K Number
K173829
Device Name
NeoLSD MSMS kit
Date Cleared
2018-07-18

(212 days)

Product Code
Regulation Number
862.1488
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeoLSD MSMS Kit is intended for the quantitative measurement of the activity of the enzymes acid-pglucocerebrosidase (ABG), acid-sphingomyelinase (ASM), acid-a-glucosidase (GAA), B-galactocerebrosidase (GALC), α-galactosidase A (GLA) and α-L-iduronidase (IDUA) in dried blood spots (DBS) from newborn babies. The analysis of the enzymatic activity is intended as an aid in screening newborns for the following lysosomal storage disorders (LSD) respectively; Gaucher Disease, Nieman-Pick A/B Disease, Pompe Disease, Fabry Disease, and MPS I Disease.

Device Description

The NeoLSD MSMS test system uses mass spectrometry to quantitatively measure the activity of six lysosomal enzymes simultaneously from a dried blood spot sample. The NeoLSD MSMS test system is comprised of:

  1. NeoLSD MSMS kit, including substrates, internal standards, solutions and controls
  2. Waters TQD MSMS instrument comprised of,
    a. Waters 1525 sample pump
    b. Waters 2777c autosampler
    c. Waters MassLynx v4.1 firmware C.
    d. Power cables, tubing, syringes, connection cables
  3. Waters NeoLynx v4.1 software and computer with monitor
  4. PerkinElmer MSMS Workstation Software

The NeoLSD MSMS kit evaluates enzyme activities by measuring the product generated when an enzyme reacts with a synthesized substrate to create a specific end product. The activities of the six lysosomal enzymes present in a 3.2 mm punch from a dried blood spot (DBS) are simultaneously measured by the NeoLSD MSMS kit. The punches are incubated with the assay reagent mixture which contains;
. six substrates, one corresponding to each lysosomal enzyme
. six stable-isotope mass-labeled internal standards (IS) each designed to chemically resemble each product generated
. a buffer to maintain the reaction pH, and to carry inhibitors to limit activity from competing enzymes if present and additives to enhance the targeted enzyme reactions.

AI/ML Overview

The NeoLSD MSMS Kit is intended for the quantitative measurement of the activity of six lysosomal enzymes (acid-β-glucocerebrosidase (ABG), acid-sphingomyelinase (ASM), acid-α-glucosidase (GAA), β-galactocerebrosidase (GALC), α-galactosidase A (GLA), and α-L-iduronidase (IDUA)) in dried blood spots (DBS) from newborn babies. The analysis of enzymatic activity serves as an aid in screening newborns for Gaucher Disease, Niemann-Pick A/B Disease, Pompe Disease, Krabbe Disease, Fabry Disease, and MPS I Disease.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the performance metrics reported, such as linearity ranges, precision (reproducibility %CV), and LoQ values. The screening performance, particularly sensitivity and specificity, are key for a screening tool.

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance (NeoLSD MSMS Kit)
Linear RangeBroad enough to cover physiological and pathological rangesIDUA: 0.34 – 17.2 µmol/L/hGAA: 0.44 – 24.2 µmol/L/hABG: 0.69 – 20.1 µmol/L/hGLA: 0.97 – 20.9 µmol/L/hASM: 0.90 – 20.5 µmol/L/hGALC: 0.63 – 6.3 µmol/L/h
Lower Limit of Quantitation (LoQ)Low enough to detect deficient enzyme activity (within acceptable CV%)IDUA: 0.44 µmol/L/h (CV% at LoQ: 18.2%)GAA: 0.63 µmol/L/h (CV% at LoQ: 17.5%)ABG: 0.69 µmol/L/h (CV% at LoQ: 21.7%)GLA: 0.97 µmol/L/h (CV% at LoQ: 17.5%)ASM: 0.90 µmol/L/h (CV% at LoQ: 20.0%)GALC: 0.34 µmol/L/h (CV% at LoQ: 20.6%)
Reproducibility (%CV)Within acceptable limits for a diagnostic assay (e.g., <20-30%)Within-Laboratory CV% RangeIDUA: 4.7 – 6.9%GAA: 4.2 – 5.5%ABG: 11.6 – 13.8%GLA: 5.0 – 13.3%ASM: 7.3 – 11.0%GALC: 7.9 – 19.5%Between-Laboratory CV% RangeIDUA: 4.4 – 8.1%GAA: 3.5 – 7.6%ABG: 4.7 – 15.8%GLA: 5.6 – 8.4%ASM: 1.8 – 6.6%GALC: 2.1 – 7.0%Overall Reproducibility CV% RangeIDUA: 6.9 – 10.0%GAA: 5.6 – 9.4%ABG: 13.0 – 21.0%GLA: 8.6 – 15.7%ASM: 7.6 – 11.4%GALC: 9.3 – 20.7%
Sensitivity (overall)High, to minimize false negatives in screening (e.g., >90%)92.9% (76.5%-99.1%) (excluding invalid and lost-to-follow-up, including 2 Fabry females that were false negatives) With female Fabry subjects excluded, the test system has no false negative results for any of the enzymes.
Specificity (overall)High, to minimize false positives (e.g., >95%)99.4% (99.1%-99.6%) (excluding invalid and lost-to-follow-up)
False Positive Rate (overall)Low, to minimize unnecessary follow-up (e.g., <5%)0.6% (0.4% - 0.9%)
False Negative Rate (overall)Very low, critical for screening (e.g., <1%)7.1%* (0.9% - 23.5%) (*includes 2 Fabry females). When female Fabry subjects are excluded, the test system has no false negative results for any of the enzymes.
InterferenceMinimal, or clearly identified and manageableSeveral potential interferents identified (e.g., Glucose, Hematocrit, Hemoglobin, Triglycerides, EDTA), with their effects and implications described. For most, the interferences are not pronounced enough to impair affected/unaffected separation or occur at clinically irrelevant concentrations. Specific warnings are provided for high glucose, hematocrit, and triglyceride levels near cut-off values.

Note: The document provides performance metrics, implying these are the acceptance criteria that the device has met or is expected to meet for its intended use as a newborn screening aid.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Screening Performance Study:

    • Routine Samples: 4011 newborn specimens (retrospective, 4 years old, used for follow-up of clinical status).
    • Confirmed LSD Positive Samples (enriched): 30 newborn DBS specimens (from the site's biobank, ranging from 5.8 to 17.6 years of age).
    • Total Test Set: 4041 specimens (4011 routine + 30 confirmed positive).
  • Data Provenance:

    • Routine Samples: Retrospective routine newborn screening samples, 4 years old, from an EU (European) newborn screening laboratory.
    • Confirmed LSD Positive Samples: From the site's biobank (likely the same EU lab), with ages ranging from 5.8 to 17.6 years.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test set.

Instead, the ground truth for the 4011 routine samples was established based on:

  • Clinical outcome: "Clinical outcome was used as a comparator for all samples, including the 4011 routine screening samples, as derived from the civil registry status and national hospital registry. Subject´s survival at 4 years of age without LSD diagnosis or clinical signs suggestive of an LSD was used as clinical confirmation of an unaffected newborn."
  • For the 30 confirmed LSD positive samples: Their status was "known" as "confirmed LSD positive newborn DBS specimens."

Therefore, the ground truth relies on clinical follow-up data and prior confirmed diagnoses, rather than a panel of experts adjudicating each case for the study.

4. Adjudication Method for the Test Set

No explicit "adjudication method" in the sense of expert review (e.g., 2+1, 3+1) is described for the test set. The ground truth was established by:

  • Clinical outcome and registry data for routine samples to determine "unaffected" status.
  • Known (prior confirmed) diagnoses for the "confirmed positive" samples.
  • Screening algorithm: For routine samples, those below the initial cut-off were re-tested in duplicate to classify as normal, presumptive positive, or invalid.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a diagnostic kit that quantitatively measures enzyme activity, not an interpretative imaging AI tool that assists human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

Yes, the screening performance study essentially represents a standalone (algorithm only) performance for the NeoLSD MSMS kit. The device measures enzyme activity, and the "screening results" (positive, negative, invalid) are derived directly from these quantitative measurements compared against predefined cut-off values and a re-testing algorithm. While a human laboratory technician performs the assay, the interpretation of the results as "screen positive" or "screen negative" is determined by the device's output and the established algorithm, without human interpretative judgment affecting the individual sample classification.

7. The Type of Ground Truth Used

The ground truth used was primarily:

  • Outcomes data/Clinical Confirmation: For the 4011 routine samples, "Subject´s survival at 4 years of age without LSD diagnosis or clinical signs suggestive of an LSD was used as clinical confirmation of an unaffected newborn." This is a form of clinical outcome data.
  • Pathology/Confirmed Diagnosis: For the 30 enriched samples, they were "confirmed LSD positive" specimens, indicating a definitive medical diagnosis.

8. The Sample Size for the Training Set

The document describes studies for establishing reference ranges and calibration, but it does not explicitly describe a "training set" in the context of machine learning model development. The development of this assay likely involved extensive analytical validation (e.g., linearity, LoQ, interference) and establishing reference ranges using large sample sets, which might be considered analogous to a training or development phase for defining assay parameters and cut-offs.

  • Reference Range Establishment:
    • EU site: 5041 newborn samples were tested to establish cut-off values. These were "retrospective routine newborn screening samples" from newborns 0-30 days of age.
    • US Site A: 5251 newborn DBS specimens, newborns ≤ 4 days.
    • US Site B: 5053 newborn DBS specimens, newborns ≤ 7 days.

These large cohorts were used to determine population distributions, medians, and percentiles to set initial and retest cut-off values. While not a "training set" for an AI algorithm, they serve a similar purpose in defining the operational parameters for the device's classification logic.

9. How the Ground Truth for the Training Set Was Established

Given that there isn't a "training set" for an AI model, the "ground truth" for establishing the reference ranges and cut-offs was based on:

  • Population Distribution: Statistical analysis of enzyme activity levels in large cohorts of presumably healthy newborns (5041 from EU, 5251 from US Site A, 5053 from US Site B).
  • Expert-defined Percentiles: The initial cut-off values were based conservatively on "0.1 - 0.3 percentile of enzyme activity distribution and converted to a percentage of population median activity," which reflects expert consensus on appropriate thresholds for screening. The "retest cut-off values were set 5% lower from the initial cut-off percentage."

This process is standard for establishing normal ranges and screening cut-offs for diagnostic assays and involves statistical methods and clinical expert judgment in setting initial thresholds.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

July 18, 2018

Wallac Oy, a subsidiary of PerkinElmer Kay Taylor VP, Regulatory, Quality, Medical & Scientific Affairs 940 Winter Street Waltham, MA 02451

Re: K173829

Trade/Device Name: NeoLSD MSMS kit Regulation Number: 21 CFR 862.1488 Regulation Name: Lysosomal storage disorder newborn screening test system Regulatory Class: Class II Product Code: POW, POT, POT, POV, OCL, OCM Dated: July 9, 2018 Received: July 10, 2018

Dear Kay Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173829

Device Name

NeoLSD MSMS kit

Indications for Use (Describe)

The NeoLSD MSMS Kit is intended for the quantitative measurement of the activity of the enzymes acid-pglucocerebrosidase (ABG), acid-sphingomyelinase (ASM), acid-a-glucosidase (GAA), B-galactocerebrosidase (GALC), α-galactosidase A (GLA) and α-L-iduronidase (IDUA) in dried blood spots (DBS) from newborn babies. The analysis of the enzymatic activity is intended as an aid in screening newborns for the following lysosomal storage disorders (LSD) respectively; Gaucher Disease, Nieman-Pick A/B Disease, Pompe Disease, Fabry Disease, and MPS I Disease.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

This 510k Summary information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510 (k) number is K173829

Date: July 18, 2018

Submitted by: PerkinElmer Inc. 940 Winter Street Waltham, MA 02451

Wallac Oy, a subsidiary of PerkinElmer Inc. Mustionkatu 6 Turku, Finland 20750

Kay A. Taylor Contact Person: Tel: 317 418-1735

Trade Name: NeoLSD MSMS Kit

Common Name: NeoLSD MSMS Kit

Regulation: 21 CFR 862.1488

Classification Name: Lysosomal storage disorder newborn screening test system

Classification: 75 Chemistry

Product Code: PQW, PQT, PQU, PQV, QCL, QCM

Predicate device: Baebies SEEKER 4-Plex Assay (IDUA/GAA/GBA/GLA) (DENT50035)

{4}------------------------------------------------

Device Description:

The NeoLSD MSMS test system uses mass spectrometry to quantitatively measure the activity of six lysosomal enzymes simultaneously from a dried blood spot sample. The NeoLSD MSMS test system is comprised of:

    1. NeoLSD MSMS kit, including substrates, internal standards, solutions and controls
      The NeoLSD MSMS Kit will contain sufficient reagents and consumables to perform 960 assays (10 x 96-well plates) as listed in the following table.
ComponentDescription
Kit insertInstructions for use
QC certificateQC certificate showing the kit lot specific Kit Control results determined bythe manufacturer, and the 1 SD limits.
Internal Standards Substrate Mix1 vial or several vials of stable-isotope standards and designed substrates,The dried substrates and internal standards are a mixture of the 6 syntheticsubstrates, the corresponding 6 stable-isotope labeled internal standards,and sodium oleate
DBS controlsC1, C2, C3 control levels on DBS cassettes, manufactured from human bloodwith a hematocrit value of 45-50%.
Assay buffer1 bottle of 40 mL buffer, ready-for-use succinate buffered (pH 4.7) saltsolution
Extraction SolutionEthyl acetate
Flow Solvent reconstitutionsolventThe ready-for-use Flow Solvent contains acetonitrile, water, and formicacid.
Incubation/Sampling plate20 x 96-well microplate, U-bottomed
Extraction plate10 x 96-deep well microplate
Aluminum foil microplate covers20 adhesive aluminum foil microplate covers
Microplate covers10 x adhesive microplate covers
Plate barcode labels30 x plate barcodes
    1. Waters TQD MSMS instrument comprised of,
    • a. Waters 1525 sample pump
    • b. Waters 2777c autosampler
    • Waters MassLynx v4.1 firmware C.
    • d. Power cables, tubing, syringes, connection cables

{5}------------------------------------------------

    1. Waters NeoLynx v4.1 software and computer with monitor
    1. PerkinElmer MSMS Workstation Software

The NeoLSD MSMS kit evaluates enzyme activities by measuring the product generated when an enzyme reacts with a synthesized substrate to create a specific end product. The activities of the six lysosomal enzymes present in a 3.2 mm punch from a dried blood spot (DBS) are simultaneously measured by the NeoLSD MSMS kit. The punches are incubated with the assay reagent mixture which contains;

  • . six substrates, one corresponding to each lysosomal enzyme
  • . six stable-isotope mass-labeled internal standards (IS) each designed to chemically resemble each product generated
  • . a buffer to maintain the reaction pH, and to carry inhibitors to limit activity from competing enzymes if present and additives to enhance the targeted enzyme reactions.

{6}------------------------------------------------

Comparison Chart:

Comparison of the NeoLSD MSMS Kit with the predicate.

NeoLSD MSMS Kit
CharacteristicsProposed DevicePredicate (DEN150035)
IntendedUse/Indications forUseThe NeoLSD MSMS Kit is intended for the quantitativemeasurement of the activity of the enzymes acid-β-glucocerebrosidase (ABG), acid-sphingomyelinase (ASM),acid-α-glucosidase (GAA), β-galactocerebrosidase (GALC), α-galactosidase A (GLA) and α-L-iduronidase (IDUA) in driedblood spots (DBS) from newborn babies. The analysis of theenzymatic activity is intended as an aid in screening newbornsfor the following lysosomal storage disorders (LSD)respectively; Gaucher Disease, Niemann-Pick A/B Disease,Pompe Disease, Krabbe Disease, Fabry Disease, and MPS IDisease.The SEEKER System, including the SEEKER Instrument and theSEEKER LSD Reagent Kit-IDUA GAA GBA GLA for use on the SEEKERInstrument, is intended for quantitative measurement of theactivity of α-L-iduronidase, α-D-glucosidase, β-glucocerebrosidaseand α-D-galactosidase A from newborn dried blood spot specimensas an aid in screening newborns for Mucopolysaccharidosis Type I,Pompe, Gaucher and Fabry diseases. Reduced activity of theseenzymes may be indicative of these lysosomal storage diseases. Theenzymes measured using the SEEKER LSD Reagent Kit-IDUA GAA GBA GLA and their associated lysosomal storagediseases are listed below..Enzyme (abbreviation) Diseaseα-L-iduronidase (IDUA) Mucopolysaccharidosis Type I (MPS I)α-D-glucosidase (GAA) Pompeβ-glucocerebrosidase (GBA) Gaucherα-D-galactosidase A (GLA) Fabry
Test MethodologyQuantitative mass spectrometric enzymatic activity assayQuantitative fluorimetric enzymatic activity assay
Instrument /Software PlatformWaters TQD instrument with MassLynx v4.1 firmware, withWaters 1525 sample pump, with Waters 2777c autosampler,with Waters NeoLynx v4.1 software and with the PerkinElmerMSMS Workstation SoftwareSeeker Instrument with Spot Logic Software
Sample TypeSamePunch from dried blood spot specimen

{7}------------------------------------------------

CharacteristicsProposed DevicePredicate (K071649)
Reportable Range(µmol/L/h)IDUA: 0.34 – 17.2IDUA: 2.77 – 50.75 µmol/L/h
GAA: 0.44 – 24.2GAA: 2.18 – 94.66 µmol/L/h
ABG: 0.69 – 20.1 (identified as GBA in predicate)GBA: 2.14 – 73.24 µmol/L/h
GLA: 0.97 – 20.9GLA: 4.88 – 153.74 µmol/L/h
ASM: 0.90 – 20.5ASM: not detected by predicate
GALC: 0.63 – 6.3GALC: not detected by predicate
Lower Limits ofMeasure (µmol/L/h)IDUA: LoB=0.059, LoD=0.24, LoQ=0.34IDUA (µmol/L/h): LoB (1.78), LoD (2.77), LoQ (2.77)
GAA: LoB=0.073, LoD=0.39, LoQ=0.44GAA (µmol/L/h): LoB (0.50), LoD (2.18), LoQ (2.18)
ABG: LoB=0.165, LoD=0.63, LoQ=0.69GBA (µmol/L/h): LoB (0.72), LoD (1.07), LoQ (1.85)
GLA: LoB=0.476, LoD=0.97, LoQ=0.97GLA (µmol/L/h): LoB (1.96), LoD (3.18), LoQ (4.88)
ASM: LoB=0.110, LoD=0.27, LoQ=0.90ASM: not detected by predicate
GALC: LoB=0.106, LoD=0.34, LoQ=0.63GALC: not detected by predicate
Calibrators /StandardsMolecular Weights & Concentrations of Internal Standards:
IDUA: 430.26 / 15.0 μΜCalibrant A: (0.0375 μΜ)
GAA: 502.32 / 24.0 μΜCalibrant B: (0.0750 μΜ)
ABG: 390.38 / 20.0 μΜCalibrant C: (0.1500 μΜ)
GLA: 488.31 / 24.0 μΜCalibrant D: (0.3000 μΜ)
ASM: 404.40 / 15.0 μΜ
GALC: 416.40 / 10.0 μΜ
Controls3 levels of control material, human blood based4 levels of control material, human blood based
Expected Values(µmol/L/h)
N 0.1% 0.2% 0.3% 2.5% 97.5% IDUA 5041 2.06 2.55 2.62 3.82 13.2 GAA 5041 2.33 2.69 2.92 4.28 17.5 ABG 5041 2.85 3.16 3.33 4.74 20.1 GLA 5041 3.04 3.37 3.62 4.81 25.8 ASM 5041 2.12 2.21 2.37 3.15 12.9 GALC 5041 0.43 0.52 0.56 0.95 9.34N 0.1% 0.5% 50% 99.5% 99.9% IDUA 102,399 3.68 5.63 18.37 46.23 >50.75 GAA 102,392 7.61 10.07 25.98 64.56 77.41 GBA 102,397 6.69 8.20 20.24 51.03 64.53 GLA 102,371 7.56 9.44 26.73 123.36 >153.74

{8}------------------------------------------------

Subjects age 7-31 days
N0.1%0.5%50%99.5%99.9%
IDUA7,1773.154.8218.82>50.75>50.75
GAA7,1775.047.7521.5373.01>94.66
GBA7,1775.166.6016:5654.83>73.24
GLA7,1765.146.3919.7096.19134.27
Subjects age >14 days
N0.1%0.5%ર૦%99.5%99.9%
IDUA7,4472.833.9319.41>50.75>50.75
GAA7,4474.406.4317.7654.6872.21
GBA7,4474.315.8314.6145.2164.54
GLA7,4474.885.3614.6467.34113.61
ASM: not detected by predicate
GALC: not detected by predicate
Reproducibility
Sample Activity%CV RangeSample Activity%CV Range
Range (µmol/L/h)Range (µmol/L/h)
IDUA0.59 – 9.116.9 – 10.0IDUA3.53 - 24.0614.2 — 28.5
GAA1.29 — 10.775.6 - 9.4GAA4.29 – 27.3712.0 – 17.0
ABG1.03 – 8.0813.0 – 21.0ABG2.84 – 15.0015.7 — 38.0
GLA1.04 — 12.498.6 - 15.7GLA6.94 – 52.669.4 — 16.3
ASM2.47 – 9.117.6 - 11.4ASM: not detected by predicate
GALC0.37 – 3.979.3 — 20.7GALC: not detected by predicate

{9}------------------------------------------------

Summary of Studies:

Reproducibility:

Reproducibility of the NeoLSD MSMS assay was determined on three different TQD MSMS instrument at three sites, two external and one internal. The reproducibility is based on 75 determinations for each level of a six member panel for each enzyme: in each laboratory 5 plates were measured over 5 working days using one kit lot and each plate having 5 replicates per sample per enzyme. The results of reproducibility, between- and within-laboratory precisions are presented below.

EnzymeSampleNMeanWithin LaboratoryBetween LaboratoryReproducibility
SDCV%SDCV%SDCV%
ABG*2751.030.1413.80.1615.80.2221.0
ABG*3752.780.3713.30.269.30.4516.2
ABG*4757.460.8912.00.7710.31.1815.8
ABG*5757.930.9211.60.465.91.0313.0
ABG*6758.081.0312.80.384.71.1013.6
ASM*2752.470.218.70.124.80.249.9
ASM*3753.930.4311.00.112.90.4511.4
ASM*4755.540.447.90.376.60.5710.3
ASM*5757.290.567.60.476.40.7310.0
ASM*6759.110.677.30.171.80.697.6
GALC1750.370.0719.50.037.00.0820.7
GALC2750.890.1112.40.066.90.1314.2
GALC3751.300.2015.70.043.20.2116.0
GALC4751.760.2514.30.042.10.2514.4
GALC5753.760.349.10.225.90.4110.8
GALC6753.970.317.90.205.00.379.3
IDUA*2750.590.035.80.058.10.0610.0
IDUA*3752.250.156.90.104.40.188.2
IDUA*4755.880.355.90.274.60.447.5
IDUA*5758.890.434.80.455.00.626.9
IDUA*6759.110.434.70.545.90.697.5
GLA1751.040.109.50.065.60.1211.1
GLA2751.230.1613.30.108.40.1915.7
GLA3753.420.175.00.247.00.308.6
GLA4755.100.326.20.306.00.448.6
GLA57510.310.525.10.828.00.979.4

{10}------------------------------------------------

67512.490.907.20.866.91.2510.0
2751.290.075.50.107.60.129.4
3752.320.104.20.146.10.177.4
GAA*4755.360.264.80.244.50.356.6
5757.570.364.80.537.00.648.5
67510.770.464.30.383.50.605.6

(*) The results of sample 1 for ABG, ASM, IDUA and GAA were outside the measuring range and therefore not included.

Limit of Blank, Detection and Quantification:

The Limits of Blank, Detection and Quantitation were determined in accordance with CLSI document EP17-A2. The Limit of Blank (LoB) for the NeoLSD MSMS kit is defined as the 95th percentile of a distribution of blank samples determined with 120 determinations for the TQD MSMS instrument. The Limit of Detection (LoD) is based on 150 determinations of low level samples using the Waters TQD MSMS instrument.

EnzymeLoB(µmol/L/h)LoD(µmol/L/h)
ABG0.1650.63
ASM0.1100.27
GALC0.1060.34
IDUA0.0590.24
GLA0.4760.97
GAA0.0730.39

The Limit of Quantitation (LoQ) is defined as the lowest activity fulfilling the total CV% requirement of the assay. For ABG, GLA and IDUA the CV% requirement is < 40%, for ASM and GAA < 30% and for GALC < 50%. If the imprecision criterion was met for activities below LoD, the LoQ was set to be equal to the LoD. The table shows the LoQ results and imprecision observed at these activities:

EnzymeABGASMGAAGALCGLAIDUA
LoQ (µmol/L/h)0.690.900.630.340.970.44
SD at LoQ0.150.180.110.070.170.08
CV% at LoQ21.7%20.0%17.5%20.6%17.5%18.2%

Linearity:

The linearity was determined in accordance with CLSI document EP06-A using the TQD MSMS instrument. For the six enzymes, the method has been demonstrated to be linear as shown in the following table.

Confidential

{11}------------------------------------------------

EnzymeLinear Range Lower Limit (µmol/L/h)Linear Range Upper Limit (µmol/L/h)
ABG0.6920.1
ASM0.9020.5
GALC0.636.3
IDUA0.3417.2
GLA0.9720.9
GAA0.4424.2

Interference:

The NeoLSD MSMS kit was evaluated for interference in accordance with CLSI document EP07-A2. The substances potentially interfering with the MSMS analysis (mass overlaps ±1 Da of the target analytes) were studied at different concentrations in Neo MSMS Flow Solvent using the assay FIA-MSMS method. The substances were found not to interfere in the MSMS analysis at the concentrations indicated in the following table.

Tested substancePotentialInterferenceMWCorrespondingNeoLSD analyteCorrespondingNeoLSD exactmassConcentration of testedsubstance(µmol/L)
Pantoprazole383.08ABG P383.341.1, 3.3 and 10
Meropenem383.15ABG P383.341.1, 3.3 and 10
Felodipine383.07ABG P383.340.09, 1.1, and 3.3
Cetirizine (M+2 peak)390.15ABG IS390.38
S-(5'-Adenosyl)-L-methionine398.14ASM P397.36
PTH-(E-phenylthiocarbamyl)lysine398.12ASM P397.361.1, 3.3 and 10
Sulfasalazine/sulfadiazine398.07ASM P397.363.3, 10 and 754
Perphenazine (M+2 peak)405.15ASM IS404.40
Lisinopril405.23ASM IS404.40
Miconazole (M+2 peak)415.98GALC IS416.401.1, 3.3 and 10
Spironolactone416.20GALC IS416.40
Calcitriol416.33GALC IS416.40
Domperidone425.16IDUA P425.23
Kanamycin484.24GLA P483.273.3, 10 and 124

The substances potentially interfering with the assay were added to whole blood with three LSD enzyme activities (deficient, cut-off and normal). The substances indicated in the table below were found not to interfere with the assay at the concentration indicated.

{12}------------------------------------------------

Substances found not to interfere with NeoLSD assay.

Tested substanceAdded concentration of tested substance (in blood)
Bilirubin (Unconjugated)10 mg/dL
Bilirubin (Conjugated)15 mg/dL
Albumin (HSA; from human serum)2.7, 3.2, 4.0 g/dL*
Acetaminophen5.5 mg/dL
Calcifediol10.5 µg/dL
Chlorhexidine digluconate0.04 %
Galactose15 mg/dL
  • For albumin, the concentration range includes endogenous and added albumin levels.

In this study, the following potential interferents (and enzymes affected) were identified.

Glucose interference on GAA: Glucose was found to interfere with the assay by decreasing the measured GAA activity. Glucose at level of 0.50 g/dL decreases GAA activity level of 11.3 µM/h by 17%. Glucose at level of 0.75 g/dL decreases GAA activity levels of 1.44 and 2.87 µM/h by 0.40 µM/h and 23%, respectively. Thus, glucose concentration above 0.25 g/dL with GAA may cause a false positive screening result for a specimen with measured GAA activity close to the cut-off value. However, the observed glucose concentration shown to interfere with GAA is clearly beyond the endogenous reference interval for glucose that has been reported to be for neonates (0-1 months) from 0.055 to 0.115 g/dL in whole blood. Note: Preterm infants typically with very-low birth weight have a high risk of hyperglycemia (blood glucose level > 0.18 g/dL) due to glucose infusion.

Hematocrit interference on ABG and GALC: Hematocrit at level of 35% decreases ABG activity levels of 1.10, 2.92 and 8.48 µM/h by 0.49 µM/h, 25% and 18%, respectively. This may cause a false positive screening result for a specimen with measured ABG activity close to the cut-off value. Also, hematocrit at level of 65% increases ABG activity levels of 2.92 and 8.48 µM/h by 29% and 28%, respectively. At hematocrit levels ≥65%, the measured ABG activity levels could be increased as much as by 35%. Therefore, at hematocrit levels of ≥65% the interference could result in misclassification of a patient with an ABG result near the cut-off value as 'normal' when in fact patient should be classified as having presumed ABG deficiency. For samples up to 30% above the ABG cut-off value, with known or suspected high hematocrit levels (≥ 65%); testing by an alternate method that is not subject to hematocrit interference is recommended.

{13}------------------------------------------------

Hematocrit at level of 35% increases GALC activity level of 3.38 µM/h by 22%; while at levels of 65% hematocrit decreases GALC activity levels of 0.75 and 1.02 µM/h by 24% and 20%, respectively. The decrease in GALC activity may cause a false positive screening result for a specimen with measured GALC activity close to the cut-off value. The inhibitory effect of high hematocrit values to GALC activity has been reported.

Hemoglobin interference on ABG, ASM and IDUA:

Hemoglobin was found to interfere with the assay by increasing the measured ABG, ASM and IDUA activity.

ABG: Hemoglobin at level of 16.6 g/dL increases ABG activity level of 2.36 µM/h by 27% when compared to a sample having 15.3 g/dL of hemoglobin at level of 18.0 g/dL hemoglobin increases ABG activity level of 0.65 µM/h by 0.34 µM/h when compared to a sample having 15.5 g/dL of hemoglobin at level of 18.9 g/dL increases ABG activity level of 7.68 µM/h by 30% when compared to a sample having 15.1 g/dL of hemoglobin.

ASM: Hemoglobin at level of 17.8 g/dL hemoglobin increases ASM activity level of 1.57 µM/h by 26% when compared to a sample having 15.3 g/dL of hemoglobin at level of 18.0 g/dL increases ASM activity level of 0.41 µM/h by 0.20 µM/h when compared to a sample having 15.5 g/dL of hemoglobin at level of 20.1 g/dL increases ASM activity level of 4.78 µM/h by 16% when compared to a sample having 15.1 g/dL of hemoglobin.

IDUA: Hemoglobin at level of 18.0 g/dL increases IDUA activity level of 1.18 µM/h by 19% when compared to a sample having 15.5 g/dL of hemoglobin.

Although interference by hemoglobin was observed with concentrations within the newborn reference ranges (12.0 – 22.0 g/dL for hemoglobin), it is concluded based on the external study results that the interferences are not pronounced enough to impair the separation of the affected and unaffected cases.

Triglyceride interference on GAA, GLA and IDUA: Intralipid (Triglyceride) was found to interfere with the assay increasing the measured GAA, GLA and IDUA activity.

GAA: Intralipid at level of 0.23 g/dL increases GAA activity of 2.98 and 7.10 µM/h by 27% and 18%, respectively. Intralipid at level of 0.75 g/dL increases GAA activity level of 1.05 µM/h by 0.52 µM/h.

{14}------------------------------------------------

GLA: Intralipid at level of 0.30 g/dL increases GLA activity level of 7.21 µM/h by 16%; while at level 0.38 g/dL Intralipid increases GLA activity level of 3.63 µM/h by 25%. Intralipid at level of 0.75 g/dL increases GLA activity level of 1.37 µM/h by 0.65 µM/h.

IDUA: Intralipid at level of 0.38 g/dL increases IDUA activity level of 8.18 µM/h by 19% and Intralipid at level of 0.75 g/dL increases IDUA activity levels of 1.17 and 2.67 µM/h by 20%.

Note! High triglyceride concentrations (hypertriglyceridemia) in newborns due to medication effects or pathological conditions may cause a false negative screening result for a specimen with measured GAA, GLA or IDUA activity close to the cut-off value.

EDTA interference on ABG and ASM: EDTA at level 0.10 g/dL increases ABG activity level of 1.80 µM/h by 0.41 µM/h. EDTA at level of 0.04 g/dL decreases ASM activity level of 3.74 µM/h by 20%; while at level of 0.10 g/dL EDTA decreases ASM activity level of 1.18 µM/h by 16%. EDTA at a level of 0.20 g/dL decreases ASM activity level of 0.44 μM/h by 0.17 μM/h.

The concentration of EDTA found to interfere with AGB and ASM were far beyond the typical therapeutic dosage 3.4 µmol/L (0.10 mg/dL) [12]. Also, the direct application of blood from the heel-puncture to the DBS paper is intended for newborn screening eliminating the risk of contamination.

Reference Range:

A prospective clinical study using four-year-old retrospective routine newborn screening samples was conducted with the NeoLSD MSMS test system to screen for the lysosomal enzymes ABG, ASM, GAA, GALC, GLA, and IDUA.

Over the course of three months 5041 newborn samples were tested at a European (EU) newborn screening laboratory to establish cut-off values. Samples in the reference range study were from newborns ranging 0-30 days of age. The initial cut-off values were based conservatively on 0.1 - 0.3 percentile of enzyme activity distribution and converted to a percentage of population median activity. The retest cut-off values were set 5% lower from the initial cut-off percentage. The cut-off percentages were applied to daily medians established based on the initial routine sample results for the day.

The cut-off percentages were not adjusted during the study. The initial and retest cut-off values used in the clinical study are shown below.

{15}------------------------------------------------

EnzymenRange*Enzyme activity (µmol/L/h)Lower percentilesInitial cutoffRetest cutoff
MeanMedian0.1%0.2%0.3%
ABG50412.07 - 66.310.359.602.853.163.3335%30%
ASM50411.64 - 35.26.385.862.122.212.3740%35%
GALC50410.22 - 54.93.633.100.430.520.5620%15%
IDUA50410.45 - 35.47.557.252.062.552.6230%25%
GLA50411.48 - 94.811.069.723.043.373.6240%35%
GAA50411.55 - 34.89.288.742.332.692.9230%25%

Descriptive statistics for the EU site

  • Some results were outside the measuring range and cannot be considered accurate

Additional newborn population distributions were determined at two different newborn screening laboratories in the United States. Site A analyzed 5251 and Site B analyzed 5053 newborn dried blood spot specimens submitted for routine testing. Descriptive statistics for the samples are shown below. Samples tested in Site A and Site B were collected from newborns of ≤ 4 days and ≤ 7 days, respectively.

Descriptive statistics for the US Site A

Enzyme activity (µmol/L/h)
EnzymenRange*MeanMedianLower percentiles
0.1%0.2%0.3%
ABG52511.39 – 77.910.529.672.002.162.36
ASM52511.10 – 24.34.704.441.491.601.64
GALC52510.52 – 85.05.494.630.860.961.02
IDUA52510.09 – 18.76.316.080.851.401.56
GLA52511.69 – 12116.9114.393.444.424.89
GAA52510.75 - 43.010.219.622.162.382.56
  • Some results were outside the measuring range and cannot be considered accurate
EnzymenRange*MeanMedianLower percentiles
0.1%0.2%0.3%
ABG50531.68 – 76.511.410.42.943.233.33
ASM50531.02 – 28.45.445.131.711.791.98
GALC50530.53 – 33.04.674.200.680.820.86
IDUA50530.9 – 21.86.656.391.591.982.09
GLA50530.96 – 59.513.312.03.994.334.43
GAA50531.11 – 36.411.010.42.532.883.37

Descriptive statistics for the US Site B

{16}------------------------------------------------

  • Some results were outside the measuring range and cannot be considered accurate

Screening Performance:

The screening performance of the NeoLSD MSMS kit was determined in a prospective clinical study of four-year-old retrospective routine samples at an EU newborn screening laboratory. Over the twomonth study duration 4041 newborn samples were tested were from newborns ≤ 29 days old. All 4011 routine samples were 4 years old enabling the follow-up of clinical status at the age 4 years. Due to the low incidence of LSDs, the study population was enriched confirmed LSD positive newborn DBS specimens obtained from the site's biobank, these samples ranged from 5.8 to 17.6 years of age. Sample inclusion criteria were the same as in the reference range study.

The 4011 routine samples were tested according to the defined algorithm to determine the enzyme activities for sample interpretation. Site specific cut-off percentages determined prior to the clinical study were applied as percent of daily median activity. The routine specimens having enzyme activities below the cut-off values in the initial testing were re-tested in duplicate. The final results (normal, presumptive positive, invalid result) were classified after the initial testing for samples above the cut-off values and after the final testing for samples below the initial cut-off level.

Confirmed LSD positive newborn specimens were tested once, and assessed using the retest cut-off values for the enzymes. Given the known status of the confirmed LSD positive newborn specimens, processing the samples through the full testing algorithm provided no additional information nor changed the sample interpretation.

Clinical outcome was used as a comparator for all samples, including the 4011 routine screening samples, as derived from the civil registry status and national hospital registry. Subject´s survival at 4 years of age without LSD diagnosis or clinical signs suggestive of an LSD was used as clinical confirmation of an unaffected newborn. Subjects who did not have follow-up information available up to 4 years were classified as lost to follow-up.

Results

A total of 4041 specimens (4011 routine newborn specimens from confirmed positive LSD cases) were tested in the clinical study. The following is a summary of the samples tested.

RoutineABGASMGALCIDUAGLAGAATotal
Screened samples4011401140114011401140114011
Below initial cutoff15111233121088
Screen positive301055426
Unaffected301055325

Summary of the samples tested in the pivotal study.

{17}------------------------------------------------

Wallac Oy, PerkinElmer Inc Confidential

Lost to follow up0000011
Retest rate0.40%0.27%0.30%0.17%0.85%0.30%2.2%
False positive rate0.07%0.00%0.10%0.07%0.27%0.10%0.6%
Confirmed PositiveABGASMGALCIDUAGLAGAATotal
Screened samples511055430
Below initial cutoff511053428
Screen positive311053426
Invalid2000002
Screen negative0000202

Two confirmed positive Gaucher specimens were categorized as "invalid result". In routine newborn screening, if a specimen result is categorized as invalid result, a new dried blood spot specimen should be obtained and retesting performed using age-specific cut-off values.

In total 86 subjects had emigrated before the age of 4 years and therefore their clinical status could not be confirmed. None of the subjects in the lost to follow-up cohort had LSD diagnosis or showed signs and symptoms of any of the screened LSDs at the time of being lost to follow up.

The screening performance results for the 4041 specimens, including the 30 confirmed LSD positive specimens, tested with the NeoLSD MSMS kit are shown below.

Outcome, all 6 enzymes
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
Positive2625152
NeoLSDMSMS kitNegative2*3900853987
Invalid2002
Total303925864041
Performance estimates (invalid and lost-to-follow-up excluded)
SensitivityFalse-negative rateSpecificityFalse-positive rate
92.9 %7.1%*99.4 %0.6 %
(76.5%-99.1%)(0.9% - 23.5%)(99.1% - 99.6%)(0.4% - 0.9%)

Screening performance of NeoLSD MSMS kit

  • includes 2 Fabry females
Outcome, Gaucher (ABG)
--------------------------

{18}------------------------------------------------

Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive3306
Negative03947864033
Invalid2002
Total53950864041
Outcome, Niemann-Pick A/B (ASM)
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive1001
Negative03952864038
Invalid0202
Total13954864041
Outcome, Krabbe (GALC)
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive104014
Negative03939864025
Invalid0202
Total103945864041
Outcome, MPS I (IDUA)
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive5308
Negative03945864031
Invalid0202
Total53950864041
Outcome, Fabry (GLA)
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive311014
Negative23937864025
Invalid0202
Total53950864041

{19}------------------------------------------------

Outcome, Pompe (GAA)
Screening ResultsAffectedUnaffectedLost-to-follow-upTotal
NeoLSDMSMS kitPositive4419
Negative03945854030
Invalid0202
Total43951864041

With the female Fabry subjects excluded the NeoLSD MSMS test system has no false negative results for any of the enzymes; and a false positive range of ABG (0.07%), ASM (0%), GALC (0.10%), IDUA (0.07 %), GLA (0.27 %), GAA ( 0.10%). . The retest rate was 6.6%, of which only 2.2% was due to test failure (e.g. quality control exceeded limits). Remaining retests (4.4%) were due to the screening algorithm, which requires results below initial cut-off value for any or the six enzymes to be repeated in duplicate.

§ 862.1488 Lysosomal storage disorder newborn screening test system.

(a)
Identification. A lysosomal storage disorder newborn screening test system is intended to measure lysosomal enzyme levels obtained from dried blood spot specimens on filter paper from newborns as an aid in screening newborns for a lysosomal storage disorder.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include information that demonstrates the performance characteristics of the device, including:
(i) Study results that adequately demonstrate the clinical validity of the device, which must include information supporting the link between the analyte being measured and the condition being screened. The clinical validity of the device must be demonstrated in a clinical validation study using either well-characterized prospectively or retrospectively obtained clinical specimens from the intended use population. Testing in the clinical validation study must be performed by operators representative of the types of operators intended to use the test. The study design of the clinical validation study must assess the effects of sample collection and processing steps on test performance. Confirmed positive specimens must have a diagnosis based on confirmatory diagnostic methods or clinically meaningful information regarding the status of the subject must be obtained.
(ii) The reference interval in the normal newborn population for the analyte or analytes measured by the device.
(iii) Study results demonstrating the level of carryover or drift affecting the device performance.
(iv) Study results demonstrating the concentrations of the limit of blank, limit of detection, and limit of quantitation of the device. Sample concentrations below the limit of quantitation must not be reported by the device.
(v) Study results, which must be collected using sample panels from at least three reagent lots and at least three instruments over more than 20 testing days, demonstrating the imprecision of the device. The sample panels must consist of blood spot specimens with a range of analyte concentrations that span the reportable range of the device and must include samples with concentrations in the screen positive range, samples with concentrations at each cutoff, and samples with concentration in the normal range.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A warning that indicates that the test is not intended to diagnose lysosomal storage disorders.
(ii) A warning that indicates that test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, and clinical evaluation as appropriate.
(iii) Detailed information on device performance, including the false positive rate and the false negative rate observed in the clinical study.
(iv) Information on device performance in any relevant subgroup (
e.g., age of newborn at time of sample collection, birth weight, sex, gestational age, race, ethnicity) observed in the clinical study.