K Number
K152514
Manufacturer
Date Cleared
2015-11-18

(77 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Clickfine AutoProtect Pen Needle for use with pen injector devices for the injection of fluids, including insulin and exenatide.

Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.

Device Description

The pen needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of fluids. Each needle is individually packaged in a sealed protective container with a peel tab.

The pen needle is used by peeling back the peel tab and snapping or screwing the housing onto the threaded end of the pen-injector. The back end of the cannula punctures the septum of the drug reservoir in the pen-injector. The outer protective cap is then removed. When the injection is needed the needle is inserted into the chosen site. While inserting the needle into the skin, the safety shield glides into the housing enabling the needle to penetrate the skin barrier and the subcutaneous tissue. While the safety shield glides into the housing the safety mechanism is activated. The pen-injector delivers the medicinal product through the needle.

After the injection, in order to remove the needle from the skin, the user withdraws the pen injector from the skin. As the pen and needle is withdrawn from the skin, the safety shield glides back in its initial position, completely covering the needle, where it will remain locked. The safety shield is designed to automatically cover the needle (providing passive protection) to minimize the risk of accidental needle-stick injury. Once the Clickfine AutoProtect Pen Needle is in the locked mode, it can no longer be used. The red safety lock indicator tells the user that the safety lock has been activated. The needle is detached from the injection device and disposed of in accordance with local regulations. For each subsequent injection, another disposable needle must be used.

AI/ML Overview

This document describes a 510(k) premarket notification for the "Clickfine AutoProtect Pen Needle" (K152514). The purpose of this submission is to expand the range of needle lengths to include a 5 mm needle. The document references various ISO standards and internal protocols for performance testing to demonstrate substantial equivalence to the predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

TestRequirement (Acceptance Criteria)Reported Device Performance
Needle Assembly
Needle assembly attachment shall fit and function with needle-based injection systems specified in ISO 11608-1Clause 4.2.1 of ISO 11608-2: The pen needles were tested for compatibility with needle-based injection system as specified in ISO 11608-1.Pass
Tubing dimensionsClause 4.2.2, Table 1 of ISO 11608-2: Length of patient end: +/- 1.25 mm; Length of cartridge end: 5.7 – 7.0 mm.Pass
Flow rateClause 4.3 of ISO 11608-2: Measurement of flow rate in accordance with Annex A.Pass
Bond between hub and needle tubeClause 9 of ISO 11608-2 and clause 13.1 of ISO 7864: The union of the hub and needle tube shall not break for at least 5 sec. while a force of at least 22 N is applied.Pass
Needle pointsClause 4.5 of ISO 11608-2: Visually sharp at a 2.5x magnification; designed to minimize coring and fragmentation.Pass
LubricationClause 4.7 of ISO 11608-2: No visible droplets on the outside surface of the needle tube.Pass
Dislocation of measuring point at the patient-endClause 8 of ISO 11608-2: maximum allowable dislocation acc. Clause 4.8, Table 2 for 5mm: 0.65mm and for 8mm: 0.9mm.Pass
Functional compatibility with needle-based injection systemsClause 11 of ISO 11608-2: Needle assembly torque: 0.07 +/- 0.01 Nm; Needle hub removal: less than 0.100 Nm; Dose accuracy: for doses ≤ 20 ml the calculated values were within ± 0.01 ml of the targeted dose; for doses > 20 ml the calculated values were within ± 5 % of the targeted dose.Pass
Tubing Characteristics (ISO 9626/A1)
Tubing characteristicsTable 1 of ISO 9626/A1: The tubing is made of austenitic stainless steel acc. one of the designated types given in Table 1 of ISO 9626/A1.Pass
Tubing diametersTable 2 of ISO 9626/A1: Tubing dimensions meet outer and inner diameter requirements as defined in table 2 of ISO 9626/A1.Pass
StiffnessClause 9 of ISO 9626: When tested in accordance with annex C of ISO 9626 the tubing shall show a deflection not greater than the relevant value given in table 3 of ISO 9626/A1.Pass
Resistance to breakageClause 10 of ISO 9626: When tested in accordance with annex D of ISO 9626 and Table 4 of ISO 9626/A1 the tubing shall not break.Pass
Limits for acidity and alkalinityClause 6 of ISO 9626: Tested in accordance with annex A and extract preparation of the tubing in accordance with Annex B: correction for the volume of titrant required for the control fluid shall not be more than 0.04 ml NaOH or 0.12 ml HCl to reach the end-point titration.Pass
Resistance to corrosionClause 11 of ISO 9626: When tested in accordance with annex E, the immersed half of the tubing shall show no evidence of corrosion resulting from the test.Pass
Device Integrity (ISO 7864)
Freedom of defectsClause 11.3 in ISO 7864: When examined by normal or corrected-to-normal vision, the needle shall appear straight and of regular cross-section and wall thickness. (This requirement is listed twice in the document, here and below for ISO 7864)Pass
Limits for extractable metalsClause 6 of ISO 7864: When tested with a recognized method the content of lead, tin, zinc and iron shall be less than 5mg/l. The cadmium content shall be lower than 0.1 mg/l.Pass
Freedom of defects (re-listed)Clause 11.3 of ISO 7864: When examined by normal or corrected vision, the needle lube shall appear straight and of regular cross-section and wall thickness.Pass
Sharps Injury Protection (ISO 23908)
Activation of the sharps injury protectionIn accordance with ISO 23908 clause 5.2 the sharps injury protection feature shall be able to be activated by an appropriate force that eases actuation and avoids unintended actuation. An appropriate force was determined using a risk-based approach in accordance with ISO 14971.Pass
Verification of safety feature activationIn accordance with ISO 23908, clause 5.3, needles were tested to provide evidence that once the safety feature was activated, they resist axial and lateral forces so as to prevent unintended exposure to the sharps.Pass
Device design verification for needle tip protection after activationIn accordance to clause 5.4 of ISO 23908 it was demonstrated that a sphere of 6mm radius did not come into contact with the tip of the cannula, either before or immediately after triggering the safety mechanism.Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

The document does not explicitly state the sample sizes for each specific test or the country of origin of the data. However, it indicates that the testing was conducted as part of "Design Control Activities" by Ypsomed AG, which is a Swiss company. The testing is prospective in nature, as it is conducted to verify the performance of the modified device before it is marketed. The data provenance can be inferred as originating from Ypsomed AG's internal testing facilities, likely in Switzerland, to demonstrate compliance with international standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is a medical device, not an AI/imaging device requiring expert adjudication for ground truth of imaging data. The "ground truth" for these tests is established by objective measurements against predefined acceptance criteria from international standards (ISO). Therefore, no human experts were used to establish a "ground truth" in the way it would be for, for example, classifying medical images. The expertise lies in the engineers and technicians performing the tests and interpreting the results against the standard requirements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable in the context of this device. The testing involves objective measurements and functional assessments against standards, not subjective interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a physical medical device (pen needle), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device. The "performance" described is the device's mechanical and functional compliance with established safety and quality standards, not an algorithm's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The ground truth is based on established international ISO standards (e.g., ISO 11608-2, ISO 9626, ISO 7864, ISO 23908) which define specific, measurable requirements for physical and functional characteristics of hypodermic needles and sharps injury protection devices. These standards represent scientific and engineering consensus on acceptable device performance and safety.

8. The sample size for the training set:

Not applicable. This is a physical medical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

Not applicable. This is a physical medical device, not an AI model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2015

Ypsomed AG Mr. Stephan Affolter Head of Quality System & Regulatory Affairs Brunnmattstrasse 6, CH- 3401 Burgdorf SWITZERLAND

Re: K152514

Trade/Device Name: Clickfine AutoProtect Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: II Product Code: FMI Dated: October 20, 2015 Received: October 22, 2015

Dear Mr. Affolter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image shows a signature and the name "Tina Kiang". There is a large, looping signature to the left of the name. Below the name, there is a small "-S".

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152514

Device Name Clickfine AutoProtect Pen Needle

Indications for Use (Describe)

The Clickfine AutoProtect Pen Needle for use with pen injector devices for the injection of fluids, including insulin and exenatide.

Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

8 510(k) Summary for K152514

Submitted By/Contact Person:Stephan AffolterHead of Quality System & Regulatory AffairsYpsomed AGBrunnmattstrasse 6CH - 3401 BurgdorfSwitzerlandTel. 0041-34 424 3382Fax 0041-34 424 4122E-mail: stephan.affolter@ypsomed.com
Alternative Contact:Lee LeichterPresidentP/L Biomedical10882 Stonington AvenueFort Myers, FL 33913USATel. (239) 244-1448Fax. (815) 550-0162E-mail: leichter@plbiomedical.com
Date Prepared:September 23, 2015
8.1 Trade/Proprietary Name:Clickfine AutoProtect Pen Needle

Injection Pen Needle

8.3 Classification Name:

II 1 . . 1 1

8.3 Classification Name:

8.2 Common/Usual Name:

Hypodermic single lumen needle

8.4 Classification: Regulation 21 CFR Part 880.5570 - Hypodermic single lumen needle was published in the Federal Register on October 21, 1980 after review by the General Hospital and Personal Use Devices Classification, classifying Hypodermic single lumen needles in Class II.

Panel: 80 Procode: FMI

8.5 Predicate Device: K131123 Clickfine AutoProtect Pen Needle

8.6 Purpose of Submission: To widen the range of the needle lengths and include a 5 mm Clickfine AutoProtect pen needle.

{4}------------------------------------------------

8.7 Device Description

The pen needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of fluids. Each needle is individually packaged in a sealed protective container with a peel tab.

The pen needle is used by peeling back the peel tab and snapping or screwing the housing onto the threaded end of the pen-injector. The back end of the cannula punctures the septum of the drug reservoir in the pen-injector. The outer protective cap is then removed. When the injection is needed the needle is inserted into the chosen site. While inserting the needle into the skin, the safety shield glides into the housing enabling the needle to penetrate the skin barrier and the subcutaneous tissue. While the safety shield glides into the housing the safety mechanism is activated. The pen-injector delivers the medicinal product through the needle.

After the injection, in order to remove the needle from the skin, the user withdraws the pen injector from the skin. As the pen and needle is withdrawn from the skin, the safety shield glides back in its initial position, completely covering the needle, where it will remain locked. The safety shield is designed to automatically cover the needle (providing passive protection) to minimize the risk of accidental needle-stick injury. Once the Clickfine AutoProtect Pen Needle is in the locked mode, it can no longer be used. The red safety lock indicator tells the user that the safety lock has been activated. The needle is detached from the injection device and disposed of in accordance with local regulations. For each subsequent injection, another disposable needle must be used.

8.8 Intended Use

The intended use of the modified device remains the same as the predicate device Clickfine AutoProtect Pen Needles K131123:

The Clickfine AutoProtect Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin and exenatide.

Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.

  • 8.9 Technological Characteristics and Substantial Equivalence discussion For the 5 mm Clickfine AutoProtect, except for the material of the needle shield, all of the material, processes and operating principles are identical to the currently marketed pen needles.

{5}------------------------------------------------

CharacteristicCurrent Ypsomed Clickfine AutoProtectInjection Pen Needle 8 and 6 mmModified YpsomedClickfine AutoProtectInjection Pen Needle5 mm
510(k)K131123The Clickfine AutoProtect Pen Needle isintended for use with pen injector devices forthe injection of fluids, including insulin andexenatide.N/A
Indications foruseAdditionally, the attached safety shieldautomatically locks in place and reduces theoccurrence of accidental needle sticks fromthe patient end of the needle. The shield alsoserves to hide the needle before and afterinjection.Unchanged
OTCOTCUnchanged
Needle length6 and 8 mm5, 6 and 8 mm
Needle gauge31, 30 and 29 gaugeUnchanged
Needle outerdiameter0.25, 0.3 and 0.33 mmUnchanged
Normal walled needle:$\ge$ 0.114, $\ge$ 0.133 and$\ge$ 0.133 mmUnchanged
Needle innerdiameterThin walled needle:$\ge$ 0.125, 0.165 and$\ge$ 0.190 mmUnchanged
Needle grindFront and back bevelUnchanged
Needle tip:3-bevelUnchanged
Method ofattachment topen"Snap-on" & "Twist-on” attachment andTwist-off removalUnchanged
NeedlematerialStainless steelUnchanged
Needle/hubbondAdhesive bondedUnchanged
Hub materialPPUnchanged
HousingWhite colored PPUnchanged
Needle shieldBlue colored MBS, using the Zylar 530granulateBlue colored MBS,using the Zylar 550granulate
Safety lockindicatorRed colored PPUnchanged
SpringStainless steelUnchanged
CharacteristicCurrent Ypsomed Clickfine AutoProtectInjection Pen Needle 8 and 6 mmModified YpsomedClickfine AutoProtectInjection Pen Needle5 mm
OuterprotectivecontainermaterialBlue colored PPUnchanged
Peel tabPETUnchanged
SterilizationmethodGamma irradiationUnchanged
Sterilizationconditions25 kGy17.5 kGy
IFULeafletUnchanged

Comparison to Predicate Device

{6}------------------------------------------------

8.10 Performance Data

As part of the Design Control Activities, Ypsomed evaluated the potential risks, hazards or hazardous situations, potential clinical impact on the patient, identified the appropriate risk control measure and acceptability criteria associated with each change in the modified devices. Then, each risk was mitigated through appropriate verification and/or validation activities which were determined to be the testing specified in the following international and internal standards and protocols. This testing confirmed that the subject devices are substantially equivalent to the predicate devices.

The Clickfine AutoProtect Pen Needles have met the requirements of the relevant sections of the following standards:

  • ISO 11608-2:2012 Needle based injection systems for medical use -. Requirements and test methods - Part 2: Needles
  • ISO 9626:1991/Amd.1:2001 Stainless steel needle tubing for the manufacture of medical devices
  • ISO 7864:1993 Sterile hypodermic needles for single use
  • . ISO 23908:2011 Sharps injury protection - Requirements and test methods
TestRequirementResult
Needle assemblyattachment shallfit and functionwith needlebased injectionsystemsspecified inISO 11608-1Clause 4.2.1 of ISO 11608-2The pen needles were tested for compatibility withneedle based injection system as specified in ISO11608-1Pass
TubingdimensionsClause 4.2.2, Table 1 of ISO 11608-2Length of patient end: +/- 1.25 mmLength of cartridge end: 5.7 – 7.0 mmPass
TestRequirementResult
Flow rateClause 4.3 of ISO 11608-2:Measurement of flow rate in accordance withAnnex A.Pass
Bond betweenhub and needletubeClause 9 of ISO 11608-2 and clause 13.1 of ISO7864The union of the hub and needle tube shall notbreak for at least 5 sec. while a force of at least 22N is appliedPass
Needle pointsClause 4.5 of ISO 11608-2Visually sharp at a 2.5x magnification; designed tominimize coring and fragmentationPass
LubricationClause 4.7 of ISO 11608-2No visible droplets on the outside surface of theneedle tube.Pass
Dislocation ofmeasuring pointat the patient-endClause 8 of ISO 11608-2:maximum allowable dislocation acc. Clause 4.8,Table 2 for 5mm: 0.65mm and for 8mm: 0.9mmPass
Functionalcompatibilitywith needlebased injectionsystemsClause 11 of ISO 11608-2:Needle assembly torque: 0.07 +/- 0.01 NmNeedle hub removal: less than 0.100 NmDose accuracy:for doses ≤ 20 ml the calculated values were within± 0.01 ml of the targeted dose;for doses > 20 ml the calculated values were within± 5 % of the targeted dose.Pass
TubingcharacteristicsTable 1 of ISO 9626/A1:The tubing is made of austenitic stainless steel acc.one of the designated types given in Table 1 of ISO9626/A1.Pass
TubingdiametersTable 2 of ISO 9626/A1Tubing dimensions meet outer and inner diameterrequirements as defined in table 2 of ISO 9626/A1.Pass
StiffnessClause 9 of ISO 9626:When tested in accordance with annex C of ISO9626 the tubing shall show a deflection not greaterthan the relevant value given in table 3 of ISO9626/A1.Pass
Resistance tobreakageClause 10 of ISO 9626:When tested in accordance with annex D of ISO9626 and Table 4 of ISO 9626/A1 the tubing shallnot breakPass
TestRequirementResult
Limits foracidity andalkalinityClause 6 of ISO 9626Tested in accordance with annex A and extractpreparation of the tubing in accordance with AnnexB: correction for the volume of titrant required forthe control fluid shall not be more than 0.04 mlNaOH or 0.12 ml HCl to reach the end-pointtitration.Pass
Resistance tocorrosionClause 11 of ISO 9626:When tested in accordance with annex E, theimmersed half of the tubing shall show no evidenceof corrosion resulting from the test.Pass
Freedom ofdefectsClause 11.3 in ISO 7864:When examined by normal or corrected-to-normalvision, the needle shall appear straight and ofregular cross-section and wall thicknessPass
Limits forextractablemetalsClause 6 of ISO 7864:When tested with a recognized method the contentof lead, tin, zinc and iron shall be less than 5mg/l.The cadmium content shall be lower than 0.1 mg/l.Pass
Freedom ofdefectsClause 11.3 of ISO 7864:When examined by normal or corrected vision, theneedle lube shall appear straight and of regularcross-section and wall thickness.Pass
Activation of thesharps injuryprotectionIn accordance with ISO 23908 clause 5.2 the sharpsinjury protection feature shall be able to beactivated by an appropriate force that easesactuation and avoids unintended actuation. Anappropriate force was determined using a risk-basedapproach in accordance with ISO 14971.Pass
Verification ofsafety featureactivationIn accordance with ISO 23908, clause 5.3, needleswere tested to provide evidence that once the safetyfeature was activated, they resist axial and lateralforces so as to prevent unintended exposure to thesharps.Pass
Device designverification forneedle tipprotection afteractivationIn accordance to clause 5.4 of ISO 23908 it wasdemonstrated that a sphere of 6mm radius did notcome into contact with the tip of the cannula, eitherbefore or immediately after triggering the safetymechanismPass

{7}------------------------------------------------

{8}------------------------------------------------

The 5 mm Clickfine AutoProtect Pen Needles were tested, and met all acceptance criteria, for all the requirements as provided above. Based on the results Y psomed concluded that the device performance is acceptable for the product.

8.11 Clinical Data

{9}------------------------------------------------

Clinical Data was not required to establish substantial equivalence for these devices.

  • 8.12 Conclusion
    Ypsomed AG concludes based on the information presented that the subject device is substantially equivalent to the predicate product legally marketed in the USA.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).