(312 days)
No
The summary describes a standard vital signs monitor with basic data storage, display, and alarming functions. There is no mention of AI, ML, or any advanced analytical capabilities that would suggest the use of such technologies. The performance studies focus on electrical safety, EMC, and software verification/validation, not on the performance of an AI/ML algorithm.
No
Explanation: The device is described as a monitor intended for monitoring, storing, reviewing, and alarming for physiological parameters. Its capabilities include displaying measuring data and producing alarms for patient care, but it does not describe any therapeutic or treatment function.
No
The device is intended for monitoring, storing, and reviewing physiological parameters and generating alarms, not for diagnosing a condition or disease.
No
The device description explicitly states it measures physiological parameters like NIBP, SpO2, PR, and TEMP, which requires hardware sensors and components. The performance studies also mention electrical safety and EMC testing, which are relevant to hardware.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for "monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters". These parameters (NIBP, SpO2, PR, TEMP) are measured directly from the patient's body, not from a sample of bodily fluid or tissue in vitro (outside the body).
- Device Description: The description reinforces that the device measures physiological parameters directly from the patient.
- Lack of IVD Characteristics: There is no mention of analyzing samples, reagents, or any other components typically associated with in vitro diagnostics.
IVD devices are used to examine specimens derived from the human body (such as blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening. This device measures vital signs directly from the patient, which falls under the category of patient monitoring devices, not IVDs.
N/A
Intended Use / Indications for Use
The monitor is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments.
Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP.
The monitor is not intended for MRI environments. TEMP module is not intended for neonates.
Product codes
MWI, DQA, DXN, FLL
Device Description
The iM3 has three work modes, Monitor, Spot and Round, to measure physiological parameters, including non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), and Quick TEMP/Infrared TEMP (TEMP).
• For the Monitor mode, the user can do continuous measurement, monitoring, alarming and manager patient data.
• For the Spot mode, also call spot check mode, the user can measure parameters quickly without creating a patient within the instrument workflow.
The Round mode is designed to support hospital staff when doing ward rounds. The Round mode is similar as Spot mode. The difference between Round mode and Spot mode is that for one ward round, the iM3 can additionally store one ward round record automatically (the latest measurement) or manually (user can choose the measure results).
iM3 capabilities include storing, displaying measuring data. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately.
The iM3can connect with EDAN’s MFM-CMS Central Monitoring System to display the iM3 data on a remote work station. The MFM-CMS system received FDA 510(k) clearance on June 21, 2013(K120727)
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults, pediatrics and neonates
Intended User / Care Setting
trained healthcare professionals in hospital environments.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical data: Electrical safety and electromagnetic compatibility (EMC) and Software Verification and Validation Testing were performed.
Electrical safety and EMC testing were conducted on the iM3 Vital Signs Monitor. The system complies with ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-8: 2006, IEC 60601-2-49: 2011, IEC 80601-2-30:2009, ISO 80601-2-56: 2009 and ISO 80601-2-61:2011 standards for safety and the IEC 60601-1-2: 2014 standard for EMC. Coexistence testing was performed and adhere to the Radio Frequency Wireless Technology in Medical Devices guidance document recommendations.
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The performance testing demonstrated that the subject devices are as safe and as effective as the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
December 21, 2018
Edan Instruments, Inc. % Alice Yang Regulatory Engineer Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94089
Re: K180380
Trade/Device Name: iM3 vital signs monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI, DQA, DXN, FLL Dated: November 21, 2018 Received: November 27, 2018
Dear Alice Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S5
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
K180380
Device Name iM3 vital signs monitor
Indications for Use (Describe)
The monitor is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments.
Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP.
The monitor is not intended for MRI environments. TEMP module is not intended for neonates.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
3
510(K) Summary
Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
1. Submitter: | Edan Instruments, Inc. |
---|---|
#15 Jinhui Road, Jinsha Community, | |
Kengzi Sub-District, Pingshan District, | |
Shenzhen, 518122 P.R.China. | |
Tel: +86(0755) 26858736 | |
Fax: +1 (408) 418-4059 | |
Contact person: | Alice Yang |
Preparing date: | December 21, 2018 |
2. Device name and classification: | Device Name: Vital Signs Monitor |
Model: iM3 | |
Classification Name/ Product code: | |
21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm)/MWI | |
21 CFR 870.1130 Noninvasive blood pressure measurement system/DXN | |
21 CFR 870.2700, Oximeter/DQA | |
21 CFR 880.2910 Clinical electronic thermometer/FLL | |
Regulatory Class: Class II | |
3. Predicate Device(s): | 1) Edan Instruments, Inc, Vital Signs Monitor Model M3, K131818(Primary) |
2) Edan Instruments, Inc, Patient Monitor Model V5 V6 V8, K160981 (Reference) | |
4. Device Description: | The iM3 has three work modes, Monitor, Spot and Round, to measure physiological parameters, including non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), and Quick TEMP/Infrared TEMP (TEMP). |
• For the Monitor mode, the user can do continuous measurement, monitoring, alarming and manager patient data. | |
• For the Spot mode, also call spot check mode, the user can measure parameters quickly without creating a patient within the instrument workflow. |
- The Round mode is designed to support hospital staff when doing ward .
4
rounds. The Round mode is similar as Spot mode. The difference between Round mode and Spot mode is that for one ward round, the iM3 can additionally store one ward round record automatically (the latest measurement) or manually (user can choose the measure results).
iM3 capabilities include storing, displaying measuring data. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately.
The iM3can connect with EDAN's MFM-CMS Central Monitoring System to display the iM3 data on a remote work station. The MFM-CMS system received FDA 510(k) clearance on June 21, 2013(K120727)
Not all sensors are intended for neonatal patients. The intended Note: population for each sensor is detailed in the section 6 below.
-
- Indication for Use The monitor is intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments. Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP. The monitor is not intended for MRI environments. TEMP module is not intended for neonates.
6. Predicate Device Comparison
The subject device shares the same characteristics in most items with the predicate device except in the following aspects:
Item | |||
---|---|---|---|
iM3 | M3 | V5/V6/V8 | |
K# | Current Submission | K131818 | K160981 |
Intended Use | |||
Description | The monitor is intended to be | ||
used for monitoring, storing, | |||
recording, and reviewing of, | |||
and to generate alarms for, | |||
multiple physiological | |||
parameters of adults, pediatrics | The monitor is intended to be | ||
used by qualified physicians or | |||
personnel professionally | |||
trained | |||
and it is for monitoring adults, | |||
pediatrics and neonates in | |||
hospital environments. | The monitors are intended to be | ||
used for monitoring, storing, | |||
and reviewing of, and to | |||
generate alarms for, multiple | |||
physiological parameters of | |||
adults, pediatrics and neonates. | |||
The monitors are intended for |
Table 1Comparison to predicates
5
K180380 Page 3 of 11
| | and neonates. The monitor is
intended for use by trained
healthcare professionals in
hospital environments.
Monitored parameters include:
NIBP, SpO2, PR (pulse rate),
Quick TEMP/Infrared TEMP.
The monitor is not intended
for MRI environments. TEMP
module is not intended for
neonates. | This monitor is used to
monitor vital signals for
patients and is suitable for use
in hospital
environments including
out-patient department, wards
and NICU.
Monitored parameters include:
NIBP, SpO2, pulse rate, Quick
TEMP/Infrared TEMP. | use by trained healthcare
professionals in hospital
environments.
The monitored physiological
parameters include: ECG,
respiration (RESP), temperature
(TEMP), oxygen saturation of
arterial blood (SpO2), pulse rate
(PR), non-invasive blood
pressure (NIBP), invasive blood
pressure (IBP), carbon dioxide
(CO2), cardiac output (C.O.),
anesthetic gas (AG), bispectral
index (BIS), respiration
mechanics (RM) and impedance
cardiography (ICG). |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended patient
population | adult, pediatric and neonatal
patients | adult, pediatric and neonatal
patients | Same |
| Intended
application
environment | Hospital environment. | Hospital environment. | Same |
NIBP monitor (EDAN Module)
| Principle of
Operation | Oscillation | Oscillation | |
---|---|---|---|
Measurement | |||
Range | Adult | ||
Systolic 40 to 270 mmHg | |||
Diastolic 10 to 215 mmHg | |||
Mean 20 to 235 mmHg |
Pediatric
Systolic 40 to 230 mmHg
Diastolic 10 to 180 mmHg
Mean 20 to 195 mmHg
Neonate
Systolic 40 to 135 mmHg
Diastolic 10 to 100 mmHg | Adult
Systolic 40 to 270 mmHg
Diastolic 10 to 215 mmHg
Mean 20 to 235 mmHg
Pediatric
Systolic 40 to 200 mmHg
Diastolic 10 to 150 mmHg
Mean 20 to 165 mmHg
Neonate
Systolic 40 to 135 mmHg
Diastolic 10 to 100 mmHg | |
| | | | |
| | Mean 20 to 110 mmHg | Mean 20 to 110 mmHg | |
| Accuracy | Maximum average error: ±5 mmHg | Maximum average error: ±5 mmHg | |
| | Maximum standard deviation: 8 mmHg | Maximum standard deviation: 8 mmHg | |
| | | | |
| Resolution | 1 mmHg | 1 mmHg | |
| Maximum
measuring period | Adult/Pediatric 120s
Neonate 90s | Adult/Pediatric 120s
Neonate 90s | |
| | Adult 297±3 mmHg | Adult 297±3 mmHg | |
| Overpressure
protection | Pediatric 245±3 mmHg | Pediatric 240±3 mmHg | |
| | Neonate 147±3 mmHg | Neonate 147±3 mmHg | |
| Mode | Manual, Auto, Continuous | Manual, Auto, Continuous | |
| Measuring interval
in AUTO Mode | 1/2/3/4/5/10/15/30/60/90/120/
180/240/360/480 mins | 1/2/3/4/5/10/15/30/60/90/120/
240/480 mins | |
| Continuous | 5min | 5min | |
| PR from NIBP | | | |
| Measurement range | 40 to 240 bpm | 40 bpm to 240 bpm | |
| Accuracy | ±3 bpm or ±3.5%,
whichever is greater | ±3 bpm or 3.5%,
whichever is greater | |
| | | | |
| Resolution | 1 bpm | 1 bpm | |
| NIBP monitor(Suntech Module) | | | |
| Principle of
Operation | oscillation | | Same |
| | | | |
| | Adult
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg- 220 mmHg | | Same |
| | Pediatric
SYS: 40 mmHg- 230 mmHg
DIA: 20 mmHg- 160 mmHg
MAP: 26 mmHg– 183 mmHg | | |
| Measurement
Range | Neonate
SYS: 40 mmHg – 130 mmHg
DIA: 20 mmHg– 100 mmHg
MAP: 26 mmHg – 110 mmHg | | |
| | | | |
| Accuracy | Maximum average error: | | Same |
| | | | |
| | ±5mmHg | | |
| | Maximum standard deviation: | | |
| | 8mmHg | | |
| Resolution | 1mmHg | | Same |
| | Adult 130s | | Same |
| Maximum
measuring period | Pediatric 120s | | |
| | Neonate 75s | | |
| Overpressure
protection | Adult /Pediatric Saturation
Adult/pediatric, non-motion
conditions
70 to 100% ±2 %
0-69% unspecified
Neonate
70 to 100% ±3%
0-69% unspecified
Pulse Rate
Adult and Neonate
25 to 300 bpm ±2bpm
(non-motion conditions) | Saturation
Adult/pediatric, non-motion
conditions
70 to 100% ±2 %
0-69% unspecified
Neonate
70 to 100% ±3%
0-69% unspecified
Pulse Rate
Adult and Neonate
25 to 300 bpm ±2bpm
(non-motion conditions) | |
| Resolution | SpO2 1%
Pulse Rate 1 bpm | SpO2 1%
Pulse Rate 1 bpm | |
| Sensor | Emitted Light Energy: ≤15 mW | Emitted Light Energy: ≤15 mW | |
| SpO2 monitor (NELLCOR Module) | | | |
| Optional Spo2 | NELLCOR SpO2 module | NELLCOR SpO2 module | |
| module | | | |
| TEMP
monitor(EDAN
quick temp
module) | | | |
| Measuring range | Monitor mode: 25 °C ~45 °C
Predict mode: 35.5 °C ~42 °C | Monitor mode: 25 °C ~45 °C
Predict mode: 35.5 °C ~42 °C | |
| Working
temperature | 10°C ~ 40°C | 10°C ~ 40°C | |
| Sensor type | Oral /axillary /rectal | Oral /axillary /rectal | |
| Adjustable range of
alarm limits | 35.5 °C ~42 °C | 35.5 °C ~42 °C | |
| Resolution | 0.1 °C | 0.1 °C | |
| Accuracy | Monitor mode: ±0.1 °C
(25 °C ~ 45 °C ) | Monitor mode: ±0.1 °C
(25 °C ~ 45 °C ) | |
| Response time |