AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)-Extended Cuff is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU) is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

Device Description

Not Found

AI/ML Overview

This document is a 510(k) premarket notification decision letter from the FDA for medical gloves. It describes the "Indications for Use" for two types of gloves: "Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)- Extended Cuff" and "Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU)".

The acceptance criteria for the devices are based on their resistance to permeation by various chemotherapy drugs, measured by breakthrough time. The study conducted to prove the devices meet these criteria is a chemotherapy drug permeation test.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Device: Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)- Extended Cuff

Chemotherapy Drug (Concentration)Acceptance Criteria (Breakthrough Time)Reported Device Performance
Carmustine (3.3 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.20.3 minutes
Cisplatin (1.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Cyclophosphamide (20 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Dacarbazine (10.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Doxorubicin HCl (2.0mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Etoposide (20.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Fluorouracil (50.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Methotrexate (25 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Mitomycin C (0.5 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Paclitaxel (6.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Thiotepa (10.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.100.6 minutes
Vincristine Sulfate (1.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes

Device: Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU)

Chemotherapy Drug (Concentration)Acceptance Criteria (Breakthrough Time)Reported Device Performance
Carmustine (3.3 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.43.8 minutes
Cisplatin (1.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Cyclophosphamide (20 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Dacarbazine (10.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Doxorubicin HCl (2.0mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Etoposide (20.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Fluorouracil (50.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Methotrexate (25 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Mitomycin C (0.5 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Paclitaxel (6.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes
Thiotepa (10.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.97.8 minutes
Vincristine Sulfate (1.0 mg/ml)Not explicitly stated as "acceptance criteria", but the measured time is the performance.> 240 minutes

Note: The document states "Please note that Carmustine and Thiotepa have extremely low permeation times of 20.3 minutes and 100.6 minutes" for VBLU gloves, and "Please note that Carmustine and Thiotepa have extremely low permeation times of 43.8 minutes" for ABLU gloves. This implies that while these times are lower than other drugs, they are considered acceptable for appropriate use with awareness of the breakthrough time.


The provided document does not contain the following information typically associated with AI/software medical device studies:

  • 2. Sample size used for the test set and the data provenance (This is a physical product test, not an AI model test.)
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (Not applicable to a physical product permeation test.)
  • 4. Adjudication method (Not applicable.)
  • 5. Multi-reader multi-case (MRMC) comparative effectiveness study (Not applicable, as this is not an AI diagnostic device.)
  • 6. Standalone (algorithm only without human-in-the-loop performance) study (Not applicable, as this is not an AI algorithm.)
  • 7. The type of ground truth used (For this type of device, the "ground truth" is the empirically measured breakthrough time of the chemical through the glove material under controlled laboratory conditions, typically following standards like ASTM D6978. The document does not specify the exact standard or methodology, but it's an objective measurement.)
  • 8. The sample size for the training set (Not applicable, as this is not an AI model.)
  • 9. How the ground truth for the training set was established (Not applicable.)

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 2, 2018

Hartalega Sdn. Bhd. Nurual Kong Ouality Assurance Senior Manager No. 7. Kawasan Perusahaan Suria Bestari Jaya, 45600 My

Re: K180786

Trade/Device Name: Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Aqua Blue (ABLU), Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs Violet Blue (VBLU) - Extended Cuff Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: October 11, 2018 Received: October 15, 2018

Dear Nurual Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K180786

Device Name

Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)- Extended Cuff

Indications for Use (Describe)

The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Violet Blue (VBLU)-Extended Cuff is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

The list of Chemotherapy Drugs tested (with breakthrough times) is as below:

Carmustine (3.3 mg/ml)20.3 minutes
Cisplatin (1.0 mg/ml)> 240 minutes
Cyclophosphamide (20 mg/ml)> 240 minutes
Dacarbazine (10.0 mg/ml)> 240 minutes
Doxorubicin Hydrochloride (2.0mg/ml)> 240 minutes
Etoposide (20.0 mg/ml)> 240 minutes
Fluorouracil (50.0 mg/ml)> 240 minutes
Methotrexate (25 mg/ml)> 240 minutes
Mitomycin C (0.5 mg/ml)> 240 minutes
Paclitaxel (6.0 mg/ml)> 240 minutes
Thiotepa (10.0 mg/ml)100.6 minutes
Vincristine Sulfate (1.0 mg/ml)> 240 minutes

Please note that Carmustine and Thiotepa have extremely low permeation times of 20.3 minutes and 100.6 minutes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|× Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K180786

Device Name

Sterile Nitrile Powder Free Examination Gloye Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU)

Indications for Use (Describe)

The Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs - Aqua Blue (ABLU) is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

The list of Chemotherapy Drugs tested (with breakthrough times) is as below:

Carmustine (3.3 mg/ml)43.8 minutes
Cisplatin (1.0 mg/ml)> 240 minutes
Cyclophosphamide (20 mg/ml)> 240 minutes
Dacarbazine (10.0 mg/ml)> 240 minutes
Doxorubicin Hydrochloride (2.0mg/ml)> 240 minutes
Etoposide (20.0 mg/ml)> 240 minutes
Fluorouracil (50.0 mg/ml)> 240 minutes
Methotrexate (25 mg/ml)> 240 minutes
Mitomycin C (0.5 mg/ml)> 240 minutes
Paclitaxel (6.0 mg/ml)> 240 minutes
Thiotepa (10.0 mg/ml)97.8 minutes
Vincristine Sulfate (1.0 mg/ml)> 240 minutes

Please note that Carmustine and Thiotepa have extremely low permeation times of 43.8 minutes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.