K Number
K163170
Device Name
Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office Pro, Capillus 302, Capillus 312, and Capillus 352
Manufacturer
Date Cleared
2017-01-31

(78 days)

Product Code
Regulation Number
890.5500
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.
Device Description
The Capillus272 Pro and Capillus272 OfficePro both consist of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner, and configured for portable use with rechargeable battery and adapter (Capillus272 Pro) and salon/spa use (Capillus272 OfficePro). The Capillus82, Capillus202, Capillus302, Capillus312, and Capillus352 are exactly the same as the Capillus272 Pro with the exception of the number of diodes which are respective of model (e.g. 82 diodes in Capillus82). As stated in prior submissions, the devices emit an audible tone at the beginning and end of a therapy session, indicating that therapy has begun (2 short beeps) or ended (one long beep). The portable systems are powered by rechargeable Li-Ion battery cells assembled into a proprietary battery pack. Both the battery are fully compliant to recognized, international standards. The OfficePro is directly powered by the DC charger, which is fully compliant to recognized, international standards.
More Information

103368, 142573, 141567

No
The device description and performance studies focus on the laser technology and usability testing, with no mention of AI or ML.

Yes
The device is intended to treat Androgenetic Alopecia and promote hair growth, which are therapeutic claims.

No

Explanation: The "Intended Use" section states the device is "intended to treat Androgenetic Alopecia and promote hair growth," indicating a therapeutic rather than diagnostic function. There is no mention of the device being used to identify or diagnose medical conditions.

No

The device description clearly states that the device consists of physical components including laser diodes, a helmet, a battery, and an adapter. It is a hardware device that utilizes light therapy.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to treat Androgenetic Alopecia and promote hair growth. This is a therapeutic purpose, not a diagnostic one.
  • Device Description: The device is a laser dome that emits light. It does not involve the examination of specimens derived from the human body (like blood, urine, tissue, etc.), which is the core characteristic of an IVD.
  • Lack of IVD-related terms: The document does not mention any terms typically associated with IVDs, such as "in vitro," "specimen," "assay," "reagent," "diagnostic," "analyte," etc.
  • Performance Studies: The performance studies focus on usability and compliance with laser safety standards, not on the accuracy or reliability of a diagnostic test.

In summary, the Capillus laser domes are medical devices intended for therapeutic use (hair growth), not for diagnosing diseases or conditions using in vitro methods.

N/A

Intended Use / Indications for Use

The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

Product codes

OAP

Device Description

The Capillus272 Pro and Capillus272 OfficePro both consist of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner, and configured for portable use with rechargeable battery and adapter (Capillus272 Pro) and salon/spa use (Capillus272 OfficePro). The Capillus82, Capillus202, Capillus302, Capillus312, and Capillus352 are exactly the same as the Capillus272 Pro with the exception of the number of diodes which are respective of model (e.g. 82 diodes in Capillus82).

As stated in prior submissions, the devices emit an audible tone at the beginning and end of a therapy session, indicating that therapy has begun (2 short beeps) or ended (one long beep). The portable systems are powered by rechargeable Li-Ion battery cells assembled into a proprietary battery pack. Both the battery are fully compliant to recognized, international standards. The OfficePro is directly powered by the DC charger, which is fully compliant to recognized, international standards.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

The testing was administered to 55 adult subjects of any age (>18).

Intended User / Care Setting

Over-The-Counter Use (21 CFR 801 Subpart C) / Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

The testing was administered to 55 adult subjects of any age (>18), educational background, race, disease status present or pre-education about the testing to be performed. This was designed to provide for the broadest test criteria without any bias being imposed, thereby assessing the real-world capability of the average adult, "retail customer" who might wish to purchase a device without the benefit of a physician or other qualified health care provide counsel as is the case with prescription devices. The study was conducted in two (2) parts specifically intended to demonstrate that the device is appropriate for consumer use by lay persons (see data and report in Section 13). The first part addressed self-selection (box labeling), while the second part addressed the content of the user manual including warnings and precautions.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the devices. The Capillus202, Capillus272 Pro, 272 OfficePro, Capillus302, Capillus312, and Capillus352 conform to the standard IEC-60825-2007-03. This IEC standard is a recognized and accepted standard by the FDA. The guidance document for this accepted standard is found in the Federal register, July 26, 2001 (volume 66, Number 144) [page 39049 39050]. This report validates for the Capillus82, Capillus202, Capillus272 Pro, 272 OfficePro, Capillus312, and Capillus352 the laser class of 3R which establishes the AEL (accessible emission limits) as 5 milliWatts maximum. The charger conforms to IEC 61959.

The data presented in this submission is restricted to self-selection and usability to include comprehension of user instructions, warnings, and precautions. The results of a presubmission request (Q160385) verified the objectives that must be fulfilled for the Capillus family of devices to be classified for over-the-counter (OTC) sale. One of the responses for the Q-submission indicated that a minimum of an 80% success rate must be achieved for the intended use of OTC to be granted. The testing was administered to 55 adult subjects of any age (>18), educational background, race, disease status present or pre-education about the testing to be performed. This was designed to provide for the broadest test criteria without any bias being imposed, thereby assessing the real-world capability of the average adult, "retail customer" who might wish to purchase a device without the benefit of a physician or other qualified health care provide counsel as is the case with prescription devices. The study was conducted in two (2) parts specifically intended to demonstrate that the device is appropriate for consumer use by lay persons (see data and report in Section 13). The first part addressed self-selection (box labeling), while the second part addressed the content of the user manual including warnings and precautions. The Capillus usability test demonstrated a pass rate of 94.6% (self-selection) and 90.9% pass rate for user instruction questionnaire, satisfying the FDA's requirements for clearance (sale) as Over-the-Counter Intended Use.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

The Capillus usability test demonstrated a pass rate of 94.6% (self-selection) and 90.9% pass rate for user instruction questionnaire, satisfying the FDA's requirements for clearance (sale) as Over-the-Counter Intended Use.

Predicate Device(s)

103368, 142573, 141567

Reference Device(s)

K143199; K150578; K151516; K153618, K122248; K140931

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

0

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 31, 2017

Capillus, LLC Patricia Schnoor Director, Ouality & Regulatory Affairs 1715 NW 82nd Avenue Miami. Florida 33126

Re: K163170

Trade/Device Name: Capillus 82, Capillus 202, Capillus 272 Pro, 272 Office Pro, Capillus 302, Capillus 312, and Capillus 352 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: November 4, 2016 Received: November 14, 2016

Dear Patricia Schnoor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

1

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Image /page/1/Picture/7 description: The image shows the signature of Jennifer R. Stevenson. The text "Sincerely" is written above the name. The name is written in a larger font size than the word "Sincerely". The letters "FDA" are faintly visible in the background.

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K163170

Device Name

Devillus Laser Domes: Capillus202, Capillus272 Pro, Capillus272 Office Pro, Capillus302, Capillus312, and Capillus352

Indications for Use (Describe)

Indications for Use (Describe)
The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

Capillus, LLC

Date Prepared:November 04, 2016
Submit t er's Cont act Inform at i on:
Name:Patricia Schnoor
Address:Capillus LLC
1715 NW 202nd Avenue
Miami, FL 33126
Establishment
Registration#3010123655
Telephone:
Facsimile:(786) 888-1874
(305) 418-7581
Name of Device and Name / Address of Sponsor:
Trade Name:Capillus82, Capillus202, Capillus272 Pro, 272 OfficePro,
Capillus302, Capillus312, and Capillus352
Common or Usual Name:Lamp, non-heating, for promotion of hair growth
Classification Name:Infrared lamp per 21 CFR 890.5500
Classification Code:OAP (Laser, Comb, Hair)
Sponsor Contact Information:Patricia Schnoor
Capillus LLC
1715 NW 82nd Avenue
Miami, FL 33126
Telephone:(786) 888-1874

Predicate Devices:

Device Trade NameManufacturer510(k)
Hairmax LasercombLexington International103368, 142573
iGrow IIApira141567

Intended Use / Indications for Use:

The Capillus laser domes 82, 202, 272 Pro, 272 OfficePro, 302, 312, and 352, are intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood -Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

4

Technological Characteristics

The Capillus272 Pro and Capillus272 OfficePro both consist of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner, and configured for portable use with rechargeable battery and adapter (Capillus272 Pro) and salon/spa use (Capillus272 OfficePro). The Capillus82, Capillus202, Capillus302, Capillus312, and Capillus352 are exactly the same as the Capillus272 Pro with the exception of the number of diodes which are respective of model (e.g. 82 diodes in Capillus82).

As stated in prior submissions, the devices emit an audible tone at the beginning and end of a therapy session, indicating that therapy has begun (2 short beeps) or ended (one long beep). The portable systems are powered by rechargeable Li-Ion battery cells assembled into a proprietary battery pack. Both the battery are fully compliant to recognized, international standards. The OfficePro is directly powered by the DC charger, which is fully compliant to recognized, international standards.

Performance Data:

Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criteria were met by the devices. The Capillus202, Capillus272 Pro, 272 OfficePro, Capillus302, Capillus312, and Capillus352 conform to the standard IEC-60825-2007-03. This IEC standard is a recognized and accepted standard by the FDA. The guidance document for this accepted standard is found in the Federal register, July 26, 2001 (volume 66, Number 144) [page 39049 39050]. This report validates for the Capillus82, Capillus202, Capillus272 Pro, 272 OfficePro, Capillus312, and Capillus352 the laser class of 3R which establishes the AEL (accessible emission limits) as 5 milliWatts maximum. The charger conforms to IEC 61959.

The data presented in this submission is restricted to self-selection and usability to include comprehension of user instructions, warnings, and precautions. The results of a presubmission request (Q160385) verified the objectives that must be fulfilled for the Capillus family of devices to be classified for over-the-counter (OTC) sale. One of the responses for the Q-submission indicated that a minimum of an 80% success rate must be achieved for the intended use of OTC to be granted. The testing was administered to 55 adult subjects of any age (>18), educational background, race, disease status present or pre-education about the testing to be performed. This was designed to provide for the broadest test criteria without any bias being imposed, thereby assessing the real-world capability of the average adult, "retail customer" who might wish to purchase a device without the benefit of a physician or other qualified health care provide counsel as is the case with prescription devices. The study was conducted in two (2) parts specifically intended to demonstrate that the device is appropriate for consumer use by lay persons (see data and report in Section 13). The first part addressed self-selection (box labeling), while the second part addressed the content of the user manual including warnings and precautions. The Capillus usability test demonstrated a pass rate of 94.6% (self-selection) and 90.9% pass rate for user instruction questionnaire, satisfying the FDA's requirements for clearance (sale) as Over-the-Counter Intended Use.

Just as for Capillus272 Pro (K143199) and reference devices, there are no reported adverse events for this technology.

5

Substantial Equivalence

The requirements set for substantial equivalence in the traditional sense, do not have applicability in this process as the purpose of this submission is expansion of indications for use to include OTC sale capability. However, The Hairmax Lasercomb and iGrow Hair Growth System (predicate devices) have been cleared with the Intended Use of Over-the-Counter, and the Capillus devices can demonstrate the minimum pass-rate for usability by lay persons. The Capillus devices utilize the same technology and share the same risk profile as the predicates. The Capillus family of devices referenced in this application are the same technology used by all LLLT devices cleared under device code OAP. The Capillus272 Pro, Capillus272 OfficePro, Capillus202, Capillus202, Capillus302, Capillus312 and Capillus352 are as safe and effective as the predicate devices, as well as other reference devices in its class.

Capillus DevicesiGrowOAP (General)
K143199; K150578; K151516; K153618K122248; K140931N/A
LLLT Device TypeLLLT Device TypeLLLT Device Type
Prescription UsePrescription UsePrescription and OTC
Intended Use - Androgenic AlopeciaIntended Use - Androgenic AlopeciaIntended Use -Androgenic Alopecia
Contain Laser Diodes- Class 3RContain Laser Diodes- Class 3R and LEDsContain Laser Diodes- Class 3R; LEDs, or a
combination of both
Helmet/Cap DesignHelmet/Cap DesignHelmet/Cap, comb, or flat Design
650nm650nm650nm
Marketing clearance for FemalesMarketing clearance for Males and FemalesMarketing clearance for both genders
Passive Use-Hands FreePassive Use-Hands FreePassive Use-Hands-Free, or comb
272 Laser Diodes21 Laser Diodes; 30 LEDsVarious configurations of lasers, LEDs or both
OAP ClassificationOAP ClassificationOAP Classification
Classification Name -Infrared LampClassification Name -Infrared LampClassification Name - Infrared Lamp
Common Usage Name - Lamp, Non-HeatingCommon Usage Name - Lamp, Non-HeatingCommon Usage Name - Lamp, Non- Heating
General & Plastic Surgery CommitteeGeneral & Plastic Surgery CommitteeGeneral & Plastic Surgery Committee
Fitzpatrick Skin Phototypes - I- IVFitzpatrick Skin Phototypes - I- IVFitzpatrick Skin Phototypes - I -IV
Ludwig-Savin I-II (females)
Norwood Hamilton IIA-V (males)Ludwig-Savin I-II (females)
Norwood Hamilton IIA-V (males)Ludwig-Savin I-II (females); Norwood
Hamilton IIA-V (males); or both
Efficacy Rates - High Compared to PlaceboEfficacy Rates -34-41%Various; trial results are all significant
increase with respect to placebo.
Treatment- 17 weeks, every other day
(indefinite)Treatment- 17 weeks, every other day
(indefinite)Treatment - ~16 weeks, every other day
(indefinite)
Device Class IIDevice Class IIDevice Class II

For these reasons, the Capillus family of laser domes satisfy the FDA's substantial equivalence requirements with respect to risk profile, intended use, and technological and design characteristics. Additionally, no new safety or efficacy concerns are raised due to the minor differences present between devices.