(29 days)
INDIGO Aspiration Catheters and Separators: As part of the INDIGO™ Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
INDIGO Aspiration Tubing: As part of the INDIGO™ Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Pump MAX.
Penumbra Pump MAX: The Penumbra Pump MAX is indicated as a vacuum source for the Penumbra Aspiration Systems.
The INDIGO Aspiration System is comprised of several devices: INDIGO Aspiration Catheter, INDIGO SeparatorTM, INDIGO Aspiration Tubing, INDIGO Pump/Canister Tubing, Penumbra Aspiration Pump. The Aspiration Catheter is introduced through a guide catheter or long femoral sheath and into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used in parallel with the INDIGO Separator and an aspiration source to separate the embolus or thrombus and aspirate it from the occluded vessel. The INDIGO Separator is deployed from the INDIGO Aspiration Catheter. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion. This facilitates aspiration and debulking of the thrombus and reduces or eliminates the endovascular clot burden. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Pump/Canister Tubing. The INDIGO Separator is provided with an introducer and torque device. The INDIGO Aspiration Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The devices are visible under fluoroscopy.
The provided text describes the INDIGO™ Aspiration System and its components (INDIGO Aspiration Catheters, Separators, Tubing, and Penumbra Pump MAX). The acceptance criteria outlined and the study performance reported are primarily for the Separator 8 component of the system.
Here's the breakdown of the requested information based on the provided text:
Acceptance Criteria and Reported Device Performance
| Attribute | Specification | Acceptance Criteria | Reported Performance |
|---|---|---|---|
| Dimensional/Visual Inspection | Units meet all inspection criteria for release of finished goods (clinically acceptable) product. | Effectively, 100% compliance with inspection criteria | 100% Pass |
| Simulated Use (Peripheral Access, Vessel Access Entry Performance & Clot Removal) | Effectiveness of devices to remove clots and Aspiration Catheter does not collapse under vacuum when delivered through a tortuous anatomical model. | Effectively, 100% effectiveness/no collapse | 100% Pass |
| Separator Shape (for Subject Separator 8) | Must be diamond shaped | 100% Must meet Specification | 100% Pass |
| Advanced Separator 8 and Wire Joint Break Force | $\ge$ 0.67 lb | 100% Must meet Specification | 100% Pass |
| Core Wire PTFE Coating Integrity (after simulated use) | Coating has not delaminated, peeled, or flaked | 100% Must meet Specification | 100% Pass |
| Particulate Testing with Heparinized Saline (Separator) | Maximum number of particles: - $\ge$ 10 μm will be $\le$ 6000 particles - $\ge$ 25 μm will be $\le$ 600 particles. - $\ge$ 75 μm & $\ge$ 125 μm will be recorded. | 100% Must meet Specification | 10μm: 100% Pass 25μm: 100% Pass Data recorded for informational purposes only (for $\ge$ 75 μm & $\ge$ 125 μm) |
Study Details for Testing Subject Separator 8:
-
Sample Size used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify a numerical sample size for individual acceptance criteria tests. It generally refers to "units used in this Design Verification testing" for dimensional/visual inspection and "devices used for mechanical testing."
- Data Provenance: The studies are described as "bench-top testing" and were selected in accordance with EN ISO 10993-1 guidelines and conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. This indicates in-vitro testing rather than human or animal studies, and the data originates from internal verification testing. No country of origin is explicitly stated beyond the company's location (Alameda, CA, USA). It is retrospective in the sense that the testing was performed, and results compiled for this 510(k) submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The testing described is primarily mechanical and physical in nature, with "clinically acceptable product" being a criterion for visual inspection, implying expert judgment, but no details on who made these judgments or their qualifications are given.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This information is not provided. The assessment appears to be based on direct measurement or observation against predetermined specifications, not multi-reader adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This document describes the 510(k) submission for a physical medical device (aspiration system/catheter components), not an AI/software device that assists human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, a standalone (algorithm only) performance study was not done. The device is a physical medical device, not a software algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the physical and mechanical tests, the "ground truth" is defined by pre-determined specifications and international standards (e.g., ISO 10555-1) and internal design requirements. For example, for "Separator Shape," the ground truth is "diamond shaped." For "Break Force," the ground truth is "$\ge$ 0.67 lb." For "Simulated Use," it is the device's demonstrated ability to remove clots and not collapse.
-
The sample size for the training set:
- This concept is not applicable here. This is a physical medical device. There is no training set in the context of an algorithm or AI.
-
How the ground truth for the training set was established:
- This question is not applicable as there is no training set for a physical medical device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 1, 2016
Penumbra, Inc. Ms. Aditi Kolla Regulatory Affairs Manager One Penumbra Place Alameda. CA 94502
Re: K161523
Trade/Device Name: INDIGO™ Aspiration System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: June 1, 2016 Received: June 2, 2016
Dear Ms. Kolla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K161523
Device Name INDIGO™ Aspiration System
Indications for Use (Describe)
INDIGO Aspiration Catheters and Separators:
As part of the INDIGO™ Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
INDIGO Aspiration Tubing:
As part of the INDIGO™ Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Pump MAX.
Penumbra Pump MAX: The Penumbra Pump MAX is indicated as a vacuum source for the Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the subject Separator 8.
1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
1.2 Sponsor Contact Information
Aditi Kolla Regulatory Affairs Specialist Phone: (510) 995-2010 Fax: (510) 217-6414 Email: akolla@penumbrainc.com
1.3 Date of Preparation of 510(k) Summary
June 24, 2016
Device Trade or Proprietary Name 1.4
INDIGO™ Aspiration System
Device Classification 1.5
Regulatory Class: II Classification Panel: Cardiovascular Classification Name: Catheter, Embolectomy Regulation Number: 21 CFR 870.5150 Product Code: DXE
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside. The logo appears to be for a company or organization called Penumbra.
Predicate and Reference Devices 1.6
| Predicate Device | |||
|---|---|---|---|
| 510(k) Number | Clearance Date | Name of Predicate Device | Name ofManufacturer |
| K160533 | May 24, 2016 | Separator 8 | Penumbra, Inc. |
| Reference Devices | |||
| 510(k) Number | Clearance Date | Name of Predicate Device | Name ofManufacturer |
| K100826 | 13 July 2010 | PX 400 Microcatheter | Penumbra, Inc. |
| K082290 | 31 October 2008 | Neuron 6F DeliveryCatheters 053 and 070 | Penumbra, Inc. |
1.7 Predicate Comparison
| Common Device Name | Penumbra Embolectomy Aspiration System | |
|---|---|---|
| Trade Name | INDIGO™ Aspiration System | |
| 510(k) No. | K160533 | K161523 |
| Classification | Class II, DXE | Class II, DXE |
| Intended Use | INDIGO Aspiration Cathetersand SeparatorsAs part of the PenumbraEmbolectomy Aspiration System(INDIGO™ Aspiration System),the INDIGO Aspiration Cathetersand Separators are indicated forthe removal of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venoussystems.INDIGO Aspiration TubingAs part of the PenumbraEmbolectomy Aspiration System(INDIGO™ Aspiration System),the INDIGO Sterile AspirationTubing is indicated to connect theINDIGO Aspiration Catheters tothe Penumbra Pump MAX.Penumbra Pump MAXThe Penumbra Pump MAX isindicated as a vacuum source forthe Penumbra AspirationSystems. | Same |
| SeparatorMaterials | Separator 8 | Subject Separator 8 (90 cm, 150 cm) |
| Core Wire | Nitinol | Same |
| Coating | PTFE | Same |
| Distal External Coil | Platinum Alloy (92%Pt/8%W) | Same |
| Tip External Coil | Platinum Alloy (92%Pt/8%W) | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font and is the main focus of the image. The red circle with the white "P" is located to the right of the word "Penumbra" and serves as a logo for the company.
| Common Device Name | Penumbra Embolectomy Aspiration System | |
|---|---|---|
| Trade Name | INDIGO™ Aspiration System | |
| 510(k) No. | K160533 | K161523 |
| Proximal Coil | Stainless Steel 304 | Pebax 63D - colorant (green for 90 cmlength/orange for 150 cm length) |
| Solder Joint | Gold Solder [Au/Sn] & SilverSolder [SN/Ag] | Same |
| Separator Cone Tip | Pebax 40D, Transparent Blue(or)Pebax 35D with 40% AlldyneTungsten and 5% Titanium Oxide(TiO2) | Same |
| Dimensions | ||
| Total Length | 150 cm | 90 cm, 150 cm |
| Working Length | 96cm | 50cm, 96cm |
| Flex Length | 30 cm | 20 cm |
| Cone Shape | teardrop | diamond |
| Distal Cone OD | 0.068" (1.73mm) | 0.072" (1.83mm) |
| Center Section OD | 0.021" (0.533mm) | Same |
| Cone Length | 5mm | 7mm |
| Accessories | Torque Device & IntroducerSheath | Same |
| Packaging Materials& Configuration | Commonly utilized forinterventional devices | Same |
| Aspiration Source | Aspiration Pump | Same |
| Sterilization | EO | Same |
| Shelf-Life | 36 Months | Same |
Device Description 1.8
The INDIGO Aspiration System is comprised of several devices:
- INDIGO Aspiration Catheter
- INDIGO SeparatorTM
- INDIGO Aspiration Tubing ●
- INDIGO Pump/Canister Tubing .
- Penumbra Aspiration Pump
The Aspiration Catheter is introduced through a guide catheter or long femoral sheath and into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used in parallel with the INDIGO Separator and an aspiration source to separate the embolus or thrombus and aspirate it from the occluded vessel. The INDIGO Separator is deployed from the INDIGO Aspiration Catheter. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion. This facilitates aspiration and debulking of the thrombus and reduces or eliminates the endovascular clot burden. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Aspiration Pump, which is connected using the INDIGO
{6}------------------------------------------------
Aspiration Tubing and the INDIGO Pump/Canister Tubing. The INDIGO Separator is provided with an introducer and torque device. The INDIGO Aspiration Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The devices are visible under fluoroscopy.
1.9 Intended Use
INDIGO Aspiration Catheters and Separators
As part of the INDIGO™ Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
INDIGO Aspiration Tubing
As part of the INDIGO™ Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Pump MAX.
Penumbra Pump MAX
The Penumbra Pump MAX is indicated as a vacuum source for the Penumbra Aspiration Systems.
1.10 Summary of Non-Clinical Data
Included in this section is a description of the testing, which substantiates the substantial equivalence of subject Separator 8 to the predicate device:
- Design Verification (Bench-Top Testing) .
- Shelf-life .
The subject Separator 8 met all predetermined requirements.
1.10.1 Biocompatibility Testing
Biocompatibility testing previously performed on the predicate and reference devices listed in Section 1.6 substantiate the biocompatibility of the subject Separator 8. Studies were selected in accordance with EN ISO 10993-1 guidelines (Biological Evaluation of Medical Devices). All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices.
{7}------------------------------------------------
1.10.2 Bench-top Testing
Testing was based on the design specifications, risk analysis and available guidance documents. These guidance documents include:
- Guidance on Premarket Notification [510(k)] Submission for Short-Term and . Long-Term Intravascular Catheters (FDA - 1995)
- EN ISO 10555-1:2013 Sterile, single-use intravascular catheters Part 1: ● General Requirements
The physical and mechanical properties of subject Separator 8 devices were assessed using standard test methods and pre-determined acceptance criteria. Devices used for mechanical testing were assembled and packaged in the controlled production environment and sterilized twice using an ethylene oxide sterilization cycle. All established acceptance criteria were met. The summary of the testing performed on subject Separator 8 is listed below:
| Attribute | Specification | AcceptanceCriteria | Results |
|---|---|---|---|
| Dimensional/Visual Inspection | These evaluations confirm that the unitsused in this Design Verification testingmeet all inspection criteria for release offinished goods (clinically acceptable)product. | 100% Pass | |
| Simulated Use(Peripheral Access,Vessel Access EntryPerformance & ClotRemoval) | Simulated use testing of the AspirationCatheter and Separator was performed withaccessory devices in an anatomicalVascular Flow model which simulated thetortuosity of the peripheral vasculature.Devices were delivered through the tortuousanatomical model to evaluate theeffectiveness of the devices to remove clotsand that the Aspiration Catheter does notcollapse under vacuum. | 100% Pass | |
| Separator Shape | Must be diamondshaped | 100% MustmeetSpecification | 100% Pass |
| Advanced Separator8 and Wire JointBreak Force | $\ge$ 0.67 lb | 100% MustmeetSpecification | 100% Pass |
| Core Wire PTFECoating Integrity | Coating has notdelaminated,peeled, or flaked | 100% MustmeetSpecification | 100% Pass |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending beyond the circle.
| Attribute | Specification | AcceptanceCriteria | Results |
|---|---|---|---|
| after simulated use | |||
| Particulate Testingwith HeparinizedSaline (Separator) | The maximumnumber of particles:$\geq$ 10 μm will be $\leq$6000 particles$\geq$ 25 μm will be $\leq$600 particles. | 100% Must meetSpecification | 10μm100% Pass25μm100% Pass |
| $\geq$ 75 μm & $\geq$ 125μm will be recorded | Data was recorded for informationalpurposes only |
All testing met specification. The results of the tests appropriately address the physical and mechanical performance expectations of the device. Based on these overall results, the physical and mechanical properties of the subject Separator 8 devices are acceptable for the intended use and substantially equivalent to the predicate device.
1.11 Sterilization
The sterilization data from previous pre-market notification listed in Section 1.6 are directly applicable to the subject devices and no additional testing was required or performed. The subject devices were adopted into the existing validated Ethylene Oxide (EO) sterilization cycle as per TIR28, Annex A. Therefore, the subject Separator 8 has proved to be sterile in accordance with EN ISO 11135:2014.
1.12 Shelf-life
The subject Separator 8 has proved to have a 36 month shelf-life.
1.13 Summary of Substantial Equivalence
The subject Separator 8 device is substantially equivalent to the predicate devices with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.
§ 870.5150 Embolectomy catheter.
(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).