(221 days)
AESKUSLIDES® ANCA is an indirect immunofluorescence assay utilizing human neutrophil granulocyte coated slides, fixed with Ethanol or Formalin, as a substrate for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA) in human serum by manual microscopy or with the HELIOS® AUTOMATED IFA SYSTEM.
This in vitro diagnostic assay is used as an aid for the diagnosis of ANCA-associated vasculitides (AAV) in conjunction with other clinical and laboratory findings. All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
AESKUSLIDES ANCA is an indirect immunofluorescence assay utilizing human neutrophil granulocyte coated slides, fixed with Ethanol or Formalin, as a substrate for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA) in human serum by manual microscopy or with the HELIOS® AUTOMATED IFA SYSTEM.
This in vitro diagnostic assay is used as an aid for the diagnosis of ANCA-associated vasculitides (AAV) in coniunction with other clinical and laboratory findings. All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
Slides coated with human neutrophil granulocytes for autoantibody detection are fixated by two different methods: ethanol (EtOH) fixation or formalin fixation. Ethanol fixation allows cell components to move through the cells after the fixation process. Formalin fixation causes cellular components to cross-link (a movement of cellular components is abrogated and the patterns are distinct). By processing serum on both Ethanol and Formalin-fixed slides, the user can confirm if the pattern is C-, P-, or A-ANCA, according to the table below.
This document outlines the acceptance criteria and supporting studies for the AESKUSLIDES® ANCA Ethanol and AESKUSLIDES® ANCA Formalin devices, which are indirect immunofluorescence assays for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA).
1. Table of Acceptance Criteria and Reported Device Performance
The provided document details various performance studies with corresponding acceptance criteria and results. Below is a summary of these:
| Study/Metric | Acceptance Criteria | Reported Device Performance (Overall/Range) |
|---|---|---|
| Serum Stability (Freeze/Thaw) | Pos/Neg/Overall Agreement: > 85%All patterns found correctlyPattern Agreement: > 85%FI allowed to differ max +/- 1 from expectedFI Agreement: > 85% | Pos/Neg/Overall Agreement: 100% for both Ethanol and Formalin. All positive samples found positive, all negative found negative.Pattern Agreement: 100% for both Ethanol and Formalin. All patterns found as expected.FI Agreement: 100% for both Ethanol and Formalin. No deviations > +/-1 FI observed. |
| Long Term Serum Stability | Positive sera found positive, negative sera found negative throughout testing period.Correct patterns found.FI allowed to differ max +/- 1 from expected value at each time point. | All criteria fulfilled. Positive samples found positive, negative samples found negative. All patterns found correctly at all test time points.FI did not differ more than +/- 1 level from expected values. |
| Within-Lab Precision | Positive sera found positive, negative sera found negative.Reported FI allowed to differ max. +/- 1 level within study.HELIOS (% Positives of positive samples excl. borderline): > 80%Reader Confirmation (% Positives of positive samples excl. borderline): > 90%Manual (% Positives of positive samples excl. borderline): > 90%Same criteria for % Negatives of negative samples. | ANCA Ethanol:HELIOS: % Positive 86.7-100%, % Negative 83.3-86.7%.User Confirmation: % Positive 96.7-100%, % Negative 93.3-100%.Manual: % Positive 90-100%, % Negative 100%.ANCA Formalin:HELIOS: % Positive 100%, % Negative 90-96.7%.User Confirmation: % Positive 100%, % Negative 90-100%.Manual: % Positive 100%, % Negative 91.7-100%.All acceptance criteria met. FI within +/- 1 level. Pattern consistent >95% for B/C, >85% for A. |
| Between-Lab Precision | Positive, Negative, Overall, Pattern, FI Agreements:Method A: ≥ 70%Method B, C: ≥ 90% (borderline samples excluded in some calculations) | ANCA Ethanol:Overall-Between-Lab (Method A): 92.5% to 96.5%.Overall-Between-Lab (Method B): 98.5% to 100%.Overall-Between-Lab (Method C): 100%.ANCA Formalin:Overall-Between-Lab (Method A): 89.8% to 94.7%.Overall-Between-Lab (Method B): 95.8% to 98.6%.Overall-Between-Lab (Method C): 98.2% to 99.4%.All acceptance criteria met for all methods and sites, exceptions noted for Method A on Formalin (Site 3 Negative Agreement 61.7% but justified). |
| Between-Operator Agreement | All agreements > 90% for Method B, C (borderline samples excluded for Ethanol)HELIOS not applicable for Between-Operator | ANCA Ethanol:Overall-Between-Operator (Method B): 97.1% to 100%.Overall-Between-Operator (Method C): 100%.ANCA Formalin:Overall-Between-Operator (Method B): 93.5% to 99%.Overall-Between-Operator (Method C): 97.5% to 100%.All acceptance criteria met. |
| Single-Operator Agreement | All agreements > 90% for Method B, C (borderline samples excluded for Ethanol)HELIOS not applicable for Single-Operator | ANCA Ethanol:Overall-Single-Operator (Method B): 96.7% to 100%.Overall-Single-Operator (Method C): 100%.ANCA Formalin:Overall-Single-Operator (Method B): 92.7% to 99%.Overall-Single-Operator (Method C): 97% to 100%.All acceptance criteria met. |
| Instrument Precision | All agreements > 70% for Method A (HELIOS) | ANCA Ethanol:Overall-Instrument to Instrument (Method A): 91.3% to 99.2%.ANCA Formalin:Overall-Instrument to Instrument (Method A): 89% to 90.7%.All acceptance criteria met, except for ANCA Formalin negative agreement at site 3 (61.7% instead of 70%), which was addressed. |
| Lot to Lot Precision | Positive, Negative, Overall, Total Pattern, Single Pattern (C/P/A), FI Agreements: > 85% | ANCA Ethanol (combined readers):Positive agreement: 96.3% to 100%.Negative agreement: 100%.Overall agreement: 96.9% to 100%.Pattern agreement: 100%.FI agreement: 97.8% to 100%.ANCA Formalin (combined readers):Positive agreement: 99.5% to 100%.Negative agreement: 96.2% to 99.0%.Overall agreement: 98.4% to 99.7%.Pattern agreement: 100%.FI agreement: 97.5% to 99.1%.All acceptance criteria met. |
| Carry Over | Pos/Neg/Overall Agreement: All positive sera found positive, all negative found negative.All patterns found correctly.FI allowed to differ max +/- 1 from expected value. | All samples fulfilled criteria. No carry over was observed from well to well. All positive samples identified as positive, all negative as negative. All patterns identified correctly. |
| Time Extension Study | Pos/Neg/Overall Agreement: All positive sera found positive, all negative found negative.FI allowed to differ max +/- 1 from expected value. | All acceptance criteria fulfilled. All positive identified as positive, all negative as negative. All patterns found as expected. FI did not deviate more than +/-1 level. |
| Interfering Substances | Pos/Neg/Overall Agreement: > 90%.Pattern Agreement: > 90%.FI allowed to differ max +/- 1 from expected value; FI Agreement: > 90%. | ANCA Ethanol: Positive Agreement 97-100%; Negative Agreement 100%; Overall Agreement 98-100%; Pattern Agreement 97-100%; FI Agreement 98-100%.ANCA Formalin: Positive Agreement 100%; Negative Agreement 92-100%; Overall Agreement 98-100%; Pattern Agreement 100%; FI Agreement 100%.All acceptance criteria met. No interference detected. |
| Accelerated Stability Report | Positive, Negative, Overall, Total Pattern, FI Agreements: > 85% | ANCA Ethanol (both readers): Positive agreement 89.3-96.4%; Negative agreement 92.9-100%; Overall agreement 90.6-96.9%; Pattern agreement 86.7-95.9%; FI agreement 95.5-98.2%.ANCA Formalin (both readers, borderline excl.): Positive agreement 91.3-97.5%; Negative agreement 100%; Overall agreement 95-98.6%; Pattern agreement 91.3-97.5%; FI agreement 96.4-98.6%.All acceptance criteria met. Claims shelf life 24+3 months for Ethanol, 18 months for Formalin. |
| Real Time Stability Report | Positive, Negative, Overall, Pattern, FI Agreements: > 85% | ANCA Ethanol (both readers): Positive agreement 94.8-100%; Negative agreement 96.7-100%; Overall agreement 95-100%; Pattern agreement 94.8-100%; FI agreement 99.2-100%.ANCA Formalin (both readers): Positive agreement 88-90.7%; Negative agreement 97.8-100%; Overall agreement 91.7-94.2%; Pattern agreement 88-90.7%; FI agreement 99.6-100%.All acceptance criteria met for 3 months (ongoing study). |
| In Use Stability Report | Positive, Negative, Overall, Total Pattern, FI Agreements: > 85% | ANCA Ethanol (both readers): Positive agreement 98.6-100%; Negative agreement 100%; Overall agreement 98.8-100%; Pattern agreement 94.3-99.3%; FI agreement 100%.ANCA Formalin (both readers, borderline excl.): Positive agreement 97.5-98.8%; Negative agreement 100%; Overall agreement 98.3-99.2%; Pattern agreement 97.5-98.8%; FI agreement 98.3-100%.All acceptance criteria met for 6 weeks. Claims In Use Stability of 6 weeks. |
| Transport Stability Report | Positive, Negative, Overall, Total Pattern, FI Agreements: > 85% | Performed by Accelerated Stability Report data, demonstrating resistance to 37°C for at least 2 weeks. All criteria fulfilled. |
| Method Comparison (vs. Predicate) | Diagnostic sensitivity & specificity for ANCA Ethanol higher than predicate.Diagnostic sensitivity & specificity for ANCA Formalin comparable to predicate.Positive, Negative, Overall Agreements acceptable (67.1%, 88.3%, 79.3% for Ethanol; 80.5%, 91.8%, 89.9% for Formalin). | ANCA Ethanol: Sensitivity 48.5% (new) vs. 36.4% (predicate); Specificity 69.3% (new) vs. 55.2% (predicate). PPV 35.8% vs 22.2%, NPV 79.3% vs 71.1%. Agreements: Positive 67.1%, Negative 88.3%, Overall 79.3%.ANCA Formalin: Sensitivity 50.0% (new) vs. 37.9% (predicate); Specificity 90.7% (new) vs. 91.5% (predicate). PPV 65.3% vs 61.0%, NPV 83.7% vs 80.7%. Agreements: Positive 80.5%, Negative 91.8%, Overall 89.9%.All stated criteria met. New device comparable or better. |
| Method Comparison (A, B, C) | Positive, Negative, Overall Agreements between different methods: > 85%.Positive, Negative, Overall, Pattern Agreements (for clinical study): > 80%. (For Method A: > 70%). | ANCA Ethanol (Combined Readers):Method C vs B: Positive 86.2-90.6%, Negative 97.6-99.5%, Overall 91.7-95.5%, Pattern 82.5-89.2%.Method B vs A: Positive 79-89.6%, Negative 98.3-99%, Overall 92.2-96.8%, Pattern 81-85.6%.Method C vs A: Positive 70.7-82.4%, Negative 94.5-99.1%, Overall 89.7-93%, Pattern 77.8-81.8%.ANCA Formalin (Combined Readers):Method C vs B: Positive 86.6-89.1%, Negative 90.8-97.6%, Overall 90.3-93.3%, Pattern 82.1-87.5%.Method B vs A: Positive 79.8-99%, Negative 77.1-95.9%, Overall 83.8-95.9%, Pattern 76.1-90.8%.Method C vs A: Positive 73.1-95.6%, Negative 71.8-95%, Overall 78.2-90.5%, Pattern 69.7-79.6%.All acceptance criteria met, with one pattern agreement (C vs A Formalin) slightly below (69.7%) but addressed. |
| Endpoint Titer Comparison | Percentage of samples that differ max +/- 1 titer level: ≥ 90%Titer Agreement: ≥ 80% | ANCA Ethanol (All Readers Combined):Within-Lab (Method B): 95.1% within +/-1 titer level.Within-Lab (Method C): 95.4% within +/-1 titer level.Between-Lab (Method B): 82.3-93.3% Titer Agreement.Between-Lab (Method C): 87.0-98.3% Titer Agreement.ANCA Formalin (All Readers Combined):Within-Lab (Method B): 94.7% within +/-1 titer level.Within-Lab (Method C): 96.1% within +/-1 titer level.Between-Lab (Method B): 79-86.7% Titer Agreement.Between-Lab (Method C): 80.3-93.7% Titer Agreement.All acceptance criteria met, with one Method B site comparison (79%) for Formalin slightly below but addressed. |
| Expected Values/Reference Range | Low number of positive samples in healthy donors consistent with literature. | ANCA Ethanol: 6/150 (4%) and 3/150 (2%) positive results for Readers 1 and 2, respectively.ANCA Formalin: 6/150 (4%) and 4/150 (2.7%) positive results for Readers 1 and 2, respectively.Low numbers correlate well with literature. |
Note on Borderline Samples: Several studies (e.g., Within-Lab Precision, Between-Lab Precision, Accelerated Stability, In Use Stability) explicitly mention the handling of "borderline" samples (very low positive samples that can be evaluated as negative). For certain calculations, results are presented both including and excluding these samples, with justification for lower agreement when included. This indicates a robust statistical approach for handling results near the decision threshold.
2. Sample Sizes Used for the Test Set and Data Provenance
Total Clinical Samples: 630 clinical samples were used for the Clinical Evaluation and Method Comparison studies.
Provenance:
- 510 clinical samples were sourced from 10 BioBanks in the US (BioChain, BioReclamationIVT, Bioserve, ConversantBio, Cureline, DiscoveryLifeSciences, iSpecimen, Precision for Medicine, ProMedDx, and Vitrologic). These samples were selected based on diagnosis to reflect important conditions for the study.
- 120 serum samples were sourced from a German University Hospital to complement rare but important diagnoses (70 Wegener's Granulomatosis, 25 MPA, 25 Churg-Strauss Syndrome).
- Retrospective/Prospective: The document does not explicitly state whether the studies were retrospective or prospective. However, the nature of acquiring samples from biobanks and the use of de-identified diagnoses strongly suggests a retrospective data collection approach for the main clinical sample set.
- Healthy Donor Samples: An additional panel of 150 sera from healthy donors was used for the Expected Values/Reference Range study: 100 from Germany and 50 from the US.
The document states that the diagnosis criteria of the samples were in agreement with diagnostic standards used in the U.S and Germany (e.g., ACR criteria), and that the US sample set was selected to contain different ethnic groups to reflect the US population. All samples were checked for purity, volume, and contaminations and deemed suitable for the study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The ground truth for the clinical sample set was established based on the "diagnosis criteria of the different samples [that] have been made in agreement with diagnostic standards used in the U.S and Germany." A written statement from different serum suppliers is available on request. This implies that the initial diagnosis (ground truth) was established by medical professionals (e.g., clinicians) at the originating institutions (biobanks, university hospital) based on clinical and laboratory findings, prior to their inclusion in this study. The document does not specify the number or specific qualifications of these initial diagnosing experts.
For the subsequent "reading" or "evaluation" of slides within the various performance studies (e.g., Within-Lab, Between-Lab, Method Comparison, Stability studies), two independent readers/experts were consistently used. The qualifications of these readers are generally referred to as "trained personnel" or "trained operator." For instance, the intended use statement explicitly says "All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel." and "The device is for use by a trained operator in a clinical laboratory setting." Specific details on years of experience or board certification (e.g., "radiologist with 10 years of experience") are not provided for these internal study readers.
4. Adjudication Method for the Test Set
For the "reading" or "evaluation" of slides in the various performance studies:
- The studies consistently involved two independent readers.
- The results of these two readers were often calculated and presented separately as well as combined.
- There is no explicit mention of an "adjudication" process (e.g., a 2+1 or 3+1 method) where a third, senior expert would resolve discrepancies between the two initial readers to establish a final ground truth for the study. Instead, the analysis focuses on the agreement between the readers and their agreement with either the expected reference values (for analytical studies) or comparison methods (for clinical studies). The concept of "User Confirmation" (Method B) implies that human oversight is always required for the automated results, but not necessarily a formal adjudication of discordant human reads.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
Yes, a multi-reader, multi-case (MRMC) comparative effectiveness study was done. This is evident in the "Method Comparison of Method A, B, and C and clinical study" section (pages 85-96). This study compared:
- Method C (Manual): Manual processing and manual reading by two independent readers.
- Method B (Reader Confirmation): Automated processing/imaging by HELIOS, with manual reading of digital images by two independent readers.
- Method A (HELIOS): Automated processing/imaging by HELIOS, with automated positive/negative classification by the HELIOS Vasculitis Pattern Plus software.
The study was conducted at three different study sites (two US, one German) using the entire 630-sample set, with two independent readers at each site for manual and reader confirmation methods.
Effect Size (AI vs. Human-in-the-Loop):
The document does not present the effect size in terms of how much human readers improve with AI vs. without AI assistance. Instead, it compares the performance (agreements and diagnostic sensitivities/specificities) between:
- Human reading of traditionally processed slides (Method C).
- Human reading of AI-processed images (Method B).
- AI-only classification (Method A).
The statement that "All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel" (intended use) and the acceptance criteria for Method A (lower agreement accepted for automated-only results, e.g., >70% compared to >90% for human reads) consistently emphasize that the AI is an aid that requires human confirmation. The data implicitly supports that humans (Methods B and C) perform better than the standalone AI (Method A) in certain aspects (higher agreement percentages, higher diagnostic performance metrics for Human vs HELIOS in Formalin especially). For example, in ANCA Formalin, Method C vs A pattern agreement was only 69.7% while C vs B was 82.1-87.5%, highlighting the current limitations of standalone AI pattern recognition and the value of human reading (even with automated imaging).
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Yes, standalone performance of the algorithm (HELIOS Vasculitis Pattern Plus software, referred to as Method A) was evaluated.
- Method A processed slides automatically, acquired images, and performed automated reading/interpretation.
- Its performance was compared against manual reading (Method C) and human reading of automated images (Method B) in the "Method Comparison of Method A, B, and C" study.
- As noted in point 5, the acceptance criteria for Method A were lower (e.g., ≥ 70% agreement) compared to the human-in-the-loop methods (≥ 90%). The results show that Method A generally achieved these lower thresholds but performed less ideally for pattern recognition.
7. The Type of Ground Truth Used
The ground truth used several types throughout the studies:
- Clinical Diagnosis (Outcomes Data / Expert Consensus): For the diagnostic sensitivity and specificity calculations (Method comparison against predicate, and Method A, B, C comparison), the ground truth for patient samples was their established "diagnosis" (e.g., ANCA-associated vasculitis (AAV), other diseases like SLE, RA, etc.). This diagnosis was made in agreement with US and German diagnostic standards (e.g., ACR criteria), implying a form of expert consensus based on clinical and laboratory findings.
- Expected Results/Reference Values (Expert Consensus): For analytical performance studies (e.g., Precision, Stability, Carry Over, Time Extension, Interfering Substances), ground truth was often defined as "expected results" or "expected values" for specific samples (e.g., positive/negative status, specific pattern, fluorescence intensity). These expected values were likely established by experienced operators / experts during the initial characterization of the control and study samples. The repeated use of "correctly found" implies agreement with a pre-established reference.
- Negative Healthy Donor Panel: For the "Expected Values/Reference Range" study, healthy donor samples confirmed to be negative for ANCA were used to establish a reference range, implicitly serving as a negative ground truth.
There is no mention of pathology or direct biopsy results as ground truth, which is typical for diagnoses like vasculitis given the nature of ANCA testing.
8. The Sample Size for the Training Set
The document does not provide a specific sample size for the "training set" of the HELIOS AUTOMATED IFA SYSTEM. The provided information focuses on the validation of the device, particularly the performance evaluation of the final device using various test sets. The software's pattern recognition uses "SVM (Support Vector Machine) technology," which implies a machine learning approach. However, details on the dataset used to train this SVM model are not disclosed in this document.
9. How the Ground Truth for the Training Set Was Established
Similarly, since the training set details are not provided, the method for establishing its ground truth is also not explicitly described. For machine learning models like SVMs used in pattern recognition, the training data would typically be images with associated labels (ground truth) that are a result of expert annotation or consensus. Given the context of manual reading by "trained personnel" and the need for "confirmation by trained personnel," it is highly probable that the ground truth for any training set would have been established by multiple expert pathologists or laboratory professionals specializing in indirect immunofluorescence interpretation, likely through a consensus or adjudication process. However, this is an inference based on industry practice and the provided context, not a direct statement in the document.
{0}------------------------------------------------
Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a bold, sans-serif font. The word "FDA" is in a blue box.
March 23, 2018
Aesku.Diagnostics GmbH & Co. KG Sandra Reuter Regulatory Affairs Manager Mikro-Forum-Ring 2 Wendelsheim, Rheinland-Pfalz55234 Germany
Re: K172461
Trade/Device Name: AESKUSLIDES® ANCA Ethanol AESKUSLIDES® ANCA Formalin
Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MOB Dated: July 31, 2017 Received: August 14, 2017
Dear Sandra Reuter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
510(k) Summary
- A. 510(k) Number: K172461
- B. Purpose for Submission: New Device
C. Measurand:
Anti-neutrophil cytoplasmic autoantibodies (ANCA)
D. Type of Test:
Qualitative and semi-quantitative, indirect immunofluorescence
E. Applicant:
AESKU.Diagnostics GmbH & Co.KG Mikro-Forum-Ring 2 55234 Wendelsheim Germany
Contact Person: Dr. Sandra Reuter Regulatory Affairs registrations@aesku.com p: +49 6734 9622 0
prepared on March 23rd, 2018
F. Proprietary and Established Names:
AESKUSLIDES® ANCA Ethanol AESKUSLIDES® ANCA Formalin
G. Regulatory Information:
-
Regulation section: §CFR 866.5660 - Multiple autoantibodies immunological test system
-
Classification: Class II
-
Product code: MOB - Anti-neutrophil cytoplasmic antibodies (ANCA)
-
Clinical use Immunology (82)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
H. Intended Use:
-
- Intended use(s):
AESKUSLIDES® ANCA is an indirect immunofluorescence assay utilizing human neutrophil granulocyte coated slides, fixed with Ethanol or Formalin, as a substrate for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA) in human serum by manual microscopy or with the HELIOS® AUTOMATED IFA SYSTEM.
- Intended use(s):
This in vitro diagnostic assay is used as an aid for the diagnosis of ANCA-associated vasculitides (AAV) in conjunction with other clinical and laboratory findings. All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
-
- Indication(s) for use: Same as intended use
-
- Special conditions for use statement(s):
-
- For prescription use only
-
This device is only for use with reagents that are indicated for use with the device.
-
The device is for use by a trained operator in a clinical laboratory setting.
-
All software-aided results must be confirmed by the trained operator.
5.For use only by manual microscopy or with HELIOS® AUTOMATED IFA SYSTEM.
l. Device Description:
AESKUSLIDES ANCA is an indirect immunofluorescence assay utilizing human neutrophil granulocyte coated slides, fixed with Ethanol or Formalin, as a substrate for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA) in human serum by manual microscopy or with the HELIOS® AUTOMATED IFA SYSTEM.
This in vitro diagnostic assay is used as an aid for the diagnosis of ANCA-associated vasculitides (AAV) in coniunction with other clinical and laboratory findings. All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
Slides coated with human neutrophil granulocytes for autoantibody detection are fixated by two different methods: ethanol (EtOH) fixation or formalin fixation. Ethanol fixation allows cell components to move through the cells after the fixation process. Formalin fixation causes cellular components to cross-link (a movement of cellular components is abrogated and the patterns are distinct). By processing serum on both Ethanol and Formalin-fixed slides, the user can confirm if the pattern is C-, P-, or A-ANCA, according to the table below.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray.
| Ethanol-fixed Result | Formalin-Fixed Result | Pattern |
|---|---|---|
| Cytoplasmic (C-ANCA) | Cytoplasmic (C-ANCA) | C-ANCA |
| Perinuclear (P-ANCA) | Cytoplasmic (C-ANCA) | P-ANCA |
| Perinuclear (P-ANCA) | Negative / unclear | Confirm with ANA |
| Mixed (P-ANCA + C-ANCA) | Negative / weak positive | A-ANCA |
| Very Perinuclear | A-ANCA | A-ANCA |
ANCA Formalin test is not intended to be used by itself, but in conjunction with ANCA Ethanol test.
The HELIOS Vasculitis Pattern Plus software is able to detect both C- and P-ANCA. A-ANCA are reported as undefined positive. All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
Each kit of ANCA Ethanol and ANCA Formalin contains (Quantity depends on product variant):
- Slides, each containing 6 or 12 wells coated with human neutrophils (ethanol fixation)/ human neutrophils (formalin fixation) cells
- -2.0 ml/ 4.0 ml vial containing Fluorescein (FITC) labelled Anti-human Antibody IgG conjugate in a solution of BSA, ready for use
- -0.5 ml vial of positive control containing human serum (diluted), ready for use
- -0.5 ml vial of negative control containing diluted human serum, ready for use
- 8.0 ml vial of mounting medium containing a solution of glycerol and PBS, ready for use
- 70 ml bottle of sample buffer, containing BSA, PBS and ready for use -
- -100 ml bottle of wash buffer, concentrated buffer 1:10 in distilled water, containing PBS
| Standard Ref. | Description | Tests |
|---|---|---|
| 54.100.US | ANCA Ethanol (12 wells) | 120 |
| 54.101.US | ANCA Formalin (12 wells) | 120 |
| 54.050.US | ANCA Ethanol (6 wells) | 60 |
| 54.051.US | ANCA Formalin (6 wells) | 60 |
| 54.100.US.Demo | ANCA Ethanol (12 wells) | 24 |
| 54.100.US.Bulk5 | ANCA Ethanol (12 wells) | 600 |
| 54.100.US.Bulk10 | ANCA Ethanol (12 wells) | 1200 |
| 54.101.US.Demo | ANCA Formalin (12 wells) | 24 |
| 54.101.US.Bulk5 | ANCA Formalin (12 wells) | 600 |
| 54.101.US.Bulk10 | ANCA Formalin (12 wells) | 1200 |
{6}------------------------------------------------
Not provided in the kit:
| Ref. | Reagent | Quantity/ Volume | Description | Readyto use | |
|---|---|---|---|---|---|
| EBIFA | EvansBlue0.2% | 1x | 3ml | Capped white: Blue colouredsolutionContaining: PBS, EvansBlue.Dilute the Evans Blue 0.2%1:3000 in 1x WBIFA | NO |
HELIOS AUTOMATED IFA SYSTEM (K153117) or equivalent manual microscope.
J. Substantial Equivalence Information:
-
- Predicate device name(s):
NOVA Lite ANCA
- Predicate device name(s):
-
- Predicate 510(k) number(s): K961340
-
- Comparison with predicate device:
Table 7: comparison table with predicate device
| Item | PredicateNOVA Lite ANCA | AESKUSLIDES® ANCA | |
|---|---|---|---|
| Similarities | Intended Use | NOVA Lite® ANCA is an indirectimmunofluorescent assay for thescreening and semi-quantitativedetermination of anti-neutrophilcytoplasmic antibodies (ANCA) inhuman serum. The presence ofantineutrophil cytoplasmicantibodies can be used inconjunction with other serologicaltests and clinical findings aids inthe assessment of varioussystemic vasculitides. | AESKUSLIDES ANCA is an indirectimmunofluorescence assay utilizing humanneutrophil granulocyte coated slides, fixed withEthanol or Formalin, as a substrate for thequalitative and semi-quantitative determination ofanti-neutrophil cytoplasmic autoantibodies (ANCA)in human serum by manual microscopy or with theHELIOS® AUTOMATED IFA SYSTEM.This in vitro diagnostic assay is used as an aid forthe diagnosis of ANCA-associated vasculitides(AAV) in conjunction with other clinical andlaboratory findings.All suggested results obtained with the HELIOSAUTOMATED IFA SYSTEM must beconfirmed by trained personnel. |
| Methology | Immunofluorescence assay (IFA) | same as Predicate | |
| Procedure | Standard IFA techniquequalitative and semi-quantitativetiter | same as Predicate | |
| Results | Serum | same as Predicate | |
| Samples Matrix | anti-neutrophil cytoplasmicautoantibodies (ANCA) | same as Predicate | |
| Analyte | ethanol-fixed human neutrophil /formalin-fixed human neutrophil | same as Predicate | |
| Antigen | FITC | same as Predicate | |
| FluorescenceMarker | cANCA Positive, pANCA Positiveand one negative control | same as Predicate | |
| Controls | Anti-Human IgG Conjugate | same as Predicate | |
| conjugate | 1:20 | same as Predicate | |
| Screeningdilution | 2 - 8 °C | same as Predicate | |
| Storage |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside. To the right of the square, the text "AESKU.DIAGNOSTICS" is written in gray. Below the text, the phrase "THE DIAGNOSTIC TOOL THAT WORKS" is written in a smaller, lighter gray font.
| Slides | 6 or 12 wells coated with antigen | same as Predicate | |
|---|---|---|---|
| Differences | Shelf life | 18 months | 24 months for ANCA Ethanol18 months for ANCA Formalin |
| ManualInterpretation ofresult | Manual fluorescence microscopy | Manual fluorescence microscopy or with HELIOSw/ trained operator verification | |
| Automatedinterpretation ofresults | N/A | HELIOS w/ trained operator verification |
K. Standard/ Guidance Document Referenced (if applicable):
Table 8: List of Standards / Guidance Documents
| # | Standards Title | Version |
|---|---|---|
| 1 | ISO 14971 - Medical Devices - Application of risk management to medicaldevices | SecondEdition |
| 2 | IEC 62366 - Medical Devices - Part 1: Application Of Usability Engineering ToMedical Devices [Including CORRIGENDUM 1 (2016)] | Edition 1.0 |
| 3 | 15223-1 - Medical Devices - Symbols To Be Used With Medical Device Labels,Labelling, and Information to be supplied - Part 1: General Requirements | SecondEdition |
| 4 | Interference Testing in Clinical Chemistry | EP07-A2 |
| 5 | Evaluation of Detection Capability for Clinical Laboratory MeasurementProcedures | EP17-A2 |
| 6 | Evaluation of Stability of In Vitro Diagnostic Reagents | EP25-A |
| 7 | Defining, Establishing and Verifying Reference Intervals in the ClinicalLaboratory | EP28-A3c |
| 8 | Evaluation of the Linearity of Quantitative Measurement Procedures AStatistical Approach | EP06-A |
| 9 | Procedures for the Handling and Processing of Blood Specimens for CommonLaboratory Tests; Approved Guideline-Fourth Edition | GP44-A4 |
| 10 | Guidance for Industry and FDA Staff: Recommendations for Anti-NuclearAntibody (ANA) Test System Premarket (510k) Submissions | - |
| 11 | Factors to Consider Regarding Benefit-Risk in Medical Device ProductAvailability, Compliance, and Enforcement Decisions | - |
| 12 | Applying Human Factors and Usability Engineering to Medical Devices | - |
L. Test Principle
AESKUSLIDES® ANCA is an indirect fluorescent antibody assay utilizing human neutrophil granulocyte coated slides, fixed with Ethanol or Formalin, as a substrate for the qualitative and semi-quantitative determination of anti-neutrophil cytoplasmic autoantibodies (ANCA) in human serum by manual microscopy or with HELIOS® AUTOMATED IFA SYSTEM. This in vitro diagnostic assay is used as an aid for the diagnosis of ANCA-associated vascultitides (AAV) in conjunction with other clinical and laboratory findings. All suggested results obtained with the HELIOS® AUTOMATED IFA SYSTEM must be confirmed by trained personnel.
Manual interpretation of AESKUSLIDES® ANCA Ethanol:
The two main patterns seen on an ethanol-fixed substrate are perinuclear (P-ANCA) and cytoplasmic (C-ANCA):
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image of a medical symbol on the left. To the right of the square is the text "AESKU.DIAGNOSTICS" in a gray sans-serif font. Below the text is the tagline "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
- -C-ANCA presents as coarse speckled cytoplasmic fluorescence, often with accentuated staining between the nuclear lobes. This pattern is characteristic for antibodies reacting with PR3.
- -P-ANCA presents as perinuclear staining with or without nuclear extension. This pattern is usually characteristic for antibodies reacting with MPO. Note that anti-nuclear antibody (ANA) positive samples (containing anti-DNA/histones) may react with the nuclei of ethanol-fixed neutrophils, causing nuclear staining, and may mask or mimic the P-ANCA pattern(s).
A third pattern, less commonly seen, is called atypical ANCA (A-ANCA or X-ANCA):
- । A-ANCA presents as a cytoplasmic and perinuclear or very perinuclear staining on ethanol-fixed neutrophil substrate and usually becomes negative on formalin fixed substrate.
Manual interpretation of AESKUSLIDES® ANCA Formalin:
On formalin fixed substrate, both MPO and PR3 antibodies appear as coarse cytoplasmic granular staining with interlobular accentuation.
Results from both AESKUSLIDES® ANCA Ethanol and AESKUSLIDES® ANCA Formalin provide further information on the antibodies present in the serum.
The fluorescence intensity level is the intensity of the specific fluorescence expressed as a numeric value. These values, if present, are reported as a number between "0" (no specific fluorescence) and "4+" (very strong visible reaction).
| Intensity | Interpretation |
|---|---|
| 4+ | high positivemaximal fluorescence, very strong visible reaction; brilliant yellow-green |
| 3+ | positivestrong visible reaction; less brilliant as 4+; yellow-green fluorescence |
| 2+ | positivemoderate visible reaction; definite but dull yellow-green fluorescence |
| 1+ | positiveweak visible reaction, very dim subdued fluorescence |
| 0 | negativeno specific fluorescence |
AESKU recommends a screening dilution of 1:20, followed by serial dilutions for semiquantitative determinations and suggests each laboratory establish its own screening dilution and titration scheme based on its population and instrumentation.
Qualitative evaluation
A serum dilution is considered negative for ANCA antibodies if the cells exhibit < 1+ fluorescence of the cytoplasm or nucleus. Likewise, a serum dilution is considered positive for ANCA antibodies if the cells exhibit ≥ 1+ fluorescence of the cytoplasm or nucleus. A sample is considered positive for ANCA antibodies if it exhibits ≥ 1+ fluorescence of the cytoplasm or nucleus at a sample dilution of 1:20 or greater. Operators should report all titers and specific fluorescence staining seen.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image appears to be a stylized representation of a chromosome. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger and bolder than "DIAGNOSTICS". Below the company name is the tagline "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Semi-quantitative evaluation
The endpoint titer is defined as the highest sample dilution factor for which specific fluorescence of the cytoplasm or nucleus is identifiable.
The titers are classified as:
- 1:20 and 1:40 are considered low titers
- 1:80 and 1:160 are considered medium titers
- 1:320 and greater are considered high titers
Automated instrument interpretation of test results
Software
After slides are processed by the HELIOS®, digital images of representative fields of view in the well are captured and stored on the computer system. The HELIOS Pattern Recognition software recognizes the pattern of the captured image by using SVM (Support Vector Machine) technology. After image pre-processing, feature extraction and classification, the software delivers the results. All suggested results given by the HELIOS® software must be confirmed within the Result Confirmation tool by a trained operator.
The HELIOS PATTERN RECOGNITON software tool identifies the following immunofluorescence patterns, abbreviated below:
| Abbreviation | Pattern |
|---|---|
| P-ANCA | perinuclear |
| C-ANCA | cytoplasmic |
| unk | unknown (undefined positive) |
| neg | negative |
The HELIOS DEVICE SOFTWARE examines the fluorescence intensity and uses an analysis algorithm which takes exposure and pixel frequency into account. It examines relevant regions of the captured images, such as the fluorescence cell area and subtracts the background, to provide an assessment of positive or negative results.
| Intensity | Interpretation |
|---|---|
| + | positive |
| - | negative |
Trained operators must confirm all suggestions. Intensity and End Point Titers are identified by the HELIOS DEVICE SOFTWARE and pattern suggestions are made by the HELIOS PATTERN RECOGNITION tool.
AESKU recommends a screening dilution of 1:20, followed by serial dilutions for semiquantitative determinations and suggests each laboratory establish its own screening dilution and titration scheme based on its population
{10}------------------------------------------------
M. Performance Characteristics:
1. Analytical performance
During the study 4 different methods have been applied:
| Method | Processing | Imaging | Reading/Evaluation odSlide | Alternate Name ofMethod |
|---|---|---|---|---|
| A | Automated | Automated | Automated (SoftwareInterpretation) | HELIOS |
| B | Automated | Automated | Manual (read of digitalimage) | HELIOS UserEvaluation |
| C | Manual | Manual | Manual (read ofmicroscope field) | ManualAESKUSLIDESANCA |
| D | Manual | Manual | Manual (read ofmicroscope field) | Manual - PredicateNOVA Lite ANCA |
Method D is the predicate for method C
In the clinical study the method comparison was: Method D vs Method C Method C vs. Method A, B Method B vs Method A
Nomenclature and acronyms used in studies
Table 9: Nomenclature and acronyms
| Nomenclature | Description/ Also known as |
|---|---|
| AESKUSLIDES® ANCAHELIOS® AUTOMATEDIFA SYSTEM | ANCAHELIOS |
| Manual or Manual Reading | Results obtained by the operator afterreading and interpreting the slides with atraditional fluorescence microscope |
| Method A | HELIOS + AESKUSLIDES ANCAw/slides read by HELIOS |
| Method B | HELIOS + AESKUSLIDES ANCAw/slides read by User |
| Method C | Manual AESKUSLIDES ANCA w/slidesread by User |
| Method D | Manual Nova Lite ANCA w/slides readby User |
| P-ANCA | perinuclear |
| C-ANCA | cytoplasmic |
| unk | unknown (undefined positive) |
| ANCA | anti-neutrophil cytoplasmicautoantibodies |
| Negative IFA score/result,negative | neg, negative |
| Positive IFA score/result | pos, positive |
| Diseases Abbreviation | Diseases Name |
| SLE | Systemic lupus erythematosus |
| MPA | Microscopic polyangiitis |
| UC | Ulcerative colitis |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white stylized image of a medical device. To the right of the square, the text "AESKU.DIAGNOSTICS" is displayed in gray, with "AESKU" in a larger, bolder font than "DIAGNOSTICS". Below the company name, the tagline "THE DIAGNOSTIC TOOL THAT WORKS" is written in a smaller, lighter gray font.
| PMR | Polymyalgia rheumatica |
|---|---|
| SpA | Spondyloarthropathy |
| RA | Rheumatoid Arthritis |
| AAV | ANCA-associated vasculitides |
| HCV | Hepatitis C Virus |
| HBV | Hepatitis B Virus |
Serum sample:
630 Clinical samples have been used for the Clinical Evaluation and Method Comparison studies described herein.
An exhaustive search for US Clinical Samples from 10 BioBanks (BioChain, BioReclamationIVT, Bioserve, ConversantBio, Cureline, DiscoveryLifeSciences, iSpecimen, Precision for Medicine, ProMedDx, and Vitrologic) has been performed and 510 Clinical Samples have been found. The samples have been selected according to their diagnosis, to reflect all important diagnosis for this study (see table). The sample set has been completed with 120 serum samples from a German University Hospital to complement some rare, but important diagnosis (70 sera from patients with Wegener's Granulomatosis, 25 sera from patients with MPA and 25 sera from patients with Churg-Strauss Syndrome).
| Study Sample Set | |||
|---|---|---|---|
| Sample Definition | Diagnosis | n | |
| TargetDiagnosis | ANCAassociatedVasculitis(AAV) | Wegener's Granulomatosis | 80* |
| Microscopic polyangiitis (MPA) | 29* | ||
| Churg-Strauss Syndrome | 26* | ||
| CharacterizedMPO/PR3/ANCApositive sera | PR3 positive | 25 | |
| MPO positive | 25 | ||
| P-ANCA positive | 50 | ||
| C-ANCA positive | 20 | ||
| Controldiagnosis | AutoimmuneLiverDiseases | Autoimmune Hepatitis (AIH) | 8 |
| Autoimmune Hepatitis/ Primary biliarycholangitis | 6 | ||
| Primary biliary cholangitis (PBC) | 11 | ||
| Primary sclerosing cholangitis (PSC) | 10 | ||
| Controldiagnosis | Inflammatoryboweldiseases | Ulcerative Colitis | 71 |
| Crohn's disease | 40 | ||
| Inflammatory bowel disease (IBD) | 9 | ||
| Otherrheumaticdiseases | Rheumatoid Arthritis (RA) | 12 | |
| Systemic Lupus Erythematosus (SLE) | 30 | ||
| Scleroderma | 19 | ||
| Myositis | 1 | ||
| Infections | Hepatitis C Virus (HCV ) | 16 | |
| Hepatitis B Virus (HBV ) | 8 | ||
| Polymyalgia rheumatica (PMR) | 18 |
Table 10: Serum samples set for clinical and method comparison studies
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside, resembling a stylized medical symbol. To the right of the square, the text "AESKU.DIAGNOSTICS" is displayed in a gray, sans-serif font. Below the company name, the tagline "THE DIAGNOSTIC TOOL THAT WORKS" is written in a smaller, lighter gray font.
| VasculitidesNOTassociatedwith ANCA | Giant cell artertitis (GCA) | 4 | |
|---|---|---|---|
| Purpura | 1 | ||
| Leukemia | Lymphoma, Myeloma | 15 | |
| Otherdiagnosis | Celiac disease | 10 | |
| Chronic kidney disease | 32 | ||
| Sinusitis | 22 | ||
| Asthma | 32 | ||
| Sum | 630 |
*samples from Germany: 70 sera from patients with Wegener's Granulomatosis, 25 sera from patients with MPA and 25 sera from patients with Churg-Strauss Syndrome
The diagnosis criteria of the different samples have been made in agreement with diagnostic standards used in the U.S and Germany. A written statement from different serum suppliers is available on request. Standard criteria are for example ACR criteria.
The US sample set was also selected to contain different ethnic groups (e.g. White, Black/Black African, Asian, Hispanic) to reflect the ethnic composition of the US population as good as possible.
The samples have been checked for purity and volume by visual inspection and for further contaminations by an in house pre-screening before inclusion into the study. All 630 sera passed and were determined suitable for the study.
All serum samples which have been used for this studies have been shipped and stored at -20°C. Upon arrival from the Biobanks they were thawed once to aliquot them for the different studies. Subsequently, samples were frozen again. Therefore, samples underwent 2-3 freezing-thawing cycles before they were used in the studies.
To confirm that repeated thawing and freezing and also long term storage at -20°C has no influence on the samples we have performed the following two studies: All serum samples which have been used for this studies have been shipped and stored at -20°C. Upon arrival from the Biobanks they were thawed once to aliquot them for the different studies. Subsequently, samples were frozen again. Therefore, samples underwent 2-3 freezing-thawing cycles before they were used in the studies.
To confirm that repeated thawing and freezing and also long term storage at -20°C has no influence on the samples we have performed the following two studies:
Serum Stability AESKUSLIDES® ANCA
Procedure:
Serum Stability was shown by testing 10 serum samples (7 positive, 3 negative). For each serum, one aliquot has been prepared that serves as control (no freezethawing). A second aliquot for each serum has been prepared that underwent 4 freeze-thaw cycles. Then, all samples and controls have been analyzed on AESKUSLIDES® ANCA Ethanol and ANCA Formalin according to the IFUs. Slides were read by two independent readers. Each freeze/thaw sample was compared to the respective control sample.
{13}------------------------------------------------
| Table 11: Serum sample set for serum stability | AESKUSLIDES ANCA Formalin | |||||
|---|---|---|---|---|---|---|
| SampleID | Pos/Neg | PatternANCAEthanol | Grading | Pos/Neg | PatternANCAFormalin | Grading |
| 1 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 2 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 3 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 4 | Pos | C-ANCA | low positive | Pos | C-ANCA | medium positive |
| 5 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 6 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 7 | Pos | A-ANCA | medium positive | Neg | Neg | negative |
| 8 | Neg | Neg | negative | Neg | Neg | negative |
| 9 | Neg | Neg | negative | Neg | Neg | negative |
| 10 | Neg | Neg | negative | Neg | Neg | negative |
Acceptance criteria:
- Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative. Overall Agreement: > 85%.
- -All patterns have to be found correctly.
- -Pattern Agreement: > 85%
- -FI is allowed to differ maximum + or - 1 from the expected value. FI Agreement: > 85%.
Results
All samples fulfilled the above criteria. No differences were observed between sera thawed a single time and sera that had undergone 4 cycles of freeze/thawing:
- -Positive/Negative/Overall Agreement was 100% for both AESKUSLIDES® ANCA Ethanol and Formalin. All positive samples have been found positive, all negative samples have been found negative.
- -Pattern Agreement was 100% for both AESKUSLIDES® ANCA Ethanol and Formalin. All patterns have been found as expected.
- FI agreement was 100% for both AESKUSLIDES® ANCA Ethanol and Formalin. -No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.
Conclusions
All criteria are fulfilled in each test. The results show that multiple freeze-thaw cycles have no effect on test results.
{14}------------------------------------------------
Long Term Serum Stability AESKUSLIDES® ANCA
To show that long term storage of serum samples at -20°C has no effect on performance and results of AESKUSLIDES® ANCA Ethanol and ANCA Formalin.
Procedure:
Nine different serum samples have been aliquoted and stored at -20°C for a time period of at least 12 months. Frozen aliquots of the different sera have been thawed and assayed on AESKUSLIDES® ANCA at the indicated time points. All tests have been performed manually according to the IFU and subsequently analyzed at the microscope.
| SampleID | AESKUSLIDES ANCA EthanolExpected Result | AESKUSLIDES ANCA FormalinExpected Result | ||||
|---|---|---|---|---|---|---|
| Pos/Neg | Pattern | FI | Pos/Neg | Pattern | FI | |
| 1 | Pos | P | 3 | Pos | C | 3 |
| 2 | Pos | C | 3 | Pos | C | 3 |
| 3 | Pos | A | 1 | Neg | Neg | 0 |
| 4 | Pos | C | 1 | Pos | C | 1 |
| 5 | Neg | 0 | 0 | Neg | 0 | 0 |
| 6 | Neg | 0 | 0 | Neg | 0 | 0 |
| 7 | Neg | 0 | 0 | Neg | 0 | 0 |
| 8 | Pos | P | 3 | Pos | P | 2 |
| 9 | Pos | P | 3 | Pos | C | 3 |
Table 12: Serum sample set for long term serum stability
Acceptance Criteria:
Positive sera have to be found positive and negative sera have to be found negative throughout the whole testing period. Correct patterns have to be found. The fluorescence intensity (FI) is allowed to differ maximum +/- 1 level from the expected value at each test time point.
Results:
Positive samples have been found positive, negative samples have been found negative, and all patterns have been found correctly at all test time points. FI did not differ more than +/- 1 level from the expected values.
| AESKUSLIDES ANCA Ethanol | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SampleID | month 0 | month 3 | month 6 | month 8 | month 10 | month 14 | ||||||
| Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | |
| 1 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
| 2 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
| 3 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 |
| 4 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 |
| 5 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 6 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 7 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 8 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
| 9 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
Table 13: Long Term Serum Stability AESKUSLIDES ANCA Ethanol
{15}------------------------------------------------
| AESKUSLIDES ANCA Formalin | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SampleID | month 0 | month 2 | month 5 | month 8 | month 10 | month 12 | ||||||
| Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | Pos/Neg | FI | |
| 1 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 2 | Pos | 2 | Pos | 2 |
| 2 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
| 3 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 4 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 | Pos | 1 |
| 5 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 6 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 7 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 | Neg | 0 |
| 8 | Pos | 1 | na | na | Pos | 2 | na | na | Pos | 1 | Pos | 2 |
| 9 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 | Pos | 3 |
Table 14: Long Term Serum Stability AESKUSLIDES® ANCA Formalin
Conclusion:
All acceptance criteria have been fulfilled for the tested sera. The results show that long term storage of sera at -20°C over a time period of at least 12 months has no effect on performance and test results of AESKUSLIDES® ANCA.
Precision
a) Within-Lab Precision/Repeatability AESKUSLIDES® ANCA
Within-Lab Precision of AESKUSLIDES® ANCA Ethanol and ANCA Formalin processed on HELIOS® (Method A - HELIOS suggestions and Method B - Reader Confirmation of HELIOS images) and processed manually (Method C), based on CLSI Guideline EP12-A2.
Sample Set
10 samples were assayed on AESKUSLIDES® ANCA Ethanol and Formalin. The sample set includes borderline, low, medium, high positive, and negative samples. The samples are characterized in the following table:
| SampleID | Result | Pattern | Grading |
|---|---|---|---|
| 1 | Pos | P | high positive |
| 2 | Pos | P | high positive |
| 3 | Pos | P | medium positive |
| 4 | Pos | P | medium positive |
| 5 | Pos | C | borderline positive |
| 6 | Pos | C | high positive |
| 7 | Pos | C | borderline positive |
| 8 | Pos | C | medium positive |
| 9 | Neg | negative | negative |
| 10 | Neg | negative | negative |
Table 15: Sample table for Precision study_ANCA Ethanol_Within-Lab Samnle AESKUSLIDES ANCA Ethanol
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image of a zipper inside of it. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
| Sample ID | AESKUSLIDES ANCA Formalin | ||
|---|---|---|---|
| Result | Pattern | Grading | |
| 1 | Pos | C | medium positive |
| 2 | Pos | C | high positive |
| 3 | Pos | C | medium positive |
| 4 | Pos | C | high positive |
| 5 | Pos | C | low positive |
| 6 | Pos | C | high positive |
| 7 | Pos | C | medium positive |
| 8 | Pos | C | high positive |
| 9 | Neg | negative | negative |
| 10 | Neg | negative | negative |
Table 16: Sample table for Precision study ANCA Formalin Within-Lab
Procedure
10 samples have been tested on five days, two runs per day, three replicates per sample per run, resulting in 30 data points for each sample.
This Within-lab precision study was done for three different Methods:
| Method A (HELIOS) | Processing of slides and image recording at HELIOS +HELIOS result suggestion |
|---|---|
| Method B (ReaderConfirmation) | HELIOS images recorded in Method A + result analysisby two independent readers |
| Method C (Manual) | Manual processing of slides and result analysis at themicroscope by two independent readers |
All runs have been performed according to the respective IFUs. 1 positive and 1 negative control (kit controls) were included in each run. Samples were tested at 1:20 dilution. Results were analyzed by two independent readers.
Data Analysis
Positive/Negative Classification and Pattern was recorded for each sample in each run and for each Method. Pattern suggestion in Method A (HELIOS) is provided by the Vasculitis Pattern Plus software tool.
For Method C (Manual performance) fluorescence intensity was also reported.
As an estimate of the imprecision of the three methods A, B and C the percentages of positive and negative results will be calculated for each sample:
- % Positive (number of positive calls divided by the total number of data points for each sample)
- % Negative (number of negative calls divided by the total number of data points per sample)
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white stylized image of a zipper inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in a gray sans-serif font, with the word "AESKU" being larger and bolder than "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Results of the two readers (Method B/C) will be calculated and displayed separately as well as combined.
Acceptance Criteria
Positive sera have to be found positive, and negative sera have to be found negative*.
- Reported FI is allowed to differ max. ± 1 level within the study ●
- Calculations should meet the following criteria:
| Estimate of imprecision | HELIOS | ReaderConfirmation | Manual |
|---|---|---|---|
| % Positives of positive samples (excluding borderlinesamples) | > 80% | > 90% | > 90% |
| % Negatives of negative samples | > 80% | > 90% | > 90% |
borderline samples are samples with analyte concentration near the cutoff. These samples are very low positive. They can be evaluated also negative in certain cases, depending on the reader, the subjective manner of result analysis and normal test variances.
For Method A we accept lower percentage of positive/negative results for positive and negative samples, respectively, because we state in the IFU that all results generated by the HELIOS (Method A) must be confirmed by trained personnel.
Results
Results are presented as % positive and % negative results of each sample. Percentages were calculated for the two readers separately (30 replicates for each sample) and for both readers combined (60 data points per sample). All predetermined acceptance criteria were met.
| SampleID | HELIOS | User Confirmation | Manual | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| N | n Neg | n Pos | n Neg | n Pos | n Neg | n Pos | |||||
| Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader2 | ||||
| S1 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S2 | 30 | 4 | 26 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S3 | 30 | 1 | 29 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S4 | 30 | 2 | 28 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S5 | 30 | 23 | 7 | 27 | 19 | 3 | 11 | 27 | 19 | 3 | 11 |
| S6 | 30 | 3 | 27 | 1 | 1 | 29 | 29 | 0 | 0 | 30 | 30 |
| S7 | 30 | 28 | 2 | 19 | 10 | 11 | 20 | 3 | 17 | 27 | 13 |
| S8 | 30 | 2 | 28 | 3 | 3 | 27 | 27 | 0 | 0 | 30 | 30 |
| S9 | 30 | 26 | 4 | 26 | 28 | 4 | 2 | 30 | 30 | 0 | 0 |
| S10 | 30 | 25 | 5 | 30 | 30 | 0 | 0 | 30 | 30 | 0 | 0 |
Table 17: Within-Lab Precision AESKUSLIDES ANCA Ethanol number of positive and negative results
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the company name, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray.
Table 18: Within-Lab Precision AESKUSLIDES ANCA Ethanol percentages of positive and negative results
| SampleID | N | HELIOS | User Confirmation | Manual | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % Negative | % Positive | % Negative | % Positive | % Negative | % Positive | ||||||
| Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader 2 | ||||
| S1 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S2 | 30 | 13.3 | 86.7 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S3 | 30 | 3.3 | 96.7 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S4 | 30 | 6.7 | 93.3 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S5 | 30 | 76.7 | 23.3 | 90.0 | 63.3 | 10.0 | 36.7 | 90.0 | 63.3 | 10.0 | 36.7 |
| S6 | 30 | 10.0 | 90.0 | 3.3 | 3.3 | 96.7 | 96.7 | 0 | 0 | 100 | 100 |
| S7 | 30 | 93.3 | 6.7 | 63.3 | 33.3 | 36.7 | 66.7 | 10 | 56.7 | 90.0 | 43.3 |
| S8 | 30 | 6.7 | 93.3 | 10.0 | 10.0 | 90.0 | 90.0 | 0 | 0 | 100 | 100 |
| S9 | 30 | 86.7 | 13.3 | 86.7 | 93.3 | 13.3 | 6.7 | 100 | 100 | 0 | 0 |
| S10 | 30 | 83.3 | 16.7 | 100 | 100 | 0 | 0 | 100 | 100 | 0 | 0 |
Table 19: Within-Lab Precision AESKUSLIDES ANCA Ethanol percentages of positive and negative results for both readers combined
| SampleID | N(Method A) | HELIOS | N(Method B/C) | User Confirmation | Manual | |||
|---|---|---|---|---|---|---|---|---|
| % Negative | % Positive | % Negative | % Positive | % Negative | % Positive | |||
| S1 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S2 | 30 | 13.3 | 86.7 | 60 | 0 | 100 | 0 | 100 |
| S3 | 30 | 3.3 | 96.7 | 60 | 0 | 100 | 0 | 100 |
| S4 | 30 | 6.7 | 93.3 | 60 | 0 | 100 | 0 | 100 |
| S5 | 30 | 77 | 23 | 60 | 76.7 | 23.3 | 76.7 | 23.3 |
| S6 | 30 | 10.0 | 90.0 | 60 | 3.3 | 96.7 | 0 | 100 |
| S7 | 30 | 93 | 7 | 60 | 48.3 | 51.7 | 33.3 | 66.7 |
| S8 | 30 | 7 | 93 | 60 | 10.0 | 90.0 | 0 | 100 |
| S9 | 30 | 86.7 | 13.3 | 60 | 90.0 | 10.0 | 100 | 0 |
| S10 | 30 | 83.3 | 16.7 | 60 | 100 | 0 | 100 | 0 |
Table 20: Within-Lab Precision AESKUSLIDES ANCA Formalin number of positive and negative results
| N | HELIOS | User Confirmation | Manual | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| SampleID | n Neg | n Negn Pos | n Neg | n Pos | |||||||
| n Pos | Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader2 | Reader1 | Reader2 | |||
| S1 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S2 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S3 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S4 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S5 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S6 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S7 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S8 | 30 | 0 | 30 | 0 | 0 | 30 | 30 | 0 | 0 | 30 | 30 |
| S9 | 30 | 29 | 1 | 27 | 30 | 3 | 0 | 29 | 26 | 1 | 4 |
| S10 | 30 | 27 | 3 | 30 | 27 | 0 | 3 | 30 | 28 | 0 | 2 |
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white graphic inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" also in gray.
Table 21: Within-Lab Precision AESKUSLIDES ANCA Formalin percentages of positive and negative results
| SampleID | N | HELIOS | User Confirmation | Manual | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| %Negative | %Positive | % NegativeReader1 | % NegativeReader2 | % PositiveReader1 | % PositiveReader2 | % NegativeReader1 | % NegativeReader2 | % PositiveReader1 | % PositiveReader2 | ||
| S1 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S2 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S3 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S4 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S5 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S6 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S7 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S8 | 30 | 0 | 100 | 0 | 0 | 100 | 100 | 0 | 0 | 100 | 100 |
| S9 | 30 | 96.7 | 3.3 | 90.0 | 100 | 10.0 | 0 | 97 | 86.7 | 3 | 13.3 |
| S10 | 30 | 90.0 | 10.0 | 100 | 90.0 | 0 | 10.0 | 100 | 93.3 | 0 | 6.7 |
Table 22: Within-Lab Precision AESKUSLIDES ANCA Formalin percentages of positive and negative results for both readers combined
| SampleID | N(Method A) | HELIOS | N(Method B/C) | User Confirmation | Manual | |||
|---|---|---|---|---|---|---|---|---|
| %Negative | % Positive | %Negative | % Positive | %Negative | % Positive | |||
| S1 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S2 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S3 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S4 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S5 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S6 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S7 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S8 | 30 | 0 | 100 | 60 | 0 | 100 | 0 | 100 |
| S9 | 30 | 96.7 | 3.3 | 60 | 95.0 | 5.0 | 91.7 | 8.3 |
| S10 | 30 | 90.0 | 10.0 | 60 | 95.0 | 5.0 | 96.7 | 3.3 |
For manual testing (Method C) of AESKUSLIDES ANCA Ethanol and Formalin fluorescence intensities were at maximum ±1 level within one run (triplicates) and between the runs. For samples found to be positive, pattern determination was consistent for more than 95% of the replicates, using Method B and C, whereas for Method A more than 85% of the determined patterns correspond to each other.
Conclusion
All pre-determined acceptance criteria were met.
b) Between-Lab Precision/Reproducibility AESKUSLIDES ANCA
To assess Between-Lab Precision of AESKUSLIDES ANCA Ethanol and ANCA Formalin processed on HELIOS (Method A - HELIOS suggestions and Method B -Reader Confirmation of HELIOS images) and processed manually (Method C).
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger and bolder than "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Sample Set
10 samples were assayed on AESKUSLIDES ANCA Ethanol and ANCA Formalin. The sample set includes borderline, low, medium, high positive, and negative samples. The samples are characterized in the following table:
| SampleID | AESKUSLIDES ANCA Ethanol | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 1 | Pos | P | high positive | ||||||
| 2 | Pos | P | high positive | ||||||
| 3 | Pos | P | medium positive | ||||||
| 4 | Pos | P | medium positive | ||||||
| 5 | Pos | C | borderline positive | ||||||
| 6 | Pos | C | high positive | ||||||
| 7 | Pos | C | borderline positive | ||||||
| 8 | Pos | C | medium positive | ||||||
| 9 | Neg | negative | negative | ||||||
| 10 | Neg | negative | negative |
Table 23: Sample table for Precision study_ANCA Ethanol_Between-Lab
| Sample ID | AESKUSLIDES ANCA Formalin | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Result | Pattern | Grading | |||||||
| 1 | Pos | C | medium positive | ||||||
| 2 | Pos | C | high positive | ||||||
| 3 | Pos | C | medium positive | ||||||
| 4 | Pos | C | high positive | ||||||
| 5 | Pos | C | low positive | ||||||
| 6 | Pos | C | high positive | ||||||
| 7 | Pos | C | medium positive | ||||||
| 8 | Pos | C | high positive | ||||||
| 9 | Neg | negative | negative | ||||||
| 10 | Neg | negative | negative |
Procedure
10 samples have been tested on five days, two runs per day, three replicates per sample per run, at three different study sites. Two study sites were in the US, one study site was in Germany, with one HELIOS device per study site. Results were analyzed by two different readers per study site, resulting in a total of 90 data points per sample per reader.
This Between-Lab Precision study was done for three different Methods:
| Method A (HELIOS) | Processing of slides and image recording at HELIOS +HELIOS result suggestion |
|---|---|
| Method B (ReaderConfirmation) | HELIOS images recorded in Method A + result analysisby two independent readers |
| Method C (Manual) | Manual processing of slides and result analysis at themicroscope by two independent readers |
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol inside the green square appears to be a stylized representation of a medical or diagnostic tool.
All runs have been performed according to the respective IFUs. 1 positive and 1 negative control (kit controls) were included in each run. Samples were tested at 1:20 dilution. Results were analyzed by two independent readers.
Data Analysis
Positive/negative classification and pattern was recorded for each sample in each run and for each method. Pattern suggestion in Method A (HELIOS) is provided by the Vasculitis Pattern Plus software tool (ANCA).
For Method C (Manual performance) fluorescence intensity was also reported.
The following % agreements including 95% Cl (confidence intervals) will be calculated for each Method A, B and C:
- -Positive % agreement (across all positive samples: number of correctly found samples divided through number of total positive samples)
- -Negative % agreement (across all negative samples: number of correctly found samples divided through number of total negative samples)
- -Pattern % agreement (across all across all positive samples: number of correctly found samples divided through number of total positive samples)
- -Overall % agreement (across all positive and negative samples: number of correctly found samples divided through number of total samples)
- -Fluorescence Intensity % agreement (across all samples, only for Method C: number of correctly found samples divided through number of total samples)
Results of the two readers will be calculated separately as well as combined.
Acceptance Criteria
Positive sera have to be found positive, and negative sera have to be found negative.
Patterns have to be found correctly.
Reported FI is allowed to differ max. ± 1 level from the expected value.
Agreements should meet the following criteria:
| Type of Agreement (borderline positive samples excluded*) | Method A | Method B | Method C |
|---|---|---|---|
| Positive Agreement | ≥ 70% | ≥ 90% | ≥ 90% |
| Negative Agreement | ≥ 70% | ≥ 90% | ≥ 90% |
| Overall Agreement | ≥ 70% | ≥ 90% | ≥ 90% |
| Pattern Agreement | ≥ 70% | ≥ 90% | ≥ 90% |
| Fluorescence Intensity Agreement | --- | --- | ≥ 90% |
- borderline samples are very low positive samples that can be evaluated also negative in certain cases, depending on the reader, the subjective manner of result analysis and normal test variances. Therefore, they may distort positive and pattern agreement, resulting in lower agreements. Therefore, agreements are calculated once including those borderline samples and once excluding them.
For Method A we accept a lower agreement (>70%), because we state in the IFU that all results generated by the HELIOS (Method A) must be confirmed by trained personnel.
{22}------------------------------------------------
From the results of those study, the following will be calculated:
- Between-Lab Precisions, 2. Between-Operator Precisions, 3. Single-Operator Precisions, and 4. Instrument Precision
Results
Results are presented in the tables below. Agreements for AESKUSLIDES® ANCA Ethanol were calculated 1. including the results of the two borderline positive samples, 2. excluding the results of those two samples. For AESKUSLIDES ANCA Formalin results were only calculated once for all tested sera, since no borderline sera were included.
All predetermined acceptance criteria were met.
Between-Site Agreement ANCA Ethanol
Manual (Method C) Between-Site Agreement (Calculation including borderline samples):
| Number of samples | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 894 | 66 | 960 | 948 | 12 | 960 | 948 | 12 | 960 | 1362 | 78 | 1440 |
| Negative Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 360 | 0 | 360 |
| Overall Agreement | 1134 | 66 | 1200 | 1188 | 12 | 1200 | 1188 | 12 | 1200 | 1722 | 78 | 1800 |
| Pattern Agreement | 894 | 66 | 960 | 945 | 15 | 960 | 945 | 15 | 960 | 1359 | 81 | 1440 |
| FI Agreement | 1427 | 13 | 1440 | 1416 | 24 | 1440 | 1416 | 24 | 1440 | 2136 | 24 | 2160 |
Table 25: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline samples). Number of correctly/incorrectly found samples Method C
Table 26: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 93.1 | 91.9 | 98.8 | 94.6 |
| (91.3 - 94.6) | (90 - 93.4) | (97.8 - 99.3) | (93.3 - 95.6) | |
| Negative Agreement | 100 | 100 | 100 | 100 |
| (98.4 - 100) | (98.4 - 100) | (98.4 - 100) | (98.9 - 100) | |
| Overall Agreement | 94.5 | 93.5 | 99 | 95.7 |
| (93.1 - 95.7) | (92 - 94.8) | (98.3 - 99.4) | (94.6 - 96.5) | |
| Pattern Agreement | 93.1 | 91.6 | 98.4 | 94.4 |
| (91.3 - 94.6) | (89.6 - 93.2) | (97.4 - 99.1) | (93.1 - 95.5) | |
| FI Agreement | 99.1 | 99.2 | 98.3 | 98.9 |
| (98.5 - 99.5) | (98.6 - 99.6) | (97.5 - 98.9) | (98.4 - 99.3) |
{23}------------------------------------------------
Manual (Method C) Between-Site Agreement (Calculation excluding borderline samples):
| Number of samples | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 720 | 0 | 720 | 720 | 0 | 720 | 720 | 0 | 720 | 1080 | 0 | 1080 |
| Negative Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 360 | 0 | 360 |
| Overall Agreement | 960 | 0 | 960 | 960 | 0 | 960 | 960 | 0 | 960 | 1440 | 0 | 1440 |
| Pattern Agreement | 720 | 0 | 720 | 717 | 3 | 720 | 717 | 3 | 720 | 1077 | 3 | 1080 |
| FI Agreement | 949 | 11 | 960 | 938 | 22 | 960 | 938 | 22 | 960 | 1418 | 22 | 1440 |
Table 27: Between-Site Agreement_ANCA Ethanol_(Calculation excluding borderline
Table 28: Between-Site Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 100 | 100 | 100 | 100 |
| (99.5 - 100) | (99.5 - 100) | (99.5 - 100) | (99.6 - 100) | |
| Negative Agreement | 100 | 100 | 100 | 100 |
| (98.4 - 100) | (98.4 - 100) | (98.4 - 100) | (98.9 - 100) | |
| Overall Agreement | 100 | 100 | 100 | 100 |
| (99.6 - 100) | (99.6 - 100) | (99.6 - 100) | (99.7 - 100) | |
| Pattern Agreement | 100 | 99.6 | 99.6 | 99.7 |
| (99.5 - 100) | (98.8 - 99.9) | (98.8 - 99.9) | (99.2 - 99.9) | |
| FI Agreement | na | na | na | na |
Reader Confirmation (Method B) Between-Site Agreement (Calculation including borderline samples):
Table 29: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline samples). Number of correctly/incorrectly found samples Method B
| Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Number ofSamples | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Pos | 869 | 91 | 960 | 952 | 8 | 960 | 877 | 83 | 960 | 1349 | 91 | 1440 | |
| Neg | 234 | 6 | 240 | 240 | 0 | 240 | 234 | 6 | 240 | 354 | 6 | 360 | |
| Overall | 1103 | 97 | 1200 | 1192 | 8 | 1200 | 1111 | 89 | 1200 | 1703 | 97 | 1800 | |
| Total Pattern | 837 | 123 | 960 | 897 | 63 | 960 | 850 | 110 | 960 | 1292 | 148 | 1440 | |
| FI | na | na | na | na | na | na | na | na | na | na | na | na |
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white abstract design on the left. To the right of the square is the company name, "AESKU.DIAGNOSTICS," in gray text. Below the company name is the tagline "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Table 30: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 90.5(88.5 - 92.2) | 99.2(98.4 - 99.6) | 91.4(89.4 - 93) | 93.7(92.3 - 94.8) |
| Negative Agreement | 97.5(94.7 - 98.8) | 100(98.4 - 100) | 97.5(94.7 - 98.8) | 98.3(96.4 - 99.2) |
| Overall Agreement | 91.9(90.2 - 93.3) | 99.3(98.7 - 99.7) | 92.6(91 - 93.9) | 94.6(93.5 - 95.6) |
| Total Pattern Agreement | 87.2(84.9 - 89.2) | 93.4(91.7 - 94.8) | 88.5(86.4 - 90.4) | 89.7(88 - 91.2) |
| FI Agreement | na | na | na | na |
Reader Confirmation (Method B) Between-Site Agreement (Calculation excluding borderline samples):
Table 31: Between-Site Agreement_ANCA Ethanol_(Calculation excluding borderline samples). Number of correctly/incorrectly found samples Method B
| Number ofSamples | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Pos | 712 | 8 | 720 | 720 | 0 | 720 | 712 | 8 | 720 | 1072 | 8 | 1080 |
| Neg | 234 | 6 | 240 | 240 | 0 | 240 | 234 | 6 | 240 | 354 | 6 | 360 |
| Overall | 946 | 14 | 960 | 960 | 0 | 960 | 946 | 14 | 960 | 1426 | 14 | 1440 |
| Total Pattern | 680 | 40 | 720 | 665 | 55 | 720 | 685 | 35 | 720 | 1015 | 65 | 1080 |
| FI | na | na | na | na | na | na | na | na | na | na | na |
Table 32: Between-Site Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 98.9(97.8 - 99.4) | 100(99.5 - 100) | 98.9(97.8 - 99.4) | 99.3(98.5 - 99.6) |
| Negative Agreement | 97.5(94.7 - 98.8) | 100(98.4 - 100) | 97.5(94.7 - 98.8) | 98.3(96.4 - 99.2) |
| Overall Agreement | 98.5(97.6 - 99.1) | 100(99.6 - 100) | 98.5(97.6 - 99.1) | 99(98.4 - 99.4) |
| Total Pattern Agreement | 94.4(92.5 - 95.9) | 92.4(90.2 - 94.1) | 95.1(93.3 - 96.5) | 94(92.4 - 95.3) |
| FI Agreement | na | na | na | na |
{25}------------------------------------------------
HELIOS (Method A) Between-Site Agreement (Calclation including borderline samples):
| samples). Number of correctly/incorrectly found samples Method A | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Number ofSamples | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Pos | 411 | 69 | 480 | 472 | 8 | 480 | 415 | 65 | 480 | 649 | 71 | 720 |
| Neg | 102 | 18 | 120 | 111 | 9 | 120 | 111 | 9 | 120 | 162 | 18 | 180 |
| Overall | 513 | 87 | 600 | 583 | 17 | 600 | 526 | 74 | 600 | 811 | 89 | 900 |
| Total Pattern | 298 | 182 | 480 | 345 | 135 | 480 | 309 | 171 | 480 | 476 | 244 | 720 |
| FI | na | na | na | na | na | na | na | na | na | na | na | na |
Table 33: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline
Table 34: Between-Site Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method A
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 85.6(82.2 - 88.5) | 98.3(96.7 - 99.2) | 86.5(83.1 - 89.2) | 90.1(87.7 - 92.1) |
| Negative Agreement | 85(77.5 - 90.3) | 92.5(86.4 - 96) | 92.5(86.4 - 96) | 90(84.7 - 93.6) |
| Overall Agreement | 85.5(82.5 - 88.1) | 97.2(95.5 - 98.2) | 87.7(84.8 - 90.1) | 90.1(88 - 91.9) |
| Total Pattern Agreement | 62.1(57.7 - 66.3) | 71.9(67.7 - 75.7) | 64.4(60 - 68.5) | 66.1(62.6 - 69.5) |
| FI Agreement | na | na | na | na |
HELIOS (Method A) Between-Site Agreement (Calclation excluding borderline samples):
| Table 35: Between-Site Agreement_ANCA Ethanol_(Calculation excluding borderlinesamples). Number of correctly/incorrectly found samples Method A | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | |||||||||
| Number ofSamples | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers |
| Pos | 342 | 18 | 360 | 352 | 8 | 360 | 346 | 14 | 360 | 520 | 20 | 540 |
| Neg | 102 | 18 | 120 | 111 | 9 | 120 | 111 | 9 | 120 | 162 | 18 | 180 |
| Overall | 444 | 36 | 480 | 463 | 17 | 480 | 457 | 23 | 480 | 682 | 38 | 720 |
| Total Pattern | 278 | 82 | 360 | 308 | 52 | 360 | 292 | 68 | 360 | 439 | 101 | 540 |
| FI | na | na | na | na | na | na | na | na | na | na | na | na |
Table 35: Between-Site Agreement ANCA Ethanol (Calculation excluding borderline
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo has a green square on the left with a white image inside. The text "AESKU.DIAGNOSTICS" is in gray, with the word "AESKU" being larger than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Table 36: Between-Site Agreement ANCA Ethanol (Calculation excluding borderline samples). % Agreements (95% CI) Method A
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 95(92.2 - 96.8) | 97.8(95.7 - 98.9) | 96.1(93.6 - 97.7) | 96.3(94.3 - 97.6) |
| Negative Agreement | 85(77.5 - 90.3) | 92.5(86.4 - 96) | 92.5(86.4 - 96) | 90(84.7 - 93.6) |
| Overall Agreement | 92.5(89.8 - 94.5) | 96.5(94.4 - 97.8) | 95.2(92.9 - 96.8) | 94.7(92.8 - 96.1) |
| Total Pattern Agreement | 77.2(72.6 - 81.3) | 85.6(81.5 - 88.8) | 81.1(76.7 - 84.8) | 81.3(77.8 - 84.4) |
| FI Agreement | na | na | na | na |
For Between-site Agreement of ANCA Ethanol, all acceptance critieria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites. All agreements were > 70% for Method A (HELIOS suggestion) for all study sites.
Overall-Between-Lab agreements ranged from 92.5% to 96.5% for Method A (HELIOS), from 98.5% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
Between-Operator Agreement ANCA Ethanol
Manual (Method C) Between-Operator Agreement (Calculation including borderline samples):
| Number of Samples | Site 1 | Site 2 | Site 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 414 | 66 | 480 | 480 | 0 | 480 | 468 | 12 | 480 |
| Negative Agreement | 120 | 0 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Overall Agreement | 534 | 66 | 600 | 600 | 0 | 600 | 588 | 12 | 600 |
| Pattern Agreement | 414 | 66 | 480 | 480 | 0 | 480 | 465 | 15 | 480 |
| FI Agreement | 720 | 0 | 720 | 707 | 13 | 720 | 709 | 11 | 720 |
| Table 37: Between-Operator Agreement_ANCA Ethanol_(Calculation including borderline | |
|---|---|
| samples). Number of correctly/incorrectly found samples Method C |
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Table 38: Between-Operator Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 86.3(82.9 - 89) | 100(99.2 - 100) | 97.5(95.7 - 98.6) |
| Negative Agreement | 100(96.9 - 100) | 100(96.9 - 100) | 100(96.9 - 100) |
| Overall Agreement | 89(86.2 - 91.3) | 100(99.4 - 100) | 98(96.5 - 98.9) |
| Pattern Agreement | 86.3(82.9 - 89) | 100(99.2 - 100) | 96.9(94.9 - 98.1) |
| FI Agreement | 100(99.5 - 100) | 98.2(96.9 - 98.9) | 98.5(97.3 - 99.1) |
Manual (Method C) Between-Operator Agreement (Calculation excluding borderline samples):
Table 39: Between-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). Number of correctly/incorrectly found samples Method C
| Site 1 | Site 2 | Site 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Number of Samples | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers |
| Positive Agreement | 360 | 0 | 360 | 360 | 0 | 360 | 360 | 0 | 360 |
| Negative Agreement | 120 | 0 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Overall Agreement | 480 | 0 | 480 | 480 | 0 | 480 | 480 | 0 | 480 |
| Pattern Agreement | 360 | 0 | 360 | 360 | 0 | 360 | 357 | 3 | 360 |
| FI Agreement | 480 | 0 | 480 | 469 | 11 | 480 | 469 | 11 | 480 |
Table 40: Between-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 100(98.9 - 100) | 100(98.9 - 100) | 100(98.9 - 100) |
| Negative Agreement | 100(96.9 - 100) | 100(96.9 - 100) | 100(96.9 - 100) |
| Overall Agreement | 100(99.2 - 100) | 100(99.2 - 100) | 100(99.2 - 100) |
| Pattern Agreement | 100(98.9 - 100) | 100(98.9 - 100) | 99.2(97.6 - 99.7) |
| FI Agreement | 100(99.2 - 100) | 97.7(95.9 - 98.7) | 97.7(95.9 - 98.7) |
{28}------------------------------------------------
Image /page/28/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white stylized image of a molecule or antibody inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in a dark gray, sans-serif font. Below the company name is the tagline "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter gray font.
Reader Confirmation (Method B) Between-Operator Agreement (Calculation including borderline samples):
Table 41: Between-Operator Agreement_ANCA Ethanol_(Calculation including borderline samples). Number of correctly/incorrectly found samples Method B
| Number of Samples | Site 1 | Site 2 | Site 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Pos | 397 | 83 | 480 | 472 | 8 | 480 | 480 | 0 | 480 |
| Neg | 114 | 6 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Overall | 511 | 89 | 600 | 592 | 8 | 600 | 600 | 0 | 600 |
| Total Pattern | 395 | 85 | 480 | 442 | 38 | 480 | 455 | 25 | 480 |
| FI | na | na | na | na | na | na | na | na | na |
Table 42: Between-Operator Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 82.7(79.1 - 85.8) | 98.3(96.7 - 99.2) | 100(99.2 - 100) |
| Negative Agreement | 95(89.5 - 97.7) | 100(96.9 - 100) | 100(96.9 - 100) |
| Overall Agreement | 85.2(82.1 - 87.8) | 98.7(97.4 - 99.3) | 100(99.4 - 100) |
| Total Pattern Agreement | 82.3(78.6 - 85.4) | 92.1(89.3 - 94.2) | 94.8(92.4 - 96.4) |
| FI Agreement | na | na | na |
Reader Confirmation (Method B) Between-Operator Agreement (Calculation excluding borderline samples):
Table 43: Between-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). Number of correctly/incorrectly found samples Method B
| Number of Samples | Site 1 | Site 2 | Site 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Pos | 352 | 8 | 360 | 360 | 0 | 360 | 360 | 0 | 360 |
| Neg | 114 | 6 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Overall | 466 | 14 | 480 | 480 | 0 | 480 | 480 | 0 | 480 |
| Total Pattern | 350 | 10 | 360 | 330 | 30 | 360 | 335 | 25 | 360 |
| FI | na | na | na | na | na | na | na | na | na |
{29}------------------------------------------------
Image /page/29/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol in the green square appears to be a stylized representation of a medical or diagnostic tool.
Table 44: Between-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 97.8(95.7 - 98.9) | 100(98.9 - 100) | 100(98.9 - 100) |
| Negative Agreement | 95(89.5 - 97.7) | 100(96.9 - 100) | 100(96.9 - 100) |
| Overall Agreement | 97.1(95.2 - 98.3) | 100(99.2 - 100) | 100(99.2 - 100) |
| Total Pattern Agreement | 97.2(95 - 98.5) | 91.7(88.4 - 94.1) | 93.1(89.9 - 95.3) |
| FI Agreement | na | na | na |
For Between-Operator Agreement of ANCA Ethanol, all acceptance criteria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites.
Overall- Between-Operator agreements ranged from 97.1% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
Single-Operator Agreement ANCA Ethanol
Manual (Method C) Single-Operator Agreement (Calculation including borderline samples):
| Table 45: Single-Operator Agreement_ANCA Ethanol_(Calculation including borderline | |
|---|---|
| samples). Number of correctly/incorrectly found samples Method C |
| Number ofSamples | Site 1 | Site 2 | Site 3 | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | ||||||||||
| FoundCorrectly | FoundIncorrectl | ExpectedNumbers | FoundCorrectly | FoundIncorrectl | ExpectedNumbers | FoundCorrectly | FoundIncorrectl | ExpectedNumbers | |||||||
| PositiveAgreement | 210 | 30 | 240 | 204 | 36 | 240 | 240 | 0 | 240 | 228 | 12 | 240 | 240 | 0 | 240 |
| NegativeAgreement | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 |
| OverallAgreement | 270 | 30 | 300 | 264 | 36 | 300 | 300 | 0 | 300 | 288 | 12 | 300 | 300 | 0 | 300 |
| PatternAgreement | 210 | 30 | 240 | 204 | 36 | 240 | 240 | 0 | 240 | 228 | 12 | 240 | 237 | 3 | 240 |
| FI Agreement | 360 | 0 | 360 | 360 | 0 | 360 | 351 | 9 | 360 | 356 | 4 | 360 | 353 | 7 | 360 |
{30}------------------------------------------------
Image /page/30/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white medical symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
| Table 46: Single-Operator Agreement_ANCA Ethanol_(Calculation including borderline | ||
|---|---|---|
| samples). % Agreements (95% CI) Method C |
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 | |||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |
| Positive Agreement | 87.5(82.7 - 91.1) | 85(79.9 - 89) | 100(98.4 - 100) | 100(98.4 - 100) | 95(91.5 - 97.1) | 100(98.4 - 100) |
| Negative Agreement | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) |
| Overall Agreement | 90(86.1 - 92.9) | 88(83.8 - 91.2) | 100(98.7 - 100) | 100(98.7 - 100) | 96(93.1 - 97.7) | 100(98.7 - 100) |
| Pattern Agreement | 87.5(82.7 - 91.1) | 85(79.9 - 89) | 100(98.4 - 100) | 100(98.4 - 100) | 95(91.5 - 97.1) | 98.8(96.4 - 99.6) |
| FI Agreement | 100(98.9 - 100) | 100(98.9 - 100) | 97.5(95.3 - 98.7) | 98.9(97.2 - 99.6) | 98.9(97.2 - 99.6) | 98.1(96 - 99.1) |
Manual (Method C) Single-Operator Agreement (Calculation excluding borderline samples):
| Number of Samples | Site 1 | Site 2 | Site 3 | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |||||||||||||
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 |
| Negative Agreement | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 |
| Overall Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 |
| Pattern Agreement | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 177 | 3 | 180 |
| FI Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 231 | 9 | 240 | 238 | 2 | 240 | 236 | 4 | 240 | 233 | 7 | 240 |
Table 47: Single-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline than C
Table 48: Single-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method C
| Site 1 | Site 2 | Site 3 | ||||
|---|---|---|---|---|---|---|
| % Agreement (95%CI) | Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 |
| Positive Agreement | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) |
| Negative Agreement | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) |
| Overall Agreement | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) |
| Total PatternAgreement | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 98.3(95.2 - 99.4) |
| FI Agreement | 100(98.4 - 100) | 100(98.4 - 100) | 96.3(93 - 98) | 99.2(97 - 99.8) | 98.3(95.8 - 99.4) | 97.1(94.1 - 98.6) |
{31}------------------------------------------------
Image /page/31/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Reader Confirmation (Method B) Single-Operator Agreement (Calculation including borderline samples):
Table 49: Single-Operator Agreement_ANCA Ethanol_(Calculation including borderline samples). Number of correctly/incorrectly found samples Method B
| Number ofSamples | Site 1 | Site 2 | Site 3 | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1FoundCorrectly | Reader 1FoundIncorrectly | Reader 1ExpectedNumbers | Reader 2FoundCorrectly | Reader 2FoundIncorrectly | Reader 2ExpectedNumbers | Reader 3FoundCorrectly | Reader 3FoundIncorrectly | Reader 3ExpectedNumbers | Reader 4FoundCorrectly | Reader 4FoundIncorrectly | Reader 4ExpectedNumbers | Reader 5FoundCorrectly | Reader 5FoundIncorrectly | Reader 5ExpectedNumbers | Reader 6FoundCorrectly | Reader 6FoundIncorrectly | Reader 6ExpectedNumbers | |
| PositiveAgreement | 190 | 50 | 240 | 207 | 33 | 240 | 236 | 4 | 240 | 236 | 4 | 240 | 240 | 0 | 240 | 240 | 0 | 240 |
| NegativeAgreement | 56 | 4 | 60 | 58 | 2 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 |
| OverallAgreement | 246 | 54 | 300 | 265 | 35 | 300 | 296 | 4 | 300 | 296 | 4 | 300 | 300 | 0 | 300 | 300 | 0 | 300 |
| PatternAgreement | 190 | 50 | 240 | 205 | 35 | 240 | 221 | 19 | 240 | 221 | 19 | 240 | 228 | 12 | 240 | 227 | 13 | 240 |
| FI Agreement | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na |
Table 50: Single-Operator Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 | |||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |
| Positive Agreement | 79.2(73.6 - 83.8) | 86.3(81.3 - 90) | 98.3(95.8 - 99.4) | 98.3(95.8 - 99.4) | 100(98.4 - 100) | 100(98.4 - 100) |
| Negative Agreement | 93.3(84.1 - 97.4) | 96.7(88.6 - 99.1) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) |
| Overall Agreement | 82(77.3 - 85.9) | 88.3(84.2 - 91.5) | 98.7(96.6 - 99.5) | 98.7(96.6 - 99.5) | 100(98.7 - 100) | 100(98.7 - 100) |
| Total PatternAgreement | 79.2(73.6 - 83.8) | 85.4(80.4 - 89.3) | 92.1(88 - 94.9) | 92.1(88 - 94.9) | 95(91.5 - 97.1) | 94.6(91 - 96.8) |
| FI Agreement | na | na | na | na | na | na |
{32}------------------------------------------------
Image /page/32/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image of a zipper inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Reader Confirmation (Method B) Single-Operator Agreement (Calculation excluding borderline samples):
Table 51: Single-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). Number of correctly/incorrectly found samples Method B
| Number of Samples | Site 1 | Site 2 | Site 3 | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1Found Correctly | Reader 1Found Incorrectly | Reader 1Expected Numbers | Reader 2Found Correctly | Reader 2Found Incorrectly | Reader 2Expected Numbers | Reader 3Found Correctly | Reader 3Found Incorrectly | Reader 3Expected Numbers | Reader 4Found Correctly | Reader 4Found Incorrectly | Reader 4Expected Numbers | Reader 5Found Correctly | Reader 5Found Incorrectly | Reader 5Expected Numbers | Reader 6Found Correctly | Reader 6Found Incorrectly | Reader 6Expected Numbers | |
| Positive Agreement | 176 | 4 | 180 | 176 | 4 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 | 180 | 0 | 180 |
| Negative Agreement | 56 | 4 | 60 | 58 | 2 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 |
| Overall Agreement | 232 | 8 | 240 | 234 | 6 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 |
| Pattern Agreement | 176 | 4 | 180 | 174 | 6 | 180 | 165 | 15 | 180 | 165 | 15 | 180 | 168 | 12 | 180 | 167 | 13 | 180 |
| FI Agreement | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na |
Table 52: Single-Operator Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 | |||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |
| Positive Agreement | 97.8(94.4 - 99.1) | 97.8(94.4 - 99.1) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) | 100(97.9 - 100) |
| Negative Agreement | 93.3(84.1 - 97.4) | 96.7(88.6 - 99.1) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) |
| Overall Agreement | 96.7(93.6 - 98.3) | 97.5(94.7 - 98.8) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) |
| Total PatternAgreement | 97.8(94.4 - 99.1) | 96.7(92.9 - 98.5) | 91.7(86.7 - 94.9) | 91.7(86.7 - 94.9) | 93.3(88.7 - 96.1) | 92.8(88 - 95.7) |
| FI Agreement | na | na | na | na | na | na |
For Single-Operator Agreement of ANCA Ethanol, all acceptance criteria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites.
Overall- Single-Operator agreements ranged from 96.7% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
{33}------------------------------------------------
Instrument Precision ANCA Ethanol
HELIOS (Method A) Instrument Agreement (Calculation including borderline samples):
Table 53: Instrument Agreement_ANCA Ethanol_(Calculation including borderline samples). Number of correctly/incorrectly found samples Method A
| Number ofSamples | Site 1HELIOS 1 | Site 2HELIOS 2 | Site 3HELIOS 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| PositiveAgreement | 177 | 63 | 240 | 234 | 6 | 240 | 238 | 2 | 240 |
| NegativeAgreement | 51 | 9 | 60 | 51 | 9 | 60 | 60 | 0 | 60 |
| OverallAgreement | 228 | 72 | 300 | 285 | 15 | 300 | 298 | 2 | 300 |
| PatternAgreement | 131 | 109 | 240 | 167 | 73 | 240 | 178 | 62 | 240 |
| FI Agreement | na | na | na | na | na | na | na | na | na |
Table 54: Instrument Agreement_ANCA Ethanol_(Calculation including borderline samples). % Agreements (95% CI) Method A
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| HELIOS 1 | HELIOS 2 | HELIOS 3 | |
| Positive Agreement | 73.8(67.8 - 78.9) | 97.5(94.7 - 98.8) | 99.2(97 - 99.8) |
| Negative Agreement | 85(73.9 - 91.9) | 85(73.9 - 91.9) | 100(94 - 100) |
| Overall Agreement | 76(70.9 - 80.5) | 95(91.9 - 96.9) | 99.3(97.6 - 99.8) |
| Pattern Agreement | 54.6(48.3 - 60.8) | 69.6(63.5 - 75.1) | 74.2(68.3 - 79.3) |
| FI Agreement | na | na | na |
{34}------------------------------------------------
Image /page/34/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white stylized image of a molecule inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger and bolder than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
HELIOS (Method A) Instrument Agreement (Calculation excluding borderline samples):
| Number ofSamples | Site 1HELIOS 1 | Site 2HELIOS 2 | Site 3HELIOS 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| PositiveAgreement | 168 | 12 | 180 | 174 | 6 | 180 | 178 | 2 | 180 |
| NegativeAgreement | 51 | 9 | 60 | 51 | 9 | 60 | 60 | 0 | 60 |
| OverallAgreement | 219 | 21 | 240 | 225 | 15 | 240 | 238 | 2 | 240 |
| PatternAgreement | 131 | 49 | 180 | 147 | 33 | 180 | 161 | 19 | 180 |
| FI Agreement | na | na | na | na | na | na | na | na | na |
Table 55: Instrument Agreement_ANCA Ethanol_(Calculation excluding borderline samples). Number of correctly/incorrectly found samples Method A
Table 56: Instrument Agreement_ANCA Ethanol_(Calculation excluding borderline samples). % Agreements (95% CI) Method A
| % Agreement (95% CI) | Site 1HELIOS 1 | Site 2HELIOS 2 | Site 3HELIOS 3 |
|---|---|---|---|
| Positive Agreement | 93.3(88.7 - 96.1) | 96.7(92.9 - 98.5) | 98.9(96 - 99.7) |
| Negative Agreement | 85(73.9 - 91.9) | 85(73.9 - 91.9) | 100(94 - 100) |
| Overall Agreement | 91.3(87 - 94.2) | 93.8(89.9 - 96.2) | 99.2(97 - 99.8) |
| Pattern Agreement | 72.8(65.9 - 78.8) | 81.7(75.4 - 86.6) | 89.4(84.1 - 93.1) |
| FI Agreement | na | na | na |
For Instrument to Instrument Agreement with ANCA Ethanol, all acceptance criteria were met. All agreements were >70% for all study sites.
Overall- Instrument to Instrument agreements ranged from 91.3% to 99.2%.
We clearly state that all results suggested by HELIOS have to be confirmed by a trained reader.
{35}------------------------------------------------
Between-Site Agreement ANCA Formalin
Manual (Method C) Between-Site Agreement:
Table 57: Between-Site Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method C
| AESKUSLIDES ANCA Formalin Manual Performance (Method C)Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Number ofSamples | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers |
| PositiveAgreement | 960 | 0 | 960 | 945 | 15 | 960 | 945 | 15 | 960 | 1425 | 15 | 1440 |
| NegativeAgreement | 233 | 7 | 240 | 240 | 0 | 240 | 233 | 7 | 240 | 353 | 7 | 360 |
| OverallAgreement | 1193 | 7 | 1200 | 1185 | 15 | 1200 | 1178 | 22 | 1200 | 1778 | 22 | 1800 |
| PatternAgreement | 960 | 0 | 960 | 945 | 15 | 960 | 945 | 15 | 960 | 1425 | 15 | 1440 |
| FI Agreement | 897 | 3 | 900 | 1166 | 34 | 1200 | 869 | 31 | 900 | 1466 | 34 | 1500 |
Table 58: Between-Site Agreement_ANCA Formalin % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 100 | 98.4 | 98.4 | 99 |
| (99.6 - 100) | (97.4 - 99.1) | (97.4 - 99.1) | (98.3 - 99.4) | |
| Negative Agreement | 97.1 | 100 | 97.1 | 98.1 |
| (94.1 - 98.6) | (98.4 - 100) | (94.1 - 98.6) | (96 - 99.1) | |
| Overall Agreement | 99.4 | 98.8 | 98.2 | 98.8 |
| (98.8 - 99.7) | (97.9 - 99.2) | (97.2 - 98.8) | (98.2 - 99.2) | |
| Pattern Agreement | 100 | 98.4 | 98.4 | 99 |
| (99.6 - 100) | (97.4 - 99.1) | (97.4 - 99.1) | (98.3 - 99.4) | |
| FI Agreement | 99.7 | 97.2 | 96.6 | 97.7 |
| (99 - 99.9) | (96.1 - 98) | (95.2 - 97.6) | (96.8 - 98.4) |
Reader Confirmation (Method B) Between-Site Agreement:
Table 59: Between-Site Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method B
| Number of Samples | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 931 | 29 | 960 | 924 | 36 | 960 | 953 | 7 | 960 | 1404 | 36 | 1440 |
| Negative Agreement | 224 | 16 | 240 | 226 | 14 | 240 | 230 | 10 | 240 | 340 | 20 | 360 |
| Overall Agreement | 1155 | 45 | 1200 | 1150 | 50 | 1200 | 1183 | 17 | 1200 | 1744 | 56 | 1800 |
| Pattern Agreement | 914 | 46 | 960 | 898 | 62 | 960 | 944 | 16 | 960 | 1378 | 62 | 1440 |
| FI Agreement | na | na | na | na | na | na | na | na | na | na | na |
{36}------------------------------------------------
Image /page/36/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white stylized image of a zipper in the left side of the logo. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger and bolder than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 97(95.7 - 97.9) | 96.3(94.9 - 97.3) | 99.3(98.5 - 99.6) | 97.5(96.6 - 98.2) |
| Negative Agreement | 93.3(89.4 - 95.9) | 94.2(90.4 - 96.5) | 95.8(92.5 - 97.7) | 94.4(91.6 - 96.4) |
| Overall Agreement | 96.3(95 - 97.2) | 95.8(94.5 - 96.8) | 98.6(97.7 - 99.1) | 96.9(96 - 97.6) |
| Pattern Agreement | 95.2(93.7 - 96.4) | 93.5(91.8 - 94.9) | 98.3(97.3 - 99) | 95.7(94.5 - 96.6) |
| FI Agreement | na | na | na | na |
Table 60: Between-Site Agreement ANCA Formalin % Agreements (95% Cl) Method B
HELIOS (Method A) Between-Site Agreement (Calclation including borderline samples):
Table 61: Between-Site Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method A
| Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Number of Samples | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers |
| Positive Agreement | 461 | 19 | 480 | 456 | 24 | 480 | 475 | 5 | 480 | 696 | 24 | 720 |
| Negative Agreement | 102 | 18 | 120 | 83 | 37 | 120 | 93 | 27 | 120 | 139 | 41 | 180 |
| Overall Agreement | 563 | 37 | 600 | 539 | 61 | 600 | 568 | 32 | 600 | 835 | 65 | 900 |
| Pattern Agreement | 422 | 58 | 480 | 404 | 76 | 480 | 440 | 40 | 480 | 633 | 87 | 720 |
| FI Agreement | na | na | na | na | na | na | na | na | na | na | na |
| Table 62: Between-Site Agreement ANCA Formalin. % Agreements (95% Cl) Method A | ||
|---|---|---|
| % Agreement (95% CI) | Site 1 vs Site 2 | Site 2 vs Site 3 | Site 1 vs Site 3 | All Sites |
|---|---|---|---|---|
| Positive Agreement | 96(93.9 - 97.5) | 95(92.7 - 96.6) | 99(97.6 - 99.6) | 96.7(95.1 - 97.7) |
| Negative Agreement | 85(77.5 - 90.3) | 69.2(60.4 - 76.7) | 77.5(69.2 - 84.1) | 77.2(70.6 - 82.7) |
| Overall Agreement | 93.8(91.6 - 95.5) | 89.8(87.2 - 92) | 94.7(92.6 - 96.2) | 92.8(90.9 - 94.3) |
| Pattern Agreement | 87.9(84.7 - 90.5) | 84.2(80.6 - 87.2) | 91.7(88.9 - 93.8) | 87.9(85.3 - 90.1) |
| FI Agreement | na | na | na | na |
For Between-site Agreement of ANCA Formalin, all acceptance criteria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites. All agreements were > 70% for Method A (HELIOS suggestion) for all study sites.
Overall-Between-Lab agreements ranged from 89.8% to 94.7% for Method A (HELIOS), from 95.8% to 98.6% for Method B (Reader Confirmation), and from 98.2% to 99.4% for Method C (Manual).
{37}------------------------------------------------
Between-Operator Agreement ANCA Formalin
Manual (Method C) Between-Operator Agreement (Calculation including borderline samples):
Table 63: Between-Operator Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method C
| Number of Samples | Site 1 | Site 2 | Site 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Pos | 480 | 0 | 480 | 480 | 0 | 480 | 465 | 15 | 480 |
| Neg | 113 | 7 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Overall | 593 | 7 | 600 | 600 | 0 | 600 | 585 | 15 | 600 |
| Single Pattern C-ANCA | 480 | 0 | 480 | 480 | 0 | 480 | 465 | 15 | 480 |
| FI | 300 | 0 | 300 | 597 | 3 | 600 | 569 | 31 | 600 |
Table 64: Between-Operator Agreement_ANCA Formalin. % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 100(99.2 - 100) | 100(99.2 - 100) | 96.9(94.9 - 98.1) |
| Negative Agreement | 94.2(88.4 - 97.1) | 100(96.9 - 100) | 100(96.9 - 100) |
| Overall Agreement | 98.8(97.6 - 99.4) | 100(99.4 - 100) | 97.5(95.9 - 98.5) |
| Single Pattern C-ANCA Agreement | 100(99.2 - 100) | 100(99.2 - 100) | 96.9(94.9 - 98.1) |
| FI Agreement | 100(98.7 - 100) | 99.5(98.5 - 99.8) | 94.8(92.8 - 96.3) |
{38}------------------------------------------------
Reader Confirmation (Method B) Between-Operator Agreement:
Table 65: Between-Operator Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method B
| Site 1 | Site 2 | Site 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Number of Samples | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers |
| Positive Agreement | 480 | 0 | 480 | 451 | 29 | 480 | 473 | 7 | 480 |
| Negative Agreement | 114 | 6 | 120 | 110 | 10 | 120 | 116 | 4 | 120 |
| Overall Agreement | 594 | 6 | 600 | 561 | 39 | 600 | 589 | 11 | 600 |
| Pattern Agreement | 480 | 0 | 480 | 434 | 46 | 480 | 464 | 16 | 480 |
| FI Agreement | na | na | na | na | na | na | na | na | na |
Table 66: Between-Operator Agreement_ANCA Formalin. % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| Positive Agreement | 100(99.2 - 100) | 94(91.5 - 95.8) | 98.5(97 - 99.3) |
| Negative Agreement | 95(89.5 - 97.7) | 91.7(85.3 - 95.4) | 96.7(91.7 - 98.7) |
| Overall Agreement | 99(97.8 - 99.5) | 93.5(91.2 - 95.2) | 98.2(96.7 - 99) |
| Pattern Agreement | 100(99.2 - 100) | 90.4(87.5 - 92.7) | 96.7(94.7 - 97.9) |
| FI Agreement | na | na | na |
For Between-Operator Agreement of ANCA Formalin, all acceptance critieria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites.
Overall- Between-Operator agreements ranged from 93.5% to 99% for Method B (Reader Confirmation), and from 97.5% to 100% for Method C (Manual).
{39}------------------------------------------------
Single-Operator Agreement ANCA Formalin
Manual (Method C) Single-Operator Agreement:
Table 67: Single-Operator Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method C
| Site 1 | Site 2 | Site 3 | |||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Number of Samples | Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |||||||||||||
| Found Correctly | Found Incorrectl | Expected Numbers | Found Correctly | Found Incorrectl | Expected Numbers | Found Correctly | Found Incorrectl | Expected Numbers | Found Correctly | Found Incorrectl | Expected Numbers | Found Correctly | Found Incorrectl | Expected Numbers | Found Correctly | Found Incorrectl | Expected Numbers | ||
| Positive Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 234 | 6 | 240 | 231 | 9 | 240 | |
| Negative Agreement | 59 | 1 | 60 | 54 | 6 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | 60 | 0 | 60 | |
| Overall Agreement | 299 | 1 | 300 | 294 | 6 | 300 | 300 | 0 | 300 | 300 | 0 | 300 | 294 | 6 | 300 | 291 | 9 | 300 | |
| Pattern Agreement | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 240 | 0 | 240 | 234 | 6 | 240 | 231 | 9 | 240 | |
| FI Agreement | 300 | 0 | 300 | 300 | 0 | 300 | 300 | 0 | 300 | 297 | 3 | 300 | 293 | 7 | 300 | 276 | 24 | 300 |
Table 68: Single-Operator Agreement_ANCA Formalin. % Agreements (95% CI) Method C
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 | |||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |
| Positive Agreement | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 97.5(94.7 - 98.8) | 96.3(93 - 98) |
| Negative Agreement | 98.3(91.1 - 99.7) | 90(79.9 - 95.3) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(94 - 100) |
| Overall Agreement | 99.7(98.1 - 99.9) | 98(95.7 - 99.1) | 100(98.7 - 100) | 100(98.7 - 100) | 98(95.7 - 99.1) | 97(94.4 - 98.4) |
| Single Pattern C-ANCAAgreement | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 100(98.4 - 100) | 97.5(94.7 - 98.8) | 96.3(93 - 98) |
| FI Agreement | 100(98.7 - 100) | 100(98.7 - 100) | 100(98.7 - 100) | 99(97.1 - 99.7) | 97.7(95.3 - 98.9) | 92(88.4 - 94.6) |
{40}------------------------------------------------
Reader Confirmation (Method B) Single-Operator Agreement (Calculation including borderline samples):
| Table 69: Single-Operator Agreement ANCA Formalin Number of correctly/incorrectly found | |
|---|---|
| samples Method B |
| Number ofSamples | Site 1 | Site 2 | Site 3 | |||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |||||||||||||||||||
| Found | Correctly | FoundIncorrectly | ExpectedNumbers | Found | Correctly | FoundIncorrectly | ExpectedNumbers | Found | Correctly | FoundIncorrectly | ExpectedNumbers | Found | Correctly | FoundIncorrectly | ExpectedNumbers | Found | Correctly | FoundIncorrectly | ExpectedNumbers | Found | Correctly | FoundIncorrectly | ExpectedNumbers | |
| PositiveAgreement | 240 | 0 | 240 | 240 | 0 | 240 | 223 | 17 | 240 | 228 | 12 | 240 | 235 | 5 | 240 | 238 | 2 | 240 | ||||||
| NegativeAgreement | 57 | 3 | 60 | 57 | 3 | 60 | 55 | 5 | 60 | 55 | 5 | 60 | 59 | 1 | 60 | 57 | 3 | 60 | ||||||
| OverallAgreement | 297 | 3 | 300 | 297 | 3 | 300 | 278 | 22 | 300 | 283 | 17 | 300 | 294 | 6 | 300 | 295 | 5 | 300 | ||||||
| PatternAgreement | 240 | 0 | 240 | 240 | 0 | 240 | 218 | 22 | 240 | 216 | 24 | 240 | 232 | 8 | 240 | 232 | 8 | 240 | ||||||
| FI Agreement | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na | na |
Table 70: Single-Operator Agreement_ANCA Formalin. % Agreements (95% CI) Method B
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 | |||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 3 | Reader 4 | Reader 5 | Reader 6 | |
| Positive Agreement | 100(98.4 - 100) | 100(98.4 - 100) | 92.9(89 - 95.5) | 95(91.5 - 97.1) | 97.9(95.2 - 99.1) | 99.2(97 - 99.8) |
| Negative Agreement | 95(86.3 - 98.3) | 95(86.3 - 98.3) | 91.7(81.9 - 96.4) | 91.7(81.9 - 96.4) | 98.3(91.1 - 99.7) | 95(86.3 - 98.3) |
| Overall Agreement | 99(97.1 - 99.7) | 99(97.1 - 99.7) | 92.7(89.1 - 95.1) | 94.3(91.1 - 96.4) | 98(95.7 - 99.1) | 98.3(96.2 - 99.3) |
| Pattern Agreement | 100(98.4 - 100) | 100(98.4 - 100) | 90.8(86.5 - 93.9) | 90(85.6 - 93.2) | 96.7(93.6 - 98.3) | 96.7(93.6 - 98.3) |
| FI Agreement | na | na | na | na | na | na |
For Single-Operator Agreement of ANCA Formalin, all acceptance critieria were met. All agreements were >90% for Method C (Manual) and Method B (Reader Confirmation) for all study sites.
Overall- Single-Operator agreements ranged from 92.7% to 99% for Method B (Reader Confirmation), and from 97% to 100% for Method C (Manual).
{41}------------------------------------------------
Instrument Precision ANCA Formalin
HELIOS (Method A) Instrument Agreement:
Table 71: Instrument Agreement_ANCA Formalin_Number of correctly/incorrectly found samples Method A
| Number of Samples | Site 1 | Site 2 | Site 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| HELIOS 1 | HELIOS 2 | HELIOS 3 | |||||||
| Positive Agreement | 240 | 0 | 240 | 221 | 19 | 240 | 235 | 5 | 240 |
| Negative Agreement | 56 | 4 | 60 | 46 | 14 | 60 | 37 | 23 | 60 |
| Overall Agreement | 296 | 4 | 300 | 267 | 33 | 300 | 272 | 28 | 300 |
| Pattern Agreement | 229 | 11 | 240 | 193 | 47 | 240 | 211 | 29 | 240 |
| FI Agreement | na | na | na | 0 | 0 | 0 | 0 | 0 | 0 |
Table 72: Instrument Agreement_ANCA Formalin. % Agreements (95% CI) Method A
| % Agreement (95% CI) | Site 1 | Site 2 | Site 3 |
|---|---|---|---|
| HELIOS 1 | HELIOS 2 | HELIOS 3 | |
| Positive Agreement | 100(98.4 - 100) | 92.1(88 - 94.9) | 97.9(95.2 - 99.1) |
| Negative Agreement | 93.3(84.1 - 97.4) | 76.7(64.6 - 85.6) | 61.7(49 - 72.9) |
| Overall Agreement | 98.7(96.6 - 99.5) | 89(85 - 92.1) | 90.7(86.8 - 93.5) |
| Pattern Agreement | 95.4(92 - 97.4) | 80.4(74.9 - 84.9) | 87.9(83.2 - 91.5) |
| FI Agreement | na | 0 | 0 |
For Instrument to Instrument Agreement with ANCA Formalin, all acceptance criteria were met, except for negative agreement at site 3. There, negative agreement was 61.7% whereas negative agreement at site 1 and was 93.3% and 76.7%. Sometimes it may occur, that HELIOS focusses falsly on unspecific, fluorescing particles, thereby reporting the sample as positive, while the reader can still see from the images that the cells are negative. Therefore, we clearly state that all results suggested by HELIOS have to be confirmed by a trained reader.
Overall- Instrument to Instrument agreements ranged from 89% to 90.7%.
{42}------------------------------------------------
Image /page/42/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image of a chromosome inside. To the right of the square is the company name in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font below.
Conclusions
ANCA Ethanol
Overall-Between-Lab agreements ranged from 92.5% to 96.5% for Method A (HELIOS), from 98.5% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
Overall- Between-Operator agreements ranged from 97.1% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
Overall- Single-Operator agreements ranged from 96.7% to 100% for Method B (Reader Confirmation), and was 100% for Method C (Manual).
Overall- Instrument to Instrument agreements ranged from 91.3% to 99.2%.
ANCA Formalin
Overall-Between-Lab agreements ranged from 89.8% to 94.7% for Method A (HELIOS), from 95.8% to 98.6% for Method B (Reader Confirmation), and from 98.2% to 99.4% for Method C (Manual).
Overall- Between-Operator agreements ranged from 93.5% to 99% for Method B (Reader Confirmation), and from 97.5% to 100% for Method C (Manual).
Overall- Single-Operator agreements ranged from 92.7% to 99% for Method B (Reader Confirmation), and from 97% to 100% for Method C (Manual).
Overall- Instrument to Instrument agreements ranged from 89% to 90.7%.
c) Lot to Lot Precision/Reproducibility Study AESKUSLIDES® ANCA A lot to lot reproducibility study was performed using three reagent lots each for AESKUSLIDES® ANCA Ethanol and ANCA Formalin. 12 sera were tested:
| SampleID | Result | Pattern ANCAEthanol | Grading | Result | Pattern ANCAFormalin | Grading |
|---|---|---|---|---|---|---|
| 1 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 2 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 3 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 4 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 5 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 6 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 7 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 8 | Pos | C-ANCA | low positive | Pos | C-ANCA | low positive |
| 9 | Pos | A-ANCA | high positive | Neg | Neg | negative |
| 10 | Pos | A-ANCA | medium positive | Neg | Neg | negative |
| 11 | Neg | Neg | negative | Neg | Neg | negative |
| 12 | Neg | Neg | negative | Neg | Neg | negative |
Table 73: Serum sample set for Lot to Lot Precision Study
{43}------------------------------------------------
Image /page/43/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font size and lighter shade of gray.
Each of the 12 serum samples was assayed 10 times on each reagent lot to give a total of 30 replicates per serum sample.
Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.
Agreements were calculated between the three reagent lots (lot 1 vs lo2 , lot 2 vs lot 3, lot 1 vs lot 3), for individual as well as for both readers combined.
Acceptance Criteria:
- Positive sera have to be found positive, and negative sera have to be found negative.
- -Correct pattern has to be found.
- -Reported FI is allowed to differ max. ± 1 level from the expected value.
- -Agreements should meet the following criteria:
| Type of Agreement | % Agreement |
|---|---|
| Positive Agreement | > 85% |
| Negative Agreement | > 85% |
| Overall Agreement | > 85% |
| Total Pattern Agreement | > 85% |
| Single Pattern Agreement P-ANCA | > 85% |
| Single Pattern Agreement C-ANCA | > 85% |
| Single Pattern % Agreement A-ANCA | > 85% |
| Fluorescence Intensity Agreement | > 85% |
Results:
Table 74a: Lot to Lot Precision AESKUSLIDES ANCA Ethanol, Lot to lot agreements – Lot 1 vs Lot 2:
| % Agreement (95% CI)Lot 1 vs Lot 2 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement(97.1 - 100) | 100(97.1 - 100) | 92.9(87.4 - 96.1) | 96.3(93.3 - 98) |
| Negative agreement(88.6 - 100) | 100(88.6 - 100) | 100(83.9 - 100) | 100(92.9 - 100) |
| Overall agreement(97.7 - 100) | 100(97.7 - 100) | 93.8(88.9 - 96.6) | 96.9(94.3 - 98.3) |
| Single Pattern C-ANCAagreement(92.9 - 100) | 100(92.9 - 100) | 100(92.9 - 100) | 100(96.3 - 100) |
| Single Pattern P-ANCAagreement(94 - 100) | 100(94 - 100) | 100(94 - 100) | 100(96.9 - 100) |
| Single Pattern A-ANCAagreement(83.9 - 100) | 100(83.9 - 100) | 100(83.9 - 100) | 100(91.2 - 100) |
| Total Pattern agreement(97.1 - 100) | 100(97.1 - 100) | 100(97.1 - 100) | 100(98.5 - 100) |
| Fl agreement(96.5 - 99.9) | 99.4(96.5 - 99.9) | 98.8(95.6 - 99.7) | 99.1(97.3 - 99.7) |
{44}------------------------------------------------
Table 74b: Lot to Lot Precision AESKUSLIDES ANCA Ethanol, Lot to lot agreements – Lot 2 vs Lot 3:
| % Agreement (95% CI)Lot 2 vs Lot 3 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement | 100(97.1 - 100) | 100(97.1 - 100) | 100(98.5 - 100) |
| Negative agreement | 100(88.6 - 100) | 100(88.6 - 100) | |
| Overall agreement | 100(97.7 - 100) | 100(97.7 - 100) | |
| Single Pattern C-ANCAagreement | 100(92.9 - 100) | 100(92.9 - 100) | |
| Single Pattern P-ANCAagreement | 100(94 - 100) | 100(94 - 100) | |
| Single Pattern A-ANCAagreement | 100(83.9 - 100) | 100(83.9 - 100) | |
| Total Pattern agreement | 100(97.1 - 100) | 100(97.1 - 100) | |
| Fl agreement | 98.1(94.6 - 99.4) | 97.5(93.7 - 99) |
FI = Fluorescence Intensity
Table 74c: Lot to Lot Precision AESKUSLIDES ANCA Ethanol, Lot to lot agreements – Lot 1 vs Lot 3:
| % Agreement (95% CI)Lot 1 vs Lot 3 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement | 100(97.1 - 100) | 92.9(87.4 -96.1) | 96.3(93.3 - 98) |
| Negative agreement | 100(88.6 - 100) | 100(83.9 - 100) | 100(92.9 - 100) |
| Overall agreement | 100(97.7 - 100) | 93.8(88.9 -96.6) | 96.9(94.3 - 98.3) |
| Single Pattern C-ANCAagreement | 100(92.9 - 100) | 100(92.9 - 100) | 100(96.3 - 100) |
| Single Pattern P-ANCAagreement | 100(94 - 100) | 100(94 - 100) | 100(96.9 - 100) |
| Single Pattern A-ANCAagreement | 100(83.9 - 100) | 100(83.9 - 100) | 100(91.2 - 100) |
| Total Pattern agreement | 100(97.1 - 100) | 100(97.1 - 100) | 100(98.5 - 100) |
| Fl agreement | 100(97.7 - 100) | 100(97.7 - 100) | 100(98.8 - 100) |
{45}------------------------------------------------
| % Agreement (95% CI) | Lot 1 vs Lot 2 | Lot 2 vs Lot 3 | Lot 1 vs Lot 3 |
|---|---|---|---|
| Positive agreement | 96.3(93.3 - 98) | 100(98.5 - 100) | 96.3(93.3 - 98) |
| Negative agreement | 100(92.9 - 100) | 100(94 - 100) | 100(92.9 - 100) |
| Overall agreement | 96.9(94.3 - 98.3) | 100(98.8 - 100) | 96.9(94.3 - 98.3) |
| Single Pattern C-ANCAagreement | 100(96.3 - 100) | 100(96.3 - 100) | 100(96.3 - 100) |
| Single Pattern P-ANCAagreement | 100(96.9 - 100) | 100(96.9 - 100) | 100(96.9 - 100) |
| Single Pattern A-ANCAagreement | 100(91.2 - 100) | 100(91.2 - 100) | 100(91.2 - 100) |
| Total Pattern agreement | 100(98.5 - 100) | 100(98.5 - 100) | 100(98.5 - 100) |
| FI agreement | 99.1(97.3 - 99.7) | 97.8(95.6 - 98.9) | 100(98.8 - 100) |
| Table 75: Overview Lot to lot agreements (combined readers): |
|---|
FI = Fluorescence Intensity
All agreements fulfilled the above criteria.
Positive agreement ranged from 96.3% to 100% for AESKUSLIDES ANCA Ethanol for both readers.
Negative agreement for ANCA Ethanol was 100% for both readers.
Overall agreement for ANCA Ethanol ranged from 96.9% to 100% for combined readers.
Pattern agreement for ANCA Ethanol was 100% for both readers, for all single pattern and total pattern agreement.
FI agreement for ANCA Ethanol ranged from 97.8% to 100% for combined readers.
{46}------------------------------------------------
Image /page/46/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol inside the green square appears to be a stylized representation of a medical or diagnostic tool.
Table 76a: Lot to Lot Precision AESKUSLIDES ANCA Formalin, Lot to lot agreements – Lot 1 vs Lot 2
| % Agreement (95% CI)Lot 1 vs Lot 2 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement | 100(96.6 - 100) | 99.1(94.9 - 99.8) | 99.5(97.4 - 99.9) |
| Negative agreement | 98(89.7 - 99.7) | 94.3(84.6 - 98.1) | 96.2(90.5 - 98.5) |
| Overall agreement | 99.4(96.5 - 99.9) | 97.5(93.7 - 99) | 98.4(96.4 - 99.3) |
| Single Pattern C-ANCA agreement | 100(96.3 - 100) | 100(96.2 - 100) | 100(98.1 - 100) |
| Total Pattern agreement | 100(96.6 - 100) | 100(96.5 - 100) | 100(98.2 - 100) |
| Fl agreement | 98.8(95.6 - 99.7) | 96.9(92.9 - 98.7) | 97.8(95.6 - 98.9) |
FI = Fluorescence Intensity
Table 76b: Lot to Lot Precision AESKUSLIDES ANCA Formalin, Lot to lot agreements – Lot 2 vs Lot 3:
| % Agreement (95% CI)Lot 2 vs Lot 3 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement | 100(96.6 - 100) | 100(96.6 - 100) | 100(98.3 - 100) |
| Negative agreement | 100(92.9 - 100) | 98(89.7 - 99.7) | 99(94.6 - 99.8) |
| Overall agreement | 100(97.7 - 100) | 99.4(96.5 - 99.9) | 99.7(98.3 - 99.9) |
| Single Pattern C-ANCAagreement | 100(96.3 - 100) | 100(96.3 - 100) | 100(98.1 - 100) |
| Total Pattern agreement | 100(96.6 - 100) | 100(96.6 - 100) | 100(98.3 - 100) |
| Fl agreement | 99.4(96.5 - 99.9) | 98.8(95.6 - 99.7) | 99.1(97.3 - 99.7) |
{47}------------------------------------------------
Table 76a: Lot to Lot Precision AESKUSLIDES ANCA Formalin, Lot to lot agreements – Lot 1 vs Lot 3:
| % Agreement (95% CI)Lot 1 vs Lot 3 | Reader 1 | Reader 2 | CombinedReaders |
|---|---|---|---|
| Positive agreement | 100(96.6 - 100) | 100(96.5 - 100) | 100(98.3 - 100) |
| Negative agreement | 98(89.7 - 99.7) | 94.3(84.6 - 98.1) | 96.2(90.5 - 98.5) |
| Overall agreement | 99.4(96.5 - 99.9) | 98.1(94.6 - 99.4) | 98.8(96.8 - 99.5) |
| Single Pattern C-ANCAagreement | 100(96.3 - 100) | 100(96.2 - 100) | 100(98.1 - 100) |
| Total Pattern agreement | 100(96.6 - 100) | 100(96.5 - 100) | 100(98.3 - 100) |
| Fl agreement | 98.1(94.6 - 99.4) | 96.9(92.9 - 98.7) | 97.5(95.1 - 98.7) |
FI = Fluorescence Intensity
Table 77: Overview Lot to Lot agreements (combined readers):
| % Agreement (95% CI) | Lot 1 vs Lot 2 | Lot 2 vs Lot 3 | Lot 1 vs Lot 3 |
|---|---|---|---|
| Positive agreement | 99.5(97.4 - 99.9) | 100(98.3 - 100) | 100(98.3 - 100) |
| Negative agreement | 96.2(90.5 - 98.5) | 99(94.6 - 99.8) | 96.2(90.5 - 98.5) |
| Overall agreement | 98.4(96.4 - 99.3) | 99.7(98.3 - 99.9) | 98.8(96.8 - 99.5) |
| Single Pattern C-ANCA agreement | 100(98.1 - 100) | 100(98.1 - 100) | 100(98.1 - 100) |
| Total Pattern agreement | 100(98.2 - 100) | 100(98.3 - 100) | 100(98.3 - 100) |
| FI agreement | 97.8(95.6 - 98.9) | 99.1(97.3 - 99.7) | 97.5(95.1 - 98.7) |
FI = Fluorescence Intensity
All agreements fulfilled the above criteria.
Positive agreement for AESKUSLIDES ANCA Formalin ranged from 99.5% to 100% for combined readers.
Negative agreement for ANCA Formalin ranged from 96.2% to 99.0% for both readers.
Overall agreement for ANCA Formalin ranged from 98.4% to 99.7% for combined readers.
Pattern agreement for ANCA Formalin was 100% for both readers, for all single pattern and total pattern agreement.
FI agreement for ANCA Formalin ranged from 97.8% to 99.1%% for combined readers.
{48}------------------------------------------------
Image /page/48/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green symbol on the left, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Conclusion:
Overall agreement for AESKUSLIDES ANCA Ethanol ranged from 96.9% to 100% and was acceptable. Overall agreement for AESKUSLIDES ANCA Formalin ranged from 98.4% to 99.7% and was acceptable.
Carry over:
To investigate and evaluate the carry over of samples from one reaction well to the following reaction well when slides are processed by HELIOS.
Three high positive samples were run on HELIOS in alternate with a negative serum sample. Results were confirmed by a reader. Tests were performed in accordance with the respective IFUs.
| Sample ID | Pos/Neg | AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | |||
|---|---|---|---|---|---|---|
| Pattern ANCA Ethanol | Grading | Pos/Neg | Pattern ANCA Formalin | Grading | ||
| 1 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 2 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 3 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 4 | Neg | Neg | negative | Neg | Neg | negative |
Table 78: Serum sample set for Carry Over study
Acceptance criteria:
- -Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
- -All patterns have to be found correctly.
- -FI is allowed to differ maximum + or - 1 from the expected value.
Results:
Table 79: Results Carry over for AESKUSLIDES ANCA
| SampleID | AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | ||||
|---|---|---|---|---|---|---|
| Result | Pattern | FI | Result | Pattern | FI | |
| 1 | Pos | P | 3 | Pos | C | 3 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
| 2 | Pos | C | 3 | Pos | C | 3 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
| 3 | Pos | P | 3 | Pos | C | 3 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
| 4 | Neg | Neg | 0 | Neg | Neg | 0 |
All samples fulfilled the above criteria. No carry over has been observed from well to well. All positive samples were identified as positive. All negative samples were identified as negative. All patterns have been identified correctly.
{49}------------------------------------------------
Image /page/49/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray.
Conclusions
No carry over could be observed.
Time Extension Study
To investigate the stability of AESKUSLIDES ANCA Ethanol and ANCA Formalin performance upon shortening and elongation of the recommended incubation times for samples and conjugate.
Procedure:
AESKUSLIDES ANCA Ethanol and ANCA Formalin was assayed with 10 samples on HELIOS with the following incubation times:
-
- 15 min sera incubation + 15min incubation of conjugate
-
- 30 min sera incubation + 30min incubation of conjugate
-
- 45 min sera incubation + 45min incubation of conjugate
Tabel 80: Samples used for Time Extension
| SampleID | Pos/Neg | AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | |||
|---|---|---|---|---|---|---|
| PatternANCAEthanol | Grading | Pos/Neg | PatternANCAFormalin | Grading | ||
| 1 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 2 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 3 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 4 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 5 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 6 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 7 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 8 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 9 | Neg | Neg | negative | Neg | Neg | negative |
| 10 | Neg | Neg | negative | Neg | Neg | negative |
Tests were performed in accordance with the respective IFUs. Results have been confirmed by a trained reader.
Results of the 30 min protocol will be compared to the results of the 15 min and 45 min protocol.
Acceptance criteria:
- -Pos/Neg/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
- -FI is allowed to differ maximum + or - 1 from the expected value.
{50}------------------------------------------------
Image /page/50/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image of a zipper inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font below.
AESKUSLIDES ANCA 510(k) Traditional Submission
Results
| IncubationTimes | 2 x 15 Minutes | 2 x 30 Minutes | 2 x 45 Minutes | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample ID | Result | Pattern | FI | Result | Pattern | FI | Result | Pattern | FI |
| PC-C | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| PC-P | Pos | P | 3 | Pos | P | 3 | Pos | P | 3 |
| NC | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
| Sample 1 | Pos | P | 3 | Pos | P | 3 | Pos | P | 3 |
| Sample 2 | Pos | P | 2 | Pos | P | 3 | Pos | P | 2 |
| Sample 3 | Pos | P | 3 | Pos | P | 3 | Pos | P | 3 |
| Sample 4 | Pos | P | 3 | Pos | P | 3 | Pos | P | 3 |
| Sample 5 | Pos | P | 2 | Pos | P | 3 | Pos | P | 3 |
| Sample 6 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 7 | Pos | C | 2 | Pos | C | 2 | Pos | C | 2 |
| Sample 8 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 9 | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
| Sample 10 | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
Table 81: Results Time extension study AESKUSLIDES ANCA Ethanol
FI = Fluorescence Intensity
| IncubationTimes | 2 x 15 Minutes | 2 x 30 Minutes | 2 x 45 Minutes | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample ID | Result | Pattern | FI | Result | Pattern | FI | Result | Pattern | FI |
| PC-C | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| PC-P | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| NC | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
| Sample 1 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 2 | Pos | C | 2 | Pos | C | 3 | Pos | C | 2 |
| Sample 3 | Pos | C | 3 | Pos | C | 2 | Pos | C | 3 |
| Sample 4 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 5 | Pos | C | 2 | Pos | C | 3 | Pos | C | 2 |
| Sample 6 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 7 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 8 | Pos | C | 3 | Pos | C | 3 | Pos | C | 3 |
| Sample 9 | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
| Sample 10 | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg | Neg |
Table 82: Results Time extension study AESKUSLIDES ANCA Formalin
FI = Fluorescence Intensity
All acceptance criteria were fulfilled in each assay. All positive samples were identified as positive. All negative samples were identified as negative. All patterns have been found as expected. Fluorescence intensities did not deviate more than +/-1 level.
{51}------------------------------------------------
Image /page/51/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol inside the green square appears to be a stylized representation of a medical or diagnostic tool.
Conclusions
The results demonstrate stability of the AESKUSLIDES ANCA Ethanol, as well as AESKUSLIDES ANCA Formalin during the tested incubation times of sample and conjugate.
Detection limit
Not applicable
Analytical specificity
Interfering substances:
The interference study was performed according to CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Interference by 10 different substances was assessed by testing 14 serum samples.
Table 83: Samples used for Interfering substances study
| AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | |||||
|---|---|---|---|---|---|---|
| SampleID | Pos/Neg | PatternANCAEthanol | Grading | Pos/Neg | Pattern ANCAFormalin | Grading |
| 1 | Pos | P-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 2 | Pos | P-ANCA | high positive | Pos | C-ANCA | high positive |
| 3 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 4 | Pos | C-ANCA | low positive | Pos | C-ANCA | high positive |
| 5 | Pos | A-ANCA | medium positive | Neg | Neg | negative |
| 6 | Pos | A-ANCA | low positive | Neg | Neg | negative |
| 7 | Neg | Neg | negative | Neg | Neg | negative |
| 8 | Neg | Neg | negative | Neg | Neg | negative |
| 9 | Pos | P-ANCA | low positive | Pos | C-ANCA | low positive |
| 10 | Pos | P-ANCA | low positive | Pos | C-ANCA | low positive |
| 11 | Pos | P-ANCA | low positive | Pos | C-ANCA | low positive |
| 12 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 13 | Pos | C-ANCA | low positive | Pos | C-ANCA | low positive |
| 14 | Pos | C-ANCA | low positive | Pos | C-ANCA | low positive |
{52}------------------------------------------------
Image /page/52/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image inside the square looks like a stylized letter "Y" with horizontal lines across it. To the right of the square, the text "AESKU.DIAGNOSTICS" is written in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" written in a smaller, lighter gray font below.
The following substances have been tested with the indicated concentrations:
| Table 84: Concentrations of Interfering substances | ||
|---|---|---|
| Interfering Substance | Minimum finalConcentrationtested | Maximum finalConcentrationtested |
| Bilirubin conjugated | 0.1 mg/mL | 0.4 mg/mL |
| Bilirubin unconjugated | 0.1 mg/mL | 0.4 mg/mL |
| Hemoglobin | 2.5 mg/mL | 5.0 mg/mL |
| Triglycerides | 5 mg/mL | 20 mg/mL |
| RF IgM | 200 IU/mL | 400 IU/mL |
| Rituximab | 0.5 mg/mL | 2.0 mg/mL |
| Methylprednisolone | 0.2 mg/mL | 0.8 mg/mL |
| Cyclophosphamide | 1.0 mg/mL | 4.0 mg/mL |
| Methotrexate | 0.025 mg/mL | 0.1 mg/mL |
| Azathioprine | 0.0075 mg/mL | 0.03 mg/mL |
Table 84: Concentrations of Interfering substances
Interfering substances were spiked in each of the 14 samples in two different concentrations. Controls were prepared for each serum sample by spiking in only the respective amount of diluent without interfering substances. Spiked serum samples and their controls have been tested in triplicates for each concentration of interferent with AESKUSLIDES ANCA Ethanol and ANCA Formalin. All tests have been performed manually according to the IFUs. Results have been analyzed by two independent readers at the microscope.
The results (positive/negative classification, pattern, fluorescence intensity (FI)) of the spiked samples were compared to the results of the respective controls. Agreements have been calculated by dividing the number of correctly found results through the number of total results for spiked samples.
Acceptance criteria:
- -Pos/Neq/Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative.
- -Overall Agreement: > 90%.
- -All patterns have to be found correctly. Pattern Agreement: > 90%
- -FI is allowed to differ maximum + or - 1 from the expected value. FI Agreement: > 90%.
{53}------------------------------------------------
Image /page/53/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white graphic inside on the left. To the right of the graphic is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger than "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Results
All samples fulfilled the above criteria. No significant differences were observed between sera spiked with different concentrations of the interferents and their controls:
ANCA Ethanol:
-
Positive Agreement ranged from 97% to 100% for all tested interferents and ● concentrations. It was 100% for all tested substances and concentrations, except for 5 mg/ml
Triglycerides with Reader 2 (97%), 200 IU/ml RF with both readers (99%), and 1 mg/ml Cyclophosphamide with both readers (99%). -
Negative Agreement was 100% for all tested interferents and concentrations.
-
Overall Agreement ranged from 98% to 100% for all tested interferents and concentrations. It was 100% for all tested substances and concentrations, except for 5 mg/ml Triglycerides with Reader 2 (98%), 200 IU/ml RF with both readers (99%), and 1 mg/ml Cyclophosphamide with both readers (99%).
-
. Pattern Agreement ranged from 97% to 100% for all tested interferents and concentrations. It was 100% for all tested substances and concentrations, except for 5 mg/ml Triglycerides with Reader 2 (97%), 200 IU/ml RF with both readers (99%), 1 mq/ml Cyclophosphamide with both readers (99%), and 0.0075 mg/ml Azathioprine with both readers (99%).
-
. FI agreement ranged from 98% to 100% for all tested interferents and concentrations. It was 100% for all tested substances and concentrations, except for 5 mg/ml Triglycerides with Reader 2 (98%), and 1 mq/ml Cyclophosphamide with both readers (99%).
ANCA Formalin:
- . Positive Agreement was 100% for all tested interferents and concentrations.
- Negative Agreement ranged from 92% to 100% for all tested interferents and . concentrations. It was 100% for all tested substances and concentrations, except for 0.025 mq/ml Methotrexate with Reader 1 (92%) and with Reader 2 (96%).
- Overall Agreement ranged from 98% to 100% for all tested interferents and . concentrations. It was 100% for all tested substances and concentrations, except for 0.025 mg/ml Methotrexate with Reader 1 (98%) and with Reader 2 (99%).
- Pattern Agreement was 100% for all tested interferents and concentrations. ●
{54}------------------------------------------------
- FI Agreement was 100% for all tested interferents and concentrations. ●
Conclusions
All criteria are fulfilled in each test. No interference was detected with the tested substances, up to the maximal concentrations indicated above.
Assay cut-off:
AESKU recommends a screening dilution of 1:20. followed by serial dilutions for semiquantitative determinations. but suggests each laboratory establish its own screening dilution and titration scheme based on its population. The titers of 1:10 and 1:20 are considered low titers. 1:40 and 1:80 are considered medium titers. and 1:160 and greater are considered high titers.
Stability studies
Accelerated Stability Report
3 lots of complete kits (including slides, controls, conjugate, & sample buffer) of AESKUSLIDES ANCA Ethanol and Formalin were stored at 37°C for 6 weeks. At t=0, 1 week, 2 weeks, 3 weeks, 4 weeks, 5 weeks, and 6 weeks, a kit was manually tested on the same set of 8 samples:
| Sample ID | AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | ||||
|---|---|---|---|---|---|---|
| Pos/Neg | Pattern ANCA Ethanol | Grading | Pos/Neg | Pattern ANCA Formalin | Grading | |
| 1 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 2 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 3 | Pos | C-ANCA | low positive | Pos | C-ANCA | low positive |
| 4 | Pos | P-ANCA | high positive | Pos | C-ANCA | medium positive |
| 5 | Pos | P-ANCA | medium positive | Pos | C-ANCA | borderline positive |
| 6 | Pos | P-ANCA | low positive | Pos | Neg | Neg |
| 7 | Pos | A-ANCA | medium positive | Neg | Neg | Neg |
| 8 | Neg | Neg | Neg | Neg | Neg | Neg |
Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.
The stability data is covering the stability of all components of the kit. Agreements were calculated for individual and combined readers. Results of the kit controls were not included in the agreement calculations.
{55}------------------------------------------------
Image /page/55/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo appears to be for a company that specializes in diagnostic tools.
Acceptance Criteria:
- . Positive sera have to be found positive, and negative sera have to be found negative.
- Correct pattern has to be found. ●
- Reported FI is allowed to differ max. ± 1 level from the expected value. .
- Agreements should meet the following criteria: ●
| Type of Agreement | % Agreement |
|---|---|
| Positive Agreement | > 85% |
| Negative Agreement | > 85% |
| Overall Agreement | > 85% |
| Total Pattern Agreement | > 85% |
| Fluorescence Intensity Agreement | > 85% |
{56}------------------------------------------------
Image /page/56/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo is simple and professional, conveying a sense of trust and reliability.
AESKUSLIDES ANCA 510(k) Traditional Submission
Results:
Table 85: Accelerated Stability AESKUSLIDES ANCA Ethanol – Number of Samples:
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 3 / 98 | 5 / 98 | 8 / 196 | 9 / 98 | 12 / 98 | 21 / 196 | 2 / 98 | 5 / 98 | 7 / 196 |
| Negative Agreement | 1 / 14 | 1 / 14 | 2 / 28 | 0 / 14 | 0 / 14 | 0 / 28 | 0 / 14 | 0 / 14 | 0 / 28 |
| Overall Agreement | 4 / 112 | 6 / 112 | 10 / 224 | 9 / 112 | 12 / 112 | 21 / 224 | 2 / 112 | 5 / 112 | 7 / 224 |
| Pattern Agreement | 3 / 98 | 5 / 98 | 8 / 196 | 11 / 98 | 15 / 98 | 26 / 196 | 2 / 98 | 8 / 98 | 10 / 196 |
| FI Agreement | 3 / 112 | 4 / 112 | 7 / 224 | 4 / 112 | 6 / 112 | 10 / 224 | 2 / 112 | 2 / 112 | 4 / 224 |
FI = Fluorescence Intensity
Table 86: Accelerated Stability AESKUSLIDES ANCA Ethanol – % Agreement:
| Lot 1 | Lot 2 | Lot 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| % Agreement(95% CI) | Reader 1 | Reader 2 | both Readers | Reader 1 | Reader 2 | both Readers | Reader 1 | Reader 2 | both Readers |
| Positive Agreement | 96.9(91.4 - 99) | 94.9(88.6 - 97.8) | 95.9(92.2 - 97.9) | 90.8(83.5 - 95.1) | 87.8(79.8 - 92.9) | 89.3(84.2 - 92.9) | 98(92.9 - 99.4) | 94.9(88.6 - 97.8) | 96.4(92.8 - 98.3) |
| Negative Agreement | 92.9(68.5 - 98.7) | 92.9(68.5 - 98.7) | 92.9(77.4 - 98) | 100(78.5 - 100) | 100(78.5 - 100) | 100(87.9 - 100) | 100(78.5 - 100) | 100(78.5 - 100) | 100(87.9 - 100) |
| Overall Agreement | 96.4(91.2 - 98.6) | 94.6(88.8 - 97.5) | 95.5(92 - 97.6) | 92(85.4 - 95.7) | 89.3(82.2 - 93.8) | 90.6(86.1 - 93.8) | 98.2(93.7 - 99.5) | 95.5(90 - 98.1) | 96.9(93.7 - 98.5) |
| Pattern Agreement | 96.9(91.4 - 99) | 94.9(88.6 - 97.8) | 95.9(92.2 - 97.9) | 88.8(81 - 93.6) | 84.7(76.3 - 90.5) | 86.7(81.3 - 90.8) | 98(92.9 - 99.4) | 91.8(84.7 - 95.8) | 94.9(90.9 - 97.2) |
| FI Agreement | 97.3(92.4 - 99.1) | 96.4(91.2 - 98.6) | 96.9(93.7 - 98.5) | 96.4(91.2 - 98.6) | 94.6(88.8 - 97.5) | 95.5(92 - 97.6) | 98.2(93.7 - 99.5) | 98.2(93.7 - 99.5) | 98.2(95.5 - 99.3) |
{57}------------------------------------------------
Image /page/57/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image inside the square looks like a vest with lines going across it. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Agreement calculations for AESKUSLIDES ANCA Ethanol included all data points of the whole test time span of 6 weeks at 37°C:
All agreements fulfilled the above criteria.
Positive agreement for AESKUSLIDES ANCA Ethanol for both readers ranged from 89.3% to 96.4% for all 3 reagent lots.
Negative agreement for ANCA Ethanol for both readers ranged from 92.9% to 100% for all 3 reagent lots.
Overall agreement for ANCA Ethanol for both readers ranged from 90.6% to 96.9% for all 3 reagent lots.
Pattern agreement for ANCA Ethanol for both readers ranged from 86.7% to 95.9% for all 3 reagent lots.
FI agreement for ANCA Ethanol for both readers ranged from 95.5% to 98.2% for all 3 reagent lots.
No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.
{58}------------------------------------------------
Image /page/58/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green symbol on the left, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The overall design is clean and professional, conveying a sense of expertise in diagnostics.
AESKUSLIDES ANCA 510(k) Traditional Submission
Table 87: Accelerated Stability AESKUSLIDES ANCA Formalin - Number of Samples:
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 4 / 50 | 14 / 50 | 18 / 100 | 4 / 50 | 12 / 50 | 16 / 100 | 5 / 50 | 16 / 50 | 21 / 100 |
| Negative Agreement | 0 / 30 | 0 / 30 | 0 / 60 | 0 / 30 | 0 / 30 | 0 / 60 | 0 / 30 | 0 / 30 | 0 / 60 |
| Overall Agreement | 4 / 80 | 14 / 80 | 18 / 160 | 4 / 80 | 12 / 80 | 16 / 160 | 5 / 80 | 16 / 80 | 21 / 160 |
| Pattern Agreement | 4 / 50 | 14 / 50 | 18 / 100 | 4 / 50 | 12 / 50 | 16 / 100 | 5 / 50 | 16 / 50 | 21 / 100 |
| FI Agreement | 0 / 80 | 3 / 80 | 3 / 160 | 0 / 80 | 2 / 80 | 2 / 160 | 1 / 80 | 4 / 80 | 5 / 160 |
FI = Fluorescence Intensity
Table 88: Accelerated Stability AESKUSLIDES ANCA Formalin – % Agreement:
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 92(81.2 - 96.8) | 72(58.3 - 82.5) | 82(73.3 - 88.3) | 92(81.2 - 96.8) | 76(62.6 - 85.7) | 84(75.6 - 89.9) | 90(78.6 - 95.7) | 68(54.2 - 79.2) | 79(70 - 85.8) |
| Negative Agreement | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) |
| Overall Agreement | 95(87.8 - 98) | 82.5(72.7 - 89.3) | 88.8(82.9 - 92.8) | 95(87.8 - 98) | 85(75.6 - 91.2) | 90(84.4 - 93.8) | 93.8(86.2 - 97.3) | 80(70 - 87.3) | 86.9(80.8 - 91.3) |
| Pattern Agreement | 92(81.2 - 96.8) | 72(58.3 - 82.5) | 82(73.3 - 88.3) | 92(81.2 - 96.8) | 76(62.6 - 85.7) | 84(75.6 - 89.9) | 90(78.6 - 95.7) | 68(54.2 - 79.2) | 79(70 - 85.8) |
| FI Agreement | 100(95.4 - 100) | 96.3(89.5 - 98.7) | 98.1(94.6 - 99.4) | 100(95.4 - 100) | 97.5(91.3 - 99.3) | 98.8(95.6 - 99.7) | 98.8(93.3 - 99.8) | 95(87.8 - 98) | 96.9(92.9 - 98.7) |
{59}------------------------------------------------
Image /page/59/Picture/0 description: The image contains a logo for AESKU.DIAGNOSTICS. The logo features a green symbol on the left, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The overall design is clean and professional, suggesting a company in the medical or diagnostic field.
AESKUSLIDES ANCA 510(k) Traditional Submission
Table 89: Accelerated Stability AESKUSLIDES ANCA Formalin – Number of Samples, borderline positive samples excluded:
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 0 / 40 | 5 / 40 | 5 / 80 | 0 / 40 | 2 / 40 | 2 / 80 | 1 / 40 | 6 / 40 | 7 / 80 |
| Negative Agreement | 0 / 30 | 0 / 30 | 0 / 60 | 0 / 30 | 0 / 30 | 0 / 60 | 0 / 30 | 0 / 30 | 0 / 60 |
| Overall Agreement | 0 / 70 | 5 / 70 | 5 / 140 | 0 / 70 | 2 / 70 | 2 / 140 | 1 / 70 | 6 / 70 | 7 / 140 |
| Pattern Agreement | 0 / 40 | 5 / 40 | 5 / 80 | 0 / 40 | 2 / 40 | 2 / 80 | 1 / 40 | 6 / 40 | 7 / 80 |
| FI Agreement | 0 / 70 | 3 / 70 | 3 / 140 | 0 / 70 | 2 / 70 | 2 / 140 | 1 / 70 | 4 / 70 | 5 / 140 |
FI = Fluorescence Intensity
Table 90: Accelerated Stability AESKUSLIDES ANCA Formalin – % Agreement, borderline positive samples excluded:
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 100(91.2 - 100) | 87.5(73.9 - 94.5) | 93.8(86.2 - 97.3) | 100(91.2 - 100) | 95(83.5 - 98.6) | 97.5(91.3 - 99.3) | 97.5(87.1 - 99.6) | 85(70.9 - 92.9) | 91.3(83 - 95.7) |
| Negative Agreement | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) | 100(88.6 - 100) | 100(88.6 - 100) | 100(94 - 100) |
| Overall Agreement | 100(94.8 - 100) | 92.9(84.3 - 96.9) | 96.4(91.9 - 98.5) | 100(94.8 - 100) | 97.1(90.2 - 99.2) | 98.6(94.9 - 99.6) | 98.6(92.3 - 99.7) | 91.4(82.5 - 96) | 95(90 - 97.6) |
| Pattern Agreement | 100(91.2 - 100) | 87.5(73.9 - 94.5) | 93.8(86.2 - 97.3) | 100(91.2 - 100) | 95(83.5 - 98.6) | 97.5(91.3 - 99.3) | 97.5(87.1 - 99.6) | 85(70.9 - 92.9) | 91.3(83 - 95.7) |
| FI Agreement | 100(95.4 - 100) | 96.3(89.5 - 98.7) | 97.9(93.9 - 99.3) | 100(95.4 - 100) | 97.5(91.3 - 99.3) | 98.6(94.9 - 99.6) | 98.8(93.3 - 99.8) | 95(87.8 - 98) | 96.4(91.9 - 98.5) |
{60}------------------------------------------------
Image /page/60/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside of what appears to be a vest. To the right of the image is the company name in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below.
Agreement calculations for AESKUSLIDES ANCA Formalin included all data points of a test time span of 4 weeks at 37°C:
The sample set for AESKUSLIDES ANCA Formalin included one sample (sample 5) that was borderline positive. For this sample we expect either a low positive result (Fl=1) or a negative result (Fl=0). Therefore, this sample leads to many false negatives and therefore has a big impact on positive agreement and on pattern agreement which is not due to stability issues of the assay.
Therefore, we did an alternative agreement calculation excluding this borderline positive sample. In this calculation all acceptance criteria have been met:
Positive agreement for AESKUSLIDES ANCA Formalin for both readers ranged from 91.3% to 97.5% for all 3 reagent lots.
Negative agreement for ANCA Formalin for both readers was 100% for all 3 reagent lots.
Overall agreement for ANCA Formalin for both readers ranged from 95% to 98.6% for all 3 reagent lots.
Pattern agreement for ANCA Formalin for both readers ranged from 91.3% to 97.5% for all 3 reagent lots.
FI agreement for ANCA Formalin for both readers ranged from 96.4% to 98.6% for all 3 reagent lots.
No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.
Conclusion:
For AESKUSLIDES ANCA Ethanol all acceptance criteria were met for a time span of 6 weeks at 37°C. Therefore, according to Arrhenius equation, we claim a shelf life of at least 24 months (24+3 months) at 2-8°C.
For AESKUSLIDES ANCA Formalin all acceptance criteria were met for a time span of 4 weeks at 37°C. Therefore, according to Arrhenius equation, we claim a shelf life of at least 18 months at 2-8°C.
{61}------------------------------------------------
Image /page/61/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font size and lighter shade of gray. The overall design is clean and professional, conveying a sense of trust and reliability.
Real Time Stability Report
Shelf life of AESKUSLIDES ANCA Ethanol has been claimed to be at least 24 months at 2-8°C storage temperature based on the results of the Accelerated Stability Study of AESKUSLIDES ANCA Ethanol. Shelf life of AESKUSLIDES ANCA Formalin has been claimed to be at least 18 months at 2-8°C storage temperature based on the results of the Accelerated Stability Study of AESKUSLIDES ANCA Formalin. To confirm this, a real time / long term stability study over a time period of 27 months is conducted.
Procedure:
Three different reagent lots of AESKUSLIDES ANCA Ethanol and ANCA Formalin will be stored at 2-8°C over a time period of 27 months. The same eight sera will be assayed at each of the following time points:
- Starting point of the stability study to
- 3 months after to t
- 6 months after to t2
- 12 months after to t3
- 18 months after to t4
- 24 months after to t5
- 27 months after to te
Sample description:
| AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | |||||
|---|---|---|---|---|---|---|
| Sample ID | ExpectedResult | ExpectedPattern | Grading | ExpectedResult | ExpectedPattern | Grading |
| Sample 1 | Pos | C | high positive | Pos | C | high positive |
| Sample 2 | Pos | C | medium positive | Pos | C | medium positive |
| Sample 3 | Pos | C | low positve | Pos | C | low posiitve |
| Sample 4 | Pos | P | high positive | Pos | C | medium positive |
| Sample 5 | Pos | P | medium positive | Pos | C | low posiitve |
| Sample 6 | Pos | P | low positve | Neg | Neg | negative |
| Sample 7 | Pos | A | medium positive | Neg | Neg | negative |
| Sample 8 | Neg | Neg | negative | Neg | Neg | negative |
Each serum is tested in triplicates in each run.
At time point to three runs will be performed for each of the three different kit lots to give a total of 9 repetitions per serum per lot.
At the following time points (t, to ts) only two runs will be performed for each of the three different kit lots to give a total of 6 repetitions per serum per lot.
All tests are performed manually according to the IFU. Results are analyzed by two independent readers at the microscope.
{62}------------------------------------------------
Image /page/62/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image of a vest inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below.
The results (pos/neg, pattern, Fl) at each time point and for each sample are compared to the respective results, determined at start of the stability study (to) and will be displayed as % agreements (pos/neq/overall agreement, pattern agreement. Fl agreement).
The results of both readers are combined for agreement calculations. Results of the kit controls are not included in the agreement calculations.
The stability data is covering the stability of all (undiluted) components of the kit. For each test time point only the minimal needed amount of concentrated wash buffer will be removed from the original bottle and diluted to give 1x wash buffer.
Acceptance criteria:
Positive/Neqative /Overall Agreement: All positive sera have to be found positive and all negative sera have to be found negative. Agreement is calculated from the number of correctly and incorrectly found samples, for positive samples (positive agreement), negative samples (negative agreement) and total samples (overall agreement).
Positive/Negative/Overall Agreement > 85%.
Patterns have to be found correctly.
Pattern Agreement > 85%.
FI is allowed to differ maximum + or - 1 from the expected value. Agreement is calculated from the number of correctly and incorrectly found FI values (of all tested samples).
FI Agreement > 85%.
Interpretation for Real time stability:
Acceptance criteria have to be fulfilled over the whole time period.
Results:
At the time of preparation of this report, only the results for time points to and to have been acquired so far, covering a time period of 3 months at 2-8°C.
Real time stability study is still ongoing. Further time points to be tested are 6, 12, 18, 24, and 27 months.
The results of the In Use Stability Study of AESKUSLIDES ANCA Ethanol and ANCA Formalin described elsewhere, can also serve as a real time stability study, covering a time period of 6 weeks at 2-8°C. In this In Use study the same sera have been used as for the herein described real time / long term stability study.
{63}------------------------------------------------
Image /page/63/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo is clean and professional, with a focus on the company's name and its diagnostic capabilities.
AESKUSLIDES ANCA 510(k) Traditional Submission
| Number of Samples | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 101 | 4 | 105 | 105 | 0 | 105 | 105 | 0 | 105 |
| Negative Agreement | 15 | 0 | 15 | 15 | 0 | 15 | 15 | 0 | 15 |
| Overall Agreement | 116 | 4 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
| Pattern Agreement | 101 | 4 | 105 | 105 | 0 | 105 | 104 | 1 | 105 |
| FI Agreement | 119 | 1 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
Table 91: Real time Stability AESKUSLIDES ANCA Ethanol (Reader 1) - Number of Samples:
FI = Fluorescence Intensity
Table 92: Real time Stability AESKUSLIDES ANCA Ethanol (Reader 2) – Number of Samples:
| Number of Samples | Reader 2 | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | ||||||||
| FoundCorrectly | Found Incorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | ||
| Positive Agreement | 98 | 7 | 105 | 105 | 0 | 105 | 101 | 4 | 105 | |
| Negative Agreement | 14 | 1 | 15 | 15 | 0 | 15 | 15 | 0 | 15 | |
| Overall Agreement | 112 | 8 | 120 | 120 | 0 | 120 | 116 | 4 | 120 | |
| Pattern Agreement | 98 | 7 | 105 | 105 | 0 | 105 | 100 | 5 | 105 | |
| FI Agreement | 119 | 1 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
{64}------------------------------------------------
Image /page/64/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
AESKUSLIDES ANCA 510(k) Traditional Submission
Table 93: Real time Stability AESKUSLIDES ANCA Ethanol (Combined Readers) – Number of Samples:
| Number of Samples | Combined Readers | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |||||||
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Positive Agreement | 199 | 11 | 210 | 210 | 0 | 210 | 206 | 4 | 210 |
| Negative Agreement | 29 | 1 | 30 | 30 | 0 | 30 | 30 | 0 | 30 |
| Overall Agreement | 228 | 12 | 240 | 240 | 0 | 240 | 236 | 4 | 240 |
| Pattern Agreement | 199 | 11 | 210 | 210 | 0 | 210 | 204 | 6 | 210 |
| FI Agreement | 238 | 2 | 240 | 240 | 0 | 240 | 240 | 0 | 240 |
{65}------------------------------------------------
Image /page/65/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 |
|---|---|---|---|
| Positive Agreement | 96.2(90.6 - 98.5) | 100(96.5 - 100) | 100(96.5 - 100) |
| Negative Agreement | 100(79.6 - 100) | 100(79.6 - 100) | 100(79.6 - 100) |
| Overall Agreement | 96.7(91.7 - 98.7) | 100(96.9 - 100) | 100(96.9 - 100) |
| Pattern Agreement | 96.2(90.6 - 98.5) | 100(96.5 - 100) | 99(94.8 - 99.8) |
| FI Agreement | 99.2(95.4 - 99.9) | 100(96.9 - 100) | 100(96.9 - 100) |
Table 94: Real time Stability AESKUSLIDES ANCA Ethanol (Reader 1) – %Agreement:
FI = Fluorescence Intensity
Table 95: Real time Stability AESKUSLIDES ANCA Ethanol (Reader 2) – %Agreement:
| % Agreement(95% CI) | Reader 2 | ||
|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |
| Positive Agreement | 93.3(86.9 - 96.7) | 100(96.5 - 100) | 96.2(90.6 - 98.5) |
| Negative Agreement | 93.3(70.2 - 98.8) | 100(79.6 - 100) | 100(79.6 - 100) |
| Overall Agreement | 93.3(87.4 - 96.6) | 100(96.9 - 100) | 96.7(91.7 - 98.7) |
| Pattern Agreement | 93.3(86.9 - 96.7) | 100(96.5 - 100) | 95.2(89.3 - 97.9) |
| FI Agreement | 99.2(95.4 - 99.9) | 100(96.9 - 100) | 100(96.9 - 100) |
FI = Fluorescence Intensity
Table 96: Real time Stability AESKUSLIDES ANCA Ethanol (Combined Readers) – %Agreement:
| % Agreement(95% CI) | Combined Readers | ||
|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |
| Positive Agreement | 94.8(90.9 - 97.1) | 100(98.2 - 100) | 98.1(95.2 - 99.3) |
| Negative Agreement | 96.7(83.3 - 99.4) | 100(88.6 - 100) | 100(88.6 - 100) |
| Overall Agreement | 95(91.5 - 97.1) | 100(98.4 - 100) | 98.3(95.8 - 99.4) |
| Pattern Agreement | 94.8(90.9 - 97.1) | 100(98.2 - 100) | 97.1(93.9 - 98.7) |
| FI Agreement | 99.2(97 - 99.8) | 100(98.4 - 100) | 100(98.4 - 100) |
{66}------------------------------------------------
Image /page/66/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol inside the green square appears to be a stylized representation of a medical or diagnostic tool.
Agreement calculations for AESKUSLIDES ANCA Ethanol included all data points of the test over a time period of 3 months at 2-8°C:
All agreements fulfilled the above criteria.
Positive agreement for AESKUSLIDES ANCA Ethanol for both readers ranged from 94.8% to 100% for all 3 reagent lots.
Negative agreement for ANCA Ethanol for both readers ranged from 96.7% to 100% for all 3 reagent lots.
Overall agreement for ANCA Ethanol for both readers ranged from 95.0% to 100% for all 3 reagent lots.
Pattern agreement for ANCA Ethanol for both readers ranged from 94.8% to 100% for all 3 reagent lots.
FI agreement for ANCA Ethanol for both readers ranged from 99.2% to 100% for all 3 reagent lots.
Just one deviation of fluorescence intensity greater than +/-1 from the initial value has been observed for each reader.
{67}------------------------------------------------
Image /page/67/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image inside the square looks like a vest with a zipper. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font.
AESKUSLIDES ANCA 510(k) Traditional Submission
| Number of Samples | Reader 1 | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |||||||
| Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | Found Correctly | Found Incorrectly | Expected Numbers | |
| Positive Agreement | 72 | 3 | 75 | 71 | 4 | 75 | 67 | 8 | 75 |
| Negative Agreement | 45 | 0 | 45 | 45 | 0 | 45 | 44 | 1 | 45 |
| Overall Agreement | 117 | 3 | 120 | 116 | 4 | 120 | 111 | 9 | 120 |
| Pattern Agreement | 72 | 3 | 75 | 71 | 4 | 75 | 67 | 8 | 75 |
| FI Agreement | 120 | 0 | 120 | 120 | 0 | 120 | 120 | 0 | 120 |
Table 97: Real time Stability AESKUSLIDES ANCA Formalin (Reader 1) – Number of Samples:
FI = Fluorescence Intensity
Table 98: Real time Stability AESKUSLIDES ANCA Formalin (Reader 1) – Number of Samples:
| Number of Samples | Reader 2 | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |||||||
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Positive Agreement | 64 | 11 | 75 | 64 | 11 | 75 | 65 | 10 | 75 |
| Negative Agreement | 45 | 0 | 45 | 45 | 0 | 45 | 44 | 1 | 45 |
| Overall Agreement | 109 | 11 | 120 | 109 | 11 | 120 | 109 | 11 | 120 |
| Pattern Agreement | 64 | 11 | 75 | 64 | 11 | 75 | 65 | 10 | 75 |
| FI Agreement | 120 | 0 | 120 | 119 | 1 | 120 | 120 | 0 | 120 |
{68}------------------------------------------------
Image /page/68/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" also in gray.
AESKUSLIDES ANCA 510(k) Traditional Submission
Table 99: Real time Stability AESKUSLIDES ANCA Formalin (Combined Readers) – Number of Samples:
| Number of Samples | Combined Readers | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | Lot 2 | Lot 3 | |||||||
| FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | FoundCorrectly | FoundIncorrectly | ExpectedNumbers | |
| Positive Agreement | 136 | 14 | 150 | 135 | 15 | 150 | 132 | 18 | 150 |
| Negative Agreement | 90 | 0 | 90 | 90 | 0 | 90 | 88 | 2 | 90 |
| Overall Agreement | 226 | 14 | 240 | 225 | 15 | 240 | 220 | 20 | 240 |
| Pattern Agreement | 136 | 14 | 150 | 135 | 15 | 150 | 132 | 18 | 150 |
| FI Agreement | 240 | 0 | 240 | 239 | 1 | 240 | 240 | 0 | 240 |
{69}------------------------------------------------
Image /page/69/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with rounded corners, inside of which is a white image of a vest with a zipper. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the word "AESKU" being larger than the word "DIAGNOSTICS". Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
| % Agreement(95% CI) | Reader 1Lot 1 | Lot 2 | Lot 3 |
|---|---|---|---|
| Positive Agreement | 96(88.9 - 98.6) | 94.7(87.1 - 97.9) | 89.3(80.3 - 94.5) |
| Negative Agreement | 100(92.1 - 100) | 100(92.1 - 100) | 97.8(88.4 - 99.6) |
| Overall Agreement | 97.5(92.9 - 99.1) | 96.7(91.7 - 98.7) | 92.5(86.4 - 96) |
| Pattern Agreement | 96(88.9 - 98.6) | 94.7(87.1 - 97.9) | 89.3(80.3 - 94.5) |
| FI Agreement | 100(96.9 - 100) | 100(96.9 - 100) | 100(96.9 - 100) |
Table 100: Real time Stability AESKUSLIDES ANCA Formalin (Reader 1) – %Agreement:
FI = Fluorescence Intensity
Table 101: Real time Stability AESKUSLIDES ANCA Formalin (Reader 2) – %Agreement:
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 |
|---|---|---|---|
| Positive Agreement | 85.3(75.6 - 91.6) | 85.3(75.6 - 91.6) | 86.7(77.2 - 92.6) |
| Negative Agreement | 100(92.1 - 100) | 100(92.1 - 100) | 97.8(88.4 - 99.6) |
| Overall Agreement | 90.8(84.3 - 94.8) | 90.8(84.3 - 94.8) | 90.8(84.3 - 94.8) |
| Pattern Agreement | 85.3(75.6 - 91.6) | 85.3(75.6 - 91.6) | 86.7(77.2 - 92.6) |
| FI Agreement | 100(96.9 - 100) | 99.2(95.4 - 99.9) | 100(96.9 - 100) |
FI = Fluorescence Intensity
Table 102: Real time Stability AESKUSLIDES ANCA Formalin (Combined Readers) – %Agreement:
| % Agreement(95% CI) | Combined Readers | Lot 1 | Lot 2 | Lot 3 |
|---|---|---|---|---|
| Positive Agreement | 90.7(84.9 - 94.4) | 90(84.2 - 93.8) | 88(81.8 - 92.3) | |
| Negative Agreement | 100(95.9 - 100) | 100(95.9 - 100) | 97.8(92.3 - 99.4) | |
| Overall Agreement | 94.2(90.4 - 96.5) | 93.8(89.9 - 96.2) | 91.7(87.5 - 94.5) | |
| Pattern Agreement | 90.7(84.9 - 94.4) | 90(84.2 - 93.8) | 88(81.8 - 92.3) | |
| FI Agreement | 100(98.4 - 100) | 99.6(97.7 - 99.9) | 100(98.4 - 100) |
{70}------------------------------------------------
Image /page/70/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Agreement calculations for AESKUSLIDES ANCA Formalin included all data points of the test over a time period of 3 months at 2-8°C:
All agreements fulfilled the above criteria.
Positive agreement for AESKUSLIDES ANCA Formalin for both readers ranged from 88.0% to 90.7% for all 3 reagent lots.
Negative agreement for ANCA Formalin for both readers ranged from 97.8% to 100% for all 3 reagent lots.
Overall agreement for ANCA Formalin for both readers ranged from 91.7% to 94.2% for all 3 reagent lots.
Pattern agreement for ANCA Formalin for both readers ranged from 88.0% to 90.7% for all 3 reagent lots.
FI agreement for ANCA Formalin for both readers ranged from 99.6% to 100% for all 3 reagent lots.
Just one deviation of fluorescence intensity greater than +/-1 from the initial value was observed for reader 2.
Conclusion:
All acceptance criteria were met. All agreements were found to be > 85%. The presented data support stability for AESKUSLIDES ANCA Ethanol and ANCA Formalin for at least 3 months at 2-8°C, so far. Other test time points will follow to support the claimed shelf life of 24 months (based on the accelerated stability study of AESKUSLIDES ANCA described elsewhere).
In Use Stability Report
In Use stability testing is designed to predict the period of time that a product remains viable once placed into use. Therefore, In Use stability of AESKUSLIDES ANCA was tested over a period of 6 weeks.
Procedure:
Three different reagent lots were stored at 2-8°C over a time period of 6 weeks. Tests have been performed at the following time points: t0, t1, t2, t4, t6 (t1= after 1 week, t2= after 2 weeks, t4= after 4 weeks, t6 after 6 weeks) for each reagent lot. After the first test at time point t0, kits were incubated at the indicated temperature (2-8°C). Each kit component (with the exception of slide foils) had been opened and all buffers (except wash buffer) had been prepared at time point t0. Eight serum samples have been tested at each time point:
{71}------------------------------------------------
Image /page/71/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image inside the square looks like a stylized letter "Y" with horizontal lines across it. To the right of the square, the text "AESKU.DIAGNOSTICS" is written in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" written in a smaller, lighter gray font below.
| SampleID | AESKUSLIDES ANCA Ethanol | AESKUSLIDES ANCA Formalin | ||||
|---|---|---|---|---|---|---|
| Pos/Neg | Pattern ANCAEthanol | Grading | Pos/Neg | Pattern ANCAFormalin | Grading | |
| 1 | Pos | C-ANCA | high positive | Pos | C-ANCA | high positive |
| 2 | Pos | C-ANCA | medium positive | Pos | C-ANCA | medium positive |
| 3 | Pos | C-ANCA | low positive | Pos | C-ANCA | low positive |
| 4 | Pos | P-ANCA | high positive | Pos | C-ANCA | medium positive |
| 5 | Pos | P-ANCA | medium positive | Pos | C-ANCA | borderline positive |
| 6 | Pos | P-ANCA | low positive | Pos | C-ANCA | borderline positive |
| 7 | Pos | A-ANCA | high positive | Neg | Neg | negative |
| 8 | Neg | Neg | negative | Neg | Neg | negative |
Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.
The stability data is covering the stability of all components of the kit, except for diluted wash buffer and opened slides. Diluted wash buffer is stable for 1 week at 2-8°C, as has been shown in a wash buffer stability study. This is explicitly written in the IFU. Also noted in the IFU is the recommendation to process slides immediately upon opening the silde foil.
Agreements were calculated for individual and combined readers. Results of the kit controls were not included in the agreement calculations.
Acceptance Criteria:
- Positive sera have to be found positive, and negative sera have to be found negative. ●
- . Correct pattern has to be found.
- Reported FI is allowed to differ max. ± 1 level from the expected value.
- . Agreements should meet the following criteria:
| Type of Agreement | % Agreement |
|---|---|
| Positive Agreement | > 85% |
| Negative Agreement | > 85% |
| Overall Agreement | > 85% |
| Total Pattern Agreement | > 85% |
| Fluorescence Intensity Agreement | > 85% |
{72}------------------------------------------------
Image /page/72/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo is clean and professional, with a focus on the company's name and its diagnostic tools.
AESKUSLIDES ANCA 510(k) Traditional Submission
Results:
Table 103: In Use Stability AESKUSLIDES ANCA Ethanol – Number of Samples:
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 1 / 70 | 1 / 70 | 2 / 140 | 1 / 70 | 1 / 70 | 2 / 140 | 0 / 70 | 0 / 70 | 0 / 140 |
| Negative Agreement | 0 / 10 | 0 / 10 | 0 / 20 | 0 / 10 | 0 / 10 | 0 / 20 | 0 / 10 | 0 / 10 | 0 / 20 |
| Overall Agreement | 1 / 80 | 1 / 80 | 2 / 160 | 1 / 80 | 1 / 80 | 2 / 160 | 0 / 80 | 0 / 80 | 0 / 160 |
| Pattern Agreement | 1 / 70 | 1 / 70 | 2 / 140 | 4 / 70 | 4 / 70 | 8 / 140 | 0 / 70 | 1 / 70 | 1 / 140 |
| FI Agreement | 0 / 80 | 0 / 80 | 0 / 160 | 0 / 80 | 0 / 80 | 0 / 160 | 0 / 80 | 0 / 80 | 0 / 160 |
FI = Fluorescence Intensity
Table 104: In Use Stability AESKUSLIDES ANCA Ethanol – % Agreement:
| %Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | both Readers | Reader 1 | Reader 2 | both Readers | Reader 1 | Reader 2 | both Readers | |
| PositiveAgreement | 98.6(92.3 - 99.7) | 98.6(92.3 - 99.7) | 98.6(94.9 - 99.6) | 98.6(92.3 - 99.7) | 98.6(92.3 - 99.7) | 98.6(94.9 - 99.6) | 100(94.8 - 100) | 100(94.8 - 100) | 100(97.3 - 100) |
| NegativeAgreement | 100(72.2 - 100) | 100(72.2 - 100) | 100(83.9 - 100) | 100(72.2 - 100) | 100(72.2 - 100) | 100(83.9 - 100) | 100(72.2 - 100) | 100(72.2 - 100) | 100(83.9 - 100) |
| OverallAgreement | 98.8(93.3 - 99.8) | 98.8(93.3 - 99.8) | 98.8(95.6 - 99.7) | 98.8(93.3 - 99.8) | 98.8(93.3 - 99.8) | 98.8(95.6 - 99.7) | 100(95.4 - 100) | 100(95.4 - 100) | 100(97.7 - 100) |
| PatternAgreement | 98.6(92.3 - 99.7) | 98.6(92.3 - 99.7) | 98.6(94.9 - 99.6) | 94.3(86.2 - 97.8) | 94.3(86.2 - 97.8) | 94.3(89.1 - 97.1) | 100(94.8 - 100) | 98.6(92.3 - 99.7) | 99.3(96.1 - 99.9) |
| FIAgreement | 100(95.4 - 100) | 100(95.4 - 100) | 100(97.7 - 100) | 100(95.4 - 100) | 100(95.4 - 100) | 100(97.7 - 100) | 100(95.4 - 100) | 100(95.4 - 100) | 100(97.7 - 100) |
{73}------------------------------------------------
Image /page/73/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside. The image inside the square looks like a vest with lines going across it. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the text "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below it.
Agreement calculations for AESKUSLIDES ANCA Ethanol included all data points of the whole test time span of 6 weeks at 2-8°C:
All agreements fulfilled the above acceptance criteria.
Positive agreement for AESKUSLIDES ANCA Ethanol for both readers ranged from 98.6% to 100% for all 3 reagent lots.
Negative agreement for ANCA Ethanol for both readers was 100% for all 3 reagent lots.
Overall agreement for ANCA Ethanol for both readers ranged from 98.8% to 100% for all 3 reagent lots.
Pattern agreement for ANCA Ethanol for both readers ranged from 94.3% to 99.3% for all 3 reagent lots.
FI agreement for ANCA Ethanol for both readers was 100% for all 3 reagent lots.
No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.
{74}------------------------------------------------
Image /page/74/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo appears to be for a company that specializes in diagnostic tools.
AESKUSLIDES ANCA 510(k) Traditional Submission
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 13 / 60 | 21 / 60 | 34 / 120 | 11 / 60 | 16 / 60 | 27 / 120 | 11 / 60 | 19 / 60 | 30 / 120 |
| Negative Agreement | 0 / 20 | 0 / 20 | 0 / 40 | 0 / 20 | 0 / 20 | 0 / 40 | 0 / 20 | 0 / 20 | 0 / 40 |
| Overall Agreement | 13 / 80 | 21 / 80 | 34 / 160 | 11 / 80 | 16 / 80 | 27 / 160 | 11 / 80 | 19 / 80 | 30 / 160 |
| Pattern Agreement | 13 / 60 | 21 / 60 | 34 / 120 | 11 / 60 | 16 / 60 | 27 / 120 | 11 / 60 | 19 / 60 | 30 / 120 |
| FI Agreement | 0 / 80 | 0 / 80 | 0 / 160 | 0 / 80 | 0 / 80 | 0 / 160 | 0 / 80 | 2 / 80 | 2 / 160 |
Table 105: In Use AESKUSLIDES ANCA Formalin – Number of Samples (including borderline samples 5 and 6):
FI = Fluorescence Intensity
Table 106: In Use Stability AESKUSLIDES ANCA Formalin – % Agreement (including borderline samples 5 and 6):
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 78.3 | 65 | 71.7 | 81.7 | 73.3 | 77.5 | 81.7 | 68.3 | 75 |
| (66.4 - 86.9) | (52.4 - 75.8) | (63 - 79) | (70.1 - 89.4) | (61 - 82.9) | (69.2 - 84.1) | (70.1 - 89.4) | (55.8 - 78.7) | (66.6 - 81.9) | |
| Negative Agreement | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | |
| Overall Agreement | 83.8 | 73.8 | 78.8 | 86.3 | 80 | 83.1 | 86.3 | 76.3 | 81.3 |
| (74.2 - 90.3) | (63.2 - 82.1) | (71.8 - 84.4) | (77 - 92.1) | (70 - 87.3) | (76.6 - 88.1) | (77 - 92.1) | (65.9 - 84.2) | (74.5 - 86.5) | |
| Pattern Agreement | 78.3 | 65 | 71.7 | 81.7 | 73.3 | 77.5 | 81.7 | 68.3 | 75 |
| (66.4 - 86.9) | (52.4 - 75.8) | (63 - 79) | (70.1 - 89.4) | (61 - 82.9) | (69.2 - 84.1) | (70.1 - 89.4) | (55.8 - 78.7) | (66.6 - 81.9) | |
| FI Agreement | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 97.5 | 98.8 |
| (95.4 - 100) | (95.4 - 100) | (97.7 - 100) | (95.4 - 100) | (95.4 - 100) | (97.7 - 100) | (95.4 - 100) | (91.3 - 99.3) | (95.6 - 99.7) |
{75}------------------------------------------------
Image /page/75/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the company name, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
AESKUSLIDES ANCA 510(k) Traditional Submission
| No of samples(Incorrect/total) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 0 / 40 | 2 / 40 | 2 / 80 | 0 / 40 | 1 / 40 | 1 / 80 | 0 / 40 | 2 / 40 | 2 / 80 |
| Negative Agreement | 0 / 20 | 0 / 20 | 0 / 40 | 0 / 20 | 0 / 20 | 0 / 40 | 0 / 20 | 0 / 20 | 0 / 40 |
| Overall Agreement | 0 / 60 | 2 / 60 | 2 / 120 | 0 / 60 | 1 / 60 | 1 / 120 | 0 / 60 | 2 / 60 | 2 / 120 |
| Pattern Agreement | 0 / 40 | 2 / 40 | 2 / 80 | 0 / 40 | 1 / 40 | 1 / 80 | 0 / 40 | 2 / 40 | 2 / 80 |
| FI Agreement | 0 / 60 | 0 / 60 | 0 / 120 | 0 / 60 | 0 / 60 | 0 / 120 | 0 / 60 | 2 / 60 | 2 / 120 |
Table 107: In Use AESKUSLIDES ANCA Formalin - Number of Samples (excluding borderline samples 5 and 6):
FI = Fluorescence Intensity
Table 108: In Use Stability AESKUSLIDES ANCA Formalin – % Agreement (excluding borderline samples 5 and 6):
| % Agreement(95% CI) | Lot 1 | Lot 2 | Lot 3 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | Reader 1 | Reader 2 | bothReaders | |
| Positive Agreement | 100 | 95 | 97.5 | 100 | 97.5 | 98.8 | 100 | 95 | 97.5 |
| (91.2 - 100) | (83.5 - 98.6) | (91.3 - 99.3) | (91.2 - 100) | (87.1 - 99.6) | (93.3 - 99.8) | (91.2 - 100) | (83.5 - 98.6) | (91.3 - 99.3) | |
| Negative Agreement | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | (83.9 - 100) | (83.9 - 100) | (91.2 - 100) | |
| Overall Agreement | 100 | 96.7 | 98.3 | 100 | 98.3 | 99.2 | 100 | 96.7 | 98.3 |
| (94 - 100) | (88.6 - 99.1) | (94.1 - 99.5) | (94 - 100) | (91.1 - 99.7) | (95.4 - 99.9) | (94 - 100) | (88.6 - 99.1) | (94.1 - 99.5) | |
| Pattern Agreement | 100 | 95 | 97.5 | 100 | 97.5 | 98.8 | 100 | 95 | 97.5 |
| (91.2 - 100) | (83.5 - 98.6) | (91.3 - 99.3) | (91.2 - 100) | (87.1 - 99.6) | (93.3 - 99.8) | (91.2 - 100) | (83.5 - 98.6) | (91.3 - 99.3) | |
| FI Agreement | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 96.7 | 98.3 |
| (94 - 100) | (94 - 100) | (96.9 - 100) | (94 - 100) | (94 - 100) | (96.9 - 100) | (94 - 100) | (88.6 - 99.1) | (94.1 - 99.5) |
{76}------------------------------------------------
Image /page/76/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Agreement calculations for AESKUSLIDES ANCA Formalin included all data points of a test time span of 6 weeks at 2-8°C:
The sample set for AESKUSLIDES ANCA Formalin included two samples (sample 5 and 6) that were borderline positive. For these samples we expect either a low positive result (Fl=1) or a neqative result (Fl=0). Therefore, these samples lead to many "false negatives" when they are defined as "positive" for the calculation (or to many "false positives" when they are defined as "negative" for the calculation). These "false negatives" (or "false positives") have a big impact on positive agreement and on pattern agreement which is due to normal test variances that have a bigger impact on borderline samples than on samples that are clearly negative or clearly positive. Therefore, the lower positive (and pattern) agreements we observed with the agreement calculation for ANCA Formalin are not due to stability issues of the assay.
Therefore, we did an alternative agreement calculation excluding these two borderline positive samples. In this calculation all acceptance criteria have been met:
Positive agreement for AESKUSLIDES ANCA Formalin for both readers ranged from 97.5% to 98.8% for all 3 reagent lots.
Negative agreement for ANCA Formalin for both readers was 100% for all 3 reagent lots.
Overall agreement for ANCA Formalin for both readers ranged from 98.3% to 99.2% for all 3 reagent lots.
Pattern agreement for ANCA Formalin for both readers ranged from 97.5% to 98.8% for all 3 reagent lots.
FI agreement for ANCA Formalin for both readers ranged from 98.3% to 100% for all 3 reagent lots. No deviations of fluorescence intensities greater than +/-1 from the initial value have been observed.
Conclusion:
For AESKUSLIDES ANCA Ethanol all acceptance criteria were met for a time span of 6 weeks at 2-8°C. Therefore, we claim an In Use Stability of 6 weeks at 2-8°C storage temperature.
For AESKUSLIDES ANCA Formalin all acceptance criteria were met for a time span of 6 weeks at 2-8°C. Therefore, we claim an In Use Stability of 6 weeks at 2-8°C storage temperature.
{77}------------------------------------------------
Image /page/77/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image inside, and the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
Transport Stability Report
To determined transport stability of AESKUSLIDES ANCA, data were taken from Accelerated Stability Report, as these conditions used in the study simulate a worst case scenario of transport of the kits.
The kit components were incubated at 37°C over a period of 6 weeks.
Acceptance Criteria:
- Positive sera have to be found positive, and negative sera have to be found negative. ●
- Correct pattern has to be found.
- Reported FI is allowed to differ max. ± 1 level from the expected value. ●
- . Agreements should meet the following criteria:
| Type of Agreement | % Agreement |
|---|---|
| Positive Agreement | > 85% |
| Negative Agreement | > 85% |
| Overall Agreement | > 85% |
| Total Pattern Agreement | > 85% |
| Fluorescence Intensity Agreement | > 85% |
Conclusion:
All criteria were fulfilled in each test. Therefore, we claim that the kit can resist a temperature of 37°C for at least 2 weeks.
For more information please see at "Accelerated Stability Report".
{78}------------------------------------------------
Image /page/78/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The logo is clean and professional, with a focus on the company's name and its diagnostic capabilities.
2. Comparison studies
Method comparison with predicate device:
To show comparable performance of AESKUSLIDES ANCA (Ethanol and Formalin) and the predicate assay NOVA Lite ANCA (Ethanol and Formalin) of INOVA.
Procedure:
A method comparison study with 507 serum samples (comprising 132 serum samples from patients with AAV and 375 samples from patients with other diseases) was performed in house. A detailed description of the study sample set can be found on page 12-79, table 110.
Healthy donor samples have been excluded from this study.
For each assay the sample set was processed manually and analyzed (positive/negative classification, pattern classification) at the microscope (Method C).
All assays were performed according to the respective IFUs. Sample were screened at a dilution of 1:20.
From the results of both assays the relative sensitivity, specificity, positive (%), negative (%) and overall agreements (%) were calculated.
Furthermore, diagnostic sensitivity and specificity were calculated for both assays, and the cross reactivities for different diagnoses are shown in a table.
Results:
Results Predicate comparison ANCA Ethanol
Diagnostic Sensitivity and Specificity
Table 109a: Method comparison with predicate device ANCA Ethanol Diagnostic Sensitivity and Specificity Number of positive and negative results:
| Diagnostic Sensitivity & Specificity | Diagnosis | |||
|---|---|---|---|---|
| AAVs | Other Diseases | Total | ||
| AESKUSLIDES ANCA Ethanol | Pos | 64 | 115 | 179 |
| Neg | 68 | 260 | 328 | |
| Total | 132 | 375 | 507 | |
| Predicate Assay | Pos | 48 | 168 | 216 |
| Neg | 84 | 207 | 291 | |
| Total | 132 | 375 | 507 |
{79}------------------------------------------------
Image /page/79/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The symbol in the green square appears to be a stylized representation of a medical or diagnostic tool.
Table 109b: Method comparison with predicate device ANCA Ethanol Diagnostic Sensitivity and Specificity PPV/NPV:
| % DiagnosticSensitivity &Specificity (95% CI) | % Sensitivity(95% CI)AAVs (n= 132) | % Specificity(95% CI)OD (n=375) | PredictiveValue of aPositive TestResult (PPV) | Predictive Value ofa Negative TestResult (NPV) |
|---|---|---|---|---|
| AESKUSLIDES ANCAEthanol | 48.5(40.1 - 56.9) | 69.3(64.5 - 73.8) | 35.8% | 79.3% |
| Predicate Assay | 36.4(28.7 - 44.8) | 55.2(50.1 - 60.2) | 22.2% | 71.1% |
Sensitivity of AESKUSLIDES ANCA Ethanol is higher (48.5%) than that calculated for the predicate assay (36.4%). Specificity for AESKUSLIDES ANCA Ethanol was 69.3%, whereas for the predicate assay a specificity of 55.2% was calculated. Values reported in literature are variable and spread over a wide range, depending on the serum panel used in the study. The ANCA titer may be associated with disease activity and may also disappear upon treatment. The AAV cohort tested in this study was not selected for disease activity or treatment.
The higher sensitivity and specificity of AESKUSLIDES ANCA Ethanol go along with a higher PPV (35.0% compared to 22.2% for the predicate) and NPV (79.3% vs 71.1%).
The results for both assays are shown in detail in the cross-reactivity table below. With AESKUSLIDES ANCA Ethanol fewer cross-reactivity are seen. Cross-reactivity could be detected in autoimmune liver diseases (up to 50%), inflammatory bowel diseases (up to 50.7%) and also other rheumatic diseases (SLE, RA, Scleroderma: up to 47.4%). The predicate assay detected positives in similar diagnoses but with higher number of positives.
{80}------------------------------------------------
Image /page/80/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
AESKUSLIDES ANCA
510(k) Traditional Submission
| AESKUSLIDES ANCA Ethanol | Predicate | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Diagnosis | n | n Pos | %Pos | n Neg | %Neg | n Pos | %Pos | n Neg | %Neg | ||
| TargetDiagnosis | ANCA associatedVasculitis (AAV) | Wegener's Granulomatosis | 78 | 33 | 42.3% | 45 | 57.7% | 20 | 25.6% | 58 | 74.4% |
| MPA | 29 | 22 | 75.9% | 7 | 24.1% | 20 | 69.0% | 9 | 31.0% | ||
| Churg-Strauss Syndrome | 25 | 9 | 36.0% | 16 | 64.0% | 8 | 32.0% | 17 | 68.0% | ||
| Autoimmune Hepatitis (AIH) | 8 | 3 | 37.5% | 5 | 62.5% | 5 | 62.5% | 3 | 37.5% | ||
| Autoimmune LiverDiseases | Autoimmune Hepatitis/ Primary biliarycholangitis | 6 | 3 | 50.0% | 3 | 50.0% | 6 | 100.0% | 0 | 0.0% | |
| Primary biliary cholangitis (PBC) | 11 | 2 | 18.2% | 9 | 81.8% | 8 | 72.7% | 3 | 27.3% | ||
| Primary sclerosing cholangitis (PSC) | 10 | 5 | 50.0% | 5 | 50.0% | 5 | 50.0% | 5 | 50.0% | ||
| Ulcerative Colitis | 71 | 36 | 50.7% | 35 | 49.3% | 42 | 59.2% | 29 | 40.8% | ||
| Inflammatory boweldiseases | Crohn's disease | 40 | 16 | 40.0% | 24 | 60.0% | 26 | 65.0% | 14 | 35.0% | |
| Inflammatory bowel disease (IBD) | 9 | 4 | 44.4% | 5 | 55.6% | 5 | 55.6% | 4 | 44.4% | ||
| Rheumatoid Arthritis (RA) | 12 | 4 | 33.3% | 8 | 66.7% | 9 | 75.0% | 3 | 25.0% | ||
| Other rheumaticdiseases | Systemic Lupus Erythematosus (SLE) | 30 | 14 | 46.7% | 16 | 53.3% | 22 | 73.3% | 8 | 26.7% | |
| Controldiagnosis | Scleroderma | 19 | 9 | 47.4% | 10 | 52.6% | 13 | 68.4% | 6 | 31.6% | |
| Myositis | 1 | 0 | 0.0% | 1 | 100.0% | 0 | 0.0% | 1 | 100.0% | ||
| Infections | Hepatitis C Virus (HCV ) | 16 | 3 | 18.8% | 13 | 81.3% | 5 | 31.3% | 11 | ||
| Hepatitis B Virus (HBV ) | 8 | 0 | 0.0% | 8 | 100.0% | 0 | 0.0% | 8 | |||
| Vasculitides NOTassociated with ANCA | Polymyalgia rheumatica (PMR) | 18 | 3 | 16.7% | 15 | 83.3% | 5 | 27.8% | 13 | 72.2% | |
| Giant cell artertitis (GCA) | 4 | 0 | 0.0% | 4 | 100.0% | 2 | 50.0% | 2 | 50.0% | ||
| Purpura | 1 | 1 | 100.0% | 0 | 0.0% | 1 | 100.0% | 0 | 0.0% | ||
| Leukemia | Lymphoma, Myeloma | 15 | 3 | 20.0% | 12 | 80.0% | 4 | 26.7% | 11 | 73.3% | |
| Other diagnosis | Celiac disease | 10 | 0 | 0.0% | 10 | 100.0% | 0 | 0.0% | 10 | 100.0% | |
| Chronic kidney disease | 32 | 2 | 6.3% | 30 | 93.8% | 3 | 9.4% | 29 | 90.6% | ||
| Sinusitis | 22 | 4 | 18.2% | 18 | 81.8% | 4 | 18.2% | 18 | 81.8% | ||
| Asthma | 32 | 3 | 9.4% | 29 | 90.6% | 3 | 9.4% | 29 | 90.6% | ||
| Sum | 507 | 179 | 35.3% | 328 | 64.7% | 216 | 42.6% | 291 | 57.4% |
Table 110: Method comparison with predicate device ANCA Ethanol. Number and positive samples in the different tested diagnosis
{81}------------------------------------------------
Relative sensitivity and specificity
Table 110a: Method comparison with predicate device ANCA Ethanol Relative Sensitivity and Specificity Number of positive and negative results and % agreements:
| Method Comparison | Predicate | |||
|---|---|---|---|---|
| Pos | Neg | Total | ||
| AESKU ANCAEthanol | Pos | 145 | 34 | 179 |
| Neg | 71 | 257 | 328 | |
| Total | 216 | 291 | 507 |
| Method Comparison Predicate vs AESKU | % Agreement (95% CI) |
|---|---|
| Positive Agreement | 67.1(60.6 - 73) |
| Negative Agreement | 88.3(84.1 - 91.5) |
| Overall Agreement | 79.3(75.5 - 82.6) |
Positive, negative and overall agreement between both assays were acceptable (67.1%, 88.3% and 79.3%). The rather low positive agreement was due to the fact, that the predicate assay detects less positive results.
Conclusions
AESKUSLIDES ANCA Ethanol showed higher diagnostic sensitivity (48.5%) and specificity (69.3%) compared to the predicate assay NOVA Lite provided by INOVA (36.4%, 55.2%). This is due to the fact, that AESKU assay detects more positives in the AAV cohort, and less positives in the other disease groups.
This study shows that the AESKU assay is comparable or better than the predicate device.
{82}------------------------------------------------
Results Predicate comparison ANCA Formalin
Diagnostic Sensitivity and Specificity
Table 111: Method comparison with predicate device_ANCA Formalin_Diagnostic Sensitivity and Specificity Number of positive and negative results:
| Diagnostic Sensitivity &Specificity | Diagnosis | |||
|---|---|---|---|---|
| AAVs | Other Diseases | Total | ||
| AESKUSLIDESANCA Formalin | Pos | 66 | 35 | 101 |
| Neg | 66 | 340 | 406 | |
| Total | 132 | 375 | 507 | |
| Predicate Assay | Pos | 50 | 32 | 82 |
| Neg | 82 | 343 | 425 | |
| Total | 132 | 375 | 507 |
Table 112: Method comparison with predicate device_ANCA Formalin_Diagnostic Sensitivity and Specificity PPV/NPV:
| % Diagnostic Sensitivity &Specificity (95% CI) | % Sensitivity(95% CI) | % Specificity(95% CI) | PredictiveValue of aPositive TestResult (PPV) | Predictive Value ofa Negative TestResult (NPV) |
|---|---|---|---|---|
| AAVs (n= 132) | OD (n=375) | |||
| AESKUSLIDES ANCAFormalin | 50(41.6 - 58.4) | 90.7(87.3 - 93.2) | 65.3% | 83.7% |
| Predicate Assay | 37.9(30.1 - 46.4) | 91.5(88.2 - 93.9) | 61.0% | 80.7% |
Sensitivity of AESKUSLIDES ANCA Formalin is higher (50.0%) than that calculated for the predicate assay (37.9%) in this study. Specificities for both assays were comparable (90.7% vs 91.5%). PPVs and NPVs of both assays were comparable. However, values for AESKUSLIDES were slightly higher.
The cross-reactivity table shows equivalent results for both assays. Cross-reactivity could be detected in AIH (25.0%), PSC (20.0%), Ulcerative Colitis (16.9% and SLE (20.0%)
{83}------------------------------------------------
Image /page/83/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray. The logo appears to be for a company that specializes in diagnostic tools.
AESKUSLIDES ANCA
510(k) Traditional Submission
| Table 113: Method comparison with predicate device_ANCA Formalin. Number and percent positive samples in the different tested diagnosis | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| AESKUSLIDES ANCA Formalin | Predicate | ||||||||||
| Diagnosis | n | n Pos | %Pos | n Neg | %Neg | n Pos | %Pos | n Neg | %Neg | ||
| TargetDiagnosis | ANCA associatedVasculitis (AAV) | Wegener's Granulomatosis | 78 | 45 | 57.7% | 33 | 42.3% | 37 | 47.4% | 41 | 52.6% |
| MPA | 29 | 15 | 51.7% | 14 | 48.3% | 12 | 41.4% | 17 | 58.6% | ||
| Churg-Strauss Syndrome | 25 | 6 | 24.0% | 19 | 76.0% | 1 | 4.0% | 24 | 96.0% | ||
| Autoimmune LiverDiseases | Autoimmune Hepatitis (AIH) | 8 | 2 | 25.0% | 6 | 75.0% | 1 | 12.5% | 7 | 87.5% | |
| Autoimmune Hepatitis/ Primary biliarycholangitis | 6 | 0 | 0.0% | 6 | 100.0% | 0 | 0.0% | 6 | 100.0% | ||
| Primary biliary cholangitis (PBC) | 11 | 0 | 0.0% | 11 | 100.0% | 0 | 0.0% | 11 | 100.0% | ||
| Primary sclerosing cholangitis (PSC) | 10 | 2 | 20.0% | 8 | 80.0% | 0 | 0.0% | 10 | 100.0% | ||
| Inflammatorybowel diseases | Ulcerative Colitis | 71 | 12 | 16.9% | 59 | 83.1% | 10 | 14.1% | 61 | 85.9% | |
| Crohn's disease | 40 | 1 | 2.5% | 39 | 97.5% | 0 | 0.0% | 40 | 100.0% | ||
| Inflammatory bowel disease (IBD) | 9 | 1 | 11.1% | 8 | 88.9% | 1 | 11.1% | 8 | 88.9% | ||
| Controldiagnosis | Other rheumaticdiseases | Rheumatoid Arthritis (RA) | 12 | 1 | 8.3% | 11 | 91.7% | 1 | 8.3% | 11 | 91.7% |
| Systemic Lupus Erythematosus (SLE) | 30 | 6 | 20.0% | 24 | 80.0% | 8 | 26.7% | 22 | 73.3% | ||
| Scleroderma | 19 | 0 | 0.0% | 19 | 100.0% | 2 | 10.5% | 17 | 89.5% | ||
| Myositis | 1 | 0 | 0.0% | 1 | 100.0% | 0 | 0.0% | 1 | 100.0% | ||
| Infections | Hepatitis C Virus (HCV ) | 16 | 2 | 12.5% | 14 | 87.5% | 3 | 18.8% | 13 | ||
| Hepatitis B Virus (HBV ) | 8 | 0 | 0.0% | 8 | 100.0% | 0 | 0.0% | 8 | |||
| Vasculitides NOTassociated withANCA | Polymyalgia rheumatica (PMR) | 18 | 2 | 11.1% | 16 | 88.9% | 1 | 5.6% | 17 | ||
| Giant cell artertitis (GCA) | 4 | 0 | 0.0% | 4 | 100.0% | 2 | 50.0% | 2 | |||
| Leukemia | Purpura | 1 | 1 | 100.0% | 0 | 0.0% | 1 | 100.0% | 0 | ||
| Lymphoma, Myeloma | 15 | 2 | 13.3% | 13 | 86.7% | 0 | 0.0% | 15 | |||
| Other diagnosis | Celiac disease | 10 | 0 | 0.0% | 10 | 100.0% | 0 | 0.0% | 10 | ||
| Chronic kidney disease | 32 | 2 | 6.3% | 30 | 93.8% | 1 | 3.1% | 31 | |||
| Sinusitis | 22 | 1 | 4.5% | 21 | 95.5% | 1 | 4.5% | 21 | 95.5% | ||
| Asthma | 32 | 0 | 0.0% | 32 | 100.0% | 0 | 0.0% | 32 | 100.0% | ||
| Sum | 507 | 101 | 19.9% | 406 | 80.1% | 82 | 16.2% | 425 | 83.8% |
Table 113: Method comparison with predicate device ANCA Formalin. Number and percent positive samples in the different tested diagnosis
{84}------------------------------------------------
Relative sensitivity and specificity
Table 114: Method comparison with predicate device ANCA Formalin Relative Sensitivity and Specificity Number of positive and negative results and % agreements:
| Method Comparison | Predicate | |||
|---|---|---|---|---|
| Pos | Neg | Total | ||
| AESKUSLIDESANCA Formalin | Pos | 66 | 35 | 101 |
| Neg | 16 | 390 | 406 | |
| Total | 82 | 425 | 507 |
| Method Comparison ANCAFormalin | % Agreement(95% CI) |
|---|---|
| Positive Agreement | 80.5(70.6 - 87.6) |
| Negative Agreement | 91.8(88.8 - 94) |
| Overall Agreement | 89.9(87 - 92.3) |
Overall agreement between both assays was good (89.9%). The same holds true for the negative agreement (91.8%). The relatively low positive agreement (80.5%) is due to the fact, that the AESKU assay detects more samples as positive.
Conclusions
AESKUSLIDES ANCA Formalin showed a diagnostic sensitivity (50.0%) compared to the predicate assay NOVA Lite provided by INOVA (37.9%). However, the diagnostic sensitivity was comparable between the two (90.7% vs 91.5%).
This study shows that the AESKU assay is comparable or better than the predicate device.
Method Comparison of Method A, B, and C and clinical study
To show comparable performance of AESKUSLIDES ANCA Ethanol and ANCA Formalin when processed manually and on HELIOS.
Procedure:
A clinical study with the same 630 serum samples (comprising 135 serum samples from patients with AAV, 120 characterized MPO/PR3/ANCA positive sera and 375 samples from patients with other diseases) was performed on 3 different study sites (2 US sites, one German site). A detailed description of the study sample can be found on page 12-87/88, table 118.
{85}------------------------------------------------
At each study site the sample set was processed
- manually and analyzed (positive/negative classification) at the microscope by two independent readers (Method C)
- on HELIOS and HELIOS images were analyzed by two independent readers at the PC (Method B)
- on HELIOS and positive/negative classification was done by the HELIOS Vasculitis Pattern Plus Software (ANCA) (Method A)
All assays were performed according to the IFU.
For each study site, the results of the single and combined readers with the different Methods were compared to each other and % agreements were calculated:
Method C (Manual) vs Method B (Reader Confirmation)
Method B (Reader Confirmation) vs Method A (HELIOS)
Method C (Manual) vs Method A (HELIOS).
Furthermore, the number of positive samples found for each tested diagnosis for all study sites and all methods will be compared as well as the calculated diagnostic sensitivities and specificities. For the calculation of diagnostic sensitivity and specificity the 120 characterized MPO/PR3/ANCA positive sera have been excluded.
Acceptance Criteria:
Positive, Negative, and Overall Agreements between the different Methods should be >85%.
Positive, Negative, Overall and Pattern Agreements between the different methods were set to be >80%. This is due to the fact that unknown clinical samples often results in mixed patterns, which are not easy to characterize. For example, in general one reader may report a sample with unspecific fluorescence as positive undefined, whereas another evaluates it as negative. Thus, the evaluation of ANCA slides is very subjective and strongly dependent on the reader's experience. Therefore, we expect and accept variances up to 15% in this clinical study.
For Method A we accept a lower agreement (>70%). Comparing Methods B vs A and C vs A. In general, method A (HELIOS Vasculitis Pattern Plus software) detects less positive samples. In addition, positive samples are mostly classified as positive undefined. This can be explained by the complex patterns, which can be distinguished by the readers at the microscope or at the screen, but to limited extent by the software. Because of this, all results generated by the HELIOS (Method A) must be confirmed by trained personnel.
{86}------------------------------------------------
Results:
Results Clinical study and method comparison ANCA Ethanol
Table 115: Method Comparison of Method A, B and C_ANCA Ethanol_Single Reader Agreements
| % Agreement(95% CI) | Method C vs B | Method B vs A | Method C vs A | ||||
|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 1 | Reader 2 | Reader 1 | Reader 2 | ||
| Site 1 | PositiveAgreement | 86.5(80.9 - 90.6) | 85.9(80.2 - 90.1) | 79(72.3 - 84.5) | 79(72.3 - 84.5) | 70.8(64 - 76.8) | 70.7(63.9 - 76.7) |
| NegativeAgreement | 99.8(98.7 - 100) | 99.3(98 - 99.8) | 98.3(96.6 - 99.1) | 98.3(96.6 - 99.1) | 99.1(97.7 - 99.6) | 98.9(97.4 - 99.5) | |
| TotalAgreement | 95.7(93.8 - 97) | 95.2(93.3 - 96.6) | 93.2(90.9 - 94.9) | 93.2(90.9 - 94.9) | 90.5(87.9 - 92.5) | 90.3(87.8 - 92.4) | |
| Site 2 | PositiveAgreement | 88.5(83.4 - 92.1) | 92.8(88.4 - 95.6) | 92.6(88 - 95.6) | 86.9(81.6 - 90.8) | 83.2(77.5 - 87.6) | 81.6(75.8 - 86.3) |
| NegativeAgreement | 98.6(96.9 - 99.3) | 96.7(94.5 - 98) | 98.6(97.1 - 99.4) | 99.3(97.9 - 99.8) | 97.9(96 - 98.9) | 96.9(94.8 - 98.2) | |
| TotalAgreement | 95.2(93.3 - 96.6) | 95.4(93.5 - 96.8) | 96.8(95.1 - 97.9) | 95.2(93.3 - 96.6) | 93(90.8 - 94.8) | 91.9(89.5 - 93.8) | |
| Site 3 | PositiveAgreement | 90.2(85.6 - 93.5) | 85.1(79.5 - 89.4) | 79.5(73.6 - 84.3) | 83.8(78.2 - 88.2) | 76(70 - 81.1) | 79(72.7 - 84.1) |
| NegativeAgreement | 96(93.7 - 97.6) | 91.3(88.2 - 93.6) | 99(97.5 - 99.6) | 98.4(96.6 - 99.2) | 98.3(96.5 - 99.2) | 94.5(91.9 - 96.3) | |
| TotalAgreement | 94(91.8 - 95.6) | 89.4(86.7 - 91.5) | 92.2(89.9 - 94.1) | 93.7(91.5 - 95.3) | 90.3(87.8 - 92.4) | 89.7(87.1 - 91.8) |
Table 116: Method Comparison of Method A, B and C_ANCA Ethanol_Combined Reader Agreements
| % Agreement (95% CI) | Method C vs B | Method B vs A | Method C vs A | |
|---|---|---|---|---|
| Positive Agreement | 86.2(82.3 - 89.3) | 79(74.4 - 83.1) | 70.8(66 - 75.1) | |
| Site 1 | Negative Agreement | 99.5(98.8 - 99.8) | 98.3(97.2 - 98.9) | 99(98.1 - 99.5) |
| Total Agreement | 95.5(94.2 - 96.5) | 93.2(91.6 - 94.4) | 90.4(88.6 - 91.9) | |
| Positive Agreement | 90.6(87.4 - 93) | 89.6(86.3 - 92.3) | 82.4(78.5 - 85.8) | |
| Site 2 | Negative Agreement | 97.6(96.4 - 98.5) | 99(98 - 99.5) | 97.4(96.1 - 98.3) |
| Total Agreement | 95.3(94 - 96.4) | 96(94.8 - 97) | 92.5(90.9 - 93.8) | |
| Positive Agreement | 87.9(84.4 - 90.6) | 81.6(77.6 - 85) | 77.4(73.1 - 81.1) | |
| Site 3 | Negative Agreement | 93.6(91.7 - 95) | 98.7(97.7 - 99.3) | 96.3(94.8 - 97.4) |
| Total Agreement | 91.7(90 - 93.1) | 92.9(91.4 - 94.2) | 90(88.2 - 91.5) |
{87}------------------------------------------------
| % Agreement (95% CI) | Method C vs B | Method B vs A | Method C vs A |
|---|---|---|---|
| Site 1 | 87(85 - 88.7) | 81(78.8 - 82.9) | 77.8(75.4 - 79.9) |
| Site 2 | 89.2(87.5 - 90.7) | 84.7(82.6 - 86.5) | 81.8(79.7 - 83.8) |
| Site 3 | 82.5(80.5 - 84.2) | 85.6(83.7 - 87.4) | 80.3(78.2 - 82.3) |
| All sites | 86.2(85.2 - 87.2) | 83.8(82.6 - 84.8) | 80(78.7 - 81.2) |
Table 117: Method Comparison of Method A, B and C – Pattern Agreements
{88}------------------------------------------------
Image /page/88/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white image inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text is the phrase "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font. The image inside the green square appears to be a stylized representation of a medical or diagnostic tool.
AESKUSLIDES ANCA 510(k) Traditional submission
Table 118: Method comparison. Number and percent positive samples in the different tested diagnosis_ANCA Ethanol
| Site 1 | Site 2 | Site 3 | ||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CrossReactivity | Manual Performance | Reader Confirmation at HELIOS | HELIOSsuggestion | Manual Performance | Reader Confirmation at HELIOS | HELIOSsuggestion | Manual Performance | Reader Confirmation at HELIOS | HELIOSsuggestion | |||||||||||||||||||||||
| Diagnosis | n | Reader 1 | Reader 2 | Reader 1 | Reader 2 | n | Reader 3 | Reader 4 | Reader 3 | Reader 4 | n | Reader 5 | Reader 6 | Reader 5 | Reader 6 | |||||||||||||||||
| nPos | %Pos | nPos | %Pos | nPos | %Pos | nPos | %Pos | n | %Pos | nPos | %Pos | nPos | %Pos | nPos | %Pos | nPos | %Pos | n | %Pos | nPos | %Pos | nPos | %Pos | nPos | %Pos | nPos | %Pos | |||||
| Wegener's | 80 | 13 | 16.3 | 13 | 16.3 | 13 | 16.3 | 13 | 16.3 | 11 | 13.8 | 20 | 25.0 | 22 | 27.5 | 17 | 21.2 | 21 | 26.2 | 17 | 21.2 | 19 | 23.7 | 21 | 26.2 | 20 | 25 | 22 | 27.5 | 17 | 21.2 | |
| MPA | 29 | 18 | 62.1 | 18 | 62.1 | 16 | 55.2 | 16 | 55.2 | 14 | 48.3 | 19 | 65.5 | 18 | 62.0 | 20 | 68.9 | 20 | 68.9 | 20 | 68.9 | 19 | 65.5 | 19 | 65.5 | 19 | 65.5 | 19 | 65.5 | 18 | 62.0 | |
| Churg-Strauss | 26 | 2 | 7.7 | 2 | 7.7 | 2 | 7.7 | 2 | 7.7 | 0 | 0.0 | 5 | 19.2 | 4 | 15.3 | 5 | 19.2 | 5 | 19.2 | 6 | 23.0 | 6 | 23.0 | 6 | 23.0 | 8 | 30.7 | 8 | 30.7 | 4 | 15.3 | |
| C-ANCA | 20 | 14 | 70.0 | 14 | 70.0 | 13 | 65.0 | 13 | 65.0 | 11 | 55.0 | 17 | 85.0 | 17 | 85 | 16 | 80 | 16 | 80 | 14 | 70 | 14 | 70 | 13 | 65 | 13 | 65 | 14 | 70 | 13 | 65 | |
| P-ANCA | 50 | 24 | 48.0 | 23 | 46.0 | 24 | 48.0 | 24 | 48.0 | 25 | 50.0 | 23 | 46.0 | 23 | 46 | 23 | 46 | 25 | 50 | 24 | 48 | 25 | 50 | 25 | 50 | 25 | 50 | 26 | 52 | 25 | 50 | |
| MPO | 25 | 18 | 72.0 | 18 | 72.0 | 17 | 68.0 | 17 | 68.0 | 18 | 72.0 | 23 | 92.0 | 22 | 88 | 19 | 76 | 22 | 88 | 18 | 72 | 22 | 88 | 20 | 80 | 20 | 80 | 19 | 76 | 18 | 72 | |
| PR3 | 25 | 5 | 20.0 | 5 | 20.0 | 3 | 12.0 | 3 | 12.0 | 3 | 12.0 | 9 | 36.0 | 7 | 28 | 8 | 32 | 7 | 28 | 7 | 28 | 15 | 60 | 12 | 48 | 10 | 40 | 10 | 40 | 11 | 44 | |
| AIH | 8 | 5 | 62.5 | 5 | 62.5 | 5 | 62.5 | 5 | 62.5 | 4 | 50.0 | 3 | 37.5 | 3 | 37.5 | 2 | 25 | 3 | 37.5 | 2 | 25 | 2 | 25 | 1 | 12.5 | 3 | 37.5 | 1 | 12.5 | 1 | 12.5 | |
| AIH/PBC | 6 | 4 | 66.7 | 4 | 66.7 | 4 | 66.7 | 4 | 66.7 | 3 | 50.0 | 3 | 50.0 | 3 | 50 | 3 | 50 | 3 | 50 | 2 | 33.3 | 1 | 16.6 | 4 | 66.6 | 3 | 50 | 2 | 33.3 | |||
| PBC | 11 | 3 | 27.3 | 3 | 27.3 | 3 | 27.3 | 3 | 27.3 | 2 | 18.2 | 2 | 18.2 | 2 | 18.1 | 3 | 27.2 | 4 | 36.3 | 1 | 9.09 | 4 | 36.3 | 2 | 18.1 | 4 | 36.3 | 3 | 27.2 | 1 | 9.0 | |
| PSC | 10 | 5 | 50.0 | 5 | 50.0 | 5 | 50.0 | 5 | 50.0 | 3 | 30.0 | 5 | 50.0 | 5 | 50 | 5 | 50 | 5 | 50 | 4 | 40 | 1 | 10 | 4 | 40 | 4 | 40 | 3 | 30 | |||
| Ulcerative Colitis | 71 | 28 | 39.4 | 28 | 39.4 | 16 | 22.5 | 16 | 22.5 | 15 | 21.1 | 30 | 42.3 | 30 | 42.2 | 28 | 39.4 | 29 | 40.8 | 28 | 39.4 | 32 | 45.0 | 27 | 38.0 | 34 | 47.8 | 25 | 35.2 | 27 | 38.0 | |
| Crohn's | 40 | 10 | 25.0 | 9 | 22.5 | 8 | 20.0 | 8 | 20.0 | 2 | 5.0 | 13 | 32.5 | 11 | 27.5 | 9 | 22.5 | 7 | 17.5 | 6 | 15 | 13 | 32.5 | 12 | 30 | 12 | 30 | 10 | 25 | 9 | 22.5 | |
| IBD | 9 | 1 | 11.1 | 1 | 11.1 | 1 | 11.1 | 1 | 11.1 | 0 | 0.0 | 4 | 44.4 | 4 | 44.4 | 4 | 44.4 | 4 | 44.4 | 3 | 33.3 | 5 | 55.5 | 3 | 33.3 | 5 | 55.5 | 3 | 33.3 | 3 | 33.3 | |
| RA | 12 | 1 | 8.3 | 1 | 8.3 | 1 | 8.3 | 1 | 8.3 | 1 | 8.3 | 1 | 8.3 | 2 | 16.6 | 0 | 0 | 1 | 8.33 | 0 | 0 | 4 | 33.3 | 4 | 33.3 | 1 | 8.3 | 1 | 8.3 | 1 | 8.3 | |
| SLE | 30 | 16 | 53.3 | 16 | 53.3 | 15 | 50.0 | 15 | 50.0 | 12 | 40.0 | 14 | 46.7 | 14 | 46.6 | 14 | 46.6 | 14 | 46.6 | 13 | 43.3 | 16 | 53.3 | 12 | 40 | 14 | 46.6 | 12 | 40 | 10 | 33.3 | |
| Myositis | 1 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Scleroderma | 19 | 10 | 52.6 | 10 | 52.6 | 10 | 52.6 | 10 | 52.6 | 6 | 31.6 | 5 | 26.3 | 8 | 42.1 | 4 | 21.0 | 9 | 47.3 | 5 | 26.3 | 8 | 42.1 | 0 | 0 | 10 | 52.6 | 9 | 47.3 | 5 | 26.3 | |
| HBV | 8 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 12.5 | 1 | 12.5 | 0 | 0 | 0 | 0 | 0 | 0 | |
| HCV | 16 | 3 | 18.8 | 3 | 18.8 | 2 | 12.5 | 2 | 12.5 | 3 | 18.8 | 3 | 18.8 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 3 | 18.7 | 2 | 12.5 | |
| GCA | 4 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| PMR | 18 | 4 | 22.2 | 4 | 22.2 | 3 | 16.7 | 3 | 16.7 | 3 | 16.7 | 2 | 11.1 | 2 | 11.1 | 0 | 0 | 1 | 5.5 | 0 | 0 | 3 | 16.6 | 3 | 16.6 | 2 | 11.1 | 2 | 11.1 | 1 | 5.5 | |
| Purpura | 1 | 1 | 100 | 1 | 100 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | 1 | 100 | |
| Leukemia | 15 | 3 | 20.0 | 4 | 26.7 | 2 | 13.3 | 2 | 13.3 | 1 | 6.7 | 2 | 13.3 | 2 | 13.3 | 2 | 13.3 | 2 | 13.3 | 2 | 13.3 | 3 | 20 | 2 | 13.3 | 3 | 20 | 2 | 13.3 | 2 | 13.3 | |
| Celiacdisease | 10 | 1 | 10.0 | 1 | 10.0 | 1 | 10.0 | 1 | 10.0 | 1 | 10.0 | 0 | 0.0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| CKD | 32 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 0 | 0 | 1 | 3.1 | 3 | 9.3 | 1 | 3.1 | 3 | 9.3 | 1 | 3.1 | |
| Asthma | 32 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 0 | 0.0 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 2 | 6.2 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 2 | 6.2 | 1 | 3.1 | |
| Sinusitis | 22 | 1 | 4.5 | 1 | 4.5 | 1 | 4.5 | 1 | 4.5 | 1 | 4.5 | 2 | 9.1 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 | 2 | 9.0 |
{89}------------------------------------------------
Image /page/89/Picture/0 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, it says "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
AESKUSLIDES ANCA 510(k) Traditional submission
{90}------------------------------------------------
Image /page/90/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller, lighter gray font.
AESKUSLIDES ANCA 510(k) Traditional submission
| % Diagnostic Sensitivity & Specificity(95% CI) | % Sensitivity(95% CI) | % Specificity(95% CI) | % Sensitivity(95% CI) | % Specificity(95% CI) | |||
|---|---|---|---|---|---|---|---|
| AAV (n= 135) | OD (n=375) | AAV (n= 135) | OD (n=375) | ||||
| Site1 | Manual | Reader 1 | 24.4(18 - 32.3) | 73.9(69.2 - 78.1) | CombinedReaders | 24.4(19.7 - 29.9) | 73.9(70.6 - 76.9) |
| Reader 2 | 24.4(18 - 32.3) | 73.9(69.2 - 78.1) | |||||
| ReaderConfirmationat HELIOS | Reader 1 | 23(16.7 - 30.7) | 78.9(74.5 - 82.8) | CombinedReaders | 23(18.3 - 28.3) | 78.9(75.9 - 81.7) | |
| Reader 2 | 23(16.7 - 30.7) | 78.9(74.5 - 82.8) | |||||
| HELIOS | 18.5(12.9 - 25.9) | 84.5(80.5 - 87.8) | |||||
| Site2 | Manual | Reader 3 | 32.6(25.3 - 40.9) | 75.5(70.9 - 79.5) | CombinedReaders | 32.6(27.3 - 38.4) | 75.2(72 - 78.2) |
| Reader 4 | 32.6(25.3 - 40.9) | 74.9(70.3 - 79.1) | |||||
| ReaderConfirmationat HELIOS | Reader 3 | 31.1(23.9 - 39.4) | 78.1(73.7 - 82) | CombinedReaders | 32.6(27.3 - 38.4) | 77.1(73.9 - 79.9) | |
| Reader 4 | 34.1(26.6 - 42.4) | 76(71.4 - 80) | |||||
| HELIOS | 31.9(24.6 - 40.1) | 79.7(75.4 - 83.5) | |||||
| Site3 | Manual | Reader 5 | 32.6(25.3 - 40.9) | 72(67.3 - 76.3) | CombinedReaders | 33.3(28 - 39.2) | 75.5(72.3 - 78.4) |
| Reader 6 | 34.1(26.6 - 42.4) | 78.9(74.5 - 82.8) | |||||
| ReaderConfirmationat HELIOS | Reader 5 | 34.8(27.3 - 43.2) | 72.3(67.5 - 76.6) | CombinedReaders | 35.6(30.1 - 41.4) | 74.7(71.4 - 77.6) | |
| Reader 6 | 36.3(28.7 - 44.7) | 77.1(72.5 - 81) | |||||
| HELIOS | 28.9(21.9 - 37) | 80.8(76.5 - 84.5) |
Table 119: Diagnostic Sensitivity and Specificity ANCA Ethanol
Note: The 120 characterized MPO/PR3/ANCA positive sera have been excluded from the calculation of diagnostic sensitivity and specificity.
Summary of Results:
Positive, Negative, Overall and Pattern Agreements for Method Comparison between Method C (Manual) and Method B (Reader Confirmation of HELIOS images) were > 80%. All acceptance criteria were met.
Positive Agreement of Method C vs B ranged from 85.1% to 92.8%.
{91}------------------------------------------------
Negative Agreement of Method C vs B ranged 91.3% to 99.8%.
Overall Agreement of Method C vs B ranged from 89.4% to 95.7%.
Pattern Agreement of Method C vs B ranged from 82.5% to 89.2%.
The results show that Methods C (Manual) and B (Reader Confirmation of HELIOS images) give equivalent results. Furthermore, the results of the 3 study site were comparable with each other.
Agreements for Comparison between Method B (Reader Confirmation of HELIOS images) and Method A (HELIOS positive/negative classification by the Vasculitis Pattern Plus software) were as follows:
Positive Agreement of Method B vs A ranged from 79.0% to 92.6%.
Negative Agreement of Method B vs A ranged from 98.3% to 99.3%.
Overall Agreement of Method B vs A ranged from 92.2% to 96.8%.
Pattern Agreement of Method B vs A ranged from 81.0% to 85.6%.
According to this, acceptance criteria for comparison of Method B vs A were met. The same holds true for the comparison of Method C (Manual) vs A (HELIOS suggested results).
Positive Agreement of Method C vs A ranged from 70.7% to 83.2%.
Negative Agreement of Method C vs A ranged from 94.5% to 99.1%.
Overall Agreement of Method C vs A ranged from 89.7% to 93.0%.
Pattern Agreement of Method C vs A ranged from 77.8% to 81.8%.
Overall, the crossreactivity table (table No.96) shows few crossreactivities for Method A, B and C. The most crossreactivities could be seen in the autoimmune liver disease group (AIH, PBC, PSC: up to 66.7%) and in inflammatory bowel diseases (Ulcerative Colitis, Crohn's disease: up to 45.0%). However, these values are in accordance with literature, where it is stated that ANCAs could also be associated with inflammatory bowel disease and autoimmune liver disease.
{92}------------------------------------------------
Results Clinical study and method comparison ANCA Formalin
Table 120: Method Comparison of Method A, B and C_ANCA Formalin_Single Reader Agreements
| % Agreement (95% CI) | Method C vs B | Method B vs A | Method C vs A | ||||
|---|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | Reader 1 | Reader 2 | Reader 1 | Reader 2 | ||
| Site 1 | PositiveAgreement | 87.5(80.9 - 92) | 85.7(79.1 - 90.5) | 84.9(78.2 - 89.8) | 82.2(75.2 - 87.5) | 75(67.1 - 81.5) | 71.4(63.7 - 78.1) |
| NegativeAgreement | 94.5(92.2 - 96.2) | 95.9(93.7 - 97.3) | 99.2(97.9 - 99.7) | 98.3(96.8 - 99.2) | 94.7(92.4 - 96.4) | 95.2(93 - 96.8) | |
| TotalAgreement | 93(90.8 - 94.8) | 93.5(91.3 - 95.2) | 95.9(94 - 97.2) | 94.6(92.6 - 96.1) | 90.5(87.9 - 92.5) | 89.7(87.1 - 91.8) | |
| Site 2 | PositiveAgreement | 86.8(79.9 - 91.6) | 87(81 - 91.4) | 88.7(81.9 - 93.2) | 73.6(66.7 - 79.5) | 79.8(72.1 - 85.9) | 71(63.6 - 77.4) |
| NegativeAgreement | 97.6(95.9 - 98.6) | 92.1(89.3 - 94.2) | 94.1(91.6 - 95.8) | 98(96.3 - 98.9) | 92.6(90 - 94.6) | 94.6(92.2 - 96.3) | |
| TotalAgreement | 95.4(93.5 - 96.8) | 90.8(88.3 - 92.8) | 93(90.7 - 94.7) | 91.1(88.6 - 93.1) | 90(87.4 - 92.1) | 88.6(85.8 - 90.8) | |
| Site 3 | PositiveAgreement | 89.6(83.8 - 93.5) | 88.6(83.2 - 92.4) | 99.4(96.6 - 99.9) | 98.7(96.1 - 99.5) | 98.7(95.4 - 99.6) | 92.9(88.3 - 95.8) |
| NegativeAgreement | 94.9(92.6 - 96.6) | 86.3(82.8 - 89.2) | 72.2(67.9 - 76) | 82.7(78.7 - 86.1) | 70.7(66.5 - 74.6) | 73(68.7 - 76.9) | |
| TotalAgreement | 93.6(91.5 - 95.3) | 87(84.1 - 89.4) | 79.2(75.8 - 82.2) | 88.4(85.7 - 90.7) | 77.6(74.2 - 80.7) | 78.9(75.5 - 81.9) |
Table 121: Method Comparison of Method A, B and C_ANCA Formalin_Combined Reader Agreements
| % Agreement(95% CI) | Method C vs B | Method B vs A | Method C vs A | |
|---|---|---|---|---|
| Site 1 | PositiveAgreement | 86.6(82.1 - 90.1) | 83.6(78.9 - 87.4) | 73.1(67.7 - 78) |
| NegativeAgreement | 95.2(93.7 - 96.4) | 98.8(97.8 - 99.3) | 95(93.4 - 96.2) | |
| TotalAgreement | 93.3(91.7 - 94.5) | 95.2(93.9 - 96.3) | 90.1(88.3 - 91.6) | |
| Site 2 | PositiveAgreement | 86.9(82.6 - 90.3) | 79.8(74.9 - 83.9) | 74.9(69.6 - 79.5) |
| NegativeAgreement | 94.9(93.4 - 96.1) | 95.9(94.5 - 97) | 93.6(91.9 - 95) | |
| TotalAgreement | 93.1(91.5 - 94.4) | 92.1(90.4 - 93.4) | 89.3(87.4 - 90.9) | |
| Site 3 | PositiveAgreement | 89.1(85.3 - 92) | 99(97.4 - 99.6) | 95.6(92.8 - 97.3) |
| NegativeAgreement | 90.8(88.7 - 92.5) | 77.1(74.2 - 79.7) | 71.8(68.9 - 74.7) | |
| TotalAgreement | 90.3(88.5 - 91.8) | 83.8(81.6 - 85.7) | 78.2(75.9 - 80.4) |
{93}------------------------------------------------
Table 122: Method Comparison of Method A, B and C_ANCA Formalin_Pattern Agreements
| % Agreement (95% CI) | Method C vs B | Method B vs A | Method C vs A |
|---|---|---|---|
| Site 1 | 82.1(79.9 - 84.2) | 90.8(89.1 - 92.3) | 77.9(75.5 - 80.1) |
| Site 2 | 87.5(85.5 - 89.2) | 84(81.9 - 86) | 79.6(77.3 - 81.7) |
| Site 3 | 82.1(79.8 - 84.1) | 76.1(73.7 - 78.4) | 69.7(67.1 - 72.2) |
| All sites | 83.9(82.7 - 85) | 83.7(82.4 - 84.8) | 75.7(74.3 - 77.1) |
{94}------------------------------------------------
Image /page/94/Picture/0 description: The image contains the logo for AESKU.DIAGNOSTICS. The logo features a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font size.
AESKUSLIDES ANCA 510(k) Traditional submission
Table 123: Method comparison. Number and percent positive samples in the different tested diagnosis_ANCA Formalin
| Site 1Site 2 | Site 3 | ||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Cross Reactivity | Manual Performance | Reader Confirmation atHELIOS | HELIOS | Manual Performance | Reader Confirmation at | HELIOS | HELIOS | Manual Performance | Reader Confirmation atHELIOS | HELIOSsuggestion | |||||||||||||||||||||
| Reader 1 | Reader 2 | Reader 1 | Reader 2 | suggestion | Reader 3 | Reader 4 | Reader 3 | Reader 4 | suggestion | Reader 5Reader 6 | Reader 5 | Reader 6 | |||||||||||||||||||
| Diagnosis | n | n Pos | %Pos | nPos | % Pos | nPos | %Pos | n Pos | % Pos | n Pos | %Pos | 0Pos | %Pos | nPos | 0%Pos | 0Pos | %Pos | nPos | %Pos | n Pos | %Pos | n Pos | %Pos | n Pos | % Pos | n Pos | % Pos | nPos | %Pos | n Pos | %Pos |
| Wegener's | 80 | રૂપ | 48.8 | 42 | 52.5 | 45 | 56.3 | 43 | 53.8 | 41 | 51.3 | રૂદિ | 45.0 | રૂત્વ | 48.8 | 31 | 38.8 | 37 | 46.3 | 37 | 46.3 | 38 | 47.5 | 51 | 63.8 | 38 | 47.5 | 50 | 62.5 | 56 | 70.0 |
| MPA | 29 | 13 | 44.8 | 12 | 41.4 | 15 | 51.7 | 14 | 48.3 | 14 | 48.3 | 11 | 37.9 | 16 | 55.2 | 13 | 44.8 | 17 | 58.6 | 14 | 48.3 | 12 | 41.4 | 13 | 44.8 | 13 | 44.8 | 14 | 48.3 | 19 | 65.5 |
| Churg-Strauss | 26 | 2 | 7.7 | 3 | 11.5 | 7 | 26.9 | 4 | 15.4 | ઉ | 23.1 | 15.4 | 5 | 19.2 | 5 | 19.2 | 34.6 | 6 | 23.1 | 3.8 | 1 | 3.8 | 2 | 7.7 | प | 15.4 | 11 | 42.3 | |||
| C-ANCA | 20 | 14 | 70.0 | 15 | 75.0 | 14 | 70.0 | 15 | 75.0 | 14 | 70.0 | 15 | 75.0 | 17 | 85.0 | 15 | 75.0 | 15 | 75.0 | 13 | 65.0 | 19 | 95.0 | 19 | 95.0 | 19 | 95.0 | 18 | 95.0 | 20 | 100 |
| P-ANCA | 50 | 20 | 40.0 | 20 | 40.0 | 18 | 36.0 | 19 | 38.0 | 17 | 34.0 | 17 | 34.0 | 18 | 36.0 | 16 | 32.0 | 23 | 46.0 | 20 | 40.0 | 22 | 44.0 | 25 | 50.0 | 20 | 40.0 | 25 | 50.0 | 32 | 64.0 |
| MPO | 25 | 8 | 32.0 | 7 | 28.0 | 8 | 32.0 | 7 | 28.0 | 3 | 12.0 | 28.0 | 15 | 60.0 | 28.0 | 68.0 | 12 | 48.0 | 19 | 76.0 | 19 | 76.0 | 16 | 64.0 | 18 | 76.0 | 24 | 96.0 | |||
| PR3 | 25 | 14 | 56.0 | 16 | 64.0 | 14 | 56.0 | 14 | 56.0 | 15 | 60.0 | ਤ | 52.0 | 16 | 64.0 | 13 | 52.0 | 16 | 64.0 | 17 | 68.0 | 19 | 76.0 | 20 | 80.0 | 15 | 60.0 | 20 | 80.0 | 23 | 92.0 |
| AlH | 8 | 4 | 50.0 | 2 | 25.0 | ব | 50.0 | 2 | 25.0 | 2 | 25.0 | 37.5 | 3 | 37.5 | 62.5 | 50.0 | 3 | 37.5 | 2 | 25.0 | 2 | 25.0 | 3 | 37.5 | 37.5 | ব | 50.0 | ||||
| AIH/PBC | ଚ | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | O | 0.0 | 0 | 0.0 | 16.7 | 0.0 | 16.7 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | ব | 66.7 | |||||
| РВС | 11 | O | 0.0 | O | 0.0 | 0 | 0.0 | O | 0.0 | 0 | 0.0 | 18.2 | 0.0 | 18.2 | 0.0 | 0 | 0.0 | 2 | 18.2 | 2 | 18.2 | 2 | 18.2 | 27.3 | 4 | 36.4 | |||||
| PSC | 10 | O | 0.0 | 1 | 10.0 | 0 | 0.0 | 1 | 10.0 | 0 | 0.0 | 30.0 | ব | 40.0 | 40.0 | 30.0 | O | 0.0 | 3 | 30.0 | 3 | 30.0 | 3 | 30.0 | 70.0 | 8 | 80.0 | ||||
| UlcerativeColitis | 71 | 6 | 8.5 | 8 | 11.3 | ব | 5.6 | 8 | 11.3 | प | 5.6 | 5 | 7.0 | 7 | 0.9 | ব | 5.6 | 5.6 | 5 | 7.0 | 7 | ರಿ. ಇ | 9 | 12.7 | 14 | 19.7 | 27 | 38.0 | 37 | 52.1 | |
| Crohn`sDisease | 40 | O | 0.0 | 2 | 5.0 | 0 | 0.0 | വ | 5.0 | 0 | 0.0 | 2.5 | 2 | 5.0 | 0 | 0.0 | 2 | 5.0 | 0 | 0.0 | O | 0.0 | 3 | 7.5 | 2 | 5.0 | 11 | 27.5 | 13 | 32.5 | |
| IBD | 9 | 0 | 0.0 | 1 | 11.1 | 0 | 0.0 | 1 | 11.1 | 0 | 0.0 | 0.0 | 11.1 | 0.0 | 11.1 | 0 | 0.0 | 0 | 0.0 | 11.1 | 0 | 0.0 | 11.1 | 1 | 11.1 | ||||||
| RA | 12 | 1 | 8.3 | 1 | 8.3 | 8.3 | 1 | 8.3 | 1 | 8.3 | 0.0 | 0 | 0.0 | () | 0.0 | 0.0 | 8.3 | 0 | 0.0 | 8.3 | O | 0.0 | 0.0 | 0 | 0.0 | ||||||
| SLE | 30 | 5 | 16.7 | র্ব | 13.3 | 5 | 16.7 | 3 | 10.0 | 3 | 10.0 | 5 | 16.7 | 23.3 | 5 | 16.7 | 23.3 | 5 | 16.7 | 2 | 6.7 | प | 13.3 | 5 | 16.7 | 23.3 | 6 | 20.0 | |||
| Myositis | 1 | O | 0.0 | O | 0.0 | O | 0.0 | O | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | O | 0.0 | 0.0 | O | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | |||
| Scleroderma | 19 | 3 | 15.8 | 3 | 15.8 | 21.1 | 2 | 10.5 | 3 | 15.8 | 0.0 | 5.3 | 5.3 | 21.1 | 2 | 10.5 | 21.1 | 3 | 15.8 | 4 | 21.1 | 3 | 15.8 | 8 | 42.1 | ||||||
| HBV | 8 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | O | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | |||
| HCV | 16 | త | 18.8 | 2 | 12.5 | 3 | 18.8 | വ | 12.5 | 1 | 6.3 | 6.3 | 2 | 12.5 | 6.3 | 2 | 12.5 | 3 | 18.8 | 2 | 12.5 | ব | 25.0 | 2 | 12.5 | 25.0 | ട | 37.5 | |||
| GCA | イ | 1 | 25.0 | 1 | 25.0 | 25.0 | 1 | 25.0 | 25.0 | 0.0 | 0 | 0.0 | () | 0.0 | 0.0 | O | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | 1 | 25.0 | |||||
| PMR | 18 | 1 | 5.6 | 1 | 5.6 | 5.6 | 1 | 5.6 | 1 | 5.6 | 16.7 | 3 | 16.7 | 5.6 | 2 | 11.1 | 5.6 | O | 0.0 | 0 | 0.0 | 0 | 0.0 | 0.0 | 0 | 0.0 | |||||
| Purpura | 1 | O | 0.0 | O | 0.0 | O | 0.0 | O | 0.0 | 0 | 0.0 | 0.0 | 100 | O | 0.0 | 100 | O | 0.0 | 100 | 100 | 1 | 100 | 100 | 1 | 100 | ||||||
| Leukemia | 15 | O | 0.0 | O | 0.0 | 0 | 0.0 | O | 0.0 | 0 | 0.0 | 0 | 0.0 | 6.7 | O | 0.0 | 6.7 | 6.7 | 0 | 0.0 | 2 | 13.3 | 2 | 13.3 | ব | 26.7 | 7 | 46.7 | |||
| Celiacdisease | 10 | 1 | 10.0 | 1 | 10.0 | 10.0 | 1 | 10.0 | 10.0 | 0.0 | 10.0 | O | 0.0 | 30.0 | O | 0.0 | O | 0.0 | 0 | 0.0 | O | 0.0 | O | 0.0 | 3 | 30.0 | |||||
| CKD | 32 | O | 0.0 | 3 | 9.4 | 0 | 0.0 | 3 | 9.4 | 0 | 0.0 | 3.1 | 2 | 6.3 | 0.0 | 9.4 | 0 | 0.0 | 3.1 | 3.1 | O | 0.0 | 3.1 | 2 | 6.3 | ||||||
| Asthma | 32 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | 0 | 0.0 | 0 | 0.0 | O | 0.0 | 3.1 | O | 0.0 | O | 0.0 | 0 | 0.0 | 1 | 3.1 | 1 | 3.1 | 1 | 3.1 | |
| Sinusitis | 22 | O | 0.0 | 1 | 4.5 | 0 | 0.0 | 1 | 4.5 | 0 | 0.0 | 4.5 | 4.5 | () | 0.0 | 6 | 27.3 | O | 0.0 | O | 0.0 | O | 0.0 | 0 | 0.0 | O | 0.0 | 0 | 0.0 |
{95}------------------------------------------------
Image /page/95/Picture/0 description: The image shows the logo for AESKU.DIAGNOSTICS. The logo consists of a green square with a white symbol inside, followed by the text "AESKU.DIAGNOSTICS" in gray. Below the text, there is a tagline that reads "THE DIAGNOSTIC TOOL THAT WORKS" in a lighter shade of gray.
AESKUSLIDES ANCA 510(k) Traditional submission
| Table 124: Diagnostic Sensitivity and Specificity_ANCA Formalin | ||||||||
|---|---|---|---|---|---|---|---|---|
| % Diagnostic Sensitivity & Specificity(95% CI) | % Sensitivity(95% CI) | % Specificity(95% CI) | ||||||
| AAV (n= 135) | OD (n=375) | AAV (n= 135) | OD (n=375) | |||||
| Site 1 | Manual | Reader 1 | 40(32.1 - 48.4) | 93.1(90 - 95.2) | CombinedReaders | 41.1(35.4 - 47.1) | 92.3(90.1 - 94) | |
| Reader 2 | 42.2(34.2 - 50.7) | 91.5(88.2 - 93.9) | ||||||
| ReaderConfirmationat HELIOS | Reader 1 | 49.6(41.3 - 58) | 93.3(90.3 - 95.4) | CombinedReaders | 47.4(41.5 - 53.4) | 92.7(90.6 - 94.3) | ||
| Reader 2 | 45.2(37 - 53.6) | 92(88.8 - 94.3) | ||||||
| HELIOS | 45.2(37 - 53.6) | 95.2(92.5 - 96.9) | ||||||
| Site 2 | Manual | Reader 3 | 37.8(30 - 46.2) | 93.1(90 - 95.2) | CombinedReaders | 41.1(35.4 - 47.1) | 91.7(89.5 - 93.5) | |
| Reader 4 | 44.4(36.3 - 52.9) | 90.4(87 - 93) | ||||||
| ReaderConfirmationat HELIOS | Reader 3 | 36.3(28.7 - 44.7) | 93.6(90.7 - 95.7) | CombinedReaders | 41.5(35.8 - 47.4) | 90.9(88.7 - 92.8) | ||
| Reader 4 | 46.7(38.5 - 55.1) | 88.3(84.6 - 91.1) | ||||||
| HELIOS | 42.2(34.2 - 50.7) | 94.4(91.6 - 96.3) | ||||||
| Site 3 | Manual | Reader 5 | 37.8(30 - 46.2) | 93.6(90.7 - 95.7) | CombinedReaders | 43(37.2 - 48.9) | 92(89.8 - 93.7) | |
| Reader 6 | 48.1(39.9 - 56.5) | 90.4(87 - 93) | ||||||
| ReaderConfirmationat HELIOS | Reader 5 | 39.3(31.4 - 47.7) | 89.6(86.1 - 92.3) | CombinedReaders | 44.8(39 - 50.8) | 85.1(82.3 - 87.4) | ||
| Reader 6 | 50.4(42 - 58.7) | 80.5(76.2 - 84.2) | ||||||
| HELIOS | 63.7(55.3 - 71.3) | 71.7(67 - 76.1) |
Table 124: Diagnostic Sensitivity and Specificity_ANCA Formalin
Note: The 120 characterized MPO/PR3/ANCA positive sera have been excluded from the calculation of
diagnostic sensitivity and specificity.
{96}------------------------------------------------
Image /page/96/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image inside of what appears to be a vest. To the right of the image is the company name in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below.
Summary of Results:
Positive, Negative, Overall and Pattern Agreements for Method Comparison between Method C (Manual) and Method B (Reader Confirmation of HELIOS images) were > 80%. All acceptance criteria were met.
Positive Agreement of Method C vs B ranged from 85.7% to 89.6%.
Negative Agreement of Method C vs B ranged 86.3% to 97.6%.
Overall Agreement of Method C vs B ranged from 87.0% to 95.4%.
Pattern Agreement of Method C vs B ranged from 82.1% to 87.5%.
The results show that Methods C (Manual) and B (Reader Confirmation of HELIOS images) give equivalent results. Furthermore, the results of the 3 study site were comparable with each other.
Agreements for Comparison between Method B (Reader Confirmation of HELIOS images) and Method A (HELIOS positive/negative classification by the Vasculitis Pattern Plus software) were as follows:
Positive Agreement of Method B vs A ranged from 73.6% to 99.4%.
Negative Agreement of Method B vs A ranged from 72.2% to 99.2%.
Overall Agreement of Method B vs A ranged from 79.2% to 95.9%.
Pattern Agreement of Method B vs A ranged from 76.1% to 90.8%.
According to this, acceptance criteria for comparison of Method B vs A were met.
The comparison of Method C (Manual) vs A (HELIOS suggested results) gave good results within the acceptance criteria for positive, negative and overall agreement. In the case of pattern agreement, 69.7% was reached. This is just below the acceptance criteria and we tolerate the deviation since this is a slight and unique occurrence.
Positive Agreement of Method C vs A ranged from 71.0% to 98.7%.
Negative Agreement of Method C vs A ranged from 70.7% to 95.2%.
Overall Agreement of Method C vs A ranged from 77.6% to 90.5%.
Pattern Agreement of Method C vs A ranged from 69.7% to 79.6%.
The Cross Reactivity Table (table no.: 101) shows that there were only few crossreactivities for Method A, B and C, which is also reflected by the high specificity of this methods (90%). The most crossreactivities could be seen in the Autoimmune Liver Disease Group (AIH, PBC, and PCS: up to 62.5%), in the inflammatory bowel disease group (Crohn' Disease, Ulcerative Colitis, IBD: up to 11.3%) and in SLE (up to 23%). These results reflect values reported in the literature. Method A (site 3) has a lower specificity as described and therefore, higher crossreactivities can be seen.
{97}------------------------------------------------
Conclusion:
Method C and Method B show substantial equivalence. AESKUSLIDES ANCA Ethanol and Formalin perform substantially equivalent when processed manually and on HELIOS with Reader Confirmation of HELIOS images, as shown by the method comparison study, the calculated diagnostic sensitivities and specificities and the cross reactivity table.
For Method A we found a lower agreement compared to Method B and C. This can be explained by the complex patterns, which can be distinguished by the readers at the microscope or at the screen, but to limited extent by the software.
Our results show that results generated at HELIOS must always be confirmed by trained readers. The Vasculitis Pattern Plus software can support the reader to preselect positive samples, regardless of the detected pattern. According to the international recommendations on ANCA testing and the literature, it is recommended to further characterize ANCA positive IFA results by another method, for example PR3- and/or MPO-ELISA.
Endpoint Titer comparison:
To evaluate Within-Lab Precision and Between-Lab Precision of endpoint titer determination on AESKUSLIDES ANCA Ethanol and ANCA Formalin processed on HELIOS and processed manually, based on CLSI Guideline EP12-A2.
Procedure:
Five serum samples with endpoint titers between 1:10 and 1:320 were diluted serially, ranging from 1:10 up to 1:1280 sample dilution, depending on the expected enpoint titer displayed in the table below. For each sample at least 4 different dilutions around the expected endpoint titer were made. Those serum dilutions were assayed on AESKUSLIDES ANCA Ethanol and Formalin and analyzed on HELIOS (Method B, Reader Confirmation) and manually at the microscope (Method C) by two independent readers. For each of the two methods, each serum dilution was assayed on five days, with two runs per day and three replicates per run, resulting in 30 repetitions per serum dilution per reader.
| SampleID | Pos/Neg | PatternANCA Ethanol | Endpoint titerANCA Ethanol | PatternANCA Formalin | Endpoint titerANCAFormalin |
|---|---|---|---|---|---|
| 1 | Pos | C-ANCA | 10 | C-ANCA | 20 |
| 2 | Pos | C-ANCA | 160 | C-ANCA | 160 |
| 3 | Pos | C-ANCA | 160 | C-ANCA | 320 |
| 4 | Pos | P-ANCA | 80 | C-ANCA | 20 |
| 5 | Pos | P-ANCA | 320 | C-ANCA | 40 |
All runs have been performed according to the respective IFUs. 1 positive and 1 negative control (kit controls) were included in each run.
The described study has been conducted at three different study sites to allow the calculation of Between-Lab Precision.
{98}------------------------------------------------
Data Analysis
Positive/Negative Classification and Pattern were recorded for each sample dilution in each run and for each Method. For Method C (Manual performance) fluorescence intensity was also reported.
Determination of endpoint titer:
For each replicate of the dilution series endpoint titer was reported as the reciprocal of the highest dilution at which the sample was classified as positive.
After pairwise comparison of the datapoints Within-lab precision was calculated for Methods B and C, for each of the three study sites.
Data will be presented as
- percentage of samples that differ maximum ±1 titer level from each . other and
- . percentage of samples that are > ±1 titer level away from each other.
For Between-Lab Precision the titer agreements (including 95% CI) between the three study sites were determined for each Method B and C (site 1 vs site 2, site 2 vs site 3, site 1 vs site 3).
Acceptance Criteria
- . For each serum reported endpoint titers are allowed to differ maximum ±1 titer level from each other
- . Calculations should meet the following criteria:
| Calculations | Method B | Method C |
|---|---|---|
| Percentage of samples that differ maximum ±1 titer level | ≥ 90% | ≥ 90% |
| Titer Agreement | ≥ 80% | ≥ 80% |
Results - Within-Lab Precision of endpoint titer determination
Results of endpoint titer determination for each of the three replicates per sample dilution can be found in the appendix.
For calculation of Within-Lab Precision the percentages of samples that are within and outside +/-1 titer level were calculated for the two readers separately (30 replicates for each sample dilution) and for all readers combined. The results of Within-Lab Precision for study sites 1, 2 and 3 are presented in the tables below.
{99}------------------------------------------------
AESKUSLIDES ANCA Ethanol
Table 125a: Number and percentages of samples that are within and outside +/-1 titer level – Method B and C
| AESKUSLIDESANCA Ethanol -Method B | Number ofsamples within+/-1 titer level | Number ofsamples> +/-1 titer level | Totalnumber | Percentage ofsamples thatare +/-1 titerlevel | Percentage ofsamples> +/-1 titer level | |
|---|---|---|---|---|---|---|
| Site 1 | Reader 1 | 147 | 3 | 150 | 98.0% | 2.0% |
| Reader 2 | 139 | 11 | 150 | 92.7% | 7.3% | |
| Site 2 | Reader 3 | 139 | 11 | 150 | 92.7% | 7.3% |
| Reader 4 | 138 | 12 | 150 | 92.0% | 8.0% | |
| Site 3 | Reader 5 | 149 | 1 | 150 | 99.3% | 0.7% |
| Reader 6 | 144 | 6 | 150 | 96.0% | 4.0% | |
| All readers combined | 856 | 44 | 900 | 95.1% | 4.9% |
| AESKUSLIDESANCA Ethanol -Method C | Number ofsamples within+/-1 titer level | Number ofsamples> +/-1 titerlevel | Totalnumber | Percentage ofsamples thatare +/-1 titerlevel | Percentage ofsamples> +/-1 titer level | |
|---|---|---|---|---|---|---|
| Site 1 | Reader 1 | 143 | 7 | 150 | 95.3% | 4.7% |
| Reader 2 | 144 | 6 | 150 | 96.0% | 4.0% | |
| Site 2 | Reader 3 | 143 | 7 | 150 | 95.3% | 4.7% |
| Reader 4 | 142 | 8 | 150 | 94.7% | 5.3% | |
| Site 3 | Reader 5 | 146 | 4 | 150 | 97.3% | 2.7% |
| Reader 6 | 141 | 9 | 150 | 94.0% | 6.0% | |
| All readers combined | 859 | 41 | 900 | 95.4% | 4.6% |
Acceptance criteria for Within-Lab Precision of endpoint titer determination on ANCA Ethanol were met. Percentage of samples that were within +/-1 titer level was 95.1% for Method B and 95.4% for Method C, for combined readers from all study sites. Considering all three study sites and readers separately, percentage of samples that were within +/-1 titer level ranged from 92.0% to 99.3% for Method B (Reader Confirmation) and from 94.0% to 97.3% for Method C (Manual). Results for Methods B and C were comparable.
Pos/Neg classification and pattern determination were consistent for 100% of the replicates.
{100}------------------------------------------------
AESKUSLIDES ANCA Formalin
Table 125b: Number and percentages of samples that are within and outside +/-1 titer level – Method B and C
| AESKUSLIDESANCA Formalin -Method B | Number ofsamples within+/-1 titer level | Number ofsamples> +/-1 titer level | Totalnumber | Percentage ofsamples that are+/-1 titer level | Percentage ofsamples> +/-1 titer level | |
|---|---|---|---|---|---|---|
| Site 1 | Reader 1 | 145 | 5 | 150 | 96.7% | 3.3% |
| Reader 2 | 145 | 5 | 150 | 96.7% | 3.3% | |
| Site 2 | Reader 3 | 147 | 3 | 150 | 98.0% | 2.0% |
| Reader 4 | 135 | 15 | 150 | 90.0% | 10.0% | |
| Site 3 | Reader 5 | 145 | 5 | 150 | 96.7% | 3.3% |
| Reader 6 | 135 | 15 | 150 | 90.0% | 10.0% | |
| All readers combined | 852 | 48 | 900 | 94.7% | 5.3% |
| AESKUSLIDESANCA Formalin -Method C | Number ofsamples within+/-1 titer level | Number ofsamples> +/-1 titer level | Totalnumber | Percentage ofsamples that are+/-1 titer level | Percentage ofsamples> +/-1 titer level | |
|---|---|---|---|---|---|---|
| Site 1 | Reader 1 | 145 | 5 | 150 | 96.7% | 3.3% |
| Reader 2 | 148 | 2 | 150 | 98.7% | 1.3% | |
| Site 2 | Reader 3 | 145 | 5 | 150 | 96.7% | 3.3% |
| Reader 4 | 144 | 6 | 150 | 96.0% | 4.0% | |
| Site 3 | Reader 5 | 149 | 1 | 150 | 99.3% | 0.7% |
| Reader 6 | 134 | 16 | 150 | 89.3% | 10.7% | |
| All readers combined | 865 | 35 | 900 | 96.1% | 3.9% |
Acceptance criteria for Within-Lab Precision of endpoint titer determination on ANCA Formalin were met. Percentage of samples that were within +/-1 titer level was 94.7% for Method B and 96.1% for Method C, for combined readers from all study sites.
Considering all three study sites and readers separately, percentage of samples that were within +/-1 titer level ranged from 90.0% to 98.0% for Method B (Reader Confirmation) and from 89.3% to 99.3 % for Method C (Manual). Only for reader 6 in Method C the percentage of samples was slightly below our acceptance criterion (89.3% instead of 90%). We tolerate this deviation, since this only concerns one reader out of six. However, this result shows that positive classification of samples near its endpoint titer is subjective and highly depends on the experience of the reader.
Results for Methods B and C were comparable.
Pos/Neg classification and pattern determination were consistent for 100% of the replicates.
Results - Between-Lab Precision of endpoint titer determination
In order to evaluate Between-Lab Precision of endpoint titer determination, titer agreements (including 95% Cls) were determined between the three sites (site 1 vs site 2, site 2 vs site 3, site 1 vs site 3) for each Method B and C. Results are displayed in the following tables.
{101}------------------------------------------------
AESKUSLIDES ANCA Ethanol
Table 126a: % Titer agreement between three study sites – Method B and C
| % Titer agreement(95% CI) | Site 1 vs. Site 2 | Site 1 vs. Site 3 | Site 2 vs. Site 3 |
|---|---|---|---|
| Method B(ReaderConfirmation) | 93.3(89.9 - 95.6) | 86.7(82.4 - 90.1) | 82.3(77.6 - 86.2) |
| Method C(Manual) | 98.3(96.2 - 99.3) | 88.7(84.6 - 91.8) | 87(82.7 - 90.3) |
For Between-lab Precision of endpoint titer determination on ANCA Ethanol, all acceptance criteria were met. Titer agreements comparing the three study sites ranged from 82.3% to 93.3% for Method B (Reader Confirmation) and from 87.0% to 98.3% for Method C (Manual). With manual reading of the slides (Method C) slightly higher titer agreements have been achieved.
AESKUSLIDES ANCA Formalin
Table 126b: % Titer agreement between three study sites - Method B and C
| % Titer agreement(95% CI) | Site 1 vs. Site 2 | Site 1 vs. Site 3 | Site 2 vs. Site 3 |
|---|---|---|---|
| Method B(ReaderConfirmation) | 86.7(82.4 - 90.1) | 79(74 - 83.2) | 82.7(78 - 86.5) |
| Method C(Manual) | 93.7(90.3 - 95.9) | 80.3(75.5 - 84.4) | 82.7(78 - 86.5) |
Titer agreement between the three study sites ranged from 79.0% to 86.7% for Method B (Reader Confirmation) and from 80.3% to 93.7% for Method C (Manual). For Between-Lab Precision of endpoint titer determination on ANCA Formalin, all acceptance criteria were met, except for Method B site 1 vs. site 3. Here, titer agreement was slightly below the acceptance criterion (79.0% instead of 80%). However, we tolerate this deviation, since at study site 3, consistently higher endpoint titers have been obtained than at study sites 1 and 2. This is also reflected by the generally lower agreements when comparing site 3 to the other study sites. This study shows that the evaluation of sample titer is subjective and can vary from laboratory to laboratory, from microscope to microscope. Thus, we suggest for each laboratory to perform a titer control using the positive control or other well-known samples to balance variations in endpoint titer determination.
Results for Methods B and C were comparable when contrasting Site 1 vs Site 3 and Site 2 vs Site 3. When comparing Site 1 vs Site 2 titer agreements were slightly higher for Method C (Manual).
{102}------------------------------------------------
Conclusion
Pre-determined acceptance criteria were met. Method B (Reader Confirmation) as well as Method C (Manual) showed a good and acceptable repeatability (Within-Lab Precision) and reproducibility (Between-Lab Precision) for enpoint titer determination.
3. Clinical cut-off:
See assay cut-off
4. Expected values/Reference range
Expected values for ANCA were analyzed with a panel of 150 sera from healthy donors: 100 from Germany (56 females,44 males, mean age 31 years, range of 18-58 years, Ethnicity: German) and 50 from the US (17 females, 33 males, mean age 34 years, range of 19-62 years, Ethnicity: 30 Black, 10 Hispanic, 10 White).
Slides were processed manually according to the IFU and subsequently analyzed at the microscope by two independent readers.
Results:
With AESKUSLIDES ANCA Ethanol, there were 6 positive results for reader 1 and 3 positive results for reader 2, and with AESKUSLIDES ANCA Formalin, there were 6 positive results for reader 1 and 4 positive results for reader 2.
| Table 127. Normal Range Study, number and percent positive and negative samples. | ||||||
|---|---|---|---|---|---|---|
| Normal RangeStudy | ANCA Ethanol | ANCA Formalin | ||||
| Pos (%) | Neg (%) | Total | Pos (%) | Neg (%) | Total | |
| Reader 1 | 6 (4) | 144 (96) | 150 | 6 (4) | 144 (96) | 150 |
| Reader 2 | 3 (2) | 147 (98) | 150 | 4 (2.7) | 146 (97.3) | 150 |
Table 127: Normal Range Study, number and percent positive and negative samples:
{103}------------------------------------------------
| Normal Range Study | ANCA Ethanol | ANCA Formalin | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample ID | Sample Origin | Reader 1 | Reader 2 | Reader 1 | Reader 2 | ||||
| Pos/Neg | ANCA Pattern FI | Pos/Neg | ANCA Pattern FI | Pos/Neg | ANCA Pattern FI | Pos/Neg | ANCA Pattern FI | ||
| NC 008 | USA | Pos | A 2 | Pos | A 2 | Pos | C 1 | Pos | C 1 |
| NC 010 | USA | Neg | Neg 0 | Neg | Neg 0 | Neg | Neg 0 | Pos | C 1 |
| NC 028 | USA | Pos | P 1 | Pos | P 1 | Neg | Neg 0 | Neg | Neg 0 |
| NC 085 | Germany | Pos | A 1 | Pos | A 1 | Neg | Neg 0 | Neg | Neg 0 |
| NC 109 | Germany | Neg | Neg 0 | Neg | Neg 0 | Pos | C 1 | Pos | C 1 |
| NC 123 | Germany | Pos | C 1 | Neg | Neg 0 | Pos | C 1 | Neg | Neg 0 |
| NC 126 | Germany | Neg | Neg 0 | Neg | Neg 0 | Pos | C 1 | Neg | Neg 0 |
| NC 127 | Germany | Pos | P 1 | Neg | Neg 0 | Neg | Neg 0 | Neg | Neg 0 |
| NC 139 | Germany | Pos | C 1 | Neg | Neg 0 | Pos | C 1 | Neg | Neg 0 |
| NC 144 | Germany | Neg | Neg 0 | Neg | Neg 0 | Pos | C 1 | Pos | C 1 |
Table 128: Normal Range Study, positive samples:
FI = Fluorescence Intensitiy
Only one sample (NC 008) has been found positive in both ANCA Ethanol (FI = 2) and ANCA Formalin (FI = 1) assays by both readers. 2 samples (NC 123, NC 139) have been found positive in both assays (Fl=1) by reader 1 and negative by reader 2. All other samples have been found positive either in the ANCA Ethanol or in the ANCA Formalin assay with an FI of 1.
Conclusion:
The low number of positive samples we found with AESKUSLIDES ANCA Ethanol and ANCA Formalin in the tested healthy population correlate well with numbers reported in the literature.
N. Instrument Name
HELIOS AUTOMATED IFA SYSTEM
O. System Description
The HELIOS AUTOMATED IFA SYSTEM is an automated system including pipetting unit with microscope and software that acquires. stores and displays digital images of stained indirect immunofluorescence slides. The HELIOS DEVICE SOFTWARE is designed to support input of results from the AESKUSLIDES into electronic laboratory data management systems.
The HELIOS AUTOMATED IFA SYSTEM should only be used with AESKUSLIDES assays that are cleared or approved for use on the instrument. All suggested results obtained with the HELIOS DEVICE SOFTWARE must be confirmed by trained personnel.
{104}------------------------------------------------
Image /page/104/Picture/1 description: The image is a logo for AESKU.DIAGNOSTICS. The logo features a green square with a white image of a vest inside. To the right of the square is the text "AESKU.DIAGNOSTICS" in gray, with the words "THE DIAGNOSTIC TOOL THAT WORKS" in a smaller font below.
1. Modes of Operation:
HELIOS: automated processing with automated imaging and automated reading HELIOS User Eval: automated processing with automated imaging and manual reading
HELMED Mode: automated processing with manual imaging and manual reading Manual Mode: manual processing with manual imaging and manual reading
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
No
2. Software:
General
The HELIOS AUTOMATED IFA SYSTEM is an automated system for immunofluorescence processing with an integrated camera with an optic (microscope) and software for routine laboratory use by professional users under controlled environmental conditions.
The HELIOS camera with an optic (microscope) automatically focuses on the areas of biological material on the slides. Pictures are stored and evaluated by the HELIOS DEVICE SOFTWARE. The software performs a positive/negative of the cells. The software suggests the result as a qualitative result (positive. negative). All suggested results obtained with the HELIOS AUTOMATED IFA SYSTEM and the HELIOS DEVICE SOFTWARE must be confirmed by trained personnel before the result is finalized.
Level of Concern
The Level of concern was determined according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11. 2005. Following the recommendations of this standard. the HELIOS DEVICE SOFTWARE is assigned moderate level of concern.
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types.
- Specimen Identification: 3. Manual sample identification and/or Barcode
-
- Specimen Sampling and Handling: Not applicable
{105}------------------------------------------------
5. Calibration:
There is no calibration of the instrument by the user.
6. Quality Control:
Positive and negative controls are supplied with the assay reagents (see assay description above).
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
§ 866.5660 Multiple autoantibodies immunological test system.
(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).