K Number
DEN150035
Device Name
SEEKER System
Manufacturer
Date Cleared
2017-02-03

(548 days)

Product Code
Regulation Number
862.1488
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SEEKER System, including the SEEKER Instrument and the SEEKER LSD Reagent Kit-IDUA|GAA|GBA|GLA for use on the SEEKER Instrument, is intended for quantitative measurement of the activity of a-L-iduronidase, a-D-glucosidase, Bglucocerebrosidase and a-D-galactosidase A from newborn dried blood spot specimens as an aid in screening newborns for Mucopolysaccharidosis Type I, Pompe, Gaucher and Fabry diseases. Reduced activity of these enzymes may be indicative of these lysosomal storage diseases. The enzymes measured using the SEEKER LSD Reagent Kit-IDUA|GAA|GBA|GLA and their associated lysosomal storage diseases are listed below.

Enzyme (abbreviation)Disease
α-L-iduronidase (IDUA)Mucopolysaccharidosis Type I (MPS I)
α-D-glucosidase (GAA)Pompe
β-glucocerebrosidase (GBA)Gaucher
α-D-galactosidase A (GLA)Fabry
Device Description

The SEEKER System employs digital microfluidic technology to measure multiple lysosomal enzymatic activities quantitatively from newborn dried blood spot specimens. The following components are provided:

    1. SEEKER Instrument (including USB and power cables), Desktop PC with monitor, keyboard, mouse, and the Spot Logic software.
    1. SEEKER LSD Reagent Kit IDUA|GAA|GBA|GLA containing enzyme specific substrate reagents, dried blood spot extraction buffer, reaction stop buffer, 4 levels of calibrators and quality control dried blood spots containing 4 levels of quality control (OC) material. Each Reagent Kit contains sufficient consumables for 1440 tests. The contents of the kit are listed below:
ComponentContentsQuantity
Quality controldried blood spotsQC-Base Pool (QCBP)1x15 spots
QC-Low (QCL)1x15 spots
QC-Medium (QCM)1x15 spots
QC-High (QCH)1x15 spots
Enzyme substratesIDUA9x100 µL
GAA9x100 µL
GBA9x100 µL
GLA9x100 µL
CalibratorsCalibrant A (CALA)9x50 µL
Calibrant B (CALB)9x50 µL
Calibrant C (CALC)9x50 µL
Calibrant D (CALD)9x50 µL

Other components needed to run tests include the following:

ComponentContentsQuantity
Other ReagentsExtraction Buffer (EXT)9 x 30 mL
Filler Fluid9 x 10 mL
CartridgeSEEKER cartridge1 x 36

The composition of the enzyme reagents and buffers are summarized below:

Reagent (Description)Composition
IDUA(α-L-iduronidase substrate)2 mM 4-MU-α-L-iduronide sodium salt3mM D-saccharolactone0.04 M acetate buffer, pH 3.520 mM methyl β-cyclodextrin0.01% Tween20
GAA(α-D-glucosidase substrate)5 mM 4-MU-α-glucopyranoside12 µM acarbose0.04 M acetate buffer, pH 3.820 mM methyl β-cyclodextrin0.01% Tween20
GBA(β-glucocerebrosidase substrate)16 mM 4-MU-β-glucopyranoside0.05 M/0.1M citrate phosphate buffer, pH 5.20.01% Tween201.5% sodium taurocholate
GLA(α-D-galactosidase A substrate)10 mM 4-MU-α-galactopyranoside145 μM N-acetyl galactosamine0.04 M acetate buffer, pH 4.620 mM methyl β-cyclodextrin0.01% Tween20
Stop Buffer STB(reaction stopping buffer)0.6 M NaHCO3, pH 11.0 in 0.04% Tween 20
Extraction Buffer EXT(dried blood spot extraction buffer)0.1% Tween 20 in water
Filler Fluid(medium for droplet movement)0.1% Triton X-15 in 5cSt silicone oil

The Seeker Calibrators are supplied as part of the SEEKER LSD Reagent Kit -IDUA|GAA|GBA|GLA. The calibrators consist of 4 levels of aqueous preparation of 4methylumbelliferone sodium salt (4-MU) in 0.6M sodium bicarbonate buffer, pH 11.0 with 0.01% Tween 20. The concentration of 4-MU in each of the 4 calibrators is indicated in the table below:

Calibrator LevelConcentration of 4-MU
Calibrant A (CAL A)0.0375 µM
Calibrant B (CAL B)0.0750 µM
Calibrant C (CAL C)0.1500 µM
Calibrant D (CAL D)0.3000 µM

The quality control dried blood spots include 4 levels of control material: OC Low (QCL), QC Medium (QCM) and QC High (QCH). The composition of the 4 quality control dried blood spot (DBS) is summarized below. QCBP is used to fill empty wells on a cartridge.

Quality Control LevelComposition
QCBPHeat inactivated human serum, adjusted to ~50% hematocritusing human red blood cells
QCL5% cord blood and 95% heat inactivated serum, adjusted to~50% hematocrit using human red blood cells
QCM50% cord blood and 50% heat inactivated serum adjusted to~50% hematocrit using human red blood cells
QCHHuman umbilical cord blood, adjusted to ~50% hematocritusing human red blood cells

The enzymatic activity values for the quality control DBS measured by the manufacturer are given on the lot specific quality control certificate included in each assay kit for all levels except QCBP. Each laboratory should establish its own mean and acceptable ranges for the quality control materials.

3. SEEKER Cartridges

    1. Finnipipette Novus 8-channel automatic pipette 1-10 uL
    1. Finnipipette Novus 1-channel automatic pipette 10-100 uL
AI/ML Overview

The SEEKER System is a device designed for quantitative measurement of the activity of α-L-iduronidase (IDUA), α-D-glucosidase (GAA), β-glucocerebrosidase (GBA), and α-D-galactosidase A (GLA) from newborn dried blood spot specimens. This serves as an aid in screening newborns for Mucopolysaccharidosis Type I (MPS I), Pompe, Gaucher, and Fabry diseases.

Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the SEEKER System are implicitly defined by the analytical and clinical performance standards required for its De Novo classification as a Class II device with special controls. These controls mandate thorough demonstration of performance characteristics, including clinical validity, reference intervals, carry-over, detection limits, and imprecision. The clinical validity is primarily demonstrated by the false positive and false negative rates observed in the pivotal clinical study.

Table: Acceptance Criteria (Implied by Regulatory Requirements) and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Pivotal Phase Study)Notes
Clinical Validity (False Positive Rate)Acceptable false positive rates to minimize unnecessary confirmatory testing and emotional burden.IDUA: 0.035%GAA: 0.092%GBA: 0.047%GLA: 0.097%These rates are considered acceptable, particularly given the benefits of early detection and the fact that newborns with "presumed normal" results after risk assessment were not followed up, potentially leading to an overestimation of the false positive rate.
Clinical Validity (False Negative Rate)Acceptable false negative rates to ensure affected newborns are detected and receive timely therapy.IDUA: 0%GAA: 0%GBA: 0%GLA: 0% (Note: 2 newborns that would have been FN at 7.0µmol/L/hr cutoff were TP at 8.0µmol/L/hr cutoff. Labeling specifies known limitations for female Fabry and late-onset Pompe).The reported 0% false negative rate for IDUA, GAA, and GBA is based on active surveillance and the assumption that early-onset cases would be reported. However, some limitations are disclosed in the labeling, acknowledging potential false negatives due to clinical variability (e.g., late-onset Pompe, female Fabry patients). The adjustment of cutoffs over time also indicates complexities in achieving a perfect 0% FN rate. The note about GLA false negatives suggests that the 0% reported might be dependent on the specific cutoff used at the time.
Imprecision (Reproducibility)Reproducibility across different instruments, reagent lots, and days, within acceptable limits. (No explicit %CV target stated, but general medical device standards suggest low CVs for reliable quantitative assays).IDUA: 14.2% - 28.5% CV (depending on concentration)GAA: 12.0% - 17.0% CVGBA: 15.7% - 38.0% CVGLA: 9.4% - 16.3% CVThese CVs are generally considered acceptable for screening assays, especially at lower concentrations where biological variability or technical limitations might be higher. The CLSI EP5-A3 guideline was followed.
Linearity/Reportable RangeLinear response over the intended measurement range, covering normal and disease-associated values.IDUA: 2.77 to 50.75 µmol/L/hGAA: 2.18 to 94.66 µmol/L/hGBA: 2.14 to 73.24 µmol/L/hGLA: 4.88 to 153.74 µmol/L/h. Max deviation from linearity <10%.Acceptable linearity was demonstrated. Outliers were noted but did not impact overall linearity claims, though their presence highlights the importance of robust outlier handling in routine testing.
Detection Limits (LoQ)Limit of Quantitation (LoQ) should be below or at relevant clinical cutoffs to ensure accurate measurement in the critical low activity range.IDUA: LoQ = 2.77 µmol/L/h (some clinical cutoffs were below this, e.g., High Risk 1.5 µmol/L/h)GAA: LoQ = 2.18 µmol/L/hGBA: LoQ = 1.85 µmol/L/hGLA: LoQ = 4.88 µmol/L/h (some clinical cutoffs were below this, e.g., 14+ days High Risk 3.0 µmol/L/h)The LoQ values are provided. A significant limitation noted is that some clinical cutoffs for IDUA and GLA were below the LoQ, meaning that results in this critical range might not be reliably quantifiable, introducing an element of uncertainty in diagnosis for values between LoQ and these lower cutoffs. This is explicitly recognized as a limitation in the study findings.
Analytical SpecificityMinimal interference from common endogenous and exogenous substances.Most common interferents (bilirubin, galactose, glucose, EDTA, heparin, intralipid) showed <+10% bias. IDUA specific interferences: total protein at >88 mg/mL and hematocrit at 66% caused significant negative bias; heparin increased imprecision.Generally good, but specific interferences identified for IDUA (high protein, high hematocrit, heparin) are crucial for laboratories to consider. These are included in the package insert.
Carry-overDemonstrate minimal or no significant impact from carry-over.Most assays showed minimal bias (<10%). GLA showed up to 27% bias in one specific condition, leading to a recommendation for retesting.Carry-over was identified for GLA under specific conditions (high GLA sample preceding a borderline sample), leading to a specific retesting recommendation in the package insert. This directly addresses the identified risk.

The studies described predominantly cover the analytical performance requirements (reproducibility, linearity, detection limits, analytical specificity, carry-over) and the clinical performance (false positive/negative rates).

2. Sample Size Used for the Test Set and Data Provenance

The primary clinical effectiveness study (pivotal phase) involved screening 105,599 newborns.

Data Provenance: The study was performed over 24 months at a public state health laboratory that performs newborn screening. This implies the data were collected in the United States. The study was prospective in nature, as it involved actively screening newborns using the device over a defined period (pivotal phase 18 months). A preliminary study, prior to the pilot/pivotal studies, used approximately 13,000 presumed normal de-identified specimens and 29 known affected specimens, suggesting a mix of retrospective (de-identified) and potentially prospective collection for the known affected cases.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the "number of experts" or their specific "qualifications" involved in establishing the ground truth for the clinical test set in a centralized, adjudicatory manner (e.g., as consensus readers for imaging). Instead, the ground truth was established through confirmatory diagnostic methods and clinical follow-up within the standard newborn screening program.

For patients referred for follow-up:

  • "Babies considered to still be at high risk for the screened condition after the risk assessment were referred for confirmatory testing."
  • "Confirmed positive specimens must have a diagnosis based on confirmatory diagnostic methods or clinically meaningful information regarding the status of the subject must be obtained."

The "risk assessment" itself was performed by the laboratory, and would involve experienced personnel (e.g., laboratory directors, genetic counselors, metabolic specialists). While not explicitly listed as "experts establishing ground truth," these individuals make the critical decisions leading to confirmatory diagnosis. The state public health laboratory also had an active surveillance program to track false negative results, which likely involves experts in metabolic diseases.

4. Adjudication Method for the Test Set

The adjudication method for the clinical test set was an elaborate, multi-step process ingrained in the newborn screening workflow of the state health laboratory:

  • Initial Screening: DBS cards were tested in singlicate.
  • Retesting for Borderline Results: DBS with at least one enzyme below the borderline cutoff were retested in duplicate.
  • Visual Outlier Evaluation: The laboratory evaluated retest results for "visual outliers." If found, additional testing was performed.
  • Average Calculation: The average of valid retest results (excluding visual outliers) was calculated.
  • Risk Assessment: If the average was below the high-risk cutoff, a risk assessment was performed by the laboratory. This assessment included review of other newborn screening enzyme results, sample quality, transfusion status, health status, transit time, gestational age, family history, and age at sample collection.
  • Referral for Confirmatory Testing: Babies still considered at high risk after the risk assessment were referred for confirmatory diagnostic methods.
  • Presumed Normal: Babies with an average above the high-risk cutoff after retest, or those not referred after risk assessment, were presumed normal.

This is a complex, stepwise adjudication process, rather than a simple 2+1 or 3+1 consensus. It involves a combination of algorithmic cutoffs, repeat testing, visual inspection, and comprehensive clinical review.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not performed for the SEEKER System. This device is a quantitative enzymatic activity assay for newborn screening, not an imaging-based diagnostic aid or an AI system intended to be used directly by human readers in an interpretive capacity. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" does not apply in the context of this device. The SEEKER System performs the assay autonomously using digital microfluidic technology.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the primary function of the SEEKER System is standalone (algorithm only) in measuring enzyme activities. The results are reported as quantitative values (micromoles of 4-MU produced / liter of blood / hour of incubation).

However, the "human-in-the-loop" component comes into play with the interpretation of these quantitative results within the newborn screening program's workflow:

  • The device generates raw data, which is then processed by its software (Spot Logic).
  • The laboratory applies specific cutoff values (high risk, borderline), which were adjusted and refined during the study period by human experts based on observed performance and factors like age.
  • Human judgment is involved in the "risk assessment" process for borderline cases.
  • Human decision-making leads to referral for confirmatory testing.

So, while the initial measurement is standalone, the screening decision (positive/negative/retest/referral) is a combination of the device's quantitative output and a multi-step human-defined protocol.

7. The Type of Ground Truth Used

The ground truth used for determining true positives and true negatives in the clinical study was primarily based on confirmatory diagnostic methods and clinical follow-up/outcomes data.

  • For positive cases, the "True Positives" are those confirmed by "other confirmatory diagnostic methods." This implies biochemical, genetic, or clinical diagnoses consistent with the respective lysosomal storage diseases.
  • For negative cases, the "Presumptive False Positives" are those referred but not confirmed, or those deemed normal after risk assessment. The absence of reported false negatives through the active surveillance program and the assumption that early-onset cases would be clinically identified also contribute to the ground truth for true negatives (i.e., those who screened negative and did not later develop symptoms or receive a diagnosis).

8. The Sample Size for the Training Set

The document does not explicitly delineate a "training set" in the context of typical machine learning or AI models with a distinct number of samples. This device is an enzymatic assay system, not an algorithm that undergoes iterative training on a dataset.

However, the "cutoffs" used by the laboratory for interpreting the results can be thought of as parameters that were "tuned" or "trained" based on observational data:

  • "A preliminary study was performed by the public state health laboratory prior to the pilot/pivotal studies to establish initial cutoff values. Approximately 13,000 presumed normal de-identified specimens and 29 known affected specimens were analyzed using the proposed device."
  • "The cutoff values were adjusted during the course of the study in order to reduce the false negative and false positive rate and also to take into account seasonal changes..."

These initial 13,029 specimens and the ongoing refinement using data from the 48,813 newborns in the pilot phase could be considered analogous to a "training" or "development" dataset for establishing the operational parameters (cutoffs) of the screening algorithm.

9. How the Ground Truth for the Training Set was Established

For the "training" data (preliminary study and pilot phase used for cutoff adjustments):

  • Known affected specimens (29 samples): These likely had their diagnosis confirmed by established diagnostic methods (e.g., enzyme assays, genetic testing) or clinical presentation from relevant patient populations prior to being used in the preliminary study.
  • Presumed normal de-identified specimens (13,000 samples): These were likely samples from healthy newborns collected and de-identified, implicitly considered "normal" based on the absence of a disease diagnosis through routine clinical channels.
  • Ongoing Adjustment: During the pilot and pivotal phases, the clinical outcomes (confirmatory diagnoses, absence of symptoms) informed the continuous adjustment of cutoffs, acting as a feedback mechanism for refining the test's interpretive ground truth. The "active surveillance program" for false negatives also served to validate the accuracy of the negative screening decisions.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR SEEKER System DECISION SUMMARY

A. DEN Number:

DEN150035

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the SEEKER System

C. Measurand:

α-L-iduronidase (IDUA), α-D-glucosidase (GAA), β-glucocerebrosidase (GBA), and
α-D-galactosidase A (GLA)

D. Type of Test:

Quantitative fluorimetric enzymatic activity assay

E. Applicant:

Baebies, Inc.

F. Proprietary and Established Names:

SEEKER System

G. Regulatory Information:

    1. Regulation:
  • 21 CFR 862.1488
    1. Classification:

Class II (special controls)

    1. Product code(s): PQW PQT PQU PQV

{1}------------------------------------------------

    1. Panel: Chemistry (75)

H. Indications For Use:

    1. Indications for Use:
      The SEEKER System, including the SEEKER Instrument and the SEEKER LSD Reagent Kit-IDUA|GAA|GBA|GLA for use on the SEEKER Instrument, is intended for quantitative measurement of the activity of a-L-iduronidase, a-D-glucosidase, Bglucocerebrosidase and a-D-galactosidase A from newborn dried blood spot specimens as an aid in screening newborns for Mucopolysaccharidosis Type I, Pompe, Gaucher and Fabry diseases. Reduced activity of these enzymes may be indicative of these lysosomal storage diseases. The enzymes measured using the SEEKER LSD Reagent Kit-IDUA|GAA|GBA|GLA and their associated lysosomal storage diseases are listed below.
Enzyme (abbreviation)Disease
α-L-iduronidase (IDUA)Mucopolysaccharidosis Type I (MPS I)
α-D-glucosidase (GAA)Pompe
β-glucocerebrosidase (GBA)Gaucher
α-D-galactosidase A (GLA)Fabry
    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only. ●
    • For prescription use only. ●
    • This test is not intended to diagnose lysosomal storage disorders.
    • Reduced activity for any of the four enzymes should be confirmed by other confirmatory diagnostic methods.
    • Test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, and clinical evaluation as appropriate.
    • Conditions that are known to cause anomalous results are : ●
      • Specimen spot not uniformly saturated with blood. о
      • Specimen spot punched too close to the edge of the blood spot. O
      • Poorly collected and improperly dried specimens. O
      • Non-eluting blood spot due to deterioration of specimen caused by exposure to o heat and humidity.
    • Also. the SEEKER LSD Reagent Kit IDUA|GAA|GBA|GLA may result in: ●
      • False negatives by not detecting Fabry disease in females". o

4 Clinical and Laboratory Standards Institute (CLSI). Blood Collection on Filter Paper for Newborn Screening Programs: Approved Standard - Sixth Edition. CLSI document NBS01-A6 (ISBN 1-56238-883-5 (Print); ISBN 1-56238-884-3 [Electronic]). Clinical and Laboratory Standards Institute, 950 West Valley Road, Suite 2500, Wayne, Pennsylvania 19087 USA. 2013.

2 Linthorst GE, Poorthuis BJHM, Hollak CEM. Enzyme activity for determination of presence of Fabry disease in women results in 40% false-negative results. Journal of the American College of Cardiology. 2008; 51(21):2082.

{2}------------------------------------------------

  • False positives by identifying pseudo deficiencies and carriers as affected for o MPS I, Gaucher and Pompe diseases3.
  • False negative by not detecting certain late onset forms for Pompe disease . O
  • o Increased rate of false positives when the dried blood spot specimen is exposed to high temperature (≥45℃) during shipping.
    1. Special instrument requirements: SEEKER Instrument

I. Device Description:

The SEEKER System employs digital microfluidic technology to measure multiple lysosomal enzymatic activities quantitatively from newborn dried blood spot specimens. The following components are provided:

    1. SEEKER Instrument (including USB and power cables), Desktop PC with monitor, keyboard, mouse, and the Spot Logic software.
    1. SEEKER LSD Reagent Kit IDUA|GAA|GBA|GLA containing enzyme specific substrate reagents, dried blood spot extraction buffer, reaction stop buffer, 4 levels of calibrators and quality control dried blood spots containing 4 levels of quality control (OC) material. Each Reagent Kit contains sufficient consumables for 1440 tests. The contents of the kit are listed below:
ComponentContentsQuantity
Quality controldried blood spotsQC-Base Pool (QCBP)1x15 spots
QC-Low (QCL)1x15 spots
QC-Medium (QCM)1x15 spots
QC-High (QCH)1x15 spots
Enzyme substratesIDUA9x100 µL
GAA9x100 µL
GBA9x100 µL
GLA9x100 µL
CalibratorsCalibrant A (CALA)9x50 µL
Calibrant B (CALB)9x50 µL
Calibrant C (CALC)9x50 µL
Calibrant D (CALD)9x50 µL

3 Hopkins PV, Campbell C, Klug T, Rogers S, Rabum-Miller J, Kiesling J. Lysosomal storage disorder screening implementation: findings from the first six months of full population pilot testing in Missouri. Journal of Pediatics. 2015; 166(1): 172-177.

4 Chiang S-C, Hwu W-L, Lee N-C, Hsu L-W, Chien Y-H. Algorithm for Pompe disease newbom screening: Results from the Taiwan screening program. Mol Genet Metab. 2012;106(3):281-286.

{3}------------------------------------------------

ComponentContentsQuantity
Other reagentsStop Buffer (STB)9x500 uL

Other components needed to run tests include the following:

ComponentContentsQuantity
Other ReagentsExtraction Buffer (EXT)$9 \times 30$ mL
Filler Fluid$9 \times 10$ mL
CartridgeSEEKER cartridge$1 \times 36$

The composition of the enzyme reagents and buffers are summarized below:

Reagent (Description)Composition
IDUA(α-L-iduronidase substrate)2 mM 4-MU-α-L-iduronide sodium salt3mM D-saccharolactone0.04 M acetate buffer, pH 3.520 mM methyl β-cyclodextrin0.01% Tween20
GAA(α-D-glucosidase substrate)5 mM 4-MU-α-glucopyranoside12 µM acarbose0.04 M acetate buffer, pH 3.820 mM methyl β-cyclodextrin0.01% Tween20
GBA(β-glucocerebrosidase substrate)16 mM 4-MU-β-glucopyranoside0.05 M/0.1M citrate phosphate buffer, pH 5.20.01% Tween201.5% sodium taurocholate
GLA(α-D-galactosidase A substrate)10 mM 4-MU-α-galactopyranoside145 μM N-acetyl galactosamine0.04 M acetate buffer, pH 4.620 mM methyl β-cyclodextrin0.01% Tween20
Stop Buffer STB(reaction stopping buffer)0.6 M NaHCO3, pH 11.0 in 0.04% Tween 20
Extraction Buffer EXT(dried blood spot extraction buffer)0.1% Tween 20 in water
Filler Fluid(medium for droplet movement)0.1% Triton X-15 in 5cSt silicone oil

The Seeker Calibrators are supplied as part of the SEEKER LSD Reagent Kit -IDUA|GAA|GBA|GLA. The calibrators consist of 4 levels of aqueous preparation of 4methylumbelliferone sodium salt (4-MU) in 0.6M sodium bicarbonate buffer, pH 11.0 with 0.01% Tween 20. The concentration of 4-MU in each of the 4 calibrators is indicated in the table below:

{4}------------------------------------------------

Calibrator LevelConcentration of 4-MU
Calibrant A (CAL A)0.0375 $\mu$ M
Calibrant B (CAL B)0.0750 $\mu$ M
Calibrant C (CAL C)0.1500 $\mu$ M
Calibrant D (CAL D)0.3000 $\mu$ M

The quality control dried blood spots include 4 levels of control material: OC Low (QCL), QC Medium (QCM) and QC High (QCH). The composition of the 4 quality control dried blood spot (DBS) is summarized below. QCBP is used to fill empty wells on a cartridge.

Quality Control LevelComposition
QCBPHeat inactivated human serum, adjusted to ~50% hematocritusing human red blood cells
QCL5% cord blood and 95% heat inactivated serum, adjusted to~50% hematocrit using human red blood cells
QCM50% cord blood and 50% heat inactivated serum adjusted to~50% hematocrit using human red blood cells
QCHHuman umbilical cord blood, adjusted to ~50% hematocritusing human red blood cells

The enzymatic activity values for the quality control DBS measured by the manufacturer are given on the lot specific quality control certificate included in each assay kit for all levels except QCBP. Each laboratory should establish its own mean and acceptable ranges for the quality control materials.

3. SEEKER Cartridges

    1. Finnipipette Novus 8-channel automatic pipette 1-10 uL
    1. Finnipipette Novus 1-channel automatic pipette 10-100 uL

J. Standard/Guidance Documents Referenced:

CLSI EP5-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition

CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline

CLSI EP07-A2: Interference Testing in Clinical Chemistry, Approved Guideline-Second Edition

CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement

{5}------------------------------------------------

Procedures: Approved Guideline

CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline

K. Test Principle:

The SEEKER System automates enzymatic analysis for newborn screening using digital microfluidic technology. The Seeker Cartridge integrates and automates all the liquidhandling steps involved in an assay using sub-microliter droplets as reaction vessels. The cartridge manipulates the reagents from the SEEKER LSD Reagent Kit-IDUA|GAA|GBA|GLA and is controlled by the SEEKER Instrument through the Spot Logic software.

The activities of the lysosomal enzymes (IDUA, GAA, GBA and GLA) are measured in DBS extracts by an end point method using synthetic fluorescent substrates. The fluorescent substrates use 4-methylumbelliferone as the fluorophore, which has an excitation peak at 365 nm and emission peak at 460 nm. The synthetic substrates are hydrolyzed by their corresponding enzyme at acidic pH conditions to release free 4-methylumbelliferone (4-MU).

The substrate for IDUA is 4-methylumbelliferyl a-L-iduronide (4-MU-α-IDUA). The chemical D-Saccharolactone is used to selectively inhibit endogenously present Bglucuronidase, which is active for a stereoisomer of 4-MU-α-IDUA (impurity in chemical synthesis) . The reaction is summarized below:

IDUA/pH 3.5 4-MU-∞-L-iduronide -----------------------------------------------------------------------------------------------------------------------------------------------------------

The substrate for GAA is 4-methylumbelliferyl a-D-glucoside (4-MU-a-Gluc). The chemical acarbose is used to selectively inhibit endogenously present maltose glucoamylase which is also active for 4-MU-α-Gluc67,899. The reaction is summarized below:

GAA/pH 3.8 4-MU-x-D-glucoside -----------------------------------------------------------------------------------------------------------------------------------------------------------

5 Chamoles N, Blanco, M, Gaggioli D. Diagnosis of a-L-iduronidase deficiency in dried blood spots on filter paper: the possibility of newborn diagnosis. Clinical Chemistry. 2001; 47:780-781.

6 Winchester B, Bali D, Bodamer OA, Caillaud C, Christiensen E, Cooper A., et al. Methods for a prompt and reliable laboratory diagnosis of Pompe disease: Report from an international consensus meeting. Molecular Genetics and Metabolism.2008; 93(3):275-281.

&#x27; Goldstein JL, Young SP, Changela M, et al. Screening for Pompe disease using a rapid dried blood spot method: experience of a clinical diagnostic laboratory. Muscle Nerve. 2009;40(1):32-36.

8 Li Y, Scott CR, Chamoles NA, Chavami A, Pinto BM, Turecek F, et al. Direct multiplex assay of lysosomal enzymes in dried blood spots for newborn screening. Clinical Chemistry. 2004; 50(10):1785-1796.

3 Kallwass H, Carr C, Gerrein J, Titlow M, Pomponio R, Bali D, et al. Rapid diagnosis of late-onset Pompe disease by fluorometric assay of alpha-glucosidase activities in dried blood spots. Molecular Genetics and Metabolism. 2007: 90(4):449-452.

{6}------------------------------------------------

The substrate for GBA is 4-methylumbelliferyl ß-D-glucopyranoside (4-MU-ß-Gluc). The surfactant sodium taurocholate (NaTC) is present in the buffer and is required to activate the enzyme. The reaction is summarized below:

GBA/pH 5.2/NaTC 4-MU-β-D-glucopyranoside -----------------------------------------------------------------------------------------------------------------------------------------------------

The substrate for GLA is 4-methylumbelliferyl α-D-galactopyranoside (4-MU-α-Gal). The chemical N-acetyl-D-galacosamine is used to inhibit endogenously present a-Nacetylgalactosaminidase which is also active for 4-MU-α-Gal. The reaction is summarized below:

GLA/pH 4.6 4-MU-x-D-galactopyranoside ---------------------------------------------------------------------------------------------------------------------------------------------------

For all the enzymatic assays the reaction is stopped using a high pH stop buffer (sodium bicarbonate, pH 11). The high pH is not optimal for the enzymes and effectively stops the substrate turnover. 4-MU is also not fluorescent at the reaction pH (3.5-5.2) and is highly fluorescent at the stopped condition (pH 11) with an excitation at 365 nm and emission at 460 nm.

4-MU (acid pH, non-fluorescent) --- 4-MU (fluorescent, Ex 365nm/Em 460nm)

All enzymatic reactions are performed at 37°C. After a pre-specified incubation time the free 4-MU is measured using a UV fluorimeter on the SEEKER Instrument. The fluorescence value of the 4-MU product is converted to a 4-MU concentration using a 4-MU calibration curve. The amount of 4-MU generated, after correction for substrate background and nonenzymatic hydrolysis, is proportional to the enzyme concentration. Substrate background and non-enzymatic hydrolysis is estimated by substituting the dried blood spot extract in the reaction with extraction buffer. Enzymatic activity is reported as micromoles of 4-MU produced / liter of blood / hour of incubation.

L. Performance Characteristics:

    1. Analytical performance:

a. Reproducibility/Precision:

A study to estimate the imprecision of the IDUA assay was performed using 4 instruments and 3 reagent lots, and performed testing during 21 non-consecutive days, with 2 runs per day and 2 punches from each DBS specimen per run. The samples were prepared by mixing human umbilical cord blood and heat inactivated serum (which has no endogenous lysosomal enzymes) to create samples with different enzyme concentrations. The hematocrit of was adjusted to ~50% to mimic newborn specimens and the samples were spotted onto Ahlstrom 226 grade filter paper. The study was based

{7}------------------------------------------------

on the recommendations in the CLSI EP5-A3 guideline. A total of 336 replicate results for each specimen were tested. There were several invalid test results for each sample for all assays. The results of the study are summarized below. In the following table the repeatability (or within-run precision) estimate includes the instrument run as the component of variability and the reproducibility estimate includes the instrument run, reagent lot, instrument and day as the components of variability.

Meanumol/L/h*NRepeatability(%CV)BetweenLot(%CV)BetweenInstrument(%CV)BetweenDay(%CV)Reproducibility(%CV)
3.5333422.7%15.6%0%7.4%27.2%
6.2233526.4%11.1%0%0.0%28.5%
12.0933415.4%11.0%0%0.0%18.8%
24.063359.0%9.8%0%0.0%14.2%

Results for the IDUA test

*There were a total of 7 invalid replicates obtained in the study resulting in a total N of less than 336 replicates per sample.

Results for the GAA test

Meanumol/L/h*NRepeatability(%CV)BetweenLot(%CV)BetweenInstrument(%CV)BetweenDay(%CV)Reproducibility(%CV)
4.2933115.6%5.8%0%14%17.0%
6.2733415.2%4.1%0%0.0%15.8%
9.593359.9%7.0%0%0.0%12.0%
18.0633413.6%5.9%0%0.0%14.8%
27.3733511.3%6.3%1.6%0.0%12.9%
  • There were a total of 11 invalid replicates obtained in the study resulting in a total N of less than 336 replicates per sample.

Results for the GBA test

Meanumol/L/h*NRepeatability(%CV)BetweenLot(%CV)BetweenInstrument(%CV)BetweenDay(%CV)Reproducibility(%CV)
2.8433134.9%12.7%1.8%2.5%38.0%
3.4733413.5%10.7%2.3%0.0%18.4%
5.0733511.0%11.2%0%0.0%16.6%
8.5533411.6%10.4%1.4%0.0%15.8%
15.0033511.3%11.4%2.1%1.2%15.7%

*There were a total of 11 invalid replicates obtained in the study resulting in a total N of less than 336 replicates per sample.

{8}------------------------------------------------

Meanumol/L/h*NRepeatability(%CV)BetweenLot(%CV)BetweenInstrument(%CV)BetweenDay(%CV)Reproducibility(%CV)
6.9433115.4%5.6%0%1.7%16.3%
9.8033410.4%6.4%0%2.9%13.6%
15.323357.7%7.2%0%3.1%11.5%
28.763347.8%4.8%0%0.0%9.4%
52.663358.8%4.1%1.3%2.7%10.6%

Results for the GLA test

  • There were a total of 11 invalid replicates obtained in the study resulting in a total N of less than 336 replicates per sample.

b. Linearity/assay reportable range:

Linearity studies were performed following the CLSI EP6-A Guideline using one lot of reagents. The samples used were dried blood spot specimens spotted onto Ahlstrom 226 grade filter paper. For the first study, a high sample pool was prepared by spiking human recombinant enzymes into heat inactivated serum (to remove endogenous lysosomal enzymes), and adjusting hematocrit to ~50% to correspond to the hematocrit of neonates. Eleven intermediate levels were prepared by diluting the high sample with heat inactivated serum and then adjusting hematocrit to ~50%. These 12 samples were tested in replicates of 12 using 4 instruments. A second study was performed using native enzymes. In this study, the high sample was prepared using human umbilical cord blood and the intermediate samples were made by mixing the high sample with heat inactivated serum and adjusting hematocrit to ~50%. A total of 12 samples were tested in replicates of 12 using 4 instruments.

The degree of nonlinearity was assessed by analyzing the second and third order polynomial regression. For any series where the polynomial fit was a better fit statistically, the maximum deviation from linearity was <10%. Acceptable linearity was demonstrated for the following ranges in the different studies:

AnalyteStudy 1Study 2
IDUA3.02 to 50.75 µmol/L/h2.61 to 19.40 µmol/L/h
GAA7.53 to 94.66 µmol/L/h2.12 to 21.79 µmol/L/h
GBA7.86 to 73.24 µmol/L/h2.14 to 10.42 µmol/L/h
GLA22.12 to 153.74 µmol/L/h4.38 to 45.34 µmol/L/h

Based on the results of the linearity studies and the limit of quantitation (LoQ) studies, the sponsor claims the following reportable ranges:

AnalyteReportable range
IDUA2.77 to 50.75 umo1/L/h
GAA2.18 to 94.66 umol/L/h
GBA2.14 to 73.24 umol/L/h
GLA4.88 to 153.74 umol/L/h

{9}------------------------------------------------

Four statistically high outlier test results were observed in the second study (please refer to section O below). Including these outliers in the analyses did not impact the claimed deviation from linearity. However, these 4 test results are outside the claimed imprecision performance of the test.

Test results below the LoQ are reported as "<LoQ (value)" with the actual LoQ value in parenthesis. Test results above the reportable range are reported as "> (value)" with the actual upper reportable limit in parenthesis. The following information is provided in the package insert:

It is recommended to retest specimens with activity above the upper end of the range to ensure that the elevated activity is not due to other pre-analytical causes such as specimen contamination. If the result of the retest is still above the upper end of the linear range, the specimen should be considered presumed normal.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Calibrator (traceability and value assignment): Since there is no reference material for 4-methyllumbelliferone, the calibrators are traceable to internal master calibrators. These master calibrators are prepared gravimetrically and value assigned using UV spectrophotometric analysis. Product calibrators are produced in the same manner as master calibrators. Product calibrators are value assigned by testing multiple replicates against the master calibrator lot. Specifications for calibrator target values and linear regression must be met. In addition, calibrators and reagents are tested together using DBS samples and predefined acceptance criteria for bias and precision must be met.

Quality control samples (value assignment): The Seeker Quality Control spots (QCL, QCM and QCH) are analyzed in multiple replicates on multiple cartridges. The sponsor calculates and provides mean enzymatic activity values and standard deviation for the quality control samples with each reagent kit. Each laboratory should establish their own mean and standard deviation values and determine their own acceptable range of enzymatic activity for quality control samples.

Calibrator stability: Accelerated stability studies were performed and supported that the calibrators are stable for 12 months when stored at -80°C. Real time studies are ongoing to support the claim based on the accelerated stability studies. The protocols were reviewed and found acceptable.

Ouality control sample stability: The sponsor performed accelerated stability studies and claims that the controls are stable for 12 months when stored at -80°C. Real time studies are ongoing to support the claim based on the accelerated stability studies. The protocols were reviewed and found acceptable.

Stability protocols designed to monitor reagent stability (which include the calibrators

{10}------------------------------------------------

and quality control samples) were reviewed and found acceptable.

The reagents (including calibrators and controls) are stable for 90 minutes after removing the aliquots from the freezer. The protocols for determining the in-use stability claim were reviewed and found acceptable.

Sample (DBS) stability: The sponsor provided information to support the following claims:

  • Lysosomal enzymes (IDUA, GAA, GBA and GLA) are stable for up to 5 days at ● 10°C at low humidity (20% RH) and high humidity (80% RH) levels.
  • . IDUA, GAA and GBA are stable for up to 5 days at 25℃ and 50% RH. There is a moderate loss in GLA activity (up to 18%) after 5 days at 25°C and 50% RH.
  • At 45°C and low humidity (20% RH), there is a moderate loss in IDUA (up to ● 32%) and GAA activity (up to 32%), and significant loss in GBA (up to 46%) and GLA activity (up to 66%).
  • At 45°C and high humidity (80% RH) there is significant loss in IDUA (up to . 81%), GAA (up to 84%), GBA (up to 69%) and GLA activity (up to 90%).
  • Lysosomal enzymes (IDUA, GAA, GBA and GLA) ●

The following limitation is in the labelling:

The SEEKER LSD Reagent Kit - IDUA|GAA|GBA|GLA may result in: Increased rate of false positives when the dried blood spot specimen is exposed to high temperature (≥45°C) during shipping.

  • d. Detection limit:
    The Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Ouantitation (LoO) for the test system was determined. The analysis of the LoB, LoD and LoQ were performed in accordance with CLSI document EP17-A2.

To determine the LoB, the reaction blank was used to prepare DBS specimens on Ahlstrom 226 grade filter paper. 80 replicates of the blank were tested using each of 3 reagent lots, 6 cartridges and 1 analyzer. The LoB was determined for each lot using the non-parametric classical approach described in the guideline. The α was set at 5%. The LoB for the worst performing lot is reported in the package insert.

To determine the LoD, low analyte DBS activity specimens were prepared by mixing human umbilical cord blood and heat inactivated serum, adjusting the hematocrit to ~50% and then spotting it onto Ahlstrom 226 grade filter paper. The DBS samples (8 per lot) were tested in 8 replicates using the 3 reagent lots, 6 cartridges and 1 analyzer (for a total of 64 replicates per lot). The LoD was determined for each lot following the parametric classical approach described in the guideline. The ß was set at 5%. The sponsor observed 1 statistically high outlier test result in this study (please refer

{11}------------------------------------------------

to section O below) and removed it from the analysis. This test result is outside the claimed imprecision performance of the test. The LoD for the worst performing lot is reported in the package insert.

To determine the LoQ, 8 specimens with varying enzyme activity were prepared by mixing human umbilical cord blood and heat inactivated serum. The hematocrit was then adjusted to ~50% and spotted onto Ahlstrom 226 grade filter paper. The DBS specimens were tested in 16 replicates. 3 reagent lots, 12 cartridges and 4 analyzers were used in the study. The LoQ was determined using the precision profile of each test and defined as the lowest concentration where the total imprecision was ≤ 1.5 umol/L/h or 20% CV, whichever was greater. Four statistically high outlier test results were observed and removed from the analysis (please refer to section O below). These 4 test results are outside the claimed imprecision performance of the test. The LoO for the worst performing lot is reported in the package insert. The sponsor also provided the impression observed with the worst performing lot at the claimed LoQ concentration for each enzyme.

Detection Capability (µmol/L/h)IDUAGAAGBAGLA
LoB1.780.500.721.96
LoD2.772.181.073.18
Detection Capability (µmol/L/h)IDUAGAAGBAGLA
LoQ2.772.181.854.88
Standard deviation at LoQ1.000.300.481.15
%CV at LoQ36.1%13.9%26%23.6%

The following information is provided in the package insert:

e. Analytical specificity:

Analytical specificity studies were performed following the recommendations in the CLSI EP7-A2 Guideline. A base pool was prepared to have enzyme concentrations around the cutoffs (high risk or borderline) for each enzyme (~ 5 umol/L/h for IDUA, ~ 6 umol/L/h for GAA, ~ 3 µmol/L/h for GBA and ~ 6 µmol/L/h for GLA) and the hematocrit was adjusted to 50% to mimic neonatal samples. The control pools were prepared by spiking in the appropriate solvent into the base pool. The test pools were prepared by spiking in the interfering substance at an appropriate concentration into the base pool. For the hematocrit interference studies, control and test pools were prepared by adjusting hematocrit to a target of 50% (control pool), 60% (test pool, level 2) and 70% (test pool, level 3). Test and control pools were spotted onto Ahlstrom 226 grade filter paper.

The control and test pools were tested in replicates of 30 and the means and the imprecision of the results for the test pools were compared to the means and the imprecision of the control pools. Bias estimates were calculated. The following substances at the listed concentrations did not interfere with the performance of the test:

{12}------------------------------------------------

SubstanceConcentration% bias observed
Unconjugatedbilirubin342 μM<+10%
Conjugated bilirubin342 µM<+10%
D-galactose0.84 mM<+10%
D-glucose55 mM<=II%
EDTA3.4 µM<+10%
Heparin3000 U/L<+13%
Intralipid1000 mg/dL<+10%

Total protein (at 120 mg/dL) did not interfere with the GAA, GBA, and GLA assay (max bias observed was ± 13%). Hematocrit (at 70%) did not interfere with the GAA, GBA, and GLA assay (max bias observed was ≤± 10%)

The following information is presented in the package insert to describe interferences observed with the IDUA assay:

For IDUA total protein at 75.8mg/mL did not interfere with the IDUA assay. However total protein at 88 mg/mL resulted in a bias of -16% and total protein at 100.3 mg/mL resulted in a bias of -31%.

For IDUA hematocrit values lower than 59% did not interfere with the test result. However at 66% hematocrit, bias of -38% was observed.

For IDUA, heparin at 3000 U/L may lead to an increased in imprecision.

In this study, 15 statistical outlier test results (5 IDUA, 2 GAA, 6 GBA and 2 GLA) were removed from the analysis (please refer to section O below). These 15 test results are outside the claimed imprecision of the test.

    1. Comparison studies:
    • a. Method comparison with predicate device: Not applicable, there is no predicate device.
    • b. Matrix comparison: This device is only intended to be used with DBS specimens collected from neonates.
    1. Clinical studies:

An evaluation of the test system was performed over the course of 24 months at a public state health laboratory that performs newborn screening. During the study 275 newborns were referred for confirmatory diagnosis. Of these 73 newborns were confirmed to have one of the 4 diseases. The 73 newborns included 1 newborn with MPS 1, 17 newborns with Pompe disease, 3 newborns with Gaucher disease and 52 newborns with Fabry

{13}------------------------------------------------

Incidence from StudyPublished Incidence
MPS I (IDUA)1:153,6971:54,000 - 1:185,000
Pompe (GAA)1:9,0411:28,000
Gaucher (GBA)1:51,2321:57,000
Fabry (GLA)1:2,9561:1,500 - 1:13,000

disease. The incidence from the study is summarized in the table below along with the published incidence.

The study consisted of two phases (delineated retrospectively once the study was complete): a pilot study of 6 months that included screening of 48,813 newborns and an 18 month pivotal phase where 105,599 newborns were screened. During the first 6 months of testing the device was modified several times and therefore the sponsor defined this period as a pilot phase because of the significant modifications to the device (please refer to the FDA Advisory Panel executive summary prepared for the August 10, 2016 Meeting of the Clinical Chemistry and Clinical Toxicology Devices Panel for information on the device modifications found online at:

http://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/ MedicalDevicesAdvisoryCommittee/ClinicalChemistryandClinicalToxicologyDevicesPa nel/ucm511565.htm).

A preliminary study was performed by the public state health laboratory prior to the pilot/pivotal studies to establish initial cutoff values. Approximately 13,000 presumed normal de-identified specimens and 29 known affected specimens were analyzed using the proposed device. The laboratory selected two cutoffs to evaluate test results in the pilot/pivotal studies: a borderline cutoff and a high risk cutoff for each enzyme. The cutoffs were selected to try to ensure that all known affected specimens would be detected and to minimize the false positive rate. The cutoffs used at the start of the study are shown below:

EnzymeHigh Risk (µmol/L/hr)Borderline (µmol/L/hr)
IDUA4.05.0
GAA8.010.0
GBA4.57.0
GLA5.57.0

The cutoff values were adjusted during the course of the study in order to reduce the false negative and false positive rate and also to take into account seasonal changes, i.e., increased heat and humidity that was found to affect the enzymatic activity during transport. The laboratory also instituted an additional set of cutoffs for specimens based on the age of collection as it was observed that enzymatic activity values decreased with the age of the newborn for GAA, GBA and GLA (please refer to FDA's executive summary referenced above for more information on cutoff modifications). The final high risk and borderline cutoff values (in (umol/L/hr) used at the end of the clinical study period are shown below:

{14}------------------------------------------------

Enzyme1-6 days of age7-13 days of age14+ days of age
High RiskBorderlineHigh RiskBorderlineHigh RiskBorderline
IDUA1.55.01.55.01.55.0
GAA7.210.04.510.04.510.0
GBA5.57.04.07.04.07.0
GLA7.09.05.05.03.05.0

The laboratory's screening procedure is described briefly below. (For detailed information, please refer to FDA's Advisory Panel executive summary at http://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/ MedicalDevicesAdvisoryCommittee/ClinicalChemistryandClinicalToxicologyDevicesPa nel/ucm511565.htm.)

The DBS cards from each neonate were tested in singlicate. Results were interpreted as follows:

    1. All valid test results above the borderline cutoff (and below an upper threshold set by the laboratory) were considered presumed normal.
  • Test results above the upper threshold were considered invalid. All DBS with 2. invalid test results were retested in singlicate.
    1. DBS with at least one enzyme below the borderline cutoff were retested in duplicate and the test results were evaluated by the laboratory for "visual outliers". (Note: If visual outliers were found, additional testing was performed.) If the average of the results (excluding test results considered visual outliers) was below the high risk cutoff a risk assessment was performed by the laboratory. The risk assessment included review of other newborn screening enzyme results, sample quality, transfusion status, health status, transit time, gestational age, family history and age at sample collection. Babies considered to still be at high risk for the screened condition after the risk assessment were referred for confirmatory testing. If the average of the results (excluding test results considered visual outliers) was above the high risk cutoff, then the babies were presumed normal.
  • Specimens collected from premature newborns or newborns aged < 24 hours at 4. the time the sample was collected automatically mandated a repeat screen (i.e., a new DBS card was collected from the baby) and the screening procedure was repeated.

A total of 105,599 newborns were screened during the pivotal phase. Five hundred ten (510) newborns were excluded from analysis since they had no valid DBS (collected <24 hours, no recorded age at collection, poor quality) or valid test results. A presumptive false positive rate was calculated by applying the cutoffs at the end of the study to the entire pivotal phase results. The pivotal phase study results are summarized below.

{15}------------------------------------------------

IDUAGAAGBAGLA
Total Newborns105,089
1st test below borderline8441,3987531,419
Average test result below high risk3910452134
Not referred after risk assessment10604479
Presumed Affected and referred forfollow-up2944855
True Positives07228a
Refused/moved2014
Presumptive False Positives(including newborns notreferred after risk assessment)379749102
Presumptive False Positive Rate0.035%0.092%0.047%0.097%
Presumptive False Negative Rate0000

4The number of true positives for Fabry newborns identified during the study was 30. Two newborns were identified as true positives while the lab was using a cutoff of 8.0umol/L/hr (compared to the final cutoff of 7.0umol/L/hr) to account for seasonal variation in the enzyme activity. Using a cutoff of 7.0µmol/L/hr could lead to false negatives.

The pivotal phase study results are summarized according to the age of the newborn at the time of sample collection in the tables below. The analysis is conducted by applying the cutoffs at the end of the study to the entire pivotal phase results:

IDUAIDUAIDUA
Age of newborn at time of sample collection(days)1 to 67 to 1314+
Total Newborns92,3345,3287,427
1st test below borderline49162159
Average test result below high risk2926
Not referred after risk assessment215
Presumed Affected and referred for follow-up2711
True Positives000
Refused/moved200
Presumptive False Positives (includingnewborns not referred after risk assessment)2726
Presumptive False Positive Rate0.029%0.038%0.081%
Presumptive False Negatives000

{16}------------------------------------------------

GAAGAAGAA
Age of newborn at time of sample collection(days)1 to 67 to 1314+
Total Newborns92,3345,3287,427
1 st test below borderline615123486
Average test result below high risk6827
Not referred after risk assessment2912
Presumed Affected and referred for follow-up3915
True Positives700
Refused/moved000
Presumptive False Positives (includingnewborns not referred after risk assessment)6127
Presumptive False Positive Rate0.066%0.038%0.094%
Presumptive False Negatives000
GBAGBAGBA
Age of newborn at time of sample collection(days)1 to 67 to 1314+
Total Newborns92,3345,3287,427
1 st test below borderline39966171
Average test result below high risk2915
Not referred after risk assessment1914
Presumed Affected and referred for follow-up701
True Positives200
Refused/moved100
Presumptive False Positives (includingnewborns not referred after risk assessment)2315
Presumptive False Positive Rate0.025%0.019%0.067%
Presumptive False Negatives000
GLAGLAGLA
Age of newborn at time of sample collection(days)1 to 67 to 1314+
Total Newborns92,3345,3287,427
1st test below borderline662122508
Average test result below high risk9795
Not referred after risk assessment4385
Presumed Affected and referred for follow-up5410
True Positives2800

{17}------------------------------------------------

GLAGLAGLA
Refused/moved400
Presumptive False Positives (includingnewborns not referred after risk assessment)6595
Presumptive False Positive Rate0.070%0.169%0.067%
Presumptive False Negatives200

Number of tests performed: A total of 120,118 DBS from 105,089 newborns were tested during the pivotal phase. The number of DBS is higher than the number of newborns since some newborns needed repeat screens and new DBS cards were collected. A total of 137,153 individual tests were performed on the 120,118 specimens. Of these 137,153 individual tests 7785 tests were test results (demonstrating an invalid test rate of 5.7%). However, one of the reasons for invalid test results is when the test is above an upper threshold. The remaining 9.250 tests were due to retesting triggered by activity values below borderline cutoff for one of the assays (as required by the screening protocol).

Note that during the clinical study the upper threshold for each enzyme was set above the reportable range of the assays. Since the completion of the clinical study, the upper thresholds for the assays have been changed to match the top value of the reportable range of the assays. Therefore, the invalid test rate due to a value above the upper threshold may be higher than observed during the clinical study.

The following is a summary of the limitations of the study included in the package insert:

Risk Assessment: Because test screen positive newborns that were presumed normal following the risk assessment process were not sent for follow up, clinical truth for these test screen positive newborns was not confirmed and false negatives are possible.

Acquiring repeat specimens: During the clinical study, the laboratory was not permitted to rescreen and collect new samples based on the results of the proposed device. Any additional screens were a result of premature infants, sick infants, infant samples collected at age <24 hours or other routine newborn screening results. Therefore, some samples that were suspected to be of poor quality (when for example all of the LSD test results were low) were presumed normal, according the laboratory's risk assessment procedure, and a repeat screen could not be collected.

Cutoffs below LoO: During the pivotal phase, two of the cutoffs used for the IDUA assay were below the limit of quantitation for IDUA (LOQ=2.77 umol/L/hr). A total of 49 newborns had IDUA test results between the high risk cutoff at the time of testing and the LOO and were presumed normal by the laboratory's screening procedure.

Additionally, one of the age specific cutoffs (14+ days of age) used for the GLA

{18}------------------------------------------------

assay was below the limit of quantitation (LOQ=4.88 umol/L/hr). A total of 50 newborns had GLA test results between the high risk cutoff at the time of testing and the LOO and were presumed normal by the laboratory's screening procedure.

Regarding the false negative rate of the test, the state public health laboratory has an active surveillance program to track any reports of false negative results to the contracted metabolic centers they use for confirming diagnosis of any of these LSDs. Based on information from this surveillance program there were no known false negative results due to an incorrect screening result during the 2 year study. It is presumed that the newborns screened during the clinical study with early onset disease would have been reported to one of these metabolic centers.

The package insert includes the following information about the false negative rate:

Certain late onset forms for Pompe disease may have GAA enzymatic activity in the normal range and result in a false negative 10.

For female Fabry disease patients GLA enzyme activity is highly variable and it could overlap with the normal range. Therefore. GLA enzyme activity for females should be interpreted with caution as some female carriers can have enzyme activity in the normal range and result in a false negative.

On August 10, 2016, a meeting of the Clinical Chemistry and Clinical Toxicology Devices Panel of the Medical Devices Advisory Committee was held to help inform FDA on the safety and effectiveness of the SEEKER System. The panel unanimously agreed that the probable benefits of the device outweighed the probable risks. Please refer to the transcript of the meeting found online at:

http://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/ MedicalDevicesAdvisoryCommittee/ClinicalChemistryandClinicalToxicologyDevicesPa nel/ucm511565.htm).

    1. Expected Values
      The 0.1th, 0.5th, 50th, 99.5th and 99th percentile values from the testing of presumed normal newborns with the proposed device at the state public health laboratory are summarized below for each age group. All values are in umol/L/hr.
IDUAGAAGBAGLA
%ileN=102,399N=102,392N=102,397N=102,371
0.1%3.687.616.697.56
0.5%5.6310.078.209.44

Reference limits for newborns age 1-6 days

10 Chiang S-C, Hwu W-L, Lee N-C, Hsu L-W, Chien Y-H. Algorithm for Pompe disease newborn screening: Results from the Taiwan screening program. Mol Genet Metab. 2012;106 (3):281-286.

{19}------------------------------------------------

IDUAGAAGBAGLA
50.0%18.3725.9820.2426.73
99.5%46.2364.5651.03123.36
99.9%>50.7577.4164.53>153.74

Reference limits for newborns age 7-13days

IDUAGAAGBAGLA
%ileN=7,177N=7,177N=7,177N=7,176
0.1%3.155.045.165.14
0.5%4.827.756.606.39
50.0%18.8221.5316.5619.70
99.5%>50.7573.0154.8396.19
99.9%>50.75>94.66>73.24134.27

Reference limits for newborns age >14 days

IDUAGAAGBAGLA
%ileN=7,447N=7,447N=7,447N=7,447
0.1%2.834.404.314.88
0.5%3.936.435.835.36
50.0%19.4117.7614.6114.64
99.5%>50.7554.6845.2167.34
99.9%>50.7572.2164.54113.61

The labeling for the SEEKER System, states that each laboratory should establish their own reference ranges and cutoffs.

M. Instrument Name:

SEEKER Instrument

N. System Description:

1. Modes of Operation:

Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device? Yes X

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission: Yes

{20}------------------------------------------------

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes X__ or No ________________________________________________________________________________________________________________________________________________________

2. Specimen Identification:

Specimens are extracted in a 96-well plate and the extracts from one plate are transferred to two SEEKER cartridges. The software will prompt the user to enter the sample and reagent layout information for that cartridge.

3. Specimen Sampling and Handling: Specimens are processed according to the package insert instructions.

    1. Calibration:
      The 4 calibrators are run with each cartridge. The device software calculates a slope, intercept and R2 by linear regression for every run. If the calibration for a run is flagged as "failed" no activity values are reported for the entire run and the entire run must be repeated.
    1. Quality Control:
      QC samples are provided with the kit. Each cartridge should include at least one punch of two levels of OC (OCL and OCM) to determine the validity of the run. These OC samples are subjected to the same protocol steps as the newborn specimen and the enzymatic activity for all 4 assays is determined. Mean enzymatic activity values and standard deviation is provided with each reagent included in the reagent kit. Each laboratory should establish mean and standard deviation values and determine their own acceptable range of enzymatic activity for QC samples.

The instrument also checks the non-enzymatic hydrolysis value for each enzyme against a pre-defined range. If the values are outside this range, the activity is not calculated and an "n/a" result is reported for the activity.

Test results above the upper threshold (which is the upper limit of the reportable range) are highlighted and these samples should be considered invalid and retested.

O. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:

Outliers: In the analytical studies statistical high outliers were observed at a rate of 0.103% for IDUA and GAA, 0.144% for GBA and 0.092% for GLA. These test results were outside the expected performance of the assays and were excluded from the analysis where noted above. If they occur during routine testing, these incorrectly elevated test results could result in a false negative test. The sponsor stated that because of the low prevalence of the diseases and the low frequency of the observed outliers, the likelihood that an affected neonate would obtain an incorrectly elevated test result is very low.

{21}------------------------------------------------

Carry-over: Carry-over was evaluated using two specimen layouts for each test: one with carry-over and another with no carry-over. The carry-over layout included DBS with concentrations representing affected samples and DBS with concentrations representing normal to high enzyme activity (samples 1 to 4). The no carry-over layout only included DBS with concentrations representing affected samples (no carry-over). Carry-over was estimated as the percent bias between the average values of representative affected specimens obtained between the carry-over and no carry-over layout. One result determined to be a statistical outlier was observed. The results without the outlier are summarized below. In the table below the results of the affected samples with and without carry-over are reported in the columns labeled "affected sample" and the results of the samples with normal to high activity are reported in the columns labeled "high sample".

AssayAffected SampleHigh SampleCarryoverµmol/L/hrBias
Carry-overSampleNMeanµmol/L/hrCarry-overSampleNMeanµmol/L/hr
Nocarryover1504.07
IDUA1644.0018813.55-0.07-2%
2644.2328826.160.164%
4644.3443253.260.277%
Nocarryover115*8.87
GAA1638.7718816.59-0.10-1%
2609.2428430.280.374%
4648.9743258.890.101%
Nocarryover1496.37
GBA1646.3418813.62-0.030%
2637.0428825.110.6710%
3626.7233236.180.356%
4647.2343249.570.8613%
GLANocarryover1529.11
1649.2318830.600.111%
264/63*11.59/10.28*28858.992.48/1.17**27%/13%*
36410.3933283.401.2814%
46411.07432120.791.9521%

*For the GAA study, one run failed and was not repeated.

**For the GLA study one test result identified as a statistical outlier was removed from the analysis.

Based on the results, the following recommendation is included in the package insert:

{22}------------------------------------------------

A specimen with GLA activity within +2umol/L/hr of the borderline cutoff should be retested if a specimen with GLA activity >120umol/L/hr is present in the same specimen column (cartridge has 12 columns and 4 rows).

P. Proposed Labeling:

The labeling is sufficient and satisfies the requirements of 21 CFR Parts 801 and 809, and the special controls for this type of device.

Q. Patient Perspectives:

This submission did not include specific information on patient perspectives for this device. However, during the August 10, 2016 meeting of the Clinical Chemistry and Clinical Toxicology Devices Panel of the Medical Devices Advisory Committee for the SEEKER System, multiple patients (or the parents of patients) spoke during the public comment period. All patients (or the parents of patients) were supportive of the authorization for marketing of this device. See the transcript of the meeting found online at: http://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/Me dicalDevicesAdvisoryCommittee/ClinicalChemistryandClinicalToxicologyDevicesPanel/uc m511565.htm

{23}------------------------------------------------

R. Identified Risks to Health and Required Mitigations:

Identified Risks to HealthRequired Mitigations
Inaccurate test results that provide falsenegative test results could lead to anewborn to not be detected as a possiblelysosomal storage disorder case and to bedelayed from timely therapy.Special Controls (1) and (2)
Inaccurate test results that provide falsepositive test results could lead a newbornto have unnecessary additionalconfirmatory testing and to add emotionalburden to the family of the newborn.Special Controls (1) and (2)

S. Benefit/Risk Analysis

Summary
Summary of theBenefit(s)Newborns may potentially benefit from the use of the device as anaid in screening because it might assist practitioners in makingearlier decisions that diagnostic testing is needed for patients withMPSI, Pompe, Gaucher, and Fabry disease which can lead toearlier potentially beneficial treatment.
Summary of theRisk(s)Associated device risks include erroneous false negative results dueto device or user error or false positive results due to device or usererror. A false positive result could result in unnecessary additionaltesting for the newborn and add emotional burden to the family ofthe newborn. A patient with a false negative result could fail to bedetected as a possible LSD case and be delayed from timelytherapy.
Summary of OtherFactorsOn August 10, 2016, a meeting of the Clinical Chemistry andClinical Toxicology Devices Panel of the Medical DevicesAdvisory Committee was held to help inform FDA of the safetyand effectiveness of the SEEKER System. Clinicians includinggeneticists and experts in the field of clinical chemistry andbiostatistics were members of the panel. The panel unanimouslyagreed (10 out of 10) that the probable benefits of the deviceoutweighed the probable risks for the proposed intended use andprovided recommendations for the types of information that shouldappear in the labeling for this device to provide a reasonableassurance of the safety and effectiveness of the device. The panelstated that they supported the granting of this de novo given the

{24}------------------------------------------------

real-world large clinical study performed and the lack of currentlyavailable FDA cleared/approved devices for lysosomal storagedisease screening in newborns (unmet need), and the potential forstandardization in testing between states. Given the results of thepivotal clinical study and the special controls incorporating thepanel's recommendations, which include demonstrating thescreening performance of the device and including information ontest performance including the false positive rate, and the falsenegative rate observed in the clinical study in the labeling, willmitigate the risk of erroneous false positive and false negativeresults, the probable benefits of this device outweigh the probablerisks for this device.
ConclusionsDo the probable benefits outweigh the probable risks?

Yes, the probable benefits outweigh the probable risks of this device in light of the special controls assigned, along with general controls, including design controls.

T. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 862.1488. FDA believes that special controls, along with the applicable general controls, provide reasonable assurance of the safety and effectiveness of the device type. This device is classified under the following:

Product Code:PQW, PQT, PQU, PQV
Device Type:Lysosomal storage disorder newborn screening test system
Class:II (special controls)
Regulation:21 CFR 862.1488

(a) Identification. A lysosomal storage disorder newborn screening test system is intended to measure lysosomal enzyme levels obtained from dried blood spot specimens on filter paper from newborns as an aid in screening newborns for a lysosomal storage disorder.

(b) Classification. Class II (special controls). A lysosomal storage disorder newborn screening test system must comply with the following special controls:

    1. Premarket notification submissions must include information that demonstrates the performance characteristics of the device, including:
    • (i) Study results that adequately demonstrate the clinical validity of the device, which shall include information supporting the link between the analyte being measured and the condition being screened. The clinical validity of the device must be demonstrated in a clinical validation study using either well characterized prospectively or retrospectively obtained clinical specimens from the intended use

{25}------------------------------------------------

population. Testing in the clinical validation study must be performed by operators representative of the types of operators intended to use the test. The study design of the clinical validation study must assess the effects of sample collection and processing steps on test performance. Confirmed positive specimens must have a diagnosis based on confirmatory diagnostic methods or clinically meaningful information regarding the status of the subject must be obtained.

  • (ii) The reference interval in the normal newborn population for the analyte or analytes measured by the device.
  • (iii)Study results demonstrating the level of carry-over or drift affecting the device performance.
  • (iv)Study results demonstrating the concentrations of the limit of blank, limit of detection, and limit of quantitation of the device. Sample concentrations below the limit of quantitation should not be reported by the device.
  • (v) Study results, which shall be collected using sample panels from at least three reagent lots and at least three instruments over more than 20 testing days, demonstrating the imprecision of the device. The sample panels must consist of blood spot specimens with a range of analyte concentrations that span the reportable range of the device and must include samples with concentrations in the screen positive range, samples with concentrations at each cutoff, and samples with concentrations in the normal range.
    1. Your 21 CFR 809.10 compliant labeling for this device must include:
    • (i) A warning that reads "This test is not intended to diagnose lysosomal storage disorders."
    • (ii) A warning that reads "Test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, and clinical evaluation as appropriate."
    • (iii)Detailed information on device performance, including the false positive rate and the false negative rate observed in the clinical study.
    • (iv)Information on device performance in any relevant subgroup (e.g., age of newborn at time of sample collection, birth weight, sex, gestational age, race, ethnicity) observed in the clinical study.

§ 862.1488 Lysosomal storage disorder newborn screening test system.

(a)
Identification. A lysosomal storage disorder newborn screening test system is intended to measure lysosomal enzyme levels obtained from dried blood spot specimens on filter paper from newborns as an aid in screening newborns for a lysosomal storage disorder.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include information that demonstrates the performance characteristics of the device, including:
(i) Study results that adequately demonstrate the clinical validity of the device, which must include information supporting the link between the analyte being measured and the condition being screened. The clinical validity of the device must be demonstrated in a clinical validation study using either well-characterized prospectively or retrospectively obtained clinical specimens from the intended use population. Testing in the clinical validation study must be performed by operators representative of the types of operators intended to use the test. The study design of the clinical validation study must assess the effects of sample collection and processing steps on test performance. Confirmed positive specimens must have a diagnosis based on confirmatory diagnostic methods or clinically meaningful information regarding the status of the subject must be obtained.
(ii) The reference interval in the normal newborn population for the analyte or analytes measured by the device.
(iii) Study results demonstrating the level of carryover or drift affecting the device performance.
(iv) Study results demonstrating the concentrations of the limit of blank, limit of detection, and limit of quantitation of the device. Sample concentrations below the limit of quantitation must not be reported by the device.
(v) Study results, which must be collected using sample panels from at least three reagent lots and at least three instruments over more than 20 testing days, demonstrating the imprecision of the device. The sample panels must consist of blood spot specimens with a range of analyte concentrations that span the reportable range of the device and must include samples with concentrations in the screen positive range, samples with concentrations at each cutoff, and samples with concentration in the normal range.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A warning that indicates that the test is not intended to diagnose lysosomal storage disorders.
(ii) A warning that indicates that test results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, and clinical evaluation as appropriate.
(iii) Detailed information on device performance, including the false positive rate and the false negative rate observed in the clinical study.
(iv) Information on device performance in any relevant subgroup (
e.g., age of newborn at time of sample collection, birth weight, sex, gestational age, race, ethnicity) observed in the clinical study.