K Number
K122470
Manufacturer
Date Cleared
2013-04-10

(240 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™ a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

Device Description

The PENTAX EPK-i5010 video processor consists of a video system, integrated light source, monitor, and ancillary equipment. This processor is intended for endoscopic diagnostic, treatment and video observation. It is intended to process electrical signals from a video endoscope.

The PENTAX EPK-i5010 offers an optional digital, post-processing imaging enhancement technology called the PENTAX i-Scan™ modes 1, 2, and 3, which is intended to give the user an enhanced view of the texture of the mucosal surface and blood vessels. i-Scan 1 provides the user with a view that sharpens surface vessels and enhances surface texture of the mucosa. i-Scan 2 provides the user with increased visibility of blood vessels while also providing the same enhancements to the mucosa achieved in i-Scan 1. i-Scan 3 provides the user with increased visibility of blood vessels including dimly illuminated far-field regions while also providing the same enhancement to the mucosa achieved in i-Scan 1. The user can select either white light image or i-Scan modes by pressing a pre-programmed button on the scope, by using a pre-programmed foot pedal or by pressing a keyboard button. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.

The EPK-i5010 video processor incorporates the hardware of the EPK-i5020 video processor model (K 113873) and the PENTAX i-Scan™ technology.

The EPKi-5010 is compatible with PENTAX flexible and rigid k-Series and i-Series videoscopes. The subject premarket notification is specific for gastrointestinal videoscopes and colonovideoscopes.

AI/ML Overview

The provided text describes the PENTAX EPK-i5010 Video Processor and its i-Scan™ technology, intended for endoscopic diagnosis, treatment, and video observation. However, it does not contain specific acceptance criteria for "device performance" in terms of diagnostic accuracy metrics (e.g., sensitivity, specificity, AUC) or a detailed study specifically designed to prove these criteria are met for the i-Scan technology itself.

Instead, the document focuses on:

  • Substantial Equivalence: The primary aim of the 510(k) submission is to demonstrate that the PENTAX EPK-i5010 is substantially equivalent to legally marketed predicate devices, meaning it does not introduce new questions of safety or effectiveness.
  • Safety: The clinical studies mentioned are primarily to establish the safety of the device, not its diagnostic performance against specific acceptance criteria for imaging enhancement.

Based on the provided text, here is an analysis of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Not explicitly stated for diagnostic performance)Reported Device Performance (Focus on Safety and Equivalence)
Implicit Acceptance Criteria:Reported Performance:
- No safety issues associated with the device.- "Out of the 975 patients in the seven (7) clinical studies conducted with the PENTAX EPK-i video processor there were no safety issues reported."
- Substantially equivalent to predicate devices regarding intended use, method of operation, material, or design, ensuring no adverse effect on safety or effectiveness.- "The EPK-i5010 does not incorporate any significant changes in the Intended Use, in the Method of Operation, Material or Design that could be considered to affect the safety or effectiveness of the Medical Device" compared to predicate devices (Olympus EVIS EXERA III VIDEO SYSTEM, PENTAX EPK-i5020 Video Processor).
- Software validation performed according to IEC 62304.- Software validation performed in accordance with IEC 62304.
- Electrical testing according to IEC 60601.- Electrical testing according to IEC 60601.
- Image Enhancement Analysis (method not specified).- Image Enhancement Analysis was conducted. (No quantitative performance metrics reported for this analysis in the provided text).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The text mentions 7 clinical studies for safety evaluation:

  • Sample Size: A total of 975 patients.
  • Data Provenance:
    • Germany (3 studies)
    • Japan (1 study)
    • Korea (3 studies)
  • Retrospective/Prospective: Not specified, but "clinical studies" typically imply prospective data collection for safety and efficacy endpoints. However, without further details, it cannot be definitively determined from this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not provide this information. The clinical studies mentioned were for safety, and there's no indication of a diagnostic performance evaluation with expert-established ground truth for the i-Scan technology.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not provide this information as it does not detail a diagnostic performance study requiring ground truth adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, the document does not describe a MRMC comparative effectiveness study. The i-Scan technology is described as an "optional adjunct" and "not intended to replace histopathological sampling," implying it's a visual enhancement tool rather than a diagnostic AI that would be compared with human performance or AI assistance.
  • Effect Size: Not applicable, as no MRMC study or quantitative diagnostic performance study is described.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

No, the document does not describe a standalone performance study for the i-Scan algorithm. The i-Scan is an imaging enhancement technology integrated into a video processor, intended to be used by a clinician during endoscopy. It's not presented as a standalone diagnostic algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document does not specify a type of ground truth because it does not detail a diagnostic performance study where ground truth would be needed for evaluating the i-Scan's diagnostic accuracy. The safety studies would primarily rely on adverse event reporting and clinical observation for safety endpoints.

8. The sample size for the training set

The document does not provide this information. As a post-processing imaging enhancement technology, the i-Scan likely uses algorithms that are developed and refined, but the document does not specify a "training set" in the context of machine learning model development or validation.

9. How the ground truth for the training set was established

The document does not provide this information as it does not mention a training set or the establishment of ground truth for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the word "PENTAX" in large, bold letters. Above the word "PENTAX" is the code "K122470". Below the word "PENTAX" is the text "PENTAX Medical Company A Division of PENTAX of America, Inc."

510(K) Summary

APR 1 0 2013

The following summary is provided in accordance with 21 CFR 807.92:

Date: April 08, 2013

Sponsor Information:

PENTAX Medical Company, A Division of PENTAX America, Inc. 3 Paragon Drive Montvale, New Jersey 07645-1782

Contact Person:

Name:Krishna Govindarajan
Title:Regulatory Manager
Telephone:(201) 251-2300 x 2125
Facsimile:(201) 251-4117
E-Mail:Krishna.govindarajan@pentaxmedical.com

Manufacturer:

HOYA Corporation PENTAX Yamagata Factory 4-1 Hinode-Cho Nagai-Shi, Japan, 993-0012

FDA Establishment Registration #3003782606

Identification of the Proposed Device:

Proprietary/Trade Name:PENTAX EPK-i5010 Video Processor
Common/Usual Name:Endoscopic Video Processor and Light Source
Classification Name:Endoscopic video imaging system/component,gastroenterology-urology
Regulation Number:21 CFR Part 876.1500
Regulation Description:Endoscope and accessories
Medical Specialty:Gastroenterology/Urology
Regulatory Class:Class II
Product Code:FET and GCT

Device Description:

The PENTAX EPK-i5010 video processor consists of a video system, integrated light source, monitor, and ancillary equipment. This processor is intended for endoscopic diagnostic, treatment and video observation. It is intended to process electrical signals from a video endoscope.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for PENTAX Medical Company. The logo is in black and white and features the word "PENTAX" in large, bold letters. Above the word "PENTAX" is the code "K122470". Below the word "PENTAX" is the text "PENTAX Medical Company A Division of PENTAX of America, Inc."

The PENTAX EPK-i5010 offers an optional digital, post-processing imaging enhancement technology called the PENTAX i-Scan™ modes 1, 2, and 3, which is intended to give the user an enhanced view of the texture of the mucosal surface and blood vessels. i-Scan 1 provides the user with a view that sharpens surface vessels and enhances surface texture of the mucosa. i-Scan 2 provides the user with increased visibility of blood vessels while also providing the same enhancements to the mucosa achieved in i-Scan 1. i-Scan 3 provides the user with increased visibility of blood vessels including dimly illuminated far-field regions while also providing the same enhancement to the mucosa achieved in i-Scan 1. The user can select either white light image or i-Scan modes by pressing a pre-programmed button on the scope, by using a pre-programmed foot pedal or by pressing a keyboard button. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.

The EPK-i5010 video processor incorporates the hardware of the EPK-i5020 video processor model (K 113873) and the PENTAX i-Scan™ technology.

The EPKi-5010 is compatible with PENTAX flexible and rigid k-Series and i-Series videoscopes. The subject premarket notification is specific for gastrointestinal videoscopes and colonovideoscopes.

PENTAX Medical Company considers video endoscope families, defined as a grouping of endoscopes with the same: Intended diagnostic/ therapeutic affect for given anatomy (with respect to the indication for procedure), method(s) of introduction (trans oral, trans rectal, etc). Technical characteristics (design, materials, etc), and range of descriptive characteristics (specifications). Changes within the families are considered not-significant changes and mechanics for the description and grouping of endoscope families include 510(k) consideration for changes.

, Intended Use:

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adiunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

Technology Characteristics:

The EPK-i5010 Video Processor has the same fundamental scientific technology as commercially available endoscope video processor and the substantially equivalent predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the text "K122470" at the top, followed by the word "PENTAX" in large, bold letters. Below that, it says "PENTAX Medical Company" and "A Division of PENTAX of America, Inc." in a smaller font.

Substantial Equivalence Determination:

1. Summary of Nonclinical Tests

The PENTAX EPK-i5010 Video Processor complies with voluntary standards as detailed in this premarket notification submission. The following quality assurance measures were applied to the design and development:

  • Animal Model Testing ●
  • . Image Enhancement Analysis
  • Endoscope Optical Testing ●
  • Risk Management (Analysis & Mitigation) ●
  • Requirements Reviews .
  • Design Reviews .
  • Unit Testing (Software & Hardware Verification) .
  • Integration testing (Software & Hardware Verification) 0
  • t Performance testing (Software & Hardware Verification)
  • Safety testing (System level Verification & Validation) .
  • t Simulated use testing (System level QA Validation)
  • Electrical testing according to IEC 60601 ●
  • Software validation performed in accordance with IEC 62304 .

2. Summary of Clinical Tests

A total of seven (7) clinical studies are presented in this review to help establish the safety of the device. The studies were conducted in Germany (3), Japan (1) and Korea (3). An overall total of 975 patients were evaluated in the 7 clinical studies. All of the 7 studies used the PENTAX EPK-i processor with i-Scan technology.

Four of the studies were conducted on the esophagus totaling 350 patients and three of the studies were conducted on the colon, totaling 625 patients. It should be noted that out of the 975 patients in the seven (7) clinical studies conducted with the PENTAX EPK-i video processor there were no safety issues reported.

3. Predicate Device:

PENTAX EPK-i5010 Video Processor is substantially equivalent to the following devices:

Legally Marketed Device(s)510(k) #
EVIS EXERA III VIDEO SYSTEMK112680
PENTAX EPK-i5020 Video ProcessorK113873

The subject device is also substantially equivalent to other legally marketed video processors for use with endoscopes including, but not limited to the Olympus predicate K100584. The EPK-i5010 video processor incorporates the hardware of the EPK-i5020 video processor model and the optional i-Scan™ technology. All other functions and specifications between the two processors are substantially identical.

The similarities and differences between the Olympus EVIS EXERA II and III Video System and the PENTAX EPK-i5010 Video Processor can be seen in the Table 1.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the text "K122470" at the top, followed by the word "PENTAX" in large, bold letters. Below that, it says "PENTAX Medical Company" and "A Division of PENTAX of America, Inc."

When compared to the predicate devices, the EPK-i5010 does not incorporate any significant changes in the Intended Use, in the Method of Operation, Material or Design that could be considered to affect the safety or effectiveness of the Medical Device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized human figure with outstretched arms, overlaid with three parallel lines that curve upwards, resembling a stylized bird or wing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 22, 2013

PENTAX of America, Inc. PENTAX Medical Company % Robert Schiff, Ph.D., RAC, CQA, FRAPS President Schiff & Company, Inc. 1129 Bloomfield Avenue WEST CALDWELL NJ 07006

Re: K122470 Trade/Device Name: PENTAX EPK-i5010 Video Processor Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: PEA Dated: March 6, 2013 Received: March 7, 2013

Dear Dr. Schiff:

This letter corrects our substantially equivalent letter of April 10, 2013.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Robert Schiff. Ph.D., RAC, COA, FRAPS

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21

CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K122470

Device Name: PENTAX EPK-i5010 Video Processor

Indications for Use:

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™ a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -S

(Division Sign-Off) Division of Reproductive. Gastro-Renal, and Urological Devices K122470 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Page 1 of 1

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.