510(k) Data Aggregation
K Number
K223902Device Name
FibroScan® device (Models: 502 Touch, 530 Compact, 430 Mini+, 230, and 630)
Manufacturer
Echosens
Date Cleared
2023-03-02
(64 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® device (Models: 502 Touch, 530 Compact, 430 Mini+, 230, and 630) is intended to measure liver stiffness (E) using Vibration Controlled Transient Elastography™ (VCTE™) at 50 Hz shear wave frequency and liver ultrasound attenuation coefficient (CAP™)* at 3.5 MHz. FibroScan® 630 Expert is also intended to measure spleen stiffness using VCTE™ at 100 Hz shear wave frequency.
FibroScan liver stiffness measurements (LSM) by VCTE™ may aid the physician in determining the likelihood of cirrhosis and may be used, taken in context with other clinical and laboratory data, as an aid in the assessment of liver fibrosis. FibroScan CAP™ measurements may be used, taken in context with other clinical and laboratory data, as an aid in the assessment of hepatic steatosis.
FibroScan® is indicated as a non-invasive aid for the clinical management, diagnosis, and monitoring of adult and pediatric patients with confirmed or suspected liver disease, as part of an overall assessment of the liver. Results in the pediatric population should be interpreted while considering the clinical condition and the overall patient profile.
The FibroScan® device is intended for use by healthcare professionals in hospitals, clinics or any facility where healthcare is provided.
*CAP™ refers to ultrasound attenuation coefficient (originally defined as Controlled Attenuation Parameter). CAP™ on S+ probe is only available with SmartExam capability.
Device Description
FibroScan® System consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology and is designed to perform non-invasive measurements of liver/spleen shear wave speed and estimate tissue stiffness. The probe, containing a mechanical vibrator, produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver/spleen. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
Ask a Question
K Number
K212035Device Name
FibroScan 230
Manufacturer
Echosens
Date Cleared
2021-07-30
(30 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® 230 is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The FibroScan® 230 is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness, and CAP may be used as an aid in the clinical management of pediatric patients with liver disease.
Device Description
FibroScan® System and its probes is an active non-implantable medical device using ultrasound and based on Vibration- Controlled Transient Elastography (VCTE™) technology. It is designed to perform non-invasivemeasurements of liver shear wave speed and estimate tissue stiffness. The probe, containing a mechanical vibrator, produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness.
The focus of this submission is the new FibroScan® 230, which separates the FibroScan® System into two parts: a smaller equipment unit (which includes the hand-held probes) and the FibroScan® application, installed on the user's computer. The FibroScan® Application displays the user interface of the system and interacts with the equipment unit through a USB connection. The FibroScan® Application may also interact with the Echosens Cloud through an internet network. Only the FibroScan® 230 equipment and FibroScan® application are included in the FibroScan® 230 system while the Echosens cloud and the end-user computer are not part of the FibroScan® 230 medical device.
Ask a Question
K Number
K203273Device Name
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, 430 Mini+, and 630)
Manufacturer
Echosens
Date Cleared
2021-03-25
(139 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® Family of Products (Models: 502 Touch, 530 Compact, 430 Mini+, and 630) is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
The FibroScan® Family of Products (Models: 502 Touch, 530 Compact, 430 Mini+, and 630) is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness, and CAP* may be used as an aid in the clinical management of pediatric patients with liver disease.
FibroScan® 630 (Expert) is also indicated for noninvasive measurement in the spleen of 100 Hz shear wave speed and estimates of stiffness that may be used as an aid to diagnosis, monitoring and clinical management of adult patients with liver disease, as part of an overall assessment of the liver.
*CAP for pediatric patients with liver disease is only available with SmartExam capability on FibroScan® Models: 530 Compact, 430 Mini+, and 630
Device Description
FibroScan® System consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology and is designed to perform non-invasive measurements of liver/spleen shear wave speed and estimate tissue stiffness. The probe, containing a mechanical vibrator, produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver/spleen. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
The focus of this submission is an updated version of FibroScan software (CLPC 4.1) for all previously cleared FibroScan® Family of Products. The software version CLPC 4.1 involves the following changes:
- Streamlined software platform across between all devices
- Simplified user interface
- Functional enhancements:
- Continuous CAP (Controlled Attenuation Parameter) measurement (CAPc)
- CAPc applied to S+ probe
- SmartDepth adjustment of measurement depth to patient anatomy
- Improved probe localization step
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, 430 Mini+, and 630) has the same intended use as the previously cleared FibroScan models. The submission expands the indications for use of the system to include use of the S+ probe for estimation of CAP.
Ask a Question
K Number
K200655Device Name
FibroScan® 630
Manufacturer
Echosens
Date Cleared
2020-07-16
(126 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® 630 is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease.
FibroScan® 630 is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness may be used as an aid in the clinical management of pediatric patients with liver disease.
FibroScan® 630 (Expert) is also indicated for noninvasive measurement in the spleen of 100 Hz shear wave speed and estimates of stiffness that may be used as an aid to diagnosis, monitoring and clinical management of adult patients with liver disease, as part of an overall assessment of the liver.
Device Description
FibroScan® system consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology and is designed to perform non-invasive measurements of liver/spleen shear wave speed and estimates of tissue stiffness. The probe containing a mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver/spleen. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
The focus of this submission is the new FibroScan® 630, available in two configurations: FibroScan® 630 'Prime' (FS630P) and FibroScan® 630 'Expert' (FS630E). Both FS630P and FS630E include improvements of the FibroScan appearance and user interface.
The FS630P version has the same indications as the previously cleared models and is only indicated for non-invasive measurement of 50 Hz shear wave speed in the liver and estimates of its stiffness, as well as determination of a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The FS630E includes an expansion of the intended use to include noninvasive measurement of shear wave speed at 100Hz in the spleen and estimates of its stiffness as well as an inclusion of a B-Mode ultrasound imaging system to help the user locate the liver or spleen.
Ask a Question
K Number
K181547Device Name
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+)
Manufacturer
Echosens
Date Cleared
2018-07-09
(27 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+) is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+) is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.
Device Description
FibroScan® system consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology, and is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. The probe containing a mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
The focus of this submission is the addition of the S+ probe to models FibroScan® 530 Compact and FibroScan® 430 Mini+. The S+ probe addresses a smaller anatomic size of pediatric patients, while using the same principle of operation, intended use and methodology (i.e. application to patient, signal measurement, processing and display), design, materials, manufacturing and testing processes as the previously cleared M+ and XL+ probes. In addition, the S+ probe developed for the 530 Compact and 430 Mini+ models is similar to the S+ probe cleared for the FibroScan® 502 Touch model of the FibroScan® Family of Products.
Ask a Question
K Number
K173034Device Name
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+)
Manufacturer
Echosens
Date Cleared
2017-11-14
(47 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+) is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.
FibroScan® Family of Products (Models: 502 Touch, 530 Compact, and 430 Mini+) is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.
Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.
Device Description
FibroScan® system consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology, and is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. The probe containing a mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
Ask a Question
K Number
K172142Device Name
FibroScan 430 Mini+
Manufacturer
Echosens
Date Cleared
2017-09-13
(58 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® 430 Mini+ system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.
FibroScan® 430 Mini+ is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).
The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.
Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.
Device Description
FibroScan® 430 Mini+, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The FibroScan® 430 Mini+ CAP (Controlled Attenuation Parameter, ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness.
Ask a Question
K Number
K160524Device Name
FibroScan 530 Compact
Manufacturer
ECHOSENS
Date Cleared
2016-03-18
(22 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
ECHOSENS
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® 530 Compact system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.
FibroScan® 530 Compact is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.
Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.
Device Description
FibroScan® 530 Compact, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The FibroScan® 530 Compact CAP (Controlled Attenuation Parameter, ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness.
Ask a Question
K Number
K150239Device Name
FibroScan
Manufacturer
Echosens
Date Cleared
2015-09-01
(211 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
Echosens
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness through internal structures of the body.
FibroScan® is indicated for noninvasive measurement of shear wave speed and estimate of stiffness at 50 Hz in the liver. The shear wave speed may be used as an aid to clinical management of pediatric and adult patients with liver disease.
Device Description
FibroScan® system consists of a system unit and a hand-held probe. It is based on Vibration-Controlled Transient Elastography (VCTE™) technology, and is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. The probe containing a mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. Ultrasound is used to track the shear (elastic) wave, measure its speed and provide estimated stiffness. The results are displayed on the system unit.
The focus of this submission is the expansion of the indications for use for the FibroScan system by Echosens to pediatric patients. In order to address the smaller anatomic size of pediatric patients, a new probe (S+) was developed, and the indications for use of the previously cleared M+ probe were modified. The new probe uses the same principle of operation, intended use and methodology (i.e. application to patient, signal measurement, processing and display), design, materials, manufacturing and testing processes as the previously cleared M+ and XL+ probes. The device specifications are similar to those of the predicate device. The system's software was upgraded to accommodate these changes.
Ask a Question
K Number
K150949Device Name
FibroScan
Manufacturer
ECHOSENS
Date Cleared
2015-06-03
(56 days)
Product Code
IYO, ITX
Regulation Number
892.1560Why did this record match?
Applicant Name (Manufacturer) :
ECHOSENS
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FibroScan® system is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffhess as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body.
FibroScan® is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter). The shear wave speed and stiffness, and CAP may be used as an aid to clinical management of adult patients with liver disease.
Shear wave speed and stiffness may be used as an aid to clinical management of pediatric patients with liver disease.
Device Description
FibroScan®, based on Vibration-Controlled Transient Elastography (VCTE™) technology, is designed to perform non-invasive measurements of liver shear wave speed and estimates of tissue stiffness. A mechanical vibrator produces low-amplitude elastic waves that travel through the skin and intercostal space into the liver. The speed of propagation of the shear (elastic) wave is measured using ultrasounds. A new FibroScan® parameter labeled CAP (Controlled Attenuation Parameter), ranging between 100 and 400 decibels per meter (dB/m), provides an estimation of the total aforementioned ultrasonic wave attenuation (forward and return paths) at 3.5 MHz, measured concomitantly with tissue stiffness.