K Number
K171553
Date Cleared
2017-12-08

(192 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Unit for cleaning the spray channels and lubricating the moving parts of dental and surgical handpieces, turbines, air motors and air-driven dental scalers.

Device Description

The W&H Assistina TWIN MB-302 is an electrically and pneumatically driven maintenance device with the following functions:

  • Rinsing the spray channels with cleaning solution (water basis) and drying with compressed air.
  • Lubricating the rotational internal gear parts. The oil is spread and a constant lubrication film is set up.
AI/ML Overview

The provided text describes the W&H Assistina TWIN MB-302, a device for cleaning and lubricating dental and surgical handpieces. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study of acceptance criteria and performance against those criteria.

Therefore, the requested information regarding acceptance criteria, device performance, study details (sample size, data provenance, expert involvement, adjudication, MRMC studies, standalone performance, ground truth), and training set information is not available in the provided text.

The closest information provided related to performance or testing is under the "Performance Testing" section (page 6), which lists various tests performed to demonstrate safety and effectiveness:

  • Electrical Safety Tests according to IEC 61010-1
  • Electromagnetic Compatibility Test according to IEC 61326-1
  • Software validation according to IEC 62304:2006
  • Biological Assessment and Cytotoxicity Testing of the liquids according to EN ISO 10993
  • Risk Analysis according to ISO 14971
  • Testing of cleaning performance according to ISO 15883
  • Validation of wipe disinfection according to the IFU
  • Bench Tests like Lifetime tests (Activation of start button, moving/removing of sliding door, Lifecycle test)

However, specific acceptance criteria values and the reported performance against these criteria are not detailed. The text only states that the results "demonstrate substantial equivalence" and that the biological assessment "meets the requirements," without specifying what those requirements are or the numerical results achieved. It also explicitly states "Clinical Testing: Not applicable," indicating that no clinical studies were performed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 8, 2017

W & H Dentalwerk Buermoos GmbH Mag. Anja Lindner Manager Regulatory Affairs Ignaz-Glaser-StraBe 53 5111 Buermoos AUSTRIA

Re: K171553

Trade/Device Name: W&H Assistina Twin Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EFB Dated: November 7, 2017 Received: November 9, 2017

Dear Mag. Anja Lindner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Mary S. Runner -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171553

Device Name Assistina Twin MB-302

Indications for Use (Describe)

Unit for cleaning the spray channels and lubricating the moving parts of dental and surgical handpieces, turbines, air motors and air-driven dental scalers.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a logo with the letters "W&H" in a stylized font. The logo is enclosed in a gray, rounded, hexagon-like shape. Above the logo, there is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters.

510(k) Summary

510(k) SUMMARY

SubmitterW & H DENTALWERK BÜRMOOS GMBHIgnaz-Glaser-Strasse 53A - 5111 BürmoosAustriaTel.: 0043 -6274 / 6236 -397Fax: 0043 -6274 / 6236 -55
RegistrationNumber9681479
ContactPersonMag. Anja LINDNER
Date ofSubmission7th of November, 2017
Device NameAssistina TWIN MB-302
ClassificationNameDental Handpieces and Accessories
510(k) numberK171553
RegulationNumber21 CFR 872.4200
RegulatoryclassI
Product CodeEFB
PredicateDevicesPrimary Predicate: "Care3 Plus", NSK.; cleared under K081811Reference Predicate: Assistina 301+, W&H Dentalwerk Bürmoos, clearedunder K010127.The predicate has not been subject to a design-related recall.
DeviceDescriptionThe W&H Assistina TWIN MB-302 is an electrically and pneumaticallydriven maintenance device with the following functions:- Rinsing the spray channels with cleaning solution (water basis)and drying with compressed air.- Lubricating the rotational internal gear parts. The oil is spreadand a constant lubrication film is set up.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a partial view of a graphic with the words "PEOPLE HAVE PRIORITY" in a green box at the top. Below this, there is a gray graphic element that appears to be part of a larger design, possibly a logo or a visual representation of a concept. The text "IORITY" is visible at the bottom, suggesting that it is part of the word "PRIORITY" from the phrase above.

510(k) Summary

Technical data:
Assistina TWIN MB-302
Cycle duration10 sec.
Air consumptionApprox. 40 NI/min.
Operating pressure5 – 10 bar (controlled via an integrated, automatic pressure controller)
Supply voltage100 – 240 VAC
Max. power consumption18 VA
Weight3,8 kg
Dimensions297 x 435 x 190 mm
Indications for Use:Unit for cleaning the spray channels and lubricating the moving parts of dental and surgical handpieces, turbines, air motors and air-driven dental scalers.
ItemNew devicePredicate deviceJudgment
NameW&H Assistina TWIN MB-302NSK Care3 Plus---
Technological CharacteristicsPhotoImage: W&H Assistina TWIN MB-302Image: NSK Care3 Plus---
Where usedDental practice, dental clinic.Dental practice, dental clinic.same
ManufacturerW&H Dentalwerk Bürmoos GmbHwww.wh.comNakanishi Inc.www.nsk-inc.com---
510(k)------K081811---

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters "W&H" inside of a white hexagon with rounded corners. The logo is placed on a white background. Above the logo, there is a green bar with the text "PEOPLE HAVE PRIORITY" written in white.

510(k) Summary

Indication forUseUnit for cleaning the spray channelsand lubricating the moving parts ofdental and surgical handpieces,turbines, air motors and air- drivenscalers.Care3 Plus is intended forinternal cleaning e.g. purgingof old lubricant, for themaintenance of rotatingdental and surgicalinstruments.Similar
Scope ofsubmission- Assistina TWIN MB-302device- Adapter HPI- Air supply hose- TWIN Care Set (including- lubricant, detergent andHEPA filter)- Power supply unit- Care3 unit- AC power cord- Maintenance oil- Oil supply funnel- with filter- Mist filter set- Air tube- Oil absorber sheet- O- ring set- Test bur- MagnetSimilar
Dimensions297 x 435 x 190mm (WxDxH)280 x 275 x 360mm(WxDxH)Similar

Technical Da

Technical Data:
ItemNew deviceMB-302Predicate deviceCare3 PlusJudgment
Powersupply100 – 240 V AC ± 10%120 – 230 V AC ± 10%Similar
Powerconsumption18 VA19.5 VASimilar
Air pressure5 – 10 bar3.5 – 6 barSimilar*
CleaningClean the process chambers with a dry cloth and wipe- disinfect the exterior and the process chambers.Not knownSimilar

Process liquids:

LubricantW&H Service Oil F1 MD-302 (cartridge containing200ml lubricant)*NSK maintenance oil(1 liter bottle)Similar
DetergentW&H Activefluid MC-302(cartridge containing 200mldetergent)**not applicable---

{6}------------------------------------------------

510(k) Summary

301+ only without alcohol and is therefore no dangerous good any more.
The technical principle is the same as within the predicate device.The main technological characteristics are the same or, at least, quitesimilar to those of the comparable product.The intended use is identical to the predicate device
Comparison ofthe device tothe predicatedeviceThe target field of application, the intended use, performance parameterand material are the same or, at least, quite similar to those of thepredicate device.The only differences are some technological features: different modes,different number of buttons, different number of chambers.The predicate device has no cleaning function. However, thesedifferences do not have any negative effect on the safety and/oreffectiveness of the subject device.The new device is substantially equivalent to the predicate device.
PerformanceTestingElectrical Safety Tests according to IEC 61010-1, Safety requirements forelectrical equipment for measurement, control and laboratory use – Part1: General requirements for safety
Electromagnetic Compatibility Test according to IEC 61326-1: Electricalequipment for measurement, control and laboratory use, EMCrequirements – Part 1: General requirements for safety
Software validation according to IEC 62304:2006: Medical devicesoftware - Part 1: Guidance on the application of ISO 14972 to medicaldevice software
Biological Assessment and Cytotoxicity Testing of the liquids accordingto EN ISO 10993 was performed. The evaluation meets therequirements.
Risk Analysis according to ISO 14971.
Testing of cleaning performance according to ISO 15883.Validation of wipe disinfection according to the IFU.
Different Bench Tests like Lifetime tests - Activation of start button,moving/removing of sliding door, Lifecycle test according to internalrequirements. The results demonstrate substantial equivalence in thisregards.
ClinicalTesting
ConclusionW&H considers the Assistina Twin MB-302 to be substantially equivalentto the predicate device listed above. This conclusion is based on thesimilarities in intended use, principles of operation, functional design, and

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo with the text "W&H" inside a rounded hexagon shape. Above the logo, there is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo and banner are set against a light gray background.

510(k) Summary

establishment medical use. Differences between the devices shown in the side-by-side comparison table above are minor and do not have any negative effect on equivalence.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.