K Number
K181751
Manufacturer
Date Cleared
2019-02-01

(214 days)

Product Code
Regulation Number
864.5620
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HemoCue® Hb 801 System is intended for the quantitative determination of hemoglobin in capillary or venous whole blood (K2EDTA and Li-Heparin) in point-of-care settings. The HemoCue® Hb 801 System is intended to be used to determine the hemoglobin concentration for adults, adolescents, children, and infants above 1 month old. The HemoCue® Hb 801 System is for professional in vitro diagnostic use only.

Device Description

The HemoCue® Hb 801 System provides a direct reading of the hemoglobin concentration in a sample using specially designed, single use microcuvette and an analyzer. The system can be used by non-laboratory personnel.

The HemoCue® Hb 801 System consists of the following parts:

  • An analyzer supporting the following features: .
    • O Photometric determination of hemoglobin
    • Presentation of results on a display O
    • O Wired and wireless communication (USB and Bluetooth)
  • Power supply by power adapter, chargeable or non- chargeable batteries ●
  • Single use microcuvettes (test consumable)
  • Labeling: ●
    • O Operating Manual
    • o Package Insert
    • Quick reference Guide o
    • o Labels

The microcuvette serves both as a pipette and as a measuring cuvette. No dilution or other preparation of the blood sample is required before filling of the microcuvette. A whole blood sample of approximately 10 uL is drawn into the cavity in the microcuvette by capillary action.

The measurement takes place in the analyzer, which measures the absorbance of whole blood at an Hb/ HbO2 isosbestic point. The measurement is performed directly on the whole blood through measurement of the transmitted and scattered light and using an algorithm for translation into the hemoglobin concentration of the sample.

The HemoCue® Hb 801 System is traceable to the hemiglobincyanide (HiCN) method, the international reference method according to ICSH for the determination of the hemoglobin concentration in blood.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the HemoCue® Hb 801 System based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document largely focuses on demonstrating equivalence to a predicate device and established reference methods, with direct explicit acceptance criteria only mentioned for linearity and anticoagulants. Performance is reported in terms of precision, linearity, and correlation with reference methods.

Test CategoryAcceptance Criteria (Explicit or Implied)Reported Device Performance
Precision (Whole Blood)Not explicitly stated (implied to meet clinical needs and demonstrate consistency).Multi-microcuvette lots study (Single site): - Hb 2.0-3.0 g/dL: Mean 2.43 g/dL, SD 0.05 - Hb 6.0-7.0 g/dL: Mean 6.55 g/dL, SD 0.07 - Hb 9.5-10.5 g/dL: Mean 9.96 g/dL, CV 0.68% (Repeatability), 0.71% (Within Lab), 1.11% (Reproducibility) - Hb 13.5-14.5 g/dL: Mean 14.07 g/dL, CV 0.71% (Repeatability), 0.82% (Within Lab), 1.16% (Reproducibility) - Hb 16.5-17.0 g/dL: Mean 16.87 g/dL, CV 0.60% (Repeatability), 0.73% (Within Lab), 0.95% (Reproducibility) - Hb 23.0-24.0 g/dL: Mean 23.39 g/dL, CV 0.67% (Repeatability), 0.77% (Within Lab), 0.97% (Reproducibility) Overall precision (Multi-site): - Hb 2.0-3.0 g/dL: Mean 2.30 g/dL, SD 0.03 (R), 0.04 (WL), 0.05 (R) - Hb 6.0-7.0 g/dL: Mean 6.55 g/dL, SD 0.07 (R), 0.07 (WL), 0.08 (R) - Hb 9.5-10.5 g/dL: Mean 10.24 g/dL, CV 1.04% (R), 1.17% (WL), 1.63% (R) - Hb 13.5-14.5 g/dL: Mean 13.91 g/dL, CV 0.71% (R), 0.75% (WL), 1.00% (R) - Hb 16.5-17.0 g/dL: Mean 16.75 g/dL, CV 0.52% (R), 0.63% (WL), 0.66% (R) - Hb 23.0-24.0 g/dL: Mean 23.35 g/dL, CV 0.74% (R), 0.74% (WL), 0.89% (R)
Precision (Quality Control)SD and CV calculated for repeatability, between-run, between-day, and within laboratory precision for each level were within the defined acceptance criteria.Overall precision (Multi-site, multi-lots, operators, days): All reported SD and CV values (e.g., Low control: Mean 6.34 g/dL, SD 0.05 (R), 0.04 (WL), 0.06 (R)) were within the defined acceptance criteria.
Linearity"fulfilled acceptance criteria for the non-linear error" (for range 1.0-25.6 g/dL).System determined to be linear in the range 1.0-25.6 g/dL.
Detection Limit (LoB)Not explicitly stated (calculated to be 0.26 g/dL).LoB for the HemoCue® Hb 801 System was determined to be 0.26 g/dL.
Detection Limit (LoD)Not explicitly stated (calculated to be 0.3 g/dL).LoD for the HemoCue® Hb 801 System was determined to be 0.3 g/dL.
Quantification Limit (LoQ)Total Error (TE) for each sample ≤ 0.5 g/dL.LoQ for the HemoCue® Hb 801 System was determined to be 0.5 g/dL, meeting the TE goal.
Analytical SpecificityNo significant interference at tested concentrations.Most tested substances (Acetaminophen, Ascorbic acid, Creatinine, HbCO, HbO2, Hemolysis, Ibuprofen, MetHb, Platelets, Total protein, Salicylic acid, Simvastatin, Tetracycline, Triglycerides, Urea, Uric acid, Warfarin, Normal blood pH) showed no interference at respective concentrations. Interference observed with: - Conjugated bilirubin (>23 mg/dL at 10 g/dL Hb) - Unconjugated bilirubin (>12 mg/dL at 10 g/dL Hb, >23 mg/dL at 20 g/dL Hb) - Intralipid (>214 mg/dL at 10 g/dL Hb, >483 mg/dL at 20 g/dL Hb) - Leucocytes (>260 x 10^9/L at 6.8 – 14.7 g/dL Hb) (Note: "May give elevated results in higher substance concentrations".)
Anticoagulant EquivalenceMet the acceptance criteria regarding the correlation and bias between K2EDTA and Li-Heparin.Both K2EDTA and Lithium Heparin samples fulfilled the acceptance criteria.
Capillary vs. Venous SampleMet the defined acceptance criteria and showed equivalency.Both venous and capillary samples were within the defined acceptance criteria and showed equivalency.
Method Comparison (Predicate)Linear regression analysis demonstrated comparable performance.Venous (N=264): Slope 1.00, Intercept -0.14, r 1.00 against predicate. Capillary (N=233): Slope 1.07, Intercept -0.91, r 0.96 against predicate. Pediatric samples (N=71): - HemoCue® Hb 801 vs ICSH: slope 0.95, r 0.99. - HemoCue® Hb 801 vs HemoCue® Hb 301 System: slope 0.97, r 0.99. Both samples types were within defined acceptance criteria.
Reference Range VerificationValues fall within the expected pediatric and adult reference intervals.Study results verified that the reference ranges data on the HemoCue® Hb 801 System for the subgroups fall within the defined reference ranges (e.g., Infant 9.4-14.1 g/dL, Adult Male 13.0-17.0 g/dL).

2. Sample Size Used for the Test Set and Data Provenance

  • Precision (Whole Blood):

    • Multi-microcuvette lots study: 6 hemoglobin levels, 5 operating days, 5 replicates per run, 3 microcuvette lots. Total 75 measurements per level. (Implied N = 6 levels * 75 measurements = 450).
    • Multi-site study: 3 sites, 5 operating days, 5 replicates per run, 6 hemoglobin levels. Total 75 measurements per level. (Implied N = 6 levels * 75 measurements = 450).
    • Provenance: Not explicitly stated, but clinical studies for method comparison mention primary care settings in the US and one European clinical laboratory. It is likely these precision studies were conducted in similar clinical laboratory environments. Retrospective/Prospective not specified, but likely prospective.
  • Precision (Quality Control Material):

    • Sample size: Three sites, 20 operating days, 1 lot of control material (3 levels), duplicate runs twice daily. Total 240 measurements per level (80 per site). (Implied N = 3 levels * 240 measurements = 720).
    • Provenance: Not explicitly stated, but likely clinical laboratory settings, potentially those mentioned for method comparison (US/Europe). Likely prospective.
  • Linearity:

    • Sample size: 9 hemoglobin concentrations, 15 replicates per concentration (5 replicates/analyzer). (Implied N = 9 concentrations * 15 replicates = 135).
    • Provenance: Clinical laboratory, likely prospective.
  • Detection Limit (LoB, LoD, LoQ):

    • LoB: 4 individual plasma samples. 3 analyzers, 2 microcuvette lots, 4 operating days. 1 sample analyzed each day, 3 runs per day, 2 replicates/analyzer per run. Total 72 replicates/microcuvette lot (Implied N = 2 lots * 72 replicates = 144).
    • LoD: 4 K2EDTA whole blood samples from different subjects. 3 analyzers, 2 microcuvette lots, 4 operating days. 1 sample analyzed each day, 3 runs per day, 2 replicates/analyzer per run. Total 72 replicates/microcuvette lot (Implied N = 2 lots * 72 replicates = 144).
    • LoQ: 4 K2EDTA whole blood samples from different subjects. 3 analyzers, 2 microcuvette lots, 4 operating days. 1 sample analyzed each day, 3 runs per day, 3 replicates/analyzer per run. Total 108 replicates/microcuvette lot (Implied N = 2 lots * 108 replicates = 216).
    • Provenance: Clinical laboratory, likely prospective.
  • Analytical Specificity:

    • Sample size: K2EDTA whole blood samples with adjusted Hb levels. Number of samples/subjects not explicitly stated for each interferent, but tested at two Hb concentrations (10±0.5 and 20±1.0 g/dL).
    • Provenance: Clinical laboratory, likely prospective.
  • Anticoagulant Equivalence:

    • Sample size: 120 subjects provided both K2EDTA and Li-Heparin venous whole blood. Additional 11 subjects contributed samples to adjust Hb values.
    • Provenance: Two sites, likely clinical settings, likely prospective.
  • Capillary vs. Venous Sample Equivalence:

    • Sample size: 40 subjects for direct comparison, plus 212 subjects from the method comparison study (total 252).
    • Provenance: Not explicitly stated, but likely related to the multi-site method comparison study in the US. Likely prospective.
  • Method Comparison (Predicate):

    • Sample size:
      • US Study: 264 venous samples (28 contrived) and 233 capillary samples. Total 497.
      • European Clinical Lab: 71 pediatric samples.
    • Provenance:
      • US Study: Five primary care settings in the US.
      • European Clinical Lab Study: One European clinical laboratory site.
    • Retrospective/Prospective: Likely prospective.
  • Reference Range Verification:

    • Sample size: Whole blood specimens collected. Number of individual samples not specified.
    • Provenance: 5 point-of-care locations in the US. Likely prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • No information provided regarding experts establishing ground truth for the test set. Most studies compare the device against a predicate device or the International Council for Standardization in Haematology (ICSH) reference method, which represents a gold standard, not expert consensus.

4. Adjudication Method for the Test Set

  • Not applicable/Not mentioned. The studies described are analytical performance studies comparing the device to reference methods or a predicate, not studies involving human interpretation or adjudication of results. For the method comparison studies, duplicates were used for the predicate device, and triplicates for the ICSH reference method, implying a direct comparison of numerical results rather than an adjudication process.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. The studies are focused on the analytical performance of the device in measuring hemoglobin concentrations, primarily comparing it to a predicate device and a reference method. They do not evaluate human reader performance with or without AI assistance. The device is intended for non-laboratory personnel to use for direct numerical readings.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

  • Yes, a standalone performance study was done. All the analytical performance studies (precision, linearity, detection limits, analytical specificity, anticoagulant/sample type equivalence, and method comparison with predicate/ICSH) represent standalone performance of the HemoCue® Hb 801 System. The "system" includes the analyzer which uses an algorithm for translation into hemoglobin concentration. Users are described as "non-laboratory personnel" who obtain a direct reading from the device.

7. Type of Ground Truth Used

  • The ground truth varied depending on the study:
    • Precision, Linearity, Detection Limits, Analytical Specificity: These studies establish the inherent performance characteristics of the device itself. The "ground truth" for the samples was their known or carefully prepared hemoglobin concentrations as measured by highly controlled laboratory methods.
    • Anticoagulant and Capillary vs. Venous Equivalence: The "ground truth" was the comparison between the two sample types from the same subject.
    • Method Comparison: The ground truth was established by:
      • The predicate device (HemoCue® Hb 301 System).
      • The hemiglobincyanide (HiCN) method, which is the International Council for Standardization in Haematology (ICSH) international reference method.
    • Reference Range Verification: Comparison against published reference intervals from authoritative texts (Dacie and Lewis Practical Haematology, Pediatric Reference Intervals).

8. Sample Size for the Training Set

  • Not applicable/Not mentioned. This is an IVD device that measures a specific analyte using a spectrophotometric measuring principle with a pre-defined algorithm and factory calibration. There is no mention of a "training set" in the context of machine learning or AI algorithm development as typically understood in the context of image analysis or diagnostic prediction. The algorithm for translating light measurements into hemoglobin concentration is fundamental to the device's design, not something that appears to be "trained" on a large dataset in the sense of modern AI.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable/Not mentioned. As noted above, typical "training sets" and their associated ground truth establishment methods (e.g., expert consensus labeling) are not relevant to the description of this device's development or validation. The device's fundamental measurement principle and algorithm are traceable to the HiCN method (ICSH), meaning its design and underlying calculations are based on established scientific principles for hemoglobin determination.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

February 1, 2019

HemoCue AB Maria Fagerberg Director Regulatory Affairs Kuvettgatan 1 Angelholm, Sweden SE-26271

Re: K181751

Trade/Device Name: HemoCue Hb 801 System Regulation Number: 21 CFR 864.5620 Regulation Name: Automated hemoglobin system Regulatory Class: Class II Product Code: GKR Dated: June 27, 2018 Received: July 2, 2018

Dear Maria Fagerberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181751

Device Name HemoCue® Hb 801 System

Indications for Use (Describe)

The HemoCue® Hb 801 System is intended for the quantitative determination of hemoglobin in capillary or venous whole blood (K2EDTA and Li-Heparin) in point-of-care settings. The HemoCue® Hb 801 System is intended to be used to determine the hemoglobin concentration for adults, adolescents, children, and infants above 1 month old. The HemoCue® Hb 801 System is for professional in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Enclosure 1

510(k) Summary

{4}------------------------------------------------

1. Submitter and contact information

Submitter:

Company Name:HemoCue AB
ER Number:3003044483
Address:Kuvettgatan 1SE-262 71Sweden

Official contact person:

Name:Maria Fagerberg
Function:Director Regulatory Affairs
E-mail:maria.fagerberg@hemocue.se
Phone:+46 431 48 13 46
Fax:+46 431 48 12 25

Secondary contact:

Name:Andrijana Kankaras
Function:Regulatory Affairs Specialist
Phone:+46 431 48 12 26
Fax:+46 431 48 12 25
E-mail:andrijana.kankaras@hemocue.se

Date of preparation:

Date: June 27, 2018

2. Device information

Proprietary name:HemoCue® Hb 801 System
Common name:Hemoglobin analyzing system
Regulation:Automated hemoglobin system (21 CFR § 864.5620
Product code:GKR
Classification:Class II
Common name:Hemoglobin analyzing system
Panel:Hematology

{5}------------------------------------------------

3. Device description

The HemoCue® Hb 801 System provides a direct reading of the hemoglobin concentration in a sample using specially designed, single use microcuvette and an analyzer. The system can be used by non-laboratory personnel.

The HemoCue® Hb 801 System consists of the following parts:

  • An analyzer supporting the following features: .
    • O Photometric determination of hemoglobin
    • Presentation of results on a display O
    • O Wired and wireless communication (USB and Bluetooth)
  • Power supply by power adapter, chargeable or non- chargeable batteries ●
  • Single use microcuvettes (test consumable)
  • Labeling: ●
    • O Operating Manual
    • o Package Insert
    • Quick reference Guide o
    • o Labels

The microcuvette serves both as a pipette and as a measuring cuvette. No dilution or other preparation of the blood sample is required before filling of the microcuvette. A whole blood sample of approximately 10 uL is drawn into the cavity in the microcuvette by capillary action.

The measurement takes place in the analyzer, which measures the absorbance of whole blood at an Hb/ HbO2 isosbestic point. The measurement is performed directly on the whole blood through measurement of the transmitted and scattered light and using an algorithm for translation into the hemoglobin concentration of the sample.

The HemoCue® Hb 801 System is traceable to the hemiglobincyanide (HiCN) method, the international reference method according to ICSH for the determination of the hemoglobin concentration in blood.

4. Intended use and Indications for Use

Intended Use

The HemoCue® Hb 801 System is intended for the quantitative determination of hemoglobin in capillary or venous whole blood (K2EDTA and Li-Heparin) in point-of-care settings. The HemoCue® Hb 801 System is intended to be used to determine the hemoglobin concentration for adults, adolescents, children, and infants above 1 month old. The HemoCue® Hb 801 System is for professional in vitro diagnostic use only.

Indications for Use Same as intended use.

{6}------------------------------------------------

5. Substantial Equivalence

Candidate devicePredicate device
Proprietary name:HemoCue® Hb 801 SystemHemoCue® Hb 301 System
Common name:Hemoglobin analyzing systemHemoglobin analyzing system
510(k) number/K061047
Product codeGKRGKR
ClassificationClass IIClass II
Regulation21 CFR 864.5620Automated hemoglobin system21 CFR 864.5620Automated hemoglobin system
Classification PanelHematologyHematology

The most important similarities and differences between the HemoCue® Hb 801 System (Candidate Device) and HemoCue® Hb 301 System (Predicate Device) are listed in the table below.

CharacteristicHemoCue® Hb 801 System(Candidate device)HemoCue® Hb 301 System(Predicate device)
AnalyzerImage: HemoCue® Hb 801 System AnalyzerImage: HemoCue® Hb 301 System Analyzer
MicrocuvetteImage: HemoCue® Hb 801 System MicrocuvetteImage: HemoCue® Hb 301 System Microcuvette
CharacteristicHemoCue® Hb 801 System(Candidate device)HemoCue® Hb 301 System(Predicate device)
Similarities
Intended UseThe HemoCue® Hb 801 System isintended for the quantitativedetermination of hemoglobin incapillary or venous whole blood(K2EDTA and Li-Heparin) in point-of-care settings. The HemoCue® Hb801 System is intended to be used todetermine the hemoglobinconcentration for adults, adolescents,children, and infants above 1 monthold. The HemoCue® Hb 801 Systemis for professional in vitro diagnosticuse only.The HemoCue® Hb 301 System isdesigned for quantitative point-of-carewhole blood hemoglobin determinationin primary care using a speciallydesigned analyzer, the HemoCue Hb301 Analyzer, and specially designedmicrocuvettes, the HemoCue Hb 301Microcuvettes. The HemoCue Hb 301System is for In Vitro Diagnostic useonly. The HemoCue Hb 301 Analyzeris only to be used with HemoCue Hb301 Microcuvettes.
AnalyteHemoglobinSame
Sample preparation(pre-treatment)NoneSame
Sample volume10 μLSame
Measuring principleSpectrophotometricSame
ReagentNo active ingredientsSame
CalibrationThe system is traceable to thehemiglobincyanide (HiCN) method,according to ICSH. The system isfactory calibrated and needs nofurther calibration.Same
Quality controlInternal self-test (verifying analyzerperformance)Same
Differences
Sample typeCapillary or venous whole bloodCapillary, venous or arterial wholeblood
Measuring range1-25.6g/dL0-25.6 g/dL
Connectivity• Wireless Bluetooth LowEnergy (BLE)• USBSerial port
Microcuvette insertiontechnique"Slot in"Tray
CharacteristicHemoCue® Hb 801 SystemHemoCue® Hb 301 System
(Candidate device)(Predicate device)
Image: HemoCue® Hb 801 SystemImage: HemoCue® Hb 301 System
User interface● Display● Display
● Beeper● Beeper
● Two buttons● One button
● Status LED

{7}------------------------------------------------

{8}------------------------------------------------

6. Measuring principle

Spectrophotometric

7. Standard/Guidance Referenced

  • CLSI EP05-A3; Evaluation of Precision Performance of Qualitative Measurement Methods; . Approved Guideline - Third Edition.
  • CLSI EP06-A; Evaluation of the Linearity of Quantitative Measurement Procedures: A ● Statistical Approach; Approved Guideline.
  • CLSI EP07-A2: Interference Testing in Clinical Chemistry: Approved Guideline Second . Edition.
  • CLSI EP07- Interference Testing in Clinical Chemistry; Approved Guideline -Third Edition ●
  • CLSI EP37; Supplemental Tables for Interference Testing in Clinical Chemistry First Edition ●
  • CLSI EP09-A3; Measurement procedure comparison and Bias Estimation Using Patient Samples; Approved Guideline - Third Edition.
  • CLSI EP17-A2; Evaluation of Detection Capability for Clinical Laboratory Measurement ● Procedures; Approved Guideline - Second Edition.
  • CLSI EP25-A; Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. ●
  • CLSI EP28-A3c; Defining, Establishing, and Verifying Reference Intervals in the Clinical ● Laboratory; Approved Guideline - Third Edition.

{9}------------------------------------------------

  • . CLSI H15-A3 Reference and Selected Procedures for the Quantitative Determination of Hemoglobin in Blood; Approved Standard - Third Edition.
  • IEC 61010-1: 2010 (Third Edition) + AMD1:2016, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements.
  • IEC 61010-2-101:2015 (Second Edition), Safety requirements for electrical equipment for ● measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.
  • IEC 61326-1:2012 (Second Edition), Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements.
  • IEC 61326-2-6:2012 (Second Edition), Electrical equipment for measurement, control and ● laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic medical equipment.
  • Electromagnetic Compatibility according to FCC CFR 47 Part 15 ●
  • IEC 62304:2006 (First Edition), Medical device software - Software life cycle processes
  • ISO 14971:2007 (Second Edition), Medical devices Application of risk management to ● medical devices

8. Performance Characteristics (if/when applicable)

    1. Analytical performance
  • a. Precision Whole Blood

Whole blood precision study was conducted with the purpose to verify repeatability, reproducibility, within-laboratory and between lot and site precision of the HemoCue® Hb 801 System by using venous K2EDTA whole blood samples. This study was divided in two different parts, Part 1 a multi-microcuvette lots and Part 2 a multi-site test to evaluate repeatability, within laboratory precision and reproducibility.

Multi-microcuvette study was performed at a single site over 5 operating day. Six hemoglobin levels were analyzed with 5 separate runs testing 5 replicates with three lots of HemoCue® Hb 801 Microcuvette and one HemoCue® Hb 801 Analyzers. Each blood sample was analyzed with 25 replicates per microcuvette lot, providing 75 measurements for each level.

Multi-site study was performed at 3 sites during 5 operating days. Six hemoglobin levels were analyzed with five separate runs testing five replicates at each site using three HemoCue® Hb 801 Analyzers and one lot of HemoCue® Hb 801 Microcuvette. Each blood sample was analyzed with 25 replicates using one analyzer per site, providing 75 measurements for each level.

{10}------------------------------------------------

Hb level(g/dL)Mean(g/dL)SD (g/dL)CV (%)SD (g/dL)CV (%)SD (g/dL)CV (%)
2.0-3.02.430.05-0.05-0.05-
6.0-7.06.550.07-0.08-0.08-
9.5-10.59.96-0.68-0.71-1.11
13.5-14.514.07-0.71-0.82-1.16
16.5-17.016.87-0.60-0.73-0.95
23.0-24.023.39-0.67-0.77-0.97

all precision based on all HemoCue® Hh 801 microcuvette lots

Table 2: Overall precision based on all sites

Hb level(g/dL)Mean(g/dL)RepeatabilityWithin Lab PrecisionReproducibility
SD (g/dL)CV (%)SD (g/dL)CV (%)SD (g/dL)CV (%)
2.0-3.02.300.03-0.04-0.05-
6.0-7.06.550.07-0.07-0.08-
9.5-10.510.24-1.04-1.17-1.63
13.5-14.513.91-0.71-0.75-1.00
16.5-17.016.75-0.52-0.63-0.66
23.0-24.023.35-0.74-0.74-0.89

b. Precision - Quality Control Material

A precision study was conducted at three sites over 20 operating days using three HemoCue® Hb 801 Microcuvette lots (one lot per site), three HemoCue® Hb 801 Analyzers (one analyzer per site), and 1 lot of control material (level 1= low, level 2= medium, level 3= high). Each control set was run in duplicate twice daily for 20 days, by 10 operators (minimum 2 operators per site). A total of 240 measurements (80 per level) were generated at each site. SD and CV calculated for repeatability, between-run, between-day and within laboratory precision for each level were within the defined acceptance criteria. Also, SD and CV calculated for repeatability, within-laboratory precision and reproducibility combined for all sites per level were within the defined acceptance criteria.

Table 3: Overall precision, based on all sites, all microcuvette lots, all operators and all days per level
ControlLevelMeanValue(g/dL)NumberofresultsRepeatabilityWithin Lab/sitePrecisionReproducibility
SD (g/dL)CV (%)SD (g/dL)CV (%)SD (g/dL)CV (%)
Low6.342400.050.70.040.70.060.9
Medium11.502400.050.40.050.50.060.5
High15.362400.151.00.161.00.171.1

c. Linearity

A linearity study was conducted using hemoglobin concentration prepared from a single, venous whole blood sample. A total of 9 hemoglobin concentration (0.50, 3.78, 7.05, 10.33, 13.60, 16.88, 20.15, 23.43, 26.70 g/dL) spanning the claimed measuring range of the HemoCue® Hb 801 Analyzer (1.0-25.6 g/dL) were tested with 15 replicates and analyzed using three HemoCue® Hb 801 Analyzers (5 replicates/analyzer) and one lot of HemoCue® Hb 801 Microcuvette. The statistical analysis was performed separately for each analyzer using a polynomial regression analysis using first, second and third order fits. Statistical testing was used to determine the polynomial model with the best fit by examining the standard error of the regression, Syx.

{11}------------------------------------------------

The HemoCue® Hb 801 System is linear in the range 1.0-25.6 g/dL with fulfilled acceptance criteria for the non-linear error.

d. Shelf -life

HemoCue® Hb 801 Microcuvette shelf-life was determined by using 3 lots of HemoCue® Hb 801 Microcuvette. At every analysis run, the microcuvettes were tested with 3 venous K2EDTA whole blood samples from 3 different subjects at each Hb level (a total of 9 fresh blood samples per analysis run). The shelf life of the HemoCue® Hb 801 Microcuvette is determined to 6 months (at time of premarket notification, will be extended)

e. Detection limit

Limit of Blank (LoB)

To determine Limit of Blank (LoB), plasma samples were obtained from four individual whole blood samples. The study was conducted with 3 HemoCue® Hb 801 Analyzers and 2 lots of HemoCue® Hb 801 Microcuvette over 4 operating days. One sample was analyzed each day at 3 different runs separated by at least 2 hours between each run. Two replicates/analyzer were analyzed at each run, in total 72 replicates/microcuvette lot. The data from both microcuvette lots were combined and the Anderson Darling normality test showed a non-normal distribution of the measuring results. The rank position was calculated according to the non-parametric option, using a-error 0.05 (yields p =0.95). 72 measurements resulted in a rank position of 68.9. LoB for the HemoCue® Hb 801 System was determined to be 0.26 g/dL.

Limit of Detection (LoD)

To determine Limit of Detection (LoB), four K2EDTA whole blood samples were collected from different subjects. Four independent samples were prepared and the Hb-levels were within 1-5 times the LoB value and distributed within the range 0.46 - 1.0 g/dL. The study was conducted over 4 operating days using 3 HemoCue® Hb 801 Analyzers and 2 lots of HemoCue® Hb 801 Microcuvette. Four samples were analyzed where one sample was analyzed each day at 3 different runs with at least 2 hours between each run and with 2 replicates/analyzer and run, totally 72 replicates/microcuvette lot. The data has been evaluated and showed to have equal variance and LoD was established with a ß-error at 0.05. LoD for the HemoCue® Hb 801 System was determined to be 0.3 g/dL.

Limit of Quantification (LoQ)

To determine Limit of Quantification (LoQ), four K2EDTA whole blood samples were collected from different subjects. Four independent samples were prepared and the Hb-levels of the samples were slightly above LoD, within the range 0.40-0.52 g/dL. The study was conducted with 3 HemoCue® Hb 801 Analyzers and 2 lots of HemoCue® Hb 801 Microcuvette over 4 operating days. Samples were analyzed within the interval of LoD+0.4 g/L, where one sample was analyzed each day at 3 different runs with at least 2 hours between each run and with 3 replicates /analyzer and run, totally 108 replicates /microcuvette lot. Total Error for each sample was calculated by using the RMS-model. The calculated TE for each microcuvette lot was reviewed against the

{12}------------------------------------------------

defined goal for TE 8≤ 0.5 g/dL. LoQ for the HemoCue® Hb 801 System was determined to be 0.5 g/dL.

  • f. Analytical specificity
    The purpose of the interference studies was to evaluate the effect of potential interferents on the HemoCue® Hb 801 System.

The study was performed by adjusting Hb-levels of the K2EDTA whole blood samples collected. Each sample was divided into a reference group and a test group containing potential interferent at the test concentration listed below.

For potential interferences from leucocytes and platelets, and certain patient conditions (Sickle cells, Polycythemia vera, Thalassemia/hypochromia and Anemia) the interference was evaluated by using a comparative procedure between the HemoCue® Hb 801 System and ICSH as a reference method.

The following substances have been tested at hemoglobin concentrations 10±0.5 and 20±1.0 g/dL. No interference was found at following concentrations of the substances tested.

SubstanceTest concentrationUnit
Acetaminophen1324µmol/L
Ascorbic acid342µmol/L
Creatinine442µmol/L
HbCO10%
HbO2≤ 50%
Hemolysis10g/L
Ibuprofen2425µmol/L
MetHb*25%
Platelets2000x 109/L
Total protein15g/dL
Salicylic acid4.34mmol/L
Simvastatin49µmol/L
Tetracycline34µmol/L
Triglycerides1500mg/dL
Urea42.9mmol/L
Uric acid1.4mmol/L
Warfarin32.5µmol/L

*Multiple concentrations of MetHb up to 25% were tested and did not interfere with hemoglobin measurement at Hb-level 10±0..

Normal blood pH and above, up to 8, at Hb level 10±0.5 or 20±1.0 g/dL do not interfere with the system.

{13}------------------------------------------------

SubstanceConcentration (unit)Hb concentration (g/dL)Result
Conjugatedbilirubin>23 (mg/dL)10Interfering*
ConjugatedbilirubinUp to 40 (mg/dL)20Non-interfering
Unconjugatedbilirubin>12 (mg/dL)10Interfering*
Unconjugatedbilirubin>23 (mg/dL)20Interfering*
Intralipid>214 (mg/dL)10Interfering*
Intralipid>483 (mg/dL)20Interfering*
Leucocytes>260 x 109/L6.8 – 14.7Interfering*

The following substances have been tested to determine the interfering concentration.

  • May give elevated results in higher substance concentrations.
  • g. Within Measurement Equivalence Tests: Anticoagulants
    The purpose of this study was to demonstrate the equivalence between the anticoagulants K2EDTA and Li-Heparin on the HemoCue® Hb 801 System. The study was conducted at two sites, over 5 operating days at each site, using one HemoCue® Hb 801 Analyzer and one lot of HemoCue® Hb 801 Microcuvettes. Each out of 120 subjects provided one fresh venous K2EDTA and one fresh venous Li-Heparin whole blood samples. Additional samples were collected from 11 subjects to adjust Hb value to high and low Hb values. The regression analysis was performed by using replicate 1 from the sample in the Li-Heparin tube vs. replicate 1 from the sample from the K2EDTA tube.

Both the K2EDTA and Lithium Heparin samples fulfils the acceptance criteria regarding the correlation and bias between the two anticoagulants used on the HemoCue® 801 System.

  • h. Within Measurement Equivalence Tests: Capillary vs. Venous Sample
    The purpose of this study was to verify the equivalency of venous and capillary whole blood samples on the HemoCue® Hb 801 System. Capillary and venous whole blood samples were collected from 40 subjects, within an as wide range as possible. Additional data points from 212 subjects were taken from the method comparison study. The regression analysis was performed by using replicate from the venous sample vs. replicate from the capillary sample from the same subject. The venous K2EDTA blood sample was collected directly after collection of the capillary sample. Both the venous and capillary samples were within the defined acceptance criteria and showed equivalency when used on the HemoCue® Hb 801 System.

  • i. Assay cut-off
    N/A

{14}------------------------------------------------

    1. Method Comparison Study
  • a. Method comparison with predicate device

Method comparison studies were performed to evaluate the accuracy of the HemoCue® Hb 801 System in the hands of intended users in point-of-care facilities when testing capillary and/or venous blood from individuals 1 month of age and older. The studies were performed at five primary care settings in the US over 73 operating days.

The study included a total of 264 venous and 233 capillary whole blood samples. Both venous and capillary whole blood samples were tested at each site using the HemoCue® Hb 801 System and the predicate device, HemoCue® Hb 301 System (comparative method, in duplicate). The studies were performed on six HemoCue® Hb 801 Analyzers and four lots of HemoCue® Hb 801 Microcuvette by 13 operators. 6 % of the total number of samples (28/497 samples) were adjusted (contrived) in order to cover the lower and upper ends of the HemoCue® Hb 801 System's Analytical Measuring Range (AMR).

The results for all venous samples respectively all capillary samples are summarized in the table below. Linear regression analysis demonstrated comparable performance between the HemoCue® Hb 801 System and HemoCue® Hb 301 System.

Sample typeTotaln =Contrivedn =Slope(95% CI)Intercept(95% CI)r
K2EDTAVenous264281.00(0.99 to 1.01)-0.14(-0.26 to -0.03)1.00
Capillary233NA1.07(1.02 to 1.12)-0.91(-1.54 to -0.28)0.96

Table 4: Method comparison study overall summary, all sites combined

In addition, 71 pediatric samples were tested at one European clinical laboratory site over 7 operating days. Each blood sample was analyzed with 2 replicates on the HemoCue® Hb 801 System, with 2 replicates on the predicate device HemoCue® Hb 301 System, and with 3 replicates with the reference method ICSH.

Regression analysis demonstrated comparable performance between the HemoCue® Hb 801 System and ICSH and between the HemoCue® Hb 801 System and HemoCue® Hb 301 System.

In the linear regression analysis of the relationship between HemoCue® Hb 801 System and reference method ICSH, have a slope 0.95 and a correlation coefficient (r) 0.99.

In the linear regression analysis of the relationship between HemoCue® Hb 801 System and the predicate device HemoCue® Hb 301 System, have a slope 0.97 and a correlation coefficient (t) 0.99.

{15}------------------------------------------------

Both the venous and capillary samples were within the defined acceptance criteria. HemoCue® Hb 801 System provides a quantitative Hb measurement in both venous and capillary whole blood samples for adults and children above 1 month old and support use at point-of-care settings.

b. Matrix comparison

For Within Measurement Equivalence Tests on Anticoagulants and Capillary vs. Venous sample, refer to "Performance Characteristics" Section above.

3. Clinical studies

  • Clinical Sensitivity a. N/A
  • b. Clinical specificity N/A
  • Other clinical supportive data (when a. and b. are not applicable): C. N/A

4. Clinical cut-off

N/A

5. Reference range

Reference ranges were verified by testing whole blood specimens collected and tested at 5 point-ofcare locations in the US on the HemoCue® Hb 801 System. The values of the collected samples have been compared against the reference ranges to assess if the obtained values fall within the expected pediatric and adult reference intervals1.2.

The study results verified that the reference ranges data on the HemoCue® Hb 801 System for the subgroups fall within the defined reference ranges below.

Subject groupAgeHb, g/dL
Infant>1 month -2 years9.4 - 14.1
Child>2 - 12 years11.0 - 15.5
Adolescent>12 - 21 years10.9 - 15.1
Adult Male>21 years13.0-17.0
Adult Female>21 years12.0-15.0

Table 5: Defined reference range

1 Dacie and Lewis Practical Haematology, Elsevier Limited, 11th Edition, 2011 and references herein 2 Soldin, S. J. Pediatric Reference Intervals, AACC Press; 7th edition, 2011

{16}------------------------------------------------

6. Instrument name

HemoCue® Hb 801 System

7. System Description

  • a. Mode of operation
    The HemoCue® Hb 801 System provides a direct reading of the hemoglobin concentration in a sample using specially designed, single use microcuvette and an analyzer.

The microcuvette serves both as a pipette and as a measuring cuvette. No dilution or other preparation of the blood sample is required before filling of the microcuvette. A whole blood sample of approximately 10 uL is drawn into the cavity in the microcuvette by capillary action.

The measurement takes place in the analyzer, which measures the absorbance of whole blood at an Hb/ HbO2 isosbestic point. The measurement is performed directly on the whole blood through measurement of the transmitted and scattered light and using an algorithm for translation into the hemoglobin concentration of the sample.

The HemoCue® Hb 801 System is be traceable to the hemiglobincyanide (HiCN) method, the international reference method according to ICSH for the determination of the hemoglobin concentration in blood.

  • b. Software
    Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is "moderate" level of concern.

  • c. Specimen Identification
    There is no specimen identification function for the HemoCue® Hb 801 System.

  • d. Specimen Sampling and Handling
    A whole blood sample of approximately 10 uL is drawn into the cavity in the HemoCue® Hb 801 Microcuvette by capillary action. To perform a hemoglobin reading, a filled microcuvette is inserted into the microcuvette holder in the HemoCue® Hb 801 Analyzer.

  • e. Calibration
    The HemoCue® Hb 801 System is factory calibrated and needs no further calibration.

{17}------------------------------------------------

f. Quality Controls

The HemoCue® Hb 801 Analyzer has an internal quality control, a self-test. It automatically verifies that the optronic unit of the analyzer is working properly every time the analyzer is turned on, when the microcuvette holder is put back into place after removal, and every hour when in use.

If an external quality control is required by local or other regulations, commercially available controls recommended by HemoCue should be used.

8. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above

Cleaning and Disinfection Study

A cleaning and disinfection study was conducted to validate virucidal efficacy using the selected disinfectant with the recommended disinfection protocol. Super Sani-Cloth®Germicidal Disposable Wipe (EPA Registration No. 9480-4), a ready to use pre-saturated towelette, demonstrated complete inactivation of Duck Hepatitis B virus (surrogate for Human Hepatitis B virus) for all tested materials, following two-minutes exposure at room temperature.

A robustness study was performed to demonstrate that the HemoCue® Hb 801 System is robust to multiple cleaning and disinfection cycles by using Super Sani-Cloth® Germicidal Disposable Wipe.

9. Proposed Labeling

The labeling is sufficient and it satisfies the requirements of 21CFR Part 809.10.

10. Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 864.5620 Automated hemoglobin system.

(a)
Identification. An automated hemoglobin system is a fully automated or semi-automated device which may or may not be part of a larger system. The generic type of device consists of the reagents, calibrators, controls, and instrumentation used to determine the hemoglobin content of human blood.(b)
Classification. Class II (performance standards).