data in the medical diagnostic sense, but rather on direct physical and chemical measurements.

8. The sample size for the training set

This information is not applicable. The DIASAFE®plusUS is a physical device, not an AI or machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

January 10, 2019

Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451

Re: K182367 Trade/Device Name: DIASAFE®plusUS Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water Purification System for Hemodialysis Regulatory Class: II Product Code: FIP Dated: December 4, 2018 Received: December 6, 2018

Dear Denise Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182367

Device Name DIASAFE@plusUS

Indications for Use (Describe)

The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate from pre-treated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g., RO system, central delivery system).

The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

The DIASAFE@plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font and is also blue.

510(K) SUMMARY 5.

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.

Submitter's Information 5.1.

Name:Fresenius Medical Care Renal Therapies Group, LLC
Address:920 Winter StreetWaltham, MA02451-1457
Phone:(781) 699-4479
Fax:(781) 699-9635
Contact Person:Denise OppermannSenior Director, Regulatory Affairs – Devices
Preparation Date:31 August 2018

5.2. Device Name

Trade Name:DIASAFE® plusUS
Common Name:Subsystem, water purification
Regulation Name:Water purification system for hemodialysis
Regulatory Class:Class II per 21 CFR §876.5665
Product Code:FIP
Classification Panel:Gastroenterology/Urology

Legally Marketed Predicate Device 5.3.

The legally marketed predicate device is the Dialysis Machine with Dialysate Ultrafilter (K944767), hereinafter referred to as "Diasafe Dialysate Ultrafilter". This predicate has not been subject to a design-related recall.

Device Description 5.4.

5.4.1. Device Identification

The DIASAFE® plusus (P/N F00007039) is the subject of this Traditional 510(k).

5.4.2. Device Characteristics

The DIASAFE®plusus is a non-sterile dialysis fluid filter that produces ultrapure dialysate as defined in ANSI/AAMI/ISO 11663. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment. The DIASAFE® plusus is installed and exchanged on 2008 Series hemodialysis machines using the DIAFIX™ Lock System. The DIAFIX™ Lock System is a standard feature on 2008 Series hemodialysis machines and is installed during machine production.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue abstract symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevron shapes stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

5.4.3. Environment of Use

The DIASAFE® plusus is used in environments where acute and chronic hemodialysis is performed.

5.4.4. Brief Description of the Device

Mixed dialysate is forced through an open filter port across the fibers of DIASAFE® plusus. A bypass valve at the other end is closed which forces the dialysate across the fiber membrane. Dialysate passes through the porous membrane and into the filtrate compartment where it is allowed to flow through the uncapped side ports and into the hemodialyzer. Microorganisms and molecules too large to pass through the membrane are trapped in the fibers until they are flushed out, and/or disinfected during a routine cleaning cycle

5.4.5. Materials of Use

The DIASAFE® plusus is classified as an externally communicating, blood path indirect, prolonged contact (>24 hours to 30 days) duration, Class II (Category B) device in accordance with FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (16 June 2016).

A list of the DIASAFE® plusus components and materials is provided in Table 1.

ComponentMaterial
Capillary fibers (fiber bundle)Fresenius Polysulfone®,Polyvinylpyrrolidone
Filter housingPolypropylene
PU-Resin (potting mass)Polyurethane
Welded header (caps)Polypropylene
O-ringSilicone
Plastic tabs (tear-off brackets)Silicone
Sealing disc-ringSilicone

Table 1: DIASAFE®plusus Components and Materials

Key Performance Specifications/Characteristics 5.4.6.

A description of the key DIASAFE® plusus performance specifications is provided in Table 2.

DIASAFE®plusus Features Table 2:

FeaturesSpecifications
Filtration Rate≥ 300 mL/hr·mmHg≥ 3.75 L/min·bar
Maximum Filtration Pressure2 bar
Connection SystemDIAFIX™ Lock System

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue abstract symbol on the left, resembling three downward-pointing chevrons stacked on top of each other. To the right of the symbol, the text "FRESENIUS MEDICAL CARE" is displayed in a bold, sans-serif font, also in blue. The words "MEDICAL CARE" are stacked below "FRESENIUS".

FeaturesSpecifications
Surface Area2.2m²
Dialysis Fluid Purity (ultrapure dialysate)<0.03 EU/mL<0.1 CFU/mL

Table 2: DIASAFE®plusus Features (Continued)

ર્ડ.

જિંત Intended Use/Indications for Use

The DIASAFE®plusus is intended for the preparation of ultrapure dialysate from pre-treated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g., RO system, central delivery system).

The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

The DIASAFE®plusus can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE® plusus.

5.6. Comparison of Technological Characteristics with the Predicate Device

A comparison of the technological characteristics of the DIASAFE® plusus with the Diasafe Dialysate Ultrafilter (K944767) is provided in Table 3.

FeatureDiasafe Dialysate Ultrafilter -Predicate Device (K944767)DIASAFE®plusus - Proposed Device
Principle ofOperationThe Diasafe Dialysate Ultrafilterreduces the level of bacteria andpyrogens in the dialysate for usein acute and chronic hemodialysisthrough membrane filtration andadsorptionThe DIASAFE®plusus providesultrapure dialysate for use in acuteand chronic hemodialysis by reducingthe level of bacteria and pyrogensthrough membrane filtration andadsorption
Environment ofUseThe Diasafe Dialysate Ultrafilteris used in environments whereacute and chronic hemodialysis isperformedThe DIASAFE®plusus is used inenvironments where acute and chronichemodialysis is performed
MembraneSurface Area1.8 m²2.2 m²
Priming Volume(inside ofcapillaries)110 mL130 mL
Table 3:DIASAFE® plus US Comparison of Technological Characteristics
-------------------------------------------------------------------------------

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue chevron-like symbol on the left, followed by the text "FRESENIUS" on the first line and "MEDICAL CARE" on the second line, both in blue. The text is bold and sans-serif.

FeatureDiasafe Dialysate Ultrafilter -Predicate Device (K944767)DIASAFE®plusUS - Proposed Device
Priming Volume(outside ofcapillaries)310 mL276 mL
HemodialysisMachineConnectionSystemDialyzer connectionsDIAFIX™ Lock System
DevicePackagingPlastic overwrap with Tyvek®sterilized pouchThe filter is sealed with plastic tabs(tear-off brackets) and is packaged ina transparent Sealable PA/PEoverwrap
Bleach Exposure≤15 hours≤ 30 hours (≤ 30 chemicaldisinfection cycles)
Shelf-life3 years3 years
SterilityEthylene Oxide sterilizedNon-sterile

Table 3: DIASAFE®plusus Comparison of Technological Characteristics (Continued)

5.7. Performance Data

Testing conducted to support the determination of substantial equivalence is summarized in Table 4.

Test ConductedTest Method DescriptionAcceptance CriteriaResults / Conclusion
BacteriaRetentionDialysis fluid spiked withbacteria that exceeded theallowable limit perANSI/AAMI RD52 wasfiltered through theDIASAFE®plususThe filter shall retainbacteria to produceultrapure dialysatewith a microbialcount <0.1 CFU/mL(ANSI/AAMI/ISO11663)Pass, results withinacceptance criteria
EndotoxinRetentionDialysis fluid spiked withendotoxin that exceededthe allowable limit perANSI/AAMI RD52 wasfiltered through theDIASAFE®plususThe filter shall retainendotoxin to produceultrapure dialysatewith an endotoxincount <0.03 EU/mLPass, results withinacceptance criteria
Test ConductedTest Method DescriptionAcceptance CriteriaResults / Conclusion
(ANSI/AAMI/ISO11663)
UltrafiltrationRateDIASAFE® plus US filtersamples were evaluated foraqueous KUFThe aqueousultrafiltration rateshall be >300 mL/hrmmHg (3.75 L/minbar) at 37°CPass, results withinacceptance criteria
Dialysis FluidCompositionDialysate was passedthrough the filter atvarying flow ratessimulating typical in-useconditions. Dialysatecomposition was measuredbefore and after filtrationDialysatecomposition shall notbe adversely affectedafter passing throughthe DIASAFE® plus USPass, results withinacceptance criteria
Filter IntegrityFilter integrity wasevaluated after simulatedship testingThe filter connectors,caps, housing, andwelding shall not leakat a filtration pressure<2 bar and at aPass, results withinacceptance criteria
MembraneIntegrityMembrane integrity wasevaluated after simulatedship testingThe membrane shallbe intact at atransmembranepressure <2 bar andat a temperature of 20± 5°CPass, results withinacceptance criteria
ChemicalContaminantsThe DIASAFE® plus USwas evaluated forelemental and ionicchemical contaminantslisted in ANSI/AAMI/ISO13959:2014TheDIASAFE® plus USshould not contributeunacceptable levelsof chemicalcontaminantsPass, results withinacceptance criteriafor each definedelement/ion

Table 4: Performance Data Testing

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons.

Table 4: Performance Data Testing (Continued)

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue abstract symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The abstract symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

Biocompatibility Testing 5.7.1.

The following testing was performed to support the biological safety of the DIASAFE®plusus.

  • Chemical characterization
  • Cytotoxicity, ISO Elution Method with MEM
  • Sensitization, Guinea Pig Maximization
  • Intracutaneous Irritation
  • Material Mediated Pyrogenicity
  • Hemocompatibility, ASTM Hemolysis (Indirect) Contact ●
  • Subchronic Intravenous Toxicity
  • Risk assessment of potential toxicity ●
  • o PVP Testing

5.7.2. Stability Design Verification

The DIASAFE® plusus is provided non-sterile and will be labeled with a 36 months (3-year) product shelf life. The shelf life is supported by accelerated aging equivalent to 3-year realtime and 24 months real-time.

5.7.3. Shipping Verification

Shipping verification was conducted to support that the DIASAFE® plusys can withstand the distribution environments it is subjected to during actual use. Initial shipping verification was performed to evaluate pallet shipping configurations according to ASTM D4169-14 using an assurance level II with distribution cycle 6. A subsequent gap assessment between the ASTM D4169-14 and ASTM D4169-16 confirmed that the previous requirements of ASTM D4169-14 satisfied the requirements of ASTM D4169-16.

Additional shipping verification was then performed to evaluate single packaging shipping configurations according to ASTM D4169-16 using an assurance level II with distribution cvcles 3 and 13.

Membrane and filter integrity testing were performed after ship testing to evaluate gross mechanical damage to the fiber/filter.

5.7.4. Human Factors Testing

Formative and validation studies were conducted to address both labeling comprehension and user interaction with the DIASAFE® plusus. Execution of the human factors process is consistent with the requirements of ANSI/AAMI/IEC 62366-1:2015 and FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (03 February 2016).

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue emblem on the left and the company name on the right. The emblem is made up of three downward-pointing chevrons stacked on top of each other. The text "FRESENIUS MEDICAL CARE" is in bold, blue letters.

Software Verification and Validation Testing 5.7.5.

This section is not applicable. The DIASAFE® plusus does not contain software.

5.7.6. Animal Studies

No animal studies were performed.

5.7.7. Clinical Studies

No clinical studies were performed.

5.8. Conclusion

The intended use, principle of operation, environment of use, and design of the DIASAFE®plusus are substantially equivalent to those of the predicate device. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the DIASAFE®plusus is safe and effective for its intended use

§ 876.5665 Water purification system for hemodialysis.

(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.