K Number
K172884
Manufacturer
Date Cleared
2017-10-03

(12 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arisure™ Closed Vial Adapter is intended for use by healthcare professionals in a wide variety of healthcare environments including hospitals, healthcare facilities, and pharmacies for reconstitution or dispensing of medication. The Arisure™ Closed Vial Adapter is indicated for use with rubber-stopper medication vials for reconstitution or dispensing of medications, including chemotherapy agents.

Device Description

The Arisure Closed Vial Adapter is intended for use by healthcare professionals in a wide variety of healthcare environments including hospitals, healthcare facilities, and pharmacies for reconstitution or dispensing of medication. The Arisure Closed Vial Adapter is indicated for use with rubber-stopper medication vials for reconstitution or dispensing of medications, including chemotherapy agents.

The Arisure Closed Vial Adapter is a mechanically closed device, designed to retain vapors produced from medications in the attached vial.

AI/ML Overview

The Arisure™ Closed Vial Adapter's acceptance criteria and the study proving it meets these criteria are described below:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance CharacteristicArisure™ Closed Vial Adapter PerformanceEquivalence to Predicate (Y/N)
Misuse TestTested and meets requirementsY
Detachment Force (Horizontal)Tested and meets requirementsY
Detachment Force (Vertical)Tested and meets requirementsY
Flow RateTested and meets requirementsY
Trapped Air VolumeTested and meets requirementsY
Residual FluidTested and meets requirementsY
Vial attachment forceTested and meets requirementsY
Vial pressureTested and meets requirementsY
Valve Detachment TorqueTested and meets requirementsY
Biocompatibility (ISO 10993)Meets biological requirements (Cytotoxicity, Intracutaneous Reactivity, Sensitization, Acute Systemic Toxicity, Hemolysis, Materials Mediated Pyrogenicity)Y
Chemical CompatibilityTested and meets requirementsY
Sterile package integrity testingTested and meets requirementsY
Distribution testingTested and meets requirementsY
Shelf Life TestingTested and meets requirementsY

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each individual performance test (e.g., how many devices were tested for "Misuse Test" or "Flow Rate"). It generally states that "The following tests were conducted on the Arisure Closed Vial Adapter device to demonstrate equivalency of the performance characteristics to the predicate device(s)."

The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. It is implied that these are lab-based performance tests related to the device itself.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable. The device is a medical accessory (vial adapter), and the tests described are physical, chemical, and biological performance tests, not clinical evaluations requiring human expert interpretation for ground truth.

4. Adjudication Method for the Test Set

This information is not applicable, as the tests are objective performance metrics rather than assessments requiring human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human readers interpret results, and the Arisure™ Closed Vial Adapter is a medical accessory.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable to the Arisure™ Closed Vial Adapter. It is a physical medical device, not a software algorithm that operates without human intervention. The performance tests evaluate the device's physical and functional properties.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests described are the established scientific and engineering standards and specifications relevant to medical device performance, such as ISO 10993 for biocompatibility and internal specifications for parameters like detachment force, flow rate, trapped air volume, etc.

8. The Sample Size for the Training Set

This information is not applicable. The Arisure™ Closed Vial Adapter is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2017

Yukon Medical, LLC % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K172884

Trade/Device Name: Arisure™ Closed Vial Adapter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: December 8, 2016 Received: September 21, 2017

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tara A. Ryan -S

for Tina Kiang, PhD Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name Arisure™ Closed Vial Adapter

Indications for Use (Describe)

The Arisure™ Closed Vial Adapter is intended for use by healthcare professionals in a wide variety of healthcare environments including hospitals, healthes, and pharmacies for reconstitution or dispensing of medication. The Arisure™ Closed Vial Adapter is indicated for use with rubber-stopper medication vials for reconstitution or dispensing of medications, including chemotherapy agents.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.2.

510(k) Summary 5.

5.1. Submitter Information

Company Name:Yukon Medical, LLC
Company Address:4021 Stirrup Creek Drive
Suite 200Durham, NC 27703
Company Phone:(919) 595-8250
Company Fax:(919) 595-8251
Contact Person:Pam McNulty, Director QA and Compliance
Yukon MedicalPhone: 919-595-8253
Fax: 919-595-8251
Email: pmcnulty@yukonmedical.com
Prepared By:Soorya Kumar, Project Engineer
Yukon Medical
Phone: 919-636-6976
Fax: 919-595-8251Email: skumar@yukonmedical.com
Date SummaryPrepared:25 August, 2017
Device Identification
Trade/Proprietary Name:13mm Arisure™ Closed Vial Adapter
20mm Arisure™ Closed Vial Adapter28mm Arisure™ Closed Vial Adapter
Common Name:Vial Adapter
Classification Name:Intravascular Administration Set
Regulation Number:21 CFR 880.5440
Class:II
Classification Panel:General Hospital
Product Code:LHI, I.V. Fluid Transfer Set

{4}------------------------------------------------

5.3. Predicate Device

The Arisure Closed Vial Adapter is substantially equivalent to the following predicate device:

DeviceManufacturer510(k)Date Cleared
SmartSite® VialShieldYukon MedicalK132863October 4, 2013

5.4. Device Description

The Arisure Closed Vial Adapter is intended for use by healthcare professionals in a wide variety of healthcare environments including hospitals, healthcare facilities, and pharmacies for reconstitution or dispensing of medication. The Arisure Closed Vial Adapter is indicated for use with rubber-stopper medication vials for reconstitution or dispensing of medications, including chemotherapy agents.

The Arisure Closed Vial Adapter is a mechanically closed device, designed to retain vapors produced from medications in the attached vial.

5.5. Intended Use

Arisure Closed Vial AdapterPredicate Device
The ArisureTM Closed Vial Adapter is intended foruse by healthcare professionals in a wide variety ofhealthcare environments including hospitals,healthcare facilities, and pharmacies forreconstitution or dispensing of medication. TheArisureTM Closed Vial Adapter is indicated for usewith rubber-stopper medication vials forreconstitution or dispensing of medications,including chemotherapy agents.The SmartSite® VialShield is intended for use byhealthcare professionals in a wide variety ofhealthcare environments including hospitals,healthcare facilities, and pharmacies forreconstitution or dispensing of medication. TheSmartSite® VialShield is indicated for use withrubber-stopper medication vials for reconstitutionor dispensing of medications, includingchemotherapy agents.

The Arisure Closed Vial Adapter's intended use is equivalent to that of the SmartSite VialShield.

5.6. Indications for Use

Arisure Closed Vial AdapterPredicate Device
The Arisure™ Closed Vial Adapter is intended forThe SmartSite® VialShield is intended for use by
use by healthcare professionals in a wide variety ofhealthcare professionals in a wide variety of
healthcare environments including hospitals,healthcare environments including hospitals,
healthcare facilities, and pharmacies forhealthcare facilities, and pharmacies for
reconstitution or dispensing of medication. Thereconstitution or dispensing of medication. The
Arisure™ Closed Vial Adapter is indicated for useSmartSite® VialShield is indicated for use with
with rubber-stopper medication vials forrubber-stopper medication vials for reconstitution
reconstitution or dispensing of medications,or dispensing of medications, including
including chemotherapy agents.chemotherapy agents.

The Arisure Closed Vial Adapter's indications for use is equivalent to that of the SmartSite VialShield.

{5}------------------------------------------------

5.7. Predicate Device Comparison - Technical Characteristics

Equivalency of technical characteristics is demonstrated through a direct comparison of the Arisure Closed Vial Adapter device and the predicate devices listed in the table below.

TechnicalCharacteristicsSubject Device:Arisure Closed VialAdapterPredicate Device:SmartSite®VialShieldEquivalence(Y/N)
Hydrophobic FilterYesYesY
SpikeYesYesY
Locking ShroudYesYesY
Luer AccessArisure Neutral Valve(ANV)SmartSite® NeedleFree ValveY
Materials(Fluid Contacting)• Polycarbonate• Silicone• MABS• Acrylicmultipolymer• Silicone• PolycarbonateY

Spike

The spike is used to penetrate a standard medication vial stopper and provide fluid and filtered air paths.

The subject device and the predicate both have equivalent dual lumen spikes; one lumen for fluid transfer and the other which allows for pressure equalization with vented air.

Locking Shroud

The purpose of the locking shroud is to secure the device to a standard medication vial after the stopper is penetrated. Both the subject and predicate device utilize a locking shroud with retention tabs to ensure device security atop a vial.

Luer Access

All configurations of the subject device and predicate device use a needle free valve for luer access to the device.

Hydrophobic Filter

Both the subject and predicate device use a hydrophobic filter membrane in their respective designs. This filter serves four purposes in both devices: 1) The filter prevents particulates from leaving the devices when air is introduced; 2) The filter prevents contaminants in the surrounding environment from entering the secured drug vial; 3) The filter prevents liquid from leaving the device during misuse conditions, where the devices are inverted when liquid is injected; 4) The filter allows the air pressure in the vial to acclimate with ambient air pressure, preventing the build-up of pressure in the vial.

{6}------------------------------------------------

Materials

The subject Arisure Closed Vial Adapter device is constructed of polymeric components, of those components the following are fluid contacting:

  • . Arisure Closed Vial Adapter
    • Vial Adapter, Membrane Retention Ring, Bell Housing: o Methyl Methacrylate Acrylonitrile Butadiene Styrene
    • o Arisure Neutral Valve: Polycarbonate, Silicone, Fluorosilicone

The subject and predicate devices have been tested and meet the biological requirements outlined in ISO 10993-1.

5.8. Predicate Device Comparison - Performance Characteristics

The performance data supplied with this submission demonstrates that the Arisure Closed Vial Adapter meets all specified requirements and is substantially equivalent to the predicate device.

The following tests were conducted on the Arisure Closed Vial Adapter device to demonstrate equivalency of the performance characteristics to the predicate device(s):

  • . Misuse Test
  • Detachment Force (Horizontal) ●
  • Detachment Force (Vertical) ●
  • Flow Rate ●
  • Trapped Air Volume
  • Residual Fluid ●
  • Vial attachment force ●
  • Vial pressure ●
  • Valve Detachment Torque
  • . Biocompatibility - ISO 10993
    • Cytotoxicity by Elution Test (Cytotoxicity) o
    • Intracutaneous Reactivity (Irritation or Intracutaneous Reactivity) o
    • Maximization Test for Delayed Hypersensivity (Sensitization) o
    • Acute Systemic Toxicity (Systemic Toxicity (Acute)) O
    • Evaluation of Hemocompatibility: Interaction with Blood (Hemolysis) o
    • o Materials Mediated Pyrogenicity
  • Chemical Compatibility ●
  • Sterile package integrity testing ●
  • . Distribution testing
  • Shelf Life Testing

Note:

Both the subject and predicate devices contain needle free valves, which has been previously validated for ISO-594 compliance, microbial ingress, and multiple/extended activation.

{7}------------------------------------------------

5.9. Conclusion

Test results demonstrate that the Arisure™ Closed Vial Adapter is as safe, as effective, and performs as well as the legally marketed device designated as the predicate device.

Based on comparisons of the device's intended use, technology and performance characteristics, the subject device is substantially equivalent to the indicated predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.