(57 days)
No
The description focuses on the mechanical and material properties of a balloon catheter and does not mention any computational or analytical features that would suggest the use of AI/ML. The "Mentions AI, DNN, or ML" field is explicitly marked as "Not Found".
Yes
The device is intended for "balloon dilation of the iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG)," which describes a therapeutic intervention to treat conditions in these anatomical structures.
No
Explanation: The device is a Percutaneous Transluminal Angioplasty (PTA) catheter intended for balloon dilation, which is a therapeutic procedure to open narrowed arteries, not for diagnosing conditions.
No
The device description clearly details a physical catheter with a balloon, guidewire lumen, inflation lumen, and radiopaque markers. It also describes non-clinical performance studies related to the physical properties and function of the hardware.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information about a person's health.
- Device Function: The Pirouette High Pressure (HP) PTA Catheter is a medical device used within the body to physically dilate blood vessels. It is a therapeutic device, not a diagnostic one.
- Intended Use: The intended use clearly states "balloon dilation of the iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG)." This is a treatment procedure, not a diagnostic test.
- Device Description: The description details the physical components and function of a catheter used for angioplasty, which is a procedure performed directly on the patient's anatomy.
- Lack of Diagnostic Elements: There is no mention of analyzing biological samples, detecting biomarkers, or providing diagnostic information based on in vitro testing.
Therefore, based on the provided information, the Pirouette High Pressure (HP) PTA Catheter is a therapeutic medical device, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
Pirouette High Pressure (HP) PTA Catheter is intended for balloon dilation of the iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).
Pirouette High Pressure (HP) PTA Catheter K172033 is intended for balloon dilation of the iliac, femoral, popliteal, infrapopliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).
Product codes (comma separated list FDA assigned to the subject device)
LIT
Device Description
The Pirouette High Pressure (HP) Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating. The manifold connector and shaft design consists of a guidewire lumen allowing the catheter to track over a guidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualisation of the catheter/balloon under fluoroscopy. The catheter is compatible with 0.018-inch (0.46 mm) wire guides.
The Pirouette High Pressure (HP) PTA Catheter Family includes multiple balloon sizes ranging from 4 to 6 mm in diameter and 20 to 80mm in length. The nominal balloon diameter (mm) and the balloon length (mm) are inscribed on the guidewire hub of the manifold. The effective lengths of the balloon range from 45cm to 75cm.
Physical Description: High Pressure (HP) Percutaneous Transluminal Angioplasty Catheter (PTA Catheter).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Fluoroscopy
Anatomical Site
iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The Pirouette HP PTA Catheter was thoroughly tested on the bench to evaluate and verify that it meets the required performance specifications and to support a determination of substantial equivalence. The bench testing plan was developed with the consideration of the recommendations outlined in the applicable FDA guidance documents, tests recommended in ISO 10555-1, Intravascular catheters- Sterile and single-use catheters - Part 1: General Requirements, and 10555-4, Intravascular catheters- Sterile and single-use catheters - Part 4: Balloon dilatation catheters. Testing performed on the Pirouette included the following:
Non-Clinical Tests:
- Dimensional verification
- Balloon preparation, pushability , trackability, deployment, withdrawal . and balloon reinsertion
- Balloon rated burst pressure (RBP)
- Balloon fatique (repeated balloon inflations)
- Balloon Compliance at nominal and rated burst inflation pressure
- . Balloon Inflation/Deflation Time
- Catheter Bond Strength (Tip to balloon, balloon to proximal shaft, shaft to manifold)
- . Flexibility and Kink test
- Torque Strength
- . Radiopacity
- . Coating Integrity
- Particulate evaluation
- . Guide wire compatibility
- . Introducer sheath compatibility
The results showed that the Pirouette HP PTA Catheter met the pre-determined acceptance criteria.
Biocompatibility Tests:
Per ISO10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the Pirouette HP PTA Catheter is classified as an externally communicating device, which contacts circulating blood during a limited contact duration (≤24hours). Biocompatibility testing was not performed directly on the Pirouette HP PTA Catheter. Biocompatibility testing was leveraged from the reference device-Pirouette 018. Biocompatibility testing was conducted on Pirouette 018 per ISO 10993-1:2009.
The test results show that the Pirouette 018 PTA Catheter is biocompatible, and based on their similar characteristics these were leveraged to support the biocompatibility of the Pirouette HP Catheter.
Sterilization, Shelf life tests and Packaging validation:
- . EtO sterilzation validation
- EtO/ECH residue determination
- Shelf life testing .
- . Packaging validation
Test results show that the Pirouette HP PTA Catheter is sterility is maintained by the packaging during the entire shelf life of the device. From the results it was also concluded that the device meets the criteria for Residual Testing, i.e. EtO/ECH residues.
Clinical Performance Data:
- No clinical studies were performed for the purpose of obtaining safety and effectiveness data.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three profiles of human faces incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 31, 2017
ArraVasc Limited Ms. Carmel Doherty Design Assurance Engineer 2 Ballybrit Business Park Galway, Ireland
Re: K172033
Trade/Device Name: Pirouette High Pressure (HP) PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: June 30, 2017 Received: July 6, 2017
Dear Ms. Doherty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
1
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K172033
Device Name Pirouette High Pressure (HP) PTA Catheter
Indications for Use (Describe)
Pirouette High Pressure (HP) PTA Catheter is intended for balloon dilation of the iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) |
---|
☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Section 5.0 510(k) Summary
5.1 Submitter
| Submitter Address: | ArraVasc Limited
2 Ballybrit Business Park,
Galway
Ireland |
|--------------------|---------------------------------------------------------------------|
| Phone Number: | +353-91-758939 |
| Fax Number: | +353-91-758930 |
| Contact Person: | Carmel Doherty |
| Date Prepared: | 24 August 2017 |
5.2 Device
Device Trade Name: | Pirouette High Pressure (HP) PTA Catheter K172033 |
---|---|
Common Name: | OTW PTA catheter |
Classification Name: | Peripheral Transluminal Angioplasty Catheter |
Classification number: | 21 CFR 870.1250 |
Product code: | LIT |
Class: | II |
Classification Panel: | Cardiovascular |
Predicate Devices 5.3
| Predicate: | Mustang Balloon Dilation Catheter
Boston Scientific Corporation.
510(K) number: K103751 |
|-------------------|-----------------------------------------------------------------------------------------------|
| Reference device: | Pirouette 018 PTA Dilation Catheter
ArraVasc
510(K) number: K151153 |
5.4 Device Description
Device Description: The Pirouette High Pressure (HP) Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating. The manifold connector and shaft design consists of a guidewire lumen allowing the catheter to track over a guidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualisation of the
4
| | catheter/balloon under fluoroscopy. The catheter is compatible with
0.018-inch (0.46 mm) wire guides. |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Pirouette High Pressure (HP) PTA Catheter Family includes multiple
balloon sizes ranging from 4 to 6 mm in diameter and 20 to 80mm in
length. The nominal balloon diameter (mm) and the balloon length (mm)
are inscribed on the guidewire hub of the manifold. The effective lengths
of the balloon range from 45cm to 75cm. |
| Physical Description: | High Pressure (HP) Percutaneous Transluminal Angioplasty Catheter
(PTA Catheter). |
| 5.5 | Indication for use |
- Indications for Use Pirouette High Pressure (HP) PTA Catheter K172033 is intended for Statement: balloon dilation of the iliac, femoral, popliteal, infrapopliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).
5.6 Comparison of technological characteristics with the predicate device
- Summary The Pirouette High Pressure (HP) PTA Catheter K172033 incorporates of substantially equivalent design, packaging, fundamental technology, Technological Characteristics materials, manufacturing, sterilization and intended use as those featured in the predicate, Mustang Balloon Dilation catheter.
It has the same design, packaging, fundamental technology, materials, manufacturing, sterilization and hydrophilic coating as the reference device, Pirouette 018 PTA catheter.
5
Parameter | Characteristics |
---|---|
Classification | Class II, 21 CFR 870.1250 |
Same Classification as predicate devices. | |
Intended Use | Same intended use of balloon dilatation of the iliac, femoral, popliteal, |
infra-popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio- | |
venous Graft (AVG). The Mustang Balloon Dilatation Catheter is also | |
indicated for post-dilatation of balloon expandable and self-expanding | |
stents in the peripheral vasculature. | |
Balloon material | Similar type of material (Nylon v Nylon-Pebax blend) |
Balloon diameter | Comparable range of balloon diameters: 4 – 6 mm |
Balloon length | Comparable range of balloon lengths: 20 – 80mm |
Nominal pressure (atm) | Pirouette HP PTA Catheter has the same or higher nominal pressure: |
12 atm. | |
Rated burst pressure (atm) | Pirouette HP PTA Catheter has a comparable rated burst pressure: |
22 atm (balloon diameter 4.0- 5.0 mm); | |
20 atm (balloon diameter 6.0mm) | |
Radiopaque | |
Marker bands | Pirouette HP PTA Catheter has two Markerbands, one at distal and one |
at proximal side of the balloon, with the same function. | |
Outer shaft | Pirouette HP PTA Catheter has similar type of material: Polyamide based |
Catheter shaft | |
outer diameter | Pirouette HP PTA Catheter has the same diameter or lower: 3.9 – 4.2 Fr |
Catheter Usable | |
Lengths (cm) | Pirouette HP PTA Catheter has a comparable range of usable lengths: |
45 – 75cm | |
Recommended | |
Introducer Sheath | |
compatibility | Pirouette HP PTA Catheter has the same size or lower: 4 – 5 Fr |
Recommended | |
Guidewire diameter | Pirouette HP PTA Catheter has the same or lower compatibility; |
maximum 0.018" | |
Catheter strain relief & | |
manifold material | Pirouette HP PTA Catheter has the same type of material (Polyamide based), with the same function. |
Catheter manifold | |
design | Same design, dual lumen Y design, with the same function. |
Catheter coating | Pirouette HP PTA Catheter has similar type of coating. |
Sterilization Method | Same method; Ethylene Oxide. |
Single Use / Reusable | Single Use |
Table 1 Summary of general and technical characteristics against the predicate devices.
6
5.7 Performance Data
The Pirouette HP PTA Catheter was thoroughly tested on the bench to evaluate and verify that it meets the required performance specifications and to support a determination of substantial equivalence. The bench testing plan was developed with the consideration of the recommendations outlined in the applicable FDA guidance documents, tests recommended in ISO 10555-1, Intravascular catheters- Sterile and single-use catheters - Part 1: General Requirements, and 10555-4, Intravascular catheters- Sterile and single-use catheters - Part 4: Balloon dilatation catheters. Testing performed on the Pirouette included the following:
Non-Clinical Tests:
- . Dimensional verification
- Balloon preparation, pushability , trackability, deployment, withdrawal . and balloon reinsertion
- Balloon rated burst pressure (RBP) ●
- Balloon fatique (repeated balloon inflations) ●
- Balloon Compliance at nominal and rated burst inflation pressure
- . Balloon Inflation/Deflation Time
- Catheter Bond Strength (Tip to balloon, balloon to proximal shaft, shaft to manifold)
- . Flexibility and Kink test
- Torque Strength ●
- . Radiopacity
- . Coating Integrity
- Particulate evaluation
- . Guide wire compatibility
- . Introducer sheath compatibility
The results showed that the Pirouette HP PTA Catheter met the pre-determined acceptance criteria.
Biocompatibility Tests:
Per ISO10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the Pirouette HP PTA Catheter is classified as an externally communicating device, which contacts circulating blood during a limited contact duration (≤24hours). Biocompatibility testing was not performed directly on the Pirouette HP PTA Catheter. Biocompatibility testing was leveraged from the reference device-Pirouette 018. Biocompatibility testing was conducted on Pirouette 018 per ISO 10993-1:2009.
The test results show that the Pirouette 018 PTA Catheter is biocompatible, and based on their similar characteristics these were leveraged to support the biocompatibility of the Pirouette HP Catheter.
7
Sterilization, Shelf life tests and Packaging validation:
- . EtO sterilzation validation
- EtO/ECH residue determination ●
- Shelf life testing .
- . Packaging validation
Test results show that the Pirouette HP PTA Catheter is sterility is maintained by the packaging during the entire shelf life of the device. From the results it was also concluded that the device meets the criteria for Residual Testing, i.e. EtO/ECH residues.
Clinical Performance Data:
- No clinical studies were performed for the purpose of obtaining safety and effectiveness ● data.
5.8 Conclusions
The performance testing and comparison tabulated above demonstrates that the Pirouette High Pressure (HP) PTA Catheter K172033 device is substantially equivalent to the predicate devices.
Information contained within this submission demonstrates that the Pirouette HP PTA Catheter:
- . Has a legally marketed predicate device;
- Has the same indications for use as the identified predicate device; .
- . Incorporates similar fundamental technology, and uses accepted scientific methods and international standards to evaluate device safety and effectiveness;
- Demonstrates that the design and performance of the Pirouette HP PTA Catheter has equivalent safety and performance characteristics of predicate device, and does not raise different questions of safety and effectiveness.
Based upon the; intended use, design, performance characteristics, non-clinical performance testing performed, and comparison to legally marketed devices, it is concluded that the Pirouette HP PTA Catheter is appropriate for its intended use, and is substantially equivalent to the predicate device.