K Number
K172142
Manufacturer
Date Cleared
2017-09-13

(58 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FibroScan® 430 Mini+ system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.

FibroScan® 430 Mini+ is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).

The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.

Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.

Device Description

FibroScan® 430 Mini+, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The FibroScan® 430 Mini+ CAP (Controlled Attenuation Parameter, ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness.

AI/ML Overview

The FibroScan® 430 Mini+ system, a diagnostic ultrasound device, was evaluated for substantial equivalence to the FibroScan® 530 Compact system. The evaluation focused on the device's ability to measure shear wave speed and Controlled Attenuation Parameter (CAP) for liver stiffness and fat assessment, respectively.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the FibroScan® 430 Mini+ were based on demonstrating comparable or better performance (bias and precision) to its predicate device, FibroScan® 530 Compact, when tested on calibrated phantoms.

Performance MetricAcceptance Criteria (based on predicate FibroScan® 530 Compact)Reported Device Performance (FibroScan® 430 Mini+)
Shear Wave Speed BiasM+ probe: <11%M+ probe: ≤12%
XL+ probe: <12%XL+ probe: <11%
CAP BiasM+ probe: <7%(Not explicitly stated for 430 Mini+ but validated by identical acquisition platform)
XL+ probe: <3%(Not explicitly stated for 430 Mini+ but validated by identical acquisition platform)
Shear Wave Speed PrecisionM+ probe: <1%M+ probe: <1% (specifically [0.2%; 0.7%])
XL+ probe: <2%XL+ probe: <1% (specifically [0%; 0.7%])
CAP PrecisionM+ probe: <1%(Not explicitly stated for 430 Mini+ but validated by identical acquisition platform)
XL+ probe: <1%(Not explicitly stated for 430 Mini+ but validated by identical acquisition platform)

The study concluded that the bias and precision of the shear wave speed measured by the FibroScan® 430 Mini+ are within the same range or better than those of the predicate FibroScan® 530 Compact device. The CAP bias and precision were considered valid for the FibroScan 430 Mini+ due to the identical acquisition platform (US board and firmware), ultrasound module, and software core used in both devices.

2. Sample Size Used for the Test Set and Data Provenance

The evaluation was performed on calibrated phantoms. The document does not specify the exact number of phantoms used or the number of measurements taken, but it refers to "calibrated phantoms with known elasticity and attenuation."

The data provenance is from non-clinical testing in a controlled environment, not from human subjects or a specific country of origin. This implies it was performed as part of product development and validation by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth was established using calibrated phantoms with known elasticity and attenuation. This implies that the ground truth values (nominal shear wave speed and nominal CAP) were determined through the manufacturing and calibration process of the phantoms, rather than through expert human assessment. No human experts were involved in establishing the ground truth for this non-clinical phantom study.

4. Adjudication Method for the Test Set

Not applicable. The test set used calibrated phantoms with objectively defined "ground truth" values, so no adjudication method by human experts was required.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not conducted as this was a non-clinical phantom study to assess the performance of the device itself, rather than its impact on human reader performance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

Yes, a standalone performance evaluation was conducted. The study assessed the accuracy and precision of the FibroScan® 430 Mini+ device itself (algorithm and hardware) by measuring shear wave speed and CAP on calibrated phantoms. There was no human "in the loop" performance being evaluated; the focus was solely on the device's measurements compared to known physical properties.

7. The Type of Ground Truth Used

The ground truth used was based on the known physical properties (elasticity and attenuation) of calibrated phantoms.

8. The Sample Size for the Training Set

The document does not specify a training set size. This non-clinical study focused on performance validation against a predicate device using phantoms, not on machine learning model training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a training set for an AI/machine learning model is not mentioned in the context of this device's non-clinical evaluation. The ground truth for the device's validation was established by the known characteristics of the calibrated phantoms.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the logo is a stylized image of a caduceus, a symbol of medicine and healing, with a staff and two snakes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 13, 2017

Echosens % Zvi Ladin, Ph.D. Principal Boston MedTech Advisors, Inc. 990 Washington Street, Suite #204 DEDHAM MA 02026

Re: K172142

Trade/Device Name: FibroScan® 430 Mini+ Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: July 7, 2017 Received: July 17, 2017

Dear Dr. Ladin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Page 2 - Zvi Ladin, Ph.D.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara For

Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172142

Device Name FibroScan® 430 Mini+

Indications for Use (Describe)

The FibroScan® 430 Mini+ system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.

FibroScan® 430 Mini+ is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).

The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.

Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Diagnostic Ultrasound Intended Use

System: FibroScan® 430 Mini+

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
(Track 1 Only)(Tracks 1 & 3)
OphthalmicOphthalmic
Fetal
AbdominalPP 1, 2, 3
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricPP 1, 2
FetalImaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  1. A-mode

  2. Vibration Controlled Transient Elastography at 50 Hz

  3. Controlled Attenuation Parameter (CAPTM) et 3.5 MHz

{4}------------------------------------------------

Diagnostic Ultrasound Intended Use

Transducer: FibroScan® M+ probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalImaging& OtherFetal
AbdominalPP 1, 2, 3
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
PediatricPP 1, 2
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  1. A-mode

  2. Vibration Controlled Transient Elastography at 50 Hz

  3. Controlled Attenuation Parameter (CAPTM) et 3.5 MHz

{5}------------------------------------------------

Diagnostic Ultrasound Intended Use

Transducer: FibroScan® XL+ probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
(Track 1 Only)(Tracks 1 & 3)
OphthalmicOphthalmic
Fetal
AbdominalPP 1, 2, 3
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
FetalImaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)
Cardiac Adult
CardiacCardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  1. A-mode

  2. Vibration Controlled Transient Elastography at 50 Hz

  3. Controlled Attenuation Parameter (CAPTM) et 3.5 MHz

{6}------------------------------------------------

510(K) Summary Echosens FibroScan® 430 Mini+ System

Submitter's Name, Address, Telephone Number, Contact Persona and Date Prepared:

Manufacturer:Echosens30 Place d'Italie75013 Paris, FranceTelephone: +33 1 44 82 78 55Facsimile: +33 1 44 82 68 36
Contact Person:Zvi Ladin, Ph.D.PrincipalBoston MedTech Advisors, Inc.990 Washington StreetSuite #204Dedham, MA 02026Telephone: (781) 407 0900 x104Facsimile: (781) 407 0901Email: zladin@bmtadvisors.com
Date Prepared:July 7, 2017
Name of Device and Name/Address of Sponsor
Trade/Proprietary Name:FibroScan® 430 Mini+
Common Name:Diagnostic Ultrasound System and Accessories

Classifications:

Classification NameRegulationProduct Code
Ultrasonic Pulsed Echo Imaging System21 CFR §892.1560IYO
Diagnostic Ultrasonic Transducer21 CFR §892.1570ITX

Manufacturing Facility: Echosens 30 Place d'Italie 75013 Paris, France Telephone: +33 1 44 82 78 55 Facsimile: +33 1 44 82 68 36

Establishment Registration Number: 3010258456

Predicate Device

This submission claims substantial equivalence to FibroScan® 530 Compact (#K160524) manufactured by the sponsor and cleared on March 18, 2016.

{7}------------------------------------------------

Device Description

FibroScan® 430 Mini+, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The FibroScan® 430 Mini+ CAP (Controlled Attenuation Parameter, ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness.

Intended Use / Indications for Use

The FibroScan® 430 Mini+ system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.

FibroScan® 430 Mini+ is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.

Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.

Comparison of Technological Characteristics

The FibroScan® 430 Mini+ system is substantially equivalent to the FibroScan® 530 Compact system cleared via 510(k) #K160524. Both systems encompass a Controlled Attenuation Parameter (CAP) designed to estimate the ultrasound attenuation (forward and return paths) at the frequency of 3.5 MHz, using the M+ and XL+ transducers. Additionally, both systems provide 50Hz shear wave speed measurements and estimates of tissue stiffness.

The FibroScan® 430 Mini+ modification relates to its smaller size and conversion to a portable device. The FibroScan® 430 Mini+ system is otherwise identical to the predicate FibroScan® 530 Compact systems as related to the indications for use, operating principles, M+ and XL+ probes, materials, examination procedure, imaging capabilities, information processing, performance measurements, and manufacturing process.

Consensus Standards Used

Non-clinical testing to assure compliance with acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety were performed and have been found to conform to applicable medical device safety standards.

The system complies with the following standards:

  • . IEC 60601-2-37 Edition 2.1 2015: Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment.
  • . NEMA UD 2-2004 (R2009): Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
  • . AIUM MUS: Medical Ultrasound Safety, Third Edition
  • IEC 62127-1 Edition 1.1 2013-02: Ultrasonics -- Hydrophones -- Part 1: Measurement . And Characterization of Medical Ultrasonic Fields up to 40 Mhz

{8}------------------------------------------------

  • . IEC 62127-2 Edition 1.1 2013-02: Ultrasonics -- Hydrophones -- Part 2: Calibration for Ultrasonic Fields up to 40 Mhz
  • . IEC 62127-03 Edition 1.1 2013-05: Ultrasonics -- Hydrophones -- Part 3: Properties of Hydrophones for Ultrasonic Fields up to 40 Mhz
  • . IEC 61161 Edition 3.0 2013-01: Ultrasonics -- Power Measurement -- Radiation Force Balances and Performance Requirements
  • . AAMI / ANSI ES60601-1:2005/(R)2012: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, Mod)
  • IEC 60601-1-2 Edition 3: 2007-03: Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
  • . IEC 60601-1-6 Edition 3.1 2013-10: Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
  • . IEC 62366 Edition 1.1 2014-01: Medical Devices - Application of Usability Engineering to Medical Devices
  • . AAMI / ANSI / IEC 62304 First Edition 2006-05: Medical Device Software - Software Life Cycle Processes
  • ISO 14971 Second Edition 2007-03-01: Medical Devices - Application of Risk Management to Medical Devices

Performance Data

The accuracy and precision of the device was tested for shear wave speed and CAP on calibrated phantoms with known elasticity and attenuation.

The shear wave speed bias, i.e. the difference in the mean shear wave speed measured and the nominal shear wave of the phantom, normalized by the nominal shear wave and expressed in percent was evaluated and compared to the corresponding value reported for the predicate device. Results show that while the predicate FibroScan 530 Compact shear wave speed bias results reported minimum and maximum values of [-12.9%; -2.6%] for the M+ probe and [-14.3%; -2.3%] for the XL+ probe; the range of shear wave speed bias results measured with the FibroScan 430 Mini+ device had minimum and maximum values of [-13.1%; -1.1%] for the M+ probe and [-13.3%; -2.5%] for the XL+ probe. Therefore, the overall range of bias values for the shear wave speed in the predicate FibroScan 530 Compact is <11% and <12% for the M+ and XL+ probes respectively, while the corresponding values for shear wave speed in the FibroScan 430 Mini+ is ≤12% and <11% respectively, and can therefore be considered as similar to the predicate.

The CAP bias, i.e. the difference in the mean CAP measured and the nominal CAP of the phantom, normalized by the nominal CAP and expressed as a percentage, was evaluated. The range of CAP bias values measured on the predicate FibroScan are [-1.5%; 4.7%] for the M+ probe and [1.4%; 3.9%] for the XL+ probe. Therefore, the overall range of bias values for CAP is <7% and <3% for the M+ and XL+ probes respectively.

The results are valid for FibroScan 430 Mini+ as the same acquisition platform model (US board and firmware) with identical ultrasound module and the same software core are used on FibroScan 530 Compact and FibroScan 430 Mini+.

Moreover, this was confirmed by the shear wave speed bias testing on FibroScan 430 Mini+ that showed no difference between both models.

{9}------------------------------------------------

The sheer wave speed precision measurement, i.e. the standard deviation of the independent measurements of the shear wave speed, normalized by the reference value was evaluated and compared to the corresponding value reported for the predicate device. Results show that while the predicate FibroScan 530 Compact shear wave speed precision results reported minimum and maximum values of [0%; 0.9%] for the M+ probe and [0%; 1.5%] for the XL+ probe; the range of shear wave speed results reported for the FibroScan 430 Mini+ are [0.2%; 0.7%] for the M+ probe and [0%; 0.7%] for the XL+ probe. Therefore, the overall range of precision values for the shear wave speed in the predicate FibroScan 530 Compact is <1% and <2% for the M+ and XL+ probes respectively, while the corresponding values in the FibroScan 430 Mini+ is <1% for both probes, and can therefore be considered as better than the predicate.

The CAP precision measurement, i.e. the standard deviation of the independent measurements of the CAP, normalized by the reference value was evaluated. The range of CAP precision values measured on the predicate FibroScan are [0.6%; 1.0%] and [0.9%; 1.3%] for M+ and XL+ probes respectively. Therefore, the overall range of precision values for CAP values is <1% for both probes.

The results are valid for FibroScan 430 Mini+ as the same acquisition platform model (US board and firmware) with identical ultrasound module and the same software core are used on FibroScan 530 Compact and FibroScan 430 Mini+.

Moreover, this was confirmed by the shear wave speed precision testing on FibroScan 430 Mini+ that showed no difference between both models.

In summary, the bias and the precision of the shear wave speed measured by the FibroScan 430 Mini+ are within the same range or better than those of the predicate FibroScan 530 Compact device. Therefore, the FibroScan 430 Mini+ system was found to have a safety and effectiveness profile that is similar to the predicate FibroScan 530 Compact device.

Substantial Equivalence

The modified FibroScan® System (Model 430 Mini+) has the same intended use and indications for use, uses the same operating principle and materials, incorporates the same basic design, emits the same energy and acquires the same information as the predicate device. The differences in size, weight and internal organization of its components do not raise new or different questions of safety or efficacy.

In summary, the FibroScan® 430 Mini+ is substantially equivalent to the predicate device the FibroScan® 530 Compact (#K160524) manufactured by the sponsor and cleared on March 18, 2016.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.