(112 days)
The White Nitrile Examination Glove,Sterile, is a disposable device intended for medical purposes that is worn on the examiner Is hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards: Practice for assessment of Medical Glove to Permeation by chemotherapy Drugs
The subject device in this 510(k) Notification is White Nitrile Examination Gloves Sterile with tested for use with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, White color, powder free and sterile (Per 21 CFR 880.6250, class I). The White Nitrile Examination Glove ,Sterile, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and is used with chemotherapy drugs (Product code LZC).
The provided document is a 510(k) Premarket Notification for a medical device: "White Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim." It outlines the acceptance criteria for the glove's performance, particularly its resistance to chemotherapy drugs, and summarizes the non-clinical testing performed.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The primary acceptance criteria for the chemotherapy drug resistance claim are defined by ASTM D6978-05 standards: Practice for assessment of Medical Glove to Permeation by chemotherapy Drugs. The device's performance is reported as the "Minimum Breakthrough detection time in minutes." The acceptance criteria are implicit in the comparison to the predicate device and the standard, aiming for breakthrough times that prevent drug permeation for a reasonable duration of use. The reported "Minimum Breakthrough detection time" for most drugs is "> 240 Minutes", indicating that no permeation was detected within the maximum test duration.
| Chemotherapy Drugs and Concentration | Acceptance Criteria (Implicit) | Reported Device Performance (Minimum Breakthrough detection time in minutes, 0.01 µg/cm2/minute) |
|---|---|---|
| Carmustine (BCNU) (3.3 mg/ml) | Adequate time to prevent permeation during anticipated use | 17.1 Minutes |
| Cyclophosphamide (20 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes |
| Cisplatin (1.0 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes (from table on page 2) |
| Dacarbazine (DTIC) (10 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes (from table on page 2) |
| Doxorubicin Hydrochloride (2.0 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes |
| Etoposide (20 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes |
| Fluorouracil (50 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes |
| Paclitaxel (6.0 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes |
| Thiotepa (10 mg/ml) | Adequate time to prevent permeation during anticipated use | 55.9 minutes |
| Cytarabine (10 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes (from table on page 4) |
| Methotrexate (25 mg/ml) | Adequate time to prevent permeation during anticipated use | > 240 Minutes (from table on page 4) |
Other acceptance criteria and performance related to the general glove properties (ASTM D6319-10 Standard specification for Nitrile Examination Gloves) and Biological Evaluation on medical device:
| Property | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Tensile Strength (MPa) - Before Ageing | 14 min | 14 min |
| Ultimate Elongation (%) - Before Ageing | 500 min | 500 min |
| Tensile Strength (MPa) - After Ageing | 14 min | 14 min |
| Ultimate Elongation (%) - After Ageing | 400 min | 400 min |
| Pin-hole Level | AQL 2.5 Inspection Level G-1 | Meets ASTM D6319-10 (AQL 2.5 Inspection Level G-1 via ASTM D5151-06) |
| Residual Powder | <2.0 mg/pc | Meets ASTM D6124-06 (<2.0 mg/pc) |
| Overall Length (mm) | Min 230 mm | Meets ASTM D6319-10 |
| Width (± 5mm) - Size S | 85 mm | Meets ASTM D6319-10 |
| Width (± 5mm) - Size M | 95 mm | Meets ASTM D6319-10 |
| Width (± 5mm) - Size L | 105 mm | Meets ASTM D6319-10 |
| Width (± 5mm) - Size XL | 115 mm | Meets ASTM D6319-10 |
| Thickness at Palm (mm) | Min 0.05 mm | Meets ASTM D6319-10 |
| Thickness at Finger Tip (mm) | Min 0.05 mm | Meets ASTM D6319-10 |
| Primary Skin Irritation Test (ISO 10993-10) | Non-irritant | Non-irritant |
| Dermal Sensitization Assay (ISO 10993-10) | Non-sensitizer | Non-sensitizer |
| Sterility | Sterile | Sterile (Gamma Irradiation ISO 11137-2:2015) |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the specific sample size used for the chemotherapy drug permeation test set. However, it indicates that the testing was performed "in accordance with ASTM D6978-05 standards." This standard would dictate the sampling methodology and number of replicate tests required.
The data provenance is not explicitly mentioned (e.g., country of origin of the testing lab), but it is a premarket notification from Riverstone Resources SDN BHD in Selangor, Malaysia. The testing would have been conducted by a laboratory to generate data for this submission. The nature of the studies (material property and drug permeation) implies prospective testing specifically for this device submission, not retrospective data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the tests described are objective, laboratory-based analyses (material properties, drug permeation, biological evaluations) governed by established ASTM and ISO standards, rather than subjective evaluations requiring expert consensus or interpretation of images. The "ground truth" is determined by the results of these standardized physical and chemical tests.
4. Adjudication method for the test set
This information is not applicable for the reasons stated in point 3. The results are quantitative measurements against predefined thresholds in standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a physical examination glove, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical examination glove, not an algorithm.
7. The type of ground truth used
The ground truth used for this device's evaluation is primarily based on:
- Standardized Physical and Chemical Measurements: Adherence to ASTM D6319-10 (for general glove properties) and ASTM D6978-05 (for chemotherapy drug permeation), which involve objective measurements of tensile strength, elongation, dimensions, pinhole levels, residual powder, and breakthrough times.
- Biological Evaluation Standards: Adherence to ISO 10993-10 for primary skin irritation and dermal sensitization, which are also standardized laboratory tests with objective endpoints.
8. The sample size for the training set
This information is not applicable. There is no "training set" in the context of physical medical device testing for properties like drug permeation and material strength. The device is not an AI model requiring a training phase.
9. How the ground truth for the training set was established
This information is not applicable for the reasons stated in point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2018
Riverstone Resources SDN BHD Suresh Kumar QA Manager Lot 56, No 15, Jalan Jasmin 2 Bukit Beruntung, Selangor 48300 Malaysia
Re: K180452
Trade/Device Name: White Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: May 3, 2018 Received: May 11, 2018
Dear Suresh Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K180452
Device Name
White Nitrile Examination Gloves Sterile Tested For Used With Chemotherapy Drugs Claim
Indications for Use (Describe)
The White Nitrile Examination Glove,Sterile, is a disposable device intended for medical purposes that is worn on the examiner Is hand or finger to prevent contamination between patient and examiner.
In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards: Practice for assessment of Medical Glove to Permeation by chemotherapy Drugs
| Chemotherapy Drugs and Concentration | Minimum Breakthrough detection time in minutes, 0.01 $\mu$ g/cm2/minute |
|---|---|
| Carmustine (BCNU)(3.3 mg/ml) | 17.1 Minutes |
| Cyclophosphamide (20 mg/ml) | > 240 Minutes |
| Cisplatin (1.0 mg/ml) | > 240 Minutes |
| Dacarbazine (DTIC) (10 mg/ml) | > 240 Minutes |
| Doxorubicin Hydrochloride (2.0 mg/ml) | > 240 Minutes |
| Etoposide (20 mg/ml) | > 240 Minutes |
| Fluorouracil (50 mg/ml) | > 240 Minutes |
| Paclitaxel (6.0 mg/ml) | > 240 Minutes |
| Thiotepa (IO mg/ml) | 55.9 minutes |
The Maximum testing time is 240 Minutes. Please note that the following drugs have extremely low permeartion time: Carmustine (BCNU)(3.3mg/ml) Thiotepa ( 10mg/ml)
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
510K No : K180452
510(K) Summary White Nitrile Examination Glove,Sterile Tested for use with Chemotherapy Drugs claim
1.0 Submitter :
| Company Name | : RIVERSTONE RESOURCES SDN BHD |
|---|---|
| Company Address | : Lot 55, No 13, Jalan Jasmin 2,Kawasan Perindustrian Bukit Beruntung48300, Bukit Beruntung Selangor Malaysia |
| Contact Person | : | Mr Suresh Kumar |
|---|---|---|
| Telephone No | : | 603-60283033 |
| : | qa1@riverstone.com.my |
2.0 Preparation Date : 12th June 2018
3.0 Name of the Device
Trade Name / Proprietary Name : White Nitrile Examination Glove, Sterile tested for use with Chemotherapy Drugs claims.
Device Name : Nitrile Patient Examination gloves.
Device Classification Name : Patient Examination gloves (21 CFR 880.6250).
Device Class : Class I.
Product Code : Nitrile-LZA and LZC.
{4}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
510K No : K180452
4.0 Identification of The Legally Marketed Device :
The Primary Predicate K161187, EMG Blue Nitrile Examination Gloves Powder Free With Tested for use with Chemotherapy Drugs, and K171093 Sterile Nitrile Patient Examination Gloves, Powder Free.
5.0 Device Description
The subject device in this 510(k) Notification is White Nitrile Examination Gloves Sterile with tested for use with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, White color, powder free and sterile (Per 21 CFR 880.6250, class I).
The White Nitrile Examination Glove ,Sterile, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and is used with chemotherapy drugs (Product code LZC).
Indication for Use:
The White Nitrile Examination Glove, Sterile, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner and for use with chemotherapy drugs . In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
| Chemotherapy Drug and concentration | Minimum Breakthrough detectiontime in Minutes,µg/cm²/minute |
|---|---|
| 1)Carmustine (BCNU) (3.3mg/ml) | 17.1 minutes |
| 2)Cyclophosphamide(20mg/ml) | > 240 minutes |
| 3)Cytarabine (10mg/ml) | > 240 minutes |
| 4)Doxorubicin Hydrochloride (2 mg/ml) | > 240 minutes |
| 5)Etoposide (20mg/ml) | > 240 minutes |
| 6)Fluorouracil (50mg/ml) | > 240 minutes |
| 7)Methotrexate (25mg/ml) | > 240 minutes |
| 8) Paclitaxel (6mg/ml) | > 240 minutes |
| 9) Thiotepa (10mg/ml) | 55.9 minutes |
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time.
Carmustine (BCNU) (3.3mg/ml) - Minimum Breakthrough detection time 17.1 ug/cm2/minute.
Thiotepa (10mg/ml) - Minimum Breakthrough detection time 55.9 ug/cm²/minute.
{5}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
510K No : K180452
6.0 Specification for Nitrile gloves: 6.1 Dimension and Thickness of Gloves
| Dimension | Size S | Size M | Size L | Size XL |
|---|---|---|---|---|
| Overall Length (mm) (Minimum) | 230 | 230 | 230 | 230 |
| Width (± 5mm) | 85 | 95 | 105 | 115 |
| Thickness at Palm (mm)(Minimum) | 0.05 | 0.05 | 0.05 | 0.05 |
| Thickness at Finger Tip(mm)(Minimum) | 0.05 | 0.05 | 0.05 | 0.05 |
| Measurement | Before Ageing | After Aging at 70°C for 168 hrs @ 100°C for 22 hrs |
|---|---|---|
| Tensile Strength (MPa) | 14min | 14 Min |
| Ultimate Elongation (%) | 500min | 400min |
| Pin-hole Level | AQL 2.5 Inspection Level G-1 | AQL 2.5 Inspection Level G-1 |
6.1.2 Gloves Physical Properties and Holes
Pin-hole Level
Gloves meet all the specification listed in ASTM D 6319-10
7.0 Summary of the Technological Characteristics of the Device
There are no differences in technological characteristics of the subject device compare with the predicate device. The gloves are made from nitrile latex compound, White color, Sterile. The gloves met all the specifications in ASTM D6319-10 Standard specification for Nitrile Examination Gloves as well Biological Evaluation on medical device. Additionally, the gloves have been tested for permeability to chemotherapy drugs.
{6}------------------------------------------------
510K No : K180452
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
Acceptance Characteristics White Nitrile Sterile Nitrile EMG Blue Nitrile Assessment Examination Patient Examination Criteria Similarities Examination Glove, Sterile Gloves, Powder and tested for use with Gloves. Free with tested for Differences Powder use with chemotherapy drugs Free,K171093 chemotherapv claim.K180452 drugs, K161187 (Primary Predicate) LZA and LZA and LZC LZA LZA and LZC Subject is same LZC with 510k:161187 Subject is different with 510k :171093 Intended use A powder A powder free A powder free A powder free Same patient examination patient examination patient free patient glove is a disposable glove is a disposable examination glove examination device intended for device intended for is a disposable glove is a medical purposes medical purposes device intended for disposable that is worn on the that is worn on the medical purposes device examiner's hand or examiner's hand or that is worn on the intended for finger to prevent finger to prevent examiner's hand medical contamination contamination or finger to prevent purposes between patient and between patient and contamination that is worn examiner. examiner. between patient on the and examiner. examiner's hand or finger to prevent contaminati on between patient and examiner. The device is for The device The device is for The device is for over-the-counter over-the-counter over-the-counter is for overuse. use. use. the-counter use. Sterile Examination Sterile Examination Non Sterile Different Description glove glove Examination
{7}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung,
48300 Bukit Beruntung, Selangor
| 510K No : K180452 | Physical properties | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| made of Nitrile andcolor White andtested withchemotherapy Drugs | made of Nitrile andcolor Blue | glove made ofNitrile and colorBlue and testedwithchemotherapyDrugs | Before Aging | Meets ASTM D6319-10 | Meets ASTM D6319-10 | Meets ASTM D6319-10 | Same | ||||
| Presentation | Sterile glove areprovided inPouches | Sterile glove areprovided inPouches | Non Sterile isprovided indispenser | Different | Tensile Strength (MPa) = 14minUltimate Elongation (%) = 500minAfter Aging at 70°C for 168 hrs @100°C for 22 hrs | Tensile Strength (MPa) = 14min | |||||
| Ambidextrous | Yes | Yes | Yes | Same. | Ultimate Elongation (%) = 400min | ||||||
| Material use | NitrileSyntheticLatex | Nitrile SyntheticLatex | Nitrile SyntheticLatex | Nitrile SyntheticLatex | Similar | Freedom from pinholes | AQL 2.5 Inspection Level G-1 ASTM D6319-10 | Meets ASTM D5151-06 | Meets ASTM D5151-06 | Meets ASTM D5151-06 | |
| Colour | White | Blue | Blue | Different | Residual Powder | <2.0 mg/pc ASTM D6124-06 | Meets ASTM D6124-06 | Meets ASTM D6124-06 | Meets ASTM D6124-06 | ||
| Sterility | Sterility | SterileGamma IrradiationISO 11137-2:2015 | SterileEO sterilizationISO 11135:2014 | Non-sterile | DifferentThe difference insterilization processdid not raise newquestions of safetyand effectiveness | Biological Evaluation on Medical Device - Primary Skin Irritation Test | Non irritant- Primary Skin Irritation in RabbitISO 10993-10 | Under the conditions of this study, the test article was a non-irritant. | Under the conditions of this study, the test article was a non-irritant. | Under the conditions of this study, the test article was a non-irritant. | |
| Single used | Single used | Single used | Single used | Biological Evaluation on Medical Device- Dermal Sensitization Assay | Non - sensitizer- Skin sensitization in Giunea | Under the conditions of this study, the test article was a non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | |||
| Dimensions | OverallLength(mm)Min 230mmWidth (±5mm)Size S=85mmSize M=95mmSize L=105mmSize XL =115mmThicknessat Palm(mm)Min; 0.05mmThicknessat FingerTip (mm)Min 0.05mm | Meets ASTMD6319-10 | Meets ASTMD6319-10 | Meets ASTMD6319-10 | Same |
{8}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung,
48300 Bukit Beruntung, Selangor
510K No : K180452
{9}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung,
48300 Bukit Beruntung, Selangor
| ISO 10993-10 | ||||
|---|---|---|---|---|
| ResistanceagainstChemotherapyDrugs | StandardsPractice forAssessmentofresistanceof MedicalGlove toPermeationbyChemotherapy drugsASTMD6978-05(2013) | 1) Carmustine(3.3mg/ml or3000ppm),MinimumBreakthroughdetection time :17.1µg/cm²/minute2)Cyclophosphamide(20mg/ml or20,000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute3) Cisplatin(1.0mg/ml or1,000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute4)DoxorubicinHydrochloride(2.0mg/ml or2000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute5) Etoposide(20mg/ml or20,000ppm),MinimumBreakthroughdetection time: | 1) Carmustine(3.3mg/ml or3000ppm),MinimumBreakthroughdetection time:1.3µg/cm²/minute2)Cyclophosphamide (20mg/ml or20,000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute3) Cytarabine(100mg/ml),MinimumBreakthroughdetection time:>240µg/cm²/minute4)DoxorubicinHydrochloride(2.0mg/ml or2000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute5) Etoposide(20mg/ml or20,000ppm),MinimumBreakthroughdetection time:>240µg/cm²/minute | Different |
{10}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
510K No : K180452 >240 μg/cm²/minute 6) Flourouracil (50mg/ml or 6) Flourouracil 50,000), (50mg/ml or Minimum Breakthrough 50,000), Minimum Breakthrough detection time: detection time: >240 >240 μg/cm²/minute μg/cm²/minute 7) Methorexate (25mg/ml or 25.000ppm). 7) Paclitaxel Minimum (6mg/ml or Breakthrough 6,000ppm), detection time: > Minimum Breakthrough 240 detection time: ug/cm²/minute >240 μg/cm²/minute 8) Paclitaxel (6mg/ml or6,000ppm), 8) Thiotepa (10mg/ml or Minimum Breakthrough 10,000ppm), detection time: Minimum >240 Breakthrough ug/cm²/minute detection time: ર્સ વ μg/cm²/minute 9) Thiotepa (10mg/ml or 10,000ppm), Minimum Breakthrough detection time: 67.8 μg/cm²/minute 9) Dacarbazine (1.0 mg/ml ), Minimum Breakthrough detection time: >240 ug/cm²/minute
{11}------------------------------------------------
Lot 55, No 13, Jalan Jasmin 2, Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung, Selangor
510K No : K180452
8.0 Summary of Non-Clinical Testing
The gloves are made from nitrile latex compound, White colour,Sterile. The gloves met all the specifications in ASTM D6319-10 Standard specification for Nitrile Examination Gloves as well Biological Evaluation on medical device. Additionally, the gloves have been tested for permeability to chemotherapy drugs.
9.0 Conclusion
Based on intended uses, technological characteristics and non - clinical performance data, the White Nitrile Examination Gloves, Sterile with tested for use with Chemotherapy Drugs is as safe, as effective, and performs as well as the legally marketed predicate device, K 171093 ,Sterile Nitrile Patient Examination Gloves, Powder Free and K 161187, EMG Blue Nitrile Examination Gloves Powder Free With Tested for Use with Chemotherapy Drugs.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.