(282 days)
LensHooke X1 Semen Quality Analyzer for OTC use
The LensHooke X1 Semen Quality Analyzer used with LensHooke Semen is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
-Sperm concentration (10^6 per ml) -Total motility (PR+NP, %)
-Sperm morphology (normal forms, %)
-pH value
The LensHooke X1 Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status. It is a self-testing, in-vitro diagnostic system intended for human semen analysis of individuals at home to evaluate male fertility. The systems are intended for single person use only and should not be shared.
LensHooke X1 PRO Semen Quality Analyzer for prescription use
The LensHooke X1 PRO Semen Quality Analyzer used with LensHooke Semen Test Cassette is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
(1) Sperm concentration (10^6 per ml)
(2) Total motility (PR+NP, %)
-
Progressive motility (%)
-
Non-Progressive motility (%)
-
(3) Sperm morphology (normal forms, %)
(4) pH value
The LensHooke X1 PRO Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in healthcare professional setting to evaluate male fertility.
Semen Quality Analyzer integrates optical design and image analysis and combined with artificial intelligence image processing method, to fully automated analysis of semen quality including semen pH, sperm concentration and motility. The images are captured and recorded by cameras and with image processing methods, the locations of sperms are detected. The sperm concentration is analyzed by the sperm unit density; the sperm motility is calculated by tracing sperm trajectories and the sperm morphology is calculated by comparing head and tail percentage. Through camera, the chromatographic image of pH is captured and with image saturation and brightness analysis, the level of pH is determined.
Here's a breakdown of the acceptance criteria and the study details for the LensHooke X1 Pro Semen Quality Analyzer and LensHooke X1 Semen Quality Analyzer, based on the provided document:
Acceptance Criteria and Reported Device Performance
The document describes various non-clinical tests to establish performance, functionality, and reliability. The "performance" in the table refers to how well the device performed against the established acceptance criteria in the studies mentioned.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance/Conclusion |
|---|---|---|
| Repeatability | Calculated CV (%) for pH, Concentration, Motility, and Morphology within 10% | Within 10%: "It was demonstrated that the calculate CV (%) pH, Concentration, Motility and morphology were within 10%." |
| Reproducibility | Calculated CV (%) for pH, Concentration, Motility, and Morphology within 10% | Within 10%: "It was demonstrated that the calculate CV (%) pH, Concentration, Motility and morphology were within 10%." |
| LoB/LoD/LoQ (Semen) | LoB, LoD, LoQ established at specific concentrations | Established: LoB: 0 x 10^6/mL, LoD: 1.8 x 10^6/mL, LoQ: 7.2 x 10^6/mL. (Note: The document uses "10%mL" which likely means 10^6/mL). |
| Linearity (Semen) | Mean slope above 0.95 and R^2 values above 0.99 | Linear from 2 to 350 x 10^6/mL: "Analysis of results indicated Semen linear regression between device and reference method shows mean slope above 0.95 and R2 values above 0.99. The results support the claim that the candidate assay is linear from 2 to 350 x10^6/mL concentrations." |
| Linearity (pH) | Linear from pH 5.8 to 8.2 | Linear from pH 5.8 to 8.2: "The results support the claim that the pH is linear from pH 5.8 to 8.2." |
| Interference | Meet CLSI: EP7-A2 acceptance criteria | No significant interference: "The study results indicate that all tested 11 interference substances meet CLSI: EP7-A2 acceptance criteria for therapeutic levels of concentration. Tested substances do not cause significant interference." |
| Sample Volume (Semen) | CV within 5% and bias within 20% for semen concentration | Within criteria for 35-45 µL: "The study results demonstrate the CV is within 5% and the bias is within 20% for semen concentration. All results within criteria when specimen is 35-45 uL. Thus, the minimum sample volume is 35 uL." |
| Sample Volume (pH) | CV within 10% and bias within 10% | Within criteria for 30-80 µL: "The study results demonstrate the CV is within 10% and the bias is within 10%. All results within criteria when specimen is 30-80 uL. Thus, the minimum sample volume is 30 µL." |
| Operating Temperature & Humidity (Semen) | Bias within 10% | Operates under 15-38°C and 50-90% RH: "The results indicate basis within 10% and conclude that candidate device can operate under the environment conditions 15-38°C (59-100.4°F) and 50-90% relative humidity." |
| Operating Temperature & Humidity (pH) | Bias within 10% | Operates under 15-38°C and 50-90% RH: "The results indicate basis within 10% and conclude that candidate device can operate under the environment conditions 15-38°C (59-100.4°F) and 50-90% relative humidity." |
| Test Strip Stability (Closed Vial) | Stable test cassette strip performance for estimated shelf life | Estimated 2-year shelf life: "The study results should demonstrate stable test cassette strip performance for 8 months under accelerated 40°C condition, which can estimate shelf life of test strips for 2 years at room temperature." (The document states "should demonstrate," implying this was the goal or expectation based on the accelerated conditions, not necessarily a direct report of completion.) |
| System Accuracy & Layuser Performance | Layperson user accuracy and ease of use (via questionnaire) | Demonstrated accuracy and ease of use: "The study results demonstrate that the layperson user accuracy and ease of use (via participant questionnaire scoring) of Candidate device." Excluded findings were addressed in the study report. |
Study Information:
2. Sample size(s) used for the test set and the data provenance:
- Repeatability: 60 data points per sperm concentration/sperm motility/sperm morphology level (from 3 operator/analyzer/cassette combinations x 2 replicates x 2 runs x 5 times/day). The exact number of unique samples is not specified, but they were tested multiple times.
- Reproducibility:
- Control solutions (pH, Concentration): 3 prepared latex beads concentration and pH levels, each tested in 5 replicates by 1 operator per day, across 3 lots of test system, with 1 lot per site. The total number of data points per level would be 3 (levels) * 5 (replicates) * X (days/operators) * 3 (lots/sites). The document specifies "3 lots of test system were used and 1 lot per site," suggesting 3 sites.
- Motility and Morphology: 3 semen motile levels and 3 semen morphology levels, evaluated through 3 operator/analyzer/cassette lot combinations x 5 replicates x 5 times/day.
- LoB/LoD/LoQ: Two concentration levels, assayed in 5 replicates, with each of two lots.
- Linearity (Semen): Semen samples at 9 intervals (2 to 400 x 10^6/mL), tested in 3 replicates per lot per concentration level, using 1 analyzer and 3 lots of Cassettes.
- Linearity (pH): Semen samples at 13 pH intervals (5.8 to 8.2), tested in 3 replicates per lot per pH level, using two analyzers and 3 lots of Cassettes.
- Interference: 2 concentration levels of semen samples (50-100 and 100-200 x 10^6/mL) with 11 interference substances. 1 analyzer and 3 lots of test cassettes were used to perform 5 replicates per lot.
- Sample Volume (Semen): 3 lots of test cassette, semen samples at low and high concentrations, evaluated at 3 sample volumes (35, 40, 45 uL). Tested in 5 replicates.
- Sample Volume (pH): 3 lots of pH test, semen sample at pH 7.8, evaluated at 6 sample volumes (30, 40, 50, 60, 70, 80uL). Tested in 3 replicates per analyzer, 6 replicates in total.
- Operated Temperature and Humidity (Semen): 6 concentration levels of Semen samples, tested in 5 replicates.
- Operated Temperature and Humidity (pH): 3 pH levels of Semen samples, tested in 5 replicates.
- Test Strip Stability (Closed Vial): 3 lots of Test Cassette, and 3 Semen samples, test measurements in 3 replicates.
- System Accuracy Study and Layuser Performance Study: 106 subjects (with 100 tabulated findings).
Data Provenance: The document does not explicitly state the country of origin for the data or if it was retrospective or prospective. However, given Bonraybio Co., Ltd. is based in Taiwan and the studies involve human subjects ("subjects", "lay users"), it is likely the studies were prospective to gather new data for regulatory submission. The predicate device (SQA V) is from "Medical Electronic System, LLC," which is likely a US-based company.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For the non-clinical performance studies (repeatability, reproducibility, linearity, etc.), the reference method was the SQA-V Analyzer (for semen parameters) and a pH meter (for pH). These are established laboratory methods/devices.
- For the Layuser Performance Study, the reference method was "SQA-V Analyzer performed by POC personnel." The "POC personnel" likely refers to healthcare professionals but their specific qualifications (e.g., years of experience, type of professional) are not detailed. It mentions "Point-of-Care professionals or licensed registered nurses" in the context of the study's objective, implying these are the types of individuals who established ground truth when the SQA-V was used for reference.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
The document does not describe an adjudication method for reconciling differences in ground truth establishment. The ground truth was based on comparison to an existing predicate device (SQA-V Analyzer) and a standard laboratory instrument (pH meter).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not explicitly described in the provided text in the context of human readers improving with AI vs. without AI assistance.
- The study involved comparing the new device's performance to a predicate device (SQA-V Analyzer) and a pH meter.
- The "Layuser Performance Study" did involve human users (lay users and POC professionals) reading and using the device, but it focused on the device's accuracy and ease of use, not on measuring human performance improvement with or without AI assistance from the device. The device itself utilizes "artificial intelligence image processing" (page 5), but the studies described are device performance and user studies, not AI assistance (human-in-the-loop) comparative studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance of the algorithm (as embedded in the device) was done. All the non-clinical tests (Repeatability, Reproducibility, LoB/LoD/LoQ, Linearity, Interference, Sample Volume, Operated Temperature and Humidity, Test Strip Stability) evaluate the device's (which includes its embedded AI/image processing algorithms) performance directly against reference methods without a human interpretation step between the device's output and the final result. The device "fully automated analysis of semen quality including semen pH, sperm concentration and motility" (page 5).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Primarily comparison to an established laboratory reference method/device: The SQA-V Analyzer (for sperm concentration, motility, and morphology) and a standard pH meter (for pH). These are considered highly reliable and validated methods/devices.
8. The sample size for the training set:
The document does not specify the sample size for the training set used for the device's "artificial intelligence image processing method." It primarily focuses on the validation/test sets for demonstrating performance against a predicate.
9. How the ground truth for the training set was established:
The document does not provide information on how the ground truth for the training set was established. While it states the device uses "artificial intelligence image processing method," it does not delve into the development or training process of this AI, nor the ground truth used therein.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2018
Bonraybio Co., Ltd. % Feng-Yu Lee Principal Consultant Dynamic Biotech Inc. dba IVDD Regulatory Consultant 29222 Rancho Viejo Road, Suite 218 San Juan Capistrano, California 92675
Re: K180343
Trade/Device Name: LensHooke X1 Pro Semen Quality Analyzer, LensHooke X1 Semen Quality Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: POV Dated: January 31, 2018 Received: February 7, 2018
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Takeesha Taylor-bell -S
Takeesha Taylor-Bell Branch Chief Hematology Branch Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
For
Leonthena Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K180343
Device Name
LensHooke X1 Semen Quality Analyzer LensHooke X1 PRO Semen Quality Analyzer LensHooke Semen Test Cassette
Indications for Use (Describe) LensHooke X1 Semen Quality Analyzer for OTC use
The LensHooke X1 Semen Quality Analyzer used with LensHooke Semen is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
-Sperm concentration (10^6 per ml) -Total motility (PR+NP, %)
-Sperm morphology (normal forms, %)
-pH value
The LensHooke X1 Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status. It is a self-testing, in-vitro diagnostic system intended for human semen analysis of individuals at home to evaluate male fertility. The systems are intended for single person use only and should not be shared.
LensHooke X1 PRO Semen Quality Analyzer for prescription use
The LensHooke X1 PRO Semen Quality Analyzer used with LensHooke Semen Test Cassette is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
(1) Sperm concentration (10^6 per ml)
(2) Total motility (PR+NP, %)
-
Progressive motility (%)
-
Non-Progressive motility (%)
-
(3) Sperm morphology (normal forms, %)
(4) pH value
The LensHooke X1 PRO Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status. It is an in-vitro diagnostic system intended for human semen analysis of individuals in healthcare professional setting to evaluate male fertility.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "bonraybio". The logo consists of a stylized teal-colored plant icon on the left, followed by the company name "bonraybio" in a sans-serif font. The font is a light gray color.
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K180343
1. Submitter's Identification:
bonraybio Co., Ltd 4F., No.118, Gongye 9th Rd., Dali Dist., Taichung City 41280, Taiwan (R.O.C.) Contact Person: Clare Huang Phone Number: 886-4-2491-2385 FAX Number: 886-4-2491-2885
c/o IVDD Regulatory Consultant 29222 Rancho Viejo Road, Suite 218 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-949-218-0928
Date Summary Prepared: November 16tt, 2018
-
- Name of the Device: LensHooke Semen Test Cassette LensHooke X1 Semen Quality Analyzer LensHooke X1 PRO Semen Quality Analyzer
-
- Common or Usual Name: Semen Analysis Device
| Product Code | Classification | Regulation Section | Panel |
|---|---|---|---|
| POV;Semen Analysis Device | Class II | 21 CFR 864.5220 | Hematology80 |
4. Device Description:
Semen Quality Analyzer integrates optical design and image analysis and combined with artificial intelligence image processing method, to fully automated analysis of semen quality including semen pH, sperm concentration and motility. The images are captured and recorded by cameras and with image processing methods, the locations of sperms are detected. The sperm concentration is analyzed by the sperm unit density; the sperm motility is calculated by tracing sperm trajectories and the sperm morphology is calculated by comparing head and tail percentage. Through camera, the chromatographic
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Bonraybio. The logo consists of a teal circle with a white wheat stalk inside, followed by the text "bonraybio" in a modern, sans-serif font. The text is in a dark gray color, providing a clean and professional look.
image of pH is captured and with image saturation and brightness analysis, the level of pH is determined.
Product Information
- 4.1 For Over-the-Counter Setting:
LensHooke X1 Semen System, consist of the following devices: LensHooke X1 Semen Quality Analyzer LensHooke Semen Test Cassette C-KUP Liquefaction Test Cup LensHooke Cleaning Wipe
The LensHooke X1 Semen Quality Analyzer and LensHooke Semen Test Cassette are manufactured by bonrayBio.
4.2 For Point-of-Care Professional Setting:
LensHooke X1 PRO Semen System, consist of the following devices: LensHooke X1 PRO Semen Quality Analyzer and LensHooke Semen Test Cassette. C-KUP Liquefaction Test Cup LensHooke Cleaning Wipe
The LensHooke X1 PRO Semen Quality Analyzer and LensHooke Semen Test Cassette are manufactured by Bonraybio.
- 4.3 Consumables Description
LensHooke Semen Test Cassette
LensHooke Semen Test Cassette is a well-designed microscopic slide for the optical analyzer, LensHooke Semen Quality Analyzer. Top and bottom plastic case and pH paper are the components of LensHooke Semen Test Cassette. There are two polished windows which analyzed concentration, motility and morphology of the semen and the pH of semen respectively.
C-KUP Liquefaction Test Cup
C-KUP Liquefaction Test Cup is used to collecting semen samples to liquefaction and volume testing. Collected semen samples are applicable for semen quality analysis. Cup, cup cover and drip cover are the components of C-KUP Liquefaction Test Cup. The V-Stick on cup cover is used to check the liquefaction's status. The Scale on cup is used to check the volume of the semen sample.
LensHooke Cleaning Wipe
LensHooke Cleaning Wipe is a plastic stick with lens cotton. Using LensHooke Cleaning Wipe to clean the Test Cassette Insert Slot of LensHooke Semen Quality Analyzer. This is the cleaning and maintenance procedures usually used for microscopic analyzers.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Bonraybio. The logo consists of a teal-colored circular icon with a stylized wheat stalk inside, followed by the company name "bonraybio" in a modern, sans-serif font. The text is in a dark gray color, providing a clean and professional look.
5. Indications for Use:
- 5.1 For Over-the-Counter Setting:
LensHooke X1 Semen Quality Analyzer
The LensHooke X1 Semen Quality Analyzer used with LensHooke Semen Test Cassette is an optical device for human semen analysis which provides direct and calculated quantitative measurements for:
-Sperm concentration (10° per ml)
- -Total motility (PR+NP. %)
-Sperm morphology (normal forms, %)
-pH value
The LensHooke X1 Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status.
It is a self-testing, in-vitro diagnostic systems intended for human semen analysis of individuals at home to evaluate male fertility. The systems are intended for single person use only and should not be shared.
-
5.2 For Point-of-Care Professional Setting:
The LensHooke X1 PRO Semen Quality Analyzer used with LensHooke Semen Test Cassette is an optical device for human semen analysis which provides direct and calculated quantitative measurements for: -
(1) Sperm concentration (10° per ml)
-
(2) Total motility (PR+NP, %)
- Progressive motility (%)
- Non-Progressive motility (%)
-
(3) Sperm morphology (normal forms, %)
-
(4) pH value
The LensHooke X1 PRO Semen Quality Analyzer does not provide a comprehensive evaluation of a male's fertility status.
It is an in-vitro diagnostic system intended for human semen analysis of individuals in healthcare professional setting to evaluate male fertility.
- Predicate Device Information: 6.
LensHooke X1 and X1 PRO Semen Quality Analyzers are substantially equivalent to: SQA V, Sperm Quality Analyzer - Visual Device Company: Medical Electronic System, LLC 510(k) Number: K021746
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for "bonraybio". The logo consists of a teal circle with a white plant inside of it, followed by the name "bonraybio" in a sans-serif font. Below the logo is the text "CED".
7. Comparison to Predicate Devices:
| ProductName | LensHooke X1 PROSemen Quality Analyzer(Professional) | LensHooke X1 SemenQuality Analyzer(OTC) | Predicate (K021746)SQA-V (MES) |
|---|---|---|---|
| Intended Use | The LensHooke X1 PROSemen Quality Analyzerused with LensHookeSemen Test Cassette is anoptical device for humansemen analysis whichprovides direct andcalculated quantitativemeasurements for:(1) Sperm concentration(106 per ml)(2) Total motility (PR+NP,%)- Progressive motility (%)- Non-Progressivemotility (%)(3) Sperm morphology(normal forms, %)(4) pH valueThe LensHooke X1 PROSemen Quality Analyzerdoes not provide acomprehensive evaluationof a male's fertility status.It is an in-vitro diagnosticsystem intended forhuman semen analysis ofindividuals in healthcareprofessional setting toevaluate male fertility. | The LensHooke X1Semen Quality Analyzerused with LensHookeSemen Test Cassette is anoptical device for humansemen analysis whichprovides direct andcalculated quantitativemeasurements for:-Sperm concentration(106 per ml)-Total motility (PR+NP,%)-Sperm morphology(normal forms, %)-pH valueThe LensHooke X1Semen Quality Analyzerdoes not provide acomprehensiveevaluation of a male'sfertility status. It is a self-testing, in-vitrodiagnostic systemsintended for humansemen analysis ofindividuals at home toevaluate male fertility.The systems are intendedfor single person use onlyand should not be shared. | The SQA V is a point-of-care,electro-optical device with on-screen visualization that is used forsemen analysis.The SQA V reports:• total sperm concentration(TSC, millions/ml)• percent motility (%MOT) and %progressive motility (%PMOT)• % normal morphology(%MORPH)• motile sperm concentration(MSC, millions/mL) andprogressive MSC (PMSC)• functional sperm concentration(FSC, millions/mL) |
| Male FertilityFactor | Yes | Yes | |
| Technology | Desk-top unit consists of light sources, built-in videomicroscopy and an internal computer containingalgorithms for the assessment of semen parameters. | Desk-top unit consists of a lightsource, optical sensors, built-invideo microscopy and an internalcomputer containing algorithms forthe assessment of semenparameters. | |
| Transmissioninterface | HDMI/USB | Bluetooth/Wi-Fi | N/A |
| Intended User | Point-of-Care professional | Over-the-Counter | Professional |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for "bonraybio". The logo consists of a stylized teal-colored plant icon on the left, followed by the company name "bonraybio" in a simple, sans-serif font. The color of the text is a dark gray or black, providing a contrast against a white background.
Discussion of Non-Clinical Tests Performed for Determination of Substantial 8. Equivalence is as follows:
Verification and validation of test results were evaluated to establish the performance, functionality and reliability of LensHooke X1 Semen Quality Analyzer and LensHooke X1 PRO Semen Quality Analyzer.
The evaluation included Repeatability, Reproducibility, LoB/LoD/LoQ and linearity, interference, sample volume, operating conditions and stability in close vial.
Repeatability
The study was conducted two replicates per run, and two runs for every 4 hours, five times/day (3 operator/analyzer/cassette lot combinations × 2 replicates × 2 runs × 5 times/day = 60 data points per sperm concentration/sperm motility/sperm morphology level).
It was demonstrated that the calculate CV (%) pH, Concentration, Motility and morphology were within 10%.
Reproducibility
The evaluation was estimated by control solutions of 3 prepared latex beads concentration and pH levels (Label #Low, Normal, High). Each sample was tested in 5 replicates by 1 operator per day. 3 lots of test system were used and 1 lot per site.
For motility and morphology, due to the limited stability of semen samples, the "day" of repeatability experiment as different times of the day (every 4 hours = 1 "day"). 3 semen motile and 3 semen morphology levels (Label #Low, Normal, High) spread across the measuring interval were evaluated 3 operator/analyzer/cassette lot combinations x 5 replicate × 5 times/day.
Total 3 POC sites with 5 days/times were performed.
It was demonstrated that the calculate CV (%) pH, Concentration, Motility and morphology were within 10%.
LoB/LoD/LoQ
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Bonraybio. The logo consists of a circular icon on the left and the company name "bonraybio" on the right. The icon is a light blue circle with a white plant-like design inside. The text is in a sans-serif font and is a dark gray color.
The limit of blank and detection limits were established by assaying two concentration levels, in 5 replicates, with each of the two lots. The concentration levels were LoB: 0 10%mL, LoD: 1.8 10%mL, and LoQ: 7.2 10%/mL.
Linearity and Sensitivity
The linearity and sensitivity study for Semen was performed using one analyzer and 3 lots of Cassettes to evaluate detection range. Semen samples were prepared at 9 semen intervals ranged from low to elevated level (2 to 400 x 10%mL).
Test procedures were performed in 3 replicates per lot per concentration level and compared with SQA-V analyzer.
The linearity and sensitivity study for pH was performed using two analyzers and 3 lots of Cassettes to evaluate detection range. Semen samples were prepared at 13 pH intervals ranged from pH 5.8 to 8.2.
Test procedures were performed in 3 replicates per lot per pH level and compared with pH meter.
Analysis of results indicated Semen linear regression between device and reference method shows mean slope above 0.95 and R2 values above 0.99. The results support the claim that the candidate assay is linear from 2 to 350 x106/mL concentrations. The results support the claim that the pH is linear from pH 5.8 to 8.2.
Interference
The interference study was performed using 2 concentration levels (50-100 and 100-200 x 10%mL) of semen samples and 11 interference substances. 1 analyzer and 3 lots of test cassettes were used to perform 5 replicates per lot and compared to reference method SQA- V Analyzer.
The study results indicate that all tested 11 interference substances meet CLSI: EP7-A2 acceptance criteria for therapeutic levels of concentration. Tested substances do not cause significant interference.
Sample Volume
Semen Study
The specimen volume study was performed using 3 lots of test cassette, semen samples at low and high concentrations were used to evaluate sample volumes at 35, 40 and 45 uL. Test procedures were performed in 5 replicates, and results were compared to reference method SQA-V Analyzer.
The study results demonstrate the CV is within 5% and the bias is within 20% for semen concentration. All results within criteria when specimen is 35-45 uL. Thus, the minimum sample volume is 35 uL.
pH Study
The specimen volume study was performed using 3 lots of pH test, semen sample at pH 7.8 was used to evaluate sample volumes at 30, 40, 50, 60, 70 and 80uL. Test procedures
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Bonraybio. The logo consists of a teal-colored circular icon with a stylized wheat stalk inside, followed by the word "bonraybio" in a lowercase, sans-serif font. The text is a dark gray color.
were performed in 3 replicates per analyzer, 6 replicates in total. The results were compared to reference method pH meter.
The study results demonstrate the CV is within 10% and the bias is within 10%. All results within criteria when specimen is 30-80 uL. Thus, the minimum sample volume is 30 µL.
Operated Temperature and Humidity
Semen Study
The operated temperature and humidity effect study was evaluated under various environment factors: 15 and 38°C (59 and 100.4°F) and each with 50% and 90% relative humidity. Six concentration levels of Semen samples, 1 analyzer, and 3 lots of test cassette were used. Test procedures were performed in 5 replicates. The results were compared against reference method SQA-V Analyzer.
The results indicate basis within 10% and conclude that candidate device can operate under the environment conditions 15-38°C (59-100.4°F) and 50-90% relative humidity.
pH Study
The operated temperature and humidity effect study was evaluated under various environment factors: 15, 25 and 38°C (59, 77 and 100.4°F) and 50% to 90% relative humidity. Three pH levels of Semen samples, 1 analyzer, and 3 lots of test cassette were used. Test procedures were performed in 5 replicates. The results were compared against reference pH meter.
The results indicate basis within 10% and conclude that candidate device can operate under the environment conditions 15-38°C (59-100.4°F) and 50-90% relative humidity.
Test Strip Stability (Closed Vial)
The closed vial test stability studies were performed for Semen Test cassette to evaluate the following:
Real-Time Conditions (2 and 25°C) Accelerated Conditions (40°C)
1 Analyzer, 3 lots of Test Cassette, and 3 Semen samples were used to perform the Semen and pH stability study, test measurements in 3 replicates. All results were compared against reference method SQA-V Analyzer.
The study results should demonstrate stable test cassette strip performance for 8 months under accelerated 40℃ condition, which can estimate shelf life of test strips for 2 years at room temperature.
-
- Discussion of Clinical Tests Performed:
System Accuracy Study and Layuser Performance Study
- Discussion of Clinical Tests Performed:
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Bonraybio. The logo consists of a teal-colored circular icon with a stylized wheat stalk inside, followed by the text "bonraybio" in a sans-serif font. The text is in a dark gray or black color.
The user performance study was performed to demonstrate that English speaking and reading lay users across all educational backgrounds can easily understand and follow the labeling/user instructions to obtain accurate results while using Candidate device. The study was also performed using Point-of-Care professionals or licensed registered nurses to obtain POC test findings.
The evaluation was conducted on 106 subjects with 5 units of candidate device and accompanying user instructions, 100 subjects' findings were tabulated, excluded findings were addressed in the study report. SQA-V Analyzer performed by POC personnel was used as a reference method.
The study results demonstrate that the layperson user accuracy and ease of use (via participant questionnaire scoring) of Candidate device. The study protocol, user questionnaire and form, and study findings are attached for your review.
-
- _ Conclusions:
Results of performance evaluation of LensHooke X1 Semen Quality Analyzer and LensHooke X1 PRO Semen Quality Analyzer demonstrate that the candidate devises are substantial equivalence to the predicate device, SQA V, Sperm Quality Analyzer.
- _ Conclusions:
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”