(157 days)
The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. The system is not intended for use in the coronary, carotid, iliac, pulmonary, or renal vasculature.
Phoenix Atherectomy Plus System:
When used with the Phoenix Aspiration Pump as the vacuum source, the Phoenix 2.4mm Deflecting Atherectomy System is indicated for the removal of thrombus from vessels of the peripheral arterial vasculature.
The Phoenix 2.4mm Atherectomy Plus System comprises the currently cleared Phoenix 2.4mm Deflecting Atherectomy System (K172386) and a new Phoenix Aspiration Pump. which when used in conjunction with each other are designed for removal of thrombus from vessels of the peripheral arterial vasculature.
The Phoenix 2.4mm Deflecting Atherectomy System and the Phoenix Aspiration Pump are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The Phoenix 2.4mm Deflecting Atherectomy System comprises the Phoenix 2.4mm Deflecting Atherectomy Catheter and the Phoenix Atherectomy Handle with Wire Support Clip.
The Phoenix Catheter is an over-the-wire (OTW), multi-lumen Catheter with a cutter at the distal tip that continuously captures and clears debulked (excised) material proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectorny Catheter at its specified rotational speed, and is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Phoenix Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System. The Phoenix Catheter is compatible with commercially available 0.014" exchange length (260 cm or greater) guidewires.
For the purpose of this 510(k), the design of the Phoenix 2.4mm Deflecting Atherectomy System remains unchanged.
The Phoenix Aspiration Pump connects to the disposal outlet of the Phoenix Catheter via the pump's aspiration tubing. The exit end of the aspiration tubing is connected to a waste disposal bag to collect aspirated material. A 60 ml syringe is provided as part of the Aspiration Pump Assembly to assist in priming the pump aspiration tubing and purging the aspiration system of air. The Aspiration Pump is battery-operated and is operated by an ON/OFF switch. It serves as a vacuum source for the Aspiration System for aspirating the thrombus from the target vessel out of the Phoenix Catheter via the pump aspiration tubing and into the disposal bag. The Aspiration Pump is non-patient contacting.
The provided text is a 510(k) summary for the Phoenix 2.4mm Atherectomy Plus System. It describes the device, its intended use, and a comparison to predicate and reference devices, as well as a summary of testing performed to demonstrate substantial equivalence. However, it does not contain the specific information requested about acceptance criteria and a study that proves the device meets those criteria in the context of an AI/algorithm-driven device performance study.
The document describes the device as a mechanical atherectomy system with an aspiration pump for removing thrombus, which is a physical device, not an AI or algorithm. The "Testing Summary" section lists various engineering and preclinical tests (e.g., visual inspection, vacuum and leak test, flow rate test, system simulated use test, preclinical animal testing), which are typical for medical devices of this nature. There is no mention of an algorithm, AI, or any data-driven performance study comparing human readers with and without AI assistance, or standalone algorithm performance, or ground truth establishment based on expert consensus for image analysis.
Therefore, I cannot provide the requested information from the given text. The text does not detail an AI/algorithm-driven study with acceptance criteria.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 17, 2018
Volcano AtheroMed Inc. Ms. Jean Chang Senior Director, Operations 1530 O'Brien Drive, Suite A Menlo Park, California 94025
Re: K181877
Trade/Device Name: Phoenix 2.4mm Atherectomy Plus System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: November 15, 2018 Received: November 16, 2018
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Date: / 2018.12.17 Eleni Whatley 13:25:35 -05'00' For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K181877
Device Name Phoenix 2.4 Atherectomy Plus System
Indications for Use (Describe)
The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. The system is not intended for use in the coronary, carotid, iliac, pulmonary, or renal vasculature.
Phoenix Atherectomy Plus System:
When used with the Phoenix Aspiration Pump as the vacuum source, the Phoenix 2.4 Deflecting Atherectomy System is indicated for the removal of thrombus from vessels of the peripheral arterial vasculature.
| Type of Use (Select one or both, as applicable) | ||
|---|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ||
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Submitter Information:
| Date of 510(k) Summary Preparation: | November 29, 2018 |
|---|---|
| Name and Address of Manufacturer: | Volcano AtheroMed, Inc.1530 O'Brien Dr.Menlo Park, CA 94025 |
| Contact Person: | Jean ChangSenior Director, OperationsPhone: (650) 352 5255 |
Subject Device:
| Device Trade Name: | Phoenix® 2.4mm Atherectomy Plus System |
|---|---|
| Common Name: | Atherectomy System |
| Regulation Description: | Intraluminal Artery Stripper |
| Regulation Number: | 21 CFR 870.4875 |
| Product Code: | MCW |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
| Predicate Devices: | |
| Primary Predicate: | Jetstream Atherectomy System (K130637) |
| Reference Devices: | The Phoenix® Atherectomy System(K172386) & the Penumbra INDIGOTMAspiration System (K161523) serve asadditional reference devices |
{4}------------------------------------------------
510(k) Premarket Notification K181877
Device Description:
The Phoenix 2.4mm Atherectomy Plus System comprises the currently cleared Phoenix 2.4mm Deflecting Atherectomy System (K172386) and a new Phoenix Aspiration Pump. which when used in conjunction with each other are designed for removal of thrombus from vessels of the peripheral arterial vasculature.
The Phoenix 2.4mm Deflecting Atherectomy System and the Phoenix Aspiration Pump are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The Phoenix 2.4mm Deflecting Atherectomy System comprises the Phoenix 2.4mm Deflecting Atherectomy Catheter and the Phoenix Atherectomy Handle with Wire Support Clip.
The Phoenix Catheter is an over-the-wire (OTW), multi-lumen Catheter with a cutter at the distal tip that continuously captures and clears debulked (excised) material proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectorny Catheter at its specified rotational speed, and is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Phoenix Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System. The Phoenix Catheter is compatible with commercially available 0.014" exchange length (260 cm or greater) guidewires.
For the purpose of this 510(k), the design of the Phoenix 2.4mm Deflecting Atherectomy System remains unchanged.
The Phoenix Aspiration Pump connects to the disposal outlet of the Phoenix Catheter via the pump's aspiration tubing. The exit end of the aspiration tubing is connected to a waste disposal bag to collect aspirated material. A 60 ml syringe is provided as part of the Aspiration Pump Assembly to assist in priming the pump aspiration tubing and purging the aspiration system of air. The Aspiration Pump is battery-operated and is operated by an ON/OFF switch. It serves as a vacuum source for the Aspiration System for aspirating the thrombus from the target vessel out of the Phoenix Catheter via the pump aspiration tubing and into the disposal bag. The Aspiration Pump is non-patient contacting.
{5}------------------------------------------------
510(k) Premarket Notification K181877
Indications for Use:
The Phoenix Atherectomy System is intended for use in atherectomy of the perioheral vasculature. The system is not intended for use in the coronary, carotid, iliac, pulmonary, or renal vasculature.
Phoenix Atherectomy Plus System:
When used with the Phoenix Aspiration Pump as the vacuum source, the Phoenix 2,4mm Deflecting Atherectomy System is indicated for the removal of thrombus from vessels of the peripheral arterial vasculature.
Comparison with Predicate:
The Phoenix 2.4mm Atherectomy Plus System is the subject of this submission for removal of thrombus from peripheral arterial vasculature. The Phoenix 2.4mm Atherectomy Plus System has been assessed for substantial equivalence relative to the currently marketed Boston Scientific Jetstream Atherectomy System as the predicate device, along with consideration of two additional reference devices, the Phoenix 2.4mm Deflecting Atherectomy System and the Penumbra INDIGO Aspiration System. The Phoenix 2.4mm Atherectomy Plus System has the same design, materials, technological characteristics and principle of operation of the reference Phoenix 2.4mm Deflecting Catheter and Handle, There is no change to the currently marketed 2.4mm Deflecting Atherectomy Catheter and Handle with Wire Support Clip. The Phoenix Aspiration Pump that is used as part of the Phoenix Atherectomy Plus System is comparable to the Penumbra INDIGO Aspiration System and shares a similar mechanism of action. The proposed indications for use to add aspiration for thrombus removal for the Phoenix Atherectomy Plus System are similar to those of the predicate Boston Scientific Jetstream Atherectomy System, with respect to a combined atherectomy and thrombectomy indication, as well as the reference Penumbra INDIGO System with thrombus removal using vacuum aspiration. Any differences between the subject and predicate device were evaluated through technological comparison (also taking into account the noted reference devices) and design verification and validation testing, which demonstrated no new questions of safety and effectiveness.
The similarities and differences between the subject device and the predicate and reference devices are outlined in the table below:
{6}------------------------------------------------
| Table 1-1: Comparison Table | ||||
|---|---|---|---|---|
| Design/TechnologicalCharacteristic | SubjectPhoenix Atherectomy PlusSystem | PredicateBoston Scientific (PathwayMedical)Jetstream SystemK130637 | ReferencePhoenix Atherectomy SystemK172386 | ReferencePenumbraIndigo Aspiration SystemK161523 |
| Indications /Intended Use | The Phoenix AtherectomySystem is intended for usein atherectomy of theperipheral vasculature. Thesystem is not intended foruse in the coronary, carotid,iliac, pulmonary, or renalvasculature.Phoenix Atherectomy PlusSystem:When used with thePhoenix Aspiration Pumpas the vacuumsource, the Phoenix 2.4mmDeflecting AtherectomySystem is indicated for theremoval of thrombus fromvessels of the peripheralarterial vasculature. | The Jetstream System isintended for use in atherectomyof the peripheral vasculatureand to break apart and removethrombus from upper and lowerextremity peripheral arteries. Itis not intended for use incoronary, carotid, iliac or renalvasculature. | Intended for use in atherectomyof the peripheral vasculature. Itis not intended for use incoronary, carotid, iliac or renalvasculature. | INDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators are indicatedfor the removal of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venoussystems.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO SterileAspiration Tubing is indicated toconnect the INDIGO AspirationCatheter to the Penumbra AspirationPump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump isindicated as a vacuum source for thePenumbra Aspiration Systems. |
| Table 1-1: Comparison Table | ||||
| Design/TechnologicalCharacteristic | SubjectPhoenix Atherectomy PlusSystem | PredicateBoston Scientific (PathwayMedical)Jetstream SystemK130637 | ReferencePhoenix Atherectomy SystemK172386 | ReferencePenumbraIndigo Aspiration SystemK161523 |
| GeneralTechnologicalDesign | Phoenix Catheter with distalrotating cutter head cutsthrough target lesion, wherethe cutter geometry (helicalflutes) draws shavedmaterial proximally withinthe catheter housing as itcuts. The excised material isremoved continuously fromthe patient by mechanicalconveyance along withvacuum assist from theAspiration Pump anddeposited in a collectionbag outside the patient.Catheter deflection allowsthe catheter to cutdirectionally to a diameterlarger than the catheter.Aspiration Pump - handheldbattery powered pumpcapable of providing avacuum source of ≥ 25inchMercury (Hg) to aspirateblood and thrombus relatedmaterials. | Atherectomy via rotating distaltip advanced longitudinallythrough target lesion. The distaltip ("blades down") has cuttingflutes to remove lesion materialwhen rotating at the minimumdiameter. Simultaneousirrigation/infusion andcontinuous aspiration duringdevice activation removeexcised debris to a collectionbag outside the patient.In certain Jetstream models,rotational direction is reversedto deploy expandable blades("blades up") to achieve anincreased diameter for furtherlesion treatment. | Phoenix Catheter with distalrotating cutter cuts throughtarget lesion. The cuttergeometry (helical flutes) drawsshaved plaque proximallywithin the catheter housing as itcuts. Excised debris is removedcontinuously from the patientby mechanical conveyance anddeposited in a collection bagoutside the patient.Catheter deflection allows thecatheter to cut directionally to adiameter larger than thecatheter. | The Penumbra Aspiration System isdesigned to remove thrombus fromvasculature by aspirating the proximalside of the thrombus using continuousaspiration through the AspirationCatheter from the pump directly to thethrombus.For the aspiration source, theAspiration Catheter is used inconjunction with the Aspiration Pump,which is connected using theAspiration Tubing and the PumpCanister/Tubing.The Aspiration Catheter is used withthe Aspiration Pump to aspiratethrombus from an occluded vessel. Asneeded, the Separator may bedeployed from the Aspiration Catheterto assist with thrombus removal. TheSeparator is advanced and retractedthrough the Aspiration Catheter at theproximal margin of the primaryocclusion to facilitate clearing of thethrombus from the AspirationCatheter tip. |
| Table 1-1: Comparison Table | ||||
| Design/TechnologicalCharacteristic | SubjectPhoenix Atherectomy PlusSystem | PredicateBoston Scientific (PathwayMedical)Jetstream SystemK130637 | ReferencePhoenix Atherectomy SystemK172386 | ReferencePenumbraIndigo Aspiration SystemK161523 |
| DebulkingMechanism | Front Cutting RotationalDevice | Front Cutting RotationalDevice | Front Cutting RotationalDevice | Vacuum aspiration only |
| DebrisCollection &Removal | Active aspirationContinuous collection andremoval of thrombus bymechanical conveyance ofCatheter and vacuumprovided by aspirationpump | Active aspirationContinuous collection andremoval of excised debris viavacuum aspiration. | Continuous collection andremoval of excised debris bymechanical conveyance ofCatheter. | Continuous aspiration and removal ofthrombus via vacuum aspiration withthe catheter targeting aspiration fromthe pump to the thrombus. |
| CatheterRotationalSpeed | 10,000-12,000 RPM | ~70,000 RPM (atherectomy)~24,000 RPM (guidewireremoval) | 10,000-12,000 RPM | Not applicable |
| VacuumAspiration | ≥ 25 inch Hg Vacuum(battery powered.diaphragm pump) | > 20 inch Hg Vacuum(electrical, peristaltic pump) | Not applicable | Up to 29 inch Hg Vacuum(electrical, piston vacuum pump) |
| Power Source | 6V DC Battery for PhoenixHandle12V DC Battery forAspiration Pump | 110/220 V AC, 50/60 Hz ACpower supply from wall socket | 6V DC Battery for PhoenixHandleNot applicable for aspiration | 100-115 V AC/ 230 V AC, 50/60 HzAC power supply from wall socket forPenumbra Max Pump |
| GuidewireExchange | Over-the-wire | Over-the-wire | Over-the-wire | Over-the-wire |
| GuidewireCompatibility | 0.014" | 0.014" | 0.014" | 0.014"-0.035" |
| SheathCompatibility | 7F | 7F | 7F | 6F-8F |
| CatheterWorkingLength (s) | 125 - 127 cm | 120cm - 145cm | 125 - 127 cm | 85cm - 120 cm |
| Table 1-1: Comparison Table | ||||
| Design/TechnologicalCharacteristic | SubjectPhoenix Atherectomy PlusSystem | PredicateBoston Scientific (PathwayMedical)Jetstream SystemK130637 | ReferencePhoenix Atherectomy SystemK172386 | ReferencePenumbraIndigo Aspiration SystemK161523 |
| Tip Diameter orCrossing Profile | 2.4mm (tip diameter andcrossing diameter) | 1.6mm – 2.4mm ("bladesdown" range, tip diameter)3.0mm - 3.4mm ("blades up"range, tip diameter) | 2.4mm (tip diameter andcrossing diameter) | 2.2mm - 3.0mm |
| Minimum VesselSize for DeviceUse | ≥ 3.0mm | ≥2.5mm (XC)≥ 3.0mm (SC) | ≥ 3.0mm | ≥ 4.0mm |
| Deflectionmechanism | Deflecting Catheter:Advancing slider forwardon catheter chassis causesdistal tip to deflect | No deflection mechanism.However, certain Jetstreammodels allow for a largerdiameter cutter ("blades up")by reversing rotational directionto deploy expandable blades. | Deflecting Catheter:Advancing slider forward oncatheter chassis causes distal tipto deflect | Not applicable8F device with Angled tip |
| Steering(Directional)mechanism | Rotation of knob on handlesteers distal tip and cutterby torqueing catheter shaft | Not applicable | Rotation of knob on handlesteers distal tip and cutter bytorqueing catheter shaft | Not applicable8F Catheter with Angled Tip |
| CatheterCoating | No | No | No | Yes |
| Biocompatibility | per ISO 10993-1 | per ISO 10993-1 | per ISO 10993-1 | per ISO 10993-1 |
| Cutting TipMaterial | DLC coated Stainless Steel | Stainless Steel | DLC coated Stainless Steel | Not applicable |
| Sterilization | Ethylene Oxide | Ethylene Oxide -(Catheter/PodComponent ) | Ethylene Oxide | Ethylene Oxide |
| Single-use only | Catheter and Handle: YesAspiration Pump: Yes | Catheter and Pod- YesConsole - Reusable/electrical | Catheter and Handle: Yes | Catheter: YesAspiration Pump: Reusable/electrical |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
Testing Summary:
Since there is no change to the currently marketed Phoenix 2.4 mm Deflecting Atherectomy Catheter and Handle with Wire Support Clip (K172386 and K151145) that is used as part of the subject Phoenix Atherectomy Plus System, previous testing for biocompatibility, bench performance, sterilization, packaging and shelf-life, electrical safety and EMC, remain applicable and was not repeated.
To demonstrate the substantial equivalence of the subject Phoenix Atherectomy Plus System to the selected predicate device, the performance and technological characteristics were evaluated by completion of the following tests:
- Visual Inspection
- . Vacuum and Leak Test
- Flow Rate Test .
- System Simulated Use Test ●
- Life Cycle Test ●
- Tubing Tensile Test ●
- System Comparative Testing in Simulated Lesion
- Pump Comparative Test
- Sterilization Validation ●
- Packaging and Shelf Life
- Electrical Safety and EMC ●
- . Preclinical Animal Testing
Conclusion
The subject Phoenix Atherectomy Plus System shares the same intended use, technological characteristics, and principles of operation with the currently marketed predicate and reference devices. Additionally, the design of the 2.4mm Phoenix Deflecting Catheter and Handle components are identical to that of the currently marketed reference Phoenix Atherectomy System. The Phoenix Aspiration Pump component of the subject Phoenix Atherectomy Plus System has similar indications for use and technological characteristics as those of the reference Penumbra INDIGO Aspiration System. Any differences between the subject and predicate device were evaluated through design verification and validation testing which demonstrated no new questions of safety and effectiveness.
Based on the information submitted in this 510(k) premarket notification, the subject Phoenix 2.4mm Atherectomy Plus System is substantially equivalent to the currently marketed predicate device.
§ 870.4875 Intraluminal artery stripper.
(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).