K Number
K122809
Manufacturer
Date Cleared
2013-06-17

(277 days)

Product Code
Regulation Number
862.3100
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Advin Multi-Drug Screen Test is a one-step immunoassay for the qualitative detection of multiple drugs of abuse and/or their metabolites in human urine at the following cutoff levels:
Amphetamine (AMP) Calibrator: d-Amphetamine, Cutoff Level: 500 ng/mL
Barbiturates (BAR) Calibrator: Secobarbital, Cutoff Level: 300 ng/mL
Benzodiazepines (BZO) Calibrator: Oxazepam, Cutoff Level: 300 ng/mL
Buprenorphine (BUP) Calibrator: Buprenorphine, Cutoff Level: 10 ng/mL
Cocaine (COC) Calibrator: Benzoylecgnine, Cutoff Level: 150 ng/mL
EDDP Calibrator: 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Cutoff Level: 300 ng/mL
Ecstasy (MDMA) Calibrator: Methylenedioxymethamphetamine, Cutoff Level: 500 ng/mL
Methamphetamine (MET) Calibrator: d-Methamphetamine, Cutoff Level: 500 ng/mL
Methadone (MTD) Calibrator: Methadone, Cutoff Level: 300 ng/mL
Morphine (MOR300) Calibrator: Morphine, Cutoff Level: 300 ng/mL
Opiates (OPI) Calibrator: Morphine, Cutoff Level: 2,000 ng/mL
Oxycodone (OXY) Calibrator: Oxycodone, Cutoff Level: 100 ng/mL
Phencyclidine (PCP) Calibrator: Phencyclidine, Cutoff Level: 25 ng/mL
Propoxyphene (PPX) Calibrator: Propoxyphene, Cutoff Level: 300 ng/mL
Tricyclic-antidepressant (TCA) Calibrator: Nortriptynine, Cutoff Level: 1,000 ng/mL
Marijuana (THC) Calibrator: 11-nor-Δ9-THC-9-COOH, Cutoff Level: 50 ng/mL

The Advin Multi-Drug Screen Test consists of three formats: a Cassette, a Dip Card and a Cup, which may be configured in any combination of the drug analytes listed in the above table.

Advin Multi-Drug Screen Test is used to obtain a visual, qualitative, preliminary test result for prescription use in point of care sites, laboratory settings and is also intended for workplace and over-the-counter use. The Propoxyphene test strip is not intended for over-the-counter use.

The Advin Multi-Drug Screen Test will yield preliminary positive results when prescription drugs TCA, Barbiturates and Benzodiazepine, Methadone, Buprenorphine and Opiates are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for TCA, Barbiturates and Benzodiazepine in urine.

The Advin Multi-Drug Screen Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. Gas chromatography/mass spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry/Mass (LC/MS/MS) and High Performance Liquid Chromatography-(HPLC)-are the-preferred-confirmatory-methods-Glinical-consideration and professional judgment should be applied to any drug of abuse test result, particularly in the evaluation of a preliminary positive test result.

Device Description

The Advin Multi-Drug Screen Test is a lateral flow immunochromatographic competitive binding assay and is intended for prescription use, CLIA Waived use and OTC use. It is used as in vitro diagnostics to visually and qualitatively detect some of the most common drugs of abuse and/or their metabolites in human urine specimens. The test is based on the principle of competitive binding antigen-antibody immunochemistry. It utilizes anti-drug antibody to selectively detect elevated levels of multi-drugs and/or their metabolites in urine at a specified cutoff for each drug. These visual read tests are performed without the use of an instrument.

A drug-Positive urine specimen will not generate a colored-line in the designated test region, while a drug-Negative urine specimen or a urine specimen containing drug(s) at the concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of urine specimen has been added and membrane wicking has occurred.

Some of the Advin Multi-Drug Screen Test will be sold with either one or two adulteration test strips for the detection of adulterants added to the urine specimen. Each Adulteration strip consists of one to three reagent pads which consist of tests for oxidant, specific gravity, pH, glutaraldehyde, creatinine and nitrite.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Advin Multi-Drug Screen Test, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the "Accuracy" section, which compares the device's results to GC/MS analysis data. For each drug, the device shows an agreement percentage with GC/MS for negative samples and 100% agreement for positive samples above certain thresholds. The specific thresholds for agreement percentage for negative results are:

Drug TestReported Performance (Negative % Agreement)Reported Performance (Positive % Agreement)
AMP97.7%100%
BAR95.2%100%
BUP95.5%100%
BZO93.2%100%
COC97.7%100%
EDDP93.2%100%
MDMA95.5%100%
MET93.2%100%
MOR93.2%100%
MTD95.5%100%
OPI93.2%100%
OXY93.2%100%
PCP97.7%100%
PPX95.3%100%
TCA95.5%100%
THC97.7%100%

Additionally, for the lay-user study, the performance criteria were defined as:

  • Agreement between lay persons' results and GC/MS analysis results > 99% for concentrations at +/- 50% cutoff levels.
  • Agreement between lay persons' results and GC/MS analysis results 65% to 86% for concentrations near the cutoff levels.

The tables for each device format (Cassette, Dipcard, Cup) show that these agreement percentages were met for the tested samples.

2. Sample Size and Data Provenance

Test Set Sample Size:

  • Negative Specimens: 40 drug-free urine specimens were collected and tested for each drug.
  • Positive Specimens: At least 40 drug-positive urine specimens were obtained for each drug test.
  • Lay-Users Study: Over one hundred male and female persons age 18 or older participated. Each level of drug solution (at +/- 50% cutoff and +/- 25% cutoff) was randomly tested with at least 20 replicates by lay persons.

Data Provenance:

  • Drug-free urine specimens were collected from presumed non-user volunteers (no specific country mentioned but implied to be the US, given the submission is to the FDA and reference labs are "throughout the United States").
  • Drug-positive urine specimens were obtained from reference labs throughout the United States.
  • The study is observational/analytical in nature, evaluating device performance against a gold standard method. It can be considered a retrospective analysis of collected urine samples.

3. Number of Experts and Qualifications for Ground Truth

  • The text does not explicitly state the number of experts used.
  • The ground truth for the test set was established using Gas chromatography/mass spectrometry (GC/MS) and Liquid Chromatography / Mass Spectrometry/Mass (LC/MS/MS for TCA analysis). These are analytical chemistry methods conducted in reference laboratories, implying qualified analytical chemists/technicians were responsible for these measurements. No specific qualifications like "radiologist with 10 years of experience" are mentioned as the "interpreters" of the ground truth for this type of test.

4. Adjudication Method for the Test Set

No explicit adjudication method is mentioned. The comparison was directly between the Advin Multi-Drug Screen Test results and the ground truth established by GC/MS/LC/MS analysis. The "Analysis of Discordant Results" section indicates cases where the device result did not match the GC/MS result, suggesting these were noted rather than adjudicated by another method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There was no MRMC comparative effectiveness study involving human readers with and without AI assistance. This device is a rapid immunoassay that provides a visual, qualitative result, not an AI-powered diagnostic system. However, a "Lay-Users Study" was conducted, demonstrating that lay persons could accurately use the device, which is a form of user performance evaluation.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance study was done. The "Accuracy" section directly evaluates the performance of the Advin Multi-Drug Screen Test devices (Cassette, Dipcard, Cup) against GC/MS analysis data. This constitutes the algorithm's (immunoassay's) standalone performance without human interpretation being the primary variable. The results are reported as percentage agreements for negative and positive samples.

7. Type of Ground Truth Used

The primary ground truth used was analytical chemistry methods:

  • Gas chromatography/mass spectrometry (GC/MS)
  • Liquid Chromatography / Mass Spectrometry/Mass (LC/MS/MS) for TCA analysis

8. Sample Size for the Training Set

The document does not mention a training set. This device is an immunoassay, not a machine learning or AI-based diagnostic that would typically require a training set. The studies described are for validation and verification of the device's performance.

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned or applicable for this type of device, this information is not provided in the document.

{0}------------------------------------------------

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The Assigned 510(k) number is: K122809

Submitter:

Advin Biotech 6861 Nancy Ridge Dr., Suite #D, San Diego, CA 92121.

Tel.: 858-866-8382 Fax: 858-866-8382

Date:

April 21, 2013

Contact Person:

Edward Tung, Ph.D.

Tel. 858-668-8305 (Cell)

Product Names:

Advin Multi-Drug Screen Test Cassette Advin Multi-Drug Screen Test Dipcard Advin Multi-Drug Screen Test Cup

Common Name:

Immunochromatographic test for the qualitative detection of multi-drugs and/or their metabolites in human urine.

Device Classification:

The Advin Multi-Drug Screen Test Cassette, Dipcard and Cup are similar to other FDAcleared devices for the qualitative detection of following drugs in urine specimens:

Drug TestCalibratorCutoff
Amphetamine (AMP)d-Amphetamine500 ng/mL
Barbiturates (BAR)Secobarbital300 ng/mL
Benzodiazepines (BZO)Oxazepam300 ng/mL
Buprenorphine (BUP)Buprenorphine10 ng/mL
Cocaine (COC)Benzoylecgonine150 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL

{1}------------------------------------------------

Drug TestCalibratorCutoff
Ecstasy (MDMA)Methylenedioxymethamphetamine500 ng/mL
Methamphetamine (MET)d-Methamphetamine500 ng/mL
Methadone (MTD)Methadone300 ng/mL
Morphine (MOR300)Morphine300 ng/mL
Opiates (OPI)Morphine2,000 ng/mL
Oxycodone (OXY)Oxycodone100 ng/mL
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)Propoxyphene300 ng/mL
Tricyclic-antidepressant(TCA)Nortriptyline1,000 ng/mL
Marijuana (THC)11-nor-Δ9-THC-9-COOH50 ng/mL

These tests are used to provide only preliminary test results. The test systems have been classified as Class II devices with moderate complexity.

Classification Name:

Amphetamine Test System Methamphetamine Test System Barbiturates Test System Benzodiazepine Test system Antidepressant test system Opiate Test System Cannabinoids Test System Cocaine and cocaine metabolite Test System Methadone Test System Tricyclic Antidepressants Test System Methadone Test System Propoxyphene test system

Intended Use:

The Advin Multi-Drug Screen Test is a one-step immunoassay for the qualitative detection

of multiple drugs of abuse and/or their metabolites in human urine at the following cutoff levels:

Drug TestCalibratorCutoff
Amphetamine (AMP)d-Amphetamine500
Barbiturates (BAR)Secobarbital300
BenzodiazepinesOxazepam300
Buprenorphine (BUP)Buprenorphine10 ng/mL
Cocaine (COC)Benzoylecgnine150
EDDP2-ethylidene-1,5-dimethyl-3,3-300
Ecstasy (MDMA)Methylenedioxymethamphetamine500

{2}------------------------------------------------

Methamphetamined-Methamphetamine500
Methadone (MTD)Methadone300
Morphine (MOR300)Morphine300
Opiates (OPI)Morphine2,000
Oxycodone (OXY)Oxycodone100
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)Propoxyphene300
Tricyclic-antidepressantNortriptynine1,000ng/mL
Marijuana (THC)11-nor-Δ9-THC-9-COOH50 ng/mL

The Advin Multi-Drug Screen Test consists of three formats: a Cassette, a Dipcard and a Cup format, which may be configured in any combination of the drugs analytes listed in the above table.

Advin Multi-Drug Screen Test is used to obtain a visual, qualitative, preliminary test result for prescription use in point of care sites, laboratory settings and is also intended for workplace and over-the-counter use. The Propoxyphene test strip is not intended for over-the-counter use.

This test does not distinguish between drugs of abuse and certain medications. It will yield preliminary positive results when prescription drugs TCA, Barbiturates, Benzodiazepine, Methadone, Buprenorphine and Opiates drugs are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for these prescription drugs in urine.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. Gas chromatography/mass spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry/Mass (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are the preferred confirmatory methods.

Clinical considerations and professional judgment should always be applied to interpret any drug of abuse test result, particularly in evaluating a preliminary positive test result.

Description:

The Advin Multi-Drug Screen Test is a lateral flow immunochromatographic competitive binding assay and is intended for prescription use, CLIA Waived use and OTC use. It is used as in vitro diagnostics to visually and qualitatively detect some of the most common drugs of abuse and/or their metabolites in human urine specimens. The test is based on the principle of competitive binding antigen-antibody immunochemistry. It utilizes anti-drug antibody to selectively detect elevated levels of multi-drugs and/or their metabolites in urine at a specified cutoff for each drug. These visual read tests are performed without the use of an instrument.

A drug-Positive urine specimen will not generate a colored-line in the designated test region, while a drug-Negative urine specimen or a urine specimen containing drug(s) at the concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control, a

{3}------------------------------------------------

colored-line will always appear at the control region, indicating that proper volume of urine specimen has been added and membrane wicking has occurred.

Some of the Advin Multi-Drug Screen Test will be sold with either one or two adulteration test strips for the detection of adulterants added to the urine specimen. Each Adulteration strip consists of one to three reagent pads which consist of tests for oxidant, specific gravity, pH, glutaraldehyde, creatinine and nitrite.

Predicate Devices and 510(k) Numbers:

Innovacon Spectrum II Test Card and Test Card with Integrated Test Cup (K061718)

Comparison to a Predicate Device:

The ADVIN Multi-Drug Screen Test (Cassette, Dipcard and Cup) have following similarities compared to the above predicate devices:

  • All tests are assays intended for the qualitative detection of drugs in urine samples. .
  • All tests are intended as a screening method that provides a preliminary qualitative . analytical test result.
  • . All tests are immunochromatographic, lateral flow, competitive binding assays for the rapid detection of drug and drug metabolites with a visual, qualitative end result.
  • . All tests utilize the same basic immunoassay principles that rely on antigen/ antibody interaction to indicate a positive or a negative result.
  • All tests have the same cutoff for each of drugs with the exception of Amphetamine test. ●
FeatureAdvin Multi-Drug Screen TestPredicate devices
Indication of useA rapid chromatographicimmunoassay for the qualitativedetection of multipledrugs/metabolites in human urine.A rapid chromatographicimmunoassay for the qualitative andsimultaneous detection of multipledrugs and metabolites in humanurine.
CutoffConcntrations(ng/mL)Amphetamine 500Barbiturates 300Benzodiazepines 300Buprenorphine 10Cocaine 150Methadone 300MDMA 500THC 50Methamphetamine 500morphine 300, or opiates 2,000Oxycodone 100Phencyclidine 25Propoxyphene 300Tricyclic antidepressants 1,000,EDDP 300Amphetamine 1000 or 300Barbiturates 300Benzodiazepines 300Buprenorphine 10Cocaine 150 or 300Methadone 300MDMA 500THC 50Methamphetamine 500 or 1,000morphine 300, or opiates 2,000Oxycodone 100Phencyclidine 25Propoxyphene 300Tricyclic antidepressants 1,000EDDP 300

{4}------------------------------------------------

Intended usePrescription use, CLIA-waivedlaboratory and Over-the-CounteruseProfessionals and point of care sites
Intended specimenUrineUrine
EndpointColored LinesSame
Materials providedTest Devices (Cassette, Dipcard,Cup)Package InsertTest Devices (Test Card, Cup)Package Insert
MethodologyMembrane Particle AssaySame
Test Time5 minutesSame
FormatImmunochemical Antigen/AntibodyImmunoassaySame

Safety and Effectiveness Data:

Accuracy

The accuracy of Advin Multi-Drug Screen Test was evaluated in comparison to GC/MS analysis data. Forty drug-free urine specimens collected from presumed non-user volunteers were tested with Advin Multi-Drug Screen Test. Of these 40 negative specimens tested, all were correctly identified as negative. 10% of these negative specimens were confirmed with GC/MS as drug negative urine specimens. At least 40 drug positive urine specimens for each drug test were obtained from reference labs throughout the United States. The drug concentration in each of the positive specimen was confirmed with GC/MS and LC/MS (for TCA analysis). Summary of accuracy results on Advin Multi-Drug Screen Test Cassette, Dipcard and Cup are shown in tables below:

Drug TestTestResultDrug-freeCompared to GC/MS Analysis Data%agreement
-50% C/O to< -25% C/O-25 C/O% to C/OC/O to+25% C/O>+25% C/Oto +50% C/O>+50%C/O
AMPNeg403000097.7%
500Pos0012245100%
BARNeg401100095.2%
300Pos0025236100%
BUPNeg401100095.5%
10Pos0028032100%
BZONeg400100093.2%
300Pos0031634100%
COCNeg400300097.7%
150Pos0014153100%
EDDPNeg400100093.2%
300Pos0035233100%
MDMANeg401100095.5%
500Pos0025134100%
METNeg401000093.2%
500Pos0031351100%
MORNeg400100093.2%
300Pos0034053100%
MTDNeg400200095.5%
300Pos0024037100%
OPINeg401000093.2%

A. Summary of Accuracy Results on Advin Multi-Drug Screen Test Cassette:

{5}------------------------------------------------

2,000Pos0024340100%
OXYNeg401000093.2%
100Pos0037133100%
PCPNeg400300097.7
25Pos0013833100%
PPXNeg400100095.3%
300Pos0025233100%
TCANeg400200095.5%
1,000Pos0025728100%
THCNeg401200097.7%
50Pos0014744100%

.

B. Summary of Accuracy Results on Advin Multi-Drug Screen Test Dipcard:

:

:

Drug Testwith Cutoff(ng/mL)TestResultDrug-freeCompared to GC/MS Analysis Data
-50% C/O to< -25% C/O-25 C/O% to C/ODrug-free>+25% C/Oto +50% C/O>+50%C/ODrug-free
AMP500Neg403000097.7%
500Pos0012245100%
BAR300Neg401100095.2%
300Pos0025236100%
BUP10Neg401100095.5%
10Pos0028032100%
BZO300Neg400100093.2%
300Pos0031634100%
COC150Neg400300097.7%
150Pos0014153100%
EDDP300Neg400100093.2%
300Pos0035233100%
MDMA500Neg401100095.5%
500Pos0025134100%
MET500Neg401000093.2%
500Pos0031351100%
MOR300Neg400100093.2%
300Pos0034053100%
MTD300Neg400200095.5%
300Pos0024037100%
OPI2,000Neg401000093.2%
2,000Pos0024340100%
OXY100Neg401000093.2%
100Pos0037133100%
PCP25Neg400300097.7%
25Pos0013833100%
PPX300Neg400100095.3%
300Pos0025233100%
TCA1,000Neg400200095.5%
1,000Pos0025728100%
THC50Neg401200097.7%
50Pos0014744100%

:

,

{6}------------------------------------------------

Drug TestwithCutoff(ng/mL)TestResultDrug-freeCompared to GC/MS Analysis Data%agreement
-50% C/Oto< -25%C/O-25% C/Oto C/OC/O to+25% C/O>+25% C/Oto +50% C/O>+50%C/O
AMP500Neg403000097.7%
AMP500Pos0012245100%
BAR300Neg401100095.2%
BAR300Pos0025236100%
BUP10Neg401100095.5%
BUP10Pos0028032100%
BZO300Neg400100093.2%
BZO300Pos0031634100%
COC150Neg400300097.7%
COC150Pos0014153100%
EDDP300Neg400100093.2%
EDDP300Pos0035233100%
MDMA500Neg401100095.5%
MDMA500Pos0025134100%
MET500Neg401000093.2%
MET500Pos0031351100%
MOR300Neg400100093.2%
MOR300Pos0034053100%
MTD300Neg400200095.5%
MTD300Pos0024037100%
OPI2,000Neg401000093.2%
OPI2,000Pos0024340100%
OXY100Neg401000093.2%
OXY100Pos0037133100%
PCP25Neg400300097.7%
PCP25Pos0013833100%
PPX300Neg400100095.3%
PPX300Pos0025233100%
TCA1,000Neg400200095.5%
TCA1,000Pos0025728100%
THC50Neg401200097.7%
THC50Pos0014744100%

C. Summary of Accuracy Results on Advin Multi-Drug Screen Test Cup:

Analytical Specificity

The following compounds are detected positive in urine by the Advin Multi- Drug Screen Test Cassette, Dip card, and Cup; the concentration of each cross-reactive compound which produced a positive result is in ng/mL and the percentage of the cross-reactivity relative to
the calibrator is in parenthesis:

CompoundsConcentration (%)CompoundsConcentration (%)
AMPD-AmphetamineMDA500 (100%)8,000 (6.5%)L-amphetaminePhentermine50,000 (1%)45,000 (1.1%)
BARSecobarbitalAprobarbital300 (100%)500 (60%)AmobarbitalButabarbital2,500 (12%)100 (300%)
ButalbitalPhenobarbital300 (100%)300 (100%)CyclopentobarbitalPhentobarbital500 (60%)250 (120%)
BUPBuprenorphine10 (100%)
BZOOxazepamAlprazolamClobazamDesalkyflurazepamFlunitrazepamNitrazepamNordiazepamTemazepamTriazolam300 (100%)200 (150%)200 (150%)1,200 (25%)250 (120%)250 (120%)390 (76.9%)150 (200%)2,500 (12%)Alpha-hydroxyalprazolamBromazepamClorazepamDiazepamLorazepamLorazepam-glucuronideNordiazepamNorchlordiazepoxide1,900 (15.8%)1,000 (30%)750 (40%)1,000 (30%)3,900 (7.7%)5,000 (6%)400 (75%)500 (60%)
COCBenzoylecgonineCocaethylene150 (100%)50,000 (0.3%)CocaineEcgonine5,000 (3%)50,000 (0.3%)
EDDPEDDP300 (100%)
METd-Methamphetamined-AmphetamineMDEAMDMA500 (100%)50,000 (1%)30,000 (1.7%)3,500 (14.3%)1R,2S(-) -EphedrineL-AmphetamineMephentermine100,000 (0.5%)50,000 (1%)75,000 (0.7%)
MDMA(+/-) - MDMA(+/-)- MDEA500 (100%)500 (100%)(+/-)-MDA3,900 (12.8%)
MTDMethadone300 (100%)
MORMorphineEthylmorphineHydrocodoneLevophenolNorcodeineThebaine300 (100%)100 (300%)1,250 (24%)50,000 (0.6%)6,000 (1.9%)90,000 (0.3%)CodeineHeroineHydromorphoneMorphine 3-glucuronideOxycodone100 (300%)8,000 (37.5%)2,500 (12%)400 (75%)75,000 (0.4%)
PCPPhencyclidine25 (100%)4-hydroxy-PCP1,500 (1.7%)
OPIMorphineCodeineEthylmorphineHydromorphoneHeroine2,000 (100%)1,800 (111.1%)1500 (133.3%)5,000 (40%)11,000 (18.2%)OxycodoneMorphine-3-glucuronideHydrocodoneThebaine70,000 (2.9%)2,600 (76.9%)5,000 (40%)95,000 (2.1%)
OXYOxycodoneHydrocodoneEthylmorphine100 (100%)5,000 (2%)50,000 (0.2%)HydromorphoneOxymorphoneCodeine25,000 (0.4%)12,500 (0.8)50,000 (0.2%)
PPXPropoxyphene300 (100%)Nor-propoxyphen300 (100%)
TCA
Nortriptyline1,000 (100%)Amitriptyline4,000 (25%)
Clomipramine2,000 (50%)Doxepine1,000 (100%)
Desipramine500 (200%)Imipramine1,000 (100%)
Promethazine1,000 (100%)Trimipramine5,000 (20%)
THC
11-nor-Δ9-THC-9-COOH50 (100%)(+/-)11-hydroxy-Δ9-THC5,000 (1%)
(-)- Δ8-THC20,000 (0.3%)(-)-Δ9-THC20,000 (0.3%)

{7}------------------------------------------------

{8}------------------------------------------------

Precision:

The precision of Advin Multi-Drug Screen Test was evaluated at 3 Physician's Office Laboratory sites and 100 lay persons. Data obtained from these sites indicate that correct test results can be obtained when urine specimen with drug concentrations at +/-50% cutoff levels are tested with Advin Multi-Drug Test Cassette, Dipcard, and Cup.

Lay-Users Study

A total of over one hundred male and female persons age 18 or older participated in the lay person user study with Advin Multi -Drug Screen Test following product package insert. The drug-free urine specimens together with spiked urine specimens with different drug concentrations at +/- 50% cutoff and +/- 25% cutoff were tested by the participants. Each level of the drug solutions was randomly tested with at least 20 replicates.

The lay person user study results indicates that the agreement between the testing results from lay person and GC/MS analysis results is from over 99% for the concentrations at +/-50% cutoff levels and 65% to 86% for the concentrations near the cutoff levels and. The statistic analysis indicates that a lay person can follow the product package insert and perform Advin Multi-Drug Screen Test Cassette, Dip Card and Cup similar to professionals at POL sites with reasonable accuracy.

A Summary of the lay person user study results are listed in the tables below:

Analysis of Discordant Results with Advin Multi-drug Screen Cassette by Lay Person vs. GC/MS Analysis

Drug test cutoff(ng/mL)Number of tests(discordantresults/total results)Result of AdvinMulti-drug TestCassetteGC/MS or LCMS(ng/mL)
AMP 50012/136Positive366
7/86Negative555
BAR 3005/65Positive227
4/44Negative394
BUP 106/66Positive6.9
6/44Negative11.3

{9}------------------------------------------------

BZO 3006/65Positive209
4/41Negative374
COC 1506/68Positive113
4/42Negative186
EDDP5/66Positive212
6/42Negative406
MET 5006/62Positive394
5/44Negative646
MDMA 5006/62Positive368
6/46Negative624
MTD 3007/64Positive218
3/44Negative412
MOR 3005/62Positive255
6/47Negative318
OPI 20005/64Positive1465
4/44Negative2135
OXY6/70Positive67
3/42Negative112
PCP 257/68Positive17.2
3/41Negative35.2
TCA 10006/68Positive784
4/42Negative1210
THC 506/62Positive34.1
5/48Negative57.1

. . . . .

. . . . . . . ・

.

.

: 上

:

.

:

··

Analysis of Discordant Result with Advin Multi-drug Screen Dipcard by Lay Person vs. GC/MS Analysis . .

Drug test cutoff(ng/mL)Number of tests(discordantresults/total results)Result of AdvinMulti-drug TestDipcardGC/MS or LCMS(ng/mL)
AMP 50012/126Positive366
7/86Negative555
BAR 3005/66Positive227
6/40Negative394
BUP 104/63Positive6.9
4/42Negative11.3
BZO 3004/63Positive209
4/42Negative374
COC 1505/64Positive113
4/45Negative186
EDDP 3005/62Positive212
6/43Negative406
MET 5006/65Positive394
4/40Negative646
MDMA 5005/63Positive368
6/41Negative624
MTD 3006/63Positive218
5/42Negative412
MOR 3005/65Positive255

:

. .

{10}------------------------------------------------

4/44Negative318
OPI 20005/63Positive1465
5/41Negative2125
OXY 1005/62Positive67
4/41Negative112
PCP 255/64Positive17.2
4/42Negative35.2
TCA 10005/62Positive784
4/43Negative1210
THC 506/65Positive34.1
4/44Negative57.1

.

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000

.

Analysis of Discordant Result with Advin Multi-drug Screen Cup by Lay Person vs. GC/MS Analysis ·

·

Drug test cutoff(ng/mL)Number of tests(discordantresults/total results)Result of AdvinMulti-drug Test CupGC/MS or LCM(ng/mL)
AMP 50023/256Positive366
22/176Negative555
BAR 3005/193Positive227
6/42Negative394
BUP 104/191Positive6.9
6/44Negative11.3
BZO 3006/196Positive209
6/41Negative374
COC 1506/194Positive113
5/43Negative186
EDDP 3006/194Positive212
5/43Negative406
MET 5006/195Positive394
5/42Negative646
MDMA 5005/191Positive368
6/43Negative624
MTD 3005/193Positive218
6/44Negative412
MOR 3006/127Positive255
5/42Negative381
OPI 20005/129Positive1465
4/41Negative2125
OXY 1007/194Positive67
5/43Negative112
PCP 256/194Positive17.2
4/41Negative35.2
TCA 10005/192Positive784
6/43Negative1210
THC 506/195Positive34.1
5/42Negative57.1

{11}------------------------------------------------

Conclusion:

.

From data collected in the accuracy and precision studies, it was demonstrated that Advin Multi-Drug Screen Test is safe and effective to use when compared to predicate devices already marketed in the United States.

:

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2013

Advin Biotech C/O Edward Tung 6861 Nancy Ridge Dr., Suite #D SAN DIEGO CA 92121

Re: K122809

Trade/Device Name: Advin Multi-Drug Screen Test Cassette Advin Multi-Drug Screen Test Dipcard Advin Multi-Drug Screen Test Cup

Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: II Product Code: DKZ, DJC, DIS, JXM, DJG, LDJ, DIO, DJR, LFG, JXN, LCM Dated: June 03, 2013 Received: June 06, 2013

Dear Dr. Tung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{13}------------------------------------------------

Page 2-Dr. Tung

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney @was, Ph.D.

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Indications for Use

510(k) Number (if known): K122809

Device Name: Advin Multi-Drug Screen Test Cassette Advin Multi-Drug Screen Test Dip Card Advin Multi-Drug Screen Test Cup

:

Indications for Use:

The Advin Multi-Drug Screen Test is a one-step immunoassay for the qualitative detection of multiple drugs of abuse and/or their metabolites in human urine at the following cutoff levels:

Drug TestCalibratorCutoff Level
Amphetamine (AMP)d-Amphetamine.500 ng/mL
Barbiturates (BAR)Secobarbital300 ng/mL
Benzodiazepines (BZO)Oxazepam300 ng/mL
Buprenorphine (BUP)Buprenorphine10 ng/mL
Cocaine (COC)Benzoylecgnine150 ng/mL
EDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Ecstasy (MDMA)Methylenedioxymethamphetamine500 ng/mL
Methamphetamine (MET)d-Methamphetamine500 ng/mL
Methadone (MTD)Methadone300 ng/mL
Morphine (MOR300)Morphine300 ng/mL
Opiates (OPI)Morphine2,000 ng/mL
Oxycodone (OXY)Oxycodone100 ng/mL
Phencyclidine (PCP)Phencyclidine25 ng/mL
Propoxyphene (PPX)Propoxyphene300 ng/mL
Tricyclic-antidepressant (TCA)Nortriptynine1,000 ng/mL
Marijuana (THC)11-nor-Δ9-THC-9-COOH50 ng/mL

Prescription Use X (21 CFR Part 801 Subpart D) .And/Or ・・

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. ሬ🇿e͡§ler -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k) K122809

{15}------------------------------------------------

Indications for Use

510(k) Number (if known): K122809

Device Name: Advin Multi-Drug Screen Test Cassette Advin Multi-Drug Screen Test Dip Card Advin Multi-Drug Screen Test Cup

Indications for Use:

The Advin Multi-Drug Screen Test consists of three formats: a Cassette, a Dip Card and a Cup, which may be configured in any combination of the drug analytes listed in the above table.

Advin Multi-Drug Screen Test is used to obtain a visual, qualitative, preliminary test result for prescription use in point of care sites, laboratory settings and is also intended for workplace and over-the-counter use. The Propoxyphene test strip is not intended for over-the-counter use.

The Advin Multi-Drug Screen Test will yield preliminary positive results when prescription drugs TCA, Barbiturates and Benzodiazepine, Methadone, Buprenorphine and Opiates are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for TCA, Barbiturates and Benzodiazepine in urine.

The Advin Multi-Drug Screen Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. Gas chromatography/mass spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry/Mass (LC/MS/MS) and High Performance Liquid Chromatography-(HPLC)-are the-preferred-confirmatory-methods-Glinical-consideration and professional judgment should be applied to any drug of abuse test result, particularly in the evaluation of a preliminary positive test result.

Prescription Use X (21 CFR Part 801 Subpart D)

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

And/Or

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. Chester -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k) K122809

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).