K Number
K131818
Date Cleared
2013-08-30

(71 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vital Signs Monitor models M3 and M3A (hereinafter called monitor) is intended to be used for non-invasive continuous monitoring of SpO2 (oxygen saturation of the blood), NIBP (non-invasive blood pressure) and TEMP (temperature).

The monitor is intended to be used only under regular supervision of clinical personnel. It is applicable to adult, pediatric, and neonatal usage in hospitals, hospital type facilities and intra-hospital moves.

The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

Device Description

The Vital Signs Monitor is a portable device intended for use by health care professionals. The monitor could provide the monitoring of physiological parameters such as the non-invasive blood pressure (NIBP), the oxygen saturation of the blood (SpO2) and Rectal/Oral/Auxiliary/Ear temperature measurement.

The following lists the detailed features of the subject device.

  • LCD or LED display
  • SpO2, Pulse Rate NIBP and TEMP measurement
  • T2 or TH or F3000 module for Temp measurement
  • Nellcor or EDAN SpO2 module
  • Display numeric and waveform information simultaneously
  • Nurse call feature
  • Built-in Lithium-ion Battery
  • Suitable for adult, pediatric and neonate patients
  • Visual and audible alarm
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Edan Instruments Vital Signs Monitor Models M3 and M3A:

1. Table of Acceptance Criteria and Reported Device Performance

Based on the provided text, specific numerical acceptance criteria and corresponding reported device performance values are not explicitly stated in a clear, tabulated format for the SpO2, NIBP, or TEMP measurements. The submission focuses on substantial equivalence to predicate devices and adherence to general safety standards rather than detailed performance metrics.

The document states:

  • "The differences between the subject devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject devices. And no question is raised relating to the safety and effectiveness." (Page 3)
  • "Verification and validation testing were conducted on the Vital Signs Monitor Models M3 and M3A. This premarket notification submission demonstrates that Vital Signs Monitor Models M3 and M3A is substantially equivalent to the predicate device, and without any changes of the intended use compared to the predicate version." (Page 3)

The key acceptance criterion mentioned is substantial equivalence to predicate devices (Edan Instruments Inc.: M3 and M3A Vital Signs Monitor K120144, and The Kendall Company: K003313). The performance is implied to be at least equivalent to these predicate devices.

The only specific performance testing mentioned with regard to a standard is:

  • "Clinical Electronic Thermometers testing for the newly added F3000 Temperature Module pursuant to the guidance 'Guidance on the Content of Premarket Notification [510(K)] Submission for Clinical Electronic Thermometers' dated Mar. 1993." (Page 2)

Without the detailed results of this thermometer testing, a quantitative table cannot be constructed from the provided text.

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not explicitly state a specific sample size used for a test set or the data provenance (e.g., country of origin, retrospective/prospective).

The document mentions "Verification and validation testing were conducted," but details on these tests, including sample sizes, subjects, or data sources, are absent. The "Clinical test: Clinical testing is not required" statement indicates that no human subject clinical trials were deemed necessary for this 510(k) submission, relying instead on non-clinical testing and equivalence to predicates.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Given that "Clinical testing is not required," the concept of "ground truth established by experts" for a test set using patient data is not applicable in this context. The evaluation was primarily based on non-clinical performance against standards and comparison to predicate devices.

4. Adjudication Method for the Test Set

As there was no explicit clinical test set requiring expert ground truth, an adjudication method is not applicable and not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states "Clinical testing is not required" and focuses on substantial equivalence and non-clinical testing. Therefore, there is no discussion of human reader improvement with or without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

The device is a vital signs monitor, which is inherently a standalone measurement device that provides physiological parameters (SpO2, NIBP, TEMP). It collects and displays data directly. The provided text, however, focuses on its safety and performance relative to predicate devices and standards, rather than a specific "standalone performance study" in the context of an AI algorithm. If we interpret "standalone" as the device operating independently to deliver its intended measurements, then the entire submission implicitly supports its standalone performance. However, there's no separate study specifically labeled as "standalone performance" for an AI algorithm, as this device does not appear to primarily be an AI-driven diagnostic.

7. Type of Ground Truth Used

The primary "ground truth" implicitly relied upon for this submission is:

  • Standards: Adherence to established non-clinical electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and thermometer testing standards.
  • Predicate Device Performance: The demonstrated safety and effectiveness of the legally marketed predicate devices (K120144 and K003313). The new device's performance is asserted to be "the same or similar performance specifications" as these predicates.

8. Sample Size for the Training Set

The provided text does not mention or imply the existence of a training set. This is consistent with the device being a vital signs monitor primarily evaluated through non-clinical testing and comparison to predicates, rather than a machine learning or AI-driven diagnostic device that would typically involve a training set.

9. How the Ground Truth for the Training Set Was Established

Since there is no mention of a training set, there is no information on how its ground truth would have been established.

{0}------------------------------------------------

510(K)Summary of Safety and Effectiveness

This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92

1. Submitter:Edan Instruments, Inc3/F - B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, shekou, NanshanShenzhen, 518067 P.R. ChinaTel: 86-755-26858736Fax:86-755-26882223
Contact person:Queena ChenAUG 30 2013
Prepare Date:Jun 13th, 2013
2. Proprietary Name:Vital Signs Monitor Models M3 and M3A
Classification Name:21 CFR 870.1130 Noninvasive blood pressure measurement system21 CFR 870.2700, Oximeter21 CFR 880.2910 Clinical electronic thermometer
Product code:DQA, DXN, FLL
Classification:Class II
3. Predicate Devices:· Edan Instruments Inc.: M3 and M3A Vital Signs Monitor K120144· The Kendall Company: K003313
4. Device Description:The Vital Signs Monitor is a portable device intended for use by health careprofessionals. The monitor could provide the monitoring of physiologicalparameters such as the non-invasive blood pressure (NIBP), the oxygensaturation of the blood (SpO2) and Rectal/Oral/Auxiliary/Ear temperaturemeasurement.
The following lists the detailed features of the subject device.
· LCD or LED display· SpO2, Pulse Rate NIBP and TEMP measurement· T2 or TH or F3000 module for Temp measurement· Nellcor or EDAN SpO2 module
  • . Display numeric and waveform information simultaneously
  • . Nurse call feature

{1}------------------------------------------------

RE: K131818 Vital Sings Monitor M3 & M3A 003_ Special 510(k) Summary of Safety and Effectiveness

  • . Built-in Lithium-ion Battery
  • . Suitable for adult, pediatric and neonate patients
  • . Visual and audible alarm
  1. Intended Use: The Vital Signs Monitor models M3 and M3A (hereinafter called monitor) is intended to be used for non-invasive continuous monitoring of SpO2 (oxygen saturation of the blood), NIBP (non-invasive blood pressure) and TEMP (temperature).

The monitor is intended to be used only under regular supervision of clinical personnel. It is applicable to adult, pediatric, and neonatal usage in hospitals, hospital type facilities and intra-hospital moves.

The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both

6. Effectiveness and Safety Considerations:

Clinical test: Clinical testing is not required.

Non-clinical test:

The following safety standards are conducted on the subject device based on the predicate version:

(1) IEC 60601-1 Electrical Safety

(2) IEC 60601-1-2 Electromagnetic Compatibility

(3) Clinical Electronic Thermometers testing for the newly added F3000 Temperature Module pursuant to the guidance "Guidance on the Content of Premarket Notification [510(K)] Submission for Clinical Electronic Thermometers" dated Mar. 1993.

(4) IEC 60601-2-49 Particular requirements for the safety of multifunction patient monitoring equipment

7. Comparison to predicate device

The M3 and M3A have the following similarities to devices which have previously received 510(k) concurrences:

  • . the same intended use
  • use the similar operating principle
  • . use the same modules, including SpO2, NIBP and Quick TEMP (including infrared Ear TEMP)
  • have the same or similar performance specifications .

The differences between the subject devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject devices. And no question is raised relating to

{2}------------------------------------------------

RE: K131818 Vital Sings Monitor M3 & M3A 003_ Special 510(k) Summary of Safety and Effectiveness

the safety and effectiveness. In summary, the M3 and M3A Vital Signs Monitor described in this submission is, in our opinion, substantially equivalent to the predicate devices.

8. Substantial Equivalent Conclusion:

Verification and validation testing were conducted on the Vital Signs Monitor Models M3 and M3A. This premarket notification submission demonstrates that Vital Signs Monitor Models M3 and M3A is substantially equivalent to the predicate device, and without any changes of the intended use compared to the predicate version.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: the department's seal and the department's symbol. The seal is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. The symbol is a stylized representation of a human figure embracing a globe, symbolizing the department's mission to protect and promote the health and well-being of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6119 Silver Spring, MD 20993-0002

August 30, 2013

Edan Instruments, Inc. Queena Chen Certification Engineer 3/F - B. Nanshan Medical Equipments Park Nanhai Rd #1019 Shekou, Nanshan Shenzhen 518067 P.R. China

Re: K131818

Trade/Device Name: Vital signs monitor models M3 and M3A Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter, Blood-Pressure. Non-Nvasive Regulatory Class: Class II Product Codc: DQA, DXN, FLL Dated: July 26, 2013 Received: July 31, 2013

Dear Queena Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

OweFDAaris-S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K131818

Device Name: Vital Signs Monitor models M3 and M3A

The Vital Signs Monitor models M3 and M3A (hereinafter called monitor) is intended to be used for non-invasive continuous monitoring of SpO2 (oxygen saturation of the blood), NIBP (non-invasive blood pressure) and TEMP (temperature).

The monitor is intended to be used only under regular supervision of clinical personnel. It is applicable to adult, pediatric, and neonatal usage in hospital type facilities and intra-hospital moves.

The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

× Prescription Use (21 CFR Part 801 Subpart D)

AND/OR Over the Counter Use _ . (21 CFR Part 801 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).