K Number
K171775
Manufacturer
Date Cleared
2018-01-09

(209 days)

Product Code
Regulation Number
890.5500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Theradome LH80 PRO is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.

Device Description

The Theradome LH80 PRO for men and women is a low level coherent light therapy (LLLT) device utilised to promote hair growth. The coherent lights are contained inside a lightweight, one-size fits all helmet. The Theradome LH80 PRO uses 80 coherent light diodes in the helmet to deliver coherent light stimulation to the entire scalp for hands-free operation during treatment. An audible timer automatically turns the coherent lights off after the 20 minutes treatment is completed. This coherent light helmet is not constructed to be a safety helmet.

AI/ML Overview

The provided text describes the Theradome LH80 PRO device and its FDA submission (K171775). While it states that "all acceptance criteria were met by the devices," and refers to compliance with IEC-60825-2007-03, it does not explicitly list specific acceptance criteria or present the full details of a study with performance metrics in a way that allows for a direct table comparing acceptance criteria to reported device performance.

Instead, the document primarily focuses on establishing "substantial equivalence" to predicate devices based on similar technological characteristics and intended use. It explicitly states "Brief Description of Clinical Testing: Not applicable." This indicates that a clinical study designed to test efficacy against specific hair growth metrics was not performed for this 510(k) submission.

Therefore, many sections of your request cannot be fully answered from the provided text. I will answer what is available and indicate when information is not provided.

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the devices." It also notes compliance with standard IEC-60825-2007-03, which establishes the AEL (accessible emission limits) as 5 milliwatts maximum. However, specific, measurable "acceptance criteria" related to efficacy (e.g., percentage increase in hair count, reduction in hair loss rate) and corresponding reported performance from a specific study are not provided in this document. The focus is on safety and technical specifications, and substantial equivalence to a predicate.

Acceptance CriterionReported Device PerformanceComments
Compliance with design specificationsAll functions verified to operate as designed.General statement, specific metrics not detailed.
Compliance with IEC-60825-2007-03 (Coherent Light Class 3R)AEL (accessible emission limits) meets 5 milliwatts maximum.This is a safety standard, not an efficacy criterion.
Usability for self-selectionVerified.Details of "verification" are not provided.
Comprehension of user instructionsVerified.Details of "verification" are not provided.
Comprehension of warningsVerified.Details of "verification" are not provided.
Comprehension of precautionsVerified.Details of "verification" are not provided.
Efficacy-related metrics (e.g., hair count)Not provided in this document for the K171775 submission.The submission relies on substantial equivalence and does not report on clinical efficacy trials for this specific submission.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample size for the test set: Not applicable for clinical performance as "Clinical Testing: Not applicable" is stated. For the "self-selection and usability" tests, the sample size is not mentioned.
  • Data provenance: Not specified for the usability tests. For the technical compliance, it's typically internal testing by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not applicable as clinical efficacy studies with ground truth established by experts are explicitly stated as "Not applicable" for this submission (K171775). For usability, no information on experts is provided.

4. Adjudication Method for the Test Set

  • Not applicable, as no clinical efficacy test set is described.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done. The document states "Brief Description of Clinical Testing: Not applicable."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. The Theradome LH80 PRO is a low-level light therapy device, not an algorithm or AI system.

7. The Type of Ground Truth Used

  • For the technical performance and safety aspects, the ground truth is established by adherence to recognized standards (e.g., IEC-60825-2007-03) and design specifications.
  • For usability and self-selection, the ground truth would typically be user feedback or observation, but specific details are not provided.
  • For clinical efficacy (e.g., hair growth), no ground truth data from a clinical study is included in this submission.

8. The Sample Size for the Training Set

  • Not applicable, as this is a medical device for light therapy, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as this is not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 9, 2018

Theradome Inc. Tamim Hamid President and CEO 4900 Hopyard Rd, Ste 100 Pleasanton, California 94588

Re: K171775

Trade/Device Name: Theradome LH80 PRO Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: December 6, 2017 Received: December 11, 2017

Dear Tamim Hamid:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jennifer R. Stevenson -ਟਤ For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171775

Device Name Theradome LH80 PRO

Indications for Use (Describe)

The Theradome LH80 PRO is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.

Type of Use (Select one or both, as applicable)

Foundation Use (Part 21 CFR 201.66) - Select One
Own Firm Construction (21 CFR 211.111) - Select One

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.0 Traditional 510(k) Summary

Date Prepared: Thursday, January 4, 2018

Image /page/3/Picture/2 description: The image shows the logo for Theradome Hair Growth Treatment System. The word "theradome" is written in a combination of gray and blue colors. Above the word "theradome" are three blue curved lines, and below the word is the phrase "Hair Growth Treatment System" in blue.

Submitter's Contact Information:Sarah Mersereau, MBA PMPConsultant, Theradome Inc.4900 Hopyard Road, Suite 100Pleasanton, CA 94588Telephone: 510-714-0525Email: sarah@theradome.com
Name of Device and Name / Address of Sponsor:Theradome LH80 PROTheradome Inc.4900 Hopyard Road, Suite 100Pleasanton, CA 94588Telephone: 510-714-0525Fax: 925-374-1202Establishment Registration# 274120
Sponsor Contact Information:Tamim Hamid, PhDCEO, Theradome Inc.4900 Hopyard Road, Suite 100Pleasanton, CA 94588Telephone: 510-714-0525Fax: 925-374-1202Email: tamim@theradome.com
Device Name and ClassificationProduct/Trade Name: Theradome LH80 PROCommon or Usual Name: Theradome LH80 PROClassification Name: Infrared lamp per 21 CFR 890.5500Product Code: OAP (Coherent light, Comb, Hair)Regulation Number: 21 CFR 890.5500Panel: General and Plastic SurgeryClass: II

{4}------------------------------------------------

Predicate Devices:

Predicate510(k) NumberDevice NameManufacturer
PrimaryK122950Theradome LH80 PROTheradome Inc.
ReferenceK163170Capillus82Capillus LLC
AdditionalK162782iHelmetSlinph TechnologiesCo., Ltd

Device Trade Name Manufacturer 510(k)

Device Description Summary

The Theradome LH80 PRO for men and women is a low level coherent light therapy (LLLT) device utilised to promote hair growth. The coherent lights are contained inside a lightweight, one-size fits all helmet. The Theradome LH80 PRO uses 80 coherent light diodes in the helmet to deliver coherent light stimulation to the entire scalp for hands-free operation during treatment. An audible timer automatically turns the coherent lights off after the 20 minutes treatment is completed. This coherent light helmet is not constructed to be a safety helmet.

Intended Use / Indications for Use

The Theradome LH80 PRO is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-2; both with Fitzpatrick Skin Types I to IV.

Technological Characteristics

The Theradome LH80 PRO coherent light Helmet is a low level coherent light therapy (LLLT) device used to promote hair growth. The coherent lights are contained inside a lightweight onesize-fits-all helmet. The Theradome LH80 PRO utilizes 80 coherent light diodes to deliver coherent light stimulation to the entire scalp for hands-free operation during treatment. The device is onebutton operated, and has an audible timer that automatically turns the coherent lights off after the 20 minute treatment completes.

Performance Data

Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the devices. The Theradome LH80 PRO conforms to the standard IEC-60825-2007-03. This IEC standard is a recognized and accepted standard by the FDA. The guidance document for this accepted standard is found in the Federal register, July 26, 2001 (volume 66, Number 144) [page 39049 39050]. This report validates for the Theradome LH80 PRO the coherent light class of 3R which establishes the AEL (accessible emission limits) as 5 milliwatts maximum. The data presented in this submission is restricted to self-selection and usability to include comprehension of user instructions, warnings, and precautions.

{5}------------------------------------------------

Brief Description of NonClinical Testing

As part of establishing substantial equivalence between LH80 PRO, iHelmet, and Capillus82 devices within the LH80 PRO (K171775) 510(k) submission, the technological characteristics, such as, fluence, irradiance, treatment time, number of diodes, etc. are compared within the subject and predicate devices.

Brief Description of Clinical Testing

Not applicable.

Substantial Equivalence

Theradome LH80 PRO utilizes the same fundamental technology and comparable technological parameters as the predicates. Please see table below.

{6}------------------------------------------------

Theradome LH80 PROTheradome LH80 PROCapillus82iHelmet
- Females only
K171775K122950K163170K162782
LLLT Device TypeLLLT Device TypeLLLT Device TypeLLLT Device Type
Prescription Use/OTCPrescription Use/OTCPrescription Use/OTCPrescription Use/OTC
Intended Use -Intended Use -Intended Use -Intended Use -
Androgenic AlopeciaAndrogenic AlopeciaAndrogenic AlopeciaAndrogenic Alopecia
Contains coherentContains coherentContains coherentContains coherent
light diodes- Class 3Rlight diodes- Class 3Rlight diodes- Class 3Rlight diodes- Class 3R
Helmet/Cap DesignHelmet/Cap DesignHelmet/Cap DesignHelmet/Cap Design
678nm678nm650nm650nm
Marketing clearanceMarketing clearanceMarketing clearanceMarketing clearance
for Males and Femalesfor Femalesfor Males and Femalesfor Males and Females
Passive Use-HandsPassive Use-HandsPassive Use-HandsPassive Use-Hands
FreeFreeFreeFree
Features - One buttonFeatures - One buttonFeatures - One buttonFeatures - One button
operation, audibleoperation, audibleoperation, audibleoperation, audible
timer, status LEDtimer, status LEDtimer, status LEDtimer, status LED
Treatment time- 20Treatment time- 20Treatment time- 30Treatment time- 20
minutesminutesminutesminutes
Irradiance = sameIrradiance = sameIrradiance =Irradiance =
comparablecomparable
Fluence = sameFluence = sameFluence = comparableFluence = comparable
80 coherent light80 coherent light82 Coherent light200 Coherent light
diodesdiodesdiodesdiodes
OAP ClassificationOAP ClassificationOAP ClassificationOAP Classification
Classification Name -Classification Name -Classification Name -Classification Name -
Infrared LampInfrared LampInfrared LampInfrared Lamp
Common Usage NameCommon Usage NameCommon Usage NameCommon Usage Name
-Lamp, Non-Heating-Lamp, Non-Heating-Lamp, Non-Heating-Lamp, Non-Heating
General & PlasticGeneral & PlasticGeneral & PlasticGeneral & Plastic
Surgery CommitteeSurgery CommitteeSurgery CommitteeSurgery Committee
Fitzpatrick SkinFitzpatrick SkinFitzpatrick SkinFitzpatrick Skin
Phototypes - I- IVPhototypes - I- IVPhototypes - I- IVPhototypes - I- IV
Ludwig-Savin I-IILudwig-Savin I-IILudwig-Savin I-IILudwig-Savin I-II
(females) Norwood(females)(females) Norwood(females) Norwood
Hamilton IIA-V (males)Hamilton IIA-V (males)Hamilton IIA-V (males)Hamilton IIA-V (males)
Device Class IIDevice Class IIDevice Class IIDevice Class II

Conclusion

Based on the comparable indications for use and technological characteristics the subject device is substantially equivalent to the predicates.

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.