AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical® da Vinci Si Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system is indicated for adult and pediatric use. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Device Description

The Intuitive Surgical da Vinci Si Surgical System, model IS3000 (also known as the Endoscopic Instrument Control System) is a modification of the existing da Vinci S Surgical System, model IS2000. Essentially, the IS3000 is an updated and enhanced version of the existing IS2000 device. The new IS3000 provides a more intuitive user interface and improved ease-of-use. Overall, the model IS3000 is a computer-assisted device designed to facilitate complex surgery using a minimally invasive approach. The IS3000 consists of the following 3 main components: Surgeon Console (model SS3000), Patient Cart (model PS3000), and Vision System Cart (model VS3000).

AI/ML Overview

The provided 510(k) summary for the Intuitive Surgical da Vinci Si Surgical System (Model IS3000) does not contain information about specific acceptance criteria or a detailed study proving the device meets particular performance metrics in the way typically expected for AI/ML-based medical devices or diagnostic tools.

Instead, this submission is for an updated version of an existing surgical system, establishing substantial equivalence to a predicate device (da Vinci S Endoscopic Instrument Control System, Model IS2000). The focus is on demonstrating that the new model (IS3000) does not raise new safety or effectiveness issues and is substantially equivalent to legally marketed devices.

Therefore, many of the requested categories for AI/ML device evaluations are not applicable or not provided in this regulatory document.

However, I can extract the available information:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance (Implicit)
Safety and Effectiveness Equivalence: The device does not raise any new safety or effectiveness issues compared to the predicate device."The Intuitive Surgical da Vinci Si Surgical System (model IS3000) does not raise any new safety or effectiveness issues and is substantially equivalent to legally marketed computer-controlled surgical instrument systems...""All testing yielded acceptable results and was comparable to the predicate device."
Compliance with Technical Standards: The device meets applicable technical and FDA-recognized consensus standards.The IS3000 complies with applicable portions of:- UL/IEC 60601-1:1988, Amendment A1: 1991-11, Amendment A2: 1995 (Medical Electrical Equipment - Part 1: General requirements for safety)- IEC 60601-1-1:2000 (Medical Electrical Equipment - Part 1: General requirements for safety; Collateral standard: Safety requirements for medical electrical systems)- IEC 60601-1-2:2001 (Medical Electrical Equipment - Part 1-2: General requirements for safety; Collateral standard: Electromagnetic compatibility - requirements and tests)- IEC 60601-1-4:2000 (Medical Electrical Equipment - Part 1: General requirements for safety; Collateral standard: Programmable electrical medical systems)- IEC 60601-2-2:1998 (Medical Electrical Equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment)- IEC 60601-2-18:1996, Amendment A1: 2000 (Medical Electrical Equipment - Part 2: Particular requirements for the safety of endoscopic equipment)
Functional Equivalence: Improvements in user interface and ease-of-use do not compromise the fundamental function or safety of the device and its components (Surgeon Console, Patient Cart, Vision System Cart, EndoWrist Instruments).The IS3000 is described as "an updated and enhanced version of the existing IS2000 device" with "a more intuitive user interface and improved ease-of-use." The components serve the same fundamental purpose as the predicate. "There are no changes to the EndoWrist Instruments in this submission."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Performance (bench) testing was conducted on the IS3000." However, it does not provide details on the sample size used for this testing, the specific test cases, or the provenance of any data (as it's a surgical system, not a data-driven diagnostic in the AI/ML sense). It's bench testing, implying laboratory-based functional assessments rather than patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth, in the context of an AI/ML diagnostic's test set, is not a concept relevant to the type of bench testing described for this surgical system. The "truth" would be determined by engineering specifications and functional performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study requiring human adjudication of results or images.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a surgical control system, not an AI-assisted diagnostic or image interpretation device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a surgical system designed for human-in-the-loop operation by a surgeon.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this type of device, the "ground truth" for bench testing typically refers to engineering specifications, design requirements, and established performance parameters of the predicate device (IS2000). Compliance with these, and with technical standards, would form the basis of the "truth."

8. The sample size for the training set

Not applicable. This document describes a medical device, not an AI/ML algorithm that is "trained" on a dataset. The "training" for such a system would involve engineering design, development, and iterative testing.

9. How the ground truth for the training set was established

Not applicable. As above, this is not an AI/ML algorithm with a traditional training set.

{0}------------------------------------------------

081137

510(k) Summary

Intuitive Surgical, Inc. da Vinci Si Surgical System; model IS3000

FEB 1 8 2009

General Information

Trade nameIntuitive Surgical® da Vinci® Si™ Surgical System
Product nameIntuitive Surgical da Vinci Si Surgical System
Catalog/Model NumberModel IS3000
Common NameEndoscopic instrument control system,endoscopic instruments and accessories
ClassificationClass II
510(k) OwnerIntuitive Surgical, Inc.1266 Kifer RoadSunnyvale, CA 94086
Contact PersonNitya NarayananInternational Regulatory Project Managernitya.narayanan@intusurg.com
ManufacturerProduct will be manufactured and distributed byIntuitive Surgical, Inc. - mailing and facility address:Intuitive Surgical, Inc.950 Kifer RoadSunnyvale, CA 94086
Establishmentregistration number:2955842
Owner/Operatornumber:9028901

Summary of Substantial Equivalence

The Intuitive Surgical da Vinci Si Surgical System (model IS3000) does not raise any new safety or effectiveness issues and is substantially equivalent to legally marketed computer-controlled surgical instrument systems that are in commercial distribution, and have been determined to be substantially equivalent to devices in commercial distribution, prior to May 28, 1976.

Date: April 18, 2008

510(k) Notification INTUITIVE SURGICAL, INC.

{1}------------------------------------------------

Substantially Equivalent Devices

ManufacturerSubstantially equivalent device510(k)
Intuitive Surgical, Inc.Sunnyvale, CAda Vinci S Endoscopic InstrumentControl System (Model IS2000, V1.1),Endoscopic instruments andaccessoriesK050369K063220

Device Description

The Intuitive Surgical da Vinci Si Surgical System, model IS3000 (also known as the Endoscopic Instrument Control System) is a modification of the existing da Vinci S Surgical System, model IS2000. Essentially, the IS3000 is an updated and enhanced version of the existing IS2000 device. The new IS3000 provides a more intuitive user interface and improved ease-of-use.

Overall, the model IS3000 is a computer-assisted device designed to facilitate complex surgery using a minimally invasive approach. The IS3000 consists of the following 3 main components:

  • . Surgeon Console (model SS3000) - is the control center for the IS3000 system. While seated at the surgeon console (outside of the sterile field) and as observed through the console's stereo viewer, the surgeon controls critical aspects of the surgical procedures, including movement of the endoscopic instruments and endoscope within the operative field. Instrument and camera movements are controlled by the surgeon through use of 2 handoperated controllers residing in the console, and foot pedals at the console's base. These features allow the surgeon to be as dexterous as in "open" surgery, while operating in a minimally invasive environment. The surgeon at the console also has the option to change the surgical view from full screen mode to a multi-image mode, which displays the 3-dimensional (3-D) image of the operative field.
  • Patient Cart (model PS3000) is the operative component of the surgical . system within the sterile field. Its primary function is to support the EndoWrist® instrument arms and camera arm, during surgical procedures.
    • . EndoWrist Instruments are designed to provide surgeons with natural dexterity and a greater range of motion than even the human hand. This allows for greater precision when operating in a minimally invasive environment. EndoWrist instruments, when used with the IS3000 system, are designed to support rapid and precise suturing, dissection and tissue manipulation in surgical procedures. There are no changes to the EndoWrist Instruments in this submission.

{2}------------------------------------------------

  • Vision System Cart (model VS3000) houses the system's image processing � equipment and is operated by a person outside of the sterile field during surgery. The cart provides space for an optional touch screen monitor, as well as ancillary surgical equipment. The cart is also known as Intuitive Surgical Insite Vision System (Model VS3000) and consists of a stereo endoscope, endoscopic camera and various accessories, including a light source and light guides. The Insite Vision System provides 2 independent images that are relayed to the viewer located in the surgeon console, where they are fused to form a 3-D image (or alternatively 2-D) of the surgical field for the surgeon's reference.

Indications for Use

The Intuitive Surgical Endoscopic Instrument Control System (Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system is indicated for adult and pediatric use. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

{3}------------------------------------------------

Bench/Animal/Clinical Testing

Performance (bench) testing was conducted on the IS3000. All testing yielded acceptable results and was comparable to the predicate device.

The IS3000 complies with applicable portions of the following technical standards, or FDA-recognized consensus standards:

UL/IEC 60601-1:1988, Amendment A1: 1991-11, Amendment A2: 1995Medical Electrical Equipment - Part 1: General requirements for safety
IEC 60601-1-1:2000, Medical Electrical Equipment - Part 1:General requirements for safety; Collateral standard:Safety requirements for medical electrical systems
IEC 60601-1-2:2001, Medical Electrical Equipment - Part 1-2:General requirements for safety; Collateral standard:Electromagnetic compatibility - requirements and tests
IEC 60601-1-4:2000, Medical Electrical Equipment - Part 1:General requirements for safety; Collateral standard:Programmable electrical medical systems
IEC 60601-2-2:1998, Medical Electrical Equipment - Part 2-2:Particular requirements for the safety of high frequency surgical equipment
IEC 60601-2-18:1996, Amendment A1: 2000Medical Electrical Equipment - Part 2:Particular requirements for the safety of endoscopic equipment

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Intuitive Surgical, Inc. % Ms. Karen Uyesugi VP, Clinical and Regulatory Affairs 1266 Kifer Road Sunnyvale, California 94086

Re: K081137

Trade/Device Name: da Vinci Si Surgical System IS3000 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NAY Dated: December 23, 2008 Received: January 5, 2009

Dear Ms. Uyesugi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

EB 1 8 2009

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Ms. Karen Uyesugi

Enclosure

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)); please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Mark A. Millken

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K081137

Device Name: Intuitive Surgical® da Vinci Si Surgical System (IS3000)

Indications For Use: The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical® da Vinci Si Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surqical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system is indicated for adult and pediatric use. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED): . . .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil RP Ogale for MXM

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) Number K081137

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.