K Number
K181731
Device Name
MR Compatible Aspiration Kit
Date Cleared
2018-11-20

(144 days)

Product Code
Regulation Number
882.1480
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The MR Compatible Aspiration System is used for controlled aspiration of blood, cystic components of tumors, abscess, colloid cysts, and cerebrospinal fluid using a manual syringe during surgery of the Ventricle System or Cerebrum.
Device Description
The MCA Kit is an accessory to the cleared ClearPoint System (K171257). The MCA Kit consists of a MCA device guide, guide sheath, stylet, depth stops, cannula (with attached tube) and adapter. The MCA Kit is designed to aid a physician in the removal of specific tissue or fluid types from the brain during image guided surgery. It is MR safe, as all materials are non-metallic. The MCA Kit is comprised of single use only, sterile-packaged disposable components that are packaged together.
More Information

No
The summary describes a manual aspiration system and its mechanical compatibility and performance, with no mention of AI or ML.

Yes

This device is used for controlled aspiration of blood and other components, which is a direct medical intervention aimed at treating medical conditions. Therefore, it is a therapeutic device.

No

The device is an aspiration system used for controlled removal of various fluids and tissues from the brain during surgery, not for diagnosing conditions.

No

The device description explicitly lists physical components (device guide, guide sheath, stylet, depth stops, cannula, adapter) and describes testing related to their physical properties and performance (flow, leak, aspiration, compatibility with hardware). There is no mention of software as the primary or sole component.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens (like blood, urine, or tissue) taken from the human body outside of the body to provide information for diagnosis, monitoring, or screening.
  • Device Function: The MR Compatible Aspiration System is used during surgery to remove fluids and tissues directly from the patient's body (Ventricle System or Cerebrum) using a manual syringe. It is a surgical tool for aspiration, not a device for analyzing samples in a lab setting.

The description clearly indicates its use as a surgical accessory for aspiration during image-guided procedures, which falls outside the scope of IVD devices.

N/A

Intended Use / Indications for Use

The MR Compatible Aspiration System is used for controlled aspiration of blood, cystic components of tumors, abscess, colloid cysts, and cerebrospinal fluid using a manual syringe during surgery of the Ventricle System or Cerebrum.

Product codes

GWG

Device Description

The MCA Kit is an accessory to the cleared ClearPoint System (K171257). The MCA Kit consists of a MCA device guide, guide sheath, stylet, depth stops, cannula (with attached tube) and adapter. The MCA Kit is designed to aid a physician in the removal of specific tissue or fluid types from the brain during image guided surgery. It is MR safe, as all materials are non-metallic. The MCA Kit is comprised of single use only, sterile-packaged disposable components that are packaged together.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Ventricle System or Cerebrum

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Physician / during surgery

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Bench testing was performed to verify the MCA Kit's compatibility with the ClearPoint System, accuracy, and substantial equivalence to the predicate aspiration device.
Accuracy Testing: The company used the same ground truth procedure and test methods that were used in the accuracy testing provided in support of K142505 and K171257 for the ClearPoint System. The test set-up utilized only the scalp mount bases and SmartFrame XG, since the MCA Kit must be used with the SmartFrame XG to allow device guide insertion, and the scalp mount base is the most likely configuration to be used clinically. Placements were performed using the 12 Fr x 30 cm and 16 Fr x 30 cm MCA Kits, which represent the smallest and largest OD sizes; accordingly, the test results support the performance of all sizes of the MCA kit. Target depth was 125 mm from the base-mounting surface for all placements. Results from the company's bench accuracy tests demonstrated that the mean error was below 1.0mm, with the highest standard deviation being 0.75mm in the X direction and highest 99% upper confidence limit of 1.48mm in the X direction. The mean angular error was below 1°, with a standard deviation of 0.34° and 99% upper confidence limit of 0.72°. These values are below the 2mm and 2° accuracy limits for a stereotaxic device intended for general neurological use.

MCA-Kit Device Guide–SmartFrame XG Compatibility: Performance testing was completed to verify that the device guide and SmartFrame XG are compatible per pre-defined acceptance criteria. Testing included insertion and retention of the Device Guide and Adapter to the SmartFrame under tensile and torque loading, stability of the Guide Sheath-Stylet and Guide Sheath-Cannula under lateral loading when inserted in the SmartFrame, and accuracy with the SmartFrame. All acceptance criteria were met to demonstrate that the MCA Kit components are compatible with the cleared ClearPoint System SmartFrame XG.

Flow and Leak Testing: Testing was conducted to confirm that the device could aspirate at least 10cc of various materials simulating blood, clotted blood, cystic components of tumors, abcesses, colloid cysts, and cerebrospinal fluid without leaking or collapsing and at an acceptable flow rate. The tests demonstrated that the MCA Kit functions as intended and is substantially equivalent to the legally marketed predicate.

Aspiration Test: Testing was performed to confirm the device remains operable for a limited period of time at 29.0 inHg for the MCA Kit Cannulas Twelve Fr. and 16 Fr. Cannulas that have undergone worst-case conditioning were used for this test. Plugged cannulas were evaluated under high vacuum generated by a pump to create worst-case conditions. The change in device inner diameter (I.D.) was measured to confirm that the device's integrity was maintained during the aspiration. The device was also verified to remain undamaged after exposure to the conditions. All Cannulas tested were able to withstand collapsing, leaking, or failing in any other manner when experiencing a minimum aspiration pressure of 29.0 inHg. The measured change in the Rigid section I.D. of the Cannulas was 0.0% for the 12 Fr., and 0.8% for the 16 Fr. The flexible Tubing had a measured change in I.D. of 1.3% for the 12 Fr. and 2.9% for the 16 Fr. Cannulas. The device can operate for a limited period of time at an aspiration pressure of 29.0 inHg without failing.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K152699

Reference Device(s)

K171257, K142505

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 882.1480 Neurological endoscope.

(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

November 20, 2018

MRI Interventions, Inc. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004

Re: K181731

Trade/Device Name: MR Compatible Aspiration Kit Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: October 24, 2018 Received: October 24, 2018

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

John Marler -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K181731

Device Name MR Compatible Aspiration System

Indications for Use (Describe)

The MR Compatible Aspiration System is used for controlled aspiration of blood, cystic components of tumors, abscess, colloid cysts, and cerebrospinal fluid using a manual syringe during surgery of the Ventricle System or Cerebrum.

Type of Use (Select one or both, as applicable)

Transaction Not in 31 CFR 501.615(a) or (b)Service Transaction Not in 31 CFR 501.615(c)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

|× | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) SUMMARY MRI Interventions, Inc.'s MR Compatible Aspiration Kit

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Hogan Lovells, US LLP 555 Thirteenth Street, NW Washington, DC 20004 Phone: 202-637-5600 Fax: 202-637-5910 Contact Person: John J. Smith, M.D., J.D.

Date Prepared: November 20, 2018

Name of Device and Name/Address of Sponsor

MR Compatible Aspiration Kit ("MCA Kit")

MRI Interventions, Inc. 5 Musick Irvine, CA 92618

Common or Usual Name:

Neurological Endoscope

Classification 21 C.F.R. § 882.1480

Product Code GWG

Predicate Device Penumbra Apollo System (K152699)

Intended Use / Indications for Use

The MR Compatible Aspiration System is used for controlled aspiration of blood, cystic components of tumors, abscess, colloid cysts, and cerebrospinal fluid using a manual syringe during surgery of the Ventricle System or Cerebrum.

Device Description

The MCA Kit is an accessory to the cleared ClearPoint System (K171257). The MCA Kit consists of a MCA device guide, guide sheath, stylet, depth stops, cannula (with attached tube) and adapter. The MCA Kit is designed to aid a physician in the removal of specific tissue or fluid types from the brain during image guided surgery. It is MR safe, as all materials are non-metallic. The MCA Kit is comprised of single use only, sterile-packaged disposable components that are packaged together.

4

| Kit
Configuration | Guide Size
(Diameter) | Sheath
(Outer) | Cannula Size
(Outer Diameter) | Cannula Length
(Approximate) | Maximum Target
(mm) | Depth | |
|----------------------|--------------------------|-------------------|----------------------------------|---------------------------------|------------------------|-------|--|
| 15 Fr – Long | 15 Fr | 12 Fr | 12 Fr | 12.2" / 31 cm | 120 | | |
| 17 Fr – Long | 17 Fr | 14 Fr | 14 Fr | 12.2" / 31 cm | 120 | | |
| 19 Fr – Long | 19 Fr | 16 Fr | 16 Fr | 12.2" / 31 cm | 120 | | |
| 15 Fr – Short | 15 Fr | 12 Fr | 12 Fr | 9.4" / 24 cm | 70 | | |
| 17 Fr – Short | 17 Fr | 17 Fr | 14 Fr | 14 Fr | 9.4" / 24 cm | 70 | |
| 19 Fr – Short | 19 Fr | 19 Fr | 16 Fr | 16 Fr | 9.4" / 24 cm | 70 | |

The MCA Kit is available in the following configurations:

Performance Data

Bench testing was performed to verify the MCA Kit's compatibility with the ClearPoint System, accuracy, and substantial equivalence to the predicate aspiration device.

TestTest Method SummaryResults
Accuracy
TestingThe company used the same ground
truth procedure and test methods that
were used in the accuracy testing
provided in support of K142505 and
K171257 for the ClearPoint System.
The test set-up utilized only the scalp
mount bases and SmartFrame XG,
since the MCA Kit must be used with
the SmartFrame XG to allow device
guide insertion, and the scalp mount
base is the most likely configuration to
be used clinically. Placements were
performed using the 12 Fr x 30 cm and
16 Fr x 30 cm MCA Kits, which
represent the smallest and largest OD
sizes; accordingly, the test results
support the performance of all sizes of
the MCA kit. Target depth was 125
mm from the base-mounting surface
for all placements.Results from the company's bench
accuracy tests demonstrated that the
mean error was below 1.0mm, with the
highest standard deviation being
0.75mm in the X direction and highest
99% upper confidence limit of 1.48mm
in the X direction. The mean angular
error was below 1°, with a standard
deviation of 0.34° and 99% upper
confidence limit of 0.72°. These values
are below the 2mm and 2° accuracy
limits for a stereotaxic device intended
for general neurological use.
MCA-Kit Device
Guide–
SmartFrame
XG
CompatibilityPerformance testing was completed to
verify that the device guide and
SmartFrame XG are compatible per
pre-defined acceptance criteria.
Testing included insertion and
retention of the Device Guide and
Adapter to the SmartFrame under
tensile and torque loading, stability of
the Guide Sheath-Stylet and Guide
Sheath-Cannula under lateral loading
when inserted in the SmartFrame, and
accuracy with the SmartFrame.All acceptance criteria were met to
demonstrate that the MCA Kit
components are compatible with the
cleared ClearPoint System SmartFrame
XG.

5

| Flow and Leak
Testing | Testing was conducted to confirm that
the device could aspirate at least 10cc
of various materials simulating blood,
clotted blood, cystic components of
tumors, abcesses, colloid cysts, and
cerebrospinal fluid without leaking or
collapsing and at an acceptable flow
rate. | The tests demonstrated that the MCA
Kit functions as intended and is
substantially equivalent to the legally
marketed predicate |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aspiration Test | Testing was performed to confirm the
device remains operable for a limited
period of time at 29.0 inHg for the MCA
Kit Cannulas Twelve Fr. and 16 Fr.
Cannulas that have undergone worst-
case conditioning were used for this
test. Plugged cannulas were
evaluated under high vacuum
generated by a pump to create worst-
case conditions. The change in device
inner diameter (I.D.) was measured to
confirm that the device's integrity was
maintained during the aspiration. The
device was also verified to remain
undamaged after exposure to the
conditions. | All Cannulas tested were able to
withstand collapsing, leaking, or failing
in any other manner when experiencing
a minimum aspiration pressure of 29.0 inHg. The measured change in the
Rigid section I.D. of the Cannulas was
0.0% for the 12 Fr., and 0.8% for the 16
Fr. The flexible Tubing had a measured
change in I.D. of 1.3% for the 12 Fr. and
2.9% for the 16 Fr. Cannulas. The
device can operate for a limited period
of time at an aspiration pressure of 29.0 inHg without failing. |

The MCA Kit complies with the following recognized consensus standards:

  • . ANSI/AAMI/ISO 11137-2: Sterilization of Health Care Products – Radiation – Establishing the sterilization dose - Method VDmax
  • ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing within . a risk management process
  • ASTM F2096-11: Standard Test Method for Detecting Gross Leaks in Packaging by Internal ● Pressurization (Bubble Test)
  • ASTM D4196-16: Standard Test Method for Confirming the Sterility of Membrane Filters ●
  • . ASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

Substantial Equivalence

The Apollo System and MR Compatible Aspiration Kit have substantially similar intended use and indications for use. The table below presents a comparison of the technological characteristics of the MR Compatible Aspiration Kit and those of the predicate device.

| | Apollo System
K152699 | MR Compatible Aspiration Kit |
|---------------------------------|---------------------------------------------------------------|-------------------------------------------------------------------------------|
| Wand
(Cannula)
O.D / I.D. | 1 size:
Body = .083" / .069"
Distal Tip = .072" / .059" | 3 sizes:
12F = .158" / .138"
14F = .183" / .163"
16F = .208" / .190" |

6

| | Apollo System
K152699 | MR Compatible Aspiration Kit |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Usable Length | 27.2 cm | Maximum insertion length is 12.5cm due to height of SmartFrame |
| Basic
Operating
Principle | AC power is converted from the generator into vibrational energy to the distal tip of the wand, which removes tissue and/or fluids through and aspiration lumen | Manual actuation of the syringe plunger provides vacuum to the distal tip of the cannula which removes tissue and/or fluids through the aspiration lumen |
| Aspiration | 0–29 in.HG | 0–29 in.HG |
| Disposable
Components | Wand Collection Canister, filter and pump-canister tubing Irrigation Tubing | Cannula w/ tubing Stylet Guide Sheath Depth Stop SmartFrame Adapter and Device Guide |
| Power
Requirements | 100">–115V (60Hz),
100V (50Hz) | N/A
Manual Control |
| Navigation | Image Guidance | Image Guidance |

Conclusion

The MCA Kit and Apollo System have substantially similar intended use and indications for use, as well as similar technological characteristics. The minor technological differences in the MCA Kit compared to the Apollo System do not raise new questions of safety or effectiveness. The MCA Kit's accuracy, compatibility with the ClearPoint System SmartFrame XG, flow, leak, and aspiration testing further confirm that the device performs as intended and is substantially equivalent to the legally marketed predicate.