(168 days)
The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including, cement retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
CCM Abutment System is used with a dental implant to provide support to prosthetic restorations such as crown, bridge and overdentures in partially or fully edentulous patients. CCM Abutment System consists of UCLA Abutment and titanium abutment screws. All subject abutments are intended to be cast at zero angulations and placed on dental implant bodies which do not require angle correction.
This document is a 510(k) Premarket Notification from the FDA, asserting "substantial equivalence" of the CCM Abutment System to previously marketed devices. These types of documents focus on comparing a new device to existing ones to demonstrate similar safety and effectiveness, rather than presenting a detailed study proving the new device meets specific, pre-defined quantitative acceptance criteria through clinical trials or standalone performance evaluations (especially relevant for AI/ML devices). This document describes a dental implant abutment system, which is a physical device, not an AI/ML-driven medical device. Therefore, the information typically requested for AI/ML device performance, such as sample sizes for test sets, data provenance, expert ground truth establishment, MRMC studies, or standalone algorithm performance, is not applicable to this document.
The document primarily relies on non-clinical data to demonstrate substantial equivalence, focusing on materials, design, indications for use, and manufacturing processes compared to predicate devices.
Here's the information that can be extracted relevant to acceptance and performance for this physical device, framed in the context of "substantial equivalence" rather than "meeting acceptance criteria" in the way an AI/ML device would:
Acceptance Criteria and Device Performance (for a physical dental abutment system)
The "acceptance criteria" here are implicitly tied to demonstrating substantial equivalence to legally marketed predicate devices. The performance is assessed through non-clinical testing and comparison of attributes.
Since this is a 510(k) for a physical dental abutment, the "acceptance criteria" are not quantitative performance metrics in the same way they would be for an AI/ML device. Instead, they are about demonstrating that the device is as safe and effective as predicate devices.
Table of "Acceptance Criteria" (Implicit for 510(k)) and Reported Device Performance:
| Feature/Test | Implicit "Acceptance Criteria" (Demonstrate equivalence to predicate) | Reported Device Performance / Justification |
|---|---|---|
| Intended Use | Same as predicate devices | The CCM Abutment System is intended to restore chewing function by replacing missing teeth, supporting single or multiple-unit restorations (cement-retained, screw-retained, fixed bridgework), and is for one/two-stage surgical procedures with delayed loading. This is stated to be the same as the primary predicate (K153350) and aligns with the general use of the reference predicates (K140806, K162099, K161244). |
| Technological Characteristics (Design, Dimension, Principle of Operation, Restoration Angulations) | Similar to predicate devices, with differences not raising new safety/effectiveness questions. | Similarities: Identical machining, manufacturing process, angulation (No Angle), and principle of operation (screw-retained/cement-retained depending on part). Differences: Slight differences in specific dimensions (diameters and lengths) of various abutment parts (UCLA Abutment, Burn out core cylinders/caps, Multi-unit CCM Cylinder) compared to predicates are noted but are considered not to raise new questions of safety or effectiveness. The device maintains "No Angle" for restoration angulations, consistent with predicates. |
| Material Composition | Equivalent to predicate devices, or justified if different. | Difference: Subject device uses Co-Cr-Mo Alloy and Poly Diacetate. The primary predicate (K153350) uses Titanium Alloy and Poly Diacetate. Justification: Reference predicate (K161244) uses Co-Cr-Mo Alloy, and biocompatibility testing was performed on the subject device to support the material difference. Galvanic reaction testing was also performed between CoCr alloy and Ti Gr.5. |
| Biocompatibility | Meets established standards (ISO 10993). | Biocompatibility tests performed according to ISO 10993-1, -3, -5, -10, -11. Results "met the criteria of the standards" and "demonstrates that the subject device is biocompatible and substantial equivalence with the predicate device, K161244." |
| Sterilization | Meets established standards (ISO 17665, ANSI/AAMI ST79). | End User Steam Sterilization Test referenced from predicate K140806. Justified because "product category, material, manufacturing process, facility, and packaging is the exactly the same as the predicate." |
| Galvanic Reaction | Acceptable corrosion potential between alloys. | Testing performed for CoCr alloy and Ti Gr.5 as per "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implant and Endosseous Dental Implant abutments." Assessed corrosion potential and couple potential. "Results...met the criteria of the standards." |
| Performance Validation (General Statement) | Supposed to performance and product validations prior to release. | "The subject device has been supposed to performance and product validations prior to release." (General statement, specific test data or detailed results not provided in this summary.) |
Study Details (as applicable to a physical device 510(k) submission):
-
Sample size used for the test set and the data provenance:
- Not applicable in the context of an AI/ML device. This document does not refer to a "test set" of patient data for algorithm evaluation.
- For the non-clinical tests (biocompatibility, galvanic reaction), the "sample size" would refer to the number of physical device units tested according to the respective ISO/ASTM standards. These details are not provided in this summary but would be part of the full submission. The provenance is internal lab testing for a new device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. "Ground truth" in this context is implicitly established by the standards themselves (e.g., ISO 10993 specifications for biocompatibility) and engineering principles for mechanical functionality. Human expert consensus on a test set of images or data is not a component of this submission.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This refers to a process for resolving discrepancies in expert labeling or diagnoses, which is not relevant for this physical device testing.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a study design for evaluating diagnostic AI, not a physical dental device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This pertains to AI/ML software.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical tests, the "ground truth" is defined by the specific criteria and methodologies outlined in the referenced international standards (e.g., ISO 10993 for biocompatibility, ISO 17665 for sterilization). For galvanic reaction, it's based on "corrosion potential" and "couple potential" as assessed per FDA guidance for dental implants.
-
The sample size for the training set:
- Not applicable. This is for AI/ML model development.
-
How the ground truth for the training set was established:
- Not applicable. This is for AI/ML model development.
In summary, this document is a regulatory communication for a Class II physical medical device, not an AI/ML-driven diagnostic tool. As such, the requested details related to AI/ML study design and performance criteria are not present because they are not relevant to the type of device and submission being reviewed. The "study" proving the device meets "acceptance criteria" is the set of non-clinical bench tests and comparisons performed against established predicate devices and international standards, demonstrating "substantial equivalence."
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
March 16, 2018
InnoBioSurg Co., Ltd. April Lee Consultant Withus Group Inc. 2531 Pepperdale Drive Rowland Heights, California 91748
Re: K173120
Trade/Device Name: CCM Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 14, 2018 Received: February 14, 2018
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K173120
Device Name CCM Abutment System
Indications for Use (Describe)
The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including, cement retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary
Submitter
InnoBioSurg Co., Ltd. Bo-reum Yoo 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027 Republic of Korea Email: bryoo@ibsimplant.com Tel. +82-42-933-2879 Fax. +82-42-933-2881
Official Correspondent
Withus Group Inc April Lee 106 Superior Irvine CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
Device Information
- Trade Name: CCM Abutment System
- Common Name: Dental Abutment System
- . Classification Name: Endosseous dental implant abutment
- Product Code: NHA
- . Panel: Dental
- Regulation Number: 872.3630
- Device Class: Class II
- Date prepared: 03/16/2018
Primary Predicate
K153350, IBS Implant System by Innobiosurg Co., Ltd.
Reference Predicates
- K140806,IBS Implant System by Innobiosurg Co., Ltd
- K162099, IBS Implant System II by Innobiosurg Co., Ltd.
- K161244, s-Clean OneQ-SL Narrow Implant System by Dentis Co., Ltd.
Device Description
CCM Abutment System is used with a dental implant to provide support to prosthetic restorations such as crown, bridge and overdentures in partially or fully edentulous patients. CCM Abutment System consists of UCLA Abutment and titanium abutment screws. All subject abutments are intended to be cast at zero angulations and placed on dental implant bodies which do not require angle correction.
UCLA Abutment
UCLA Abutments are pre-manufactured prosthetic abutments directly connected to the endosseous dental implant intended for permanent restoration, for either single or multiple tooth screw retained restorations. The lower part of the abutment which connects directly to the implant is made of Co-Cr-Mo alloy, and the upper part is made of plastic (Poly Diacetate). The plastic part is dissolved once the casting is done. The abutment screw is connected to the implant by a prosthetic screw, supplied with the abutment, and made of Ti-6AL-4V Eli. UCLA Abutments are intended for no angulation and straight implantation only. The abutments are supplied non-sterile, to be steam sterilized by the user according to the labeling, and intended for single use.
{4}------------------------------------------------
Multi-unit CCM Cylinder
Multi-unit CCM Cylinders are used in conjunction with screw retained type Abutment to provide support for screw type final prosthesis, and for fabrication of custom abutment for screw retained restorations.
The subject device is compatible with the following Implants:
| K number | Compatible Implants |
|---|---|
| K140806 | IBS Implant system |
| K152520 | Magicore System |
| K153350 | IBS Implant System |
The CCM Abutment system is provided non-sterile and packaged separately. The abutments should be sterilized before use. Abutment screw (OIAS400) was cleared from K140806.
The UCLA abutments, Burn out core cylinder and cap are not intended to be cast at angulation or placed to provide angular correction. Plastic sleeve of UCLA Abutment can replace the Non-precious metal (Co-Cr Alloy). All finished devices are final abutments.
The diameters and lengths of the CCM Abutment System are below:
1) UCLA Abutment
The gingival height could be 1-4mm, angulation straight only, length of abutment post of 4-10mm. Assembled torque: 30N·cm
| No | Diameter (Ø) | Cuff | Length | Remark |
|---|---|---|---|---|
| 1 | 1 mm | 14 mm | ||
| 2 | 4 mm | 2 mm | 15 mm | |
| 3 | 3 mm | 16 mm | ||
| 4 | 4 mm | 17 mm | ||
| 5 | 1 mm | 14 mm | ||
| 6 | 4.5 mm | 2 mm | 15 mm | |
| 7 | 3 mm | 16 mm | ||
| 8 | 4 mm | 17 mm | Hex/Non-Hex | |
| 9 | 1 mm | 14 mm | ||
| 10 | 5 mm | 2 mm | 15 mm | |
| 11 | 3 mm | 16 mm | ||
| 12 | 4 mm | 17 mm | ||
| 13 | 1 mm | 14 mm | ||
| 14 | 5.5 mm | 2 mm | 15 mm | |
| 15 | 3 mm | 16 mm | ||
| 16 | 4 mm | 17 mm |
{5}------------------------------------------------
2) Burn out core cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 5 mm | 8.5 mm | Hex/Non-Hex |
| 2 | 6 mm | 8.65 mm |
3) Burn out core cap (Cemented type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 5 mm | 9.15 mm | Hex/Non-Hex |
| 2 | 6 mm | 9.15 mm |
4) Burn out core T cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 5 mm | 10.65 mm | Hex/Non-Hex |
| 2 | 6 mm | 10.8 mm |
5) Burn out core U cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 5 mm | 12.25 mm | Hex/Non-Hex |
| 2 | 6 mm | 12.4 mm |
6) Burn out core U cap (Cemented type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 5 mm | 12.95 mm | |
| 2 | 6 mm | 12.95 mm | Hex/Non-Hex |
{6}------------------------------------------------
7) Multi-unit CCM Cylinder
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|---|---|---|---|
| 1 | 4.8 mm | 12 mm | Hex/Non-Hex |
Indication for Use
The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Non-Clinical Data:
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ● ISO 10993-10:2010 and ISO 10993-11:2006.
- Galvanic Reaction Test between CoCr alloy and non-precious metal ●
Below tests were performed for predicate devices and leveraged for the subject device:
- End User Steam Sterilization Test according to ISO 17665-1:2006, -2:2009 and ANSI/AAMI ● ST79 referenced in K140806
The results of the above tests have met the criteria of the standards, and demonstrated the substantial equivalence with the predicate device.
Biocompatibility Test was conducted for the subject devices and it demonstrates that the subject device is biocompatible and substantial equivalence with the predicate device, K161244.
The galvanic reaction testing between the CoCr alloy and non-precious dental alloys, Ti Gr.5. was performed for subject device and compatible implants. As the FDA guidance of "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implant and Endosseous Dental Implant abutments", corrosion potential of each CoCr alloy and Ti Gr.5 and couple potential for assembled CoCr alloy and Ti Gr.5 were assessed.
The end user sterilization test was performed for predicate device, K140806 and leveraged for the subject device because the product category, material, manufacturing process, facility, and packaging is the exactly the same as the predicate, K140806.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
{7}------------------------------------------------
Comparison to Predicate Devices:
| UCLA Abutment | |||
|---|---|---|---|
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System | |
| 510k | N/A | K153350 | K161244 |
| Part Name | UCLA Abutment(Hex, Non-Hex) | UCLA Abutment | S-clean CCM Abutment(Narrow) |
| Material | Co-Cr-Mo AlloyPoly Diacetate | Titanium AlloyPoly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System isintended to replace missingteeth to restore chewingfunction. The CCM AbutmentSystem can be placed insupport of single or multiple-unit restorations including;cement retained, screwretained, and terminal orimmediate abutment supportfor fixed bridgework. Thissystem is for one or two stagesurgical procedures. Thissystem is intended for delayedloading. | The IBS Implant System isintended to replace missingteeth to restore chewingfunction. The IBS Implantcan be placed in support ofsingle or multiple-unitrestorations including;cement retained, screwretained, or overdenturerestorations, and terminal orimmediate abutment supportfor fixed bridgework. Thissystem is for one or twostage surgical procedures.This system is intended fordelayed loading. | The s-Clean One-Q-SL Narrow Implant System(3.0, 3.3mm) may be used as an artificial rootstructure for single tooth replacement ofmandibular central and lateral incisors andmaxillary lateral incisors. The implants may berestored immediately1) with a temporary prosthesis that is not infunctional occlusion,2) when splinted together as an artificial rootstructure for multiple tooth replacement ofmandibular incisors, or3) for denture stabilization using multipleimplants in the anterior mandible and maxilla.The implants may be placed in immediatefunction when good primary stability has beenachieved and with appropriate occlusal loading. |
| Dimension | Ø 4mm, Ø 4.5mm, Ø 5mm, Ø5.5mm (D) X 14,15,16,17 mm(L) | Ø3.5mm, Ø4mm, Ø4.5mm,Ø5mm, Ø5.5mm (D) X14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
{8}------------------------------------------------
| Restoration angulations | No Angle | No Angle | No Angle |
|---|---|---|---|
| Design | |||
| Principle of Operation | a screw retained restorationtype of abutment using a screwto fix a prosthesis | a screw retained restorationtype of abutment using ascrew to fix a prosthesis | a screw retained restoration type of abutmentusing a screw to fix a prosthesis |
| Burn out core cylinder (Hex and Non-Hex) | |||
|---|---|---|---|
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System | |
| 510k | N/A | K153350 | K161244 |
| Part Name | Burn out core cylinder | UCLA Abutment | S-clean CCM Abutment(Narrow) |
| Material | Co-Cr-Mo AlloyPoly Diacetate | Titanium AlloyPoly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System isintended to replace missing teeth torestore chewing function. TheCCM Abutment System can beplaced in support of single ormultiple-unit restorationsincluding; cement retained, screwretained, and terminal orimmediate abutment support forfixed bridgework. This system isfor one or two stage surgical | The IBS Implant System isintended to replace missing teethto restore chewing function. TheIBS Implant can be placed insupport of single or multiple-unitrestorations including; cementretained, screw retained, oroverdenture restorations, andterminal or immediate abutmentsupport for fixed bridgework.This system is for one or two | The s-Clean One-Q-SL Narrow ImplantSystem (3.0, 3.3mm) may be used as anartificial root structure for single toothreplacement of mandibular central and lateralincisors and maxillary lateral incisors. Theimplants may be restored immediately1) with a temporary prosthesis that is not infunctional occlusion,2) when splinted together as an artificial rootstructure for multiple tooth replacement ofmandibular incisors, or |
| procedures. This system isintended for delayed loading. | stage surgical procedures. Thissystem is intended for delayedloading. | 3) for denture stabilization using multipleimplants in the anterior mandible andmaxilla.The implants may be placed in immediatefunction when good primary stability hasbeen achieved and with appropriate occlusalloading. | |
| Principle ofOperation | a screw retained restoration type ofabutment using a screw to fix aprosthesis | a screw retained restoration typeof abutment using a screw to fix aprosthesis | a screw retained restoration type of abutmentusing a screw to fix a prosthesis |
| Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 8.5, 8.65mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,Ø5mm, Ø5.5mm (D) X14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: abutment | Image: abutment | Image: abutment |
| Non-Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 8.5, 8.65mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,Ø5mm, Ø5.5mm (D) X14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: abutment | Image: abutment | Image: abutment |
| Burn out core cap | |||
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System | |
| 510k | N/A | K153350 | K161244 |
| Part Name | Burn out core cap | UCLA Abutment | S-clean CCM Abutment(Narrow) |
| Material | Co-Cr-Mo AlloyPoly Diacetate | Titanium AlloyPoly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System isintended to replace missing teethto restore chewing function. TheCCM Abutment System can beplaced in support of single ormultiple-unit restorationsincluding; cement retained,screw retained, and terminal orimmediate abutment support forfixed bridgework. This system isfor one or two stage surgicalprocedures. This system isintended for delayed loading. | The IBS Implant System is intendedto replace missing teeth to restorechewing function. The IBS Implantcan be placed in support of single ormultiple-unit restorations including;cement retained, screw retained, oroverdenture restorations, andterminal or immediate abutmentsupport for fixed bridgework. Thissystem is for one or two stagesurgical procedures. This system isintended for delayed loading. | The s-Clean One-Q-SL Narrow ImplantSystem (3.0, 3.3mm) may be used as anartificial root structure for single toothreplacement of mandibular central and lateralincisors and maxillary lateral incisors. Theimplants may be restored immediately1) with a temporary prosthesis that is not infunctional occlusion,2) when splinted together as an artificial rootstructure for multiple tooth replacement ofmandibular incisors, or3) for denture stabilization using multipleimplants in the anterior mandible and maxilla.The implants may be placed in immediatefunction when good primary stability has beenachieved and with appropriate occlusalloading. |
| Principle ofOperation | a screw retained restoration typeof abutment using a screw to fixa prosthesis | a screw retained restoration type ofabutment using a screw to fix aprosthesis | a screw retained restoration type of abutmentusing a screw to fix a prosthesis |
| Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 9.15 mm(L) | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm,Ø5.5mm (D) X 14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: Design 1 | Image: Design 2 | Image: Design 3 |
| Non-Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 9.15 mm(L) | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm,Ø5.5mm (D) X 14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: Design 4 | Image: Design 5 | Image: Design 6 |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| Burn out core T Cylinder | |||
|---|---|---|---|
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System | |
| 510k | N/A | K153350 | K161244 |
| Part Name | Burn out core T cylinder | UCLA Abutment | S-clean CCM Abutment(Narrow) |
| Material | Co-Cr-Mo AlloyPoly Diacetate | Titanium AlloyPoly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System isintended to replace missing teeth torestore chewing function. The CCMAbutment System can be placed insupport of single or multiple-unitrestorations including; cementretained, screw retained, and terminalor immediate abutment support forfixed bridgework. This system is forone or two stage surgical procedures.This system is intended for delayedloading. | The IBS Implant System is intendedto replace missing teeth to restorechewing function. The IBS Implantcan be placed in support of single ormultiple-unit restorations including;cement retained, screw retained, oroverdenture restorations, andterminal or immediate abutmentsupport for fixed bridgework. Thissystem is for one or two stagesurgical procedures. This system isintended for delayed loading. | The s-Clean One-Q-SL Narrow ImplantSystem (3.0, 3.3mm) may be used as anartificial root structure for single toothreplacement of mandibular central andlateral incisors and maxillary lateralincisors. The implants may be restoredimmediately1) with a temporary prosthesis that is not infunctional occlusion,2) when splinted together as an artificialroot structure for multiple toothreplacement of mandibular incisors, or3) for denture stabilization using multipleimplants in the anterior mandible andmaxilla.The implants may be placed in immediatefunction when good primary stability hasbeen achieved and with appropriateocclusal loading. |
| Principle ofOperation | a screw retained restoration type ofabutment using a screw to fix aprosthesis | a screw retained restoration type ofabutment using a screw to fix aprosthesis | a screw retained restoration type ofabutment using a screw to fix a prosthesis |
| Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 10.65, 10.8mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm,Ø5.5mm (D) X 14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: CCM Abutment System Design | Image: IBS Implant System Design | Image: s-Clean One-Q-SL Narrow Implant System Design |
| Non-Hex | |||
| Dimension | Ø 5mm, Ø 6mm (D) X 10.65, 10.8mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm,Ø5.5mm (D) X 14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) |
| Restorationangulations | No Angle | No Angle | No Angle |
| Design | Image: dental restoration | Image: two dental implants | Image: two dental implants, one red and one white |
{12}------------------------------------------------
{13}------------------------------------------------
| Burn out core U Cap | ||||
|---|---|---|---|---|
| Subject device | Primary Predicate device | Reference Predicate device | Reference Predicate device | |
| 510k | N/A | K153350 | K162099 | K161244 |
| Device Name | CCM Abutment System | - | IBS Implant system II | s-Clean OneQ-SL Narrow ImplantSystem |
| Part Name | Burn Out Core U Cap(Hex) | - | Multiunit Ti Cylinder | S-clean CCM Abutment(Narrow) |
| Material | Co-Cr-Mo AlloyPoly Diacetate | - | Titanium Alloy | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications foruse | The CCM AbutmentSystem is intended toreplace missing teeth torestore chewing function.The CCM AbutmentSystem can be placed insupport of single ormultiple-unit restorationsincluding; cementretained, screw retained,and terminal orimmediate abutmentsupport for fixedbridgework. This systemis for one or two stagesurgical procedures. This | The IBS Implant System isintended to replacemissing teeth to restorechewing function. The IBSImplant can be placed insupport of single ormultiple-unit restorationsincluding; cementretained, screw retained, oroverdenture restorations,and terminal or immediateabutment support for fixedbridgework. This system isfor one or two stagesurgical procedures. This | Intended for use in partiallyor fully edentulousmandibles and maxillae, insupport of single ormultiple-unit restorationsincluding; cement-retained,screw- retained, oroverdenture restorations, andterminal or intermediateabutment support for fixedbridgework. This system isdedicated for one and twostage surgical proceduresand not dedicated forimmediate loading. Thissystem is intended for | The s-Clean One-Q-SL NarrowImplant System (3.0, 3.3mm) maybe used as an artificial rootstructure for single toothreplacement of mandibular centraland lateral incisors and maxillarylateral incisors. The implants maybe restored immediately1) with a temporary prosthesis thatis not in functional occlusion,2) when splinted together as anartificial root structure for multipletooth replacement of mandibularincisors, or3) for denture stabilization usingmultiple implants in the anterior |
| system is intended fordelayed loading. | system is intended fordelayed loading. | delayed loading. | mandible and maxilla.The implants may be placed inimmediate function when goodprimary stability has been achievedand with appropriate occlusalloading. | |
| Principle ofOperation | a screw retainedrestoration type ofabutment using a screwto fix a prosthesis | a screw retainedrestoration type ofabutment using a screw tofix a prosthesis | a screw retained restorationtype of abutment using ascrew to fix a prosthesis | |
| Hex | ||||
| Dimension | Ø5mm (D)X 12.95 mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,Ø5mm, Ø5.5mm X 12 mm(L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) | |
| Restorationangulations | No Angle | No Angle | No Angle | |
| Design | Image: abutment | Image: abutments | Image: abutments | |
| Non-Hex | ||||
| Dimension | Ø5mm (D)X 12.95 mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,Ø5mm, Ø5.5mm X 12 mm(L) | Ø 4.0 mm (D) x 14.5and 15 mm (L) | |
| Restorationangulations | No Angle | No Angle | No Angle | |
| Design | Image: abutment | Image: abutments | Image: abutments |
{14}------------------------------------------------
{15}------------------------------------------------
| Multi-unit CCM Cylinder | ||||
|---|---|---|---|---|
| Subject device | Primary Predicate device | Reference Predicate device | Reference Predicate device | |
| 510k | NA | K153350 | K162099 | K161244 |
| Part Name | Multi-unit CCM Cylinder (Hex, Non-Hex) | - | Multiunit Ti Cylinder | MU CCM Cylinder |
| Material | Co-Cr-Mo Alloy Poly Diacetate | - | Titanium Alloy | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. | The IBS Implant System is intended to replace missing teeth to restore chewing function. The IBS Implant can be placed in support of single or multiple-unit restorations including: cement retained, screw retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. | Intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement- retained, screw- retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures and not dedicated for immediate loading. This system is intended for delayed loading. | The s-Clean One-Q-SL Narrow Implant System (3.0, 3.3mm) may be used as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. The implants may be restored immediately 1) with a temporary prosthesis that is not in functional occlusion, 2) when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors, or 3) for denture stabilization using multiple implants in the anterior mandible and maxilla. The implants may be placed in immediate function when good primary stability has been achieved and with appropriate occlusal loading |
| Dimension | Ø 4.8 mm x 12 mm (L) | - | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm, Ø5.5mm, Ø6mm x 12 mm (L) | Ø 4.8 mm (D) x 14.25 mm (L) |
| Restorationangulations | No Angle | - | No Angle | No Angle |
| Design | Image: two dental implants | - | Image: three dental implants | Image: one dental implant |
{16}------------------------------------------------
| Abutment Screw | |||||
|---|---|---|---|---|---|
| Subject device | Primary Predicate device | Reference Predicate device | |||
| 510k | NA | K153350 | K140806 | ||
| Device Name | CCM Abutment System | IBS Implant System | |||
| Part Name | Abutment Screw | Abutment Screw | |||
| Material | Titanium Alloy | Titanium Alloy | |||
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | ||
| Indicationsfor Use | The CCM Abutment System isintended to replace missing teeth torestore chewing function. The CCMAbutment System can be placed insupport of single or multiple-unitrestorations including; cementretained, screw retained, andterminal or immediate abutmentsupport for fixed bridgework. Thissystem is for one or two stagesurgical procedures. This system isintended for delayed loading. | The IBS Implant System is intended toreplace missing teeth to restore chewingfunction. The IBS Implant can be placedin support of single or multiple-unitrestorations including; cement retained,screw retained, or overdenturerestorations, and terminal or immediateabutment support for fixed bridgework.This system is for one or two stagesurgical procedures. This system isintended for delayed loading. | The IBS Implant System is intended toreplace missing teeth to restore chewingfunction. The IBS Implant can be placed insupport of single or multiple-unit restorationsincluding; cement retained, screw retained, oroverdenture restorations, and terminal orimmediate abutment support for fixedbridgework. This system is for one or twostage surgical procedures. This system isintended for delayed loading. | ||
| design | |||||
| Diameter | ø 2 | ø 2.5 | |||
| Length | 7mm | 7.6mm | |||
| Material | identical | Titanium Alloy(ASTM F 136) |
{17}------------------------------------------------
Substantial Equivalence Discussion
Similarities:
The subject device has identical machining, manufacturing process, angulation and indication for use and similar design, dimension and technological characteristics as the predicate devices.
The subject device has been supposed to performance and product validations prior to release. Testing including biocompatibility tests has been performed to ensure the devices comply with the applicable International and US FDA Guidance.
Differences:
The differences between the subject device and the primary predicate are:
1) Material
The subject device is made of CCM alloy but the primary predicate is made of Titanium Alloy. To support this discrepancy, the predicate, K161244 is selected as reference and biocompatibility testing were performed on the subject device.
- Slight differences in abutment designs The subject device and predicate devices are detailed shape and dimension of each abutment.
To support this discrepancy, we chose our own predicates, K140806, K153350, and K162099.
Any differences in technology characteristics are accompanied by information that demonstrated the device is substantially equivalent as the predicates and do not raise different questions of safety and effectiveness than the predicate.
Conclusion
The CCM Abutment System, subject device of this submission, constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, CCM Abutment System and its predicates are substantially equivalent.
§ 872.3630 Endosseous dental implant abutment.
(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)