K Number
K180918
Device Name
Planmed Verity
Manufacturer
Date Cleared
2018-11-30

(235 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.

Device Description

The Planmed Verity is a cone beam computed tomography x-ray system for generating 3D imaging scans of extremity, head and neck anatomies. The Planmed Verity utilizes an amorphous silicon based digital image receptor to capture digital images. The receptor directly converts the incoming X-ray photons to digital image data.

The workflow with Planmed Verity is controlled from the integrated acquisition workstation and Planmed Verity Manager image acquisition and communications software. The patient information is entered manually or received from the hospital. radiology, or x-ray modality information systems (HIS, RIS, or MIS, respectively), as a format of modality worklist. Subsequently, the images are acquired, processed, and displayed for preview. After initial evaluation by the operator, the images are either printed or transferred for soft-copy review.

AI/ML Overview

The provided text describes the Planmed Verity, a Cone Beam Computed Tomography (CBCT) system, and its comparison to a predicate device for FDA 510(k) clearance. The focus of the changes in the subject device is the inclusion of new image enhancement software (ULD and CALM) and a slightly updated flat panel detector.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a clear, tabulated format. Instead, it relies on demonstrating equivalence to the predicate device and showing improvements in specific aspects. The primary acceptance criterion seems to be "substantial equivalence" to the predicate device in terms of safety and effectiveness for the stated indications for use, with additional improvements.

Acceptance Criterion (Inferred)Reported Device PerformanceStudy Demonstrating Performance
BiocompatibilityNo adverse effect on patient from material contact.Safety evaluation of acute and repeat toxicity.
Electrical Safety & EMCComplies with IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, ISO 10993-1, and IEC 60601-1-2.Electrical safety and EMC testing.
Software V&VDemonstrated to perform as intended; "Moderate" level of concern.Software verification and validation testing.
Image Quality (General)Overall image quality acceptable for all cases and image types.Clinical image evaluation by an experienced radiologist.
Image Quality (Detector)Imaging performance of new detector is essentially equal to predicate, or slightly better (MTF, DQE, noise).Comparison of manufacturer's (Varex) non-clinical physics data for detector types.
Image Quality (CALM)Improves imaging quality in most cases; reduces risk of re-takes due to patient movement; clearly reduces motion blur.Clinical test report scoring CALM vs. no CALM; phantom study with simulated motion blur.
Image Quality (ULD)Clinically suitable for imaging head and neck anatomies; lowers patient dose while maintaining/improving diagnostic image quality.Separate study at a clinic in Helsinki, Finland, where radiologists evaluated clinical image quality.
Safety and EffectivenessAs safe and effective as the predicate system for its indicated use.Overall conclusion based on all testing (biocompatibility, electrical safety, software, physical, clinical).

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Image Evaluation):

    • Sample Size: The document mentions "representative sample scans and diagnostic images" for the clinical image evaluation. It does not provide a specific numerical sample size.
    • Data Provenance: The text does not explicitly state the country of origin or whether the clinical data was retrospective or prospective for the main clinical image evaluation. However, the ULD protocol verification was performed "at a clinic in Helsinki, Finland." The clinical image evaluation was performed with a system not in final form (using 'D' detector instead of 'DX'), suggesting it might have been an early prospective study or an evaluation based on existing data.
  • Phantom Study (CALM):

    • Sample Size: Not specified (likely a single phantom with various simulated motion scenarios).
    • Data Provenance: Not specified, likely an in-house laboratory study.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • Clinical Image Evaluation: "an experienced radiologist" - One expert. Qualifications: "experienced radiologist." No further detail regarding years of experience or subspecialty is provided.
  • ULD Protocol Verification: "The radiologists evaluated the clinical image quality..." - Multiple experts (plural "radiologists"). Qualifications: "radiologists." No further details on their experience or subspecialty.

4. Adjudication Method for the Test Set

  • The document does not describe an adjudication method like 2+1 or 3+1. For the clinical image evaluation, it states "an experienced radiologist scored representative sample scans." For the ULD protocol, "The radiologists evaluated the clinical image quality." This suggests individual evaluation without a formal multi-reader adjudication process described in the text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No explicit MRMC comparative effectiveness study is mentioned comparing human readers with AI vs. without AI assistance.
    • The CALM feature study compares imaging quality with CALM vs. without CALM, implying an algorithm performance comparison, not necessarily a human reader performance study with and without that algorithm. It improves image quality, which could indirectly improve human reader performance, but this is not directly assessed in an MRMC setting.
    • The ULD protocol evaluation focused on whether the image quality was "clinically suitable," not on a direct comparison of human diagnostic accuracy with and without ULD-processed images.

6. Standalone (Algorithm Only) Performance Study

  • Yes, standalone performance (algorithm only) was implicitly evaluated for the new software features:
    • CALM: The statement "The effectiveness of the CALM feature was additionally verified with imaging a scull phantom with suitable test inserts. Simulated small and large amplitude motion blur as well as step-like distortion was added to the projection images and the effectiveness of the CALM feature to correct the motion blur was verified. Results show that the CALM algorithm clearly reduces motion blur." This is a standalone assessment of the algorithm's ability to correct motion blur using a phantom.
    • ULD: While radiologists evaluated the clinical suitability, the ULD feature itself is an image processing algorithm designed to maintain image quality at lower doses. Its effectiveness in lowering dose while producing acceptable images is a standalone characteristic.

7. Type of Ground Truth Used

  • Expert Consensus/Opinion: For the clinical image evaluation, the "experienced radiologist" provided the 'ground truth' by scoring image quality. Similarly, "the radiologists" for the ULD protocol provided their clinical suitability judgment. This relies on expert opinion.
  • Phantom Data: For the CALM feature evaluation, the 'ground truth' was derived from the known simulated motion blur introduced into the phantom images. This is a controlled experimental ground truth.
  • Non-clinical Physics Data: For the detector comparison, the ground truth was based on objective physical metrics (MTF, DQE, noise performance) provided by the manufacturer (Varex).

8. Sample Size for the Training Set

  • The document does not provide information on the sample size used for training the algorithms (ULD noise filtration, CALM motion blur reduction). This is a common omission in 510(k) summaries, which often focus on verification and validation of the final product.

9. How Ground Truth for the Training Set Was Established

  • The document does not provide information on how ground truth was established for the training data used for the ULD and CALM algorithms.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

November 30, 2018

Planmed Oy Lars Moring Regulatory Affairs Manager Sorvaajankatu 7 HELSINKI, 00880 FINLAND

Re: K180918

Trade/Device Name: Planmed Verity Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS, JAK Dated: April 5, 2018 Received: April 9, 2018

Dear Lars Moring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Hol 2. Mild

for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180918

Device Name

Planmed Verity

Indications for Use (Describe)

Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

Manufacturer

Planmed Oy Sorvaajankatu 7 FI-00880 Helsinki, Finland Phone: +358 20 7795 300 Fax: +358 20 7795 396 Contact person: Lars Moring

U.S. designated agent

Planmed USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 894 2200 Fax: (630) 894 4271 Contact person : Ed McDonough

Date Prepared: November 28, 2018

II. DEVICE

Name of Device:Planmed Verity
Common or Usual Name:Cone Beam Computed Tomography (CBCT) System
Classification Name:Computed Tomography X-ray System (CT)(21 CFR 892.1750)
Regulatory Class:II
Product Code:OAS, JAK

III. PREDICATE DEVICE

Planmed Verity cone beam computed tomography X-ray System, K143435. This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

DEVICE DESCRIPTION IV.

The Planmed Verity is a cone beam computed tomography x-ray system for generating 3D imaging scans of extremity, head and neck anatomies. The Planmed Verity utilizes an amorphous silicon based digital image receptor to capture digital images. The receptor directly converts the incoming X-ray photons to digital image data.

The workflow with Planmed Verity is controlled from the integrated acquisition workstation and Planmed Verity Manager image acquisition and communications software. The patient information is entered manually or received from the hospital. radiology, or x-ray modality information systems (HIS, RIS, or MIS, respectively), as a format of modality worklist. Subsequently, the images are acquired, processed, and displayed for preview. After initial evaluation by the operator, the images are either printed or transferred for soft-copy review.

{4}------------------------------------------------

V. INDICATIONS FOR USE

Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.

VI. COMPARISON OF INDICATIONS FOR USE

An added imaging protocol has been introduced to the system compared with the predicative device version.

The Indications for Use (IFU) statements are similar for both systems. IFU of the predicate system refers to the currently cleared system version with the indications for use covering extremities and head anatomies (mainly maxillofacial and sinus). The new version of the system may also be used for imaging of other head anatomies and the neck. This protocol uses a new patient support tray.

VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

General

Both the predicate device and the subject device are using the same basic operating principles and are technically almost identical. The subject device is an improved version of the predicate device. It has the same outer design and no changes to the physical construction visible from the outside with the exception of one added optional patient support tray. X-ray generation and control are the same. The controlling PC workstation of the subject device is integrated into the x-ray device like in the predicate system. The user interface is still the same touch screen like in the predicate device and the managing software is essentially unchanged in its controls and functionality and from the operator's perspective. All labeling material is mainly unchanged. Only the added software features are updated to the user's manual and to the technical manual.

Integrated detector

The system uses an integrated flat panel detector. The type of the detector has changed slightly to a new version of the same detector. Pixel matrixes are identical, pixel size is 127 um by 127um. Quality assurance with pixel defect acceptance criteria comparison is unchanged.

X-ray unit

Physical dimensions of the units are the same.

X-ray field size and SID

X-ray field size, SID, cone beam geometry and scanning angle are unchanged.

X-ray tube

Units use the same X-ray tube.

X-ray generator

Units use the same X-ray generator.

X-ray collimator

Collimator mechanics has been simplified with same functionality and filtration.

Differences and added features

The complete list of differences of the subject device to the predicate device is as follows:

The integrated PC's brand has been changed from HP to Dell. The Microsoft Windows operating system has been upgraded from Windows 7 to Windows 10.

The Verity Manager software newly includes image enhancement options like ULD (Ultra Low Dose) noise filtration and CALM (Correction Algorithm for Latent Movement) motion blur reduction protocols.

{5}------------------------------------------------

One optional patient supporting tray has been added for the newly introduced head and neck imaging protocols. All previously used patient support options of the predicate system are available and supported also with the subject system.

The amorphous silicon detector is replaced by a new version of the same detector type. The new version of the detector has slightly better imaging characteristics (MTF, DQE and noise performance) as the predicate device. Yet its physical construction, specifications and software interfacing are essentially the same as before.

There are no other differences of the subject system to the predicate system.

The above listed differences are deemed to be non-critical for the subject device compared to the predicate since the system is mainly unchanged. The safety of the system and the imaging performance for the intended use are maintained or improved for the new system version in comparison to the predicate. The added features have been clinically tested and approved by experienced operators and radiologists and found valuable for diagnostic use without adding risks to patient or operator safety or health. Labeling material has been updated to include the new functionality. Dosimetry information has been added to aid in judgement of the radiation safety of intended diagnostic procedures.

VIII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

To evaluate biocompatibility of the Planmed Verity CBCT device, we performed safety evaluation of acute and repeat toxicity. As conclusion, none of the compounds measured in the extracts were present at levels which could adversely affect the patient. Thus, there is no risk or concern to the patient's safety from contact with the materials of construction of this X-ray unit.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Planmed Verity CBCT system. The system complies with standards IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, ISO 10993-1 for safety, and IEC 60601-1-2 for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.

Physical laboratory testing

The performance of the system has not changed from the predicate device. Image quality is not affected by the change of the detector type from 2520D to 2520 DX version. Thus no new physical laboratory testing was deemed necessary. Instead a comparison of the manufacturer's (Varex) non-clinical physics data was performed. Both systems use the same software algorithms and interfacing software to control the flat panel detector and to compute the reconstruction. The image acquisition chain or the imaging protocols' number of frames and scan angles have not been altered in the subject system compared to the predicate device.

Clinical image evaluation

A clinical image evaluation has been performed where an experienced radiologist scored representative sample scans and diagnostic images. The overall image quality was acceptable for all cases and image types.

{6}------------------------------------------------

Parts of the clinical evaluations were performed with a system not being in final finished form of the subject device. More precisely the flat panel detector used has been of type 'D' not 'DX'. However since the imaging performance of both types is essentially equal and would not alter the scoring of the image quality seen by the radiologist we believe that all clinical testing is relevant and acceptable for the verification of the subject device.

It may also be noted that the new software features CALM and ULD can be upgraded to the predicate Verity systems. Once cleared for marketing, these features will benefit also the currently installed base of units and the patient diagnostics and radiation safety.

IX. CONCLUSIONS

The renewed UL/CB/CE testing and certification process support the safe use and imaging performance and usability of the device. Both hardware and software verification and validation demonstrate that the Planmed Verity system performs as intended in the specified use conditions. The clinical image evaluation also shows that the device performance is equal to the predicate device that is currently marketed for the same intended use. The added software features CALM and ULD improve image quality and lower the patient dose while maintaining and/or improving diagnostic image quality.

Summary of conclusions drawn from the clinical and non-clinical testing:

Since the overall construction, x-ray generation or control software have not changed significantly there are no new added risks to the subject device. The integrated flat panel detector has changed to a newer version with the same imaging performance specifications. Hence no new physical laboratory testing was deemed necessary. Only the new revision of the flat panel imaging detector performance has been compared for its non-clinical testing data and was found equal or slightly better.

The clinical tests included imaging performance scoring with use of the CALM motion blur reduction software feature, as well as use of the ULD ultra low dose imaging protocol for patient radiation dose reduction.

The clinical test report describes the CALM feature in detail. The scoring using CALM compared to not using it is presented. Based on this evaluation the CALM feature improves the imaging quality in most cases and hence it reduces the risk of re-takes because of patient movement. An added section with results of imaging performance evaluation with CALM is presented.

The effectiveness of the CALM feature was additionally verified with imaging a scull phantom with suitable test inserts. Simulated small and large amplitude motion blur as well as step-like distortion was added to the projection images and the effectiveness of the CALM feature to correct the motion blur was verified. Results show that the CALM algorithm clearly reduces motion blur.

ULD protocol was verified in a separate study at a clinic in Helsinki, Finland. The radiologists evaluated the clinical image quality of the ULD feature to be clinically suitable for imaging head and neck anatomies.

Based on the clinical tests performed it can be concluded that the subject device is as safe and effective for its indicated use as the predicate system when used as instructed.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.