K Number
K171257
Date Cleared
2017-10-20

(175 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices. The user should consult the "Navigational Accuracy" section of the User's Guide to assess if the accuracy of the system is suitable for their needs.

Device Description

The ClearPoint System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid (previously cleared under K100836), the SMARTFrame™ MRI-Guided Trajectory Frame (K100836), the SMARTFrame™ Accessory Kit (K100836) and the SMARTFrame™ Thumbwheel Extension.

The SMARTGrid and associated Marking Tool are designed to assist the physician to precisely position the burr hole as called out in the trajectory planning software. The SMARTFrame is an adjustable trajectory frame that provides the guidance and fixation for neurosurgical tools. The MRI visible fluids of the Targeting Cannula along with the fiducial markers in the base of the frame allow for trajectory feedback when the physician views the MRI images, makes changes, and confirms with subsequent MR images.

The ClearPoint System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) that meet the physician's desired imaging quality. MRI Interventions also supplies an optional head fixation frame that can be used with the ClearPoint System.

The ClearPoint Workstation includes the following:

  1. ClearPoint Workstation Software (for trajectory planning and monitoring)
  2. Laptop Computer

The hardware components of the ClearPoint System are the SMARTFrame and Accessories. They are all single use devices that are provided sterile. They include the following:

  1. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole)
    a. Marking Grid
    b. Marking Tool
  2. SMARTFrame Pack
    a. SMARTFrame (adjustable trajectory frame ("ATF") to guide and hold the neurosurgical tools, includes fiducials, Targeting Cannula, & titanium screws)
    b. Centering Device
    c. Dock
    d. Device Lock (2 different diameters)
    e. Screwdriver
  3. Rescue Screws (Extra Titanium Screws)
  4. Thumbwheel Extension
  5. Accessory Pack
    a. Peel away sheath
    b. Stylet
    c. Lancet
    d. Depth Stop
    e. Ruler
  6. Scalp Mount Base
  7. Guide Tube and Device Guide Packs (Guide Cannulas)
AI/ML Overview

Here's an analysis of the provided text regarding the ClearPoint System, focusing on the acceptance criteria and the study that proves the device meets those criteria:


Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Targeting Accuracy (Linear)$\pm$ 1.5mm @ $\leq$ 125mm
Mean Error Across Device ConfigurationsBelow 1mm
Highest Standard Deviation (Linear)0.45mm
Highest 99% Confidence Limit (Linear)1.52mm
Angular AccuracyAll below 1°
Highest Standard Deviation (Angular)0.55°
Highest 99% Confidence Limit (Angular)1.40°
Stereotaxic Device LimitsBelow 2mm (linear error) and 2° (angular error)

Study Details Proving Device Meets Acceptance Criteria

The primary study mentioned to demonstrate device performance and suitability for new indications (explicitly including DBS lead placement) is a combination of:

  • Literature and registry data for clinical use.
  • Company's bench accuracy tests.

Here are the details extracted from the document:

2. Sample Size Used for the Test Set and Data Provenance:

  • Clinical Data: Data from 1,259 procedures were available for accuracy placement. 828 of these procedures were specifically for DBS lead placement.
  • Data Provenance: This was "literature and registry data," indicating retrospective collection from 35 institutions currently using the ClearPoint System. The country of origin is not explicitly stated but implies a multi-institutional, possibly international, or at least multi-state, context given the registry nature.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not explicitly stated for the clinical registry data. It's inferred that the "literature and registry data" would have involved physicians (likely neurosurgeons) performing the procedures and assessing outcomes, but the specific number and qualifications of experts for establishing ground truth on a per-case basis are not detailed.
  • For the bench tests, no expert ground truth establishment is mentioned, as these are objective engineering measurements.

4. Adjudication Method for the Test Set:

  • This information is not explicitly stated for the clinical registry data. Given it's registry data, it's unlikely a formal centralized adjudication method like 2+1 or 3+1 was used for each case's ground truth within the registry itself. The accuracy data likely came from post-operative imaging and physician assessment in individual clinical settings.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done. The document describes a study demonstrating device accuracy and clinical effectiveness based on existing use, not a comparison of human readers with and without AI assistance. The ClearPoint System is a stereotactic guidance system, not an AI-driven image interpretation tool for diagnostic purposes where MRMC studies are typical.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, a form of standalone performance was done for the device's accuracy. The "company's bench accuracy tests" represent a standalone evaluation of the system's mechanical and software targeting capabilities, independent of real-time human interpretation or assistance during a procedure. The results (mean error, standard deviation, 99% confidence limits for linear and angular accuracy) are standalone measurements of the device's precision.

7. The Type of Ground Truth Used:

  • Clinical Data: The ground truth for the clinical effectiveness (specifically placement accuracy) in the registry data would typically be derived from post-operative imaging analysis (e.g., MRI scans) to determine the actual final position of the placed instrument (e.g., DBS lead) relative to the planned target. This is effectively outcomes data related to the physical position of the device.
  • Bench Tests: The ground truth for the bench accuracy tests would be metrological measurements against a precisely known reference target, using engineering methods to quantify linear and angular deviations.

8. The Sample Size for the Training Set:

  • This information is not applicable/not provided in the context of this 510(k) summary. The ClearPoint System is a stereotactic guidance system. While its software (version 1.6.2) is mentioned, the document does not describe it as an AI/machine learning system that requires a "training set" in the conventional sense for image analysis or diagnostic prediction. The software likely relies on algorithms for trajectory planning, visualization, and guidance, which are developed and validated through engineering principles rather than data-driven machine learning training.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As explained in point 8, the concept of a "training set" and associated "ground truth" for machine learning is not described or relevant to the information provided for the ClearPoint System. The device's software is based on established computational geometry and image processing principles for stereotactic guidance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

October 20, 2017

MRI Interventions, Inc. % John Smith, MD JD Partner Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004

Re: K171257

Trade/Device Name: ClearPoint System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: ORR, HAW Dated: September 21, 2017 Received: September 21, 2017

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Hoffmann -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171257

Device Name ClearPoint System

Indications for Use (Describe)

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices. The user should consult the "Navigational Accuracy" section of the User's Guide to assess if the accuracy of the system is suitable for their needs.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

MRI Intervention, Inc.'s ClearPoint Device

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Hogan Lovells, US LLP 555 Thirteenth Street, NW Washington, DC 20004 Phone: 202-637-5600 Fax: 202-637-5910 Contact Person: John J. Smith, M.D., J.D.

Date Prepared: September 21, 2017

Name of Device and Name/Address of Sponsor

ClearPoint System™

MRI Interventions, Inc. 5 Musick Irvine, CA 92618

Common or Usual Name:

Neurological Stereotaxic Instrument

Classification

21 C.F.R. §882.4560

Product Code

ORR, HAW

Predicate Device

MRI Interventions, Inc. ClearPoint System (K160434)

Intended Use / Indications for Use

The ClearPoint System is intended to provide stereotactic guidance for the placement and operation of instruments or devices during planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion, including deep brain stimulation (DBS) lead placement. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners and MR Conditional implants and devices. The user should consult the "Navigational Accuracy" section of the User's Guide to assess if the accuracy of the system is suitable for their needs.

{4}------------------------------------------------

Technological Characteristics

The ClearPoint System is comprised of a workstation laptop with workstation software, the SMARTGrid™ MRI-Guided Planning Grid (previously cleared under K100836), the SMARTFrame™ MRI-Guided Trajectory Frame (K100836), the SMARTFrame™ Accessory Kit (K100836) and the SMARTFrame™ Thumbwheel Extension.

The SMARTGrid and associated Marking Tool are designed to assist the physician to precisely position the burr hole as called out in the trajectory planning software. The SMARTFrame is an adjustable trajectory frame that provides the guidance and fixation for neurosurgical tools. The MRI visible fluids of the Targeting Cannula along with the fiducial markers in the base of the frame allow for trajectory feedback when the physician views the MRI images, makes changes, and confirms with subsequent MR images.

The ClearPoint System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) that meet the physician's desired imaging quality. MRI Interventions also supplies an optional head fixation frame that can be used with the ClearPoint System.

The ClearPoint Workstation includes the following:

    1. ClearPoint Workstation Software (for trajectory planning and monitoring)
    1. Laptop Computer

The hardware components of the ClearPoint System are the SMARTFrame and Accessories. They are all single use devices that are provided sterile. They include the following:

    1. SMARTGrid Pack (interacts with the Software to determine the desired location of the burr hole)
    • a. Marking Grid
    • b. Marking Tool
    1. SMARTFrame Pack
  • a. SMARTFrame (adjustable trajectory frame ("ATF") to guide and hold the neurosurgical tools, includes fiducials, Targeting Cannula, & titanium screws)

  • b. Centering Device

  • c. Dock

  • d. Device Lock (2 different diameters)

  • e. Screwdriver

    1. Rescue Screws (Extra Titanium Screws)
    1. Thumbwheel Extension

{5}------------------------------------------------

    1. Accessory Pack
    • a. Peel away sheath
    • b. Stylet
    • c. Lancet
    • d. Depth Stop
    • e. Ruler
    1. Scalp Mount Base
    1. Guide Tube and Device Guide Packs (Guide Cannulas)

Each of the above packs is sold separately. Each is intended to be used with the ClearPoint Workstation. In addition, the ClearPoint System is used with the following separately cleared or Class I, 510(k)-exempt products:

  • SmartTip MRI Hand Drill and Drill Bit Kit (K122456/K151536) ●
  • MRI Neuro Procedure Drape, with Marker Pen and Cover (K091343)
  • SmartFrame MR Fiducial (510(k)-exempt Accessory) ●

The primary purpose of this 510(k) notice is to expand the indications of the device to explicitly include use with the positioning of deep brain stimulator ("DBS") leads. In addition, since the ClearPoint System's last clearance (K160434), an additional locking slot below the top slot in the SmartFrame XG Targeting Cannula ("TC") Support has been added to allow users to retract the TC while the removable support cap is in place. There are now three total locking slots on the TC Support. The software has also been updated to version 1.6.2.

The table below presents a detailed comparison of the technological characteristics of the ClearPoint System and those of the predicate device.

ClearPoint System (Predicate)K160434ClearPoint System (Subject)
Classification21 CFR 882.456021 CFR 882.4560
Product CodeORR, HAWORR, HAW
Intended UseThe ClearPoint System is intended toprovide stereotactic guidance andoperation of instruments or devicesduring the planning and operation ofneurological procedures within the MRIenvironment and in conjunction with MRimaging. The ClearPoint System isintended as an integral part of proceduresthat have traditionally used stereotacticmethodology. These procedures includebiopsies, catheter and electrode insertion.The System is intended for use only with1.5 and 3.0 Tesla MRI scanners.The ClearPoint System is intended toprovide stereotactic guidance andoperation of instruments or devices duringthe planning and operation of neurologicalprocedures within the MRI environment andin conjunction with MR imaging. TheClearPoint System is intended as anintegral part of procedures that havetraditionally used stereotactic methodology.These procedures include biopsies,catheter, and electrode insertion, includingdeep brain stimulation (DBS) leadplacement. The System is intended for useonly with 1.5 and 3.0 Tesla MRI scanners.The user should consult the "NavigationalAccuracy" section of the User's Guide toassess if the accuracy of the system issuitable for their needs.
ClearPoint System (Predicate)K160434ClearPoint System (Subject)
EnvironmentMRI SuiteMRI Suite
SterilizationEO 10-6 SALEO 10-6 SAL
SmartGridPackMRI Planning Grid & Marking ToolMRI Planning Grid & Marking Tool
SmartFramePackSmartFrame XG, Scalp Mount Base, BoneScrews, Scalp Mount Base, Screws,Stand-Off Pins, Screwdriver, CenteringTool, Wharen Centering Guide (packagedand sold separately), Dock and Lock, RollLock Screw with Washer, Rescue Screws(packaged separately)SmartFrame XG, Scalp Mount Base,Bone Screws, Scalp Mount Base,Screws, Stand-Off Pins, Screwdriver,Centering Tool, Wharen Centering Guide(packaged and sold separately), Dockand Lock, Roll Lock Screw with Washer,Rescue Screws (packaged separately)
Hand ControllerThumbwheel Extension (Light HandController)Thumbwheel Extension (Light HandController)
Accessory KitPeel-away Sheath, Lancet, Stylet, DepthStop, RulerPeel-away Sheath, Lancet, Stylet, DepthStop, Ruler
Drill Guides4.5mm & 6.0mm4.5mm & 6.0mm
TargetingCannula ID0.0825"0.0825"
TargetingCannulaMaterialUltem and PEEKUltem and PEEK
Guide Tube /Device Guide /Drill Guide ID0.0938, 0.141, 0.191, 0.250"0.0938, 0.141, 0.191, 0.250"
Guide Tube/Device Guide /Drill GuideMaterialUltemUltem
PackagingSterile, Sealed Tray, InsideSterile Tyvek Pouch(Wharen Centering Guide and Drill andDevice Guides are Sterile in a doubleTyvek Pouch Without a Tray; WharenCentering Guide Packaging includesPVC)Sterile, Sealed Tray, InsideSterile Tyvek Pouch(Wharen Centering Guide and Drill andDevice Guides are Sterile in a doubleTyvek Pouch Without a Tray; WharenCentering Guide Packaging includesPVC)
TargetingAccuracy$\pm$ 1.5mm @ $\leq$ 125mm$\pm$ 1.5mm @ $\leq$ 125mm
Software1.61.6.2

Summary of the Technological Characteristics of the Device Compared to the Predicate Device

{6}------------------------------------------------

Performance Data

Literature and registry data demonstrating that the ClearPoint System accurately positions DBS leads in more than 500 patients demonstrates safe and effective use for the new indications. The registry data included data from 35 institutions currently using the ClearPoint System in practice. Accuracy placement data were available on 1,259 procedures, 828 of which were for DBS lead placement.

{7}------------------------------------------------

Results from the company's bench accuracy tests demonstrated that the mean error across device configurations was below 1mm, with the highest standard deviation being 0.45mm and the highest 99% confidence limit being 1.52mm. Angular errors were all below 1°, with the highest standard deviation being 0.55° and the highest 99% confidence limit being 1.40°. These observed values are all below the 2mm and 2° accuracy limits for a stereotaxic device intended for general neurological use.

Substantial Equivalence

The ClearPoint System is as safe and effective as the predicate ClearPoint System (K160434). The ClearPoint System has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device. Thus the ClearPoint System that is the subject of this submission is substantially equivalent to the previously cleared ClearPoint System.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).