AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Device Description

This 510(k) is for a labeling modification only, to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

AI/ML Overview

The provided document, K173585, describes the a 510(k) premarket notification for a labeling modification to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System (Model IS4000) and the da Vinci X Surgical System (Model IS4200). The document does not describe the acceptance criteria and study proving a device's performance against acceptance criteria in the manner typically seen for AI/ML-enabled devices. Instead, it focuses on demonstrating the substantial equivalence of adding VHR procedures to the existing robotic surgical systems by comparing the outcomes of robotic-assisted VHR to traditional open and laparoscopic VHR.

Therefore, many of the specific questions about acceptance criteria for an AI/ML device's performance, sample sizes for test sets, establishment of ground truth by experts, MRMC studies, and standalone algorithm performance, as requested in the prompt, are not applicable or directly addressed by this document. This document is about a surgical system, not an AI/ML diagnostic or prognostic tool.

However, I can extract information related to the clinical study data used to support the labeling modification, which serves a similar purpose to a performance study for a new indication.

Here's the information as it relates to the provided document:

1. A table of acceptance criteria and the reported device performance:

The document doesn't define explicit numerical "acceptance criteria" in the way one would for an AI/ML device (e.g., "sensitivity must be >X%"). Instead, it demonstrates comparability and, in some cases, superiority of the robotic-assisted VHR to existing surgical methods across several clinical outcomes and measures. The "acceptance" for this 510(k) is based on demonstrating substantial equivalence and safety and effectiveness for the new indication compared to existing methods.

The tables provided (Tables 1, 2, 3, and 4a) present the "reported device performance" in terms of clinical outcomes for robotic-assisted VHR procedures compared to open and laparoscopic VHR. Below is a summary of the key findings, which serve as the "performance" data used to support the new indication.

Summary of Device Performance (Robotic-Assisted VHR vs. Comparators)

Outcome/MeasureRobotic-Assisted vs. Open (Non-Complex VHR)Robotic-Assisted vs. Laparoscopic (Non-Complex VHR)Robotic-Assisted vs. Open (Complex VHR)Robotic-Assisted vs. Laparoscopic (Complex VHR) - Unmatched AHSQC Data & Published Literature
Length of StayComparable (2 ± 7 days vs. 2 ± 14 days)Shorter (2 ± 7 days vs. 4 ± 13 days)Shorter (2 ± 3 days vs. 5 ± 7 days)Shorter in AHSQC data (2 ± 3 days vs. 4 ± 3 days). Published literature varies (1-9.2 days).
Intraoperative ComplicationsComparableComparableComparableComparable in AHSQC data (2% vs. 11%). Published literature (0-1.5%).
Transfusions (Intra/Post-op)ComparableComparableComparableComparable (0% across all AHSQC cohorts). Published literature (0 or 51-91.7ml EBL).
Postoperative Complications (through 30 days)Comparable (9% vs. 11%)Comparable (10% vs. 11%)Comparable (24% vs. 20%)Mixed: Robotic-assisted (24%) vs. AHSQC Laparoscopic (11%). Published literature (0-60%).
Readmission Rates (through 30 days)Comparable (2% vs. 3%)Comparable (2% vs. 3%)Comparable (7% vs. 4%)Comparable in AHSQC data (7% vs. 7%). Published literature (0-13%).
Re-encounter Rates (through 30 days)Lower clinic re-encounter for robotic-assisted (29% vs. 41%). Comparable ER.ComparableLower clinic re-encounter for robotic-assisted (16% vs. 27%). Comparable ER.Mixed: Robotic-assisted (17% Clinic, 5% ER) vs. AHSQC Laparoscopic (22% Clinic, 15% ER). Published literature (0).
Reoperation Rates (through 30 days)Comparable (<1% vs. 1%)Comparable (<1% vs. 1%)Comparable (3% vs. 1%)Mixed: Robotic-assisted (3%) vs. AHSQC Laparoscopic (0%). Published literature (0-20%).
Recurrence Rates (through 30 days)Comparable (1% vs. 0%)Comparable (1% vs. 1%)Comparable (1% vs. 1%)Comparable in AHSQC data (1% vs. 0%). Published literature "Not Reported" or low.
Mortality (through 30 days)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable in AHSQC data (<1% vs. 0%). Published literature (0-1.9%).
Operative TimeLonger for robotic-assisted (e.g., 45% between 60-119 min vs. 35%).Longer for robotic-assisted (e.g., 43% between 60-119 min vs. 50%).Longer for robotic-assisted (e.g., 45% 240+ min vs. 17%).Longer for robotic-assisted (46% 240+ min vs. 19%). Published literature (125-372 min).
Conversion Raten/an/an/aRobotic-assisted (4%) vs. AHSQC Laparoscopic (70%). Published literature (0%).

Note: The document explicitly states that longer operative times for robotic-assisted cases were "not associated with increases in the complication, readmission, reoperation or mortality rates." This addresses a potential negative finding.

2. Sample sizes used for the test set and the data provenance:

  • Test Set (Clinical Data): Real-world evidence (RWE) from the AHSQC registry database.
    • Non-Complex VHR Procedures:
      • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=871 for each group.
      • Robotic-Assisted vs. Laparoscopic: Propensity score matched cohorts of N=615 for each group.
      • Data Provenance: The registry data for these procedures occurred between July 7, 2013, and January 1, 2017. The document does not specify the country of origin, but AHSQC (America's Hernia Society Quality Collaborative) suggests it is primarily from the United States. The data is retrospective as it's from a registry of past procedures.
    • Complex VHR Procedures:
      • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=297 for each group.
      • Robotic-Assisted vs. Laparoscopic: Unmatched data from the AHSQC Registry Database (Robotic-Assisted N=305, Laparoscopic N=27) and a summary of published literature of laparoscopic VHR (N=3-53 per study).
      • Data Provenance: Similar to non-complex, July 7, 2013, and January 1, 2017, primarily United States, retrospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable in the context of this document. The "ground truth" for the performance data is derived directly from clinical outcomes recorded in a surgical registry (AHSQC) and published clinical literature. These are factual clinical occurrences (e.g., length of stay, complications, readmissions, mortality) rather than subjective interpretations by experts, as would be the case for image-based diagnostic AI/ML.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. As noted above, the data are clinical outcomes from a registry, not expert-labeled data requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human "readers." The "device" (da Vinci Xi/X Surgical System) is a surgical tool. The study compares surgical approaches (robotic, open, laparoscopic), not the performance of human readers with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. The da Vinci system is a human-controlled robotic surgical system; it does not operate "stand-alone" as an algorithm would. Performance is always human-in-the-loop (i.e., the surgeon uses the system).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The ground truth for the clinical performance data (test set) is Real-World Evidence (RWE) from a clinical registry (AHSQC). This constitutes outcomes data (e.g., length of stay, complication rates, readmission rates, recurrence rates, mortality rates) directly recorded from patient care. For certain comparisons (Complex VHR Robotic-Assisted vs. Laparoscopic), it also incorporates published literature on laparoscopic VHR outcomes.

8. The sample size for the training set:

The document does not describe a "training set" in the context of machine learning. The da Vinci Xi and X Surgical Systems are existing electromechanical systems; this submission is for an expanded indication. The "training" for the system itself has already occurred and is inherent in its design and previous clearances. The data used in this submission serves as clinical validation data for the expanded Indication for Use.

9. How the ground truth for the training set was established:

Not applicable, as there is no "training set" in the AI/ML sense described.

In summary, this 510(k) submission demonstrates the substantial equivalence and safety/effectiveness of adding a new surgical indication to an already cleared robotic surgical system, not the performance of a new AI/ML device against specific quantitative acceptance criteria. The "study" here is a retrospective analysis of clinical outcomes from a large surgical registry and published literature.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left, and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".

July 19, 2018

Intuitive Surgical, Inc. % Ms. Cindy Domecus Principal, Domecus Consulting Services, LLC/ISI Chief Regulatory Advisor Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010

Re: K173585

Trade/Device Name: da Vinci Xi Surgical System (Model IS4000), da Vinci X Surgical System (Model IS4200) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: June 18, 2018 Received: June 19, 2018

Dear Ms. Domecus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Jennifer R. Stevenson -23

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K173585

Device Name

Intuitive Surgical® da Vinci® Xi Endoscopic Instrument Control System (da Vinci Xi System, Model IS4000) and Endoscopic Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K173585

Device Name

Intuitive Surgical® da Vinci® X Endoscopic Instrument Control System (da Vinci X System, Model IS4200) and Endoscopic Instruments and Accessories

(da Vinci X System, Model IS4200) and Endoscopic Instruments and Access

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary (21 CFR § 807.92(c))

I. SUBMITTER INFORMATION

Submitter:Intuitive Surgical, Inc.1266 Kifer RoadSunnyvale, CA 94086
Contact:Cindy Domecus, R.A.C. (US & EU)Principal, Domecus Consulting Services LLCChief Regulatory Advisor to Intuitive SurgicalTelephone: 650.343.4813Fax: 650.343.7822Email: domecusconsulting@comcast.net
Date Summary Prepared:July 19, 2018
II. SUBJECT DEVICE INFORMATION
Device Trade Name:da Vinci® Xi and X Surgical Systems, Model IS4000 and Model IS4200
Common Name:System, Surgical, Computer Controlled Instrument
Classification Name:Endoscope and Accessories (21 CFR §876.1500)
Regulatory Class:II
Product Code:NAY
Submission Type:Traditional 510(k)

III. PREDICATE DEVICE INFORMATION:

Predicate Device: Intuitive Surgical da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200 (K131861, K152578, K153276, K161178, K170713, K171294, K171632) Intuitive Surgical da Vinci Si Surgical System, Model IS3000 (K081137, K123463, K090993)

IV. DEVICE DESCRIPTION:

This 510(k) is for a labeling modification only, to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

{5}------------------------------------------------

V. INDICATIONS FOR USE

da Vinci Xi Surgical System

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System, Model: IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

da Vinci X Surgical System

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model: IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Precaution for Representative Uses

The demonstration of safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of cancer (overall survival, disease-free survival, local recurrence) or treatment of the patient's underlying disease/condition. Device usage in all surgical procedures should be guided by the clinical judgment of an adequately trained surgeon.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (IS4000 and IS4200) proposed in this submission.

{6}------------------------------------------------

VII. PERFORMANCE DATA

Pre-Clinical Animal Study Data

This premarket notification is supported by animal study data including the results from six (6) evaluations in a total of 24 animals demonstrating use of da Vinci Xi Surgical System in the following procedures: Hysterectomy and Salpingectomy/Oophorectomy (Adnexectomy), Pyeloplasty, Nephrectomy, Nissen Fundoplication, Colectomy and Mitral Valve Repair. These data were initially submitted in support of clearance of the da Vinci Xi Surgical System (K131861).

Real World Evidence from the AHSQC Registry Database

Real world evidence (RWE) from the AHSQC registry database support use of the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) in "Ventral Hernia Repair" (VHR) procedures that fall under the cleared "general laparoscopic surgical procedures" Indication for Use. The RWE provided included propensity matched comparative data from robotic-assisted, open and laparoscopic cohorts in noncomplex VHR procedures. Non-complex VHR procedures were defined as ventral (e.g., incisional, epigastric, umbilical) hernia repairs without myofascial release. Additionally, propensity matched comparative data from robotic-assisted and open cohorts and unmatched comparative data from robotic-assisted and laparoscopic cohorts in complex VHR procedures were also included. Complex VHR procedures were defined as incisional hernia repair plus myofascial release. Longer term endpoints beyond 30 days, such as ventral hernia repair recurrence and chronic pain were not assessed in this analysis, and thus no claims regarding these clinical outcomes data should be inferred. The results demonstrated the safety and effectiveness of robotic-assisted ventral hernia repair procedures.

Non-Complex VHR Procedures: Robotic-Assisted v. Open Surgery

Table 1 summarizes the results of the RWE that compared the results of non-complex robotic-assisted and open VHR procedures. The RWE included information comparing the use of the da Vinci System in non-complex Ventral Hernia Repair procedures with open surgery in the following key measures:

  • Length of Stay: comparable lengths of hospital stay were reported for both the robotic-assisted and open cohorts.
  • Intraoperative Complications: comparable intraoperative complication rates were reported for both the robotic-assisted and open cohorts.
  • Transfusions: comparable intraoperative blood transfusion rates were reported in both the robotic-assisted and open cohorts.
  • Postoperative Complications through 30 days: comparable postoperative complications through 30 days follow up were reported for both the robotic-assisted and open cohorts.
  • Readmission Rates through 30 days: comparable readmission rates through 30 days follow up were reported for both the robotic-assisted and open cohorts.
  • Re-encounter Rates through 30 days: a lower clinic re-encounter rate through 30 days follow up was reported for the robotic-assisted cohort (29%) as compared to the open cohort (41%). Comparable emergency room re-encounter rates through 30 days follow up were reported for both the roboticassisted and open cohorts.
  • Reoperation Rates through 30 days: comparable reoperation rates were reported for both the robotic-assisted and open cohorts.
  • Recurrence Rates through 30 days: comparable surgeon reported recurrence rates were reported for both the robotic-assisted and open cohorts through the 30-day postoperative period.
  • Mortality through 30 days: comparable mortality rates were reported in both the robotic-assisted and open cohorts through the 30-day postoperative period.

{7}------------------------------------------------

  • Operative Time: longer operative times were reported for the robotic-assisted cohort as compared to the open cohort1.
TABLE 1: Propensity Score Matched da Vinci and Open Non-Complex Ventral Hernia Repair (without
myofascial release)^
OutcomeRobotic-Assisted Cohort (N=871)Open Cohort (N=871)
Patient Demographics
Mean age, y (± SD)56 ± 1455 ± 14
Female Gender, n (%)399 (46)361 (41)
Male Gender, n (%)472 (54)510 (59)
Mean BMI, kg/ m2 (± SD)33 ± 732 ± 7
Length of Stay (days), mean ± SD2 ± 72 ± 14
Intraoperative Complications, n (%)11 (1)5 (1)
Transfusions, n (%)
Intraoperative1 (< 1)0 (0)
Postoperative0 (0)2 (< 1)
Postoperative Complications, n (%)*64 (9)75 (11)
Readmission Rates, n (%)*14 (2)21 (3)
Reoperation Rates, n (%)*3 (< 1)8 (1)
Re-encounter Rates, n (%)*
Clinic207 (29)291 (41)
Emergency Room31 (4)20 (3)
Mortality, n (%)*1 (< 1)0 (0)
Operative Time, minutes, n (%)
0 – 59111 (13)432 (50)
60 – 119396 (45)304 (35)
120 - 179247 (28)86 (10)
180 - 23981 (9)28 (3)
240+36 (4)21 (2)
Conversion Rate, n (%)5 (1)Not Applicable
Recurrence Rate, n (%)*5 (1)0 (0)

^ Includes data from the AHSQC registry for procedures that occurred between July 7, 2017 . The total number of non-complex VHR procedures in the registry during this time period was 873 and 3,979 for the robotic-assisted and open cohorts, respectively. The total number in the propensity matched cohorts is 871 as noted in the column header of the above table.

Non-Complex VHR Procedures: Robotic-Assisted v. Laparoscopic Surgery

Table 2 summarizes the results of the RWE that compared the results of non-complex robotic-assisted and open VHR procedures. The RWE included information comparing the use of the da Vinci System in non-complex Ventral Hernia Repair procedures with laparoscopic surgery in the following key measures:

  • Length of Stay: a shorter length of hospital stay was reported for the robotic-assisted cohort (2 days ± 7) as compared to the laparoscopic cohort (4 days ± 13).
  • Intraoperative Complications: comparable intraoperative complication rates were reported for both the robotic-assisted and laparoscopic cohorts.
  • Transfusions: comparable intraoperative blood transfusion rates were reported in both the robotic-assisted and laparoscopic cohorts.

4 The longer operative time for the robotic-assisted cohort was not associated with increases in the complication, readmission, reoperation or mortality rates.

{8}------------------------------------------------

  • Conversion Rate: a comparable conversion rate was reported for the robotic-assisted cohort as compared to the laparoscopic cohort.
  • Postoperative Complications through 30 days: comparable postoperative complications through 30 days follow up were reported for both the robotic-assisted and laparoscopic cohorts.
  • Readmission Rates through 30 days: comparable readmission rates through 30 days follow up were reported for both the robotic-assisted and laparoscopic cohorts.
  • Re-encounter Rates through 30 days: comparable clinic and emergency room re-encounter rates through 30 days follow up were reported for the robotic-assisted cohort as compared to the laparoscopic cohort.
  • Reoperation Rates through 30 days: comparable reoperation rates were reported for both the robotic-assisted and laparoscopic cohorts.
  • Recurrence Rates through 30 days: comparable surgeon reported recurrence rates were reported for both the robotic-assisted and laparoscopic cohorts through the 30-day postoperative period.
  • Mortality through 30 days: comparable mortality rates were reported in both the robotic-assisted and laparoscopic cohorts through the 30-day postoperative period.
  • Operative Time: longer operative times were reported for the robotic-assisted cohort as compared to the laparoscopic cohort2.

TABLE 2: Propensity Score Matched da Vinci and Laparoscopic Non-Complex Ventral Hernia Repair (without myofascial release)^

OutcomeRobotic-Assisted Cohort (N=615)Laparoscopic Cohort (N=615)
Patient Demographics
Mean age, y (± SD)55 ± 1456 ± 14
Female Gender, n (%)305 (50)284 (46)
Male Gender, n (%)310 (50)331 (54)
Mean BMI, kg/ m² (± SD)33 ± 733 ± 8
Length of Stay (days), mean ± SD2 ± 74 ± 13
Intraoperative Complications, n (%)6 (1)8 (1)
Transfusions, n (%)
Intraoperative1 (< 1)0 (0)
Postoperative0 (0)1 (< 1)
Postoperative Complications, n (%)*55 (10)59 (11)
Readmission Rates, n (%)*8 (2)17 (3)
Reoperation Rates*, n (%)1 (<1)5 (1)
Re-encounter Rates*, (%)
Clinic153 (29)181 (35)
Emergency Room21 (4)18 (4)
Mortality, n (%)*1 (<1)0 (0)
Operative Time, minutes, n (%)
0 – 5986 (14)175 (28)
60 - 119265 (43)308 (50)
120 - 179175 (28)103 (17)
180 - 23960 (10)21 (3)
240+29 (5)8 (1)
Conversion Rate, n (%)3 (< 1)14 (2)
Recurrence Rate, n (%)4 (1)4 (1)

2 The longer operative time for the robotic-assisted cohort was not associated with increases in the complication, readmission, reoperation or mortality rates.

{9}------------------------------------------------

^ Includes data from the AHSQC registry for procedures that occurred between July 7, 2013 and January 1, 2017. The total number of non-complex VHR procedures in the registry during this time period was 873 and 1,961 for the robotic-assisted and laparoscopic cohorts, respectively. The total number in the propensity matched cohorts is 615 as noted in the column header of the above table.

Complex VHR Procedures: Robotic-Assisted v. Open Surgery

Table 3 summarizes the results of the RWE that compared the results of complex robotic-assisted and open VHR procedures. The RWE included information comparing the use of the da Vinci System in complex Ventral Hernia Repair procedures with open surgery in the following key measures:

  • Length of Stay: a shorter length of hospital stay was reported in the robotic-assisted cohort (2 days ± 3) as compared to the open cohort (5 days ± 7).
  • Intraoperative Complications: comparable intraoperative complication rates were reported for both the robotic-assisted and open cohorts.
  • Transfusions: comparable intraoperative blood transfusion rates were reported in both the robotic-assisted and open cohorts.
  • Postoperative Complications through 30 days: comparable postoperative complication rates through 30 days follow up were reported for both the robotic-assisted cohort and open cohort.
  • Readmission Rates through 30 days: comparable readmission rates through 30 days follow up were reported for both the robotic-assisted and open cohorts.
  • Re-encounter Rates through 30 days: a lower clinic re-encounter rate through 30 days follow up was reported for the robotic-assisted cohort (16%) as compared to the open cohort (27%). Comparable emergency room re-encounter rates through 30 days follow up were reported for both the roboticassisted and open cohorts.
  • Reoperation Rates through 30 days: comparable reoperation rates were reported for both the robotic-assisted cohort and open cohorts.
  • Recurrence Rates through 30 days: comparable surgeon reported recurrence rates were reported for both the robotic-assisted and open cohorts through the 30-day postoperative period.
  • Mortality through 30 days: comparable mortality rates were reported in both the robotic-assisted and open cohorts through the 30-day postoperative period.
  • Operative Time: longer operative times were reported for the robotic-assisted cohort as compared to the open cohort3.

TABLE 3: Propensity Score Matched da Vinci and Open Complex Ventral Hernia Repair (with myofascial release)^

OutcomeRobotic-Assisted Cohort (N=297)Open Cohort (N=297)
Patient Demographics
Mean age, y (± SD)57 ± 1357 ± 13
Female Gender, n (%)169 (57)159 (54)
Male Gender, n (%)128 (43)138 (46)
Mean BMI, kg/ m² (± SD)33 ± 733 ± 7
Length of Stay (days), mean ± SD2 ± 35 ± 7
Intraoperative Complications, n (%)7 (2)9 (3)

3 The longer operative time for the robotic-assisted cohort was not associated with increases in the intraoperative complication, readmission, recurrence or mortality rates.

{10}------------------------------------------------

OutcomeRobotic-Assisted Cohort (N=297)Open Cohort (N=297)
Transfusions, n (%)
Intraoperative0 (0)0 (0)
Postoperative0 (0)1 (< 1)
Postoperative Complications, n (%)*66 (24)54 (20)
Readmission Rates, n (%)*19 (7)12 (4)
Reoperation Rates, n (%)*8 (3)4 (1)
Re-encounter Rates, n (%)*
Clinic44 (16)74 (27)
Emergency Room14 (5)12 (4)
Mortality, n (%)*1 (< 1)0 (0)
Operative Time, minutes, n (%)
0 – 590 (0)8 (3)
60 - 11912 (4)84 (28)
120 - 17953 (18)106 (36)
180 - 23998 (33)48 (16)
240+134 (45)51 (17)
Conversion Rate, n (%)12 (4)Not Applicable
Recurrence Rate, n (%)*4 (1)2 (1)

^ Includes data from the AHSQC registry for procedures that occurred between July 7, 2013 and January 1, 2017. The total number of complex VHR procedures in the registry during this time period was 305 and 3,106 for the robotic-assisted and open cohorts, respectively. The total number in the propensity matched in the column header of the above table.

Complex VHR Procedures: Robotic-Assisted v. Laparoscopic Surgery

Table 4a provides a summary of the results of the RWE that compared the results of complex roboticassisted and laparoscopic VHR procedures and a summary of the published literature on complex laparoscopic VHR procedures. The limited sample size of the complex laparoscopic cohort in the AHSQC Registry Database precluded matching and a comparative analysis between these cohorts in the AHSQC Registry Database. Table 4b includes the detailed data from nine (9) publications reporting on complex laparoscopic VHR procedures. These publications were selected based on specific search criteria and filters and included 6 comparative studies (LOE 2 & 3b) and 3 single-arm studies (LOE 5). The results from the literature were not matched with the available robotic-assisted data from the AHSQC Registry Database and no comparative analysis was conducted. Clinical analysis of the da Vinci Surgical Systems in robotically-assisted complex ventral hernia repair procedures raises no new issues of safety or effectiveness for the proposed labeling modification and therefore is substantially equivalent.

TABLE 4a: Unmatched** da Vinci and Laparoscopic Complex Ventral Hernia Repair from the AHSQC Registry Database (with myofascial release)

OutcomeRobotic-AssistedCohort (N=305)AHSQC LaparoscopicCohort (N=27)Summary of Table 4bPublished LiteratureLaparoscopic Cohort(N = 3 - 53)
Patient Demographics
Mean age, y (± SD)57 ± 1358 ± 1248.6 - 70 years
Female Gender, n (%)175 (57)14 (52)20 - 68%
Male Gender, n (%)130 (43)13 (48)32 - 80%
Mean BMI, kg/ m² (± SD)33 ± 734 ± 729.71 - 36 kg/ m²
Length of Stay (days), mean ± SD2 ± 34 ± 31 - 9.2 days
Intraoperative Complications, n (%)7 (2)3 (11)0 - 1.5%

{11}------------------------------------------------

OutcomeRobotic-AssistedCohort (N=305)AHSQC LaparoscopicCohort (N=27)Summary of Table 4bPublished LiteratureLaparoscopic Cohort(N = 3 - 53)
Transfusions, n (%)Transfusions: 0
Intraoperative0 (0)0 (0)Estimated Blood Loss:
Postoperative0 (0)0 (0)51 – 91.7 ml
Postoperative Complications, n (%)*66 (24)3 (11)0 – 60%
Readmission Rates, n (%)*19 (7)2 (7)0 – 13%
Reoperation Rates, n (%)*8 (3)0 (0)0 – 20%
Re-encounter Rates, n (%)*
Clinic48 (17)6 (22)0
Emergency Room14 (5)4 (15)
Mortality, n (%)*1 (< 1)0 (0)0 – 1.9%
Operative Time, minutes, n (%)
0 – 590 (0)0 (0)125 – 372 minutes
60 – 11912 (4)3 (11)
120 - 17954 (18)8 (30)
180 - 239100 (33)11 (41)
240+139 (46)5 (19)
Conversion Rate, n (%)12 (4)19 (70)0
Recurrence Rate, n (%)*4 (1)0 (0)Not Reported

** The data presented in this table are not propensity score matched and include the results for all complex VHR cases in the AHSQC registry during the access dates between July 7, 2013 and January 1, 2017 for both the robotic-assisted and laparoscopic cohorts.

{12}------------------------------------------------

K173585

Author/YearLap/EndoscopicStudy Size (N)Mean Age, y (±SD)Gender (M/F)Mean BMI,kg/m² (SD)Mean Operation Time(minutes) (SD)EBL (ml) or TranfusionsLength of Stay,days (SD)ConversionRate (%)IntraoperativeComplicationRate (%)PostoperativeComplication Rate (%)Mortality(in-hospital or30 days)ReoperationRate (30d)Re-encounterRate (30d)ReadmissionRate (30 d)RecurrenceRate (30d)
Azoury 20142559 (11)32% / 68%35.4 (8)278 (73)63 (29)4 (2)NRNR20%(Wound complications)0.0%NRNRNRNR
Belyanksy 201637066.7% / 33.3%30.1329 (OR time)91.74.70.0%0.0%33%(1/3 subjects, prolongedileus prior to discharge)0.0%0.0%0.0%0.0%NR
Belyanksy 201738NRNRNot stratified by cohortNRNRNRNRNRNR2.6% (1/38)0%0 (90d)NR(likely 0 frominference)0 (90d)NR
Daes 20145NRNRNot stratified by cohortNRNRNR1NR00.0%0.0%0.0%0.0%0.0%NR
Giurgius 20122151 (13)48% / 52%36 (7)229 (57)NR6.3 (3.6)NRNR19.0%NRNRNRNRNR
Moazzez 2013548.6 (7.9)80%/20%29.71 (3.4)372 (104)0 transfusions(No significant bleedingreported)9.2 (5.4)0.0%0.0%60% (3/5) subjects withacute (<30d)/in-hospitalcomplications0.0%20% (1/5)NRNRNR for 30 day(60% withmedian of12.6m)
Muse 20185354.440%/60%35.3224NR5NRNR13%(Wound complications)1.9%(Timeframe NR)NRNR13%NR for 30 day(15% overall)
Parker 2011864.41.7:1(reported asratio)30.5226.851 mL0 units transfused5.4NRNR0%(Wound complications,timeframe NR)1.9%NRNR0%NR
Schroeder 201343(27 incisional)5762.8% / 37.2%NR125NR3.4NR1.5%7.0%NRNRNRNRNR for 30 d(0 overallthrough f/u)

|--|

NR = Not Reported

{13}------------------------------------------------

VIII. CONCLUSION

Based on the information provided in this premarket notification, the inclusion of "Ventral Hernia Repair" procedures under the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) previously cleared "general laparoscopic surgical procedure" Indication for Use is substantially equivalent to the predicate devices. The basis of this substantial equivalence determination was an assessment of the immediate post-operative outcomes through 30 days comparing the robotic-assisted surgery ventral hernia repair versus laparoscopic ventral hernia repair.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.