K Number
K182004

Validate with FDA (Live)

Date Cleared
2018-08-21

(25 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PENTAX Medical Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Track including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are pediatric patient populations.

The PENTAX Medical Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

Device Description

The PENTAX Medical Ultrasound Video Gastroscopes used to provide visualization of, and therapeutic access to, the upper gastrointestinal tract. These endoscopes are used with cleared Pentax Video processors (a software-controlled device) and cleared Hitachi Ultrasound Scanner (a software-controlled device) to comprise the PENTAX Medical Endoscopic Ultrasound System. The endoscopes have a flexible insertion tube, a control body, PVE umbilical connector, and ultrasound scanner umbilical connector. The PVE umbilical connector will be attached to the Video Processor and has connections for illumination, video signals, air/ water/ and suction.

The ultrasound scanner umbilical connector will be attached to the ultrasound scanner unit. A sterile, single use disposable latex balloon is fitted over the convex array ultrasound transducer prior to the procedure. During an ultrasound endoscopy procedure, the latex balloon is inflated with water. The water that is contained within the balloon creates a water field that covers the transducer. The water field enables more effectively transport of ultrasonic pulses from the ultrasound transducer to the target anatomical site and back to the ultrasound transducer.

The control body includes controls for up/ down/ left/ right angulation, air/ water delivery, and an accessory inlet port. The endoscope contains light carrying bundles to illuminate the body cavity, a charge couple device (CCD) to collect endoscopic image data, and a linear or radial array ultrasound transducer to collect ultrasonic image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced.

The video processor contains a lamp that provides white light and is focused at the PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data and other screen display information are formatted and presented to the video outputs of the video processor for display. The ultrasound transducer delivers ultrasonic pulses, reflections of the pulses are received, and the signals are passed to the uitrasound scanner for processing and display. The instrument is immersible (with the use of supplied cleaning accessories) except for the ultrasound scanner connector (as described in the endoscope operator manual cleaning instructions)

AI/ML Overview

This FDA 510(k) summary describes the PENTAX Medical Endoscopic Ultrasound System. However, it does not contain the specific details required to fully address your request about acceptance criteria and study results. The document primarily focuses on demonstrating substantial equivalence to a predicate device and discusses minor design changes and compatibility with different processors and scanners.

Here's a breakdown of what can and cannot be extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document

  • does not explicitly list quantitative acceptance criteria
  • and does not provide numerical reported device performance metrics tied to such criteria.

It states: "The PENTAX Medical Endoscopic Ultrasound System has been successfully tested for its functions, performance and safety as per FDA recognized consensus standards." This is a general statement of compliance, not a presentation of specific performance data against defined thresholds.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify

  • the sample size used for performance testing (test set)
  • nor the data provenance (country of origin, retrospective/prospective).

It mentions "performance testing," but no details about the test set are provided.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications:

The document does not

  • mention the use of experts to establish ground truth
  • or their qualifications.

Given the nature of the device (endoscopic ultrasound system), ground truth for image quality and functionality would likely be assessed through objective measurements and potentially expert review of image outputs, but this is not detailed.

4. Adjudication Method for the Test Set:

Not applicable, as no information regarding human review or adjudication of data is provided.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

The document does not mention

  • an MRMC comparative effectiveness study
  • or any effect size for human readers with and without AI assistance.

The device described is an imaging system, not an AI-assisted diagnostic tool in the sense of providing automated interpretations to readers.

6. Standalone (Algorithm Only) Performance:

Not applicable. The device is a medical imaging system, not a standalone algorithm. Its performance is intrinsically linked to its hardware and software components working together to present visual and ultrasonic data to a human operator.

7. Type of Ground Truth Used:

The document does not explicitly state the type of ground truth used for any testing. For an endoscopic ultrasound system, ground truth might involve:

  • Physical phantoms: For evaluating imaging capabilities.
  • Clinical observations/pathology: For accuracy of visualization in a clinical setting (though no clinical study data is provided in this summary).
  • Developer-defined functional criteria: For verifying the various functions of the system (e.g., image clarity, angulation, accessory channel functionality).

8. Sample Size for the Training Set:

Not applicable, as the document describes an imaging device, not a machine learning model that would require a distinct training set in the typical sense. The software mentioned is for control and processing, not for AI-based interpretation.

9. How Ground Truth for the Training Set Was Established:

Not applicable, for the same reason as point 8.


Summary of what is present:

The document confirms that:

  • Software verification and validation testing were conducted as per FDA guidance.
  • The software is classified as Class A (IEC 62304:2006) with a "Moderate" level of concern.
  • Cybersecurity risks were assessed and mitigated according to FDA guidances.
  • EMC and Electrical Safety were confirmed by relevant IEC standards (IEC 60601-1-2:2007; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+AM 1:2012; and IEC 60601-2-37:2007).

However, these are tests of compliance with standards and safety, not performance against specific clinical or quantitative acceptance criteria for image quality or diagnostic accuracy. The substantial equivalence argument relies on the device having "no effect on the performance, function or general intended use" in comparison to the predicate, and that "The PENTAX Medical Endoscopic Ultrasound System is functionally equivalent to its predicate device."

To obtain the detailed information you requested about acceptance criteria and specific study results, one would typically need to review the full 510(k) submission, not just the publicly available summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 21, 2018

PENTAX of America, Inc. William Goeller Vice President, Quality and Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K182004

Trade/Device Name: PENTAX Medical Endoscopic Ultrasound System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODG, ITX Dated: July 26, 2018 Received: July 27, 2018

Dear William Goeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the FDA logo. The logo is in a light gray color. The letters FDA are large and bold.

Jeffrey W. Cooper -2 2018.08.21 13:35:18 -04'00

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182004

Device Name

PENTAX Medical Endoscopic Ultrasound System

Indications for Use (Describe)

The PENTAX Medical Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Track including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are pediatric patient populations.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below "PENTAX" is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif font. The logo is simple and professional.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.

K182004 Page 1 of 3

Applicant: PENTAX Medical

HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782

Contact: William Goeller Vice President, Quality and Regulatory Affairs PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645-1782 Telephone: (201)571-2300 ext. 2318 FAX: (201)391-4189 Email: william.goeller@pentaxmedical.com

Date Prepared: July 25, 2018

Common Name: Endoscopic Ultrasound / Ultrasound Gastroscope Name of the System: PENTAX Medical Endoscopic Ultrasound System

EG-3270UK Upper G.I. Video Scope (Convex Array Type, K162447), EG-3670URK Video Gastroscope (Radial Array Type, K130206) and EG-3870UTK (Convex Array Type, K130247) with EPK-i5010 Video Processor (K122470), EPK-i7010 Video Processor (K150618) plus Hitachi HI VISION Preirus Ultrasound Scanner (K093466) and Hitachi ARIETTA 70 Ultrasound Scanner (K134016) Regulation Number:

21 CFR Part 876.1500

Regulation Names:

Endoscope and accessories Diagnostic Ultrasound Transducer Regulatory Class: Class II Product Code: ODG, ITX Predicate Device: PENTAX Medical Endoscopic Ultrasound System (K162447)

The proposed PENTAX modification is to bundle the PENTAX Medical legacy Ultrasound Video Gastroscopes EG3270UK, EG3670URK, and EG3870UTK, and Pentax Video Processors EPK-i5010 and EPK-i7010, for use with Hitachi's HI VISION Preirus and ARIETTA 70. All the devices have been previously cleared.

This 510(k) captures some minor design changes that have occurred during the evolution of the product line that includes these three legacy ultrasound video gastroscopes. Although the changes are believed to be minor, the 510(k) is being submitted to account for technological advances in associated compatible devices and to ensure that FDA has the most current information concerning the PENTAX Medical ultrasound video gastroscopes.

The subject device has the same indications for use, composition of patient contact materials, viewing direction, image size, and reprocessing/sterilization method as the predicate. The subject device uses the same processors and peripherals as the predicate device with three additions.

PENTAX Medical Confidential

{4}------------------------------------------------

PENTAX Medical Endoscopic Ultrasound System

510(k) Summary

Image /page/4/Picture/2 description: The image shows the logo for Pentax Medical. The logo has the word "PENTAX" in red, with the word "MEDICAL" in gray underneath. The text "K182004 Page 2 of 3" is located in the bottom right corner of the image.

The main differences between the subject devices and predicate devices are as follows:

  • Bundling of EG-3670URK and EG3870UTK Ultrasound Video Gastroscopes with EG3270UK
  • Addition of the EPK-i7010 as a compatible Video Processor
  • . Addition of the Hitachi ARIETTA 70 as a compatible Ultrasound Scanner

Device Description:

The PENTAX Medical Ultrasound Video Gastroscopes used to provide visualization of, and therapeutic access to, the upper gastrointestinal tract. These endoscopes are used with cleared Pentax Video processors (a software-controlled device) and cleared Hitachi Ultrasound Scanner (a software-controlled device) to comprise the PENTAX Medical Endoscopic Ultrasound System. The endoscopes have a flexible insertion tube, a control body, PVE umbilical connector, and ultrasound scanner umbilical connector. The PVE umbilical connector will be attached to the Video Processor and has connections for illumination, video signals, air/ water/ and suction.

The ultrasound scanner umbilical connector will be attached to the ultrasound scanner unit. A sterile, single use disposable latex balloon is fitted over the convex array ultrasound transducer prior to the procedure. During an ultrasound endoscopy procedure, the latex balloon is inflated with water. The water that is contained within the balloon creates a water field that covers the transducer. The water field enables more effectively transport of ultrasonic pulses from the ultrasound transducer to the target anatomical site and back to the ultrasound transducer.

The control body includes controls for up/ down/ left/ right angulation, air/ water delivery, and an accessory inlet port. The endoscope contains light carrying bundles to illuminate the body cavity, a charge couple device (CCD) to collect endoscopic image data, and a linear or radial array ultrasound transducer to collect ultrasonic image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced.

The video processor contains a lamp that provides white light and is focused at the PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data and other screen display information are formatted and presented to the video outputs of the video processor for display. The ultrasound transducer delivers ultrasonic pulses, reflections of the pulses are received, and the signals are passed to the uitrasound scanner for processing and display. The instrument is immersible (with the use of supplied cleaning accessories) except for the ultrasound scanner connector (as described in the endoscope operator manual cleaning instructions)

Intended Use / Indications for use

The PENTAX Medical Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

Summary of Technological Characteristics:

The PENTAX Medical Endoscopic Ultrasound System is functionally equivalent to its predicate device, the PENTAX Medical Endoscopic Ultrasound System cleared by FDA in 2016. The only difference between the two devices is that the predicate can be used with the EPK-i5010 video processor and HI VISION Preirus Ultrasound scanner, whereas the subject device now additionally bundles the EG-3670URK and

PENTAX Medical Confidential

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below "PENTAX" is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif font. The logo is simple and modern.

Page 3 of 3

510(k) Summary

EG-3870UTK Ultrasound Video Gastroscopes as a family of endoscopes and the subject device is now additionally compatabile with the EPK-i7010 video processor and Hitache ARIETTA 70 Ultrasound scanner.

The additional devices have been evaluated through performance testing and raise of safety and effectiveness of the device as these differences have no effect on the performance, function or general intended use of the device.

Non-Clinical Performance Data

The PENTAX Medical Endoscopic Ultrasound System has been successfully tested for its functions, performance and safety as per FDA recognized consensus standards. The following performance data are summarized in support of the substantial equivalence determination.

Software

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software is classified as CLASS A under the Software Safety Classification per IEC 62304:2006, Medical device software- Software life cycle processes) and the software level of concern is "Moderate" based on the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Cybersecurity risks have been assessed and mitigated according to the FDA Guidances for Industry and Staff "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued October 2, 2014, and "Postmarket Management of Cybersecurity in Medical Devices." issued December 28,2016.

EMC and Electrical Safety

The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the PENTAX Medical Endoscopic Ultrasound System were confirmed by the following standards: IEC 60601-1-2:2007; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+AM 1:2012; and IEC 60601-2-37:2007.

Substantial Equivalence Discussion:

After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labeling, and sterilization method, we conclude that the subject device PENTAX Medical Endoscopic Ultrasound System is as safe and effective as the predicate device. There are no differences in indications for use and intended use between the subject and predicate device and are therefore, substantially equivalent. The technological differences in terms of design features, performance characteristics and constituent materials are not substantive.

Conclusion:

Accordingly, PENTAX Medical believes the PENTAX Medical Endoscopic Ultrasound System is substantially equivalent to the identified predicate, the PENTAX Medical Endoscopic Ultrasound System, cleared by FDA in 2016.

PENTAX Medical Confidential

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.